Subscriber access provided by Bethel University
Article
Comprehensive Target Analysis for 484 Organic Micropollutants in Environmental Waters by the Combination of Tandem Solid-Phase Extraction and Quadrupole Time-of-Flight Mass Spectrometry with Sequential Window Acquisition of All Theoretical Fragment-Ion Spectra Acquisition Kiwao Kadokami, and Daisuke Ueno Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.9b01141 • Publication Date (Web): 27 May 2019 Downloaded from http://pubs.acs.org on May 29, 2019
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Comprehensive
Target
Analysis
for
484
Organic
Micropollutants
in
2
Environmental Waters by the Combination of Tandem Solid-Phase Extraction
3
and Quadrupole Time-of-Flight Mass Spectrometry with Sequential Window
4
Acquisition of All Theoretical Fragment-Ion Spectra Acquisition
5 6
Kiwao Kadokami,1* Daisuke Ueno2
7
1 Institute
8
Hibikino, Wakamatsu, Kitakyushu, Japan
9
2 Graduate
10
of Environmental Science and Technology, The University of Kitakyushu, 1-1
School of Agriculture, Saga University, 1 Honjyo, Honjyo-machi, Saga,
Japan
11 12
*Corresponding author. Phone: +81 93 695 3739
13
E–mail addresses: kadokami@kitakyu–u.ac.jp
14 15 16 17 18 19 20
1 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
21
Abstract: There are many thousands of chemicals in use for a wide range of purposes,
22
and highly efficient analytical methods are required to monitor them for protection of
23
the environment. In order to cope with this difficult task, we have developed a novel,
24
comprehensive method for 484 substances in water samples. In this method target
25
chemicals were extracted by tandem SPE, and then were determined by
26
LC-QTOF-MS-SWATH. Targets were unambiguously identified using retention times,
27
accurate masses of a precursor and two product ions, their ion ratios and accurate
28
MS/MS spectrum. Quantitation was achieved by the internal standard method using a
29
precursor ion. Results of recovery tests at two concentrations (50 and 500 ng L-1)
30
showed average recoveries of 87.5 and 87.0% (RSD, 9.1 and 9.4%), respectively.
31
Limits of detection of half of the targets were below 1.0 ng L-1. The method was applied
32
to the influent and effluent of a sewage treatment plant, and around 100 chemicals were
33
detected. Results of examination on matrix effects using their extracts spiked with 209
34
pesticides showed that the ratios of detected amounts between the extracts and the
35
standard solution were 89.8% (influent) and 91.7% (effluent), respectively. In addition,
36
investigation on stability of calibration curves by injecting the same standards for one
37
year showed that their quantitative results did not change; average accuracy was
38
103.3 % (RSD, 10.0%), indicating that the calibration curves can be used for an
39
extended period of time without calibration and quantitative retrospective analysis can
40
be done after creating calibration curves for new targets.
41 42
Keywords: Comprehensive analysis; pesticides; pharmaceutical and personal care
43
products; SWATH; hybrid tandem mass spectrometer
44
2 ACS Paragon Plus Environment
Page 2 of 37
Page 3 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
45
Accompanying the desire for a wealthy and comfortable life, the number and volume of
46
chemical substances used worldwide have been rapidly increasing.1,
47
chemicals make their way into aquatic ecosystems. As a result, there are concerns about
48
the adverse effects of organic micro-pollutants (OMPs), especially agricultural
49
chemicals, pharmaceuticals and personal care products (PPCPs), and their metabolites
50
and decomposition products on aquatic organisms.3,
51
situation, it is important to understand the occurrences of as many OMPs as possible in
52
the aquatic environment.3, 5-7 Waterways managers are often reluctant to do the required
53
monitoring because of the cost, time and effort required to screen the huge number of
54
OMPs; this highlights the need for highly efficient, simultaneous determination
55
analytical methods. 3, 5-7
4
2
Many of these
In order to cope with this
56 57
We have developed highly efficient simultaneous analytical methods for ca. 1000
58
semi-volatile organic chemicals (SVOCs) by gas chromatography-mass spectrometry
59
(GC-MS).8 Many OMPs, however, have limited volatility and are thermally
60
decomposable substances and so are difficult to measure by GC-MS. Liquid
61
chromatography-MS (LC-MS) may be applied to analysis of such chemicals. However,
62
LC-MS has poor in-peak resolution and retention time reproducibility compared to
63
GC-MS. Moreover, although electro-spray ionization (ESI) is suitable for ionization of
64
polar substances, it is less likely to generate fragment ions than the electron ionization
65
processes used in GC-MS. To overcome these deficiencies, LC-MS-MS-multiple
66
reaction monitoring (MRM) may be used in environmental analysis9-12 due to its high
67
selectivity and sensitivity. However, since setting of measurement conditions is
68
complicated and time-consuming, this method is not suitable for screening a large
3 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 4 of 37
69
number of substances. To achieve comprehensive screening of numerous substances,
70
full-scan analysis by LC-high-resolution (HR)-MS such as LC-time-of-flight
71
(TOF)-MS5, 13-15 and LC-Orbitrap-MS3, 16 is used. However, even with the resolution of
72
LC-HR-MS, false detections occur frequently12-14 when analyzing trace amounts of
73
OMPs in complex matrices, which indicates the need to use additional fragment ions for
74
correct identification. To obtain fragments by LC-HR-MS, in-source fragmentation13-17
75
or collision-induced dissociation13, 14 in a collision chamber are used. When identifying
76
and quantifying OMPs using in-source fragmentation or collision-induced dissociation,
77
it is necessary to measure a sample twice for identification and quantitation.13-16 Also, if
78
multiple substances eluting out from a LC column at the same time are ionized, the
79
attribution of fragment ions to parent chemical is extremely difficult, which results in
80
mis-identification.15
81 82
Overall, LC-quadrupole (Q)TOF-MS is arguably one of the best instruments capable of
83
simultaneous analysis of many substances because sample identification and
84
quantitation are done using the same analytical run. There are three measurement modes
85
for simultaneous measurement in LC-QTOF-MS: all-ion-fragmentation (AIF),6,
86
data-dependent-acquisition (DDA),19, 20 and data-independent-acquisition (DIA)19, 20 but
87
the identification and quantitation of many substances using these modes have the
88
following problems. In AIF, since all precursor ions passing through a quadrupole mass
89
spectrometer
90
(mis-identification due to interference ions) as in-source fragmentation occurs
91
frequently.19, 20 With DDA only precursor ions having a minimum certain ion strength
92
are dissociated to generate product ions, which may result in environmental pollutants
are
dissociated
in
a
collision
chamber,
4 ACS Paragon Plus Environment
the
same
19, 20
problem
Page 5 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
93
with low concentrations not generating fragment ions from.19,
20Among
94
measurement modes, the most suitable one for comprehensive screening and
95
quantitation of OMPs in environmental samples and agricultural products is DIA.14, 19, 20
96
One DIA method is Sequential Window Acquisition of All Theoretical Fragment-Ion
97
Spectra Acquisition (SWATH) which has already been widely applied in proteomics,21,
98
22
99
after performing a TOF-MS scan for a fixed time, the TOF-MS scan range is divided
100
into smaller ranges by a quadrupole mass spectrometer, all precursor ions in each mass
101
range are sequentially dissociated in a collision chamber, and then product ions
102
generated are scanned in a TOF-MS. If the mass range separated is narrow, the
103
possibility of interference with co-eluted substances is low. Therefore, SWATH can
104
simultaneously perform TOF-MS scan and MS-MS scan for all peaks including small
105
peaks that DDA cannot perform, and also its selectivity is much better than that of AIF
106
due to very low interference of co-eluted substances, thus chance of miss identification
107
is low and sensitivity is high that are suitable for environmental analysis.
the three
metabolomics,23 and clinical and forensic toxicology research.24-26 In SWATH mode,
108 109
The aims of this study were to (1) develop a comprehensive screening method for 484
110
OMPs in environmental waters by LC-QTOF-MS-SWATH, and (2) to evaluate
111
performance and applicability of the method by using wastewater of a sewage treatment
112
plant (STP). We examined all the performance required for environmental analysis such
113
as selectivity, sensitivity, dynamic range, accuracy, precision, and matrix effects. In
114
addition, (3) we examined the stability of calibration curves for a year. If the slopes of
115
the calibration curves are stable, it is not necessary to update the calibration curves
116
when measuring samples, which saves cost and time, and also it is possible to create
5 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
117
quantitative methods for new substances, allowing for retrospective quantitative
118
analysis of stored LC-QTOF-MS data.
119 120
EXPERIMENTAL SECTION
121
Reagents and equipment. Chemical standards were purchased from Restek Japan
122
(Tokyo, Japan), Kanto Chemical (Tokyo, Japan) and Hayashi Pure Chemical (Osaka,
123
Japan). Analytical-grade pharmaceuticals were obtained from Kanto Chemical,
124
Funakoshi (Tokyo, Japan), Tokyo Chemical Industry (Tokyo, Japan), Wako Pure
125
Chemical Industries (Osaka, Japan), Dr. Ehrenstorfer GmbH (Augsburg , Germany),
126
LKT laboratories (St Paul, MN, USA), Sigma-Aldrich Japan (Tokyo, Japan), Toronto
127
Research Chemicals (North York, ON, Canada) and Santa Cruz Biotechnology (Dallas,
128
TX, USA). LC-MS-grade methanol and pesticide-grade dichloromethane were
129
purchased from Kanto Chemical. Stock solutions (1 mg mL-1 when possible) of each
130
substance were prepared with methanol or acetonitrile and kept at –20 °C in a freezer.
131
Multi-residue pesticide standards obtained from reagent companies were used as stock
132
solution. Working mixed standard solutions were made by diluting the stock solutions
133
with methanol. The deuterium labeled standards that were used as internal standards
134
(IS), surrogates and matrix substances that were used for evaluating matrix effects were
135
purchased from Kanto Chemical, Wellington Laboratories Japan (Tokyo, Japan),
136
Hayashi Pure Chemical, Cambridge Isotope Laboratories (Andover, MA, USA), and
137
Sigma-Aldrich Japan. The HPLC-grade ammonium acetate (1 mol L-1) used for HPLC
138
mobile phase was obtained from Wako Pure Chemical Industries. Special-grade
139
disodium hydrogenphosphate (Na2HPO4) and anhydrous sodium dihydrogenphosphate
140
(NaH2PO4) were purchased from Kanto Chemical and used to prepare the 1 mol L-1 6 ACS Paragon Plus Environment
Page 6 of 37
Page 7 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
141
NaH2PO4 – Na2HPO4 (pH 7.0) buffer solution that was used for adjusting the pH of
142
samples. LC-MS-grade water was obtained by purifying tap water in an Elga Purelab
143
Chorus 1 Analytical Research (Veolia Water, Tokyo, Japan). The SPE cartridges used
144
were Oasis HLB Plus and Waters Sep-Pak Plus AC2 (Nihon Waters, Tokyo, Japan).
145
Whatman GF/C glass fiber filters (47 mm diameter) were purchased from GE
146
Healthcare Japan (Tokyo, Japan). Millex-LG syringe filters (4 mm) were purchased
147
from Merck Millipore (Darmstadt, Germany). A GL-SPE vacuum manifold system was
148
purchased from GL Sciences (Tokyo, Japan). A Sciex ExionLC with a Sciex X500R
149
QTOF System (AB Sciex, Tokyo, Japan) was used for chemical separation and
150
determination. All glassware and plastic ware were cleaned with detergent and water,
151
washed in an ultrasonic cleaner, dried and rinsed with methanol before use.
152 153
Target chemicals. The number of the target chemicals examined in this study is 484
154
that are categorized into Table 1 and are listed in Table S1. These are comprised of
155
substances with a wide range of physicochemical properties (log Pow -1.55 – 8.53).
156 157
Analytical procedures. Sample preparation was performed according to Chau et al.15
158
In short, phosphate buffer (0.4 mL; 1 mol L-1; pH 7.0) if necessary and surrogate
159
standards (4 μg mL-1, 50 μL, Table 1 and Table S1) were added to a water sample (200
160
mL), and the mixture was then filtered with a 47 mm glass fiber filter. The aqueous
161
filtrate was then passed sequentially through an Oasis HLB cartridge and a Sep-Pak
162
AC2 cartridge at a flow rate of 10 mL min-1. The cartridges were then washed with 20
163
mL of purified water and dried by passing nitrogen through the cartridges for 40 min,
164
and then the targets were eluted from the AC2 side with 5 mL of methanol and 3 mL of
7 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
165
dichloromethane. The suspended solids (SS) remaining on the filter paper were
166
subjected to sonication extraction (Ultrasonic cleaner USK-3R, AS ONE, Osaka, Japan)
167
with 3 mL of methanol twice. After combining the eluate and the extract from SS
168
extraction, the mixture was concentrated to 400 µL under a gentle stream of nitrogen.
169
Mixed internal standards and matrix standards (standards for examining degree of the
170
matrix effects, 4 μg mL-1, 50 μL, Table 1 and Table S1) were added and the mixture
171
was reconstituted to 500 μL with methanol. The final extract was filtered with a syringe
172
filter prior to LC-QTOF-MS analysis.
173 174
Instruments and conditions. LC-QTOF-MS conditions are listed in Table 2 and Table
175
S2. LC conditions are the same as those of Chau et al.15 because the conditions are
176
suitable for measuring a large number of substances with broad physico-chemical
177
properties, although in this study we used Sciex X500R QTOF System and measured
178
the sample once only by SWATH to achieve reliable quantitation and identification
179
simultaneously. The TOF-MS scan range (m/z 50 to 1000 for 0.1 s) was divided into 22
180
ranges (SWATH window; Table S2) and each window was measured for 0.07 s by
181
MS-MS scan. The total cycle time was 1.768 s, so more than 10 sampling points for a
182
peak were obtained, which is enough for precise quantitative analysis. We can therefore
183
obtain accurate masses of a precursor ion and product ions without interference of
184
co-eluted peaks, which is most suitable for measuring a large number of substances at
185
the same time.19
186 187
Data processing. Sciex OS was used for from maintenance and control of the
188
LC-QTOF-MS instrument, measurement of samples to all data processing that is as
8 ACS Paragon Plus Environment
Page 8 of 37
Page 9 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
189
Analytical Chemistry
follows;
190 191
Building the in-house product ion mass spectral database. In order to reduce the
192
number of quantitation methods (calibration methods), seven 1 g mL-1 mixed standard
193
solutions were prepared. Each mixture was measured by Informed Data Acquisition
194
(IDA) with the LC-QTOF-MS. After measurement, each target chemical was found by
195
drawing extracted exact ion chromatograms of protonated ion [M + H] + or ammoniated
196
[M + NH4]+ adduct. Each target was manually identified using an accurate precursor ion
197
and accurate product ions. Where there were isomers in the mixture, it was difficult to
198
correctly identify the isomers using only an accurate precursor ion and accurate product
199
ions. For such cases, we distinguished the isomers using retention time obtained with an
200
ODS column and/or product ions obtained from mass spectral databases or reference
201
books and papers. After identification, the accurate product ion mass spectrum that was
202
obtained by experiment was registered in the in-house accurate mass spectral database.
203 204
Creating quantitation methods. The quantitation method for each mixed standard is
205
comprised of data for identification (retention times, a quantifier ion and two qualifier
206
ions and identification criteria) and data for quantitation (quantifier ions of targets and
207
internal standards) by using measurement data obtained by IDA. After creating
208
quantitation methods, mixed standard solutions (0.1, 1.0, 10, 100 and 1000 ng mL-1)
209
were measured by LC-QTOF-MS-SWATH. Then all targets contained in each mixed
210
standard solution were identified by a quantification method using retention times,
211
accurate masses of precursor and two product ions and their ion intensity ratios. Then
212
three internal standard calibration curves of a precursor and two product ions for each
9 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
213
target were created and registered in the quantitation method.
214 215
Identification and quantitation by quantitation methods. The software provides the
216
analyst with information to identify target substances based on retention time; accurate
217
mass precursor ion and its isotope pattern, two accurate mass product ions; and accurate
218
product ion mass spectra obtained by matching between the sample and a reference
219
product ion mass spectral database. Qualification rules are shown in Table 3. Although
220
the software provides these features, final chemical identification is under the purview
221
of the analyst and relies on the mass accuracies of a precursor and product ions, their
222
ratios, and retention time. Quantification of the identified target substance was
223
performed by the internal standard method using six internal standards (Table S1).
224
Precursor ions were usually used as the quantifier ion due to their typically having the
225
largest peak area of the three monitored ions, but if there were co-eluting peak
226
interferences, a fragment ion without interferences was used as the quantifier ion instead
227
of the precursor ion.
228 229
Recovery tests using reagent water. In order to evaluate accuracy and precision of this
230
multi-residue method, recovery tests were carried out according to the developed
231
analytical procedures at two concentrations (50 and 500 ng L-1) using reagent water.
232 233
Examination of the matrix effects using wastewater of a sewage treatment plant. It
234
is reported that the degree of the matrix effects varies according to retention times27
235
because eluted substances and their amounts also change in time. So we spiked 209
236
pesticides (100 ng each) covering all LC measurement time into the final extracts of an
10 ACS Paragon Plus Environment
Page 10 of 37
Page 11 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
237
STP influent and effluent sample and compared their detected concentrations to the
238
standard solution containing the same amounts of pesticides to examine the degree of
239
matrix effects.
240 241
Examination of stability of calibration curves. In conventional quantitative analysis,
242
calibration curves are usually updated when measuring samples. For instance, in an
243
official method for dioxin analysis,28 a representative concentration of dioxin standards,
244
usually the middle concentration of the range of calibration curve, is measured when
245
measuring samples. If the detected concentrations in this “check standard” are within ±
246
10% of the nominal concentration, update of the calibration curves is not needed.
247
Further, if the slopes of calibration curves do not change for a long time, it is also not
248
necessary to update calibration curves, which saves labor, time and cost. This
249
convention is very important for multi-residue methods because it is very difficult to
250
update calibration curves for hundred substances for every batch of samples analyzed.
251
In that context, in this study we examined stability of calibration curves by measuring of
252
209 pesticides (100 ng mL-1) using six columns including three used columns for one
253
year.
254 255
RESULTS AND DISCUSSION
256
Identification performance. In this study we divided the TOF-MS scan range (m/z 50
257
– 1000) into 22 smaller MS-MS scan ranges (Table S2), of which the MS-MS scan
258
ranges from m/z 200 to 400 are narrower than the others because more substances exist
259
in this range.19 By using this feature of SWATH, we can theoretically measure an
260
unlimited number of substances along with reducing the effects of interfering
11 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
261
substances and with high sensitivity. In addition, in order to reduce error of not
262
detection (false negative error), we used relatively loose identification criteria (Table
263
S3) because concentrations of OMPs in environmental samples are usually very low
264
compared to interfering substances. As a result, almost all of the chemicals detected in
265
the influent of a sewage treatment plant (STP) were able to be correctly identified
266
despite the sample containing a large amount of matrix by using retention times,
267
accurate precursor and two product ions, their ion ratios and precursor and product ions
268
mass spectra obtained by SWATH (Figure 1). Even with the high-resolution and smaller
269
MS-MS scan ranges provided by SWATH, a small number of target chemicals were
270
affected taken interference caused by co-eluting peaks; e.g. the precursor ion of
271
ampicillin (m/z = 350.117) (Figure 2) was interfered with, but since two product ions
272
(m/z = 160.043 and 114.037) had no interferences, the target was able to be easily
273
manually identified. In this case, quantitation was done using a product ion (m/z =
274
160.043) instead of the precursor ion. There were also some examples where a product
275
ion was interfered with, e.g. the product ion 2 of cotinine (m/z = 98.060) (Figure 3). In
276
such cases, if we used the strictest identification criteria, these substances would not be
277
found. Although almost of the targets can be correctly identified using accurate mass,
278
for some substances, particularly substances at low concentration, manual identification
279
is necessary for correct identification even when using SWATH.
280 281
Calibration curves and detection limits. The average of all coefficients of
282
determination was 0.972, and nearly 80% of the targets have dynamic range over three
283
orders of magnitude (Table S1). Method detection level (MDL) of each target was
284
estimated according to the relationship between the instrument detection level (IDL)
12 ACS Paragon Plus Environment
Page 12 of 37
Page 13 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
285
and MDL (IDL:MDL=1:4)29 and the concentration ratio (the ratio of the volume of a
286
sample (200 mL) to the volume of a final concentrate (0.5 mL): 400 times). The MDLs
287
of 236 substances out of 484 target substances were less than 1.0 ng L-1 with only 13
288
substances having a MDL higher than 100 ng L-1 (Table 3 and Table S1), showing that
289
the MLDs are good enough for environment analysis.
290 291
Recovery tests using reagent water. Eighty-one and 89% of the substances had
292
recoveries between 70 – 130% at 50 and 500 ng L-1, respectively. Average recoveries of
293
the 50 and 500 ng L-1 spikes were 87.0 and 86.4 %, respectively (RSD, 9.1 and 9.5%,
294
respectively; Table 4, Table S4-1 and Figure S1). Average recoveries of the surrogate
295
(Surrogate in Table S1) and matrix substances (Matrix substance in Table S1), which
296
were used for evaluating the extent of matrix effects, were 102.5 and 108.4%,
297
respectively (Table S4-2), (RSD, 9.0 and 10.2%, respectively). This showed there were
298
no matrix effects from the reagent water and that the extraction ability of the tandem
299
SPE method is adequate for screening a large number of substances with a broad range
300
of physicochemical properties (Figure S1). Substances with short retention times clearly
301
had lower recovery. Because polarities of substances are in inverse proportion to
302
retention times of LC with an ODS column,30 the reason for these low recoveries
303
appears to be their high water solubility. Moreover, recoveries gradually decreased with
304
increasing retention times (Figure S1), probably due to low elution efficiency of highly
305
hydrophobic substances from AC-2.
306 307
Matrix effects of wastewater of a sewage treatment plant. Matrix effects are the
308
major problem in quantitation by LC-MS-ESI.27,
31, 32
Therefore, we spiked a set of
13 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
309
deuterium-labeled surrogate compounds (Table S1) into all samples before extraction to
310
validate sample analysis through all procedures including the matrix effects. In addition,
311
we also spiked another set of deuterium-labeled compounds as matrix substance (Table
312
S1) into a concentrated extract before filtration to only evaluate the degree of the matrix
313
effects. Their recoveries in actual samples (influent and effluent of a STP) and blank
314
samples are shown in Table S5. Comparison between mean values of blanks to influents
315
shows that only one surrogate, sulfamethoxazole-d4, in STP influents was significantly
316
(p < 0.05) lower than the blanks. Further, only three matrices and one surrogate of
317
effluents were significantly (p < 0.05) lower than the blanks. However, recoveries of the
318
substances in these matrices/surrogates were only 20 to 37% those without the
319
matrix/blank effect, which is not a fatal error in screening analysis. The data for the
320
degree of the matrix effects across all LC measurement times using wastewater extracts
321
are shown in Table 5 and Table S6. Average ratios of detected amounts between the
322
extracts and the standard solution were 89.9% (influent) and 91.7% (effluent) with
323
RSDs of 22.5 and 22.7%, respectively. Moreover, the numbers of substances in the
324
influent and the effluent samples that were confirmed at concentrations in the range of
325
70 to 130% of their concentration in the standard solution were 173 and 177,
326
respectively. These results show that matrix did not affect quantitation, probably due to
327
the compensatory effect of using the internal standards.
328 329
Stability of calibration curves and retrospective quantitative analysis. The results of
330
calibration curves stability tests are shown in Table 6 and Table S7. Average
331
intermediate precision of the test pesticides was 10.4%, which was worse than 4.2% of
332
average repeatability (n = 7) of the same pesticides. The good, if not perfect, stability of
14 ACS Paragon Plus Environment
Page 14 of 37
Page 15 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
333
calibration curves has two benefits. First, it is not necessary to update calibration curves
334
at every sample analysis, which saves analytical time and cost. Second, it is possible to
335
add new targets into the database, and also perform retrospective quantitative analysis.
336
This performance could be one of the solutions of the reference standard dilemma.33 If
337
once we register new substances that are hard to obtain or are very expensive in a
338
quantitative method, we can determine them for a long time without re-procurement,
339
and so theoretically analyze an unlimited number of OMPs. Overall, the stability of
340
calibration curves may not be sufficient for regulatory purposes, but is more than
341
adequate for finding substances of high risk to human and aquatic organisms, finding
342
causes of environmental incidents, and confirming environmental safety after disasters
343
and accidents.
344 345
Application to wastewater of a sewage treatment plant. In order to confirm the
346
applicability and usefulness of the comprehensive screening method, we analyzed the
347
influent and effluent of a Japanese STP in the summer 2017. STP samples were chosen
348
because such samples are one of the most difficult environmental waters to analyze due
349
to their high matrix levels. Ninety five and 106 substances, mainly PPCPs and
350
pesticides, were detected in the influent and the effluent, respectively (Table S8). Mean
351
recoveries of the surrogates and matrix substances in the influent and the effluent were
352
85.1% (RSD, 19.2%) and 76.6% (RSD, 22.7%), respectively, which is almost the same
353
as those of the blanks using reagent water (recovery: 89.0%, RSD: 11.9%). In addition,
354
by utilizing the long term stability of calibration curves, retrospective analysis for
355
organophosphorus flame retardants was performed on the initial measurement data. As a
356
result, 9 out of 17 substances were found in the measurement data (Table S9). From
15 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
357
these results, it was confirmed that the method has sufficient applicability for
358
environmental waters, and has high usefulness because of detection of many OPMs with
359
low cost, a short time and reliable identification and quantification.
360 361
CONCLUSIONS
362
The experimental results clearly showed the effectiveness of the combination of the
363
tandem SPE and LC-QTOF-MS-SWATH method for the comprehensive target analysis
364
of 484 OMPs in environmental waters. Since the developed method utilizing SWATH
365
as the acquisition method that can measure all theoretical fragment-ion spectra, the
366
method has performance needed for future environmental monitoring34 and the
367
following advantages: ability (1) to simultaneously measure a large number of
368
substances in a short time with low cost and low labor requirements; (2) to perform
369
highly reliable identification using the excellent selectivity provided by SWATH; (3) to
370
provide reliable quantitative results by internal standard method using accurate ions
371
even high matrix samples; (4) to easily expand the number of targets until theoretically
372
an unlimited number of OMPs by utilizing the long term stability of calibration curves,
373
and to perform retrospective quantitative analysis after adding new targets. Many of the
374
target chemicals in our list are metabolized; adding metabolites and decomposition
375
products to the target list is urgently needed to enable the screening of the potentially
376
wide range of toxic metabolites in environmental waters. Such performance
377
characteristics will be very useful for finding substances posing high risk to human and
378
aquatic organisms, finding causes of environmental incidents, and confirming
379
environmental safety after disasters and accidents. In addition, the combination with
380
two comprehensive methods using LC-QTOF-MS and GC-MS,8 we can obtain a more
16 ACS Paragon Plus Environment
Page 16 of 37
Page 17 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
381
Analytical Chemistry
complete pollution picture of the environment.6
382 383
ASSOCIATED CONTENT
384
Supporting Information
385
The Supporting Information is available free of charge on the ACS Publications
386
website.
387
Table S1, List of target compounds; Table S2, SWATH windows; Table S3, Qualitative
388
rules; Table S4, Results of recovery test; Table S5, Recovery of surrogates and matrix
389
substances in influents and effluents in a sewage treatment plant; Table S6, Examination
390
results of matrix effects by comparing influent and effluent spiked with 209 pesticides
391
to pesticides standard solution; Table S7, Results of stability of calibration curves
392
(Intermediate precision); Table S8, Detected substances and their concentrations in
393
influent and effluent of a sewage treatment plant; Table S9, Results of retrospective
394
analysis of newly added substances; Figure S1, Relationship between recovery and
395
retention time.
396 397
AUTHOR INFORMATION
398
* E-mail:
[email protected]. Phone: +81-93-695-3739.
399 400
Acknowledgements
401
This study was supported by JSPS KAKENHI Grant Number JP16H02964. We are
402
grateful to Associate Professor Graeme Allinson (RMIT University, Melbourne,
403
Australia) and Dr. Mayumi Allinson (University of Melbourne, Melbourne, Australia)
404
for their kind proofreading, useful comments, and constructive suggestions on this
17 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
405
manuscript. We thank Dr. Takashi Miyawaki (Fukuoka Institute of Health and
406
Environmental Sciences) for his help to prepare mixed standard solutions of target
407
substances. We express our appreciation to Dr. Toshinari Suzuki and Dr. Yuki Kosugi
408
(Tokyo Metropolitan Institute of Public Health) and Dr. Hidenori Matsukami (National
409
Institute for Environmental Studies, Japan) for offer of standard substances.
410
REFERENCES
411
(1) Binetti, R.; Costamagna, F. M.; Marcell, I. Exponential Growth of New Chemicals
412
and Evolution of Information Relevant to Risk Control. Ann Ist Super Sanita, 2008, 44,
413
13-15.
414
(2) Bernhardt, E. S.; Rosi, E. J.; Gessner, M. O. Synthetic Chemicals as Agents of
415
Global Change. Front Ecol. Environ. 2017, 15, 84-90.
416
(3) Moschet, C.; Wittmer, I.; Simovic, J.; Junghans, M.; Piazzoli, A.; Singer, H.;
417
Stamm, C.; Leu, C.; Hollender, J. How a Complete Pesticide Screening Changes the
418
Assessment of Surface Water Quality. Environ. Sci. Technol. 2014, 48, 5423-5432.
419
(4) Eggen, R. I. L.; Hollender, J.; Joss, A.; Scharer, M.; Stamm, C. Reducing the
420
Discharge of Micropollutants in the Aquatic Environment: The Benefits of Upgrading
421
Wastewater Treatment Plants. Environ. Sci. Technol. 2014, 48, 7683-7689.
422
(5) Robles-Molina, J.; Lara-Ortega, F. J.; Gilbert-Lopez, B.; Garcia-Reyes, J. F.
423
Multi-residue Method for the Determination of Over 400 Priority and Emerging
424
Pollutants in Water and Wastewater by Solid-phase Extraction and Liquid
425
Chromatography-time-of-flight Mass Spectrometry. J. Chromatogr. A. 2014, 1350,
426
30-43.
18 ACS Paragon Plus Environment
Page 18 of 37
Page 19 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
427
(6) Moschet, C.; Lew, B. M.; Hasenbein, S.; Anumol, T.; Young, T. M. LC- and
428
GC-QTOF-MS as Complementary Tools for a Comprehensive Micropollutant Analysis
429
in Aquatic Systems. Environ. Sci. Technol. 2017, 51, 1553-1561.
430
(7) Schmidt, T. C. Recent Trends in Water Analysis Triggering Future Monitoring of
431
Organic Micro-pollutants. Anal. Bioanal. Chem. 2018, 410, 3933-3941.
432
(8) Jinya, D.; Iwamura, T.; Kadokami, K. Comprehensive Analytical Method for
433
Semi-volatile Organic Compounds in Water Samples by Combination of Disk-type
434
Solid-phase Extraction and Gas Chromatography-Mass Spectrometry Database System.
435
Anal. Sci., 2013, 29, 483-486.
436
(9) Rodil, R.; Quintana, J. B.; Lopez-Mahia, P.; Muniategui-Lorenzo, S.;
437
Prada-Rodriguez, D. Multi-residue Analytical Method for the Determination of
438
Emerging
439
Chromatography-tandem Mass Spectrometry. J. Chromatogr. A. 2009, 1216,
440
2958-2969.
441
(10) Shao, B.; Chen, D.; Zhang, J.; Wu, Y.; Sun, C. Determination of 76 Pharmaceutical
442
Drugs by Liquid Chromatography-tandem Mass Spectrometry in Slaughterhouse
443
wastewater. J. Chromatogr. A. 2009, 1216, 8312-8318.
444
(11) Gracia-Lor, E.; Sancho, J. V.; Hernandez, F. Simultaneous Determination of
445
Acidic, Neutral and Basic Pharmaceuticals in Urban Wastewater by Ultra High-pressure
446
Liquid Chromatography-tandem Mass Spectrometry. J. Chromatogr. A. 2010, 1217,
447
622-632.
448
(12) Gracia-Lor, E.; Sancho, J. V.; Hernandez, F. Multi-class Determination of Around
449
50 Pharmaceuticals, Including 26 Antibiotics, in Environmental and Wastewater
Pollutants
in
Water
by
Solid-phase
19 ACS Paragon Plus Environment
Extraction
and
Liquid
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
by
Ultra
450
Samples
451
Spectrometry. J. Chromatogr. A. 2011, 1218, 2264-2275.
452
(13) Gomez, M. J.; Gomez-Ramos, M. M.; Malato, O.; Mezcua, M.; Fernandez-Alba,
453
A. R. Rapid Automated Screening, Identification and Quantification of Organic
454
Micro-contaminants and Their Main Transformation Products in Wastewater and River
455
Waters using Liquid Chromatography–quadrupole-time-of-flight Mass Spectrometry
456
with an Accurate-mass Database. J. Chromatogr. A. 2010, 1217, 7038-7054.
457
(14) Wang, Z.; Chang, Q.; Kang, J.; Cao, Y.; Ge, N.; Fan, C.; Pang, G. F. Screening and
458
Identification Strategy for 317 Pesticides in Fruits and Vegetables by Liquid
459
Chromatography-quadrupole Time-of-flight High Resolution Mass Spectrometry. Anal.
460
Methods, 2015, 7, 6385-6402.
461
(15) Chau, H. T. C.; Kadokami, K.; Ifuku, T.; Yoshida, Y. Development
462
Comprehensive Screening Method for More Than 300 Organic Chemicals in Water
463
Samples
464
Chromatography-time-of-flight-mass Spectrometry. Environ. Sci. Pollu. Res. 2017, 24,
465
26396-26409.
466
(16) Cotton, J.; Leroux, F.; Broudin, S.; Poirel, M.; Corman, B.; Junot, C.; Ducruix, C.
467
Development and Validation of a Multiresidue Method for the Analysis of More Than
468
500 Pesticides and Drugs in Water Based on On-line and Liquid Chromatography
469
Coupled to High Resolution Mass Spectrometry. Water Research, 2016, 104, 20-27.
470
(17) Zomer, P.; Mol, H. G. J. Simultaneous Quantitative Determination, Identification
471
and
472
LC-Q-Orbitrap™-MS. Food Additives & Contaminants: Part A. 2015, 32, 1628-1636.
Using
Qualitative
a
High-performance
Combination
Screening
of
Liquid
Page 20 of 37
of
Chromatography-tandem
Solid-phase
Pesticides
in
Fruits
20 ACS Paragon Plus Environment
Extraction
and
Mass
of
and
Vegetables
a
Liquid
Using
Page 21 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
473
(18) Akiyama, Y.; Matsuoka, T.; Mitsuhashi, T. Multi-residue Screening Method of
474
Acidic Pesticides in Agricultural Products by Liquid Chromatography/Time of Flight
475
Mass Spectrometry. J. Pestic. Sci. 2009, 34, 265-272.
476
(19) Renaud, J. B.; Sabourin, L.; Topp, E.; Sumarah, M. W. Spectral Counting
477
Approach to Measure Selectivity of High-Resolution LC−MS Methods for
478
Environmental Analysis. Anal. Chem. 2017, 89, 2747-2754.
479
(20) Wong, J. W.; Wang, J.; Chow, W.; Carlson, R.; Jia, Z.; Zhang, K.; Hayward, D. G.;
480
Chang, J. S. Perspectives on Liquid Chromatography−High-Resolution Mass
481
Spectrometry for Pesticide Screening in Foods. J. Agric. Food Chem. 2018, 66,
482
9573-9581.
483
(21) Collins, B. C.; Hunter, C. L.; Liu, Y.; Schilling, B.; Rosenberger, G. R; Bader, S.
484
L.; Chan, D. W.; Gibson, B. W.; Gingras, A.; Held, J. M.; Hirayama-Kurogi, M.; Hou,
485
G.; Krisp, C. K.; Larsen, B.; Lin, L.; Liu, S.; Molloy, M. P.; Moritz, R. L.; Ohtsuki, S.;
486
Schlapbach, R.; Selevsek, N.; Thomas, S. N.; Tzeng, S.; Zhang, H.; Aebersold, R.
487
Multi-laboratory Assessment of Reproducibility, Qualitative and Quantitative
488
Performance of SWATH-mass Spectrometry. Nature Communications, 2017, DOI:
489
10.1038/s41467-017-00249-5.
490
(22) Gillet, L. C.; Navarrot, P.; Tate, S.; Rost, H.; Selevsek, N.; Reiter, L.; Bonner, R.;
491
Aebersold, R. Targeted Data Extraction of the MS/MS Spectra Generated by
492
Data-independent Acquisition: A New Concept for Consistent and Accurate Proteome
493
Analysis. Mol. Cell. Proteomics. 2012, 11, 1-17. DOI 10.1074/mcp.O111.016717.
494
(23) Hopfgartner, G.; Tonoli, D.; Varesio, E. High-resolution Mass Spectrometry for
495
Integrated Qualitative and Quantitative Analysis of Pharmaceuticals in Biological
21 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 22 of 37
496
Matrices. Anal. Bioanal. Chem. 2012, 402, 2587-2596.
497
(24) Roemmelt, A. T.; Steuer, A. E.; Poetzsch, M.; Kraemer, T. Liquid
498
Chromatography, in Combination with a Quadrupole Time-of-Flight Instrument (LC
499
QTOF), with Sequential Window Acquisition of All Theoretical Fragment-Ion Spectra
500
(SWATH) Acquisition: Systematic Studies on Its Use for Screenings in Clinical and
501
Forensic Toxicology and Comparison with Information-Dependent Acquisition (IDA).
502
Anal. Chem. 2014, 86, 11742-11749.
503
(25) Roemmelt, A. T.; Steuer, A. E.; Poetzsch, M.; Kraemer, T. Liquid
504
Chromatography, In Combination with a Quadrupole Time-of-Flight Instrument, with
505
Sequential Window Acquisition of All Theoretical Fragment-Ion Spectra Acquisition:
506
Validated Quantification of 39 Antidepressants in Whole Blood As Part of a
507
Simultaneous Screening and Quantification Procedure. Anal. Chem. 2015, 87,
508
9294-9301.
509
(26) Elmiger, M. P.; Poetzsch, M.; Steuer, A. E.; Kraemer, T. Assessment of Simpler
510
Calibration Models in the Development and Validation of a Fast Postmortem
511
Multi-analyte LC-QTOF Quantitation Method in Whole Blood with Simultaneous
512
Screening Capabilities using SWATH Acquisition. Anal. Bioanal. Chem. 2017, 409,
513
6495-6508.
514
(27) Al-Qaim, F. F.; Abduiiah, M. P.; Othman, M. R; Latip, J.; Zakaria, Z.
515
Multi-residue
516
Chromatography-time-of-flight-mass Spectrometry for the Analysis of Pharmaceutical
517
Residues in Surface Water and Effluents from Sewage Treatment Plants and Hospitals.
518
J. Chromatogr. A. 2014, 1345, 139-153.
Analytical
Methodology-based
22 ACS Paragon Plus Environment
Liquid
Page 23 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
519
(28)
JIS
K0312:
520
octachlorodibenzo-p-dioxins, tetra-through octachlorodibenzofurans and dioxin-like
521
polychlorinatedbiphenyls in industrial water and wastewater, Japanese Industrial
522
Standards Committee: Tokyo, 2008.
523
(29) Rice, E. W.; Baird, R. B.; Eaton, A. D.; Clesceri, L. S., Ed.; In Standard methods
524
for the examination of water and wastewater 22ed; American public health association,
525
American water works association, Water environment federation: Baltimore, 2012; pp
526
1-20.
527
(30) JIS 7260-117: 2006. Partition coefficient (1-octanol-water) – High performance
528
liquid chromatography (HPLC) method, Japanese Industrial Standards Committee:
529
Tokyo, 2006.
530
(31) Stahnke, H.; Reemtsma, T.; Alder, L. Compensation of Matrix Effects by
531
Postcolumn Infusion of a Monitor Substance in Multiresidue Analysis with
532
LC−MS/MS. Anal. Chem. 2009, 81, 2185-2192.
533
(32) Nurmi, J.; Pellinen, J. Multiresidue Method for the Analysis of Emerging
534
Contaminants
535
Chromatography-time-of-flight Mass Spectrometry. J. Chromatogr. A. 2011, 1218,
536
6712-6719.
537
(33) Moschet, C.; Piazzoli, A.; Singer, H.; Hollender, J. Alleviating the Reference
538
Standard Dilemma Using a Systematic Exact Mass Suspect Screening Approach with
539
Liquid Chromatography-High Resolution Mass Spectrometry. Anal. Chem. 2013, 85,
540
10312-10320.
541
(34) Perez-Fernandez, V.; Rocca, L. M.; Tomai, P.; Fanali, S.; Gentili, A. Recent
in
2008.
Method
Wastewater
for
by
determination
Ultra
23 ACS Paragon Plus Environment
of
tetra-through
Performance
Liquid
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
542
Advancements and Future Trends in Environmental Analysis: Sample Preparation,
543
Liquid Chromatography and Mass Spectrometry. Analytica. Chemica. Acta. 2017, 983,
544
9-41.
545
Figure captions
546
Figure 1. Example of a target (Ketoprofen) correctly identified by using retention times,
547
accurate precursor and two product ions, their ion ratios, and product ions mass
548
spectrum.
549
(A) Extracted ion chromatograms of a precursor ion and two product ions
550
Ion 1 (blue): precursor ion (m/z = 255.102), Ion 2 (purple): product ion 1 (m/z =
551
105.033), Ion 3 (red): product ion 2 (m/z = 209.096)
552
(B) MS/MS spectrum
553
Blue: deconvoluted spectrum from a sample, Red: spectrum from a sample, Black:
554
library spectrum
555 556 557
Figure 2. Example of a precursor ion (m/z 350.117) of a target (Ampicillin) being
558
interfered with other substance(s).
559
(C) Extracted ion chromatograms of a precursor ion and two product ions
560
Ion 1 (purple): precursor ion (m/z = 350.117), Ion 2 (blue): product ion 1 (m/z =
561
160.043), Ion 3 (red): product ion 2 (m/z = 114.037)
562
(D) MS/MS spectrum
563
Blue: deconvoluted spectrum from a sample, Red: spectrum from a sample, Black:
564
library spectrum
24 ACS Paragon Plus Environment
Page 24 of 37
Page 25 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
565 566 567
Figure 3. Example of a product ion (m/z = 98.060) of a target (Cotinine) being
568
interfered with other substance(s).
569
(E) Extracted ion chromatograms of a precursor ion and two product ions
570
Ion 1 (purple): precursor ion (m/z = 177.102), Ion 2 (red): product ion 1 (m/z =
571
80.489), Ion 3 (blue): product ion 2 (m/z = 98.060)
572
(F) MS/MS spectrum from a sample
573
25 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
574 575 576 577 578 579 580 581 582 583 584 585 586 587 588 589 590 591 592 593 594 595 596 597 598 599 600 601 602 603 604 605 606 607 608 609 610 611 612 613 614 615 616 617 618 619 620 621 622 623 624 625
Page 26 of 37
B
A Ion 1
Ion 2 Ion 3
Figure 1.
26 ACS Paragon Plus Environment
Page 27 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
626 627 628 629 630 631 632 633 634 635 636 637 638 639 640 641 642 643 644 645 646 647 648 649 650 651 652 653 654 655 656 657 658 659 660 661 662 663 664 665 666 667 668 669 670 671 672 673 674 675 676 677 678
Analytical Chemistry
B
A Ion 1
Ion 2 2 Ion 3
Figure 2.
27 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
679 680 681 682 683 684 685 686 687 688 689 690 691 692 693 694 695 696 697 698 699 700 701 702 703 704 705 706 707 708
Page 28 of 37
B
A Ion 3 Ion 1
Ion 2
Figure 3.
28 ACS Paragon Plus Environment
Page 29 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
709
Table 1. Summary of target substances Class Number* Pesticide 296 Pharmaceutical 156 Personal care product 18 Indusrial chemical 10 Others 4 Total 484 Internal standard 6 Surrogate** 4 Matrix substance*** 5 * = including metabolites; ** = substances for evaluating analysis; *** = substances for evaluating matrix effects 710 711
29 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 30 of 37
712
Table 2. LC/QTOF-MS conditions Sciex X500R QTOF Instrument system GL Science ODS-4 HP (150 mm, 2.1 mm, 3 Column µm) Column temp. 40 °C Flow rate 0.3 mL min-1 Mobile phase
A: 5 mM CH3COONH4 in H2O B: 5 mM CH3COONH4 in CH3OH
Gradient profile
Time,
Injection Volume Ion source Ionization Measurement mode TOF-MS (scan range) TOF MS/MS Collision energy ramp Mass resolution Total cycle time
B, % 5 95 2 L TurbolonSpray ESI-positive Swath 50 - 1000 Da, 0.1 s 50 - 1000 Da, 22 ranges, 0.07 s each 20 – 50 V 30000 1.768 s
min
0
30 - 40
713 714
30 ACS Paragon Plus Environment
40.01 50 5
Page 31 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
715
Table 3. Method detection limit (MDL), ng L-1 MDL Number < 0.1 25 0.1 - 1 211 1 - 10 191 10 - 100 44 > 100 13 716 717
31 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
718
Table 4. Results of recovery test n=5 Number of substances (%) Recovery, % Concentration, 500 ng L-1 Concentration, 50 ng L-1 < 20 5 (1.2) 1 (0.3) 20 - 40 4 (1.0) 4 (1.0) 40 - 60 13 (3.2) 27 (7.0) 60 - 80 75 (18.6) 89 (23.0) 80 - 100 266 (65.8) 183 (47.3) 100 - 120 38 (9.4) 73 (18.9) 3 (0.7) 10 (2.6) ≥ 120 719 720 721
32 ACS Paragon Plus Environment
Page 32 of 37
Page 33 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
722
Table 5. Examination results of matrix effects by comparing influent and effluent spiked with 209 pesticides with pesticides standard solution. Ratio of sample to standard, % (sample/standard)
Number of substance* Effluent
Influent
< 40 2 0 40 - 60 14 17 60 - 80 43 42 80 - 100 76 88 100 - 120 58 47 120 - 140 12 12 3 2 ≥ 140 * = One pesticide was excluded for evaluation due to low sensitivity. 723 724 725
33 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
726
Table 6. Intermediate precision* Intermediate precision, %