A Low-Cost Quantitative Absorption Spectrophotometer - Journal of

Sep 7, 2012 - For a more comprehensive list of citations to this article, users are encouraged to perform a search inSciFinder. Cover Image ... Stepha...
0 downloads 6 Views 1MB Size
Article pubs.acs.org/jchemeduc

A Low-Cost Quantitative Absorption Spectrophotometer Daniel R. Albert, Michael A. Todt, and H. Floyd Davis* Department of Chemistry and Chemical Biology, Baker Laboratory, Cornell University, Ithaca, New York 14853-1301, United States S Supporting Information *

ABSTRACT: In an effort to make absorption spectrophotometry available to high school chemistry and physics classes, we have designed an inexpensive visible light absorption spectrophotometer. The spectrophotometer was constructed using LEGO blocks, a light emitting diode, optical elements (including a lens), a slide-mounted diffraction grating, and a photodiode detector. The photodiode detector was mounted on a rotatable arm for wavelength selection based on elementary laws of diffraction. This simple design demonstrates basic physical principles (such as diffraction and absorption of light) that are frequently lost in commercial “black box” instruments. The homemade spectrophotometer’s performance, as measured by comparison to a commercial spectrophotometer, was shown to be sufficiently quantitative to facilitate experiments in chemistry or physics classrooms. KEYWORDS: First-Year Undergraduate/General, High School/Introductory Chemistry, Physical Chemistry, Hands-On Learning/Manipulatives, Dyes/Pigments, Laboratory Equipment/Apparatus, Quantitative Analysis, Spectroscopy, UV−Visible Absorption

A

The lowest-cost modern spectrophotometers (e.g., Vernier SpectroVis Plus) cost at least $500 and also require a PC for data acquisition and manipulation.3 To help address this deficiency, we developed a simple, rugged, and low-cost homemade visible spectrophotometer. The homemade spectrophotometer could be constructed by high school students in physics or chemistry classes using readily available components. The total cost of the homemade spectrophotometer is about $25, an eighth of the cost of a similar device4 and significantly less than commercial units. The design uses a battery pack, a white light emitting diode (LED) with current-limiting resistor (similar to those in LED flashlights), a lens, a cuvette and cuvette holder, a grating mounted on a slide, a photodiode mounted on a rotatable arm using a simple steel hinge, and a digital multimeter. For ease of alignment and use, we made an “optical table” using half of a 10 in. x 10 in. LEGO baseplate and LEGO construction blocks. It is exciting for students to perform scientific experiments with familiar materials typically thought of as toys.5 We note that the University of WisconsinMadison Materials Research Science and Engineering Center has published many exciting and educational experiments on “Exploring the Nanoworld with LEGO Bricks”.5 A photo of our homemade spectrophotometer is shown in Figure 1. A number of previous designs have been aimed at inexpensive and robust spectrophotometers. The most notable being the LEGO spectrophotometer by Knagge et al.,4 the lowcost LED spectrophotometer by Yeh and co-workers,6 Sheeline’s cell phone spectrophotometer,7 and a homemade spectrophotometer by Lema et al.8 The target audience for Knagge’s LEGO spectrophotometer is first-year graduate students and fourth-year undergraduate students, which make it too complex for typical high school and introductory

bsorption spectrophotometry is a commonly taught topic in advanced high school and introductory undergraduate chemistry laboratories. Historically, analog absorption spectrophotometers capable of measuring percent transmission at variable wavelengths (e.g., Spectronic 20) have been employed. More recently, the availability of USB-based visible spectrophotometers (such as those manufactured by Ocean Optics) have made it possible for students to rapidly obtain visible and nearUV absorption spectra and measure absorbance values at specific wavelengths. Over the past several years, we have led a 4-h module in Cornell’s Institute for Chemistry Teachers (CICT) program, held annually in the summer. This one-week program brings about 25 high school chemistry teachers from around the United States to Ithaca, NY to participate in a week of development projects. The module involving the determination of the pKa of indicators has been modified from an existing experiment1 and is intended to be used in high school or introductory chemistry laboratory courses. The existing pKa experiment may be adapted to the wide range of spectrophotometric equipment available at different schools. The flexibility of this experiment is important because access to spectrophotometers at different institutions can vary widely. The experiment is carried out by high school teachers participating in CICT program as well as undergraduate students using Ocean Optics USB2000+ spectrophotometers. These spectrophotometers are comparable to research-grade instruments and, when used carefully, allow one to determine pKa values to within 0.1 unit of literature values. Although the commercially available spectrophotometers provide accurate data, they largely represent a “black box”2 that provides the user with little insight into the elementary physics associated with spectrophotometry. Additionally, more than 75% of the high school teachers polled stated that the high cost of modern spectrophotometers was prohibitive for regular classroom use. © 2012 American Chemical Society and Division of Chemical Education, Inc.

Published: September 7, 2012 1432

dx.doi.org/10.1021/ed200829d | J. Chem. Educ. 2012, 89, 1432−1435

Journal of Chemical Education

Article

getting it to work is a quick and easy task, making the apparatus work well requires additional time and attention to detail. In our experience, the spectrophotometer can be constructed and optimized in approximately 45 min using the instructions provided in the Supporting Information.



EXPERIMENTAL SECTION Equation 1 is the Fraunhofer diffraction equation: (1)

nλ = d sin θ

where the wavelength of light (λ) diffracted at a measured angle (θ) can be calculated. The grating order is n (here n = 1), and the grating line spacing d for a 1000 line/mm grating is 1.0 × 10−6 m. According to eq 1, orange light at 589 nm (λ = 0.589 × 10−6 m) is diffracted at θ = 36°. To a reasonable approximation, one can assume the measured voltage is directly proportional to the intensity of the transmitted light, leading to the following equation for the absorbance, A:

Figure 1. Fully assembled spectrophotometer consisting of (A) LED, (B) lens, (C) cuvette holder, (D) grating, (E) rotatable detector arm, (F) photodiode, (G) digital multimeter, (H) hinge, and (I) power supply for LED.

chemistry classrooms.4 Yeh et al.’s spectrophotometer uses a different LED for each wavelength, limiting it to the availability of different wavelength LEDs and the bandwidth of the LED output.6 Sheeline’s cell phone spectrophotometer focuses, not on the determination of accurate absorbance values at specific wavelengths, but on understanding certain elements of spectrophotometer design such as dynamic range, stray light, and so forth.7 Lema’s design achieves adequate results for low cost. However, one is not able to select specific wavelengths, only colors, that is, the user can only choose a “red” part of the spectrum, but not specifically 650 nm.8 Whereas each of these spectrophotometers has its own merits, our homemade spectrophotometer offers low-cost quantitative absorption measurements at well-defined wavelengths across the visible spectrum. At the same time, it provides the opportunity for students to understand the fundamental physical principles underlying absorption spectrophotometry. When we demonstrated a prototype of the spectrophotometer to 20 CICT high school teachers, the device received strong approval. We have compared the performance of this device to that obtained using an Ocean Optics USB2000+ spectrophotometer. The quantitative performance of the homemade spectrophotometer makes it a useful tool for a variety of high school or introductory undergraduate laboratory experiments.1,9−12

⎛ V − Vd ⎞ A = log10⎜ 0 ⎟ ⎝ V − Vd ⎠

(2)

where V0 is the voltage measured with a blank sample (distilled water), V is the voltage measured with the sample of interest, and Vd is the dark signal. The dark signal, which results from stray light that reaches the photodiode, was measured by placing a cuvette containing milk into the cuvette holder. To take measurements, the angle of the detector was set using a protractor. The cuvette containing the solution was placed into the cuvette holder, and a cardboard box was placed over the setup to exclude room light. The angle and voltage readings were recorded in tabular form and then converted to wavelength and absorbance values using eqs 1 and 2. Additional detail is available as Supporting Information. Figure 2 shows



MATERIALS AND CONSTRUCTION The primary focus for the instrument development was on simplicity, ease of construction and use, and cost minimization, without sacrificing performance. The entire apparatus was built on a 5 in. × 10 in. LEGO baseplate using LEGO blocks. The light source was a 5 mm diameter white LED and 50 Ω currentlimiting resistor, powered by three AA batteries. Although a lens is built into the LED package, a second lens was added to better collimate the light. White light passes through the sample, which is held in a plastic 1 cm × 1 cm cuvette. The transmitted light is dispersed by a 1000 line/mm slide-mounted diffraction grating. The diffracted light was detected with a photodiode mounted on a rotatable arm. Wavelengths were calculated directly from the detector angle using the Fraunhofer diffraction equation. The voltage produced by the photodiode was read using a digital multimeter. Detailed instructions for constructing the spectrophotometer, construction materials (and their cost) may be found as Supporting Information. Although setting up the apparatus and

Figure 2. Absorbance at 590 nm recorded using basic bromothymol blue solutions at various indicator concentrations. For each solution, the absorbance measured using the homemade spectrophotometer is plotted versus that measured with an Ocean Optics USB2000+ spectrophotometer. Circles represent absorbance measurements; the line represents a least-squares fit.

calculated absorbance values at 590 nm for basic bromothymol blue solutions at various indicator concentrations using the homemade spectrophotometer versus absorbance values measured using an Ocean Optics USB2000+ spectrophotometer. A least-squares fit to the data in Figure 2 shows that the absorbance measurements made using the homemade spec1433

dx.doi.org/10.1021/ed200829d | J. Chem. Educ. 2012, 89, 1432−1435

Journal of Chemical Education

Article

trophotometer are in close agreement with those made using the Ocean Optics USB2000+ over the typical range of absorbance measurements up to A ∼1. As is usual for absorption spectrophotometry, it is desirable to keep absorbance values within this range as measured signals become comparable to the dark signal (i.e., become