An Alkalimeter - Analytical Chemistry (ACS Publications)

Publication Date: August 1939. ACS Legacy Archive. Note: In lieu of an abstract, this is the article's first page. Click to increase image size Free f...
1 downloads 0 Views 211KB Size
INDUSTRIAL AND ENGINEERING CHEMISTRY

462

An Improved Thermometer MARSHALL B. STANDING AND JAMES H. WIEGAND University of Michigan, Ann Arbor, Mich.

A

VOL. 11, NO. 8

and inside diameter the same as drilled in the body, is slipped on the thermometer above the stopper to prevent the stopper’s jamming into the body and turning the thermometer as the guard is screwed on. The assembled guard may be screwed on a 3/~-inch nip le which is in turn screwed into a tee, but a much neater method for permanent installations is shown in the figure. A short length of S/s-inch pipe, J , threaded at one end, is brazed or welded into a hole drilled in the main pipe, F , as shown.

COMMON laboratory problem is the installation of thermometers so that they may be easily removed and

An Alkalimeter

at the same time well guarded. In the course of research it was necessary to install ten thermometers on a double-pipe heat interchanger and a guard was developed which is believed t o have considerable merit.

W. HEINLEN HALL Bowling Green State University, Bowling Green, Ohio

The complete guard is shown in Figure 1. The body of the guard, B, is a 1-inch length of ’/*-inch hexagonal brass stock, drilled and tapped for a standard 3ls-inch pipe thread. At the bottom of the tapped hole a concentric hole is drilled through of a diameter a t least inch greater than that of the thermometer, C, being installed. In addition, three holes are drilled for 3/16-inch welding rods, D, which constitute the guards. These rods, of proper length, are threaded or soldered in place and a short piece of 1/3-inch pipe, E , is brazed or soldered inside the rods as shown. For the packing, about 3/10 inch is cut from the large end of a No. 00 one-hole rubber sto per, so that the remainder, G, will not bind in the threads d e n on the thermometer. Last, a washer, H , with g/~e-inchoutside diameter

4

Drying “J” Wafer-Aqm‘rafor Tube (%be Trap

-.

T

HE indirect determination of carbon dioxide by means of a simple alkalimeter merits consideration whenever extreme accuracy is not required and speed is of some importance. Such a n alkalimeter of somewhat original design has been used with considerable success in the author’s laboratory for the past 3 years. In it are incorporated the advantages attending the use of any of the newer high-efficiency solid desiccants (such as anhydrous calcium sulfate), instead of concentrated sulfuric acid, as well as those arising from the elimination of stopcocks,

A FIGURE 1. DETAILSOF GUARD

A is an ordinary 25-cc. distilling flask with a part of the neck cut off. B and C are ordinary test tubes shortened and drawn down. All pieces should be selected for lightness, since it is desirable that the apparatus weigh as little as possible. Ordinary rubber stop ers are used. The weigged sample (about 1.5 grams in the case of limestone) is placed in A . The desiccant is placed in B. Slightly more than enough acid to react completely with the sample is placed in tube C. The entire apparatus is assembled, small plugs being placed in the tubes leading from B and C. The whole piece is then weighed accurately. Drying tubes containing the same desiccant as that used in the apparatus are attached to B and C . To the drying tube attached to B are connected a T-tube, a water trap, and an aspirator as indicated.

ANALYTICAL EDITION

AUGUST 15, 1939

With the aspirator going slowly, the open arm of the T-tube is closed momentarily to start the flow of acid into the reaction chamber, A. It is best that the acid be added slowly with some shaking. When all the acid has been added, the open arm of the tube is closed and a slow stream of air is drawn through to remove all the carbon dioxide. After the apparatus has again reached room temperature, the plugs are attached to B and C and the apparatus is weighed again. The loss in weight represents the carbon dioxide evolved.

463

For more detailed general instructions concerning the manipulation, the reader is referred to such textsas Kolthoff and Sandell (I)*

Literature Cited (1) Kolthoff, I. M., and Sandell, E. B., "Textbook of Quantitative Inorganic Analysis," pp. 360-2, New York, Maomillan Co., 1936.

Crucible Support GILBERT E. SEIL AND H. A. HEILIGMAN, E. J. Lavino and Company, Norristown, Penna.

T

HE crucible support shown in the drawing offers many

advantages to the laboratory in which evaporations, fumings, and ignitions are done in platinum crucibles. When platinum crucibles are heated on a sand bath or supported in a triangle, many determinations are lost because of the spattering of the residue. Moreover, determinations in adjacent crucibles must be discarded because no near-by crucible is free of suspicion of contamination. The creep of salts up the sides and over the top of the crucible during the fuming of a residue is a frequent source of annoyance and error. The salts may bake to the outside of the crucible and become contaminated with the sand of the sand bath. Portions of the baked residue may crack and

It-

h

drop off. A crucible support which eliminates spattering and creep is a valuable adjunct to any laboratory. The equipment described in this article supports the crucible so that accidental overturning is impossible. The support is easily cleaned and kept clean and, unlike a sand bath, can be moved as required to any convenient location. Filter papers can be quickly and thoroughly dried without charring before ignition. The time required for carrying out a drying, an evaporation, or a fuming in a platinum crucible is reduced by one half as compared with the usual methods. The authors found, for example, that in doing several hundred fluorspar analyses, the time required for volatilizing the silicon tetrafluoride, fuming t& residue with sulfuric acid, and then igniting the calcium sulfate was cut in half by the use of this support. More important, the creep of the salts up the sides and over the top of the crucible was eliminated. No samples were lost because of creep. The crucible support described is in reality an air bath which distributes the heat around the sides as well as to the bottom of the crucible. It can be conveniently placed on one corner of a hot plate. The authors obtain excellent results with a support placed on an iron plate over a circular gas burner such as is frequently used in kitchen ranges. The size of the support and of the openings can be varied to suit the needs of the individual laboratory.

a

:

soap

'STONE

I

2

2'

I

2'

2'

PLAN VIEW

SECTIONAL VIEW

2"

2" I

I

The authors use several supports in which twenty crucibles (30-ml. capacity) can be treated simultaneously. The supports are easily made from a piece of soapstone 25 X 30 X 3 cm. (10 X 12 X 1.25 inches) drilled as shown in the drawing. First the 0.6-cm. (0.25-inch) holes are drilled at the center of each circle and at the points indicated on the circumference of each circle. Next the 2.8-cm. (1.125-inch) holes are drilled through the stone, and finally the 3.4-cm. (1.375-inch) holes are drilled 1.25 em. (0.5 inch) from the top of the stone and 0.25 inch from the bottom of the stone. The crucible will rest in the opening, and the heat will go through the holes along the sides of the crucible, thus heating the sides as well as the bottom. The crucible restslow in the opening and cannot be turned over.