Fluorometric Sniff-Cam (Gas-Imaging System) Utilizing Alcohol

Jan 23, 2018 - The dynamic range of the sniff-cam was 0.1–10 ppm which encompassed breath AcH concentrations after drinking. Finally, the sniff-cam ...
0 downloads 0 Views 3MB Size
Subscriber access provided by READING UNIV

Article

Fluorometric sniff-cam (gas-imaging system) utilizing alcohol dehydrogenase for imaging concentration distribution of acetaldehyde in breath and transdermal vapor after drinking Kenta Iitani, Toshiyuki Sato, Munire Naisierding, Yuuki Hayakawa, Koji Toma, Takahiro Arakawa, and Kohji Mitsubayashi Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.7b04474 • Publication Date (Web): 23 Jan 2018 Downloaded from http://pubs.acs.org on January 24, 2018

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

1

Fluorometric sniff-cam (gas-imaging system) utilizing alcohol dehydrogenase for imaging

2

concentration distribution of acetaldehyde in breath and transdermal vapor after drinking

3 4

Kenta Iitania, Toshiyuki Satoa, Munire Naisierdinga, Yuuki Hayakawaa, Koji Tomab, Takahiro

5

Arakawab and Kohji Mitsubayashia,b,*

6 7

a

8

5-45 Yushima, Bunkyo-ku, Tokyo 113-8510, Japan

9

b

Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, 1-

Department of Biomedical Devices and Instrumentation, Institute of Biomaterials and

10

Bioengineering, Tokyo Medical and Dental University,2-3-10 Kanda-Surugadai, Chiyoda-ku,

11

Tokyo 101-0062, Japan

12 13

* Corresponding author. Tel.: +81 3 5280 8091, Fax: +81 3 5280 8094

14

E-mail: [email protected]

15 16 17

1 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 2 of 30

18

Abstract

19

Understanding concentration distributions, release sites and release dynamics of VOCs from

20

the human is expected to lead to methods for noninvasive disease screening and assessment of

21

metabolisms. In this study, we developed a visualization system (sniff-cam) that enabled to

22

identify a spatiotemporal change of gaseous acetaldehyde (AcH) in real-time. AcH sniff-cam

23

was composed of a camera, a UV-LED array sheet, and an alcohol dehydrogenase (ADH)-

24

immobilized mesh. A reverse reaction of ADH was employed for detection of gaseous AcH

25

where a relationship between fluorescence intensity from nicotinamide adenine dinucleotide

26

and the concentration of AcH was inverse proportion; thus, the concentration distribution of

27

AcH was measured by detecting the fluorescence decrease. Moreover, image differentiation

28

method that calculated a fluorescence change rate was employed to visualize a real-time

29

change in the concentration distribution of AcH. A dynamic range of the sniff-cam was 0.1–10

30

ppm that encompassed breath AcH concentrations after drinking. Finally, the sniff-cam

31

achieved to visualize concentration distribution of AcH in breath and skin gas. A clear

32

difference of breath AcH concentration was observed between aldehyde dehydrogenase type 2

33

active and inactive subjects, which was attributed to metabolic capacities of AcH. AcH in skin

34

gas showed a similar time course of AcH concentration to the breath, and a variety of release

35

concentration distribution. Using different NADH-dependent dehydrogenases in the sniff-cam

36

could lead to a versatile method for noninvasive disease screening by acquiring

37

spatiotemporal information of various VOCs in breath or skin gas.

38 human volatilome,

39

Keywords:

40

dehydrogenase, image differential

fluorescence, spatiotemporal, acetaldehyde, alcohol

41 42

Introduction

2 ACS Paragon Plus Environment

Page 3 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

43

In the conventional and current diagnosis of diseases, the blood test has usually been 1,2

. At the same time, there has been growing interest in non-invasive

44

commonly utilized

45

diagnosis methods aiming at exhaled volatilome such as volatile organic compounds (VOCs)

46

in breath or skin gas as the alternative or auxiliary in next-generation medicines

47

VOCs are originally contained in blood as a result of specific diseases and metabolism, and

48

then they are emanated directly from skin or breath after the exchange in the lungs 7,8.

3–6

. These

49

Acetaldehyde (AcH), an intermediate of alcohol metabolism, is produced by ADH-

50

mediated reaction in the liver after ingesting alcohol 9, and it has toxicity to DNA 10,11. For

51

example, it was reported that AcH causes the mutational DNA damage in hematopoietic stem

52

cells. 12 In this study, we used AcH as a model analyte since the contained in blood is released

53

through breathing with a partition coefficient of breath : blood = 1 : 109

54

Skin gas is constantly released and can be easily collected without a mental and physical

55

burden; therefore, skin gas is considered to be more suitable for monitoring of VOCs than

56

exhaled breath. However, measurement of VOCs in skin gas is challenging since their

57

concentrations are usually much lower than the breath, e.g., ppb or sub-ppb level 15; a highly

58

sensitive system is strongly required for the detection. Conventionally, skin gas sample is

59

collected by a hermetically container such as a sample bag and measured by analytical

60

devices such as a gas chromatograph (GC) and a mass spectrometer (MS). For example,

61

Tsuda et al. and Turner et al. measured ethanol, acetone and etc. in skin gas that were released

62

from a forearm, hand, and finger and collected by a sample bag

63

ionization detector or selected ion flow tube MS. Mochalski et al. developed a room-type skin

64

gas collection system which a subject was in, and collected only emanated skin gas by

65

discharging exhaled breath out of the room; the VOCs contained in skin gas was measured

66

using a proton transfer reaction MS

67

drawbacks including complicated processes of measurement and requirement of highly skilled

16,17

13

and skin gas

14

.

using GC and flame

18

. Despite the advantages of these methods, some

3 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 4 of 30

68

personnel made them unsuitable for simple disease screening and metabolism evaluation.

69

Besides, it is difficult to investigate release sites and dynamics in detail because these

70

methods measured the total amount of released VOCs and are unable to focus on a spot. It is

71

known that VOCs’ concentration in exhaled breath and skin gas shows temporal fluctuation

72

19–21

73

important to deepen the understanding of a correlation between skin gas and physiological

74

conditions: (i) Direct emanation from blood through the dermis and the epidermis, (ii)

75

vaporization of sweat secreted from glands in a skin and (iii) production by metabolism of

76

resident bacteria on a skin surface 22–25.

, and elucidating following potential release pathways for each VOC is considered to be

77

Up to now, we have introduced a fiber optic gas sensor “bio-sniffer” that showed high

78

sensitivity and selectivity due to the substrate specificity of an enzyme, and it allowed

79

continuous measurement of VOCs 26–30. Also, applying the bio-sniffer’s technique to imaging,

80

a biofluorometric gas imaging system (sniff-cam) was developed

81

AcH sniff-cam employing a reverse reaction of ADH

82

accurately display a temporal change of AcH concentration by measuring the fluorescence

83

intensity of β-nicotinamide adenine dinucleotide (NADH) because the fluorescence

84

reached a plateau after initial decrease and did not return to a baseline although the gas

85

injection was stopped. This was attributed to measurement principle in which consumed

86

NADH has not filled up again throughout the measurement. In this study, we further advanced

87

the AcH sniff-cam to improve the responses and to enable visualize a spatiotemporal change

88

of AcH concentration in real-time. The response was improved by employing an image

89

differential method that utilized a slope of fluorescence intensity change instead. Since the

90

slope kept changing while the fluorescence intensity increased and approached plateau, the

91

slope output showed a clear peak and returned to a baseline when the gas injection was

31

. Recently, we reported

32

. However, it was difficult to

4 ACS Paragon Plus Environment

Page 5 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

92

stopped. Afterwards, the advanced AcH sniff-cam was applied to real-time visualization of

93

spatiotemporal change of AcH contained in exhaled breath and skin gas after drinking.

94 95

Experimental

96

Materials and reagents

97

Tris-HCl buffer solution (Tris, pH6.5, 0.1 M) that was used in all of the

98

experiment was made by 2-amino-2-hydroxymethyl-1,3-propanediol (99.9%, Cat. No.

99

013-16385, Wako), hydrochloric acid (35%, Cat. No. 083-03485, Wako) and ultra-pure

100

water that was generated by a Mill-Q purification system. Reagents for fabrication of

101

an ADH-immobilized mesh were ADH (EC 1.1.1.1, 369 units/mg solid, Cat. No.

102

A7011, from Saccharomyces cerevisiae, Sigma-Aldrich, USA), phosphate buffer (PB,

103

pH6.5, 0.1 M), bovine serum albumin (BSA, Cat. No. 306-13383, Wako, Japan) and

104

glutaraldehyde (GA, 25%, Cat. No. 079-00533, Wako Pure Chemical Wako, Japan). A

105

cotton mesh (100% cotton, the thickness of 1 mm, the interval of 1 mm) that was used

106

as a substrate for enzyme immobilization was purchased from Ohki Healthcare

107

Holdings (Japan). Reduced form of β-nicotinamide adenine dinucleotide (NADH, Cat.

108

No. 44326000) was purchased from Oriental Yeast (Japan). Standard gaseous AcH was

109

generated using a permeation tube (Cat. No. P-92-1, Gastec, Japan) with a standard gas

110

generator (Cat. No. PD-1B-2, Gastec, Japan). For selectivity assessment, some VOCs

111

including ethanol (99.5 %, Cat. No. 14033-00. Wako), methanol (99.8%, Cat. No. 21926-95,

112

nacalai tesque), 2-propanol (99.7, Cat. No. 166-04836, Wako), acetone (99.5%, Cat. No.

113

016-00346, Wako) were generated by a diffusion tube (No.3200, Gastec, Japan), and the other

114

such as methyl mercaptan that was done by a permeation tube (Cat. No. P-71-5, Gastec,

115

Japan). An alcohol beverage (concentration of 20%, distilled liquor) for breath and skin gas

116

experiments was purchased at a local store.

5 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 6 of 30

117 118

Construction of the AcH sniff-cam

119

Figure 1a shows a visualization principle of gaseous AcH distribution employed

120

in an AcH sniff-cam. AcH was detected as a fluorescence decrease from NADH

121

consumption which occurs through an ADH-mediated reverse reaction. In our previous

122

work, we investigated the reduction/oxidation reaction of ADH can be controlled by

123

pH condition. Also, unwanted interfere between forward and reverse reaction could not

124

occur under optimal conditions.33 NADH has an autofluorescence property that

125

exhibits absorption of excitation light at 340 nm wavelength and emission of

126

fluorescence at 490 nm wavelength. Since NADH and AcH concentrations are

127

correlated, AcH concentration can be determined by detecting fluorescence decrease of

128

NADH. The AcH sniff-cam was composed of a UV-LED array (λ of 340 nm, 9×9

129

array, 340X081SFN, Dowa, Japan), an ADH-immobilized mesh and a high-sensitive

130

camera (HEED-HARP, Pioneer, Japan) (Figure 1b). The UV-LED array and the

131

camera were aligned in the same optical, and 5 mm and 110 mm apart from the ADH-

132

immobilized mesh, respectively. The UV-LED array was connected to a power source

133

(KX-100H, Takasago, Japan) that supplied constant current 540 mA to the LED, and

134

enabled to excite NADH uniformly in the ADH-immobilized mesh. The gain

135

parameter of the high-sensitive camera was fixed at 15 in all of the experiments, and

136

video signal that was conforming to National Television System Committee (NTSC)

137

was out from composite connector of a camera controller. An analog video signal was

138

converted to a digital video format by a video encoder (Intensity shuttle USB3.0, Black

139

Magic Design, Australia). Two bandpass filters placed in front of the UV-LED array

140

(BPFex, λ of 340 ± 42.5 nm, Edmund Optics, USA) and the high-sensitive camera

141

(BPFfl, λ of 490 ± 10 nm, Asahi Spectra, Japan) excluded unwanted light to improve a 6 ACS Paragon Plus Environment

Page 7 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

142

signal to noise ratio. The ADH-immobilized mesh was fabricated by the following

143

process. First, 0.65 mg of ADH and 1.5 mg of BSA were dissolved in PB (200 µL, 0.1

144

M at pH 6.5). Then, 200 µL of the mixture was dropped and spread uniformly on a 20

145

× 20 mm2 cotton mesh substrate, and then the mesh was left in a refrigerator for 60

146

minutes. Subsequently, 10-fold diluted GA solution (32 µL, 2.5 v/v% in PB, 0.1 M at

147

pH 7.0) was dropped onto the mesh for cross-linking and left in the refrigerator for 90

148

minutes. Finally, the ADH-immobilized mesh was rinsed with 300 µL of Tris and then

149

used immediately. Standard gaseous AcH was generated by a standard gas generator

150

with a permeation tube. Clean carrier air that was compressed and filtered by a

151

compressor and air filters was supplied to the standard gas generator through

152

polytetrafluoroethylene tubes (F-8006-017, the outer diameter of 6 mm, the inner

153

diameter of 4 mm, FLON industry, Japan) with excellent chemical and adsorption

154

resistance. Flow rates of generated standard gaseous AcH and clean carrier air were

155

controlled at 40 mL/min by mass flow controllers (MFC, RK1250, KOFLOC, Japan),

156

and gas was introduced to the AcH-immobilized mesh. A glass gas outlet was placed

157

on a backside of the ADH-immobilized mesh. Incidentally, it is reported that the

158

environmental temperature affects enzyme activity and fluorescence intensity of

159

NADH34. Thus, environmental temperature was kept at 25 ± 1.5 ºC of all experiment

160

using an air conditioner. Thus, the temperature did not affect the enzyme activity and

161

fluorescence intensity of NADH.

162 163

(Figure 1 comes here)

164 165

Differential method for visualization of spatiotemporal changes of AcH

7 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 8 of 30

166

In the previous report, visualization of gaseous AcH spatial distribution was achieved

167

using AcH sniff-cam that employed a reverse reaction of ADH. However, there was a problem

168

in temporal resolution. After a decrease of fluorescence intensity on the ADH-immobilized

169

mesh, the intensity reached a plateau and did not recover to the initial value even though

170

injection of AcH was stopped. This was because there was no NADH supply once it was

171

consumed. This problem made it difficult for previous AcH sniff-cam to be used for the study

172

of gaseous AcH release dynamics. In order to solve this problem, an image differential

173

method that utilized slope of fluorescence intensity change instead of the intensity itself was

174

adopted. Application of the differential method to a video taken by the AcH sniff-cam was as

175

follows. First, an original video that showed fluorescence decrease as a result of NADH

176

consumption was converted to one in which florescence increased by subtracting a subsequent

177

frame from the first frame of the video. This process is described by the following equation:  =  −  ( > 0),

Eq.1

178

wherei i is a frame index number; FI and FD are fluorescence increasing and decreasing

179

images, respectively. Next, a simple moving average (SMA) filter described in equation 2 was

180

used to remove high-frequency noise from FI that was caused by an image sensor of the

181

camera. 1  = 





()

 ( ≥ ),

Eq.2

182

where n is the number of the frames in the SMA filter; FIk is a kth frame in the fluorescence

183

increasing video; SFI is smoothed fluorescence increasing images. Then, a differential video

184

that was composed of slopes of fluorescence intensity changes was created by differential

185

method described in equation 3.  =

 − (×∆)) ∆ = ( ∆ = 2, !"# = 30, > 60), ∆ ∆

Eq.3

8 ACS Paragon Plus Environment

Page 9 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

186

where i is a frame index number; fps is frame per s that was of 30; ∆t is a time step that was of

187

2 sec; SFI is a smoothed fluorescence increasing image; DI is a differential image.

188

Finally, high-frequency noise remaining in the DI was removed by the SMA filter. The

189

SMA filter had a trade-off relationship between noise reduction performance and response

190

time. When the n in the SMA filter increased, noises were decreasing but response time

191

became longer. Thus, an influence of the n to noise and response time was investigated with

192

various conditions (n = 0, 10, 30, 100, 300, and 600 frames) of the SMA filter. In the

193

experiment, 10 ppm of gaseous AcH was applied to the ADH-immobilized mesh wetted by

194

NADH solution (500 µM, in 300 µL of Tris at pH 6.5, 0.1 M). After creating DI, that was

195

smoothed by various conditions of SMA filters. Then, noise and response time at each

196

condition were evaluated.

197

In order to confirm the applicability of the AcH sniff-cam to the measurement of

198

breath and skin gas, sensitivity and selectivity of the AcH sniff-cam were investigated. Each

199

experiment was performed using 500 µM NADH dissolved in 0.1 M Tris at pH 6.5. For

200

evaluation of the sensitivity of the AcH sniff-cam, various concentrations of gaseous AcH

201

generated by the standard gas generator were applied to the ADH-immobilized mesh at a flow

202

rate of 40 mL/min. In the experiment, initially clean carrier air was applied for 20 s, and then

203

gaseous AcH was applied for 20 s. Finally, clean carrier air was applied again for 80 s. For

204

investigation of selectivity, typical VOCs produced by the standard gas generator were

205

applied to the AcH sniff-cam in the same procedure with the above-mentioned sensitivity

206

study, and concentrations of the gas samples were set to be those after drinking: AcH (5 ppm),

207

ethanol (100 ppm), methanol (0.1 ppm), 2-propanol (0.1 ppm), acetone (0.6 ppm), methyl

208

mercaptan (0.007 ppm), and mixture of AcH (5 ppm) and ethanol (100 ppm) were used 35,36.

9 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 10 of 30

209

Note that all of the numerical data of mean intensity and differential value were

210

calculated at the region of interest (ROI, 80 pixels × 80 pixels) that was set at the center of

211

frames in each video.

212 213

Visualization of spatiotemporal change of the AcH concentration in breath

214

Visualization of AcH in exhaled breath after drinking was taken place with subjects

215

with a different phenotype of aldehyde dehydrogenase type-2 (ALDH2). A subject with lower

216

activity of ALDH2 (ALDH2[-]) cannot catalyze AcH efficiently. In general, blood AcH

217

concentration of ALDH2[-] subject becomes higher after drinking than that of the subject with

218

high activity of ALDH2 (ALDH2[+]), thus breath AcH concentration is also expected to be

219

higher than the ALDH2[+] 37–39. These breath and skin gas experiments were conducted based

220

on the authorization of the Human Investigations Committee of Institute of Biomaterials and

221

Bioengineering, Tokyo Medical and Dental University (authorization code of 2012-6) that

222

acted up to Declaration of Helsinki. The subjects had been prohibited to smoke, take medicine

223

and alcohols since 72 hours before the experiment. The most important of the experiment with

224

a drinking was unified concentration of blood ethanol in all subject. Concentration of ethanol

225

in blood fluctuates by volume of blood that almost correlates to body weight. Therefore, each

226

subject who had been fasting for 4 hours administrated alcoholic beverage with an ethanol

227

concentration of 0.4 g/kg body weight within 15 minutes by refer to previous research.

228

the experiment, exhaled breath was directory applied to the ADH-immobilized mesh that was

229

soaked with 500 µM NADH solution (in 300 µL, 0.1 M of Tris at pH 6.5) through a breath

230

flow controller 41 that made a flow rate of the breath constant (40 mL/min). The measurement

231

was taken place at -15, 0, 15, 30, 45, 60, 90, 120, and 180 minutes after drinking. In order to

232

validate results by the AcH sniff-cam, the identical breath sample collected in a sample bag

233

(Cat. No. 2-0081-03, As One, Japan) was measured simultaneously by a gas detection tube as

40

In

10 ACS Paragon Plus Environment

Page 11 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

234

a standard comparison. The concentration of breath AcH measured by the AcH sniff-cam was

235

calculated using a calibration curve, and its spatial distribution was visualized using the image

236

differential method.

237 238

Visualization of distribution of the AcH released from palm skin

239

For skin gas visualization, a palm was chosen due to convenience for the

240

measurement. Taking an average size of palm into consideration, a larger (90 × 90 mm2)

241

ADH-immobilized mesh than that used in breath AcH visualization was prepared. Preparation

242

of this larger mesh was the same as the one for the breath experiment except for a point where

243

a total amount of chemicals was increased proportionally to the size. Briefly, a mixture of

244

13.17 mg of ADH, 20.0 mg of BSA and 1800 µL of PB was dropped and spread uniformly

245

onto the cotton mesh, then left for 60 minutes in a refrigerator. Afterwards, ADH was cross-

246

linked by 648 µL of GA in the refrigerator for 90 minutes. The mesh was then rinsed by 5000

247

µL of Tris at pH 6.5 and wetted by 500 µM NADH solution (in 0.1 M of Tris at pH 6.5, 6075

248

µL), followed by being placed in front of the camera. Figure 1d shows a skin gas injector with

249

a thickness of 3 mm was composed of a polymethylmethacrylate (PMMA) frame and a

250

honeycomb spacer for adequate imaging of spatiotemporal change of skin gas. Because it was

251

used to keep the constant distance between the ADH-immobilized mesh and the palm skin

252

contour surface. Imaging of transdermal AcH after drinking was performed by the following

253

steps. First, a background image without skin gas was recorded for 60 s, and then skin gas

254

was applied to the ADH-immobilized mesh by attaching a subject’s palm on the spacer for 20

255

s. Afterwards, resultant fluorescence on the mesh was recorded by the camera for 60 s. An

256

image of transdermal AcH distribution was made by subtracting the background image from

257

the skin gas. The measurement was taken place at -15, 30, 60, and 90 minutes after drinking,

11 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 12 of 30

258

and alcohol administration conditions were same with the above-mentioned breath

259

experiment.

260 261

Results and discussion

262

Responsivity, quantitativity, and selectivity of the AcH sniff-cam

263

Figure S-1a and S-1b show time courses of the mean intensity of the ROI and

264

differential value. An SMA filter was applied to the intensity of the ROI at different frame

265

numbers (n = 0, 4, 8, 16 and 32), and the differential values were obtained from the filtered

266

intensity. The result indicated that the response time was almost the same among the frame

267

number of 0–32. The relationship between noise and the number of frames in the SMA filter

268

is shown in Figure S-1c. The standard deviation of the baseline (time from -20 to 0 s) was

269

defined as a noise. The noises of both mean intensities of the ROI and differential value

270

decreased as increasing the number of frames. However, the performance of SMA’s high-

271

frequency noise reduction was saturated over 16 frames; thus, we used 16 frames for the SMA

272

filter to the mean intensity of the ROI in subsequent experiments. We further investigate the

273

influence of the frame number of the SMA filter on the differential value. As Figure S-2a

274

shows, the time course of the differential values becomes smoother as increasing the frame

275

numbers. Figure S-2b summarizes response time and noise as a function of the number of

276

frames. When 300 frames were used in the SMA filter, significantly small background noise

277

was obtained as well as a good response time; therefore, we decided to use 300 frames in the

278

SMA filter that was applied to the differential value.

279

Figure 2a shows fluorescence and differential images 0, 20 and 120 s after applying 10

280

ppm standard gaseous AcH to the ADH-immobilized mesh. Originally the fluorescence image

281

at 0 second was uniformly bright image due to the presence of NADH on the mesh. The

282

images in Figure 2a are the ones converted by equation 1. As a result of image differential

12 ACS Paragon Plus Environment

Page 13 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

283

method, observation of both spatial and temporal change of AcH concentration on the ADH-

284

immobilized mesh surface was achieved (supplemental movie 1). Figure 2b shows temporal

285

changes of mean intensity of the ROI and differential value. The ROI intensity has reached a

286

plateau after the initial increase and did not recover to the baseline even after AcH application

287

was stopped. In contrast, the differential value exhibited a clear peak 20 s after applying AcH

288

and returned to the baseline immediately after stopping AcH gas. Mean intensity of the ROI

289

between 90–100 s after applying the AcH was defined as a signal output (∆I); the peak of the

290

differential value was defined as a signal output of the differential value (∆D). The response

291

time of mean intensity of the ROI T90 was defined as a time which took for signal to reach

292

90% of ∆I; the response time of differential value TR was the time to reach peak maximum of

293

∆D. It was revealed that the response time was improved by 15 s (T90 = 35 s and TR = 20 s for

294

10 ppm AcH) using the image differential method. Changing of fluorescence intensity

295

continued for a while even after stopping AcH gas since a little amount of AcH remained on

296

the ADH-immobilized mesh. In contrast, a reaction speed was decreasing immediately

297

because that of ADH was sensitive to the concentration of AcH. The differential value

298

indicated the reaction speed of ADH reverse reaction. Therefore, the image differential

299

method enabled to improve response time by accurately detect changing of concentration on

300

the ADH-immobilized mesh. Note that a speed of enzymatic reaction is proportional to

301

substrate concentration only in the first-order reaction region. Thus, an improvement of

302

reaction time may not be occured for high concentration of AcH.. However, this need not be

303

taken into account in our experiment because AcH concentration in breath and skin gas are

304

lower than 10 ppm.

305 306

(Figure 2 comes here)

307 13 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 14 of 30

308

Figure 3a shows temporal changes of a differential value to various concentrations of

309

standard gaseous AcH, a calibration curve and differential images that were obtained 20 s

310

after applying 1, 5 and 10 ppm AcH. Each plot in the calibration curve was taken from the

311

peak of each differential value. As a result, the AcH sniff-cam showed a wide dynamic range

312

(0.1–10 ppm) that encompasses a concentration range of breath AcH after drinking (0.4–9.0

313

ppm). The ADH-immobilized mesh was designed for disposable by concerning sanitary of the

314

system because the sniff-cam supposed to use in medical fields. Thus, note that each plot of

315

the calibration curve in figure 3a was measured by a different ADH-immobilized mesh that

316

was prepared at the same time. The coefficient of variation of ∆D among three deferent ADH-

317

immobilized mesh was 1.5% as shown in figure S-3. Plots were fitted by the following

318

equation with a correlation coefficient of 0.997:

∆ = 0.02 × [()* (""+)]- + 0.07 × [()*(""+)] + 0.08

Eq.4

319

Figure 3b displays relative outputs ∆D to investigate selectivity of the AcH sniff-cam

320

to gaseous AcH. The relative output was obtained by normalizing the output for each VOC by

321

that for 5 ppm AcH. It was observed that the AcH sniff-cam showed output signals only from

322

samples containing AcH, which validated a high selectivity of the AcH sniff-cam due to a

323

high substrate specificity of the enzyme.

324

(Figure 3 comes here)

325 326

Visualization of breath AcH

327

Figure 4a shows differential images of 0, 20 and 120 s after applying breath samples

328

from ALDH2[+] and ALDH2[-] subjects 30 minutes after drinking. The spatiotemporal

329

change of breath AcH could be visualized similarly to standard gaseous AcH. Figure 4b

330

displays time course of those differential values. Both peaked at about the same timing and

331

returned to the baseline, and breath AcH concentration of the ALDH2[-] subject was about 3-

14 ACS Paragon Plus Environment

Page 15 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

332

fold higher (8.4 ppm) than that of the ALDH2[+] (2.8 ppm). Stagnation of recovering to initial

333

value in differentiation value at around 10 s after differentiation value reaching peak was

334

observed unlike a result of standard gaseous AcH that was shown in Figure 2. This

335

phenomenon seems originated of humidity of exhaled breath. When the breath was loaded in

336

PTFE tube, the breath was cooled immediately, and then amount of saturated water vapor was

337

decreased, finally condensation was caused on wall of the PTFE tube. After that when

338

applying the breath was stopped and starting the applying dry carrier air, AcH remaining in

339

the PTFE tube as condensation was applying to ADH-immobilized mesh again. Therefore, the

340

stagnation of decrease of differentiation value was occurred. The differential analysis was

341

very sensitive to concentration change of AcH on ADH-immobilized mesh surface. As a

342

result, unexpected influence of humidity in breath was realized. Note that this little disorder

343

on response caused by the humidity of breath did not influenced quantifying the AcH in the

344

breath as described in following discussion.

345 346

(Figure 4 comes here)

347 348

Time courses of breath AcH concentrations for ALDH2[+] and ALDH2[-] are shown

349

in Figure 5. In the graphs, vertical bar graphs represent values of the standard comparison by

350

AcH detector tubes. The AcH sniff-cam revealed that AcH concentration in exhaled breath

351

initially increased and peaked 30 minutes after drinking, and then gradually decreased with

352

time. It was reported that the concentration of AcH in blood showed a maximum 15–30

353

minutes after drinking

354

concentration was related to ethanol metabolism, and it reflected AcH blood concentration.

355

Obtained peak breath AcH concentrations for ALDH2[+] and ALDH2[-] subjects were 2.75 ±

356

0.39 ppm and 8.64 ± 0.32 ppm, respectively; these values were in good agreement with

42

. Therefore, it was considered that the change of breath AcH

15 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 16 of 30

13,43

. Also, AcH

357

reported results (ALDH2[+], 1.2 ± 1.0 ppm; ALDH2[-], 6.9 ± 2.5 ppm)

358

concentrations measured by detector tubes showed similar values and the same trend to the

359

AcH sniff-cam. These results and evidence demonstrated the sniff-cam’s capability of

360

accurate and quantitative visualization of the spatiotemporal distribution of breath AcH.

361 362

(Figure 5 comes here)

363 364

Distribution of AcH released from a palm skin

365

Figure 6a shows top and 30-degree tilted views of transdermal AcH distribution

366

released from a palm skin. Fluorescence intensity changed when applying skin gas, and this

367

image was created by merging pictures of a hand and a fluorescence distribution profile. From

368

the image, higher output was observed at around the metacarpal phalangeal joint of a second

369

finger and the thenar eminence than the other parts probably due to different emanation rate

370

over a palm skin as observed in our previous results of skin emanated ethanol that was

371

obtained by chemiluminescent sniff-cam

372

visualizing a concentration distribution of ethanol emanated from palm skin 31. More clearly

373

distribution image compared to previous one was obtained in this study. The difference was

374

seemingly made by an improvement of skin gas injector shown in figure 1d. Previously, skin

375

gas was applied to an enzyme immobilized mesh only using PMMA frame that had a 90 × 90

376

mm2 square hole. There was a possibility of touching an enzyme immobilized mesh when the

377

skin gas injector with a square hole was used; thus, skin surface had to fix in the air. Such a

378

situation, dilution, and agitation of skin gas could occur easily. In contrast, a honeycomb

379

spacer that allows fitting with skin surface at all position certainly without the possibility of

380

touching the ADH-immobilized mesh was newly added in this study. Thus, skin gas could be

381

applied to ADH-immobilized mesh directly without dilution and agitation. As the result, clear

44

. Fluorometric ethanol sniff-cam also achieved

16 ACS Paragon Plus Environment

Page 17 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

382

distribution image might have been obtained because of skin gas that reaching to the ADH-

383

immobilized mesh more increased than the previous system. However, the current AcH sniff-

384

cam cannot measure skin gas in real-time because a hand blocked the excitation light from the

385

UV-LED array when subject applied skin gas to the ADH-immobilized mesh. Therefore, the

386

image differential method could not be applied to the AcH sniff-cam. Figure 6b shows a time

387

course of mean intensity difference of the ROI ∆I from the 15 minutes before drinking. The

388

temporal change of ∆I showed a similar trend to the breath concentration. The concentration

389

of skin gas AcH was calculated by calibration curve based on difference of fluorescence

390

intensity between before and after exposure of AcH ∆I

391

exhaled breath. The results suggest that the measurement conditions of the calibration curve

392

were not fit to the measurement conditions of the skin gas. For example, standard gaseous

393

AcH has the flow-rate of 40 mL/min although flow-rate of skin gas was not constant.

32

showed higher than the one in the

394

Besides, parameters for formation of concentration distribution of AcH from palm skin

395

are still unclear. Perspiration is closely related to the release pathway of skin gas, and it is

396

reported that there is much number of sweat gland on palm skin than the other part of the

397

body.

398

occurs by measuring several biological information such as sweat rate, skin temperature and

399

concentration distribution of skin gas, simultaneously in the future.

45

Therefore, we expect to investigate mechanism in which concentration distribution

400 401

(Figure 6 comes here)

402 403

Conclusions

404

In this research, we developed the AcH sniff-cam using a reverse reaction of ADH and

405

an image differential method that allowed spatiotemporal visualization of gaseous AcH

406

concentration. After the system construction, the image differential method that calculated 17 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 18 of 30

407

reaction speed of the reverse reaction was developed in order to improve temporal response of

408

the system. Characterization of the AcH sniff-cam revealed that the AcH sniff-cam had a

409

suitable dynamic range (100–10000 ppb) and selectivity for measurement of AcH in breath

410

and skin gas. Also, the image differential method allowed to improve temporal response by 15

411

s. Finally, spatiotemporal changes of AcH concentration in exhaled breath and distribution of

412

AcH in skin gas were visualized by the AcH sniff-cam. Breath AcH concentration peaked 30

413

minutes after drinking, and it was 3-fold higher for ALDH2[-] subject than that for the

414

ALDH2[+]. The difference in AcH release rate was also clearly visualized over a palm, which

415

also peaked 30 minutes after drinking. These results demonstrated that the sniff-cam holds a

416

huge potential for deployment in simple and noninvasive metabolism evaluation and disease

417

screening by acquiring spatiotemporal information of various VOCs in breath and skin gas.

418 419

Acknowledgement

420

This work was supported by the JSPS KAKENHI Grant Numbers JP17H01759,

421

JP16J09604, and JP15H04013, the Japan Science and Technology Agency (JST), and the

422

Ministry of Education, Culture, Sports, Science and Technology (MEXT) Special Funds for

423

Education and Research “Advanced Research Program in Neo-Biology”.

424 425

Supporting Information. Brief statement in nonsentence format listing the contents of the

426

material supplied as Supporting Information.

427

171208-AcH-vis-breath-skingas.docx Graphs about optimization of image analysis

428

parameter, reproducibility of the ADH-immobilized mesh

429

supplemental movie 1.avi Video of spatiotemporal change of AcH on the ADH-immobilized

430

mesh

431

18 ACS Paragon Plus Environment

Page 19 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

432

References

433

(1)

Peeling, R. W.; Artsob, H.; Pelegrino, J. L.; Buchy, P.; Cardosa, M. J.; Devi, S.; Enria,

434

D. A.; Farrar, J.; Gubler, D. J.; Guzman, M. G.; Halstead, S. B.; Hunsperger, E.; Kliks,

435

S.; Margolis, H. S.; Nathanson, C. M.; Nguyen, V. C.; Rizzo, N.; Vázquez, S.; Yoksan,

436

S. Nat. Rev. Microbiol. 2010, 8, S30–S37.

437

(2)

Castro, C.; Gourley, M. J. Allergy Clin. Immunol. 2010, 125, S238–S247.

438

(3)

Gallagher, M.; Wysocki, C. J.; Leyden, J. J.; Spielman, A. I.; Sun, X.; Preti, G. Br. J. Dermatol. 2008, 159, 780–791.

439 440

(4)

Shirasu, M.; Touhara, K. J. Biochem. 2011, 150, 257–266.

441

(5)

Kippenberger, S.; Havlíček, J.; Bernd, A.; Thaçi, D.; Kaufmann, R.; Meissner, M. Exp. Dermatol. 2012, 21, 655–659.

442 443

(6)

van Schooten, F. J. J. Breath Res. 2012, 6, 27108.

444 445

Boots, A. W.; van Berkel, J. J. B. N.; Dallinga, J. W.; Smolinska, A.; Wouters, E. F.;

(7)

Miekisch, W.; Schubert, J. K.; Noeldge-Schomburg, G. F. . Clin. Chim. Acta 2004, 347, 25–39.

446 447

(8)

Kaneko, T.; Wang, P. Y.; Sato, A. Occup. Environ. Med. 1994, 51, 68–72.

448

(9)

Cederbaum, A. I. Clin. Liver Dis. 2012, 16, 667–685.

449

(10)

Brooks, P. J.; Theruvathu, J. A. Alcohol 2005, 35, 187–193.

450

(11)

Salaspuro, M. J. Dig. Dis. 2011, 12, 51–59.

451

(12)

Garaycoechea, J. I.; Crossan, G. P.; Langevin, F.; Mulderrig, L.; Louzada, S.; Yang, F.;

452

Guilbaud, G.; Park, N.; Roerink, S.; Nik-Zainal, S.; Stratton, M. R.; Patel, K. J. Nature

453

2018, 553, 171–177.

454

(13) Fukunaga, T.; Yamamoto, K.; Adachi, J.; Ueno, Y.; Mizoi, Y. Japanese J. alcohol Stud. drug Depend. 1989, 24, 405–416.

455 456

(14)

Sekine, Y.; Toyooka, S.; Watts, S. F. J. Chromatogr. B 2007, 859, 201–207.

19 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

457

(15)

(16)

(17)

(18)

(19)

Agapiou, A.; Amann, A.; Mochalski, P.; Statheropoulos, M.; Thomas, C. L. P. TrAC Trends Anal. Chem. 2015, 66, 158–175.

466 467

Mochalski, P.; Unterkofler, K.; Hinterhuber, H.; Amann, A. Anal. Chem. 2014, 86, 3915–3923.

464 465

Turner, C.; Parekh, B.; Walton, C.; Spanel, P.; Smith, D.; Evans, M.; Patrik, S.; Smith, D.; Evans, M. Rapid Commun. Mass Spectrom. 2008, 22, 526–532.

462 463

Tsuda, T.; Ohkuwa, T.; Itoh, H. In Gas Biology Research in Clinical Practice; KARGER: Basel, 2011; pp 125–132.

460 461

Mochalski, P.; King, J.; Unterkofler, K.; Hinterhuber, H.; Amann, A. J. Chromatogr. B 2014, 959, 62–70.

458 459

Page 20 of 30

(20)

Curran, A. M.; Rabin, S. I.; Prada, P. a.; Furton, K. G. J. Chem. Ecol. 2005, 31, 1607– 1619.

468 469

(21)

Fenske, J. D.; Paulson, S. E. J. Air Waste Manage. Assoc. 1999, 49, 594–598.

470

(22)

Schmidt, F. M.; Vaittinen, O.; Metsälä, M.; Lehto, M.; Forsblom, C.; Groop, P.-H.; Halonen, L. J. Breath Res. 2013, 7, 17109.

471 472

(23)

Kamei, T.; Tsuda, T.; Mibu, Y.; Kitagawa, S.; Wada, H.; Naitoh, K.; Nakashima, K. Anal. Chim. Acta 1998, 365, 259–266.

473 474

(24)

Hara, T.; Matsui, H.; Shimizu, H. PLoS One 2014, 9, e111833.

475

(25)

Haze, S.; Gozu, Y.; Nakamura, S.; Kohno, Y.; Sawano, K.; Ohta, H.; Yamazaki, K. J. Invest. Dermatol. 2001, 116, 520–524.

476 477

(26)

Mitsubayashi, K. Sensors Mater. 2015, 27, 1123–1130.

478 479 480

Suzuki, Y.; Ye, M.; Miyajima, K.; Arakawa, T.; Sawada, S.-I.; Kudo, H.; Akiyoshi, K.;

(27)

Toma, K.; Miyajima, K.; Sawada, S.; Arakawa, T.; Kudo, H.; Akiyoshi, K.; Mitsubayashi, K. Sensors Mater. 2016, 28, 1265.

20 ACS Paragon Plus Environment

Page 21 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

481

Analytical Chemistry

(28)

Mitsubayashi, K. Sensors Mater. 2016, 28, 1295.

482 483

(29)

(30)

(31)

(32)

(33)

(34)

(35)

Phillips, M.; Herrera, J.; Krishnan, S.; Zain, M.; Greenberg, J.; Cataneo, R. N. J. Chromatogr. B Biomed. Sci. Appl. 1999, 729, 75–88.

496 497

Chien, P.-J.; Ye, M.; Suzuki, T.; Toma, K.; Arakawa, T.; Iwasaki, Y.; Mitsubayashi, K. Talanta 2016, 159, 418–424.

494 495

Iitani, K.; Chien, P.-J.; Suzuki, T.; Toma, K.; Arakawa, T.; Iwasaki, Y.; Mitsubayashi, K. ACS Sensors 2017, 2, 940–946.

492 493

Iitani, K.; Sato, T.; Naisierding, M.; Hayakawa, Y.; Toma, K.; Arakawa, T.; Mitsubayashi, K. Analyst 2017, 142, 3830–3836.

490 491

Arakawa, T.; Sato, T.; Iitani, K.; Toma, K.; Mitsubayashi, K. Anal. Chem. 2017, 89, 4495–4501.

488 489

Chien, P.-J.; Suzuki, T.; Tsujii, M.; Ye, M.; Toma, K.; Arakawa, T.; Iwasaki, Y.; Mitsubayashi, K. Biosens. Bioelectron. 2017, 91, 341–346.

486 487

Ye, M.; Chien, P.-J.; Toma, K.; Arakawa, T.; Mitsubayashi, K. Biosens. Bioelectron. 2015, 73, 208–213.

484 485

Saito, H.; Hashimoto, Y.; Minamide, T.; Kon, T.; Toma, K.; Arakawa, T.;

(36)

Navaneethan, U.; Parsi, M. A.; Gutierrez, N. G.; Bhatt, A.; Venkatesh, P. G. K.;

498

Lourdusamy, D.; Grove, D.; Hammel, J. P.; Jang, S.; Sanaka, M. R.; Stevens, T.; Vargo,

499

J. J.; Dweik, R. A. Gastrointest. Endosc. 2014, 80, 1038–1045.

500

(37)

Yokoyama, A.; Omori, T.; Yokoyama, T. Keio J. Med. 2010, 59, 115–130.

501

(38)

Jones, A. W. Alcohol Alcohol 1995, 30, 271–285.

502

(39)

Jones, A. W. J. Anal. Toxicol. 1986, 10, 98–101.

503

(40)

Norberg, A.; Gabrielsson, J.; Jones, A. W.; Hahn, R. G. Br. J. Clin. Pharmacol. 2000,

504

49, 399–408.

21 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

505

(41)

Page 22 of 30

Arakawa, T.; Wang, X.; Kajiro, T.; Miyajima, K.; Takeuchi, S.; Kudo, H.; Yano, K.; Mitsubayashi, K. Sensors Actuators B Chem. 2013, 186, 27–33.

506 507

(42)

Jones, A.; Neiman, J.; Hillbom, M. Br. J. Clin. Pharmacol. 1988, 25, 213–221.

508

(43)

Mizoi, Y.; Hishida, S.; Ijiri, I.; Maruyama, J.; Asakura, S.; Kijima, T.; Okada, T.; Adachi, J. Alcohol. Clin. Exp. Res. 1980, 4, 354–360.

509 510

(44)

Bioelectron. 2015, 67, 570–575.

511 512

Arakawa, T.; Kita, K.; Wang, X.; Miyajima, K.; Toma, K.; Mitsubayashi, K. Biosens.

(45)

Wilke, K.; Martin, A.; Terstegen, L.; Biel, S. S. Int. J. Cosmet. Sci. 2007, 29, 169–179.

513 514

22 ACS Paragon Plus Environment

Page 23 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

515

Analytical Chemistry

Graphics and captions

516 517

Figure 1. (a) Detection principle of AcH sniff-cam that employs a reverse reaction of ADH.

518

NADH absorbs UV light (λ = 340 nm) and emits fluorescence (λ = 490 nm). (b) Schematic

519

illustration of the experimental setup for characterization of the AcH sniff-cam. In

520

measurement of the exhaled breath, breath flow controller shown in (c) was used instead of

521

gas generator and MFCAcH in (b). In order to keep spaces between a palm skin surface and the

522

ADH-immobilized mesh constant, (d) PMMA frame and honeycomb spacer were used for

523

visualization of transdermal AcH.

524 23 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 24 of 30

525 526

Figure 2. (a) Fluorescence (upper) and differential images (lower) obtained by AcH sniff-cam

527

0, 20 and 120 s after applying 10 ppm AcH. The differential images were created by applying

528

an image differential method to the fluorescence images. (b) Comparison of the signal

529

responses of fluorescence intensity and differential value. The average of the mean intensity

530

of the ROI between 90–100 s after applying the AcH was defined as a signal output (∆I); the

531

peak of differential value was defined as a signal output (∆D).

532

24 ACS Paragon Plus Environment

Page 25 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

533 534

Figure 3. (a) Calibration curve of the AcH sniff-cam to gaseous AcH. The dynamic range was

535

0.1–10 ppm with the correlation coefficient of 0.997. The inset shows time course of the

536

differential value with various concentration of AcH (0.1, 0.5, 1, 3, 5, 8 and 10 ppm). The

537

upper images are differential images 20 s after applying 1, 5 and 10 ppm of AcH. (b) The

538

selectivity of the AcH sniff-cam against typical chemical components in breath. The output of

539

each sample was normalized by that of AcH.

540 541

25 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 26 of 30

542 543

Figure 4. (a)Differential images 0, 20 and 120 s after applying exhaled breath of ALDH2[+]

544

and ALDH2[-] subject 30 minutes after drinking. (b) Time course of the corresponding

545

differential values.

546

26 ACS Paragon Plus Environment

Page 27 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

547 548

Figure 5. Time course of concentration of breath AcH after drinking obtained from (a)

549

ALDH2[+] subjects and (b) ALDH2[-] subjects. As a comparison, the concentration was

550

measured by gas detection tubes simultaneously (vertical bar).

551

27 ACS Paragon Plus Environment

Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 28 of 30

552 553

Figure 6. (a) Distribution of transdermal AcH over a palm of ALDH2[+] subject 30 minutes

554

after drinking. (b) Time course of fluorescence intensity ∆I, differences from the intensity 15

555

minutes before drinking.

556

28 ACS Paragon Plus Environment

Page 29 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

557

Analytical Chemistry

for TOC only ADH

acetaldehyde

CH3CHO +

NADH

Ex. 340 nm

+

H+

pH6.5

ethanol

CH3CH2OH + NAD+

pH10.0 Fl. 490 nm

distribution of acetaldehyde UV-LED array

ADH-immobilized mesh

camera BPF

558

29 ACS Paragon Plus Environment

ADH

Analytical Chemistry

acetaldehyde

1 3 2 3 4 Ex. 340 nm 5 6 7 8 UV-LED array 9 10 11 12 13 14 15 16 17 18 19 20 ADH-immobilized mesh 21 22 23

CH CHO +

NADH

+

H+

pH6.5 pH10.0

Page 30 of 30

ethanol

CH3CH2OH + NAD+

Fl. 490 nm

distribution of acetaldehyde

camera BPF

ACS Paragon Plus Environment