Product Review: Flow injection Analysis; Quietly Pushing Ahead

Jul 1, 2002 - Experts on flow injection analysis. (FIA) agree on one thing: The tech ... ments in the data-handling software, says Jason Gray at OI ...
0 downloads 6 Views 73KB Size
product review

Flow Injection Analysis: Quietly Pushing Ahead The field and the applications are growing, even if the technique still seems to be overlooked by some laboratory chemists. James P. Smith and Vicki Hinson-Smith xperts on flow injection analysis (FIA) agree on one thing: The technique is surprisingly under-used by laboratory chemists. Perhaps this is because automation was too expensive or took too long when FIA was introduced in the mid-1970s. But FIA enthusiasts say the game has changed since then. “This is a technique that allows chemists to easily automate and optimize well-developed wet chemical methods for routine laboratory use,” says Gary Christian of the University of Washington and previously with Alitea U.S.A., now called Global FIA, Inc. “You can even program an analyzer to switch from one analyte to another during the analysis of a batch of samples. . . . But FIA has never been a popular product of larger instrument companies, so smaller firms produce most of these analyzers.” FIA is at least fulfilling its potential in the sense that a wealth of literature describing a wide range of applications has been published. For example, FIA offers several advantages over the manual handling of solutions: It is computer-compatible, allows automated handling of solutions, and provides strict control of reaction conditions. Because sample handling is so versatile, FIA can serve as a front end to practically all spectroscopic and electrochemical detectors and to various clinical, environmental, and industrial assays. Biologists have used FIA in fluorescence microscopy for receptor

BROO SORENSEN C/O JAROMIR RUZICKA

E

and live cell studies and ultrafast drug screening and in flow cytometry for automated sample introduction and studies of fast cell–reagent kinetics. Other applications include real-time monitoring of chemical processes, automated renewal of the sensing layer in chemical transducers, and electrochemical methods, such as hydrodynamic voltammetry and ion-selective electrode measurements.

Advances in computerization, microfluidics, and hardware have facilitated the further development of new flow injection techniques. Especially noteworthy are the development of new on-line UV-digestion techniques, advanced online distillation techniques, and combined FIA–sequential injection analysis (SIA) techniques; the incorporation of ion-selective electrodes; and improve-

J U LY 1 , 2 0 0 2 / A N A LY T I C A L C H E M I S T R Y

385 A

product review

Table 1. Selected commercial flow analyzers1. Product

ASI/Eppendorf Variable Analyzers FIAlab-3500

FIAstar 5000

FloPro 4P, FloPro 9P

Company

AMKO Systems, Inc. 250 W. Beaver Creek Rd. Unit 6 Richmond Hill, Ontario Canada L4B 1C7 (905) 771-1444

FIAlab Instruments, Inc. 14400 Bel-Red Rd. Ste. 108 Bellevue, WA 98007-3926 (800) 963-1101

Foss North America 7682 Executive Dr. Eden Prairie, MN 55344 (952) 974-9892

Global FIA, Inc. P.O. Box 480 Sixth St. Fox Island, WA 98333 (253) 549-2223

URL

www.amkosystems.com

www.flowinjection.com

www.foss.dk

www.globalfia.com

Price

$10,000–$20,000

~$15,000

~$15,000–$45,000

~$10,000–$30,000

Applications

Chemical, environmental, biotech- Sample handling for FTIR, induc- Hundreds of methods—most comnology, pharmaceutical, and food tively coupled plasma MS, chemi- monly, environmental, nutrients, and beverage cal sensors, immunoassays; and industrial process water-quality or environmental analysis; proteomics

Hundreds of organic and inorganic analytes for on-line process, lab, and field analysis; industrial proprietary sample streams

Pumps and flow rates

Variable-speed, multichannel peristaltic pumps with various tubing sizes; flow rates typically 3–10 mL/min

Two 4-channel peristaltic pumps with variable speed and stand-by feature; 0.42–5.74 mL/min

Global FIA milliGAT, Cavro syringe, and Cavro peristaltic pumps; 6 nL/min–6 mL/min

Manifold and flow diameter

2–3 min for a typical analysis

0.7-mm flow diam

FIA, SIA, and zone fluidics; 0.80-mm flow diam

Sample throughput

60–120 samples/h

1–3 min/sample

Sample injection

120-position, random-access autosampler with separate rack for calibration and check standards

Gilson autosamplers

Typically 20 samples/h with 2- to Number of 100-µL samples samples and sample volume

20- to 400-µL sample volume

Variable sample number and volume

1–400 µL/s

Detectors

Include colorimeters, ISEs, conductivity sensors; external detector signals accepted if adaptor modules are used

Spectrometer-based absorbance; photomultiplier tube-based fluorescence; chemical; bioluminescence

Digital dual-wavelength photome- UV–vis absorbance, chemilumiter with automatic interchangeable nescence with liquid-core waveinterference filters guide cell, amperometric

Accessories

Selector and injection valves, peri- Heaters, autosamplers, additional staltic pump, mixing/reaction man- pumps and valves ifold, and master controller module with detector usually included; autosampler and printer available

All tubing, injection loops and coils, and gas diffusion membranes; prepacked cadmium reductor columns; pump tubes are pre-cut; reagent and waste bottles

Components and supplies; sampleconditioning systems, including membrane sampling devices and self-cleaning sample-filtering system

Special features

Flexible systems allow user to LOV manifold designed to miniachange processes or analytical turize FIA, SIA, and BIA sequences easily; systems can be fitted with various interchangeable detectors

Up to 3 channels; start up in