8 Determination of Molecular Weight and Size of Ultrahigh Molecular Weight Polymers Using Thermal Field-Flow
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
Fractionation and Light Scattering Seungho L e e a n d O h - S e u n g K w o n 3M C o m p a n y , 3M C e n t e r , St. P a u l , MN 55144
Thermalfield-flowfractionation was used to characterize ultrahigh molecular weight poly (methyl methacrylate) (PMMA) with a light-scattering detector. The influence of the differential refractive index increment (dn/dc) and the second virial coefficient (A ) on the measured molecular weight (MW) and the molecular size was investigated using a broad polystyrene standard having the nominal MW of 250,000. No significant change was observed in MW, and size with A -A could be assumed to be zero. Debye plot showed a good linearity for the entire range (0-180°) of the scattering angle. For ultrahigh MW PMMA, Debye plot was not linear, and the multiangle measurement was necessary for the extrapolation of data. Both MW and size increased with A , and thus A could not be assumed to be zero for ultrahigh MW polymers. 2
2
2
2
2
P O L Y ( M E T H Y L METHACRYLATE) ( P M M A ) has remained the most w i d e l y used material for the optic portion of the intraocular lens since it was first implanted into human eyes i n the late 1940s. It is dimensionally and chemically stable and more transparent than most other types o f optical glasses. T h e original type very high molecular weight ( M W ) P M M A . In this form, P M M A is amenable to lathe-cutting, compressioncasting, and cast-molding fabrication techniques. It can also be tumblepolished. Because of P M M A s excellent balance of properties, very little has been done until recently to develop other optic materials. Generally, acrylic polymers are brittle. M o d i f i e d acrylics having properties unattainable b y the basic unmodified compositions are n o w offered (1-3). 0065-2393/95/0247-0093$12.00/0 © 1995 American Chemical Society
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
94
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS
M W determination of these high M W polymers using size-exclusion chromatography (SEC) is sometimes difficult because portions of these materials are beyond the linear calibration range of most S E C columns. Thermal field-flow fractionation ( T h F F F ) has been used for the characterization of a wide range of organic-soluble polymers i n dilute solutions (4). In T h F F F , elution volume of a sample is a function of M W , and thus the molecular weight distribution ( M W D ) of the sample can be determined from its elution profile. T h F F F offers higher resolution than S E C i n its normal range of operation (5). T h F F F is particularly useful for characterizing very high M W polymers that are difficult to analyze using S E C (6-8) and microgel-containing polymers (8). T h F F F elution volume (or time) is a function of D / D , where D is the thermal diffusion coefficient and D is the mass diffusion coefficient (9, JO). T h e mass diffusion coefficient D of a polymer molecule i n a fluid w i t h viscosity η is given by (J J) T
T
0
(1)
where R is the universal gas constant, Τ temperature, N the Avogadro's number, M viscosity-average M W , and [η] intrinsic viscosity. If the value of D is available, the M W of a polymer can be determined directly from its T h F F F elution volume using equation 1. Values of D are not readily available and no theory exists to describe D w i t h k n o w n physicochemical parameters. A calibration is usually r e q u i r e d to determine M W D of polymers using T h F F F . It is noted that T h F F F has been used to study thermal diffusion phenomenon and to determine D (12, 13). MW-sensitive detectors (e.g., differential viscometer, light scattering detector) have been used to eliminate the need for calibration in T h F F F . W i t h a differential viscometer, the intrinsic viscosity distribution (IVD) of a polymer is measured. T h e I V D is then converted to M W D using M a r k - H o u w i n k ( M - H ) constants (14). The use of accurate M - H constant is essential i n this method. Low-angle laser light scattering ( L A L L S ) has also been used for T h F F F (15). U n l i k e viscometry, the light-scattering method measures the absolute M W of polymers directly. In multiangle laser light scattering ( M A L L S ) , the scattered light intensity is measured over a broad range of the scattering angles. Besides the M W , the mo lecular size can be measured from the angular dependence of the scat tered light intensity. A l t h o u g h M A L L S has been used i n S E C for various applications (16-18), it has not yet been used w i t h T h F F F . In this study, M A L L S was combined w i t h T h F F F to investigate the applicability of T h F F F for the characterization of ultrahigh M W polymers. A
T
T
T
T
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
8.
Ultrahigh Molecular Weight Polymers
L E E & KWON
95
Theory Theories on T h F F F and light scattering have been discussed i n numerous publications. Equations that are needed for the discussion of results are briefly reviewed here. In T h F F F , the retention ratio, R, is given b y (6) V° R = = 6\
(2)
for well-retained solutes. T h e full expression for R is somewhat com plicated (J9) and is not discussed here. V° is the channel volume and V is the elution volume. T h e retention parameter λ is related to D / D by r
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
T
ΔΤ
(3)
where A T i s the temperature drop across the channel. C o m b i n i n g equa tions 2 and 3, D / D can be calculated from the measured elution volume V . T h e M W is then determined using a calibration curve (log (D/D ) vs. log M) constructed w i t h a series of narrow standards. As M W i n creases, D decreases (eq 1), and λ decreases (eq 3). Thus i n T h F F F , l o w M W species elute earlier than high M W species. W h e n light passes through an inhomogeneous m e d i u m such as a polymer solution, it is scattered i n all directions. T h e light scattering at an angle θ by the solute is measured by the excess Rayleigh ratio Re w h i c h is defined b y T
r
T
TJ Η*
r -
(^fl,solution
J geom
^fl,solvent)
/A \
^
V*)
where i ^ o i u t i o n and I n v e n t are the intensities of the scattered light b y the solution and the solvent, respectively, and l is the intensity of the incident light. T h e geometric factor f = r^/V, where r is the distance between the scattering source and the detector and V is the scattering volume. F o r a dilute polymer solution, the excess Rayleigh ratio R is related to the weight-average molecular weight (M ) and the second virial coefficient (A ) of the polymer by (20) 0
geom
w
2
^
= MM-2A cM P ) 2
w
e
(5)
where K * is a constant defined b y (18)
Ν λ„ Α
4
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
W
96
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS
for vertically polarized light, dn/dc is the differential refractive index (RI) increment, n is the R I of the solvent at the incident wavelength λ . T h e solute concentration c (g/mL) is calculated by 0
0
Δ
(7)
η
(dn/dc)
where Δη is the difference i n R I between the solution and the pure solvent. ΡΘ is the scattering factor (or form factor) and is expressed as a power series i n s i n (Θ/2) as 2
Ρ = 1 -
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
β
+ a
a , s i n (|) 2
s i n (|)
-
4
2
a
sin ( f ) +
' ' '
6
3
(8)
w h i c h can be simplified to P,=
l - a ,
s i n (I)
(9)
2
for low scattering angles. T h e coefficient α = ( 4 x n / X ) ( r ) / 3 , where ( r ) is the %-average mean square radius of the polymer. T h e root mean square radius (or R M S radius) V ( r ) is sometimes called "radius of gyration'\ F o r each slice of the fractogram, the intensity of the scattered light is measured at a set of discrete scattering angles, and a D e b y e plot [RQJ K*c vs. s i n (0/2)] is constructed. W h e n θ = 0, ΡΘ = 1, and equation 5 becomes λ
g
2
0
0
2
g
2
z
z
g
2
z
2
M
where (R /K*c) e
tion 10, M
e=0
w
w
(1 -
2A cM ) 2
w
(10)
is the y-intercept of the D e b y e plot. B y solving equa
of the slice is obtained from
2 (intercept) 1 + V I - 8A c (intercept) 2
It is noted that the product, 2 A c M is m u c h smaller than 1 for most T h F F F (or S E C ) experiments, and equation 10 is further simplified to 2
f-^]
w
=M
W
(12)
Thus, M is directly obtained from the t/-intercept of the D e b y e plot. T h e mean square radius is obtained from the slope of the D e b y e plot w
by
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
8.
L E E àc KWON
97
Ultrahigh Molecular Weight Polymers
-3X m 0
2
0
(13)
167r n M (l - 4A cM ) 2
0
2
w
2
w
where m is the slope of the D e b y e plot at zero scattering angle, m 0
0
= d ( i V K * c ) / d ( s i n (fl/2))' 0 = 0 · 2
(
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
Experimental
Details
T h F F F . T h F F F was carried out with a Polymer Fractionator model T100 (FFFractionation, Inc., Salt Lake City, UT) equipped with a Waters model 590 pump and a 20-mL loop Rheodyne injector (Rheodyne Inc., Cotati, C A ) . Detectors were a M A L L S (Wyatt Technology model D A W N F, Santa Barbara, C A ) and a refractive index (RI; Hewlett Packard model 1037A, Palo Alto, CA) connected in series with the RI following the M A L L S . The light source of the M A L L S is a H e - N e laser (632.8 nm). The T h F F F channel is 0.0127 cm thick, 1.9 cm in breadth, and 45.6 cm long. Lightscattering data were collected and processed using the A S T R A software provided by Wyatt Technology. F o r conventional T h F F F experiments (without a light-scattering detector), a calibration curve (log (D/D ) vs. log M ) was constructed using a series of narrow polystyrene (PS) standards having M W s of up to 9.35 million D a . The calibration curve showed an excellent linearity for the entire M W range. The T h F F F - R I (refractive index) data were collected and processed using the software provided by F F F r a c tionation, Inc. T
dn/dc Measurement. A laser differential refractometer ( L D C / M i l t o n Roy model K M X - 1 6 , Riviera Beach, F L ) was used for dn/dc measurements. S E C . S E C was carried out at room temperature with an H P 1090 Chromatograph (Hewlett Packard, Palo Alto, C A ) equipped with an RI detector (Hewlett Packard model 1037A). Columns were Permagel 500-, 1 0 10 -, and 100-À columns (Column Resolution, Inc., San Jose, C A ) connected in series. A l l S E C experiments were run in tetrahydrofuran (THF) at 1.0 m L / m i n . Samples were dissolved in T H F and filtered through a 0.2-mm poly(tetrafluoroethylene) disposable filter (nonsterile, 25 mm disc). The i n jection volume was 100 m L . The column set was calibrated using a series of narrow PS standards having M W ranging from 2000 to 7.7 Χ 10 . The calibration curve (log M W vs. retention time) was obtained by fitting the data with a third-order linear regression, and the curve started deviating from linearity at the M W of ~ 5 million due to the column exclusion. 3
6
6
Materials. Narrow PS standards were obtained from Pressure C h e m ical Company (Pittsburgh, PA). A broad PS standard ( M W = 250,000) was obtained from American Polymer Standards Corp. (Mentor, O H ) . The P M M A materials were Perspex C Q U V obtained from Imperial Chemical Industries (Wilmington, D E ) and U V 52E obtained from Pharmacia Ophthalmics Inc.(Monrovia, C A ) . High-performance liquid chromatography-grade T H F from JT Baker Inc. (Phillipsburg, Ν J) was used as a carrier for all T h F F F and S E C experiments. The polymer solutions had concentrations of - 0 . 2 % (wt/vol).
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
98
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS
Results and
Discussion
The determination of polymer M W and R M S radius using light scattering measurement requires the knowledge of dn/dc and A values (see eqs 6, 7, 11, and 13). A broad PS standard having the nominal M W s of M = 100,000, M = 250,000, M = 430,000 was used to review the basics of polymer characterization using T h F F F - M A L L S - R I . F i g u r e 1 shows T h F F F elution curves of the PS standard obtained from light scattering (at 90°) and R I detector. T h F F F conditions were AT = 50 °C, flow rate = 0.3 m L / m i n w i t h the stop-flow time of 1 m i n . T h e values of dn/dc and A are available i n literature for many p o l y mer-solvent systems. F o r the polymer-solvent systems whose dn/dc and A values are not available, separate measurements are required. T h e dn/dc can be measured by differential refractometry and A by the static mode of light scattering. A range of dn/dc and A values are re ported for the P S - T H F system: dn/dc = 0.186 - 0.193 c m / g and A = 8.32 Χ 1 0 " - 2.11 Χ 1 0 " m L m o l / g at the wavelength of 633 nm (21). F o r the PS standard used i n study, the dn/dc value of 0.190 was measured using a laser differential refractometer. T h e A value was not measured separately. Table I shows the M W s and the R M S radii of the PS standard de termined w i t h different dn/dc and A values. Equations 11 and 13 i n dicate that both the M W and the R M S radius increase as A increases. N o significant changes were observed i n M W s and sizes w h e n A was varied at the constant dn/dc value of 0.2. It is noted that A c o u l d be assumed to be zero. As mentioned earlier, the product A c M is usually 2
n
w
z
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
2
2
2
2
3
4
4
2
2
2
2
2
2
2
2
M A L L S (90°)
I
0.0
Iii
.
.
.
I 5.0
.
•
.
.
I 10.0
.
L
Elution Volume (mL) Figure 1. ThFFF elution curves of a broad PS standard having nominal MWs o / M = 100,000, M = 250,000, and M = 400,000. ThFFF conditions are ΔΤ = 50 °C and flow rate = 0.3 mL/min. n
w
z
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
8.
99
Ultrahigh Molecular Weight Polymers
L E E & KWON
Table I. Molecular Weights and RMS Radii of 250,000 M W PS Standard Determined by T h F F F - M A L L S - R I Using Different dn/dc and A Values 2
dn/dc
A
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
0.200 0.200 0.200 0.200 0.200 0.200 0.190 0.180
Μ
2
1.48 1.48 1.49 1.49 1.50 1.48 1.56 1.65
0 1 X 10"-4 2 X 10"-4 4 X 10"-4 8 Χ ΙΟ-4 Ι χ ιο-4 ί χ ιο-4 ί Χ 10"-4
w
2.58 2.58 2.58 2.58 2.59 2.58 2.71 2.87
Χ ΙΟ 5 Χ ΙΟ 5 Χ ΙΟ 5
ΙΟ ΙΟ ΙΟ ΙΟ Χ ΙΟ
Χ Χ Χ Χ
Μ
M
η
5 5 5 5 5
3.99 3.99 4.00 4.00 4.01 3.99 4.20 4.44
Χ ΙΟ 5 Χ ΙΟ 5 Χ ΙΟ 5 Χ ΙΟ 5 Χ Χ Χ Χ
ΙΟ ΙΟ ΙΟ ΙΟ
5 5 5 5
*"g»n
ζ
Χ ΙΟ 5 Χ ΙΟ 5 Χ ΙΟ 5 Χ Χ Χ Χ Χ
ΙΟ ΙΟ ΙΟ ΙΟ ΙΟ
5 5 5 5 5
r
g,w
(nm)
(nm)
(nm)
15.2 15.2 15.2 15.2 15.2 15.2 15.2 15.2
20.0 20.0 20.0 20.0 20.0 20.0 20.0 20.0
25.0 25.0 25.0 25.0 25.0 25.0 25.0 25.0
much smaller than 1, and the A terms i n equations 11 and 13 are neg ligible. T h e term A cM becomes increasingly important as the M W or the concentration of the sample increases. T h e effect of A on the M W and size for ultrahigh M W P M M A materials is discussed later. The M W is inversely proportional to the product K * c (see e q 12) and the product K*c is proportional to dn/dc (see eqs 6 and 7). Thus for a given R , the calculated M W is inversely proportional to the dn/dc value used for the calculation. A s dn/dc decreases from 0.2 to 0.18 at the fixed A value of 1 X 1 0 ~ , the M W s increase proportionally as expected. Because the size is determined from the slope of the D e b y e plot, it is independent of dn/dc. Figure 2 shows the D e b y e plot for the slice at the elution volume of 3.7 m L . T h e values of dn/dc and A were taken as 0.19 and zero, 2
2
2
0
4
2
2
xlO
5
5.00 j
1
4.00
-
3.00
h
2.00
J-
1.00
-
* ^
0.00
I
0.0
1
0.25
1
1
0.50
0.75
1.00
Sin (0/2) 2
Figure 2. Debye plot for the PS standard shown in Figure 1 at the elution volume of 3.7 mL.
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
100
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS
respectively. T h e data were extrapolated using the first-order linear regression, and M W of 2.74 Χ 1 0 and R M S radius o f 20.9 n m were obtained for the slice (see eqs 11 and 13). T h e first-order least square line is shown as a solid line. A c c o r d i n g to equation 9, the D e b y e plot is linear at low scattering angles. F i g u r e 2 shows a good linearity for the entire range ( 0 - 1 8 0 ° ) of the scattering angles. T h e M and R M S radius determined for the whole distribution o f the P S standard are plotted against the elution volume i n Figures 3 and 4, respectively. F i g u r e 3 shows a good linearity between log M W and log V as expected from T h F F F theory (see eqs 1-3). A t the beginning and the end of the elution curve, the detector signal becomes too weak to measure the M W accurately. T h e noise shown at the high end of the elution volume is due to the weak R I response (relative to the lightscattering signal) as it approaches the baseline (see F i g u r e 1). F i g u r e 4 also shows the expected increase i n size w i t h increasing elution volume. T h e plot becomes noisy at the elution volume below about 3 m L . A s the molecular size becomes m u c h smaller than the wavelength o f the light source, the angular dependence of the light scattering disappears (isotropic scattering) and an accurate determination o f molecular size becomes difficult. Figure 5 shows the plot o f R M S radius versus M W on a l o g - l o g scale. T h e data for the elution volume lower than 3 m L were d r o p p e d 5
w
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
r
î.Oxio F
~
7
—
"
I
1.0x10
S3 ο
1.0x10
1.0x10 10.0
1.0 Elution Volume (mL)
Figure 3. ure 1.
MW versus elution volume for the PS standard shown in Fig
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
8.
101
Ultrahigh Molecular Weight Polymers
L E E & KWON 100.0
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
*
10.0
1.0 0
2
J
4
u
6
8
Elution Volume (mL)
Figure 4. EMS radius versus elution volume for the PS standard shown in Figure 1.
80.0
10.0 I 2x10
I
I
I
I
1
1
1
1 lxlO
s
6
Molecular Weight
Figure 5.
RMS radius versus MW for the PS standard shown in Figure 1.
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
102
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS
because of excessive noise i n the R M S radius as shown i n F i g u r e 4. T h e slope of the plot depends on the molecular density (17, 18). As the density of the molecule increases due to branching, cross-linking, or the existence of the microgels, and so on, the slope of the plot decreases, and thus the R M S radius versus M W plot can be used for polymer con formation studies (17). The first-order least-square fit of the data is shown as a dotted line. T h e slope is 0.57, w h i c h agrees w e l l w i t h the result reported elsewhere for P S - T H F system (18). T h e same T h F F F - M A L L S - R I system was used to characterize two ultrahigh M W P M M A materials (Perspex and U V 5 2 E ) . A power p r o gramming (22) was used for T h F F F operations w i t h the programming parameters of initial AT = 40 ° C , predecay time t = 5 m i n , t = —10 m i n , and the h o l d AT = 10 °C. T h e flow rate was fixed at 0.5 m L / m i n . F i g u r e 6 shows the traces from light-scattering (90°) and R I detector for the P M M A materials. F i g u r e 7 shows the D e b y e plot of the Perspex at the elution volume of 11 m L . U n l i k e the plot for the PS standard shown in F i g u r e 2, the plot is not linear: a fourth-order regression was required to fit the data. As M W increases, the A term (A cM) and thus the higher order terms i n equation 8 become increasingly important,
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
x
2
a
2
Perspex
Figure 6. Elution curves for power programmed ThFFF runs of PMMA materials with parameters tj = 5 min, t = —JO min, initial ΔΤ = 40 °C, and flow rate = 0.5 mL/min. a
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
8.
Ultrahigh Molecular Weight Polymers
LEE&KWON
103
xlO 10.0
6
7.50
5.00
υ
* *
2.50
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
0.00 0.0
0.25
0.50
0.75
1.00
Sin (0/2) 2
Figure 7.
Debye plot for PERSPEX at the elution volume of 11 mL.
and the D e b y e plot starts deviating from the linearity. Multiangle mea surement is thus necessary for the accurate determination of M W and R M S radius of ultrahigh M W polymers. A range of dn/dc and A values are also reported for the P M M A T H F system: dn/dc = 0 . 0 8 4 - 0 . 0 8 8 c m / g and A = 5.3 Χ 1 0 " to 1.1 X 10~ m L m o l / g at the wavelength of 633 n m (21). Table II shows the M W s and the R M S radii calculated for the Perspex using different d n / dc and A values. U n l i k e for the PS standard discussed earlier, both M W and R M S radius increase w i t h A . T h e A terms i n equations 11 and 13 become increasingly important as the polymer M W increases, and A can no longer be assumed to be zero. T h e calculated M W decreases w i t h increasing dn/dc, whereas the size remains unchanged. M W s and R M S radii of Perspex and U V 5 2 E obtained from T h F F F M A L L S - R I are summarized i n Table III. T h e values of dn/dc and A were taken as 0.083 and 2 Χ 10~ , respectively, for both polymers. Figures 8 and 9 show the plots of M W D and the R M S radius versus 2
3
4
4
2
2
2
2
2
2
2
4
Table II.
Molecular Weights and RMS Radii of P E R S P E X Determined by T h F F F - M A L L S - R I Using Different dn/dc and A Values 2
r
dn/dc 0.083 0.083 0.083 0.083 0.088
A
M
2
0 2 5 2 2
Χ X Χ X
10" 10~ 10~ 10~
4 4 4 4
4.31 4.42 4.61 4.42 4.15
10 10 10 10 Χ 10
Χ Χ Χ Χ
M
Μ„
n
6 6 6 6
6
5.87 6.10 6.55 6.10 5.73
10 10 10 10 Χ 10
Χ Χ Χ Χ
6 6 6 6
6
8.01 8.31 8.90 8.31 7.80
2
Χ 106 Χ Χ Χ Χ
10 10 10 10
6 6 6 6
g,n
r
g,w
r
g,w
(nm)
(nm)
(nm)
91.6 92.8 95.5 92.8 92.7
103.8 106.0 110.7 106.0 105.7
114.0 116.5 122.2 116.5 116.2
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
104
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS Table III. M W and RMS Radii Determined by T h F F F - M A L L S - R I for P M M A Materials g,w
Tg.w
(nm)
(nm)
(nm)
92.8 62.2
106.0 84.5
116.5 113.6
r
M
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
Perspex U V 52E
M
M
n
w
4.42 Χ 10 1.83 Χ 10
6
6
6.10 Χ 3.77 Χ
10 10
6
6
z
8.31 Χ 10 8.00 Χ 10
6
6
M W , respectively. In F i g u r e 9, no significant difference in the slope was found between two materials: 0.346 for Perspex and 0.354 for U V 5 2 E . Those slopes are lower than that obtained for the PS standard i n F i g u r e 5, w h i c h indicates both P M M A materials have higher molecular density than the PS standard. T h e P M M A materials were also characterized using conventional T h F F F and S E C without the use of light-scattering detector. R I traces from S E C are shown in F i g u r e 10, and the M W s determined by T h F F F and S E C are summarized i n Table I V . F o r the same sample, the M W s obtained from T h F F F are higher than those obtained from S E C . As the polymer size approaches the exclusion limit of the columns, the M W tends to be underestimated i n S E C . There is also a possibility of shear degradation as these ultrahigh M W polymers pass through the S E C columns. 1
2.0,
Molecular Weight Figure 8.
MW distributions of PMMA materials shown in Figure 6.
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
8.
105
Ultrahigh Molecular Weight Polymers
L E E & KWON
300.0 UV52E
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
Slope = 0.354
40.01 lxlO
I
I
!
•
•
•
• • i
lxlO
6
•
I I I
5xl0
7
7
Molecular Weight
Figure 9.
RMS radius versus MW for PMMA materials shown in Figure 6.
§
UV52E
0.0
5.0
10.0
15.0
20.0
Elution Time (min)
Figure 10.
SEC of PMMA materials in THF at 1 mL/min.
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
CHROMATOGRAPHIC CHARACTERIZATION OF POLYMERS
106
Table IV. M W Determined by T h F F F - R I and S E C - R I for P M M A Materials M
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
ThFFF-RI Perspex U V 52E SEC-RI Perspex U V 52E
2.77 Χ 10 1.37 Χ 10 0.832 Χ 10 0.577 Χ 10
6
6
6
6
7.11 Χ 10 3.99 Χ 10 2.14 Χ 10 1.67 Χ 10
6 6
6
6
z
11.6 Χ 10 7.24 Χ 10 3.58 Χ 10 3.06 Χ 10
6
6
6
6
Conclusions W i t h its multiangle capability, M A L L S can be used to measure the size as w e l l as the absolute M W of polymers. T h e multiangle capability seems to be particularly important for the determination of M W and R M S radius of ultrahigh M W polymers as the D e b y e plot deviates from the linearity. It is also important for the analysis of ultrahigh M W poly mers to use accurate A value as w e l l as dn/dc as the resulting M W and R M S radius tend to vary w i t h those values. E v e n w i t h an absolute M W detector such as light scattering, the accuracy of the polymer M W determined b y a separation technique depends on the resolution of the separator because the M W is calculated based on the assumption that each data slice is monodisperse. Thus, it is important to choose a separation technique that provides higher res olution for the polymers to be analyzed. T h F F F offers higher resolution than S E C for high M W polymers, particularly for polymers w i t h M W near or higher than 1 million D a . 2
Acknowledgments W e acknowledge L e n a Nilsson of W y a t t Technology for helpful discus sions on the light-scattering applications.
References 1. Apple, D . J.; Kincaid, M . C.; Mamalis, N . ; Olson, R. J. Intraocular Lenses— Evolution, Design, Complications, and Pathology; Williams and Wilkins: Bal timore, M D , 1989. 2. Bailey, D . ; Vogel, O. J. Macromol. Sci. Rev. Macromol. Chem. 1976, C-14, 267. 3. Tirrel, D . A . Polym. News 1 9 8 9 , 7, 104. 4. Giddings, J. C. Chem. Eng. News 1988, October 10, 34. 5. Gunderson, J. J . ; Giddings, J. C. Anal. Chim. Acta 1986, 189, 1. 6. Gao, Y. S.; Caldwell, K. D . ; Myers, M . N . ; Giddings, J. C. Macromolecules 1985, 18, 1272.
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.
Downloaded by FUDAN UNIV on January 13, 2017 | http://pubs.acs.org Publication Date: May 5, 1995 | doi: 10.1021/ba-1995-0247.ch008
8.
L E E & KWON
Ultrahigh Molecular Weight Polymers
107
7. Giddings, J. C.; Li, S.; Williams, P. S.; Schimpf, M . E . Makromol. Chem., Rapid Commun. 1988, 9, 817. 8. Lee, S. In Chromatography of Polymers: Characterization by SEC and FFF; Provder, T . E d . ; ACS Symposium Series 521; American Chemical Society: Washington, D C , 1993; pp 77-88. 9. Thompson, G . H . ; Myers, M . N . ; Giddings, J. C. Anal. Chem. 1969, 41, 1219. 10. Hovingh, M . E.; Thompson, G . H . ; Giddings, J. C. Anal. Chem. 1970, 42, 195. 11. Rudin, Α.; Johnston, Η. K. Polym. Lett. 1971, 9, 55. 12. Schimpf, M . E.; Giddings, J. C. J. Polym. Sci. Polym. Phys. Ed. 1989, 27B, 1317. 13. Schimpf, M . E.; Giddings, J. C. J. Polym. Sci. Polym. Phys. Ed. 1990, 28B, 2673. 14. Kirkland, J. J.; Rementer, S. W . Anal. Chem. 1992, 64, 904. 15. Martin, M . ; Hes, J. Sep. Sci. Technol. 1984, 19, 685. 16. Wyatt, P. J.; Jackson, C.; Wyatt, G . K. Am. Lab. 1988, May, 86. 17. Johann, C.; Kilz, P.J.Appl.Polym. Sci. Appl. Polym. Symp. 1991, 48, 111. 18. Wyatt, P. J.J.Liq. Chromatogr. 1991, 14(12), 2351. 19. Gunderson, J. J.; Caldwell, K. D . ; Giddings, J. C. Sep. Sci. Technol. 1984, 19(10), 667. 20. Zimm, B. H . J. Chem. Phys. 1948, 16(12), 1093. 21. Polymer Handbook, 3rd ed.; Brandrup, J . ; Immergut, Ε. H . , Eds.; Wiley: New York, 1989. 22. Williams, P. S.; Giddings, J. C. Anal. Chem. 1987, 59, 2038.
RECEIVED for review January 6, 1994. ACCEPTED revised manuscript June 13, 1994.
Provder et al.; Chromatographic Characterization of Polymers Advances in Chemistry; American Chemical Society: Washington, DC, 1995.