Chemical Characterization of Low Molecular Weight (MW < 580 Da) Soluble Microbial Products (SMPs) in an Anaerobic Membrane Bioreactor Chinagarn Kunacheva, Chencheng Le, Yan Ni Annie Soh, and David C Stuckey Environ. Sci. Technol., Just Accepted Manuscript • DOI: 10.1021/acs.est.6b05791 • Publication Date (Web): 26 Jan 2017 Downloaded from http://pubs.acs.org on January 30, 2017

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

Environmental Science & Technology is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 24

Environmental Science & Technology

1

Chemical Characterization of Low Molecular Weight (MW < 580 Da)

2

Soluble Microbial Products (SMPs) in an Anaerobic Membrane Bioreactor

3 4

Chinagarn Kunachevaa, Chencheng Lea,b, Yan Ni Annie Soha and David C. Stuckeya,c*

5 6

a

7

Institute, Nanyang Technological University, 1 Cleantech Loop, CleanTech One, Singapore

8

637141, Singapore.

9

b

Advanced Environmental Biotechnology Center, Nanyang Environment & Water Research

Division of Environmental and Water Resources Engineering, School of Civil and

10

Environmental Engineering, Nanyang Technological University, 50 Nanyang Avenue,

11

Singapore 639798, Singapore.

12

c

Department of Chemical Engineering, Imperial College London, SW7 2AZ, UK

13 14

*Corresponding author

15

E-mail: [email protected], Tel: +44 207 5945591

16 17

Abstract

18

Effluents from wastewater treatment systems contain a variety of organic compounds,

19

including end products from the degradation of influent substrates, non-biodegradable feed

20

compounds, and soluble microbial products (SMPs) produced by microbial metabolism. It is

21

important to identify the major components of these SMPs in order to understand what is in

22

wastewater effluents. In this study, physical pretreatments to extract and concentrate low

23

molecular weight SMPs (MW< 580 Da) from effluents were optimised. Liquid-liquid

24

extraction (LLE) of a 200 mL effluent sample showed the best performance using a mixture

25

of n-hexane, chloroform and dichloromethane (70 mL) for extraction. For solid phase

1 Environment ACS Paragon Plus

Environmental Science & Technology

26

extraction (SPE), two OasisHLB cartridges were connected in-line to optimise recovery, and

27

the eluted samples from each cartridge were analysed separately to avoid overlapping peaks.

28

Four solvents varying from polar through to non-polar (methanol, acetone, dichloromethane,

29

n-hexane) were selected to maximise the number of compound peaks eluted. A combination

30

of SPE (OasisHLB) followed by LLE was shown to maximise compound identification and

31

quantification. However, the compounds identified only accounted for 2.1 mg COD/L (16%

32

of total SMP as COD) because many SMPs have considerably higher MWs. Finally, the

33

method was validated by analysing a variety of different reactor effluents and feeds.

34 35

Graphical Abstract Effluent

Analysis GC-MS m/z = 30-580

5μm

Size

1μm 100k

SMP and ECP

Solid Phase Extrac on

Liquid-Liquid Extrac on

10k OH

MW < 580 O

OH

Elu on MeoH Acetone DCM Hexane

Extrac on Hexane + DCM + Chloroform

36 37 38

Key words: anaerobic digestion, membranes, characterization, dissolved organic matter,

39

soluble microbial products (SMP), GC-MS

40 41

1. Introduction

42

Biological wastewater treatment (both aerobic and anaerobic) has been the process of choice

43

for many years, although these systems are never 100% efficient, and hence significant

44

amounts of organic soluble microbial products (SMPs) are present in the effluent defining the

45

limits to treatment. These effluents contain a variety of complex organic compounds,

46

including transformation products from the (partial) degradation of influent substrates, non-

2 Environment ACS Paragon Plus

Page 2 of 24

Page 3 of 24

Environmental Science & Technology

47

biodegradable compounds in the feed, and SMPs including extracellular polymer (ECPs),

48

which are produced by microorganisms during metabolism and cell lysis 1. Many effluent

49

SMPs are degradable over time in both aerobic and anaerobic processes, although the

50

hydraulic retention time in these processes often limits complete degradation 2. Hence

51

minimizing SMP production can increase treatment performance

52

by-products 4, and reduce fouling in membrane bioreactors 5. However, to achieve these goals

53

it is important to characterise the chemical composition of SMPs in order to begin to

54

understand the metabolic pathways that create these compounds, why they do not degrade,

55

and how to reduce them in the effluent.

1, 3

, minimise chlorination

56 57

There are a limited number of papers in the literature analysing SMP composition in

58

biological systems, and only a few researchers have used advanced instruments such as gas

59

chromatography-mass spectrometry (GC-MS)

60

trap time-of-flight mass spectrometry (LC-IT-TOF-MS)

61

desorption ionization time-of-flight mass spectrometry (MALDI-TOF-MS)

62

surprisingly found long chain alkanes, alkenes, and aromatic carbons such as bis(2-ethylhexyl

63

phthalates in significant concentrations in a reactor fed with only synthetic wastewater (sugar,

64

nutrients, and minerals). Similar compounds were also found in the effluent of a submerged

65

anaerobic membrane bioreactor (SAMBR) treating municipal solid waste leachate

66

additionally, esters, phenols and amides were also found in the effluent in significant

67

concentrations. Another study in a full-scale upflow anaerobic sludge blanket (UASB)

68

reactor revealed long chain carbohydrates, and esters 8, and these compounds were not only

69

found in an anaerobic processes, but also in an aerobic sequencing batch reactor (SBR).

70

Currently SMPs are analysed using a wide range of different pretreatments and analytical

71

methodologies, and hence the data produced is often difficult to compare, and the maximum

6-8

, high-resolution liquid chromatography ion 9

3 Environment ACS Paragon Plus

, and matrix-assisted laser 7, 9

. Aquino

7

6

;

Environmental Science & Technology

72

number of compounds identified is only 27 7. A detailed analytical methodology for SMPs in

73

wastewater needs to be developed so that the data obtained by different researchers can be

74

compared. Hence, the objective of this study was to optimize pretreatment methods (liquid-

75

liquid extraction (LL) and solid phase extraction (SPE)) to maximise the extraction of solutes

76

from the effluent sample, and to develop a methodology for quantification of low molecular

77

weight compounds (MW < 580) in SMPs. This MW limit is set by GC-MS since the

78

compounds identified need to be relatively volatile, while we have used LC-TOF-MS to

79

analyse higher MW (

228

than 80% (most matches were considerably higher). Similarity index, mass spectrum and

229

retention index were all used as selection criteria for identification, and compounds with a

230

match below 80% were denoted as “unknown peaks”. Possible contamination during

231

extraction was also investigated by soaking pipette tips, gloves, tubes, and septa in the

232

solvents for an hour, and then measuring this extract using GC-MS (Figure S2).

, and based on the alkane standards (C10 – C40). The MS was operated in electron

233 234

2.7 Estimation of SMPs concentration

235

Quantifying all the SMPs identified would be very time consuming based on individual

236

compound calibrations, and hence a crude approximation was made to estimate the

237

concentration of each SMP using alkanes as representative standards. The standard solution

238

contains different alkane chain lengths (C10 – C40) which covered most of the volatility

239

range of the RTX-5MS column, and retention indices were used for qualitative analysis

240

(retention index in GC-MS analysis was calculated based on straight-chain alkanes). The

241

calibration curves consisted of five concentrations (0.1, 0.25, 0.5, 1 and 2 mg/L), and the

242

correlation (R2) values were greater than 0.99 except for C38 and C40 (Table S1).

243

10 Environment ACS Paragon Plus

Page 10 of 24

Page 11 of 24

Environmental Science & Technology

244

Alkane standards quantified the unknown compounds based on their retention time, and this

245

was done separately for each unknown using the closest alkane standard which was run in

246

every batch of analyses. Blank solvents (methanol, acetone, dichloromethane, n-hexane) were

247

also run in every batch, and the instrument identification limit (IDL) of the alkane standards

248

was evaluated for each compound based on the maximum blank concentration, and a signal-

249

to-noise ratio of 3.

250 251

3. Results and discussion

252

3.1 SMP production in the SAMBR

253

The SAMBR was fed with the low strength synthetic wastewater (492±18 mg COD/L) under

254

steady state conditions (HRT = 6 h) for 30 days. The composition and concentration of SMPs

255

and COD were monitored over time to ensure that “steady state” had been reached in SMP

256

production, and hence the data reported here is the stable composition of the SMPs after HRT

257

changes. The SAMBR performance was excellent, with an effluent SCOD as low as 13±1

258

mg/L, and 98% COD removal; VFAs were not found in any samples in this experiment.

259

100% of the SCOD was accounted for by SMPs in the effluent. Biogas contained an average

260

of 74% methane, 11% carbon dioxide, and 15% nitrogen from soluble nitrogen gas in the

261

feed, with no oxygen detected. Biogas production averaged 2.77 L/day, accounting for 323

262

mL CH4/g.COD (82% of the theoretical value at 35 °C and 1 bar -the balance being biomass

263

growth).

264 265

3.2 Optimization of liquid-liquid extraction (LLE) for SMP analysis

266

The development of this technique required the selection of suitable solvents. Although water

267

is highly polar and immiscible with most organic solvents, polar solvents like the alcohols,

268

acetone and acids do not phase separate well from water using LLE techniques because of

11 Environment ACS Paragon Plus

Environmental Science & Technology

Page 12 of 24

269

their high solubility. From 30 common solvents, eight inexpensive and easily accessible ones

270

were chosen, and all exhibited good solubility for a range of organic compounds.

271 272

Initially, a series of sample extracts with different solvents was evaluated, and the average

273

peaks and their standard deviation (SD) were obtained from a quadruplicate analysis (n = 4).

274

The results shown in Table 1 reveal that n-hexane, chloroform and dichloromethane all

275

showed good extraction capability, while no peaks were found using n-heptane and toluene.

276

Other solvents like diethyl ether, methyl tert-butyl ether and ethyl acetate had a high degree

277

of miscibility with water resulting in poor extraction.

278 279

Table 1. Efficiency of different organic solvents evaluated for extraction. Sample volume (mL)

Solvents (ratio)

Extraction solvent volume (mL)

Total number of peaks (Mean ± SD)

n-Hexane

196 ± 2.1

n-Heptane

0

Toluene

0

Chloroform

215 ± 2.8

Diethyl ether

0

Methyl tert-butyl ether

200

Dichloromethane

20

0 232 ± 2.4

Ethyl acetate

0

n-Hexane + Chloroform (1:1)

0

n-Hexane + Dichloromethane (1:1)

0

Chloroform + Dichloromethane (1:1)

260 ± 3.4

n-Hexane + Chloroform + Dichloromethane (1:1:1)

274 ± 1.3

Note: SD = standard deviation

280 281

After GC-MS analysis, it was interesting to observe that 146 identical organic compounds

282

could be identified using either n-hexane, chloroform or dichloromethane as an extractant.

283

Some of the compounds detected such as alkanes, alkenes, fatty acids, and phthalates (in

284

particular dibutyl phthalate and bis(2-ethylhexyl)phthalate) have also been reported

285

previously

7, 8, 15, 18

. There were 13, 33, and 11 organic compounds that could only be

12 Environment ACS Paragon Plus

Page 13 of 24

Environmental Science & Technology

286

extracted independently by n-hexane, chloroform and dichloromethane, respectively, and this

287

led to investigating solvent extraction using a mixture of solvents. The average number of

288

compounds extracted was higher when a mixture of chloroform and dichloromethane, as well

289

as a mixture of these solvents plus n-hexane, was used compared to individual solvents. In

290

contrast, mixtures of n-hexane and chloroform, and n-hexane and dichloromethane, were

291

very ineffective extracting solutes because both these mixtures had a high miscibility in water,

292

and thus no solutes were extracted. Therefore, a mixture of n-hexane, chloroform and

293

dichloromethane (ratio 1:1:1) was selected as the optimal extraction solvent for further study.

294 295

Secondly, since the solvents chosen were flammable and potentially toxic, the optimal

296

solvent volume should be kept to a minimum. Different volumes (20 to 100 mL) of the

297

optima solvent mixture were used to extract the same volume of effluent, with all other

298

experimental conditions fixed. Figure S3 (Supporting Information) shows that the extraction

299

efficiency was directly affected by the extracting solvent volume; both the number of peaks

300

and total peak area obtained increased with an increase in the extraction volume up to 70 mL.

301

However, a plateau in both was observed after the solvent volume exceeded 80 mL, although

302

this decrease in peak number was not statistically significant. Hence, 70 mL of solvent

303

mixture was chosen as the optimum volume for subsequent experiments.

304

Thirdly, pH is often important in the extraction of organics from water samples because it can

305

change the molecular form of the organic influencing its charge/hydrophobicity, and thus

306

affecting extraction. “Salting out” with ionic salts is one method commonly used to enhance

307

LLE performance by decreasing the aqueous solubility of the analyte, and increasing its

308

partition coefficient

309

extraction in acidic (pH < 2) and alkaline (pH >11) solutions, and the salt effect by adding

310

NaCl and MgSO4 separately (Table S2). The results demonstrate that both solution

19

. In this work, the effect of pH was determined by conducting the

13 Environment ACS Paragon Plus

Environmental Science & Technology

311

modifications result in no obvious improvement compared to pH = 7 with no salt addition,

312

and may even decrease the efficiency. Hence, the following experiments were carried out at

313

pH = 7 with no additives.

314 315

Fourthly, the influence of sample size on extraction performance was evaluated using

316

different volumes of effluent, keeping all other conditions constant. When the volume of

317

sample increased from 200 to 400 mL, the number of compounds detected did not increase

318

significantly, although the total peak area almost doubled (Table S3). In contrast, as the

319

sample volume decreased from 200 mL to 50 mL, the number of peaks and their peak area

320

declined. Since the absolute mass of an analyte will decrease with a reduction in effluent

321

sample volume, some of the analytes' concentration could drop below their detectable level.

322

Hence, a minimum sample volume of 200 mL was deemed necessary to maximise the

323

extraction of low-molecular weight SMPs.

324 325

In summary, a 70 mL mixture of n-hexane, chloroform and dichloromethane was optimal at

326

pH = 7, with no addition of salts, to extract a minimum sample volume of 200 mL (~ 2.6 mg

327

COD), and these conditions could detect the most low MW SMPs.

328 329

3.3 Optimization of solid phase extraction (SPE) for SMP analysis

330

A. Comparison of OasisHLB and ENVI-Carb cartridge

331

Two cartridges, OasisHLB and ENVI-Carb, were tested to compare their performance on

332

capturing the compounds in effluent. Effluent from the SAMBR was collected and loaded

333

onto the cartridges on the same day. The compounds identified were categorized as alkanes,

334

alkenes, esters, alcohols, nitrogenated compounds (N-compounds), phenols, and others;

335

previous researchers also found similar groups of compounds in SMPs 15, 18. The reactor feed

14 Environment ACS Paragon Plus

Page 14 of 24

Page 15 of 24

Environmental Science & Technology

336

was also analysed to determine the compounds present, and the compounds found were then

337

subtracted from the effluent sample (Table S4). Both cartridges showed similar results in

338

terms of the number of compounds they separated (OasisHLB: 73 and ENVI-Carb: 70)

339

(Figure 1), however, the compounds both cartridges captured were quite different, with only

340

about a 50% similarity. OasisHLB retained more of the higher volatility compounds than

341

ENVI-Carb, with average retention indices (retention time normalized to the retention times

342

of adjacently eluting n-alkanes: lower retention index equals higher volatility) of 1820 and

343

1942, respectively. Both cartridges showed very good performance in retaining low MW

344

compounds (MW 580

401

Da) which resulted in a lower baseline in the LLE chromatogram, and better resolution. The

402

second was that dichloromethane was present in the sample after LLE, which affected

403

adsorption in SPE. The last issue was sample volume; for LLESPE, five extracts (200 mL

404

each) with LLE were needed for further extraction by SPE (1 L). For SPELLE, 1L of

405

sample was loaded onto the OasisHLB column, and then 200mL that had passed through the

406

cartridges was extracted using LLE. Hence, SPELLE was selected as an optimised

407

pretreatment for low MW SMPs. An effluent sample was collected and split into 3 samples

408

(Samples 1, 2, 3) for statistical validation of method reproducibility. Table 2 shows the

409

number of peaks (with minimum peak area/peak height > 3000) identified using a

410

combination of SPE and LLE for extraction, and demonstrates good reproducibility across

411

the three samples. The chromatograms of sample 2, blanks and alkane standards are shown in

412

Figure S5, S6 and S7, while the lists of compounds identified are shown in Tables S4 and S5.

413 414

Table 2. Combination of SPE and LLE for the extraction of low MW compounds in SMPs. Number of peaks (SPE)

Number of peaks (LLE)

Identified

Unidentified

Total

Identified

Unidentified

Total

Total number of peaks

56 52 53

22 25 24

78 77 77

67 68 68

113 109 110

180 177 178

258 254 255

Sample 1 2 3

415 416

Note: SPE = Solid phase extraction, LLE = liquid-liquid extraction, Identified = mass spectral similarity > 90%, Unidentified = mass spectral similarity < 90%

417

18 Environment ACS Paragon Plus

Page 19 of 24

Environmental Science & Technology

418

Table 3 compares our “Standard” method to other methods in the literature. LLE was the

419

most popular pretreatment in previous research, and Aquino

420

MW SMPs. He used four solvents in LLE extraction, and 18 and 9 compounds were found in

421

a sugar feed CSTR and a SAMBR, respectively. Zhou, Wu, She, Chi and Zhang 8 used LLE

422

with dichloromethane for analysing SMPs in the effluent of UASBs treating industrial

423

wastewater, and more than 27 compounds were identified. The same group of researchers

424

found less low MW SMP in a reactor fed with glucose using a similar LLE method 15. SPE

425

was also applied to separate compounds in the effluent from a SAMBR, and 4 compounds

426

were identified using the sorbent ENV+ in SPE 7. Another study used OasisHLB as a

427

pretreatment to identify compounds in the effluent of a SAMBR treating solid waste leachate,

428

and 11 compounds were identified 6. Until now there was no information on pretreatment

429

optimization in the literature; in this study pretreatment was optimized, and an average of 77

430

and 178 compounds were identified from SPE and LLE, respectively, an order of magnitude

431

more compounds identified than in past work.

7

was the first to analyse low

432 433

Table 3. Comparison of the methods used in this study and the literature.

Type of Samples

Pretreatment

Analysis

Number of compounds

Mixed liquor and effluent from CSTR, SAMBR

LLE (hexane, monochloro benzene, dichloromethane, ethyl acetate)

GC-MS (Column: SGE Phase BPX5, m/z 40 to 600)

CSTR: 18 compounds (LLE), 9 compounds (SPE)

from

Effluent SAMBR

from

Effluent from activated sludge

References

Aquino (2004)

SAMBR: 9 compounds (LLE), 4 compounds (SPE)

SPE (sorbent ENV+)

Effluent UASB

identified

LLE (dichloromethane)

GC-MS (Column: DB 5MS, m/z 33 to 500, NIST98 and WILEY Registry 7.0)

27 compounds (major)

Zhou et al. (2009)

SPE (OasisHLB), Elution solvent: 10%methanol/90%MTBE

GC-MS (Column: SGE HT5, m/z 33 to 500, NIST05 library)

11 compounds

Trzcinski and Stuckey (2009)

Lyophillized

LC-IT-TOF-MS (MW 100 to 4,000 Da)

Not mentioned

Mesquita (2010)

19 Environment ACS Paragon Plus

et

al.

Environmental Science & Technology

Effluent UASB

from

Effluent SAMBR

from

LLE (dichloromethane)

GC-MS (NIST98 WILEY Registry 7.0)

SPE (OasisHLB), Elution solvent: methanol, acetone, dichloromethane, hexane

GC-MS (Column: RTX-5MS, m/z 33 to 580, NIST11)

LLE: mixture of n-hexane, chloroform or dichloromethane

and

13 compounds (major)

Page 20 of 24

Wu and (2010)

Zhou

SPE: 63 compounds This study LLE: 227 compounds

Note: CSTR = continuous stirred tank reactor, LC-IT-TOF-MS = High resolution liquid chromatography ion trap time-of-flight mass spectrometry, LLE = liquid-liquid extraction, SAMBR = submerged anaerobic membrane bioreactor, SPE = solid phase extraction, UASB = upflow anaerobic sludge blanket

434 435

Finally, approximate quantification was done separately for each compound using the alkane

436

with the closest retention time. A set of standards was run in and between every batch of

437

analyses to minimise instrumental error, and blank solvents (methanol, acetone,

438

dichloromethane, n-hexane) were also run for background subtraction. Combined

439

concentrations of the low MW compounds using SPE and LLE were 0.009 mg/L and 0.598

440

mg/L, respectively, and the compounds identified from both methods accounted for 2.1 mg

441

COD/L (16% of the total SMP as COD calculated from 3.46 gO2/g docosane). While

442

concentration estimation using an alkane standard may not be the most accurate way to

443

quantify SMPs, the quantification is currently the best estimate and the most practical for

444

unknown samples.

445 446

Based on the results of this study, the combination of SPELLE coupled with the GC-MS

447

was shown to be “optimum” for separating and identifying low MW (