Arrayed Time-of-Flight Mass Spectrometry for Time ... - ACS Publications

The design and operation of an arrayed time-of-flight (TOF) mass spectrometer for simultaneous data acquisition from multiple samples is described. Ve...
3 downloads 0 Views 382KB Size
Anal. Chem. 2005, 77, 3954-3959

Arrayed Time-of-Flight Mass Spectrometry for Time-Critical Detection of Hazardous Agents Timothy J. Cornish,* Miquel D. Antoine, Scott A. Ecelberger, and Plamen A. Demirev

Applied Physics Laboratory, Johns Hopkins University, Mailstop 2-217, 11100 Johns Hopkins Road, Laurel, Maryland 20723-6099

The design and operation of an arrayed time-of-flight (TOF) mass spectrometer for simultaneous data acquisition from multiple samples is described. Versions of the instrument employ sets of two or four linear or reflectron mass analyzers. They are housed in the same vacuum chamber and utilize the same laser for ion desorption. Instrument performance is illustrated in the example of a two-linear-mass-analyzer array using MALDI-MS for mixtures of commercially available proteins as well as intact microorganisms. We also describe the properties of a novel short delay time (