Compact Ultra-high Pressure Nano-flow Capillary Liquid Chromatograph Xiaofeng Zhao, Xiaofeng Xie, Sonika Sharma, Luke T Tolley, Alex Plistil, Hal E. Barnett, Martin Brisbin, Adam C. Swensen, John Calvin Price, Paul B Farnsworth, H. Dennis Tolley, Stanley D. Stearns, and Milton L. Lee Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.6b03575 • Publication Date (Web): 29 Nov 2016 Downloaded from http://pubs.acs.org on November 30, 2016
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Compact Ultra-high Pressure Nano-flow Capillary Liquid Chromatograph
2 3
Xiaofeng Zhao,†a Xiaofeng Xie,†a Sonika Sharma,† Luke T. Tolley,‡ Alex Plistil,§ Hal E.
4
Barnett,§ Martin P. Brisbin,§ Adam C. Swensen,† John C. Price,† Paul B. Farnsworth,† H. Dennis
5
Tolley,|| Stanley D. Stearns,§ and Milton L. Lee*, †
6 7
†Department of Chemistry and Biochemistry, Brigham Young University, Provo, UT 84602, USA
8 9
‡
Department of Chemistry and Biochemistry, Southern Illinois University, Carbondale, IL 62901, USA
10
§
11
||
12
†a
13
*
14 15
ABSTRACT: A compact ultra-high pressure nano-flow liquid chromatograph (LC) was
16
developed with the purpose in mind of creating a portable system that could be easily moved to
17
various testing locations or placed in close proximity to other instruments for optimal coupling,
18
such as with mass spectrometry (MS). The system utilized innovative nano-flow pumps,
19
integrated with very low volume stop-flow injector and mixing tee. The system weighed only 5.9
20
kg (13 lbs), or 4.5 kg (10 lbs) without controller, and could hold up to 1100 bar (16,000 psi)
21
pressure. The total volume pump capacity was 60 µL. In this study, the sample injection volume
22
was determined by either a 60 nL internal sample groove machined in a high pressure valve rotor
23
or by a 1 µL external sample loop, although other sample grooves or loops could be selected.
24
The gradient dwell volume was approximately 640 nL, which allowed significant reduction in
25
sample analysis time. Gradient performance was evaluated by determining the gradient step
26
accuracy. A low RSD (0.6%, n = 4) was obtained for day-to-day experiments. Linear gradient
27
reproducibility was evaluated by separating a three-component polycyclic aromatic hydrocarbon
VICI, Valco Instruments, Houston, Texas 77055, USA
Department of Statistics, Brigham Young University, Provo, UT 84602, USA These two authors contributed equally to this work.
Corresponding author:
[email protected] 1 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
28
mixture on a commercial 150 µm inner diameter capillary column packed with 1.7 µm particles.
29
Good retention time reproducibility (RSD < 0.17%) demonstrated that the pumping system could
30
successfully generate ultra-high pressures for use in capillary LC. The system was successfully
31
coupled to an LTQ Orbitrap MS in a simple and efficient way; LC-MS of a trypsin-digested
32
bovine serum albumin (BSA) sample provided narrow peaks, short dwell time and good peptide
33
coverage.
34 35 36
Introduction Ultra-high pressure liquid chromatography (UHPLC) has been around for almost two
37
decades, introduced to overcome the pressure limitations of conventional pumping systems and
38
to allow the use of very small particles as stationary phases.1-3 Due to its excellent separation
39
efficiency, short analysis time, low solvent consumption and easy coupling to mass spectrometry
40
(MS), UHPLC has become a very popular and green analytical tool, and is becoming widely
41
used for environmental, biochemical and pharmaceutical analysis.4-10 Very small particles (sub-2
42
µm) can result in improved performance in LC, as well as high speed separations.1,11-14 However,
43
without shortening the column length, the use of small particles requires high pressures. While
44
typically not an issue with small diameter capillary columns, operating at high pressures can
45
create frictional heating, which can impact the retention of solutes and, in some cases, cause
46
decomposition1,14 and loss in column efficiency.15,16 Heat dissipation can be facilitated by
47
reducing the column diameter.1,12,17 Small diameter packed fused silica columns allow fast heat
48
dissipation, low organic solvent usage and minimal waste generation.
49 50
The first UHPLC system for capillary columns packed with 1.5 µm nonporous silica particles was a home-built system reported by Jorgenson’s group.1 The same group also
2 ACS Paragon Plus Environment
Page 2 of 23
Page 3 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
51
constructed a gradient UHPLC system and separated a tryptic protein digest using 1.0 µm
52
particles at 9000 bar.2 However, the system was very large, inconvenient, and not amenable to
53
routine use. A subsequent UHPLC system was constructed by modifying a Waters Model 6000
54
pump, which was coupled to a tandem MS using a capillary packed with 1.5 µm nonporous
55
particles.18 This system demonstrated a high duty cycle, good signal-to-noise and full automation.
56
This work led to the introduction of commercial UHPLC instrumentation in 2004.19 Currently,
57
more than 20 different UHPLC systems are commercially available, with upper pressure limits
58
between 600-1500 bar.4,20,21 An excellent review by Eeltink et al.22 of advancements in UHPLC
59
instrumentation and practice were recently published. This review includes improvements in
60
column technology, pump design, sample introduction, temperature effects, and methods
61
development. UHPLC systems with pressures up to 20,000 psi, submicron porous particles and
62
automated switching valves with large external sample loops (5-10 µL) are readily available.
63
However, most of these systems are limited to laboratory bench operation, and cannot be easily
64
moved from one place to another or used for field applications. Furthermore, when coupled to
65
MS or other instruments in the laboratory, long connecting tubing is often needed, which can
66
greatly reduce the performance and throughput of the UHPLC system.
67
Compared to standard columns, e.g., 4.6-1.0 mm i.d., extra-column band broadening
68
becomes much more critical when using smaller internal diameter columns.21 Fekete et al.
69
compared the extra-column peak variance contributions of several commercially available LC
70
systems, and concluded that losses in efficiency can reach 30–55% with these UHPLC systems,
71
even when optimized.23 Further improvements in instrumental design could be achieved by using
72
smaller volume capillary fitting, smaller and shorter connecting capillary tubing and a smaller
3 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
73
volume detector cell to reduce detrimental extra-column dead-volume.21, 23-27 Furthermore, small
74
dwell volume is also highly desirable for UHPLC.21
75
The injection system for UHPLC should hold extremely high pressure, deliver a
76
reproducible small volume, eliminate all carryover and minimize brand spreading.28 Static-split
77
injection technology was reported by Jorgenson’s group, which withstood high pressures.1,2,29
78
Wu et al. used a pressure-balanced injection method to introduce sample, and found it to be
79
effective and reproducible.3 A commercial system (Waters ACQUITY UPLC) uses a flow-
80
through-needle injector.30 The gradient passes through the needle during injection, ensuring
81
complete sample transfer. In addition to the above-mentioned concerns, the detector data-
82
acquisition rate is also very important. Narrow peaks (as small as 1 s in some cases) require at
83
least 10-15 data points per peak to accurately specify resolution.21
84
Significant effort has been made in the past to miniaturize LC.31-35 However, little effort
85
has been made to miniaturize UHPLC. Recently, we reported a compact gradient LC system with
86
an LED UV-absorption detector (260 nm).32 This system meets many of the requirements for
87
field-portability, i.e., light-weight (9 lbs), low mobile phase consumption and waste generation,
88
and battery operation. The pressure generated by this pump reaches up to 8,000 psi. Although
89
nano-flow UHPLC systems are commercially available, they are not yet portable. With parallel
90
successful efforts in miniaturization of MS, compact UHPLC-MS can be expected in the future,
91
which would certainly be attractive for field analysis.36
92
An important application for UHPLC is MS-based proteomics.37 Due to the complex
93
nature of protein/peptide samples, integrated UHPLC-MS is commonly used to analyze these
94
samples. UHPLC is used prior to MS to separate complex mixtures of peptides or proteins for
95
accurate identification and quantitation. By using long columns, high pressures, small particles
4 ACS Paragon Plus Environment
Page 4 of 23
Page 5 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
96
and long gradients, peak capacities as large as 1500 have been achieved for UHPLC-MS based
97
proteomics.38 Iwasaki et al. identified 41,319 typtic peptides from 5,970 proteins using an 8 h
98
gradient and 4 m long monolithic column.39 High separation power, fast analysis, small sample
99
volume and easy coupling to MS are some of the key requirements for an UHPLC system used
100
for proteomics. As demonstrated in our previous research, one important aspect of miniature LC
101
is small dwell volume.32 For LC-MS based proteomics, small dwell volume will lead to less
102
extra-column band broadening, higher efficiency, lower carryover, faster analysis (i.e., less time
103
for sample to reach the head of the column), and higher sample throughput. In addition, a
104
compact LC will facilitate optimum positioning with the MS inlet. A miniature LC also allows
105
coupling of multiple LC units with a single MS to further improve sample throughput.40 A
106
simplified portable UHPLC system would reduce the footprint of the system, increase the duty
107
cycle, simplify maintenance, reduce costs and improve separation power by reducing extra-
108
column volume.
109
In this work, a small, portable gradient ultra-high pressure nano-flow pumping system
110
was developed and evaluated using an on-column LED-based UV-absorption detector reported
111
earlier, 41 as well as with an LTQ Orbitrap MS. The system was designed specifically for
112
capillary columns. Gradient performance and separation were evaluated using this new
113
integrated UHPLC-UV system. Efficient coupling of the nano-flow UHPLC to an Orbitrap MS
114
for analyzing trypsin-digested BSA for proteomics was demonstrated.
115
Experimental section
116
Chemicals and reagents
117 118
Water, methanol, propyl paraben, and three polycyclic aromatic hydrocarbon (PAH) standards (fluorene, anthracene, and benzo[a]pyrene) were purchased from Sigma-Aldrich (St.
5 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
119
Louis, MO, USA). Acetonitrile (ACN) was purchased from Fisher Scientific (Pittsburgh, PA,
120
USA). All solvents and chemicals were HPLC or analytical grade. Chromatographically purified
121
bovine serum albumin (Sigma-Aldrich) ≥ 95% was prepared for bottom-up mass spectrometry
122
analysis using a filter-aided sample preparation (FASP) technique.42 Approximately 500 µg of
123
protein were loaded onto a 30 kDa cut-off ultra-filtration spin filter (Vivicon) for the FASP
124
protocol. Due to the high concentration of the final BSA peptide mixture, it was diluted 100 or
125
1000 fold and cleaned once more with a C-18 solid phase extraction cartridge (Sigma-Aldrich)
126
prior to injection into the LC.
127
Instrumentation
128
The pumping system for the new gradient nano-flow UHPLC instrument (Figure 1) was
129
engineered at VICI Valco Instruments (Houston, TX, USA). Compared to our previous gradient
130
pumping system, several important modifications were made as follows: (1) the “Y” design (i.e.,
131
two pumps and one valve) was changed to two individual pumps (each with stepper motor and
132
high-pressure valve), which were connected to a separate injection valve (Figure 1); (2) a new
133
proprietary sealing material was used to seal around the piston; and (3) smaller i.d. tubing was
134
used to reduce the dwell volume. The “Y” design used a size 11 actuator, which could only hold
135
up to 8,000 psi (550 bar) pressure, while the new design uses a size 17 actuator, which is more
136
accurate and can hold much higher pressure.
137
Overall, the pumping system included two pumps, each with its own high pressure valve
138
and stepper motor, a three-way static mixing tee connected to a serpentine tube for solvent
139
mixing, and another high pressure valve (18,000 psi) with a stop-flow injector. The pumping
140
system weighed 5.9 kg (13 lbs) and could hold up to 1100 bar (16,000 psi) pressure. Each pump
141
volume capacity was 30 µL. The stop-flow injector contained a sample groove in the rotor,with
6 ACS Paragon Plus Environment
Page 6 of 23
Page 7 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
142
sample volume of 60 nL, or a 1 µL external sample loop; however, the groove or external sample
143
loop could be engineered to deliver sample volumes anywhere between 5 nL and the 10 µL
144
range. Zero dead-volume (360 µm) fittings especially designed for capillary columns were used
145
for all connections, and no splitting was employed. This system provided accurate non-split flow
146
in the low nanoliter to microliter per minute range (as described previously33), and was fully
147
software-controlled. An on-column deep UV LED (260 nm) absorption detector described
148
earlier,41 or an LTQ Orbitrap MS (described later in this section) was used for detection.
149
System evaluation
150
The methods used to determine the dwell volume and gradient step accuracy were
151
reported earlier.31 In brief, methanol and propyl paraben in methanol were used as mobile phases
152
A and B, respectively. A linear gradient was programmed through a hollow capillary (150 µm
153
i.d.) from 0 to 100% B in 10 min with an isocratic hold at 100% B for several min. An injection
154
of uracil was used to determine the capillary dead volume. The dwell volume of the system was
155
calculated by subtracting the capillary dead volume from the gradient delay volume.
156
For evaluating step accuracy, the system was programmed from 0% B to 100% B 10%
157
intervals. The gradient was stopped every 10% step for several minutes to record the UV-
158
absorption signal. The absorbance recorded by the detector increased as the proportion of B in
159
the mobile phase increased, and reached a maximum at 100% B. Every measurement was
160
repeated at least four times to determine the precision of the results. The recorded absorbance
161
value at each step was divided by the absorbance recorded at 100% B, which served as a
162
reference.
163 164
Injection carry-over was studied by recording the signal obtained by first injecting uracil (0.33 mg/mL in an ACN/water mixture, 70:30 v/v). Then, the injector was thoroughly cleaned by
7 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
165
purging with mobile phase, and a blank injection (mobile phase) was made. Isocratic elution was
166
used (ACN/H2O, 70:30 v/v), and the flow rate was 2 µL/min. These experiments were repeated
167
three times, and the detector signals were recorded. For each experiment, the peak area in the
168
blank divided by the peak area of an analyte signal, multiplied by 100%, was reported as the
169
percent carry-over.
170
Chromatographic performance was evaluated using a packed capillary column
171
(ACQUITY BEH130 C18, 1.7 µm, 100 mm × 150 µm i.d., Waters, Milford, MA) to separate a
172
PAH mixture under gradient programming conditions. The post-column capillary was removed
173
and replaced by a UV-transparent fused silica capillary (150 µm i.d.), which was used as a
174
detection window. Gradient elution was programmed from 50% A to 100% B in 11.6 min.
175
Solvent A was water and solvent B was ACN. The flow rate was 2 µL/min. The retention time
176
reproducibilities (n = 4) of components were calculated.
177
Coupling to an LTQ Orbitrap MS
178
Polyimide coated fused silica tubing (75 µm i.d. × 365 µm o.d.) was used to prepare a
179
column for coupling to the MS. Before packing, one end of the capillary was melted using a
180
small butane micro torch, and pulled into a narrow tip (approximate 25 µm i.d.) to be used for
181
electrospray ionization (ESI). Then, the capillary was packed with 1.7 µm BEH C18 particles
182
(Waters, Milford, MA) by slurry packing. After the desired length (approximately 10 cm) was
183
packed, the column was flushed with ACN before being used for MS analysis.
184
A hollow capillary (25 µm i.d. × 365 µm o.d., 15 cm total) was used for the transfer line
185
between the column with integrated ESI tip and injection valve. A VICI zero dead volume union
186
was used to connect the transfer line and the column as well as to apply the high voltage for
187
electrospray (photographs of all connections and column with integrated electrospray tip are
8 ACS Paragon Plus Environment
Page 8 of 23
Page 9 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
188
shown in the Supplemental Information, Figures S1-S2). MS detection was carried out after ESI
189
in the positive-ion full-scan mode. An LTQ Velos Pro-Orbitrap Elite tandem MS from Thermal
190
Fisher (Waltham, MA USA) was used, and the settings of the MS were as follows: spray voltage,
191
2.5 kV; capillary temperature, 325oC; and scan range, 150-2000 m/z. Nitrogen was used as
192
sheath gas (3 arb. units). Data acquisition and analysis were performed using Thermo Xcalibur
193
software Version 2.2.
194
Results and discussion
195
Pump operation
196
The UHPLC system contained a four-port injection valve to switch between sample
197
loading and injection (Figure 2). Each port served a specific purpose as described in the
198
following: S is the sample injection port, W is the waste port, C is the column port, and M is the
199
port through which the mixed mobile phase flowed to the internal sample groove and then to the
200
column and detector. SG represents the sample groove in the rotor for loading the sample.
201
Software controls the pump to switch automatically between the “load” and “inject”
202
positions. In the load position, the pumps were automatically refilled with their respective mobile
203
phases A and B at the preset rate of 20 µL/min through a 200 µm i.d. tube. At the same time, the
204
sample was introduced manually into the 60 nL sample loop. After the valve was switched to the
205
inject position (i.e., the rotor rotated by 90o), solvents from the pumps were mixed at the tee to
206
form a gradient flow. Then this gradient flow was used to carry the sample in the sample loop
207
through the analytical column to the detector for analysis.
208
Dwell volume
209 210
The dwell volume is defined as the volume from the solvent mixing point to the head of the analytical column.43 The sample experiences undesirable isocratic broadening in the dwell
9 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
211
volume. When performing fast separations in LC, there are two main concerns related to large
212
system dwell volume: (1) unreliable gradient method transfer between columns of different
213
geometries, and (2) greater than expected separation time.21 Small dwell volume is highly
214
recommended for fast analysis, especially when small i.d. columns are used, such as capillary
215
columns. Knowing the dwell volume of a new system is also very critical for transferring
216
methods from conventional LC to UHPLC.
217
The dwell volume was calculated by subtracting the capillary dead volume from the
218
gradient delay volume. The dead volume was 1.91 µL (1.91 min × 1.0 µL/min), so the dwell
219
volume of this system was calculated to be 0.64 ± 0.008 µL. This value is less than half of the
220
dwell volume measured for our previous nano-flow gradient system (1.3 µL),32 and less than the
221
delay volume of a Waters NanoACQUITY UPLC system (< 1 µL).
222
Gradient profile
223
Page 10 of 23
Good gradient accuracy and precision is a prerequisite for achieving reproducible
224
separations. To test the accuracy of the gradient, experimental step errors were plotted against
225
their corresponding theoretical step percentage values as illustrated in Figure 3A. Deviations
226
from predicted values were less than 0.8%. A gradient step accuracy of ± 1% is considered to be
227
acceptable for an LC system.32 In order to evaluate the precision of the gradient, experimental
228
RSD values were calculated for a series of step percentages and plotted against their respective
229
theoretical step percentages. As shown in Figure 3B, the RSD values (n = 4) were less than 0.6%,
230
which is comparable to commercial instruments. Therefore, the gradient performance of this
231
system was found to be excellent in terms of step accuracy and precision.
232 233
Low carry-over is important for quantitative analysis in LC. Uracil was chosen as test analyte because a non-retained marker is commonly used to evaluate the injector performance.
10 ACS Paragon Plus Environment
Page 11 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
234
Data from this work showed that the nano-flow pump injector had low carry-over (0.20 ± 0.1%).
235
This indicated that the flow-through injector is very reliable for quantitative work.
236
Reproducible retention time is a fundamental requirement for chromatographic
237
separations. Under gradient conditions, high reproducibility comes from high gradient accuracy.
238
As shown in Figure 4, highly reproducible retention times of PAHs were obtained by using the
239
UHPLC system. The RSD values were calculated to be less than 0.12%, which is better than the
240
value (1.42%) obtained from our previous “Y” design gradient pumping system.32 Day-to-day
241
reproducibility was also measured from experiments conducted over three days with three
242
injections per day. The RSD was less than 0.17% for these PAH compounds. It should be noted
243
that there was still some dead-volume between the column and UV detector, so separation
244
performance could be further improved if this dead-volume was removed.
245
Performance under high pressure and coupling with MS
246
Another nano-flow UHPLC system with 1.0 µL sample loop was coupled to an LTQ
247
Orbitrap MS for proteomic analysis. Due to the small particle size (1.7 µm) used in the analysis,
248
higher backpressure (>10,000 psi) was required. Under high pressure, solvents compress to a
249
certain extent, which leads to inaccurate flow rate and a distorted gradient. Most commercial
250
pumping systems use reciprocating designs, while some use flow and pressure feedback control
251
to deal with solvent compressibility. In our system, for simplicity, no flow and pressure sensors
252
were employed.
253
Therefore, in order to obtain the desired linear gradient, we evaluated two different ways
254
to deal with solvent compressibility. In the first approach, at the beginning, both pumps were
255
moved at the same speed until pressure was built up, and then the pump for solvent A (water)
256
was programmed to pump at full speed while the other pump was temporarily stopped to wash
11 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 12 of 23
257
all of solvent B (acetonitrile) out of the line. Sample was then injected into the flow stream and
258
the elution gradient was started. Figure 3S in the Supplemental Information shows the gradient
259
profile for operating the pumping system in this manner. After the point of sample injection, a
260
near-linear gradient was generated. For the second approach, a series of gradient steps was
261
generated at the first of the run to compensate for the solvent compressibility. As shown in
262
Figure 5A, without an initial compensation procedure, the gradient was not linear, but jumped
263
rapidly from 0 to 40% solvent B (uracil was added to solvent B for easy detection) with a steep
264
slope at the beginning. By adding two gradient steps (gradient solvent B at 25% for 10 min and
265
then 5% for 10 min), a near linear gradient from 5% to 40% was obtained as shown in Figure 5A.
266
Both gradient compensation approaches were used to separate a trypsin-digested bovine
267
serum albumin (BSA) sample. Using the method of delay injection, a 25 min gradient from 5%
268
to 45% acetonitrile was used, for the chromatogram shown in Figure 5B. The FWHM peak width
269
was approximately 10 s, the dwell volume was less than 2 µL, and an identified peptide coverage
270
of over 90% was obtained using Mascot.44 Similar results were obtained using stepwise gradient
271
compensation as shown in Figure 5C, with a gradient from 5% to 30% in 30 min and FWHM of
272
approximately 15 s. Compared to commercial nano-flow LC systems that exhibit dwell times of
273
over 10 min (> 4 µL in volume), peptide coverages of 67% to 87%, and peak widths (FWHM) of
274
approximately 20 s (3-5 µm particles),45 the dwell time in our system was less than 4 min (using
275
delayed injection), which improved the duty cycle better than 20%. Further work is currently
276
being done to implement an initial solvent compression routine into the operating procedure.
277
Conclusions
278 279
Ultra-high operating pressures for capillary LC were made possible using a newly developed LC pumping system. This new nano-flow gradient system gave highly reproducible
12 ACS Paragon Plus Environment
Page 13 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
280
retention times of PAH compounds on a capillary column packed with sub-2 µm particles.
281
Efficient gradient separation of peptides was also demonstrated when coupled to an LTQ
282
Orbitrap MS. The features of the capillary nano-flow UHPLC system described in this study
283
include: (1) it is simple to operate and maintain, (2) it consumes a minimum amount of solvent,
284
(3) it is easy to position close to the electrospray source (which greatly reduces band broadening
285
and analysis time, and improves detection limits), and (4) it provides high resolution LC of
286
complex peptide mixtures for proteomics studies. Two simple approaches were demonstrated to
287
address solvent compressibility during initial pressurization of the system for ultrahigh pressure
288
operation.
289
Supporting Information
290
S1. Photographs of a column with integrated electrospray tip.
291
S2. Photographs showing the interface between the nano-flow UHPLC system and
292 293
Orbitrap MS. S3. Gradient profile obtained when both gradient pumps were initially moved at the same
294
speed until pressure was built up, and then the pump for solvent A was programmed to pump at
295
full speed while the other pump was temporarily stopped to wash solvent B out of the line.
296 297 298
REFERENCES
299
(1) MacNair, J. E.; Lewis, K. C.; Jorgenson, J. W. Anal. Chem. 1997, 69, 983-989.
300
(2) MacNair, J. E.; Patel, K. D.; Jorgenson, J. W. Anal. Chem. 1999, 71, 700-708.
301
(3) Wu, N.; Lippert, J. A.; Lee, M. L. J. Chromatogr. A 2001, 911, 1-12.
302
(4) De Vos, J.; Broeckhoven, K.; Eeltink, S. Anal. Chem. 2016, 88, 262-278.
303
(5) Bouvier, E. S. P.; Koza, S. M. TrAC, Trends Anal. Chem. 2014, 63, 85-94. 13 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
304 305
(6) Gumustas, M.; Kurbanoglu, S.; Uslu, B.; Ozkan, S. A. Chromatographia 2013, 76, 13651427.
306
(7) Denoroy, L.; Zimmer, L.; Renaud, B.; Parrot, S. J. Chromatogr. B 2013, 927, 37-53.
307
(8) Purcaro, G.; Moret, S.; Bučar-Miklavčič, M.; Conte, L. S. J. Sep. Sci. 2012, 35, 922-928.
308
(9) Kumar, A.; Saini, G.; Nair, A.; Sharma, R. Acta Pol. Pharm. 2012, 69, 371-380.
309
(10) Howard, J. W.; Kay, R. G.; Pleasance, S.; Creaser, C. S. Bioanalysis 2012, 4, 2971-2988.
310
(11) Broeckhoven, K.; Desmet, G. TrAC, Trends Anal. Chem. 2014, 63, 65-75.
311
(12) Jorgenson, J. W. Annu. Rev. Anal. Chem. 2010, 3, 129-150.
312
(13) Anspach, J. A.; Maloney, T. D.; Colón, L. A. J. Sep. Sci. 2007, 30, 1207-1213.
313
(14) Nováková, L.; Matysová, L.; Solich, P. Talanta 2006, 68, 908-918.
314
(15) Gritti, F.; Martin, M.; Guiochon, G. Anal. Chem. 2009, 81, 3365-3384.
315
(16) Gritti, F.; Guiochon, G. J. Chromatogr. A 2007, 1138, 141-157.
316
(17) Colón, L. A.; Cintrón, J. M.; Anspach, J. A.; Fermier, A. M.; Swinney, K. A. Analyst 2004,
317
129, 503-504.
318
(18) Tolley, L.; Jorgenson, J. W.; Moseley, M. A. Anal. Chem. 2001, 73, 2985-2991.
319
(19) Mazzeo, J. R.; D. Neue, U.; Kele, M.; Plumb, R. S. Anal. Chem. 2005, 77, 460 A-467 A.
320
(20) Fekete, S.; Schappler, J.; Veuthey, J. L.; Guillarme, D. TrAC, Trends Anal. Chem. 2014, 63,
321
2-13.
322
(21) Fekete, S.; Kohler, I.; Rudaz, S.; Guillarme, D. J. Pharm. Biomed. Anal. 2014, 87, 105-119.
323
(22) De Vos, J.; Broeckhoven, K.; Eeltink, S. Anal. Chem. 2016, 88, 262-278.
324
(23) Fekete, S.; Fekete, J. J. Chromatogr. A 2011, 1218, 5286-5291.
325
(24) Wu, N.; Bradley, A. C.; Welch, C. J.; Zhang, L. J. Sep. Sci. 2012, 35, 2018-2025.
14 ACS Paragon Plus Environment
Page 14 of 23
Page 15 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
326 327 328 329 330 331
Analytical Chemistry
(25) Omamogho, J. O.; Hanrahan, J. P.; Tobin, J.; Glennon, J. D. J. Chromatogr. A 2011, 1218, 1942-1953. (26) Alexander, A. J.; Waeghe, T. J.; Himes, K. W.; Tomasella, F. P.; Hooker, T. F. J. Chromatogr. A 2011, 1218, 5456-5469. (27) Gritti, F.; Sanchez, C. A.; Farkas, T.; Guiochon, G. J. Chromatogr. A 2010, 1217, 30003012.
332
(28) Swartz, M. E. J. Liq. Chrom. Relat. Tech. 2005, 28, 1253-1263.
333
(29) Jorgenson, J. W.; Guthrie, E. J. J. Chromatogr. A 1983, 255, 335-348.
334
(30) Waters product literature.
335
(31) Baram, G. I. J. Chromatogr. A 1996, 728, 387-399.
336
(32) Sharma, S.; Plistil, A.; Barnett, H. E.; Tolley, H. D.; Farnsworth, P. B.; Stearns, S. D.; Lee,
337 338 339 340 341
M. L. Anal. Chem. 2015, 87, 10457-10461. (33) Sharma, S.; Plistil, A.; Simpson, R. S.; Liu, K.; Farnsworth, P. B.; Stearns, S. D.; Lee, M. L. J. Chromatogr. A 2014, 1327, 80-89. (34) Sharma, S.; Tolley, L. T.; Tolley, H. D.; Plistil, A.; Stearns, S. D.; Lee, M. L. J. Chromatogr. A 2015, 1421, 38-47.
342
(35) Zhou, L.; Lu, J. J.; Gu, C.; Liu, S. Anal. Chem. 2014, 86, 12214-12219.
343
(36) Snyder, D. T.; Pulliam, C. J.; Ouyang, Z.; Cooks, R. G. Anal. Chem. 2016, 88, 2-29.
344
(37) Aebersold, R.; Mann, M. Nature 2003, 422 (6928), 198-207.(38) Shen, Y. F.; Zhao, R.;
345
Belov, M. E.; Conrads, T. P.; Anderson, G. A.; Tang, K. Q.; Pasa-Tolic, L.; Veenstra, T. D.;
346
Lipton, M. S.; Udseth, H. R.; Smith, R. D. Anal. Chem. 2001, 73 (8), 1766-1775.
347 348
(39) Iwasaki, M.; Sugiyama, N.; Tanaka, N.; Ishihama, Y. J. Chromatogr. A 2012, 1228, 292297.
15 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
349
(40) Li, S.; Hao, Q.; Gounarides, J.; Wang, Y. K. J. Mass Spectrom. 2012, 47, 1074-1082.
350
(41) Sharma, S.; Tolley, H. D.; Farnsworth, P. B.; Lee, M. L. Anal. Chem. 2015, 87, 1381-1386.
351
(42) Wisniewski, J. R.; Zougman, A.; Nagaraj, N.; Mann, M., Nature methods 2009, 6 (5), 359-
352
62
353
(43) Li, Y.; Tolley, H. D.; Lee, M. L. J. Chromatogr. A 2011, 1218,1399−1408.
354
(44) Perkins, D. N.; Pappin, D. J. C.; Creasy, D. M.; Cottrell, J. S. Electrophoresis 1999, 20,
355 356 357
3551-3567. (45) Liu, Q.; Cobb, J. S.; Johnson, J. L.; Wang, Q.; Agar, J. N. J. Chromatogr. Sci. 2014, 52, 120-127.
358 359 360 361 362
16 ACS Paragon Plus Environment
Page 16 of 23
Page 17 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
363
Figure captions
364
Figure 1. (A) Cut-away drawing of a single nano-flow pump; (B) photograph of the nano-flow
365
UHPLC with LED detector.
366
Figure 2. Operating schematics of the nano-flow UHPLC injection valve. (A) Load position, and
367
(B) inject position.
368
Figure 3. (A) Gradient step accuracy (±% solvent B) and (B) precision in gradient step accuracy.
369
Figure 4. Reproducibility in retention times of of PAHs.
370
Figure 5. (A) Comparison of gradient profiles (uracil was added to monitor the concentration of
371
solvent B, acetonitrile); blue: no compensation step with pump speed set to generate a linear
372
gradient from 0% to 40%; red: two step manual compensation, 10 min at 25% B and then 10 min
373
at 5% B, was used prior to starting the sample elution gradient. (B and C) Chromatograms of a
374
trypsin-digested BSA sample using coupled UHPLC-Orbitrap MS: (B) sample injected 20 min
375
after 10 min 50:50 (A/B, water/ACN) and 10 min 95:5 wash (total ion and extracted-ion
376
chromatograms, m/z = 632. (C) Sample injected at 0 min, compensation steps of 75:25
377
(water/CAN) for 10 min and 95:5 for 10 min before linear gradient of 5% to 30% was started
378
(total-ion and extracted-ion chromatograms, m/z = 654. Experimental conditions: column with
379
integrated ESI tip, 12 cm (75 µm i.d. × 365 µm o.d.); mobile phase A, 0.1% formic acid in
380
water/ACN (95:5, v/v); mobile phase B, 0.1% formic acid in ACN; flow rate at 500 nL/min;
381
injection volume, 1.0 µL.
17 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
382 383
Figure 1
18 ACS Paragon Plus Environment
Page 18 of 23
Page 19 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
384 385
Figure 2
386
19 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408
Figure 3
20 ACS Paragon Plus Environment
Page 20 of 23
Page 21 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
409 410
Figure 4
21 ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
A
411
B
412
C
413 414
Figure 5 22 ACS Paragon Plus Environment
Page 22 of 23
Page 23 of 23