Preparative high-performance liquid chromatography - ACS Publications

J. J. DeStefano and J. J. Kirkland. Anal. Chem. , 1975, 47 (13), pp 1193A–1204a. DOI: 10.1021/ac60363a044. Publication Date: November 1975. ACS Lega...
2 downloads 0 Views 2MB Size
Instrumentation

J. J. DeStefano' arid JI. J. Kirkland2 Experimental Station E . I. du Pont de Nemours & Co. Wilmington, Del. 19898

Preparative HighmPerformanee Liqulid Chromato&aphy In Part I of this article ( 1 ) which was last month's I N S T R I M E N T A T I O N feature, the philosophy of preparative high-performance liquid chromatography was presented. In Part I1 the necessary experimental corlditions will be discussed. Optimizing the preparative system for sample capacity involves different opwating conditions than normally used for analytical separations. These differences are exemplified by comparing typical opera tinp parameters listed in Table I Columns

In preparative separations, column sample capacity usudly is increased

-I Biochemicals Ilepartmpnt Central Reiearch a i d 1)evelopment De p a r t men t

Table 1. Typical Operating Parameters for

LC

Column diameter Column packing

Mobile phase flow rate Mobile phase velocity Sample volume Detector sensitivity required Wt of injected sample

(pellicular) - 1 ml/min

0.5 cm/sec

2 liters are desirable because of the large volume of mobile phase used. Pumping systems with very high-pressure capabilit y usually are not required in prepara-

tive HPLC, since columns of larger particles (higher permeability) are used; pump ratings of 1000-3000 psi are adequate. The precision and accuracy of pumps are not critical, since analytical interpretation of the chromatogram usually is not required. Reciprocating and pneumatic (or hydraulic) amplifier pumps are preferred because of their capability for higher pumping rates and continuous solvent output. “Mini”-pneumatic amplifier pumps are particularly convenient and are relatively inexpensive. Some of the reciprocating pump systems which are available in commercial LC instruments can only deliver a maximum of -10 ml/min (the lower end of the desired preparative flow rate range). Although this limitation does not preclude their use for preparative applications, longer separation times often must be accommodated, Pulsations from certain pumps which disturb the detector baseline in analytical runs are not a serious disadvantage in preparative separations. Highly sensitive detectors usually are not required in preparative HPLC as solute concentrations generally are quite high. These high concentrations can cause problems, since frequently it is difficult to determine whether the overlapping of peaks is due to an overloaded column or a nonlinear detector response. For example, the trace for the ultraviolet photometric (UV) detector in Figure 6 suggests essentially no resolution of two components. However, the response of the less sensitive refractive index (RI) detector shows a good separation, indicating

ANALYTICAL CHEMISTRY, VOL. 47, NO. 13, NOVEMBER 1975

1199A

Introducing. . Not one. . not two. . but THREE! and Field Expandable

A UNIQUE NEW SERIES OF PROGRAMMABLE COMPUTING INTEGRATORS FROM COLUMBIA SCI ENTlFIC * Super-simple operation-operator

is guided step-

by-step by lighted instructions

“3 models, starting well below $3000 and field expandable *Sophisticated signal analysis produces accurate areas and/or heights even in complex chromatography Q

Performs all needed calculationsexternal standard, internal standard, normalization - with resDonse factors

. 1

* Provides multiple

programs so operator may easily select different sets of parameters for the various types of chromatographic runs

* Automatic calculation calibration tables

of rewonse factors d u s multiple

P

If you’ve been waiting for a full-function, programmable, computing integrator and at a sensible price, Columbia Scientific offers the “Supergrators”- three models, starting well below $3000.After many years in chromatography, Columbia Scientific engineers have produced the ultimate in an expandable integrator with true operational simplicity. For a complete brochure, price list and a demonstration, write or call today.

1200A

ANALYTICAL CHEMISTRY, VOL. 47, NO. 13, NOVEMBER 1975

3 MODELS Economical SUPER 1 Intermediate SUPER 2 Sophisticated SUPER 3

EXPANDABLE Start with the SUPER 1at well below $3000.,then “move up” as your budget permits to SUPER 2 or SUPER 3 . They’re field expandable and sensibly priced for every lab.

COLUMBIA SCIENTIFIC INDUSTRIES CORP. P.O. Box 9908, Austin, Texas 78766 (11950 Jollyville Road, 78759).Phone (512) 258-5191 TWX 910-874-1364 CIRCLE 39 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 47, NO. 13, NOVEMBER 1975

1201 A

Figure 6. Detector-overload effects (22)

that the capacity of the column has not yet been overloaded by the Sample. Although the UV detector is useful, the RI detector appears to be more generally suitable for preparative separations. Sometimes a combination of both detectors may be required for the accurate monitoring of peaks. The possible anomalies in using only a UV detector are illustrated in Figure 7 . Chromatogram A was obtained with a UV detector on 1 gram of an impure sample containing a pair of isomeric compounds. Preparative isolation of an apparent major peak 1, believed to be one of the desired components, gave 39% recovery of the injected sample. Approximately 34 and 23% by weight eluted before and after this peak, respectively, even though only very minor components were indicated later in the chromatogram. Rerunning the sample with both the RI and UV detectors produced chromatogram B. Collection of the major components detected by RI showed that the cross-hatched peaks 2 and 3 were the desired isomers, representing a total of about 66% by weight of the injected sample. Thus, the UV detector suggested peak 1 as a major component, although it was only a minor impurity with a retention time very similar to that of the desired constituents. Using both the UV and RI detectors increases the probability that the peaks of interest and possible interfering contaminants will be properly identified. Spectrophotometric detectors also are very useful for preparative separations. These devices are more versatile than single-wavelength UV photometers and have the unique capability to operate in the 210-230-nm region, 1202A

which allows the detection of many compounds not sensed a t longer wavelengths. Another advantage of spectrophotometric detectors is in the “detuning” of the wavelength from the solute absorption maximum, to decrease the sensitivity of the measurement and reduce the potential for detector overload. Other detectors for HPLC are available, but for preparative work they should be nondestructive of the sample. Transport (often called flame ionization) detectors are useful, since only a small fraction of the total column effluent is destroyed in the detection process. These devices respond to all compounds containing carbon in the reduced state. Unfortunately, present commercial devices are costly, inconvenient, and highly operatordependent. Other considerations in equipment for preparative HPLC are different from those normally required for analysis. Extra-column effects are less of a problem with large-diameter columns because of their relatively large internal volumes. Therefore, low dead-volume tubing and fittings are not as critical. Detectors should not be made with very narrow-bore tubing which severely limits the flow of mobile phase, since any restriction causes considerable back-pressure a t the flow rates needed in preparative HPLC. RI and UV detectors designed with larger bore tubing (e.g., 0.75 mm i.d.) are

available from some manufacturers. Alternatively, a stream splitter on the exit of the large-diameter column can supply a small fraction of the effluent to an analytical detector. Manual collection systems usually are adequate for isolating one or a few separated components. However, automatic collectors are convenient when a large number of components must be collected in a single run. T o conserve mobile phase, solvent systems can be equipped with a switching valve which allows uncontaminated mobile phase to either be recycled back to the pump reservoir or collected for purification and reuse. Comprehensive systems have been devised so that the mobile phase automatically is repurified and returned to the solvent reservoir in a “closed-loop” operation (23,2 4 ) . Sample injection in preparative HPLC is an important but poorly understood operation. Introduction of the sample at a point in the center of the column bed inlet apparently results in greater column efficiency; however, all the column packing does not interact with the sample, and localized overloading is a problem. Loading the sample across the entire column cross section permits better use of the total column packing with reduced column overload. Further studies of sampling effects in largediameter columns are needed. Large sample aliquots can be conve-

Figure 7. Detector anomaly in preparative isolation Column same as in Figure 2, B except: mobile phase, 1.5 % isopropanol in pentane; flow rate, 10.7 ml/min: velocity, 0.06 cm/sec; sample, 15 ml of 17 mg/ml mixture in mobile phase; RI detector, 64 X RlUFS ( 12)

ANALYTICAL CHEMISTRY, VOL. 47, NO. 13, NOVEMBER 1975

Save la BUNDLE with

Mi~iI/lid

Zoom Stereomicroscope ~

WILD M-7A. 40.218

MiniVial

Standard

MiniVial in Holder

20-rnl V i a l

THE LIQUID SCIN TILLA TllON VIAL* THAT cuirs COUNTING COSTS. 0 0

Uses less scintillatcir than any other vial. Compatible with all liquid sci nt i I lat io n cou i i t i ng systems.

Minivials have a 7-ml capacity.

You insert the s a n e amount of radioactive material as in a 20-ml vial BUT use only one-third to onehalf of the scintillation chemicals. THE SAVINGS ARE DRAMATIC.

you'll really see something.

Made of high-density polyethylene, they will not leak toluene or b u l g e , , , even w e e t s later. They have a lower background and higher efficiency than glass. Reusable vial holder.; make MiniVials compatible with all liquid scintillation counters.

You'll see exceptionally crisp, completely flat images. You'll see a superbly crafted Swiss instrument with large field diameters, ample working distance, and convenience like never before, plus stands for all requirements. Field diameter at 6x is 36.0mm;at31 x,6.6mm.Total powerrange:3xtol24x. Photomicrography? Simply insert the phototube with trinocular head under the binocular and attach the 35mm magazine (shown in photo). Or use a Polaroid@Pack Adapter. Select depth of field with built in double iris diaphragm, focus, observe and take your photo. Wild M-7A Zoom Stereomicroscope. Look into it.

EXTRA SP'ECIAL

INTRODUCTORY OFFEIRI 200 Minivials 12 Holders

$

150°

Send for your introductory offer TODAY. Immediate delivery.

For more detai Is, iask for Bulletin 95-11. N u c l e a r Associates Inc 'iJ

S

8

Write or call for Brochure M-7A.

Patent 3 888,745

NUC L E A IR ASSOCIATEIS, I N C . S u b s r d i a r r 0'

R j D I A T I O N - M E D I C A L PI7ODUCTS CORP

35 URBAN AVENUE, WESlBUF'Y N Y 11590 PHONE (516) 333 9 2 4 4

CIRCLE 165 O N READER SERVICE CARD

WILD HEERBRUOQ INSTRUMENTS, INC. FARMINGDALE, NEW YORK 11736 616-293-7400 WILD OF CANADA,881 LADY ELLEN PLACE, OTTAWA 3, CANADA WILD OF MEXICO,8.A. LONDRES 268, MEXICO 8, D. F.

CIRCLE 2 4 6 ON READER SERVICE CARD ANALYTICAL CHEMISTRY, VOL. 47. NO. 13, NOVEMBER 1975

1203A

SAFE

niently introduced with a valve without interrupting the flow of the mobile phase. Injection with a syringe normally is by the stop-flow technique whereby the sample is introduced a t atmospheric pressure; then pumping is started. Very large volumes can be delivered with a syringe by septum injection, but difficulties can occur (e.g., pieces of septum plugging the inlet). Alternatively, large sample volumes can be conveniently injected with a syringe through a high-pressure valve in a stop-flow mode (12). The volume of sample which can be introduced into the column will depend on column internal diameter, sample solubility, and the mobile phase-stationary phase combination. In a nonoverloaded column condition, this volume generally should not exceed about one-third of the volume of the earliest peak of interest. However, when the column is overloaded, this relationship no longer holds and larger volumes are tolerated. Samples of 5-20 ml or larger frequently can be used with a 100 X 2.3-cm i.d. column. I t is apparent that the equipment for preparative HPLC is not as critical as for analysis, and less sophisticated and generally lower-cost systems can be used satisfactorily. Conclusions

The Pipette for radioimmunoassay is the MLA Precision D because your hand never touches the disposable tip. Fully controlled, automatic detipping is done with a flick of the finger. The convenience of the Precision D sometimes overshadows the many serious reasons w h y MLA designed it. Here is a pipette for radioactive reagents, toxic or contaminated fluids that provides safety for the user. MLA Pipettes are available from selected Laboratory Supply Houses. The Pipetting Decision.,. .is worth your time. Use the Reader Service Number to send for our Pipette Information Pack.

Medical Laboratory Automation, Inc. 520 Nuber Ave., Mt. Vernon, N.Y. 10550 91 41664-0366

Existing knowledge of preparative HPLC techniques suggests that certain approaches different from those normally used for analysis are desirable for obtaining significant amounts of purified components: Use liquid-solid chromatography, if possible. Good analytical separations are important as “pilot” runs. Use wide-diameter columns for increased sample throughput per unit time. Columns up to a t least 2.3 cm i.d. provide excellent resolution. Use totally porous packings and higher stationary phase loadings to optimize sample capacity. Particles of 30-50 pm are a good compromise between column efficiency, operating pressure, and cost. For higher resolution with porous particles, use lower mobile velocities. For highest sample throughput per unit time, use the highest velocity permitting adequate resolution. Overload columns to increase sample throughput (but with decreased resolution). Isolation of “heartcut” portions of peaks will improve purity but decrease yield. If feasible, adjust the retention of the compound to be isolated to higher k’ values (e.g., k’ 5 5 ) to optimize both sample loading and column resolution. Use larger sample volumes of more dilute solutions to minimize column overload a t the inlet. Use continuous pumping systems

CIRCLE 150 ON READER SERVICE CARD

1204A

ANALYTICAL CHEMISTRY, VOL. 47, NO. 13, NOVEMBER 1975

(e.g., reciprocating or pneumatic amplifier pumps); pressures above 3000 psi rarely are required. Sensitive detectors usually are not necessary; dual UV and RI detectors and transport detectors generally are satisfactory. Samples are more conveniently introduced with a valve, but syringe injection can be used. Use low viscosity mobile phases and higher temperatures if sample permits. Acknowledgment

We are grateful for the useful discussions with P. E. Antle and the use of his chromatograms for Figures 2 and 7 (Part 11). The helpful comments of L. R. Snyder on this manuscript are much appreciated. Literature Cited

(1) J . J. DeStefano and J. J. Kirkland, Anal. Chem., 47,1103A (1975). (2) J. J. DeStefano. H. C. Beachell. J . Chromatogr. Sci., 8,434 (1970). (3) J. P. Wolf 111,Anal. Chem., 45,1248 1197.11.

(4)-E.Godbille and P. Devaux, J . C h r o m a togr. Sei., 12,564(1974). ( 5 ) E. Stahl, Ed., “Thin-Layer Chromatography,” 2nd ed., Springer-Verlag, New York. N.Y.. 1969. (6) F. E.Rickett, J . Chromatogr., 66,356 (1972). (7) J.F.K. Huber, Chimia (Aaran), Suppl. 24-35 (1970). (8) J . J . Kirkland, J . Chromatogr. Sei., 10, 593 (1972). (9) R. E. Majors, Anal. Chem., 44,1722 (1972). (10) J. J. Kirkland, in “Gas Chromatography 1972,”p 39,S.G. Perry, Ed., Applied Science Publ., Essex, England, 1973. (11) L. R. Snyder and J. J . Kirkland, “Introduction to Modern Liquid Chromatography,” Chap. 8, Wiley-Interscience, New York, N.Y., 1974. (12) P. E. Antle, E. I. du Pont de Nemours & Co., Wilmington, Del., unpublished studies. 1974. (13) Waters Associates, Technical Bulletin AN 72-120,1972, (14) J. J. Kirkland, Ed., ‘‘Modern Practice of Liquid Chromatography,” Wiley-Interscience, New York, N.Y., 1971. (15) L. R. Snyder and J. J. Kirkland, “Introduction to Modern Liquid Chromatography,” Wiley-Interscience, New York, N.Y., 1974. (16) J. J . Kirkland and L. R. Snyder, Manual, “Solving Problems with Modern Liquid Chromatography,”American Chemical Society, Washington, D.C., 1974. (17) L. R. Snyder, in “Modern Practice of Liquid Chromatography,” Chaps. 4,6,J. J. Kirkland, Ed., Wiley-Interscience, New York, N.Y., 1971. (18) E. Stahl, Chem. Ztg., 82,323 (1958). (19) H. Schlitt and F. Geiss, J . C h r o m a togr., 67,261 (1972). (20) L. R. Snyder, ibid.,92,223 (1974). (21) L. Rohrschneider, Anal. Chem., 45, 1241 (1973). (22) Waters Associates, Technical Bulletin D5028,July 1973. (23) W. H. Pirkle and R. W. Anderson, J . Org, Chem., 39,3901 (1974). (24) W. H. Pirkle and M. 8. Hoekstra, ibid.,p 3904.