Automated in Situ Oxygen Profiling at Aquatic ... - ACS Publications

Aug 11, 2017 - ACS AuthorChoice - This is an open access article published under an ACS AuthorChoice License, which permits ... (“mDOS”) specifica...
1 downloads 0 Views 1MB Size
Subscriber access provided by Columbia University Libraries

Article

Automated in situ oxygen profiling at aquatic-terrestrial interfaces Tanja Brandt, Michael Vieweg, Gerrit Laube, Robert Schima, Tobias Goblirsch, Jan H. Fleckenstein, and Christian Schmidt Environ. Sci. Technol., Just Accepted Manuscript • DOI: 10.1021/acs.est.7b01482 • Publication Date (Web): 11 Aug 2017 Downloaded from http://pubs.acs.org on August 13, 2017

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

Environmental Science & Technology is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 23

Environmental Science & Technology

1

Automated in situ oxygen profiling at aquatic-terrestrial interfaces

2

Tanja Brandt*1, Michael Vieweg1, Gerrit Laube1, Robert Schima2,3, Tobias Goblirsch4, Jan H.

3

Fleckenstein1, Christian Schmidt1

4

1

Department of Hydrogeology, Helmholtz Centre for Environmental Research – UFZ, Leipzig, 04318, Germany

5

2

Department of Monitoring and Exploration Technologies, Helmholtz Centre for Environmental Research – UFZ, Leipzig,

6

04318, Germany

7

3

Chair of Ocean Engineering, University of Rostock, Rostock, 18059, Germany

8

4

Faculty of Economics and Management Science, Chair of Information Management, University of Leipzig, Leipzig, 04109,

9

Germany

10 11

Corresponding author:

12

Tanja Brandt

13

Helmholtz Centre for Environmental Research – UFZ

14

Permoserstr. 15, 04318 Leipzig, Germany

15

Phone +49 341 235 1996

16

Email:[email protected]

17

1

ACS Paragon Plus Environment

Environmental Science & Technology

Page 2 of 23

18

Abstract

19

Optical sensing technologies provide opportunities for in situ oxygen sensing capable of capturing the whole range of spatial

20

and temporal variability. We developed a miniaturized Distributed Oxygen Sensor (‘mDOS’) specifically for long-term in

21

situ application in soil and sediment. The mDOS sensor system enables the unattended, repeated acquisition of time series of

22

in situ oxygen profiles at a sub-cm resolution covering a depth of up to one meter. Compared to existing approaches, this

23

provides the possibility to reveal highly variable and heterogeneous oxygen dynamics at a high, quasi-continuous resolution

24

across both scales. The applicability of the mDOS to capture both, intra- and inter-day, fine-scale variability of spatio-

25

temporal oxygen dynamics under varying hydrological conditions is exemplarily demonstrated. We specifically aim at

26

estimating the dependency between oxygen dynamics and hydrologic conditions along the measured profiles. The mDOS

27

system enables highly detailed insights into oxygen dynamics in various aquatic and terrestrial environments and in the

28

inherent transition zones between them. It thus represents a valuable tool to capture oxygen dynamics to help disentangling

29

the coupling between underlying hydrological and biogeochemical process dynamics.

30

2

ACS Paragon Plus Environment

Page 3 of 23

Environmental Science & Technology

31

1

32

Oxygen (O2) is one of the key parameters governing biogeochemical processes in terrestrial and aquatic environments. This

33

is particularly relevant for aquatic and terrestrial interfaces, where O2 dynamics and interlinked biogeochemical processing

34

exhibit a high degree of temporal and spatial variability. Pronounced transitions between water-saturated and –unsaturated

35

conditions in a sediment matrix can facilitate the occurrence of elevated turnover rates of O2 and nutrients. Examples for

36

such transitions include water table fluctuations in wetlands, riparian zones and riverbanks 1,2 as well as drying and rewetting

37

of soils and sediments3,4. A vital precondition for ultimately disentangling and quantifying O2-associated turnover rates lies

38

in the adequate description of in situ spatio-temporal O2 dynamics and their inherent variability and heterogeneity in soils

39

and sediments.

40

Optical oxygen sensing has become the standard technology particularly for in situ applications with numerous optode-based

41

oxygen sensors being commercially available. Spot optodes feature an oxygen-sensitive coating at the tip of an optical fiber

42

and have been applied extensively in situ and ex situ due to their robustness and long-term stability in challenging and harsh

43

environmental and medical applications5. Instead of combining multiple single-point measurements provided by spot

44

optodes, two-dimensional measurements can be achieved by means of planar optodes and provide the 2D O2 distribution

45

across a given area6,7. Since they utilize a camera-based system for illumination and reading, planar optodes have been

46

primarily applied in

47

illumination of the whole optode area. In situ applications of planar optodes include coastal9,10 and deep sea sediments11

48

using benthic lander systems. It should be noted that the introduction of such a transparent feature may change the soil or

49

sediment structure and thus, may not reflect the actual conditions within the interior7. However, an in situ comparison

50

between planar optodes and microprofilers revealed a good agreement for benthic lander systems11. Microprofilers provide

51

another potential solutions for spatially resolved in situ applications and employ the motorized immersion of a microneedle

52

into the matrix of interest, commonly spanning a few cm in length at very high resolutions in the µm range12–15. Both, planar

53

optodes and microprofiling systems are restricted to fine-grained, unconsolidated sediments and shallow O2 penetration

54

depths (few cm-dm)16,17.

Introduction

studies incorporating transparent vessels like flumes8, mesocosms7 or well plates6 that allow

3

ACS Paragon Plus Environment

Environmental Science & Technology

Page 4 of 23

55

The challenge remains: to combine the robustness of spot optodes in situ with the spatial resolution provided by planar

56

optodes or 1D microprofiling. The most straightforward approach entails the combination of different sensor types, e.g.

57

micro-sensors and oxygen optodes or in situ logging with hand held sensors18,19. Options to add the spatial dimension have

58

been based on either combining multiple single spot optodes, e.g. by spatial distribution along the length of the designated

59

profile1,20 or arranged through multiplexing21. A promising recent approach that has been originally developed for highly

60

conductive, i.e. coarse, sediments, is based on combining two sensing principles in one device: planar optodes and spot fiber

61

optic cables 22. Still, certain shortcomings prevail and render existing methods impractical for long-term in situ application:

62

limited possibilities for unattended monitoring as well as the size of instrumentation potentially disturbing the soil or

63

sediment structure if sensors need to be moved for conducting spatially resolved measurements, e.g. by (repeated) insertion

64

of planar optodes or microprofilers.

65

We have developed a sensor for automated profiling of the oxygen distribution at a sub-centimeter resolution, which we

66

refer to as miniaturized Distributed Oxygen Sensor (“mDOS”). The mDOS system has been designed to automatically

67

capture both, short-term, i.e. intra-day, as well as long term (inter-day) oxygen variability in situ. A novel tubular sensing

68

element and a motorized side-firing polymer optical fiber (POF) form the core element of the system, which is controlled by

69

a microcontroller. This setup enables the fully automated acquisition of oxygen profiles with the capability for conducting

70

unattended time series alongside a substantial size reduction of the sensing element to 5 mm in diameter compared to

71

previous work22. In the following, we focus on presenting our new method including a discussion of its advantages and

72

current limitations. As a proof of concept, we provide two illustrative application examples in the hyporheic zone (HZ) of

73

two different streams: intra-day dynamics in a perennial stream subject to hydropeaking as well as inter-day variability of an

74

intermittent stream during annual flow cessation.

75

4

ACS Paragon Plus Environment

Page 5 of 23

Environmental Science & Technology

76

2

77

2.1

78

The mDOS system consists of five key components: a fiber optic oxygen transmitter (Fibox 3 LCD trace, Presens GmbH,

79

Regensburg, Germany), a tubular sensing element, a side-firing polymer optical fiber (POF), a motor unit and a control unit

80

(Fig. 1a). The latter is based on an Arduino microcontroller board and manages the interactions between the transmitter and

81

the motor unit to acquire time series of oxygen profiles by repeated, automated measurements. The concept of utilizing a

82

side-firing POF has been adapted from a common principle employed in medical imaging, i.e. radial endomicroscopy23.

83

Emission and excitation light is transmitted to and from the transmitter by the side-firing POF inserted into the sensing

84

element. Briefly, single oxygen profiles are obtained by the uniform motion of the POF up- and downwards the tubular

85

sensing element while illuminating single evenly distributed spots of the outer oxygen-sensitive coating of the tubular

86

sensing element (Figure 1b).

Material and methods

Concept of the sensor

87 88

Figure 1 The mDOS oxygen profiling system: Conceptual sketch of the miniaturized Distributed Oxygen Sensor (‘mDOS’) with (a)

89

the side-firing Polymer Optical Fiber (POF) within the tubular sensing element and motor unit (to scale), control unit, fiber optic

5

ACS Paragon Plus Environment

Environmental Science & Technology

Page 6 of 23

90

oxygen transmitter and power supply (not to scale) with (b) detailed view of signal transmission by side-firing POF within the

91

tubular sensing element (not to scale) and (c) photograph of installed motor unit and tubular sensing element.

92

2.1.1

93

The tubular oxygen sensing element is made from PMMA tubes with an outer diameter of 5mm and a wall-thickness of 1

94

mm. Individual tubes of about 70 cm length, yielding a final sensing length of approximately 55 cm, were dip-coated with

95

the sensor cocktail. The sensor cocktail was prepared by dissolving 20 mg of the O2-sensitive dye PtTFPP (Porphyrin

96

Systems GbR, Lübeck, Germany) and 2.0 g of polystyrene (Carl Roth, Karlsruhe, Germany) in 10 ml of toluene

97

(chemsolute, Renningen, Germany). The custom-made dip coating unit consisted of a glass vial with a hole cap and Viton

98

septum (Viton, DuPont) and a motor unit placed above the glass vial (Figure S1). For the coating procedure, the top part of a

99

single tube was placed into the glass vial through a 4.5 mm hole punched into the septum. The diameter of the punched hole

100

was chosen to be slightly smaller than the diameter of the tube in order to seal the vial once a tube is placed inside while still

101

enabling the tube being pulled through. Approximately 1.5 ml of the sensor cocktail were added to the vial and deposited

102

onto the tube during constant upward withdrawal of tubes at a speed of 0.25 cm s-1 to yield homogenous thin sensor films

103

with a thickness of approximately 20 µm. To do so, tubes were connected to the winch of the motor unit by a thin thread.

104

Tubes were dried for approximately 12 hours until complete evaporation of toluene.

105

2.1.2

106

The excitation and emission light between the fiber optic oxygen transmitter and the tubular sensing element is guided by an

107

insulated side-firing POF (2/2.8 mm, PMMA, POFNetz GmbH, Hille, Germany) to divert the guided light 90°. This

108

configuration enables the side-wise illumination of the outer oxygen-sensitive coating of the tubular sensing element from

109

within (Figure 1b). The motorization of the POF enables quasi-continuous profile measurements along the tubular sensing

110

element.

111

The POF is incorporated into a motor unit comprised of a 12V DC gear motor. Here, the POF is tightly placed between two

112

rubber-metal buffers (REIFF GmbH, Reutlingen, Germany), one of which is directly connected to the motor serving as

113

driving wheel. The motor unit itself is mounted directly at the bottom of the main housing and spatially separated from the

Preparation of tubular sensing element

Motor unit and control unit

6

ACS Paragon Plus Environment

Page 7 of 23

Environmental Science & Technology

114

upper cable compartment to avoid entangling of the POF with the motor unit. The moving of the POF up- and downwards

115

the sensing element during measurements requires repeated winding and un-winding, i.e. fiber bending, within the upper

116

cable compartment. The inner dimensions of the housing were set in compliance with the minimal bending radius of the used

117

POF (35 mm) to reduce fiber bending and inherent loss of signal 24. The spatial resolution of individual profiles is defined by

118

the selected sampling frequency of the transmitter and the feed speed of the POF determined by the drive voltage of the

119

motor. All measurements were taken at a sampling frequency of 1 Hz and a feed speed of approximately 5 mm/s at 12 V

120

drive voltage, resulting in a step size equal to the spatial resolution of approx. 5 mm.

121

The control unit controls the excitation and detection of the optical signals by the transmitter and the motion of the motor

122

with regard to speed and direction. To control the positioning of the POF along profiles, i.e. the coated length of the tubular

123

sensing element, the raw optical signals is read from the transmitter through an analogue output. Here, a marked reduction

124

the raw signal amplitude below a certain threshold value marks the upper or lower end of the profile and triggers a reversal

125

of the POF motion between single profiles. Profiles are acquired bi-directionally and considered as independent technical

126

replicates. The whole mDOS system is put to sleep mode with the transmitter switched off between measurement runs to

127

minimize power consumption. A detailed sketch of the wiring as well as the operating script can be found in the supporting

128

information (Figure S3).

129

2.2

130

To acquire the necessary fitting parameters for final oxygen calculations, we determined the response to a) eight different O2

131

saturations between 0 and 100 % air saturation (% airsat) at a fixed temperature of 20 °C and b) over a temperature range of

132

environmentally relevant temperatures ranging from 15 to 35°C at two fixed oxygen concentrations of 0 and 100% airsat.

133

The slope of the latter is used as input parameter for temperature compensation of oxygen values which shows a different

134

extent at 0 and 100 % airsat (eq. 1, Table S1). The fitted slope of the former is used to acquire the necessary fitting

135

parameters 22 as well as the Stern-Vollmer constant (KSV) of the used two-site quenching model25 (eq. 2, Table S1), with KSV

136

indicative of the detection sensitivity S over the full dynamic range of O2 concentrations (0-100 % airsat)5. To do so, a

137

tubular sensing element was placed in a tempered bioreactor (Sixfors, Infors AG, Basel, Switzerland) to define the

Calibration and sensor performance

7

ACS Paragon Plus Environment

Environmental Science & Technology

Page 8 of 23

138

temperature and oxygen saturation accordingly. Prior to field installation, all tubular optodes were calibrated according to the

139

instructions of the transmitter company: a two-point calibration at 0 and 100 % airsat. According to their manual, we placed

140

each sensing element in a non-transparent 1 inch-bottom-sealed tube. For calibration at 100 % airsat, we placed a moist cloth

141

at the bottom of the tube, sealed the top with parafilm to minimize evaporation and let the sensing element equilibrate at

142

humid air for at least 15 min. For calibration at 0% airsat we placed each tube in de-oxygenated water and let it equilibrate

143

which took about 15 min. We found calibrations values for each batch of tubular optodes to be very comparable between

144

single tubes so that one set of calibration values was applied to all tubes of the batch. All calibrations were performed in the

145

dark to minimize interference from ambient light.

146

We evaluated the performance of our new sensor design with regard to response time, detection sensitivity and limit of

147

detection (LOD). The response time t90 was defined as the time for reaching 90 % of the true O2 value upon a change of O2

148

concentration from 100 to 0 % airsat at a constant temperature of 20.0 °C. We calculated the limit of detection (LOD) as

149

three times the standard deviation of the calculated oxygen concentration across a sensing length of 50 cm under anoxic

150

conditions.

151

2.3

152

The installation procedure as of now has been primarily developed for applications with no or only shallow water level of up

153

to approximately 5cm at the time of installation to avoid water intrusion during the assembly of the motor unit and tubular

154

sensing element. As these units of the mDOS system are very closely exposed to potential water intrusion, all connections

155

between single parts of the PVC housing of the motor unit are tightly sealed by screws and additional rubber tubing. All

156

connections between single units as well as cable outlets are also tightly sealed by screw caps. Once completely assembled,

157

this waterproof design of the mDOS system allows measurements in permanently waterlogged environments such as

158

streambeds, lake or marine sediments as well as to withstand temporary waterlogging, e.g. during flooding or rewetting

159

events.

160

First, the tubular sensing element is installed using a method adapted from the “lost cone” technique, 26 which utilizes a steel

161

casing with a removable drive point at the bottom. Both are driven into the sediment with a sledgehammer until the pointed

Experimental setup, installation and measurement procedure

8

ACS Paragon Plus Environment

Page 9 of 23

Environmental Science & Technology

162

tip reaches the desired depth of the profile. The tubular sensing element is carefully inserted into the casing to remove the

163

drive point and held in place while the steel casing is carefully removed. The described procedure can be omitted in soft and

164

fine-grained sediments where the tubular sensing element can be slowly pushed in directly in the sediment. Following initial

165

installation, the tubular sensing element is left for several hours to ensure sufficient equilibration and resettling of the

166

surrounding matrix. For the duration of experiments the tubular sensing element remains permanently installed in situ and

167

thus, in equilibration with the surrounding environment. Second, the motor unit is mounted on top of the tubular sensing

168

element and fixed to a solid steel rod. A few cobbles and gravel were placed underneath the motor unit to support the weight

169

of the motor unit and shield the exposed part of the tubular sensing element from ambient light. The presence of these

170

cobbles and gravels should only induce minor effects on the local flow field in very coarse and gravelly sediments due to the

171

natural heterogeneity of the sediment surface with regard to height and structure. Additionally, the sediment surface

172

underneath the motor unit was not visibly elevated compared to the natural, i.e. undisturbed, sediment surface in proximity

173

of the sensor system location. The installation height of the motor unit above the sediment surface can be adjusted to

174

minimize disturbance of (expected) surface water flow caused by the presence of the motor unit. Third, the other parts of the

175

mDOS system, i.e. control unit, transmitter and 12V battery, are packed into a waterproof box or casing, which is then

176

placed on an elevated location nearby the measurement location. Finally, the profiling is done manually a few times to check

177

and correct the respective threshold values used for automated measurements.

178

To correct O2 measurements for temperature, the mDOS installation was coupled to a depth-resolved temperature monitoring

179

covering the same profiling depth. Here, we use a multilevel temperature sensor (MLTS, Umwelt- and Ingenieurtechnik

180

GmbH, Dresden, Germany)

181

sediment depth and 5 cm above the sediment surface. The MLTS was installed in the same morphological unit and within

182

half a meter distance of the mDOS to minimize horizontal temperature deviation. Previous investigations revealed only

183

minimal horizontal temperature differences up to 0.5 °C for distances of several meters within a similar morphological unit

184

in the same reach 22. We applied a cubic spline interpolation to estimate spatially continuous temperature profiles between

185

the eight temperature sensors distributed along the MLTS. We did not correct any oxygen measurements for varying

186

hydrostatic pressures as pressure-related differences of phase angles are typically negligible under shallow depths of oxygen

27

which features eight circuited temperature sensors placed at 0, 15, 17, 20, 25, 35 and 55 cm

9

ACS Paragon Plus Environment

Environmental Science & Technology

Page 10 of 23

187

penetration1. All calculations for calibration and post-processing of raw data were performed using MATLAB (version

188

2010a, The Mathworks Inc., USA).

189

10

ACS Paragon Plus Environment

Page 11 of 23

Environmental Science & Technology

190

3

191

3.1

192

Based on the two-site quenching model we calculated a Stern-Volmer constant (KSV) = 0.0226 (figure S3a). We found a

193

linear inverse relationship between temperature and measured phase shifts with ΔτKτ0-1 = -0.0703 at 0 % airsat and ΔτKτ100-1 =

194

-0.1526 at 100 % airsat (Figure S3b). Measured calibration values in terms of raw phase angles varied only slightly and the

195

degree of variability was comparable both between individual tubes of one batch as well as between batches (< 1.8 %

196

relative deviation from respective mean values). The 90% response time of our sensor film upon a change in oxygen

197

concentration was t90 = 22±2 s which is faster than approx. 60 s as reported for a similar, tube-based optode system22 and

198

within the range of other PTTFPP-PS-based sensing systems20. The KSV, which defines the detection sensitivity S of the

199

sensor system, is 0.0226 over the full dynamic, i.e. non-linear, calibration range (0-100 % airsat). The observed non-linearity

200

in the relationship between the phase shifts in the absence and presence of O2 (τ0/τ) and oxygen concentration (Figure S3c)

201

results in a varying sensitivity over the calibration range with increasing S towards lower O2 concentrations28. Here,

202

sensitivity varied between Slow = 0.0144 (at 0-10 % airsat) and Shigh = 0.5599 (90-100 % airsat). The extent of variability in

203

the sensitivity could be related to light interference affecting raw phase angles: First, the influence of ambient light due to the

204

lack of a protective coating on the tubular sensing element. Second, the path for excitation and emission light likely induces

205

some scatter when passing through the 1 mm acrylic wall of the tubular sensing element22. Both characteristics have not been

206

investigated in the course of this study but should be considered in future applications. The calculated LOD of 0.25 % airsat

207

lies within the range of other conceptually similar sensor systems 20,29–31. The maximum spatial resolution obtainable without

208

signal smearing, i.e. overlapping illumination points, is theoretically equal to the size of a single illumination point which is

209

estimated to be approx. 3 mm based on the diameter of the POF (2 mm) and the scattering of light through the 1mm wall of

210

the tubular sensing element. However, it should be noted that this value should be considered merely theoretical as smearing

211

of emission light between adjacent illumination points will not only reduce the actual resolution but moreover, affect

212

measured signals and thus, calculated oxygen concentrations .

Results and discussion

Sensor calibration and temperature dependence, response time, sensitivity and limit of detection

20

11

ACS Paragon Plus Environment

Environmental Science & Technology

Page 12 of 23

213

3.2

214

Here, we present two illustrative applications to demonstrate the versatility of the mDOS to be used for intra- and inter-day

215

oxygen profiling under contrasting hydrological conditions.

216

3.2.1

217

Morphological features in streams and rivers are considered hotspots for biogeochemical reactivity, which potentially

218

account for the majority of in-stream metabolism. The spatio-temporal distribution of oxygen is of particular interest: as a

219

potential proxy for aerobic respiration as well as an indicator for aerobic zonation which can define the potential for

220

denitrification, which can only occur when most of the oxygen is consumed 32.

221

We applied the mDOS at an instream gravel bar at the Selke River, a third-order stream located in the Harz Mountains,

222

Germany (N51° 43′33.2″, E11°18′41.1″)32. The stream is subject to sub daily stage variations induced by an upstream water

223

mill inducing stage variations of up to ~ 15 cm within 15 min33. We acquired a series of O2 profile measurements at a 60 min

224

interval for a total duration of 18 hours that covered the period of night-day transition. Each measurement consisted of four

225

single profiles, i.e. technical replicates. The total profile depth was 55 cm and the mDOS system was positioned to cover a

226

sediment layer of 53 cm depth with the remaining 2 cm protruding above the sediment surface. An oxygen optode logger

227

(miniDOT, PME, San Diego, USA) was installed directly in the main channel to measure dissolved oxygen in the surface

228

water at a 5 min interval. We monitored pressure heads at a 5 minute interval in the main and side channel of the stream to

229

estimate short-term variations of head differences and stream stage variability induced by regular hydropeaking. Our aim

230

was to capture the response of the spatio-temporal O2 distribution to variation in stream stage and inherent vertical hydraulic

231

head differences (Δh).

232

We observed an oxygen zonation across the profile with an anoxic zone between 20 and 45 cm depth and oxic areas above

233

and below this anoxic layer (Figure 2b and S4 for representative scatter plots). The extent and shape of this zonation was

234

constant during night time and early morning hours which coincides with a stable water table. Higher O2 concentrations

235

below a depth of 45cm suggest pronounced lateral flowpaths. The anoxic zone is potentially a result of a local colmation

236

layer. However, extensive measurements of hydraulic conductivities at the site have not revealed a general trend of

Proof of concept in situ

Example 1 – short-term O2 dynamics in a gravel bar

12

ACS Paragon Plus Environment

Page 13 of 23

Environmental Science & Technology

237

decreasing hydraulic conductivity with depths34. Local O2 depletion is known to be a result of longer travel times to the

238

measurement points which may either be induced by small-scale heterogeneities of lower hydraulic conductivity and/or

239

longer flowpaths35.

(a)

(b)

240 241

Figure 2 (a) Pressure gradients and (b) oxygen distribution in an in-stream gravel bar of Selke river, Germany, during low flow

242

conditions in summer 2016 including a hydropeak as indicated by a sharp drop followed by partial rise of water level. For oxygen

243

data, the total profile depth was 55 cm at a spatial resolution of 5 mm with 0 cm corresponding to the sediment-air/water interface.

244

Temporal resolution of oxygen input data is 60 minutes and missing data was linearly interpolated at 10 min intervals.

245

O2 saturation in the surface water, monitored by a separate optode logger, varied around 80 % airsat at night and increased

246

gradually up to 89 % airsat during the day (data not shown). With the beginning rise of the water level around 9 AM,

247

indicative of a hydropeak, the anoxic zone was extended towards the stream surface by approximately 5 cm with the oxic-

248

anoxic gradient becoming less sharp and more gradual instead. Simultaneously, oxygen saturation below the anoxic zone

249

decreased from approx. 60 to below 40 % airsat. Considering the constant sediment temperature at this depth (Figure S5), the

250

observed decrease can likely be explained by larger travel times caused by changes in stream water level and the resulting

251

decrease in hydraulic gradients between the main and side channel surrounding the gravel bar as previously reported32,33. The 13

ACS Paragon Plus Environment

Environmental Science & Technology

Page 14 of 23

252

presence of the oxic bottom layer can be attributed to predominantly lateral flowpaths, transporting oxic stream water into

253

the gravel bar from a deeper section of the stream bed, as previously observed for another ISGB in the same stream 22. Here,

254

the formation of such flowpaths is likely facilitated by stream water infiltration from a pool (of a pool-riffle structure)

255

upstream of the ISGB. Several very narrow, less-oxygenated patches within generally oxygenated zones, e.g. at -18 cm,

256

might represent heterogeneities of reduced flow in the sediment, likely caused by the presence of small physical structures

257

like sand or stones. We consider these patches as real data as no irregularities or artefacts during calibration or measurements

258

have been encompassed.

259

The application of the mDOS revealed that O2 in the streambed is not simply a function of stream O2 variation but reveals

260

spatial patterns driven by textural heterogeneities of the streambed. These patterns would not have been detected by single

261

point observations. Besides identifying these zones, we could also observe the temporal dynamics of O2 variability not only

262

during baseflow conditions but also during individual hydrologic events, i.e. hydropeaking. Both of these features and their

263

inherent high degree of variability would not have been observed with only regular point or snapshot measurements.

264

3.2.2

265

Intermittent rivers (IRs) play an important role in aquatic biogeochemical cycling, which is particularly relevant in light of

266

future climate change scenarios indicating intensified periods of drought and more extreme weather events

267

commonly characterized by distinct hydrological phases of drying and rewetting leading to full or partial cessation of surface

268

flow and thus, transitioning between terrestrial and aquatic phases. Recent studies indicate that the short-term transition

269

between phases as well as pulsed rewetting events, e.g. by precipitation, might be particularly relevant as hot moments of

270

biogeochemical activity and carbon processing

271

and potential rewetting events to explore spatio-temporal patterns of oxygen dynamics and the role of the water level as

272

potential primarily control of these patterns.

273

A measuring program was carried out in the streambed of Fuirosos, a first order stream on the Iberian Peninsula

274

(N41°42’16.38’’, E2°34’51.75’’). Oxygen profiles of 55cm length, covering a sediment depth of 50 cm, were measured

275

repeatedly during summer 2015 (June and July) over a period of 36 days at two locations (‘upstream’ and ‘downstream’),

Example 2 – long-term O2 dynamics in an intermittent stream

3,38,39

36–38

. They are

. We acquired streambed O2 profiles over the course of flow cessation

14

ACS Paragon Plus Environment

Page 15 of 23

Environmental Science & Technology

276

separated by a pool-cascade sequence within a 80 m reach40. The interval between sampling days was extended from one day

277

at the beginning to four days towards the end of the measurement period (Figure 3b). We conducted two to three

278

measurements per sampling day during daylight between 10 a.m. and 7 p.m., thus minimizing effects of diel biogeochemical

279

fluctuations39 due to the lack of automated measurements. Each measurement consisted of at least six single profiles, i.e.

280

technical replicates, with an estimated spatial resolution of 6 mm. The data was obtained using a “first-generation” prototype

281

of the mDOS without a control unit, thus only allowing manual measurements by means of a simple switch

282

(“up/down/stop”) to control the POF movement up and down the tubular probe. Manual readings of the water level were

283

carried out with an electric contact gauge in piezometer wells near each mDOS location. The annual drought period of

284

Fuirosos stream in summer 2015 was characterized by stable dry conditions during the first three weeks at which the onset of

285

surface flow cessation coincided with day one of measurements (Fig. 3c). During drought, hydrologic conditions switch

286

from surface to subsurface flow corresponding to water-saturated to water-unsaturated sediment conditions. Several

287

precipitation events with increasing intensities up to 25 mm/day occurred towards the end of the measurement period.

15

ACS Paragon Plus Environment

Environmental Science & Technology

Page 16 of 23

(a)

(b)

(c)

288 289 290 291 292

Figure 3 Subsurface oxygen distribution during cessation of surface flow (transition from saturated to unsaturated conditions) in the streambed of Fuirosos stream (a) up- and (b) downstream of a pool-cascade sequence and (c) corresponding hydrologic conditions at both locations and local preicpitation (La Battloria, Spain). For oxygen data, the total profile depth was 55 cm at a spatial resolution of 5 mm with 0 cm corresponding to the sediment-air/water interface.

293

Both locations showed a sharp, persistent oxygen transition zone from oxic to anoxic conditions of one or two cm. Absolute

294

depths of water level cannot be linked to respective oxygen profiles due to relative height differences of the sediment surface

295

between oxygen and corresponding water level measurements. With the onset of precipitation events from day 28 onwards

296

(Fig. 3c), the measurement locations showed deviating patterns for the spatio-temporal oxygen distribution. The ‘upstream’

297

location continued with the general trend of a decreasing anoxic zone while ‘downstream’, the oxic-anoxic transition zone

298

shifted upwards remarkably by approx. 30 cm. In both locations, the transition from anoxic bottom to oxic top layers became 16

ACS Paragon Plus Environment

Page 17 of 23

Environmental Science & Technology

299

more gradual and less sharp. This pattern might have been caused by a stimulation of microbial respiration in previously dry

300

sediments that became rewetted, thus supplying water and solutes to microorganisms residing in drying sediments 41–43. Our

301

results indeed indicate the water level as primary control of oxic-anoxic zonation, whose temporal and spatial dynamics

302

could be precisely captured by the mDOS system. This water level dependency was pronounced at both locations, as

303

reflected in a clear linear relationship (R2(upstream) = 0.9709 and R2(downstream) = 0.9578) between depth of water level and

304

corresponding oxygen penetration throughout the whole drying phase (Figure S5). Interestingly, the slope deviates between

305

both locations: ‘downstream’ it is characterized by a slope of ~1, corresponding to equal rates of descending water levels and

306

oxygen penetration. In contrast, the ‘upstream’ location exhibits a slope of ~0.7, i.e. anoxic conditions prevail in the

307

unsaturated zone. This might likely be attributed to generally different hydrologic conditions between locations. In contrast

308

to the ‘downstream’ location, which was characterized by upwelling groundwater as reflected in the profound rise of water

309

levels upon rewetting events (Figure 3c), the upstream location exhibited downwelling conditions40. We found that

310

constantly falling water levels during drying of intermittent streambeds significantly influenced the subsurface spatial O2

311

distribution as revealed by long-term time series. Moreover, these time series also provided novel information on sudden

312

small-scale variability within the oxic-anoxic zonation upon rewetting events, accentuating the need for highly resolved

313

spatio-temporal oxygen profiling as provided by the mDOS system.

314

3.3

315

Overall, the mDOS has proven its suitability for automated acquisition of O2 profiles in situ to capture both, intra- and inter-

316

day variability. It allows a completely unattended acquisition of oxygen profiles and the profiling depth as well as duration

317

and frequency of measurements can be easily tailored to the specific demands of other applications. The use of a side-firing

318

fiber led to a substantial size reduction of the oxygen sensing element to 5 mm from previously 30 mm 22 which substantially

319

simplifies the installation process even in very coarse and gravelly sediments. Here, the disturbance of the natural flow field

320

by the presence of the tubular sensing element can be considered rather small since the sediment scale with regard to

321

sediment size is typically larger than the corresponding “sensing scale” referring to the diameter of the sensing element. The

322

current setup allows the motor unit to be installed sufficiently high to avoid obstruction of surface flow by the motor unit.

Applicability and future potential

17

ACS Paragon Plus Environment

Environmental Science & Technology

Page 18 of 23

323

The current need of a solid steel rod for fixation of the motor unit could be overcome by placing the motor unit on a tripod as

324

commonly employed for marine sediment profiling.

325

We also want to address current limitations and potential sources of error and uncertainty. First, the tubular sensing element

326

does not incorporate an appropriate protective layer which is troublesome for two reasons: photodegradation (bleaching) of

327

the probe itself and physical disturbance of the measurement by ambient light. Both may cause overall decreasing phase

328

angles leading to overestimated oxygen concentrations22. Among potential solutions, a thin top layer of black silicone seems

329

to be most promising5 but the integration into the manufacturing process of the tubular sensing element requires further

330

development. Previous laboratory tests revealed practical challenges, mainly to adjust the chemical composition of the

331

coating solution suitable for incorporation into the dip-coating process to yield a protective layer that is homogeneous, of

332

desired thickness and adhesive to the underlying sensing film – all while not re-dissolve the previously applied sensing layer.

333

Adhesion of a silicone-based sensing layer could be facilitated by emulsifying the polystyrene-based sensing layer with

334

silicone44. Second, optimizing the quality of the oxygen-sensitive coating of the tubular sensing element in terms of film

335

thickness and homogeneity will reduce spatial sensing variability during calibration and measurements and thus further

336

improve the overall robustness of the whole mDOS system in terms of sensing quality. Third, the current configuration of

337

the mDOS system requires regular maintenance approximately every other day for downloading of acquired data and

338

changing of batteries, depending on the chosen measurement interval. To overcome the latter, future designs will incorporate

339

an OEM board instead of the transmitter, thus allowing the integration of all components into a single architectural

340

framework. The implementation of a data model45 and an overarching cloud-based web service enables remote data access,

341

visualization and adjustments of the measurement routine close to real time. This is particularly powerful in light of recent

342

attempts to implement adaptive monitoring46 and automatic high frequency monitoring in lakes and rivers 47,48.

343

Also, we briefly want to discuss the potential of the mDOS concept for expansion to other analytes as well as the

344

combination of the system with other analytical methods. The mDOS has been designed as modular system to be potentially

345

adapted for other luminescent sensing parameters, e.g. CO2, pH or temperature (T). Additionally, it may even be expanded

346

towards the simultaneous determination of two or more analytes as reported for other optode-based systems5. Of potential

347

combinations, a dual O2/CO2 or O2/T mDOS system should be in the focus of future technical development as these are

18

ACS Paragon Plus Environment

Page 19 of 23

Environmental Science & Technology

348

particularly relevant in many biological systems because of the strong coupling between oxygen and CO2 consumption and

349

temperature effects. The principle of sandwich sensors to utilize a DGT gel as diffusive layer for the overlying optode

350

sensor49,50 could be theoretically combined with the mDOS system for simultaneous mapping of O2 and chemical species

351

dynamics, particularly trace metals and nutrients. This approach, with O2 serving as potential proxy for redox zonation, could

352

potentially expand our knowledge on the formation of biogeochemical hotspots and hot moments in soils 6,51 and streambeds

353

50,52

354

measurements of the main factors underlying oxygen dynamics in these environments, namely hydrology and diffusion, at

355

the same scale. Hydrologic conditions, e.g. in terms of travel times, can be estimated from variability of naturally occurring

356

tracers like electric conductivity33 or the Rn-222 isotope54 and coupled to oxygen concentration data to estimate aerobic

357

respiration rates.

358

4

359

We thank Peter Portius and Manuel Kositzke for the mechanical construction of the mDOS system and Astrid Harjung for

360

her valuable input and proofreading. The research was funded by the European Union's Seventh Framework for research,

361

technological development and demonstration under grant agreement no. 607150, within the Marie Curie Initial Training

362

Network “INTERFACES: Ecohydrological interfaces as critical hotspots for transformations of ecosystem exchange fluxes

363

and biogeochemical cycling".

, specifically with regard to depth-dependent high reactivity zones53. However, this requires adequate in situ

Acknowledgements

364 365

Supporting information. Details about the dip coating procedure, the control unit, equations and fitting parameters for

366

oxygen calculations, additional plots of oxygen and temperature data.

367 368

19

ACS Paragon Plus Environment

Environmental Science & Technology 369

5

370 371 372 373 374 375 376 377 378 379 380 381 382 383 384 385 386 387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408 409 410 411 412 413 414 415

(1) (2) (3)

(4) (5) (6) (7)

(8) (9) (10) (11)

(12)

(13) (14)

(15) (16) (17)

(18) (19)

Page 20 of 23

References

Rickelt, L. F.; Askaer, L.; Walpersdorf, E.; Elberling, B.; Glud, R. N.; Kühl, M. An Optode Sensor Array for LongTerm In Situ Oxygen Measurements in Soil and Sediment. J. Environ. Qual. 2013, 42 (4), 1267. Vidon, P.; Marchese, S.; Welsh, M.; McMillan, S. Short-term spatial and temporal variability in greenhouse gas fluxes in riparian zones. Environ. Monit. Assess. 2015, 187 (8). Gómez-Gener, L.; Obrador, B.; von Schiller, D.; Marcé, R.; Casas-Ruiz, J. P.; Proia, L.; Acuña, V.; Catalán, N.; Muñoz, I.; Koschorreck, M. Hot spots for carbon emissions from Mediterranean fluvial networks during summer drought. Biogeochemistry 2015, 125 (3), 409–426. Kim, D.-G.; Vargas, R.; Bond-Lamberty, B.; Turetsky, M. R. Effects of soil rewetting and thawing on soil gas fluxes: a review of current literature and suggestions for future research. Biogeosciences 2012, 9 (7), 2459–2483. Wang, X.; Wolfbeis, O. S. Optical methods for sensing and imaging oxygen: materials, spectroscopies and applications. Chem. Soc. Rev. 2014, 43 (10), 3666–3761. Rubol, S.; Dutta, T.; Rocchini, D. 2D visualization captures the local heterogeneity of oxidative metabolism across soils from diverse land-use. Sci. Total Environ. 2016, 572, 713–723. Askaer, L.; Elberling, B.; Glud, R. N.; Kühl, M.; Lauritsen, F. R.; Joensen, H. P. Soil heterogeneity effects on O2 distribution and CH4 emissions from wetlands: In situ and mesocosm studies with planar O2 optodes and membrane inlet mass spectrometry. Soil Biol. Biochem. 2010, 42 (12), 2254–2265. Precht, E.; Franke, U.; Polerecky, L.; Huettel, M. Oxygen dynamics in permeable sediments with wave-driven pore water exchange. Limnol. Oceanogr. 2004, 49 (3), 693–705. Glud, R. N.; Tengberg, A.; Kühl, M.; Hall, P. O.; Klimant, I. An in situ instrument for planar O2 optode measurements at benthic interfaces. Limnol. Oceanogr. 2001, 46 (8), 2073–2080. Wenzhöfer, F.; Glud, R. N. Small-scale spatial and temporal variability in coastal benthic O2 dynamics: Effects of fauna activity. Limnol. Oceanogr. 2004, 49 (5), 1471–1481. Glud, R. N.; Wenzhöfer, F.; Tengberg, A.; Middelboe, M.; Oguri, K.; Kitazato, H. Distribution of oxygen in surface sediments from central Sagami Bay, Japan: In situ measurements by microelectrodes and planar optodes. Deep Sea Res. Part Oceanogr. Res. Pap. 2005, 52 (10), 1974–1987. De Falco, N.; Boano, F.; Arnon, S. Biodegradation of labile dissolved organic carbon under losing and gaining streamflow conditions simulated in a laboratory flume: DOC uptake in losing and gaining streams. Limnol. Oceanogr. 2016, 61 (5), 1839–1852. Fabricius, A.-L.; Duester, L.; Ecker, D.; Ternes, T. A. New Microprofiling and Micro Sampling System for Water Saturated Environmental Boundary Layers. Environ. Sci. Technol. 2014, 48 (14), 8053–8061. GLUD, R.; GUNDERSEN, J.; JORGENSEN, B.; REVSBECH, N.; SCHULZ, H. DIFFUSIVE AND TOTAL OXYGEN-UPTAKE OF DEEP-SEA SEDIMENTS IN THE EASTERN SOUTH-ATLANTIC OCEAN - IN-SITU AND LABORATORY MEASUREMENTS. DEEP-SEA Res. PART -Oceanogr. Res. Pap. 1994, 41 (11–12), 1767– 1788. Wenzhofer, F.; Glud, R. Benthic carbon mineralization in the Atlantic: a synthesis based on in situ data from the last decade. DEEP-SEA Res. PART -Oceanogr. Res. Pap. 2002, 49 (7), 1255–1279. Wenzhöfer, F.; Holby, O.; Kohls, O. Deep penetrating benthic oxygen profiles measured in situ by oxygen optodes. Deep Sea Res. Part Oceanogr. Res. Pap. 2001, 48 (7), 1741–1755. de Beer, D.; Wenzhofer, F.; Ferdelman, T.; Boehme, S.; Huettel, M.; van Beusekom, J.; Bottcher, M.; Musat, N.; Dubilier, N. Transport and mineralization rates in North Sea sandy intertidal sediments, Sylt-Romo Basin, Wadden Sea. Limnol. Oceanogr. 2005, 50 (1), 113–127. Malcolm, I. A.; Soulsby, C.; Youngson, A. F. High-frequency logging technologies reveal state-dependent hyporheic process dynamics: implications for hydroecological studies. Hydrol. Process. 2006, 20 (3), 615–622. Quinlan, E.; Malcolm, I. A.; Gibbins, C. N. Spatio-temporal variability of dissolved oxygen within the shallow subsurface zone of a freshwater pearl mussel bed. Fundam. Appl. Limnol. Arch. Für Hydrobiol. 2014, 185 (3), 281– 294.

20

ACS Paragon Plus Environment

Page 21 of 23 416 417 418 419 420 421 422 423 424 425 426 427 428 429 430 431 432 433 434 435 436 437 438 439 440 441 442 443 444 445 446 447 448 449 450 451 452 453 454 455 456 457 458 459 460 461 462 463 464

(20) (21)

(22) (23) (24) (25) (26) (27) (28) (29) (30) (31)

(32) (33)

(34) (35) (36) (37)

(38) (39)

(40) (41)

Environmental Science & Technology Fischer, J. P.; Koop-Jakobsen, K. The multi fiber optode (MuFO): A novel system for simultaneous analysis of multiple fiber optic oxygen sensors. Sens. Actuators B Chem. 2012, 168, 354–359. Chaturvedi, P.; Hauser, B. A.; Foster, J. S.; Karplus, E.; Levine, L. H.; Coutts, J. L.; Richards, J. T.; Vanegas, D. C.; McLamore, E. S. A multiplexing fiber optic microsensor system for monitoring spatially resolved oxygen patterns. Sens. Actuators B Chem. 2014, 196, 71–79. Vieweg, M.; Trauth, N.; Fleckenstein, J. H.; Schmidt, C. Robust Optode-Based Method for Measuring in Situ Oxygen Profiles in Gravelly Streambeds. Environ. Sci. Technol. 2013, 47 (17), 9858–9865. Sun, C. W.; Lee, S. Y.; Lin, K. F. Review: Optical Scanning Probe for Optical Coherence Tomography. J. Med. Biol. Eng. 2014, 34 (2), 95–100. Argyros, A.; Lwin, R.; Large, M. C. Bend loss in highly multimode fibres. Opt. Express 2008, 16 (23), 18590– 18598. Carraway, E. R.; Demas, J. N.; DeGraff, B. A.; Bacon, J. R. Photophysics and photochemistry of oxygen sensors based on luminescent transition-metal complexes. Anal. Chem. 1991, 63 (4), 337–342. Baxter, C.; Hauer, F. R.; Woessner, W. W. Measuring Groundwater–Stream Water Exchange: New Techniques for Installing Minipiezometers and Estimating Hydraulic Conductivity. Trans. Am. Fish. Soc. 2003, 132 (3), 493–502. Schmidt, C.; Buttner, O.; Musolff, A.; Fleckenstein, J. H. A method for automated, daily, temperature-based vertical streambed water-fluxes. Fundam. Appl. Limnol. 2014, 184 (3), 173–181. Santner, J.; Larsen, M.; Kreuzeder, A.; Glud, R. N. Two decades of chemical imaging of solutes in sediments and soils – a review. Anal. Chim. Acta 2015, 878, 9–42. Perez de Vargas-Sansalvador, I. M.; Fay, C.; Fernandez-Ramos, M. D.; Diamond, D.; Benito-Lopez, F.; CapitanVallvey, L. F. LED–LED portable oxygen gas sensor. Anal. Bioanal. Chem. 2012, 404 (10), 2851–2858. Rigo, M. V.; Geissinger, P. Crossed Optical Fiber Sensor Arrays for High-Spatial-Resolution Sensing: Application to Dissolved Oxygen Concentration Measurements. J. Sens. 2012, 2012, 1–10. Amao, Y.; Miyashita, T.; Okura, I. Platinum tetrakis(pentafluorophenyl)porphyrin immobilized in polytrifluoroethylmethacrylate film as a photostable optical oxygen detection material. J. Fluor. Chem. 2001, 107 (1), 101–106. Trauth, N.; Schmidt, C.; Vieweg, M.; Oswald, S. E.; Fleckenstein, J. H. Hydraulic controls of in-stream gravel bar hyporheic exchange and reactions. Water Resour. Res. 2015, 51 (4), 2243–2263. Vieweg, M.; Kurz, M. J.; Trauth, N.; Fleckenstein, J. H.; Musolff, A.; Schmidt, C. Estimating time-variable aerobic respiration in the streambed by combining electrical conductivity and dissolved oxygen time series: Variable Respiration in the Streambed. J. Geophys. Res. Biogeosciences 2016, 121 (8), 2199–2215. Munz, M.; Oswald, S. E.; Schmidt, C. Analysis of riverbed temperatures to determine the geometry of subsurface water flow around in-stream geomorphological structures. J. Hydrol. 2016, 539, 74–87. Trauth, N.; Fleckenstein, J. H. Single discharge events increase reactive efficiency of the hyporheic zone: DISCHARGE EVENTS INCREASE REACTIVITY. Water Resour. Res. 2017, 53 (1), 779–798. Datry, T.; Larned, S. T.; Tockner, K. Intermittent Rivers: A Challenge for Freshwater Ecology. BioScience 2014, 64 (3), 229–235. Skoulikidis, N. T.; Sabater, S.; Datry, T.; Morais, M. M.; Buffagni, A.; Dörflinger, G.; Zogaris, S.; del Mar SánchezMontoya, M.; Bonada, N.; Kalogianni, E.; et al. Non-perennial Mediterranean rivers in Europe: Status, pressures, and challenges for research and management. Sci. Total Environ. 2017, 577, 1–18. Leigh, C.; Boulton, A. J.; Courtwright, J. L.; Fritz, K.; May, C. L.; Walker, R. H.; Datry, T. Ecological research and management of intermittent rivers: an historical review and future directions. Freshw. Biol. 2016, 61 (8), 1181–1199. von Schiller, D.; Graeber, D.; Ribot, M.; Timoner, X.; Acuña, V.; Martí, E.; Sabater, S.; Tockner, K. Hydrological transitions drive dissolved organic matter quantity and composition in a temporary Mediterranean stream. Biogeochemistry 2015, 123 (3), 429–446. Harjung, A.; Sabater, F.; Butturini, A. Hydrological connectivity drives dissolved organic matter processing in an intermittent stream. Limnol. - Ecol. Manag. Inland Waters 2017. Manzoni, S.; Schimel, J. P.; Porporato, A. Responses of soil microbial communities to water stress: results from a meta-analysis. Ecology 2012, 93 (4), 930–938.

21

ACS Paragon Plus Environment

Environmental Science & Technology 465 466 467 468 469 470 471 472 473 474 475 476 477 478 479 480 481 482 483 484 485 486 487 488 489 490 491 492 493 494 495 496 497 498 499

(42) (43) (44) (45)

(46)

(47)

(48)

(49)

(50) (51) (52)

(53)

(54)

Page 22 of 23

Moyano, F. E.; Manzoni, S.; Chenu, C. Responses of soil heterotrophic respiration to moisture availability: An exploration of processes and models. Soil Biol. Biochem. 2013, 59, 72–85. Febria, C. M.; Hosen, J. D.; Crump, B. C.; Palmer, M. A.; Williams, D. D. Microbial responses to changes in flow status in temporary headwater streams: a cross-system comparison. Front. Microbiol. 2015, 6. Trettnak, W.; Kolle, C.; Reininger, F.; Dolezal, C.; O’Leary, P. Miniaturized luminescence lifetime-based oxygen sensor instrumentation utilizing a phase modulation technique. Sens. Actuators B Chem. 1996, 36 (1–3), 506–512. Schima, R.; Goblirsch, T.; Salbach, C.; Franczyk, B.; Aleithe, M.; Bumberger, J.; Dietrich, P. Research in Progress: Implementation of an Integrated Data Model for an Improved Monitoring of Environmental Processes. In Business Information Systems Workshops: BIS 2016 International Workshops, Leipzig, Germany, July 6-8, 2016, Revised Papers; Abramowicz, W., Alt, R., Franczyk, B., Eds.; Springer International Publishing: Cham, 2017; pp 332–339. Blaen, P. J.; Khamis, K.; Lloyd, C. E. M.; Bradley, C.; Hannah, D.; Krause, S. Real-time monitoring of nutrients and dissolved organic matter in rivers: Capturing event dynamics, technological opportunities and future directions. Sci. Total Environ. 2016, 569–570, 647–660. Rode, M.; Wade, A. J.; Cohen, M. J.; Hensley, R. T.; Bowes, M. J.; Kirchner, J. W.; Arhonditsis, G. B.; Jordan, P.; Kronvang, B.; Halliday, S. J.; et al. Sensors in the Stream: The High-Frequency Wave of the Present. Environ. Sci. Technol. 2016, 50 (19), 10297–10307. Marcé, R.; George, G.; Buscarinu, P.; Deidda, M.; Dunalska, J.; de Eyto, E.; Flaim, G.; Grossart, H.-P.; Istvanovics, V.; Lenhardt, M.; et al. Automatic High Frequency Monitoring for Improved Lake and Reservoir Management. Environ. Sci. Technol. 2016, 50 (20), 10780–10794. Stahl, H.; Warnken, K. W.; Sochaczewski, L.; Glud, R. N.; Davison, W.; Zhang, H. A combined sensor for simultaneous high resolution 2-D imaging of oxygen and trace metals fluxes: Sandwich sensor manuscript. Limnol. Oceanogr. Methods 2012, 10 (5), 389–401. Han, C.; Ren, J.; Wang, Z.; Tang, H.; Xu, D. A novel hybrid sensor for combined imaging of dissolved oxygen and labile phosphorus flux in sediment and water. Water Res. 2017, 108, 179–188. Kuzyakov, Y.; Blagodatskaya, E. Microbial hotspots and hot moments in soil: Concept & review. Soil Biol. Biochem. 2015, 83, 184–199. Ullah, S.; Zhang, H.; Heathwaite, A. L.; Binley, A.; Lansdown, K.; Heppell, K.; Trimmer, M. In situ measurement of redox sensitive solutes at high spatial resolution in a riverbed using Diffusive Equilibrium in Thin Films (DET). Ecol. Eng. 2012, 49, 18–26. Knapp, J. L. A.; González-Pinzón, R.; Drummond, J. D.; Larsen, L. G.; Cirpka, O. A.; Harvey, J. W. Tracer-based characterization of hyporheic exchange and benthic biolayers in streams. Water Resour. Res. 2017, 53 (2), 1575– 1594. Pittroff, M.; Frei, S.; Gilfedder, B. S. Quantifying nitrate and oxygen reduction rates in the hyporheic zone using 222Rn to upscale biogeochemical turnover in rivers. Water Resour. Res. 2017, 53 (1), 563–579.

500

22

ACS Paragon Plus Environment

Page 23 of 23

Environmental Science & Technology

501 502

For Table of Contents Only

23

ACS Paragon Plus Environment