High performance method for rapid determination of Pu isotopes in soil and sediment samples by sector field inductively coupled plasma mass spectrometry Zhongtang Wang, Jian Zheng, Youyi Ni, Wu Men, Keiko Tagami, and Shigeo Uchida Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.6b04975 • Publication Date (Web): 18 Jan 2017 Downloaded from http://pubs.acs.org on January 19, 2017
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
High performance method for rapid determination of Pu
2
isotopes in soil and sediment samples by sector field
3
inductively coupled plasma mass spectrometry
4 5
Zhongtang Wang1, Jian Zheng1*, Youyi Ni1, 2, Wu Men1
6
Keiko Tagami1, Shigeo Uchida1
7 1
8
Biospheric Assessment for Waste Disposal Team,
9
National Institute of Radiological Sciences,
10
National Institutes for Quantum and Radiological Science and Technology
11
4-9-1 Anagawa, Inage, Chiba 263-8555, Japan
12 13 14
2
State Key Laboratory of Nuclear Physics and Technology, Peking University, Beijing 100871, China
15 16 17 18 19
______________________________________________________________________
20
*Corresponding author. Tel.: +81 43 2064605; Fax: +81 43 2064601.
21
E-mail address:
[email protected] 22
1
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
23
Page 2 of 28
ABSTRACT
24
Plutonium is extensively studied in radioecology (e.g. soil to plant transfer and
25
radiological assessment) and geochemistry (e.g. sediment dating). Here, we reported a
26
new chemical separation method for rapid determination of Pu in soil and sediment
27
samples, based on the following investigations: extraction behaviors of interfering
28
elements (IEs, for ICPMS measurement) on TEVA resin; decontamination of U using
29
TEVA, UTEVA and DGA resins and the impact of co-precipitation on Pu determination.
30
The developed method consists of four steps: HNO3 leaching for Pu release; CaF2/LaF3
31
co-precipitation
32
TEVA+UTEVA+DGA procedure for the removal of U, Pb, Bi, Tl, Hg, Hf, Pt and Dy;
33
and ICPMS measurement. The accuracy of this method in determining 239+240Pu activity
34
and 239Pu/240Pu and 241Pu/239Pu isotopic ratios was validated by analyzing five standard
35
reference materials (soil, fresh water sediment and ocean sediment). This method is
36
characterized by its stable and high Pu recovery (90-97% for soil; 92-98% for sediment)
37
and high decontamination factor of U (1.6 × 107) which is the highest reported for soil
38
and sediment samples. In addition, the short analytical time of 12 h and the method
39
detection limits, which are the lowest yet reported in literature, of 0.56 µBq g-1 (0.24 fg
40
g-1) for 239Pu, 1.2 µBq g-1 (0.14 fg g-1) for 240Pu, and 0.34 mBq g-1 (0.09 fg g-1) for 241Pu
for
the
removal
of
major
metals
2
ACS Paragon Plus Environment
and
U;
the
proposed
Page 3 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
41
(calculated on the basis of a 1 g soil sample) allow the rapid determination of ultratrace
42
level Pu in soil and sediment samples.
43 44
3
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
45
Page 4 of 28
Introduction The globally distributed
46
241
239
Pu (T1/2 = 2.4×104 years),
240
Pu (T1/2 = 6.5×103 years)
47
and
Pu (T1/2 = 14.4 years) background resulted from nuclear weapon detonations in
48
the last century. Meanwhile, regional Pu inputs are found at places adjacent to nuclear
49
weapon test sites and nuclear accident sites. Since its introduction into the environment,
50
attention has been given to studying Pu not only for the purpose of radiological
51
assessment due to its radiotoxicity,1,2 but also for applications using Pu as a tracer to
52
study biological and geochemical processes, e.g. soil to plant transfer,3 soil erosion and
53
sediment dating.4-6
54
In the literature, both radiometric and mass spectrometric methods have been used
55
in the determination of Pu for soil and sediment samples.7,8 In recent years, due to the
56
great advantages for both quantitative and isotopic ratio measurements with respect to
57
its easy sample preparation, relatively low cost and high sensitivity,9 inductively
58
coupled plasma mass spectrometry (ICPMS) is replacing conventional radiometric
59
methods for ultratrace Pu determination. However, ICPMS measurements are affected
60
by polyatomic interferences which generate false signals at the same m/z ratio of Pu
61
isotopes. Previous studies reported that elements such as U, Pb, Bi, Tl, Hg, Hf, Pt and
62
Dy can cause possible polyatomic interferences for the ultra-trace determination of
4
ACS Paragon Plus Environment
Page 5 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
63
239
64
should be removed during sample preparation.
Pu,
240
Pu,
241
Pu,
242
Pu by ICPMS,9-11 indicating that these interfering elements (IEs)
65
In the sample preparation for ICPMS measurements, three separation strategies
66
have been commonly used: solvent extraction, ion-exchange chromatography and
67
extraction chromatography. Among them, extraction chromatography exhibits faster
68
exchange kinetics, uses less acid and produces less hazardous waste, which explains its
69
increasing popularity.12 Due to the high retention ability of Pu,13 TEVA resin is the
70
mostly employed extraction resin in Pu studies. However, the reported methods mainly
71
focused on the decontamination of U (for mass spectrometry) and Th (for alpha
72
spectrometry), and the removal of other IEs has not been studied. In the reported
73
methods using a single TEVA column for separation, decontamination factors of U [DFs
74
(U)] on the orders of 103 - 104 were achieved.14-17 However, the DFs (U) on the orders
75
of 103 - 104 are not sufficient for ultratrace Pu analysis in high U content samples.18
76
Furthermore, there was an inconsistency in the reported methods on the utilization of
77
co-precipitation. For some researchers, co-precipitations were employed to remove the
78
matrix in soil and sediment which might interfere with the subsequent separation of Pu
79
on TEVA resin;15,19,20 while for others, no co-precipitation was applied.14,21,22 Further
80
studies need to be carried out to demonstrate the impact of co-precipitation on Pu
5
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
81
determination.
82
In a recent study, Rosenberg et al. demonstrated the separation of interfering
83
fission products, such as radiocesium, radioiodine and radiotellurium in gamma
84
spectrometry analysis of
85
extraction behaviors of IEs on TEVA resin, and quantified the decontamination abilities
86
of sample matrix elements and U for CaF2/LaF3 co-precipitation and separation using
87
TEVA, UTEVA and DGA resins. On the basis of these investigations, we proposed a
88
new analytical method to rapidly determine Pu isotopes in soil and sediment samples by
89
ICPMS for the preparedness of nuclear emergency response. Finally, we evaluated our
90
method using five standard reference materials.
239
Np using TEVA resin.23 In this study, we investigated the
91 92
Experimental section
93
Instrumentation. For the measurement of major matrix elements in soil samples,
94
an ICP-AES instrument (Activa-M, Horiba, Kyoto, Japan) was employed. A
95
SF-ICP-MS instrument (Element XR, Thermo Scientific, Bremen, Germany) equipped
96
with a Scott type spray chamber was utilized for the determination of IEs. To measure
97
ultratrace Pu isotopes, a high efficiency sample introduction system Apex-Q was
98
connected to the SF-ICPMS instrument. Detailed settings and evaluation of this system
6
ACS Paragon Plus Environment
Page 6 of 28
Page 7 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
99
can be found in our previous work.24
100
Elution experiment. An elution experiment was conducted to investigate the
101
elution efficiency of HNO3 and HCl at various acidities for the removal of IEs from
102
TEVA resin. The investigated IEs were U, Pb, Bi, Tl, Hg, Hf, Pt, Dy and Th. The elution
103
experiment was carried out as follows: (1) preconditioning TEVA resin with 5 mL of
104
tested acid solution; (2) loading 5 mL of spiked acid solution (each element at a
105
concentration of 2 ng mL-1, and at the same acidity) onto the TEVA resin cartridge; (3)
106
rinsing with another 5 mL of the tested acid solution; and (4) collecting fractions (2) and
107
(3) for ICPMS measurements. The tested molarities of HNO3 were 0.01 M, 0.05 M, 0.1
108
M, 0.5 M, 1 M, 3 M, 5 M and 8 M. The tested molarities of HCl were 0.01 M, 0.05 M,
109
0.1 M, 0.5 M, 1 M, 6 M, 9 M, and 9.5 M.
110
HNO3 leaching and CaF2/LaF3 co-precipitation. The HNO3 leaching method
111
was applied to leach Pu from soil and sediment samples. Specifically, 0.2 – 2.5 g soil or
112
sediment sample was first ashed in a muffle furnace at 450 °C for 4 h to decompose
113
organic matter.25 Then the ashed sample was transferred to a 120 mL PTFE vessel, to
114
which 10 mL conc. HNO3 and 0.57 pg 242Pu yield tracer were subsequently added. After
115
heating the vessel on a hotplate at 160 °C for 4 h, the leachate was filtered into a 50 mL
116
plastic centrifuge tube using a filter paper (Ø150 mm). Milli-Q water was added to
7
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
117
adjust the sample volume to 35 mL (ca. 3.8 M HNO3). Then 100 mg Ca (0.59 g
118
Ca(NO3)2•4H2O) and 100 mg La (0.76 g La(NO3)3•6H2O) were added, followed by the
119
addition of 2 mL of 20% TiCl3, which reduced Pu (IV) to Pu (III). After careful addition
120
of 7 mL of 46-48% HF, the suspension was mixed thoroughly and the precipitate
121
allowed to settle for 15-20 min. After centrifugation at 3000 rpm for 15 min, the
122
supernatant was discarded and the precipitate was dissolved by 20 mL of 3 M HNO3
123
with the addition of 0.5 g of H3BO3, upon which the solution was ready for plutonium
124
valence adjustment and extraction chromatographic separation. To enhance the
125
dissolution of CaF2/LaF3 precipitate in 3 M HNO3, H3BO3 was added because it reacts
126
with fluoride precipitate and generate soluble BF4-. In addition, the addition of H3BO3
127
can prevent the formation of Pu fluoride complex, thus improve the adsorption of Pu in
128
TEVA resin.
129
Pu separation using TEVA+UTEVA+DGA resins. After co-precipitation, the
130
extraction chromatographic separation step followed (Figure 1). The valence state of
131
plutonium was first adjusted to Pu (IV) by the addition of 0.3 g NaNO2 into the 50 mL
132
centrifuge tube and heated at 40 °C for 0.5 h in a water bath. Then at a flow rate of 1 mL
133
min-1, the sample solution was loaded onto a TEVA resin cartridge which had been
134
preconditioned by 10 mL of 3 M HNO3 on a polycarbonate vacuum box (Eichrom, IL,
8
ACS Paragon Plus Environment
Page 8 of 28
Page 9 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
135
USA). After sample loading, an additional 10 mL of 3 M HNO3 was used to remove Ca,
136
Fe and rare earth elements (REEs), followed by 40 mL of 1 M HNO3 to remove U, Pb,
137
Tl and Pt, and 10 mL of 9 M HCl to remove Th, Bi and Hf (at a flow rate of 2 mL min-1).
138
Before the elution of Pu, an UTEVA and a DGA resin cartridges both had been
139
preconditioned by adding 10 mL of 3 M HNO3 were connected to the TEVA resin
140
cartridge. Then 20 mL of 3 M HNO3 – 0.1 M ascorbic acid – 0.02 M Fe2+ (prepared
141
from 25% iron(II) sulfamate) was employed to reduce Pu (IV) to Pu (III) and elute Pu
142
(III) from TEVA resin (flow rate: 1 mL min-1). The eluted Pu (III) fraction passed
143
through the UTEVA resin cartridge and was retained on the DGA resin. After elution,
144
the TEVA and UTEVA resin cartridges were discarded, leaving the DGA resin cartridge
145
which was then rinsed by 30 mL 0.1 M HNO3 to remove U, Tl, Pb, Pt, Hf and Fe (flow
146
rate: 2 mL min-1). Finally, the plutonium on the DGA resin was eluted into a 50 mL
147
PTFE vial by 20 mL of 0.5 M HCl – 0.1 M NH2OH·HCl (flow rate: 1 mL min-1). The
148
eluted sample was evaporated to dryness at 250 °C and dissolved by 4 mL of aqua regia.
149
After heating the dissolved sample solution to dryness at 200 ℃, 1 mL concentrated
150
HNO3 was added and this was heated to near dryness at 250 ℃. Finally, the sample was
151
dissolved in 0.7 mL of 4% HNO3 and ready for SF-ICPMS measurements.
152
9
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
153
Results and discussion
154
Results of elution experiment. The results of the elution experiment are
155
shown in Figure S1, in which the y-axis indicates the eluted fractions of IEs during the
156
sample loading and subsequent washing steps. In HNO3 and HCl mediums, the eluted
157
fractions of U exhibited a decreasing trend as the acidities increased, in good agreement
158
with the distributions of the retention factors reported by Horwitz et al.13 Similar
159
consistency was found for Th: both datasets showed that Th was only retained on TEVA
160
resin in HNO3 for which molarity was higher than 0.5 M. These agreements
161
demonstrated the accuracy of our results and provided us with a valid base to discuss
162
the behaviors of other IEs. For Bi, it was adsorbed by TEVA resin in diluted HNO3 and
163
HCl and could only be stripped down by high concentration HNO3 (> 8 M) and HCl (>
164
9 M). In the case of Pb, no retention was shown in the whole HNO3 concentration range
165
or in diluted HCl (< 0.1 M) and only weak retention could be found in 0.5-1 M HCl.
166
Similar distribution patterns were observed for Tl, Pt and Dy; all these three IEs could
167
be eluted by HNO3 and diluted HCl (< 0.1 M). Differences appeared in 0.1- 9.5 M HCl:
168
the retention abilities of Tl and Pt increased with the acidity of HCl, while Dy was not
169
retained on TEVA in the whole HCl concentration range. In the case of Hf, it could be
170
eluted by 0.05 – 9 M HCl and 8 M HNO3. Finally, Hg could only be stripped down by
10
ACS Paragon Plus Environment
Page 10 of 28
Page 11 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
171
Analytical Chemistry
HNO3 solutions having concentrations higher than 5 M.
172
Decontamination of IEs using TEVA resin. In the literature, 1 M HNO3, 3 M
173
HNO3, and 8 M HNO3 are most frequently used for sample loading and matrix washing
174
on TEVA resin.26-31 Our results showed that 8 M HNO3 was unable to strip the most
175
important interfering element U, although it could elute Bi and Hg which 1 M and 3 M
176
HNO3 could not elute. 1 M and 3 M HNO3 did not show significant variations in the
177
washing of IEs. Thus, we selected 3 M HNO3 (20 mL) for sample loading to take
178
advantage of the highest retention factor for Pu on TEVA resin.13 After sample loading,
179
additional 10 mL of 3 M HNO3 and 40 mL of 1 M HNO3 solutions were used for the
180
washing of IEs. Various researchers have employed 6 M HCl and 9 M HCl to remove
181
Th from TEVA resin.22,26,32,33 Our results showed that both acidities HCl had similar
182
elution efficiencies in removing Th, Pb, Dy and Hf, but 9 M HCl could additionally
183
strip Bi (Figure S1). Thus, we chose 9 M HCl in our method. In summary, our approach
184
utilizing TEVA resin for the decontamination of IEs was established, and the detailed
185
description was given in the experimental section.
186
Improvement in the decontamination of U. Uranium is the key interfering
187
element in the Pu determination by ICPMS, and the decontamination of U is especially
188
essential for U-rich environmental soil and sediment samples with ultratrace level Pu.
11
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
189
To find out the decontamination ability of U by our proposed TEVA separation
190
procedure, we used U spiked standard solution (10 µg of U) for an investigation, and
191
our results showed that the DF (U) of the procedure was 2.4 × 104, which was in
192
accordance with other reported values using a single TEVA resin cartridge for Pu
193
separation.14-16 To thoroughly remove U interference on Pu measurement, the signal
194
intensities of U for soil and sediment samples in ICPMS measurement should be
195
controlled to an operational blank level (ca. 8 × 104 cps based on the sensitivity of 6 ×
196
107 cps per ng mL-1 U in our Apex-SF-ICPMS system). Considering the typical 238UH+ /
197
238
198
counting rate of
199
interference from
200
with a U concentration of 3µg g-1, we estimated a counting rate of 6.4 × 1011 cps when
201
no decontamination operation was applied, based on the sensitivity of our instrument
202
(final sample volume: 0.7 mL). Thus, a DF (U) of 8 × 106 was required to thoroughly
203
remove U interference. Obviously, passing through a single TEVA resin cartridge was
204
not sufficient.
U+ ratio in Apex-SF-ICPMS measurement (1-2 × 10-5),8 operational blank level 238
U+ could result in a UH+ signal of ca. 1 cps, indicating a negligible
238
U+ on ultratrace level Pu determination. For a 2.5-g soil sample
205
Maxwell et al.34 reported an analytical procedure which combined TEVA resin with
206
DGA resin for U removal. U and Pu (III) were stripped from TEVA resin to DGA resin
12
ACS Paragon Plus Environment
Page 12 of 28
Page 13 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
207
by 20 mL of 3 M HNO3 - 0.1 M ascorbic acid - 0.02 M Fe2+. Then Pu (III) was oxidized
208
to Pu (IV) by 5 mL of 8 M HNO3, followed by 20 mL of 0.1 M HNO3 and 10 mL 0.05
209
M HNO3 for U removal. Finally, Pu (IV) was reduced to Pu (III) and eluted by 11 mL of
210
0.02 M HCl – 0.005 M HF – 0.0001 M TiCl3. The Pu recovery of their analytical
211
method ranged from 70% to 87%. We made two improvements to simplify the
212
procedure and enhance U removal and Pu recovery. First, Pu (III) eluted from TEVA
213
resin including U was directly loaded onto the DGA resin. Because our previous study
214
had shown that trivalent actinides were retained on DGA resin during the 0.1 M HNO3
215
rinse to elute U,35 the Pu oxidation state adjustment was omitted. Second, to avoid
216
potential damage to the glassware in the sample introduction system by HF and
217
contamination of the ICPMS instrument by the highly abundant Ti, the elution reagent
218
was replaced by a novel reagent proposed in this study: 20 mL 0.5 M HCl – 0.1 M
219
NH2OH∙HCl. According to the elution curve shown in Figure S2, nearly 100% Pu was
220
recovered by 20 mL 0.5 M HCl – 0.1 M NH2OH∙HCl. After these improvements, the
221
separation step on DGA resin was simple: elution was conducted after introducing 30
222
mL of 0.1 M HNO3 on the DGA resin cartridge to remove U, Tl, Pb, Pt, Hf, and Fe.35
223
After combining the DGA separation with the TEVA procedure, a DF (U) of 2.3 × 105
224
was achieved. When one UTEVA resin cartridge was added between the TEVA and
13
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
225
DGA resins cartridges during stripping of Pu from TEVA resin to DGA resin, the DF
226
(U) reached 6.7 × 106 due to the strong U adsorption of UTEVA resin in 3M HNO3
227
medium. DF (U) of 6.7 × 106 is generally sufficient for U removal for soil and sediment
228
samples. The whole chromatographic separation procedure using TEVA+UTEVA+DGA
229
resins is summarized in the experimental section.
230
Influence of CaF2/LaF3 coprecipitation. As we discussed in the introduction,
231
there was some inconsistence in the utilization of co-precipitation before Pu separation
232
using TEVA. To find out the effect of co-precipitation, we first evaluated the matrix
233
removal efficiency by CaF2/LaF3 co-precipitation for 5 g JSAC-0471 soil samples (n=3).
234
The co-precipitated fractions of matrix elements, Al, Fe, K, Mg, Na and U are shown in
235
Figure S3, which indicated that the majority of the matrix elements in soil were
236
removed after a co-precipitation was applied. Moreover, the CaF2/LaF3 co-precipitation
237
exhibited the ability of U decontamination; about 60% of U was removed.
238
To test the impact of co-precipitation on Pu recovery, we used Japanese soil
239
samples (sample information see Yang et al.36) for an investigation. Different amounts
240
of soil samples (0 g, 0.5 g, 1 g, 2 g, n=3) were applied for the analytical method shown
241
in Figure 1. A control group of samples was also prepared by the same method except
242
for the co-precipitation part. The results of Pu recovery for both groups are shown in
14
ACS Paragon Plus Environment
Page 14 of 28
Page 15 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
243
Figure S4. For samples that were prepared employing the CaF2/LaF3 co-precipitation,
244
the Pu recoveries were steady and high (>90%, 0-2 g). On the other hand, for samples
245
without co-precipitation, the recovery of the operation blank was also higher than 90%,
246
but the recoveries for soil samples were relatively lower and unstable, ranging from
247
53% to 86%. Similar recoveries have been reported by other studies using TEVA resin
248
without co-precipitation to analyze soil and sediment samples, e.g. 46% - 80%
249
recoveries were reported by Muramatsu et al.14 and 44% - 83% recoveries were reported
250
by Nygen et al.22 High Pu recoveries for soil and sediment samples were only found in
251
those studies utilizing co-precipitations, e.g. 72% -92% reported by Varga et al.19 (CaF2
252
co-precipitation); 80% - 105% reported by Qiao et al.27 (Fe(OH)3 co-precipitation) and
253
104% ± 4.4% reported by Maxwell et al.20 (Fe(OH)3 + CeF3 co-precipitations). The
254
agreement between the reported results and our results confirmed the effect of
255
co-precipitation on the matrix removal and Pu recovery stabilization. Therefore, we
256
recommend the use of co-precipitation to obtain constantly high Pu recovery when
257
TEVA resin is employed for Pu separation for soil and sediment samples.
258
Evaluation of the analytical method. On the basis of above discussion, the
259
analytical method of ultratrace Pu determination for soil and sediment samples is
260
summarized in Figure 1. To demonstrate the accuracy of our method, 5 standard
15
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
261
reference materials including IAEA-385 (ocean sediment), NIST-4354 (fresh water
262
sediment), IAEA-soil-6 (soil), NIST-4357 (ocean sediment) and JSAC-0471 (soil) were
263
analyzed. The Pu results (239+240Pu activity, 240Pu/239Pu isotopic ratio) of these materials
264
together with the certified/information/literature values are shown in Table S1. The
265
240
266
for all materials. Meanwhile, the determined 239+240Pu activities were consistent with the
267
reported ranges, illustrating the accuracy of our method in Pu determination for ocean
268
sediment, fresh water sediment and soil. In addition, 241Pu/239Pu isotopic ratio (0.0135 ±
269
0.0004, decay corrected to 1 January 2000, n=3) was also determined for NIST-4357,
270
and was in agreement with previously reported values: 0.0131 ± 0.0010 by Bu et al.;18
271
and 0.0132 ± 0.0007 by Zhang et al.37 As shown in Table S1, stable and high Pu
272
recoveries were achieved by our method: 92% - 98% for sediment and 90% - 97% for
273
sediment. These recoveries are among the highest values reported in Pu studies.
Pu/239Pu isotopic ratios were in good accordance with the certified/literature values
274
The decontamination ability for U of our method was also assessed by soil samples,
275
and the DF (U) of 1.6 ± 0.5 × 107 (n=3) was obtained. This value was close to the
276
theoretical value 1.7 × 107 which was calculated from the DF (U) of CaF2/LaF3
277
co-precipitation (100 % / 40 % = 2.5, initial fraction divided by co-precipitated fraction)
278
and DF (U) of the TEVA+UTEVA+DGA separation procedure (6.7 × 106). Compared to
16
ACS Paragon Plus Environment
Page 16 of 28
Page 17 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
279
the previously reported DFs (U) of soil and sediment (Table 1),14,15,17,18,26,34,38 the DF
280
(U) in this study was higher than in previous studies using ion-exchange
281
chromatography or extraction chromatography. The signal intensities of ICPMS
282
measurement for other IEs were also controlled to an operational blank level, resulting
283
in high DFs for Hg (1.3×105), Tl (2.8×105), Dy (6.6×103), Pt (4.0×104), Pb (8.6×105), Bi
284
(6.0×104) and Hf (5.0×104). The instrument detection limits of
285
239
Pu,
240
Pu and
241
Pu were determined to be
286
0.06 fg mL-1, 0.05 fg mL-1 and 0.06 fg mL-1 respectively, based on the estimation of 3
287
times the standard deviation of a 4% HNO3 blank solution. The limit of detection
288
(LOD) of our analytical method was calculated in a similar way: 3 times the standard
289
deviation of the operation blanks. On the basis of analyzing 1 g soil or sediment with a
290
Pu recovery of 90%, the LODs of 239Pu, 240Pu and 241Pu were calculated to be 0.56 µBq
291
g-1 (0.24 fg g-1), 1.2 µBq g-1 (0.14 fg g-1) and 0.34 mBq g-1 (0.09 fg g-1). The LOD of
292
239
293
µBq g-1) reported for using α spectrometry.39 Compared to other reported LODs of 239Pu
294
using ICPMS for Pu determination, e.g. 1.7 fg g-1 reported by Truscott et al.,40 9 fg g-1
295
reported by Varga et al.19 and 6 fg g-1 reported by Kim et al.,41 our LOD of
296
fg g-1) was also lower, due to the higher sensitivity of our instrument and higher Pu
Pu obtained in this study (0.56 µBq g-1) was lower than the lowest LOD of 239Pu (1.2
17
ACS Paragon Plus Environment
239
Pu (0.24
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
241
Pu, our LOD (0.34 mBq g-1) was lower than the previously reported 2
297
recovery. For
298
mBq g-1 by ICPMS,42 and comparable to the lowest reported LOD of 0.1 mBq g-1 by a
299
liquid scintillation technique.43 Consequently, the lowest LOD of Pu for soil and
300
sediment samples was reported by this study.
301
The whole analytical method takes about 12 h (HNO3 leaching: 4h; filtration: 1 h;
302
coprecipitation: 1h; Pu separation on extraction resin: 2 h; sample preparation for
303
ICPMS measurements: 4 h) for 20 samples, which can be finished within two days.
304
Compared to conventional ion-exchange chromatography, which usually takes about 4-5
305
days for Pu separation,18 this method significantly shortens the analytical time. In
306
addition, this method produces a small amount of hazardous waste acid and requires
307
less evaporation of acid, greatly reducing the burden of radioactive laboratory
308
management.
309 310
Conclusions
311
In this study, we investigated the decontamination ability of IEs for the extraction
312
resin, TEVA, UTEVA and DGA. The results led to the establishment of a
313
TEVA+UTEVA+DGA separation procedure which is capable of removing IEs including
314
U, Pb, Bi, Tl, Hg, Hf, Pt, and Dy. We also assessed the effect of co-precipitation on Pu
18
ACS Paragon Plus Environment
Page 18 of 28
Page 19 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
315
analysis, and found that CaF2/LaF3 co-precipitation removed the majority of matrix
316
elements and U in soil samples, leading to the stabilized Pu recovery in the subsequent
317
extraction chromatographic separation. On the basis of these fundamental investigations,
318
we proposed an analytical method for Pu analysis in soil and sediment samples. The
319
method consists of four steps: HNO3 leaching, CaF2/LaF3 co-precipitation, the
320
TEVA+UTEVA+DGA chromatographic separation and ICPMS measurement. We
321
evaluated this method by analyzing five standard reference materials. More importantly,
322
the highest DF (U) (1.6 × 107) and lowest LODs (0.56 µBq g-1 (0.24 fg g-1) for
323
1.2 µBq g-1 (0.14 fg g-1) for
324
achieved for soil and sediment samples. In addition, the stable and high Pu recoveries
325
(90-97% for soil; 92-98% for sediment) and short analytical time (12 h) demonstrated
326
the robustness and high sample throughput in the ultratrace determination of Pu isotopes
327
for soil and sediment samples.
240
Pu, and 0.34 mBq g-1 (0.09 fg g-1) for
241
239
Pu,
Pu) were
328 329 330 331
Acknowledgements This work was supported by the Agency for Natural Resources and Energy, the Ministry of Economy, Trade and Industry (METI), Japan.
332
19
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
333
References
334
(1) Steinhauser, G. Environ. Sci. Technol. 2014, 48, 4649-4663.
335
(2) Zheng, J.; Tagami, K.; Homma-Takeda, S.; Bu, W. J. Anal. At. Spectrom. 2013, 28,
336 337 338
1676-1699. (3) Froehlich, M. B.; Dietze, M. M. A.; Tims, S. G.; Fifield, L. K. J. Environ. Radioactiv. 2016. 151 (3), 558-562.
339
(4) Xu, Y.; Qiao, J.; Hou, X.; Pan, S. Sci. Rep. 2013, 3, 3506.
340
(5) Ketterer, M. E.; Watson, B. R.; Matisoft, G.; Wilson, C. G. Environ. Sci. Technol.
341 342 343 344 345
2002, 36, 1307-1311. (6) Wu, F.; Zheng, J.; Liao, H.; Yamada, M. Environ. Sci. Technol. 2010. 44(8), 2911-2917. (7) Maxwell, S. L.; Culligan, B.; Hutchison, J. B.; McAlister, D. R. J. Radioanal. Nucl. Chem. 2015, 305(2), 599-608.
346
(8) Zheng, J.; Yamada, M. Talanta 2006, 69, 1246-1253.
347
(9) Kim, C. S.; Kim, C. K.; Martin, P.; Sansone, U. J. Anal. At. Spectrom. 2007. 22(7),
348 349 350
827-841. (10) Liao, H.; Zheng, J.; Wu, F.; Yamada, M.; Tan, M.; Chen, J. Appl. Radiat. Isotopes 2008. 66(8), 1138-1145.
20
ACS Paragon Plus Environment
Page 20 of 28
Page 21 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
351
(11) Pointurier, F.; Hémet, P.; Hubert, A. J. Anal. At. Spectrom. 2008. 23(1), 94-102.
352
(12) Vajda, N.; Kim, C. K. J. Radioanal. Nucl. Chem. 2010. 283(1), 203-223.
353
(13) Horwitz, E. P.; Dietz, M. L.; Chiarizia, R.; Diamond, H.; Maxwell, S. L.; Nelson,
354 355 356 357 358 359 360 361 362 363 364 365 366
M. R. Anal. Chim. Acta. 1995. 310(1), 63-78. (14) Muramatsu, Y.; Uchida, S.; Tagami, K.; Yoshida, S.; Fujikawa, T. J. Anal. At. Spectrom. 1999. 14(5), 859-865. (15) Qiao, J.; Hou, X.; Roos, P.; Miró, M. J. Anal. At. Spectrom. 2010. 25(11), 1769-1779. (16) Choi, M. S.; Lee, D. S.; Choi, J. C.; Cha, H. J.; Yi, H. I. Sci. Total Environ. 2006. 370(1), 262-270. (17) Godoy, M. L. D.; Godoy, J. M.; Roldao, L. A. J. Environ. Radioactiv. 2007. 97(2), 124-136. (18) Bu, W. T.; Zheng, J.; Guo, Q. J; Aono, T.; Tazoe, H.; Tagami, K.; Uchida, S.; Yamada, M. Environ. Sci. Technol. 2013, 48(1), 534-541. (19) Varga, Z.; Surányi, G.; Vajda, N.; Stefánka, Z. Radiochim. Acta. 2007, 95(2), 81-87.
367
(20) Maxwell, S. L.; Culligan, B. K. J. Radioanal. Nucl. Chem. 2006, 270(3), 699-704.
368
(21) Chamizo, E.; García-León, M.; Enamorado, S. M.; Jiménez-Ramos, M. C.; Wacker,
21
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
369 370 371 372 373
L. Atmos. Environ. 2010. 44(15), 1851-1858. (22) Nygren, U.; Rodushkin, I.; Nilssona, C.; Baxter, D.C. J. Anal. At. Spectrom. 2003, 18, 1426-1434. (23) Rosenberg, B. L.; Shozugawa, K.; Steinhauser, G. Anal. Chem. 2015, 87, 8651-8656.
374
(24) Zheng, J. J. Nucl. Radiochem. Sci. 2015, 15(1), 7-13.
375
(25) Wang, Z. T.; Yang, G. S.; Zheng, J.; Cao, L. G.; Yu, H. J.; Zhu, Y. B.; Tagami, K.;
376 377 378
Uchida, S. Anal. Chem. 2015, 87(11), 5511-5515. (26) Xu, Y. H.; Qiao, J. X.; Hou, X. L.; Pan, S. M.; Roos, P. Talanta. 2014, 119, 590-595.
379
(27) Qiao, J. X.; Hou, X. L.; Roos, P.; Miro, M. Anal. Chem. 2009, 81, 8185-8192.
380
(28) Lariviere, D.; Benkhedda, K.; Kiser, S.; Johnson, S.; Cornett, R. J. Anal.
381 382 383 384 385 386
Methods-UK, 2010. 2(3), 259-267. (29) Bisinger, T.; Hippler, S.; Michel, R.; Wacker, L.; Synal, H. A. Nucl. Instrum. Meth. B. 2010. 268(7), 1269-1272. (30) Kim, H.; Chung, K. H.; Jung, Y.; Jang, M.; Kang, M.; Choi, G. S. J. Radioanal. Nucl. Chem. 2015. 304(1), 321-327. (31) Luisier, F.; Alvarado, J. A. C.; Steinmann, P.; Krachler, M.; Froidevaux, P. J.
22
ACS Paragon Plus Environment
Page 22 of 28
Page 23 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
387 388 389 390 391 392 393 394 395
Analytical Chemistry
Radioanal. Nucl. Chem. 2009, 281(3), 425-432. (32) Chamizo, E.; Jiménez-Ramos, M. C.; Wacker, L.; Vioque, I.; Calleja, A.; García-León, M.; Garcia-Tenorio, R. Anal. Chim. Acta. 2008. 606(2), 239-245. (33) Grate, J. W.; O’Hara, M. J.; Farawila, A. F.; Douglas, M.; Haney, M. M.; Petersen, S. L.; Maiti, T. C.; Aardahl, C. L. Anal. Chem. 2011. 83(23), 9086-9091. (34) Maxwell, S. L.; Culligan, B. K.; Jones, V. D.; Nichols, S. T.; Bernard, M. A.; Noyes, G. W. Anal. Chim. Acta. 2010. 682(1), 130-136. (35) Wang, Z. T.; Zheng, J., Cao, L. G.; Tagami, K.; Uchida, S. Anal. Chem. 2016. 88(14), 7387-7394.
396
(36) Yang, G. S.; Zheng, J.; Tagami, K.; Uchida, S. Sci. Rep. 2015, 5, 9636.
397
(37) Zhang, Y.; Zheng, J.; Yamada, M.; Wu, F.; Igarashi, Y.; Hirose, K. Sci. Total
398 399 400 401 402 403 404
Environ. 2010, 408, 1139-1144. (38) Zheng, J.; Yamada, M.; Wang, Z.; Aono, T.; Kusakabe, M. Anal. Bioanal. Chem. 2004. 379(3), 532-539. (39) Jia, G. G.; Testa, C.; Desideri, D.; Guerra, F.; Roselli, C. J. Radioanal. Nucl. Chem. 1998. 230(1-2), 21-28. (40) Truscott, J. B.; Jones, P.; Fairman, B. E.; Evans, E. H. Anal. Chim. Acta. 2001. 433(2), 245-253.
23
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
405 406 407 408 409
(41) Kim, C. S.; Kim, C. K.; Lee, J. I.; Lee, K. J. J. Anal. At. Spectrom. 2000, 15, 247-255. (42) Jäggi, M.; Röllin, S.; Alvarado, J. C.; Eikenberg, J. Appl. Radiat. Isotopes 2012, 70(2), 360-364. (43) Varga, B.; Tarjan, S. (2008). Appl. Radiat. Isotopes 2008, 66(2), 265-270.
24
ACS Paragon Plus Environment
Page 24 of 28
Page 25 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
Figure caption
Figure 1 Analytical method for Pu determination in soil and sediment samples by ICPMS
25
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Figure 1
26
ACS Paragon Plus Environment
Page 26 of 28
Page 27 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48
Analytical Chemistry
Table 1 Decontamination factors (DFs) of U for soil and sediment samples reported in literature and this study Separation methods
Co-precipitation
Sample type
DF (U)
References
Dowex 1X8
No co-precipitation
soil
1-10 × 104
Muramatsu et al.14
AG 1X8 + AG 1X8
No co-precipitation
sediment
1.4 × 104
Zheng et al.37
AG 1X8 + AG MP-1M
No co-precipitation
sediment
2 × 106
Bu et al.18
TEVA
No co-precipitation
soil and sediment
2.6 × 103
Godoy et al.17
TEVA
Fe(OH)3+ Fe(OH)2
soil
> 104
Qiao et al.15
TEVA+DGA
Fe(OH)3/Ti(OH)3 + LaF3
soil
> 106
Maxwell et al.33
AG 1X4 + TEVA
Fe(OH)3+ Fe(OH)2
soil and sediment
1-100 × 103
Xu et al.25
TEVA+UTEVA+DGA
CaF2/LaF3
soil and sediment
1.6 × 107
This study