Development and Application of a User-Friendly Automated HPLC

Jan 30, 2012 - Development and Application of a User-Friendly Automated HPLC. Dilution Macro for API Route Optimization. Jia Zang, Eric Sirota, Aaron ...
0 downloads 4 Views 497KB Size
Technical Note pubs.acs.org/OPRD

Development and Application of a User-Friendly Automated HPLC Dilution Macro for API Route Optimization Jia Zang, Eric Sirota, Aaron Moment, J. Michael Williams, David M. Tellers,* and Roy Helmy* Merck Research Laboratories, Merck and Co., Inc., Rahway, New Jersey 07065, United States ABSTRACT: Concentration determination by HPLC analysis is often utilized in pharmaceutical development activities. Traditionally, manual dilutions employing volumetric glassware have to be performed to obtain accurate concentration results. To circumvent this, an automated dilution method for the Agilent 1100 HPLC system has been developed to dilute and inject the samples online. After appropriate calibration, quantitative analysis can be achieved by placing the solution samples directly into HPLC vials without additional manipulation. This method is successfully applied to active pharmaceutical ingredient (API) solubility measurements and concentration determinations of routine mother liquor samples in the range of 0.1−100 mg/mL. It is particularly useful in crystallization development where obtaining accurate concentration information quickly drives down development time and material requirements.



INTRODUCTION Monitoring reaction progress and impurity formation and identification of crystallization conditions are essential activities in active pharmaceutical ingredient (API) route development and optimization.1,2 In recent years, these particular activities have been made easier by analytical HPLC, a now common tool in most industrial laboratories. In all cases, the act of preparing a sample for HPLC analysis necessitates sampling of the substrate from reaction mixture or crystallization and subsequent diluting with a large volume of an inert solvent. This dilution step is critical to ensure the sample concentration is within the linearity range of the analytical method. While effective, this process is laborious and solvent inefficient. Utilization of new technologies or modification of existing ones offers the opportunity to improve the efficiency of these key activities.3−6 Given our increasing emphasis on the use of green chemistries and technologies, we sought to develop a userfriendly technique for all synthetic chemists that would obviate the need for reaction sampling and dilution while at the same time reducing the overall amount of solvent used in the operation.7−10 We realized that such a tool could facilitate dilution-intensive workflows such as rapid screening of crystallization conditions and facile monitoring of reaction kinetics. We herein report on a macro-driven approach to automated sample dilution that can be implemented on existing HPLC equipment. We highlight two case studies demonstrating the successful utilization of this user-friendly method for identification of crystallization conditions and reaction monitoring.11,12

Figure 1. Recommended injector program.

an, toluene, ethyl acetate, methyl tert-butyl ether, and 99.99% phosphoric acid [85% (w/w) in H2O] were used. All water used was distilled and purified by a HYDRO System (Garfield, NJ, U.S.A.). Sample Preparation. A range of up to 100 mg/mL of the samples in different solvents for concentration determination were supplied by chemists in the department of Merck Process Chemistry and Merck Chemical Process Development and Commercialization. Please note that the analyses described in this paper were done at ambient temperature. Appropriate calibration is required if sampling is done with the HPLC needle at nonambient temperatures. Apparatus and Chromatographic Conditions. HPLC vials containing inserts were utilized for sample dilutions. A preassembled HPLC vial kit with a 2-mL HPLC vial, a 300-μL drop-in insert, and a cap with silicone/PTFE septa was purchased from Analytical Sales and Services, Inc. (Pompton



EXPERIMENTAL SECTION Reagents and Chemicals. Samples of compound A-I are experimental drugs which were supplied by Merck Process Chemistry (Merck Research Laboratories, Rahway, New Jersey, U.S.A.). HPLC grade acetonitrile, isopropyl acetate, dichloromethane, ethanol, N,N-dimethylformamide, N,N-dimethylacetamide, dimethyl sulfoxide, isopropanol, 2-methyltetrahydrofur© 2012 American Chemical Society

Received: October 14, 2011 Published: January 30, 2012 250

dx.doi.org/10.1021/op200289r | Org. ProcessRes. Dev. 2012, 16, 250−254

Organic Process Research & Development

Technical Note

Figure 2. (a) HPLC vial layouts for concentration determination. For example, the injector draws the diluent from vial 11. The standard/sample is drawn from vial 1. The diluent and the standard/sample solution are mixed in an insert-contained HPLC vial at position 21. Then the diluted standard/sample solution is injected for HPLC analysis. The same procedure is performed for the remaining standards/samples at the positions illustrated. (b) HPLC vial layouts for reaction monitoring or solution stability determination. For example, at the initial time point, the injector draws the diluent from vial 11. The sample 1 is drawn from vial 1. The diluent and the sample 1 solution are mixed in an insert-contained HPLC vial at position 21. Then the diluted sample 1 solution is injected for HPLC analysis. The diluted solutions of sample 2 and sample 3 are prepared and injected for HPLC analysis in the same way at positions 31 and 41, respectively, at the initial time point. The same procedure is performed for all the samples at different time points at the positions illustrated in the figure.

dilute the samples. The autosampler is programmed to draw 1.0 μL of the sample from vial 1. The offset position of the needle was set at 2.0 mm to avoid picking up the precipitates in the sample solution. The total volume of 90.0 μL of the mixture of the diluent and the sample solution was ejected into the vial at position 21. Then the injector drew and ejected the mixture solution in the same vial at position 21 with maximum speed for 15 times.14 In order to ensure adequate mixing for the low volume solution in position 21, an insert-contained HPLC vial was utilized. HPLC Vial Layouts. As the injector program (Figure 1) shows, sample dilution was achieved by programming the injector to draw the sample, draw the diluent and then mix the two solutions into another insert-contained HPLC vial. Before starting the run, all the HPLC vials should be placed in the appropriate positions. The vials are placed in different layouts for concentration determination and reaction monitoring. For concentration determination, one empty insert-contained HPLC vial is positioned for each standard and sample. Up to 47 samples can be analyzed. For reaction monitoring, one empty insert-contained HPLC vial is positioned for each time point of the reaction sample. Multiple reaction samples can be monitored with different diluents. The same vial layout is also used for solution stability determination. Examples are shown in Figure 2. Sequence Setup. A sequence file was created to perform automated sample dilutions and injections. The sequence was as follows: The first sample/standard was diluted into another HPLC vial with insert by injection program, followed by a blank injection to avoid the interference of the sample carried over from the sample dilution. The diluted sample/standard was then injected. The same procedure was performed for the remaining samples. For reaction monitoring and solution stability determination, the flow rate was reduced to 0.2 mL/

Plains, NJ, U.S.A.). Mini-UniPrep syringeless filters were used to remove precipitates from sample solutions being prepared for automated HPLC dilutions.13 The Mini-UniPrep filter was purchased from Whatman as a preassembled filtration device with a 0.4 mL capacity vial and a plunger containing a filtration membrane. The sample solution is filtered by pressing the plunger through the sample in the 2 mL outer HPLC vial and forcing the filtrate into the 0.4 mL inner vial. The chromatographic experiments were performed using an Agilent 1100 series HPLC system equipped with an autoinjector, temperature-controlled sample tray, quaternary pump, temperature-controlled column compartment, and diode array detector. The column employed was an Agilent Eclipse Plus C18 with the following column dimensions: 50 mm × 4.6 mm, 1.8 μm. The mobile phase consisted of deionized water with phosphoric acid added in at 0.1% (v/v) [A] and HPLC-grade acetonitrile [B]. The mobile phase flow rate was 1.5 mL/min with the column temperature set at 40 °C. UV detection was set to 220 nm. A blank injection was made before each sample injection. The mobile phase gradient for the blank injection was as follows: 90:10 A/B to 5:95 A/B in 0.1 min, followed by a 0.9 min hold (the total run time was 1 min with 1 min for reequilibration). For all the samples used for concentration or solution stability determination or for reaction monitoring, a linear mobile phase gradient was employed with an injection volume of 0.5 μL. The gradient was as follows: 90:10 A/B to 10:90 A/B in 5 min, followed by a 1 min hold (the total run time was 6 min with 2 min of re-equilibration). An injector program was carried out to dilute the sample for injection. Injector Program. The injector program that was created to perform automated sample dilutions on the HPLC system is shown in Figure 1. The injector drew 89.0 μL of diluent from vial 11. The user has flexibility to select different diluents to 251

dx.doi.org/10.1021/op200289r | Org. ProcessRes. Dev. 2012, 16, 250−254

Organic Process Research & Development

Technical Note

ments (section 3.1) and then provide two case studies applying this technique to key activities in API route development: crystallization solvent screening and stability monitoring (section 3.2). 3.1. Validation and Application of the Automated HPLC Dilution Method for Concentration Determination. Linearity of the automated HPLC dilution method over a range of concentrations was evaluated using four different compounds, A, B, C, and D (Figure 3). To construct the calibration curve for each compound, a series of standard solutions were prepared and then diluted by the automated HPLC dilution system for HPLC injections. A linear regression equation was obtained for the calibration curve of each compound. The linearity of the automated HPLC dilution method was verified with determination coefficients (R2) of >0.995 for all four compounds. A comparison of manual dilution vs HPLC automated dilution was carried out by examining the solubilities of two substrates (compounds A and B) in solvents across a range of solubilities to ascertain the dynamic range of this technique. The solubilities of compound A in isopropyl acetate, dichloromethane, ethanol, water, and N,N-dimethylacetamide were examined within a solubility range 0.5-g scale of substrates using the conditions identified in Figure 6. In each case, minimal manipulation allows the chemist to map a large solubility space in a quantitative manner. Case Study 2. Reaction Monitoring and Solution Stability Determination. Knowledge of substrate stability in solution informs route liabilities and key hold points for processing. This information is typically gained via periodic, manual sampling of reaction solutions. Application of the HPLC automated dilution system allows for controlled, systematic reaction sampling and dilution using a minimal amount of solvent in an HPLC vial. To illustrate this, we highlight the monitoring of two compounds H and I in solution at approximately 100 mg/mL (Figure 7) over the course of 10 h. A nine-point data sampling from the same vial over 10 h was collected, showing that the substrates are stable for at least 10 h. The HPLC automated dilution system can be programmed to monitor multiple samples across many solvents for stability determination with different diluents at user-selected time points, providing a quick survey of potential solvents. It should also be noted that this

Figure 7. Stability hold point case study. Percent relative standard deviation (%RSD) for both measurements