Microfluidic Deans Switch for Comprehensive Two-Dimensional Gas

Jan 24, 2007 - A custom electronic circuit was developed to control the modulator. The circuit allowed the modulation sequence to be initiated by the ...
0 downloads 7 Views 195KB Size
Anal. Chem. 2007, 79, 1840-1847

Microfluidic Deans Switch for Comprehensive Two-Dimensional Gas Chromatography John V. Seeley,*,† Nicole J. Micyus,† Steven V. Bandurski,† Stacy K. Seeley,‡ and James D. McCurry§

Department of Chemistry, Oakland University, Rochester, Michigan 48309, Department of Science and Mathematics, Kettering University, Flint, Michigan 48504, and Agilent Technologies, Inc., 2850 Centerville Road, Wilmington, Delaware 19808

A microfluidic Deans switch was used as a comprehensive two-dimensional gas chromatography (GC×GC) modulator. The simplicity and wide temperature range of the Deans switch make it a promising alternative to existing modulation techniques. However, the Deans switch is a low duty cycle modulator; that is, it samples only a small portion of the primary column effluent. Like all low duty cycle modulators, the Deans switch produces inconsistent transfer of components from the primary to the secondary column if the primary peaks are undersampled. Theoretical simulations and experimental studies show that the relative standard deviation (RSD) of the fraction of material transferred from the primary column to the secondary column is less than 1% if the modulation ratio is greater than 2.5. But the RSDs increase rapidly as the modulation ratio is decreased below 2.5. Deans switch GC×GC was validated by analyzing the aromatic content of gasoline. A fast analysis (