Detecting Bacterial Surface Organelles on Single Cells using Optical Tweezers Johan Zakrisson, Bhupender Singh, Pontus Svenmarker, Krister Wiklund, Hanqing Zhang, Shoghik Hakobyan, Madeleine Ramstedt, and Magnus Andersson Langmuir, Just Accepted Manuscript • DOI: 10.1021/acs.langmuir.5b03845 • Publication Date (Web): 18 Apr 2016 Downloaded from http://pubs.acs.org on April 19, 2016
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Langmuir is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Detecting Bacterial Surface Organelles on Single Cells using
2
Optical Tweezers
3 4 5 6 7 8 9 10 11 12 13 14 15
Johan Zakrisson,† Bhupender Singh,† Pontus Svenmarker,† Krister Wiklund,† Hanqing Zhang,†Shoghik Hakobyan,‡ Madeleine Ramstedt, ‡ and Magnus Andersson†,*
16
Bacterial cells display a diverse array of surface organelles that are important for a range of
17
processes such as: intercellular communication, motility and adhesion leading to biofilm
18
formation, infections and bacterial spread. More specifically, attachment to host cells by
19
Gram-negative bacteria are mediated by adhesion pili, which are nm wide and µm long
20
fibrous organelles. Since these pili are significantly thinner than the wavelength of visible
21
light, they cannot be detected using standard light microscopy techniques. At present, there is
22
no fast and simple method available to investigate if a single cell expresses pili while keeping
23
the cell alive for further studies. In this study, we present a method to determine the presence
24
of pili on a single bacterium. The protocol involves imaging the bacterium to measure its size,
25
followed by predicting the fluid drag based on its size using an analytical model, and
26
thereafter oscillating the sample while a single bacterium is trapped by an optical tweezer to
27
measure its effective fluid drag. Comparison between the predicted and the measured fluid
28
drag thereby indicate the presence of pili. Herein, we verify the method using polymer coated
29
silica microspheres and Escherichia coli bacteria expressing adhesion pili. Our protocol, can
30
in real time and within seconds assist single cell studies by distinguishing between piliated
31
and non-piliated bacteria.
32
†
Department of Physics, Umeå University, 901 87 Umeå, Sweden
‡
Department of Chemistry, Umeå University, 901 87 Umeå, Sweden
*Corresponding author: Magnus Andersson, Department of Physics, Umeå University, Phone: +46 90 786 6336, e-mail:
[email protected] Keywords: optical tweezers, E. coli, pili, fimbriae, microsphere
Abstract
1 ACS Paragon Plus Environment
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
1. Introduction
2
Bacterial cells have a complex surface architecture with both nano- and micrometer long
3
surface organelles. These surface organelles are important for cellular functions such as;
4
adhesion, motility, transfer of genetic material, intercellular communication etc.1–3 In
5
particular, adhesion organelles expressed by bacteria are key factors for the biofilm formation
6
and for the initial attachment to host cells as well as for maintaining contact during the first
7
stages of bacterial colonization. A well-studied example is pili of pathogenic Escherichia coli
8
(E. coli), which are essential for initiation of urinary tract and gastrointestinal infections.4,5 In
9
particular, uropathogenic and enterotoxigenic E. coli, (UPEC) and (ETEC), respectively, are
10
known to express a plethora of coilable pili,6 which help the bacteria to resist shearing forces
11
in various host organs by reducing the instantaneous force on the adhesin.7 It is suggested that
12
challenging this switch by means of; proteins, chemicals and antibodies that bind to pili
13
epitopes is a possible approach of preventing bacterial adhesion.8,9 Detecting the presence of
14
pili and understanding its role in pathogenesis are thus important in the mission to develop
15
new drugs against antibiotic resistant bacteria, new antibacterial surfaces, and in vaccine
16
development.10–13
17
Detecting bacterial pili can be achieved by microscopic and molecular-biology methods.
18
Since the width of pili is in the nanometer range, it is usually not possible to visualize them
19
using quick diagnostic techniques such as light microscopy or dynamic light scattering.
20
Localization of pili is possible through the use of fluorescence microscopy techniques, while
21
electron microscopy (EM) together with atomic force microscopy provide high-resolution
22
images of pili or even smaller organelles.9,14–17 EM along with mathematical modelling have
23
been used to demonstrate the morphological states of pili together with other important
24
fundamental properties such as; numbers of pili on a cell, length of pili, rigidity, etc.18–20 In
25
addition, nuclear magnetic resonance spectroscopy (NMR) and Western blot analysis are
26
often used for indirect assessment of pili expression.9,21 2 ACS Paragon Plus Environment
Page 2 of 38
Page 3 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
One major drawback of the above-mentioned methods is that they are invasive, making it
2
impossible to perform any further analysis on the cells after detection of pili. In addition, most
3
of these methods cannot be applied to investigate the physiological and biomechanical
4
features of pili, i.e., relative extension length, unwinding force, bond kinetics, and elasticity
5
response. However, these features can in turn be assessed using force spectroscopy
6
techniques.22 Force spectroscopy offers a powerful and sensitive method to understand pili
7
mechanics and physiology at the single cell and the single organelle levels.23–25 However, the
8
expression of pili in a bacterial population ranges from high to none, which makes the
9
probability of capturing a piliated single cell for investigation extremely low,26 and thus these
10
types of experiments become tedious. Therefore, there is a need of a non-invasive method that
11
can distinguish piliated from non-piliated bacteria.
12
E. coli are rod-shaped bacterial cells with a typical length of 1-3 µm that express hundreds
13
of peritrichous pili that are 1-3 µm long and 7-10 nm wide.18,27 A high amount of pili
14
protruding out from the cell surface will increase the surface area to mass per bacterium. In a
15
fluid environment, this additional surface area causes an increased drag on the bacterium
16
compared to a “bald” bacterium of the same size that does not express any pili. Using this
17
phenomenon, we developed a simple method that can rapidly identify bacteria expressing pili.
18
Here we report that using a bright-field microscopy image together with a quick oscillation of
19
the sample with a trapped non flagellated bacterium in an optical tweezers instrument provide
20
enough information to differentiate bald bacteria from piliated bacteria. An illustration of the
21
proposed method is shown in Figure 1. Since this is a quick and non-invasive method for cell
22
sorting, the cells can directly be used for downstream applications. For examples:
23
investigating the mechanical properties of pili; expression analysis in response to various
24
external stimuli and drugs such as pilicides; bacterial pathogenesis; attachment of nano-scale
25
animate or inanimate objects on the cell surface and biofilm formation. In addition, combined
26
with flow chambers, the method could be further developed for large scale cell sorting. 3 ACS Paragon Plus Environment
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1 2
Figure 1
3
4 ACS Paragon Plus Environment
Page 4 of 38
Page 5 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
2. Materials and Methods
2
2.1 Bacterial cultures
3
E. coli strain HB10128 is a non-piliated strain, which we here refer to as having bald cells. The
4
HB101 cells were transformed with plasmids pHMG93 and pAZZ50. The transformed strain
5
HB101/pHMG93 expressed P pili 29, whereas the HB101/pAZZ50 strain expressed SII pili.30
6
Expression of pili was confirmed using scanning electron microscopy. The HB101 strain was
7
grown on Lysogeny Agar (LA), and strains HB101/pHMG93 and HB101/pAZZ50 were
8
grown on LA supplemented with 100 µg/ml ampicillin. All strains were grown at 37°C and
9
passaged once before harvesting in 1x phosphate buffered saline, pH 7.4 (1xPBS).
10
2.2 Scanning Electron Microscopy (SEM)
11
Bacterial cells were harvested and fixed in 2% paraformaldehyde for 15 minutes at room
12
temperature. The fixed samples were then cast onto poly-L-lysine coated glass slides,
13
dehydrated in series of graded ethanol (70, 80, 90, 95, 100%), critical point dried and coated
14
with an iridium layer of 5 nm. Thereafter, the samples were examined by field-emission SEM
15
(Zeiss Merlin, GmbH) using in-lens secondary electron detector at a beam accelerating
16
voltage of 4 kV and a probe current of 150 pA.
17
2.3 Chemicals
18
Silica microspheres with a diameter of 2.0 µm (Catalog Code SS04N, Manufacturer Lot
19
Number: 7829 – Bang Laboratories); 3-Sulfopropylmethacrylate, K salt (SPM) (CAS:31098-
20
21-1,
21
Copper(I)chloride (CuCl) (CAS: 7758-89-6, Sigma Aldrich); Copper(II)chloride (CuCl2)
22
(CAS: 7447-39-4, Sigma Aldrich); Initiator – (3-Trimethoxylsilyl)propyl 2-bromo-2-
23
methylpropionate (CAS: 314021-97-1, Fluorochem Lt, UK).
24
2.4 Deposition of silane initiator
25
In a small centrifuge tube, a volume of 960 µL of methanol, 15 µL of deionized (DI) water
26
and 9 µL of acetic acid were added to a volume of 25 µL of silica microsphere water
Sigma
Aldrich);
2,2'-Bipyridyl
(BiPy)
(CAS:
5 ACS Paragon Plus Environment
366-18-7,
Sigma-Aldrich);
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
suspension. The mixture was shaken for 1 hour. A volume of 20 µL of initiator was added to
2
the mixture and the centrifuge tube was placed on a rocker for 20-24 h. The tube was
3
removed from the rocker, centrifuged and the solvent was aspirated out. A volume of 1-2 mL
4
of methanol was added and the silica microspheres were re-dispersed using an ultrasonic bath.
5
We repeated the last step 5 times.
6
2.5 SPM Polymerization
7
The silica microspheres grafted with initiator31,32 were sonicated in a volume of 2 mL of
8
methanol to disperse and were thereafter degassed by bubbling with nitrogen gas (N2 gas). In
9
a round-bottomed flask, 8.6 g of SPM monomer was dissolved by stirring in a solution of
10
methanol (8 ml) and water (5 ml) and degassed with N2 gas for 15 min. To the monomer
11
solution, 325 mg of BiPy and 7.4 mg of CuCl2 were added and the mixture was degassed by
12
purging N2 gas for 30 min prior to the addition of 82.4 mg of CuCl. The polymerization
13
solution was left for 20 min under N2 gas. Thereafter, the reaction solution was syringed over
14
to the microspheres and left under N2 gas for 120 minutes. The polymerization was stopped
15
by adding DI water and bubbling with air through the solution until it turned green or blue.
16
The mixture was shaken with DI water and centrifuged at least 5 times. This process is
17
estimated to have given a thickness of the polymer brush coating on the microspheres
18
corresponding to around 100 nm.
19
2.6 Assay preparation
20
Bacteria were suspended in 1xPBS to a concentration suitable for single cell analysis. Silica
21
microspheres were similarly suspended in Milli-Q water. A sample chamber was prepared by
22
adding two pieces of double-sided Scotch adhesive tape (product no 34-8509-3289-7, 3M)
23
spaced 5 mm apart, on a 24.0 x 60.0 mm coverslip (no.1, Knittel Glass). A 20.0 x 20.0 mm
24
coverslip (no.1, Knittel Glass) was thereafter gently positioned on top of the tape, forming a
25
5.0 x 20.0 x 0.1 mm chamber. The chamber was infused with either the bacterial suspension
26
or the silica microsphere suspension by adding a few µl of suspension at one of the openings 6 ACS Paragon Plus Environment
Page 6 of 38
Page 7 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
allowing capillary forces to fill the chamber. To avoid drying of the sample, the chamber was
2
sealed by vacuum grease (DOW CORNING®) at the open ends of the chamber. The sample
3
was thereafter mounted in a sample holder that was fixed to a piezo-stage (Physik Instrument,
4
P-561.3CD stage) in the optical tweezers (OT) instrumentation. The temperature was
5
measured using a thermocouple in the sample chamber to 23.0 °C ± 0.1 °C. Moreover, the
6
suspension viscosity was assumed to only vary with temperature, thus, the viscosity was set to
7
0.932 mPas ± 0.002 mPas.
8
2.7 Optical tweezers system and measurement procedure
9
The OT was built around an inverted microscope (Olympus IX71, Olympus, Japan) equipped
10
with a high numerical aperture oil immersion objective (model: UplanFl 100X N.A. = 1.35;
11
Olympus, Japan) and a 1292 x 964 pixel camera with a cell size of 3.75 x 3.75 µm (model:
12
StingRay F-125, Allied Vision) as described in.33 Figure 2A shows a schematic illustration of
13
the OT setup used in this work, which stands in a temperature controlled room with
14
computers and controllers isolated from the room to reduce noise and vibrations. A
15
continuous wave Nd:YVO4 laser (model: Millennia IR) operating at 1064 nm was used for
16
trapping a single bacterium or silica microsphere. The trap laser was merged with a probe
17
laser (low power HeNe-laser operating at 632.8 nm) using a polarizing beam splitter cube
18
(PBSC). The light from the probe laser was refracted by the trapped object, collected by the
19
condenser and imaged onto a 2D position sensitive detector (PSD, L20 SU9, Sitek Electro
20
Optics, Sweden). The light illuminating the PSD was converted to a photocurrent and
21
thereafter converted to a voltage that was sent to a programmable low pass filter (SR640,
22
Stanford research systems), before collected by a computer and processed with an in-house
23
LabVIEW program. To minimize the amount of noise in the setup and to optimize the
24
measured time series, we used the Allan variance method described in.34
25
The following steps were carried out to detect pili on a single bacterial cell. A video
26
sequence of bright field images were initially acquired to find the geometrical dimensions of 7 ACS Paragon Plus Environment
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
the bacterium. The image that provided the largest parameter values, i.e., the largest length of
2
the long and short axis, L and w, were chosen as illustrated in Figure 1. A representative
3
image of a cell with its dimensions marked with the green arrows is shown in Figure 2B.
4
These parameters were thereafter used to calculate the model diameter, which is described in
5
detail in the theory and experimental procedure section. Prior to a measurement we
6
determined the pixel to length conversion factor (1 pixel = 37.8 ± 0.2 nm, mean ± standard
7
error (SE) on the mean, and the standard deviation (SD) was 0.7 nm, n = 11).
8
The bacterium was thereafter trapped while the sample chamber was oscillated to
9
measure its effective diameter. In the optical trap, a non-spherical object will orient itself with
10
its long axis, L, along the optical axis, as illustrated in Figure 2C. To determine its effective
11
diameter with respect to its long axis the sample was oscillated at 32 Hz with an amplitude of
12
108 nm as previously described35. The sampling frequency was chosen as a multiple of 2 x
13
(where x is a positive integer) of the oscillation frequency. We used a sampling frequency of
14
131 072 Hz ( x = 17 ), and averaged 32 consecutive data sets that were acquired for 0.25 s
15
each, giving a total acquisition time of 8 s. To reduce the influence of the wall during the
16
measurement procedure, the measurement was performed at a height of 10 µm, since the wall
17
influence at this height will be less than 3% for a spherical particle with diameter of 1 µm.36
18
To also ensure a reliable calibration, we chose cells that had a straight body shape, i.e., cells
19
with an clear body curvature were discarded to avoid dissipation of energy due to rotational
20
movement of the cell in the trap.35 Two power spectra of trapped bacteria are shown in Fig.
21
S1.
8 ACS Paragon Plus Environment
Page 8 of 38
Page 9 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1 2
Figure 2
3
9 ACS Paragon Plus Environment
Langmuir
Page 10 of 38
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
3 Theory and experimental procedure
2
3.1 Drag forces on spherical and non-spherical objects
3
The proposed method assumes that the presence of pili increases the fluid drag on a piliated
4
cell in comparison to a bald cell with the equal cell membrane surface area. To calculate the
5
drag force acting on the membrane of an E. coli bacterium, we use a mathematical model with
6
input parameters assessed from a 2D-image of the cell. The cells used in this work have a
7
cylindrical shape with spherical caps, rather than a spherical or ellipsoidal shape, as quantified
8
in the supplementary materials section, Fig. S2-S3. Fitting both an ellipse and projection of a
9
cylinder with spherical caps to the segmented area of bacteria in SEM micrographs resulted in
10
an average of 89.84 % and 94.84 % agreement (n = 15), respectively. Therefore, we used an
11
empirical model that can handle irregular shapes, like a cylinder with spherical caps, with
12
good performance.37 For cells that have ellipsoidal shape, we suggest to use the analytical
13
drag force model described by Happel and Brenner.36
14
For a non-spherical object, e.g., a rod-shaped bacterial cell, the drag force at low Reynolds
15
number can be defined as described by Leith.38 The drag force on a non-spherical object
16
using this model is given by,
17
FD = 3πηVd Mod ,
18
where η is the dynamic viscosity of the surrounding liquid, V is the velocity of the sphere
19
relative to the fluid and d Mod is the model diameter defined by
20
d Mod =
21
where An is the projected area of the object, see figure 3A, and As is the surface area of the
22
object.
(1)
2 An 2 As + , 3 π 3 π
(2)
23
In the present work, we model cells as cylinders with spherical caps, as depicted in Figure
24
3B, implying that the projected area will alter depending on the orientation of the object. For a
10 ACS Paragon Plus Environment
Page 11 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
trapped cell moving through a fluid with a velocity perpendicular to its cylinder axis, as is
2
shown in Figure 1B and the rightmost cell in Figure 3A, the projected area and surface area
3
becomes An = π w2 4 + w ( L − w ) and As = π wL respectively, where L and w, the objects
4
dimensions, are defined in Figure 1A. The model diameter of the cell is thus given by,
5
d Mod =
6
For a spherical object, i.e., where w = L , Equation (3) reduces as expected to dMod = w and the
7
drag force models given by Equation (1) becomes identical to the standard Stokes’ drag force
8
model for a sphere.
2 3
w2 w 2 + ( L − w) + wL . 4 π 3
(3)
9 10
Figure 3
11
11 ACS Paragon Plus Environment
Langmuir
Page 12 of 38
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
3.2 Measuring the effective diameter using optical tweezers
2
To compare the model diameter defined in Equation (2) with the measured effective diameter
3
of a trapped object we use force-measuring optical tweezers. The effective diameter was
4
found by analyzing the power spectral density for the position data of a trapped object. The
5
power spectral density of the trapped object was assessed from the Langevin equation, which
6
describes the Brownian motion of a trapped object and is given by,
7
mx&& ( t ) + γ 0 x& ( t ) + kx ( t ) = (2 k BT γ 0 )1 2 ξ ( t ) ,
8
where m is the mass of the particle, γ 0 is the viscous drag coefficient (assuming a laminar
9
flow) and is given by γ 0 = 3πη d Eff , k is the trap stiffness, x is the position of the trapped
10
object relative to the center of the trap, kB is Boltzmann constant, T is the temperature, ξ ( t )
11
is the normalized white noise with a zero mean value, and dEff is the effective diameter of the
12
object. In an overdamped system, the inertia of a particle can be neglected, i.e., the first term
13
in Equation (4). The power spectral density is given by the square of the time-Fourier
14
transform divided by the measurement time as,
15
P( f ) =
16
where D is the diffusion constant defined by D = k BT γ 0 , f is the frequency, and f c is the
17
corner frequency given by fc = κ ( 2πγ 0 ) .39 Since the position of the trapped particle was
18
measured in Volt we can also define D = β 2 DV , where D V is the measured diffusion
19
constant and β is the conversion factor relating the position of the object in the sample plane
20
to the measured voltage signal. By fitting equation (5) to the power spectra density of the
21
position data, we determined D V . To find the conversion factor we introduced a small
22
oscillation of the sample while recording the position of the trapped object,35 yielding,
D
π
2
(
f 2 + fc2
)
(4)
,
(5)
12 ACS Paragon Plus Environment
Page 13 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
D
A2
δ ( f − f drive ) ,
1
P( f ) =
2
where A is the oscillation amplitude and fdrive is the oscillation frequency. The conversion
3
factor is given by,
4
β = Wth Wex .
5
where W th and Wex is the theoretical and experimental power spectrum amplitude,
6
respectively. The theoretical amplitude is given by the second term in Equation (6), including
7
the delta function. The experimental amplitude Wex is given by the height at the oscillation
8
frequency relative to the fit using Equation (5) divided by the measurement time for a single
9
power spectrum, measured at the oscillation frequency. Thus, we obtain a relation for the
10
effective diameter of the trapped object, see Figure 1C, in terms of the experimentally
11
measured parameters, T, η, β and DV given by,
12
d Eff =
(
π 2 f 2 + f c2
+
) 2 (1 + fc2
2 f drive
)
(6)
(7)
k BT . 3πηβ 2 DV
(8)
13
13 ACS Paragon Plus Environment
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
4 Results and discussion
2
4.1 Verification of the method using polymer coated silica
3
microspheres
4
The fundamental hypothesis in the present work is that surface organelles, even if they are
5
very small, affect the fluid velocity around an object. This change in the fluid velocity will
6
increase the drag force on the cell, which is possible to detect using OT instrumentation. To
7
verify this hypothesis and the theory described above, we performed well controlled
8
experiments with pure silica microspheres, and silica microspheres coated with a thin polymer
9
brush layer consisting of poly(3-sulfopropymethacrylate) (SPM).31,32 Polymer brushes are
10
polymer chains tethered to a surface by one of their chain ends. For these reference
11
microspheres the brushes were synthesized using surface initiated atom transfer radical
12
polymerization, which gives rise to dense brushes with well controlled polymer thickness.40,41
13
The SPM brush is highly hydrophilic with contact angles smaller than 10 degrees.42 In water
14
solution these types of highly charged brushes swell and become completely hydrated leading
15
to a thickness in water that is 2-7 times thicker than the dry thickness.41,43 However, the exact
16
swelling depends on several factors such as ionic strength and solvent quality. The exact
17
thickness of the swollen brush on the reference microspheres was not directly quantified,
18
however, it can be estimated from the thickness of a corresponding brush on a flat surface.
19
Polymerization on a curved surface, such as a microsphere, will be less affected by crowding,
20
implying that the brush thickness on the microsphere would be slightly thicker than on a
21
corresponding flat surface. For the SPM coated microspheres used in this study the
22
polymerization time was 120 minutes, which normally produces a dry thickness of 100 nm on
23
a flat silica surface. In pure water this brush layer is expected to swell to at least 200 nm.
24
The SPM brush is here assumed to mimic surface expressed organelles on a spherical cell
25
as illustrated in Figure 4A. Both types of microspheres were imaged dried using a scanning
26
electron microscope (SEM) to visualize their surface. A representative image of a pure silica 14 ACS Paragon Plus Environment
Page 14 of 38
Page 15 of 38
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
microsphere is shown in Figure 4B together with a SPM brush coated microsphere in Figure
2
4C. Individual polymer chains and their length cannot be resolved and measured with SEM
3
due to the resolution of the SEM and since the polymer brush collapses and compacts when
4
dry. However, the image shows that the coated microsphere has an evenly distributed SPM
5
coating. According to the manufacturer, the diameter of the pure microspheres were 2.0 µm,
6
however, the microsphere size variation was not provided. To measure the microsphere size
7
variation, we used both SEM micrographs and bright-field images. Microspheres in the SEM
8
images showed a high degree of clustering wherefore we manually measured the microsphere
9
size variation, which was found to be 2.0 % (n = 25). On the bright-field images, however, we
10
applied a circle detection algorithm and calculated the microsphere size including the size
11
variation.44 This approach is briefly explained in the caption of Figure S4. It was found that
12
the average size for pure microspheres were 1.989 ± 0.003 µm (n = 138, mean ± SE, SD =
13
0.038 µm).
14
We also quantified the size of the SPM coated microspheres using bright-field images to
15
1.997 ± 0.007 µm (n = 34, mean ± SE, SD = 0.039). A comparison of the size of the pure and
16
SPM coated microspheres showed no significant difference (p = 0.26). Thus, neither of the
17
imaging methods could detect the presence of surface polymers.
18
To determine if the SPM coating could be detected using the proposed OT method we
19
first measured the effective diameter of the pure silica microspheres as a reference. The mean
20
effective diameter was assessed to 1.96 µm ± 0.02 µm (n = 25, mean ± SE, SD = 0.08 µm), as
21
presented in Figure 4D. For an individual measurement, the propagated relative errors in the
22
effective diameter considering the error on temperature, viscosity and the two measured
23
parameters β and DV , were 0.03 %, 0.2 %, 7 %, 4 %, respectively.
24
Next, the effective diameter of the SPM coated silica microspheres was assessed. The
25
highly hydrophilic SPM brush, here in its hydrated form in MilliQ water, was expected to
26
create an additional fluid drag on the microspheres. This was confirmed by the OT method 15 ACS Paragon Plus Environment
Langmuir
Tuesday, February 23, 2016 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
which resulted in an effective diameter of 2.61 µm ± 0.04 µm (n = 25, mean ± SE, SD = 0.19),
2
see Figure 4D. The increase in the effective diameter is in-line with the expected thickness of
3
the SPM brush. To evaluate the performance for surface polymer detection, the ratio of the
4
effective diameter d Eff and the model diameter d Mod , where the latter is given by Equation
5
(3) with the mean value of the diameter assessed for pure microspheres 1.989 ± 0.003 µm,
6
was calculated. The propagating error for d Mod was estimated to 3.0 % based on the
7
segmentation and fitting procedure, see supplementary materials section Propagating error
8
when quantitating the bacterial shape. The ratio d Eff / d Mod is shown in Figure 4E using a
9
kernel density estimate. The kernel density was calculated using a normal distribution with
10
100 points.
11
For pure microspheres, the ratio dEff/dMod is expected to be one, whereas higher values
12
indicate the presence of surface coating. The mean of the ratio, for the pure and coated silica
13
microspheres were 0.980 ± 0.008 (mean ± SE, SD = 0.039) and 1.31 ± 0.02 (mean ± SE, SD =
14
0.096), respectively. A statistical test (t-test: p 1.5 is required, see Figure 6B.
15
Several studies have reported that absorption of light in optical traps can damaged trapped
16
cells.45–47 We took this into consideration when designing the assay since we wanted the cells
17
to be alive after trapping. The laser was therefore run at a low laser power in the sample
18
plane, 100 mW, and cells were trapped for a short time period, 8 seconds. This gives an
19
estimated energy dose of