Rapid and automated detection of six contaminants in milk using a centrifugal microfluidic platform with two rotation axes Yiqi Chen, Yunzeng Zhu, Minjie Shen, Ying Lu, Jing Cheng, and Youchun Xu Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.9b01998 • Publication Date (Web): 24 May 2019 Downloaded from http://pubs.acs.org on May 25, 2019
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Rapid and automated detection of six contaminants in
2
milk using a centrifugal microfluidic platform with two
3
rotation axes
4 5
Yiqi Chen,†,‡ Yunzeng Zhu,†,‡ Minjie Shen, †,‡ Ying Lu†, Jing Cheng, †,‡,§ and Youchun Xu *,†,‡,§
6 7
†
8 9
‡
State Key Laboratory of Membrane Biology, Department of Biomedical Engineering, School of
Medicine, Tsinghua University, Beijing 100084, China. Collaborative Innovation Center for Diagnosis and Treatment of Infectious Diseases, Hangzhou
310003, China.
10
§ National
11 12
*Corresponding
13
E-mail address:
[email protected] (Y. Xu).
14
ABSTRACT: Antibiotic residues and illegal additives are among the most common contaminants in milk and
15
other dairy products, and they have become essential public health concerns. To ensure the safety of milk,
16
rapid and convenient screening methods are highly desired. Here, we integrated microarray technology into a
17
microfluidic device to achieve rapid, sensitive and fully automated detection of chloramphenicol, tetracyclines,
18
enrofloxacin, cephalexin, sulfonamides and melamine in milk on a centrifugal microfluidic platform with two
19
rotation axes. All the liquid reagent for immunoassay were pre-stored in the reagent chambers of the
20
microdevice and can be released on demand. The whole detection can be automatically accomplished within
21
17 min, and the limits of detection were defined as 0.92, 1.01, 1.83, 1.14, 1.96 and 7.80 μg/kg for
22
chloramphenicol, tetracycline (a typical drug of tetracyclines), enrofloxacin, cephalexin, sulfadiazine (a typical
23
drug of sulfonamides) and melamine, respectively, satisfying the national standards for maximum residue
24
limits in China. Raw milk samples were used to test the performance of the current immunoassay system, and
25
the recovery rates in the repeatability tests ranged from 80% to 111%, showing a good performance. In
26
summary, the immunoassay system established in this study can simultaneously detect six contaminants of
27
four samples in a fully automated, cost effective and easy-to-use manner, and thus has great promise as a
28
screening tool for food safety testing.
29
Corresponding Authors
30
*E-mail:
[email protected] Engineering Research Center for Beijing Biochip Technology, Beijing 102206, China
author at: Department of Biomedical Engineering, Tsinghua University School of Medicine, Beijing 100084, China. Tel.: +86 10 62796071.
1 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
31
Page 2 of 25
Introduction
32
Food safety has always been an issue of concern, and several recent food safety incidents
33
have stimulated the demand for the screening of the contaminants in milk 1. Among various
34
contaminants in dairy products, antibiotic residues are of particular interest due to their
35
frequent use in animal husbandry for infection treatment and growth promotion 2. Antibiotic
36
residues can cause severe damage to human health, including allergic reactions, drug
37
resistance, and enteric flora disturbance
38
typical illegal additive as well, because it artificially boosts the crude protein percentage.
39
Melamine and cyanuric acid mixtures can form insoluble crystals in the kidneys of babies 5.
40
Due to the severity of health risks caused by these contaminants in milk, many nations and
41
organizations have set strict regulations on residue limits for antibiotic residues and illegal
42
additives in milk. To screen these contaminants, the establishment of rapid, sensitive and
43
high-throughput detecting tools are necessary.
3,4.
In some developing countries, melamine is a
44
Several methods are currently available for the detection of antibiotic residues and
45
melamine, such as microbiological assays, high performance liquid chromatography (HPLC)
46
and immunoassays. Among these methods, microbiological assays are inexpensive but time-
47
consuming and limited in specificity and sensitivity 6. In particular, they cannot be used to
48
detect melamine. Chromatographic techniques are widely used in central laboratories 7-9 due
49
to their high sensitivity and specificity. However, they have some basic requirements,
50
including complex sample preparation, high-cost instrumentation, and skilled personnel.
51
Immunoassay-based methods, for instance, enzyme linked immunosorbent assays (ELISAs),
52
have been successfully applied to detect antibiotic residues and melamine
53
conventional ELISAs are labor-intensive and time-consuming, making them not suitable for 2 / 25
ACS Paragon Plus Environment
10-12.
But
Page 3 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
54
high-throughput screening.
55
Compared to conventional ELISAs, protein microarray is an excellent tool for the
56
simultaneous and high-throughput measurement of multiple analytes. In our previous studies
57
13-15, protein microarray had been utilized to screen and quantify the prohibited drugs in serum
58
and urine, as well as the antibiotic residues from animal tissue, and the reliability and accuracy
59
of the system was confirmed by thousands of real samples. In other studies, protein
60
microarrays were also proved as successfully screening tools
61
technology has its disadvantages as well. The operation of conventional microarray-based
62
immunoassay is sophisticated and time-consuming. To overcome these shortcomings, several
63
platforms have been developed 18-20 to simplify manual operation and shorten detection time.
64
Some of them are still semi-automatic 18, while the fully automated platforms that rely on
65
valves and pumps to control the external reagents 19-20 are still bulky and inconvenient to use.
66
Therefore, the development of a rapid, compact, automated and easy-to-use microarray-
67
based immunoassay system is still highly desired.
68
In our latest study
21,
16-17.
However, microarray
a centrifugal microfluidic platform with two rotation axes was
69
constructed, and the basic liquid operations including liquid transport, sequential release, and
70
mixing can be implemented in a simple manner, providing a new strategy for complex liquid
71
manipulation. In the present work, we retrofitted this centrifugal platform and designed a
72
matched microdevice with protein microarray for simultaneous screening and quantification
73
of antibiotic residues (chloramphenicol (CAP), tetracyclines (TCs), enrofloxacin (ENR),
74
cephalexin (CEX), sulfonamides (SAs)) and melamine (MEL) in milk samples. Liquid reagents
75
can be stored in the microdevices and released on demand, which significantly increases the 3 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
76
practicality of the platform and making the fully automated immunoassay possible. A compact
77
fluorescence detection system was designed and integrated into the platform for signal
78
recording, and four microdevices can be simultaneously processed, demonstrating its ability
79
for multiplexed target detection of multiple samples. Overall, the centrifugal immunoassay
80
system proposed in this study can successfully detect six contaminants in four milk samples
81
within 17 min, offering a great potential for rapid food safety testing.
82
Experiment
83
Reagents and materials
84
All the antibiotic standards (CAP, TCs, ENR, CEX, and SAs) were purchased from TargetMol
85
(Boston, MA). TCs and SAs are two groups of broad-spectrum antibiotics, while tetracycline
86
(TCN) and sulfadiazine (SDZ) are typical drugs of the TCs and SAs, respectively (Table S1). BSA-
87
SDZ and its antibody (Ab) were used to detect SAs, and BSA-TCN and its Ab were used to detect
88
TCs. The melamine standard was purchased from Dr. Ehrenstorfer GmbH (Augsburg,
89
Germany). Six types of target molecule conjugates (CAP-BSA, TCN-BSA, ENR-BSA, CEX-BSA,
90
SDZ-BSA, and MEL-BSA) and their corresponding murine monoclonal Abs were purchased
91
from the Huaan Magnech Bio-Tech Company Corp. (Beijing, China). BSA conjugates were used
92
as substrates to print the microarray. Stock solutions of the six contaminants were dissolved
93
in methanol or acetonitrile (analytical reagent grade) purchased from Beijing Chemical
94
Reagednts Company (Beijing, China) and stored at -20°C. Cy3-labeled goat anti-Mouse IgG was
95
purchased from Sigma-Aldrich (Shanghai, China), and Cy3-labeled BSA (Cy3-BSA) was
96
purchased from Bioss (Beijing, China). IgG produced in mice was purchased from Genia
97
(Beijing, China). The printing buffer was purchased from CapitalBio Corporation (Beijing, 4 / 25
ACS Paragon Plus Environment
Page 4 of 25
Page 5 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
98
China).
99
Glass slides were purchased from Gous Optics (Shanghai, China) and modified with epoxy
100
groups by CapitalBio. Ab solutions were prepared by serial dilution in 0.01 mol/L PBS (pH =
101
7.4), purchased from Solarbio Science & Technology Corp. (Beijing, China). PBST (PBS with 0.05%
102
Tween-20) was used as the washing buffer. Water in the experiments was purified with a Milli-
103
Q system (Millipore, Beijing, China).
104
Instrumentation
105
The immunoassay system was developed based on a centrifugal microfluidic platform
106
with two rotation axes, and the structure of the system is shown in Figure 1. The main stage
107
was driven by a servo motor with its driver (YZ-ACSD608, Yizhi Technology, Shenzhen, China).
108
Four micro-servo motors (RB-15PG, ALSRobot Technology, Harbin, China) were installed on
109
the main stage to precisely control the angular position of each microdevice on demand. Both
110
the power supply and the signal transmission were implemented through an electric slip ring
111
(MOFLON, Shenzhen, China). The pulse signals to control the micro-servo motors were
112
generated by a microcontroller (STC89C52) according to the instructions from the computer.
113
To fix the microdevices, four plastic carriers fabricated by 3D printing (WeNext Corp.,
114
Shenzhen, China) were mounted onto the micro-servo motors. The platform can be controlled
115
by a laptop with a custom-made LabVIEW program (National Instruments, Austin, TX).
116
For fluorescence detection, a green LED (530 nm, LUXEON, Alberta, Canada) with
117
corresponding lenses was used to excite the fluorescent dye. An industrial CCD (MV-BS30U,
118
CREVIS, Yongin-si, Korea), a microscope objective lens (MRH00100, Nikon, Shanghai, China),
119
and an optical filter (BD570, BODIAN Optical, Beijing, China) were used to capture the images.
5 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
120
Design and fabrication of the microdevice
121
As shown in Figure 2, the microdevice is composed of four layers, including a pressure
122
sensitive adhesive tape layer (Youbisheng Adhesive Product Corp., Hangzhou, China), a
123
polymethyl methacrylate (PMMA) layer, a double-sided adhesive layer (QL-9970-025, Wuxi
124
Bright Technology, Wuxi, China), and a glass layer. Three metal pins were inserted into the
125
microdevice as part of the three pillar valves.
126
The structure of the PMMA layer is shown in Figure 2A, which was machined by a
127
computerized numerical control (CNC) milling machine (Hongyang Laser Co., Ltd., Beijing,
128
China). A sample chamber and three reagent storage chambers (two chambers for washing
129
buffer and one chamber for secondary antibody solution) are connected to the reaction
130
chamber by microchannels. There is a pillar valve (Figure S1) located between each reagent
131
storage chamber and the reaction chamber. The pillar valve consists of a metal pillar and a
132
through-hole of the PMMA layer. The metal pillar is embedded in the through-hole of PMMA
133
layer, and the top of the connecting chamber is sealed by the PSA layer. The connecting
134
chamber (4 mm in length, 4 mm in width, 0.5 mm in depth) was machined with a spherical
135
cutter to get a concave bottom with rounded corner so that the PSA layer can be deformed to
136
seal it perfectly. The through-hole and the pillar is under interference fit to ensure no liquid
137
leakage in chip using. Once the microdevice is placed on the carrier, the pillar valves will be
138
opened because the metal pins are lifted up by the pillars on the carrier (Figure S1). Thus, the
139
liquid reagent can be released under the centrifugation as needed.
140
Among all of the chambers in the PMMA layer, the reaction chamber is perforative, and
141
a small glass slide is fixed to the reaction chamber from the backside using the double-sided
142
adhesive layer, with all the probes pre-printed and immobilized on the surface. 6 / 25
ACS Paragon Plus Environment
Page 6 of 25
Page 7 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
143
Analytical Chemistry
Printing and immobilization of the probes
144
A printing robot (Personal ArrayerTM 16, CapitalBio) was used to spot the probes onto the
145
specific area of the slides interfacing with the reaction chamber (Fig, 2A). The volume of each
146
probe was around 0.8 nL. After spotting, the epoxy group-modified glass slides were dried
147
under vacuum at 37℃ overnight. The epoxy groups of the slides would react with the free
148
amino groups of the protein, covalently linking the protein to the glass substrate.
149
Two types of microdevices were contact printed. First, different concentrations of each
150
BSA-antigen (Ag) solution were contact printed on the glass slides with six replicates to
151
optimize the spotting concentrations (Figure S2A). Second, once the spotting concentrations
152
of the Ags were determined, the six BSA-Ags of the optimal concentration and the four control
153
probes (printing buffer and BSA solution as negative controls, Cy3-BSA as a printing control,
154
and mouse-IgG as a positive and hybridization control) were printed on the microdevices for
155
detection (Figure S2B).
156
Sample preparation and recovery experiments
157
Both spiked milk samples and raw milk samples were used in the experiments. Spiked milk
158
samples were used as standards to acquire the standard curves, and raw milk samples were
159
used to test the performance of the system. Commercial milk was purchased from local
160
supermarkets (Beijing, China) to prepare the standard samples, and it was determined to be
161
free of studied contaminants by HPLC-MS.
162
The milk samples were prepared using the following protocol 20: the milk sample (2 mL)
163
was properly mixed with 15 mL of 1% trichloroacetic acid and 3 mL of acetonitrile in a 50-mL
164
polypropylene test tube by vortexing for 1 min to precipitate the proteins in milk. Then, the
165
sample was centrifuged at 7000 × g for 2 min. After the layers were thoroughly separated, the 7 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
166
middle layer was taken for subsequent analysis.
167
The microdevice-based immunoassay
168
Before use, the channel and reaction chamber of microdevices were incubated with PBST
169
(30 μL) containing 2% BSA for 1 h to minimize non-specific binding of the Ags to the
170
microdevices. Then the washing buffers and the secondary antibody solution (12 μL each)
171
were added and sealed into the microdevice, which was then stored at 4°C before use.
172
For chip using, the middle layer of the processed milk sample (1 ml) was mixed with Ab
173
solutions (9 ml, six Ab solutions with the optimal concentrations) and loaded to the sample
174
chamber. Then, the microdevice was fixed onto the carrier at the original position (α = 0°).
175
After that, the automated operating procedure for immunoassay was performed as follows
176
(Figure 3, Figure S3): the platform started to rotate, and the mixture in the sample chamber
177
was centrifuged into the reaction chamber to perform the first affinity binding reaction. At
178
this time, the microdevice was driven to tilt around the original position (α1 = -37°, α2 = +15°)
179
continuously at a frequency of 1 Hz. This step was allowed to last for 200 s to accelerate the
180
reaction. Subsequently, the microdevice tilted to the pre-determined angular position (α =
181
+58°), and the waste was driven into the waste chamber. After the first affinity binding step,
182
three more steps followed: the first washing step, the second affinity binding step, and the
183
second washing step. During each step, the corresponding reagents (washing buffer Ⅰ, the
184
secondary antibody solution and washing buffer Ⅱ , respectively) in the chambers were
185
released to the reaction chamber sequentially at specified angular positions (α = -47°, -52°,
186
and -57°), while the microdevice tilted at the same frequency (1 Hz) for 200 s, 200 s and 300
187
s, respectively.
8 / 25
ACS Paragon Plus Environment
Page 8 of 25
Page 9 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
188
It was found that the signal intensity can be enhanced by increasing the reaction time.
189
However, too long time for affinity binding will prolong the whole detection time and thus
190
restrict the practicability of the system. Therefore, the reaction time was chosen as 200 s by
191
taking the above factors into consideration.
192
Data acquisition and analysis
193
Once the on-chip immunoassay was completed, the microdevice were illuminated by LED
194
and the images captured by CCD. The differences in the median fluorescence intensity (MFI)
195
between the sample and the blank (MFISample-Blank) were extracted by the LuxScan 3.0 program
196
(CapitalBio).
197 198
Data analysis was performed by using Eq. (1): 𝑀𝐹𝐼𝐼𝐶𝑎 =
100 ― 𝑎 100
× 𝑀𝐹𝐼0 (1)
199
where a (%) is the inhibition ratio of the immunoassay, and MFI0 is the MFI at zero
200
contaminant concentration. With the data extracted from the images, standard curves for the
201
six contaminants were plotted, and the concentration of each contaminant in real samples
202
was determined by referring to these standard curves.
9 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
203
Results and discussion
204
The efficiency of the microdevice-based immunoassay
205
No extra pumps or valves were used in the current immunoassay system, and all steps
206
were performed automatically after sample loading. Because the microdevices were tilting
207
continuously during the entire process of the immunoassay, both the reactions and the
208
washing steps were performed in a dynamic manner. As shown in Table S2, with the similar
209
reaction results generated, the current immunoassay system consumes less time for affinity
210
binding and washing compared to the conventional protein chip. With the accelerated
211
reaction efficiency induced by continuously tilting of the microdevice, the entire duration of
212
detection for the immunoassay can be shorten to as less as 17 min.
213
The optimization of the concentrations of Ags and Abs
214
The concentrations of BSA-Ags and monoclonal Abs are closely related to the sensitivity
215
of an immunoassay. On one hand, the hybridization signals should be strong enough so that
216
wide linear ranges of the standard curves can be obtained. On the other hand, the Ag
217
concentrations should be kept as low as possible to lower the cost. As shown in Figure S4,
218
when the concentrations of the Ab remained constant, the fluorescence intensities increased
219
as a function of the printing concentration of BSA-Ags. Similarly, for a specific concentration
220
of BSA-Ag, the fluorescence intensities increased with the concentrations of Ab.
221
Considering both principles collectively, the spotting concentrations of CAP-BSA, TCN-BSA,
222
ENR-BSA, CEX-BSA, SDZ-BSA and MEL-BSA were determined as 0.6, 0.6, 0.2, 0.4, 0.6 and 0.4
223
mg/mL, respectively, and the optimal concentrations of the corresponding Abs were 2.0, 2.0,
224
5.0, 2.0, 1.0 and 2.0 μg/mL, respectively. For the concentration of the secondary antibody
225
solution, the increasing of it facilitates the signal intensity but the background of chip is also 10 / 25
ACS Paragon Plus Environment
Page 10 of 25
Page 11 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
226
enhanced. Therefore, an appropriate dilution ratio for the fluorescent labeled secondary
227
antibody was chosen as 1:400.
228
The determination of the specificity
229
The specificity among the BSA-Ags and their Abs was evaluated by Ag-Ab cross-reactivities
230
(CRs). Apart from the corresponding Ag-Ab at 100%, all of the other CRs were calculated using
231
the ratio of MFIs between the two hybridization signals. For example, the CR between MEL-
232
BSA and CAP-Ab was calculated by using Eq. (2): 𝑀𝐹𝐼(𝑀𝐸𝐿 ― 𝐵𝑆𝐴) 𝑎𝑛𝑑 (𝐶𝐴𝑃 ― 𝐴𝑏)
233
CR(%) = 𝑀𝐹𝐼(𝑀𝐸𝐿 ― 𝐵𝑆𝐴) 𝑎𝑛𝑑 (𝑀𝐸𝐿 ― 𝐴𝑏) × 100% (2)
234
As shown in Figure 4, all of the CRs were lower than 5%, suggesting high specificity of the
235
interaction between the six Ags and their corresponding Abs 22.
236
The reactivity of Abs to free antibiotics may be slightly different from the reactivity of Abs
237
to BSA-Abs, and the former method is a common way to evaluate the specificity between the
238
Abs and Ags 22. In our experiments, no serious cross-reactivity was found between different
239
Abs and Ags (both free antibiotics and BSA coupled antibiotics).
240 241
The CRs of the structurally related antibiotics were calculated according to Eq. (3): CR(%) =
𝐼𝐶50 (𝑆𝐷𝑍/𝑇𝐶𝑁) 𝐼𝐶50 (𝑎𝑛𝑎𝑙𝑦𝑡𝑒)
× 100% (3)
242
Cross-reactivity may be a valuable asset for protein microarray because it enables
243
detection of more analytes with limited number of antibodies. As shown in Table S1, the anti-
244
SDZ mAb exhibited high cross-reactivity against the other eight SAs, and the anti-TCN mAb
245
exhibited high cross-reactivity with the other three TCs due to their similar molecular
246
structures. Therefore, this immunoassay system can be used as a screening tool to determine
247
the total amount of SAs and TCs in milk rapidly and the other tools such as HPLC are needed
11 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
248
to further distinct the molecules with similar structures.
249
The calibration curves and sensitivity
250
Competitive immunoassays were used for the detection of six contaminants in milk. In
251
principle, the fluorescent signal intensity of corresponding probes decreases when the target
252
substance is present because the target molecules in the samples compete with the same
253
molecules immobilized on the array.
254
Spiked standard samples with a series of different concentrations for each contaminant
255
were prepared and analyzed following the operation protocol to obtain the standard curves.
256
The ratios of MFIs for each standard and the MFI for blank milk were used to generate
257
standard curves (Figure 5A). The limits of detection (LODs) were defined as the concentrations,
258
which were equivalent to 10% inhibition (IC10). The limits of the working range at the high and
259
low concentrations were defined as an inhibition of 80% (IC80) and 20% (IC20), respectively.
260
The performance of the detection, including the IC50, the LODs and the working ranges is listed
261
in Table 1.
262
The results showed that the LODs of this immunoassay system were as low as 0.92, 1.01,
263
1.83, 1.14, 1.96 and 7.80 μg/kg for CAP, TCN, ENR, CEX, SDZ and MEL, respectively, and all of
264
the working ranges were three orders of magnitude. According to the legislation of the China
265
Food and Drug Administration (CFDA), the maximum residue limits (MRLs) of TCs, CEX, SAs
266
and ENR are all 100 μg/kg, falling within the working ranges of our system. Compared to other
267
common methods, our system is rapid, automated and stable, with sufficient detection
268
performance (Table S3).
269
In the current study, the sample preparation is off-chip because some reagents and
12 / 25
ACS Paragon Plus Environment
Page 12 of 25
Page 13 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
270
operations is hard to be integrated on the chip. In the future, it is possible to realize “sample-in
271
and answer-out” detection by developing the sample pretreatment that is compatible with
272
on-chip immunoassay, which is the goal of our future work.
273
Analysis of raw milk samples
274
Raw milk samples (n = 35) provided by CapitalBio were used to evaluate the analysis
275
capabilities of the established system. The detection results were compared with data
276
acquired by HPLC-MS method, as listed in Table 2. Even though the raw samples were more
277
complex due to the existence of potential interfering substances, no false positives or false
278
negatives occurred in the experiments. The deviations of detection results between our
279
method and HPLC-MS method were all below 20%, demonstrating that our system can
280
simultaneously detect the six contaminants in raw milk samples and the results are reliable 23-
281
24.
282
microdevices to ensure the consistency of detections. In this immunoassay system, more
283
microdevices can be implanted on demand so that the calibration will be more convenient.
In practical application, the standard curves and calibration are needed for each batch of
13 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
284
Conclusion
285
In this study, an automated immunoassay system was constructed based on a centrifugal
286
microfluidic platform with two rotation axes. Simultaneous detection of six contaminants
287
(CAP, TCs, ENR, CEX, SAs and MEL) in milk on a single microdevice was achieved within 17 min,
288
and the system was able to process four samples at a time. The calibration experiments were
289
successfully performed, and the LODs and working ranges of all six contaminants were
290
determined. The applicability was tested with 35 raw milk samples, and reliable results were
291
obtained. In summary, our novel immunoassay system offers a rapid and automated method
292
for contaminants screening and quantification in milk, and this system has the potential to be
293
adapted to detect other proteins or small molecules to further aid food safety testing.
294
ASSOCIATED CONTENT
295
Supporting Information
296
Additional information as noted in text. This material is available free of charge via the Internet at
297
http:// pubs.acs.org.
298
Schematic of the pillar valve; The printing design of the microdevices; Work flow of the
299
microdevice; Optimization of the concentration of each Ag and primary Ab; Cross-Reactivity of
300
tetracyclines and sulfonamides using tetracycline (TCN) and sulfadiazine (SDZ) as standard curves,
301
respectively; Comparison of the reaction efficiency between the current immunoassay system and
302
the conventional protein chip; Comparison of the proposed method with other methods for the
303
detection of milk contaminant (TCN) (PDF).
304
Working video of the immunoassay (mp4).
14 / 25
ACS Paragon Plus Environment
Page 14 of 25
Page 15 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
305
AUTHOR INFORMATION
306
Corresponding Authors
307
*E-mail:
[email protected] 308
Acknowledgments
309
This study was supported by the National Key Research and Development Program of
310
China (2016YFC0800703), the National Natural Science Foundation of China (31500691), and
311
the Beijing Lab Foundation.
15 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
312
Figures
313
Figure 1. Schematic and photograph of the prototype of the immunoassay system (A-B). The
314
size of the prototype is 16 cm × 16 cm × 23 cm.
315 316
16 / 25
ACS Paragon Plus Environment
Page 16 of 25
Page 17 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
317
Figure 2. Structure of the microdevice. The size of the microdevice is 30 mm × 22 mm × 4.8
318
mm. (A) Top view of the PMMA layer. (B) Cross-section of the microdevice. It consists of a top
319
layer (pressure sensitive adhesive tape), a PMMA layer, a double-sided adhesive tape, a glass
320
slide with a pre-spotted probe array, and three metal pins (parts of the pillar valves). (C)
321
Photograph of the microdevice.
322 323
17 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
324
Figure 3. Work flow of the microdevice. The entire process can be divided into four parts (A-
325
D), each consisting of the following steps (1-4): 1) the microdevice is tilted to the original
326
position (α = 0°); 2) the microdevice is tilted to the specified angular positions (α = 0°, -47°, -
327
52°, -55°), and the reagent or washing buffer is driven into the reaction chamber; 3) the
328
microdevice is repeatedly tilted around the original position (α1 = -37°, α2 = +15°); and 4) the
329
microdevice is tilted (α = 65°), and the buffer is driven into the waste chamber. (A) The process
330
of the first affinity binding step. (B) The process of the first washing step. (C) The process of
331
the second affinity binding step. (D) The process of the second washing step.
332 333 18 / 25
ACS Paragon Plus Environment
Page 18 of 25
Page 19 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
334
Analytical Chemistry
Figure 4 Cross-reactivities among the six Ags and Abs.
335 336
19 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
337
Figure 5 The establishment of calibration curves. (A) Calibration curves of the six contaminants
338
in milk. (B) Representative fluorescence images of microarrays during the calibration
339
experiments with simulated standard samples with CAP. The concentrations of CAP are 1, 2,
340
5, 10, 20, 50, 100 and 200 μg/kg (from left to right and top to bottom).
341 342
20 / 25
ACS Paragon Plus Environment
Page 20 of 25
Page 21 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
343
Analytical Chemistry
Table 1. Detection performance of the system for the six contaminants (𝑥 ± s, n = 3). Analyte
LOD (μg/kg)
IC50 (μg/kg)
Working range (μg/kg)
MRL (μg/kg)
CAP
0.92
12.78
1.78-127.58
-
TCN
1.01
14.46
1.97-148.05
100
ENR
1.83
20.34
3.34-167.28
100
CEX
1.14
12.55
2.08-102.27
100
SDZ
1.96
23.80
3.66-211.39
100
MEL
7.80
266.96
18.86-4876.70
-
344
MRL stands for the maximum residue limit according to the legislations of CFDA;
345
“-” represents not allowed to exist.
21 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
346
Table 2. The analytical results of 35 raw milk samples. Sample No. S2 S5 S7 S11 S13 S14 S16 S20 S21 S28 S32 S35 Others
347
Page 22 of 25
Analytes CAP CEX MEL CAP TCN SDZ ENR CAP TCN ENR SDZ CEX
HPLC-MS (μg/kg)
Our system (μg/kg)
113.1 83.1 308.1 48.6 18.2 159.7 246.4 109.7 143.6 15.8 40.7 21.6
99.9 78.7 329.7 45.2 20.1 138.1 197.6 102.3 129.8 14.2 45.1 19.4 ND
Each was analyzed with 3 repeats; ND means not detectable.
22 / 25
ACS Paragon Plus Environment
Recovery rate (%) 88.3 94.7 107.0 93.0 110.4 86.5 80.2 93.2 90.4 89.9 110.8 89.8
Page 23 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
348
References:
349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366 367 368 369 370 371 372 373 374 375 376 377 378 379 380 381 382 383 384 385 386 387 388 389 390
[1] Song, Y.; Yu, H.; Lv, W. Risk analysis of dairy safety incidents in China. Food Control 2018, 92, 63-71. [2] Boeckel, T. P. V.; Brower, C.; Gilbert, M.; Grentfell, B.T.; Levin, S. A.; Robinson, T. P.; Teillant, A.; Laxminarayan, R. Global trends in antimicrobial use in food animals. Proc. Natl. Acad. Sci. U. S. A. 2015, 112, 5649-5654. [3] Fonseca, G. P.; Cruz, A. G.; Faria, J. A. F.; Silva, R.; Moura, M. R. L.; Carvalho, L. M. J. Antibiotic residues in Brazilian UHT milk: a screening study. Ciênc. Tecnol. Aliment. 2009, 29, 451-453. [4] Marshall, B. M.; Levy, S. B. Food Animals and antimicrobials: impacts on human health. Clin. Microbiol. Rev. 2011, 24, 718-733. [5] Hau, K. C.; Kwan, T. H.; Li, K. T. J. Melamine toxicity and the kidney. Am. Soc. Nephrol. 2009, 20, 245-250. [6] Nouws, J. F.; Van, E. H.; Loeffen, G. Schouten, J.; Keukens, H.; Smulders, I.; Stegeman, H. Suitability of the charm HVS and a microbiological multiplate system for detection of residues in raw milk at EU maximum residue levels. Vet. Q. 1999, 21, 21-27. [7] Karageorgou, E.; Armeni, M.; Moschou, I.; Samanidou, V. Ultrasound-assisted dispersive extraction for the high pressure liquid chromatographic determination of tetracyclines residues in milk with diode array detection. Food Chem. 2014, 150, 328-334. [8] Filigenzi, M. S.; Tor, E. R.; Poppenga, R. H.; Aston, L. A.; Puschner, B. The determination of melamine in muscle tissue by liquid chromatography/tandem mass spectrometry. Rapid Commun. Mass Spectrom. 2007, 21, 4027-4032. [9] Mahony, J. O.; Clarke, L.; Kennedy, R. O.; Lehotary, S. J.; Danaher, M. The use of ultra-high pressure liquid chromatography with tandem mass spectrometric detection in the analysis of agrochemical residues and mycotoxins in food - Challenges and applications. J. Chromatogr. A. 2013, 1292, 83-95. [10] Song, E.; Yu, M.; Wang, Y. HU, W.; Cheng, D.; Swihart, M. T.; Song, Y. Multi-color quantum dotbased fluorescence immunoassay array for simultaneous visual detection of multiple antibiotic residues in milk. Biosens. Bioelectron. 2015, 72, 320-325. [11] Ho, T. Y.; Chan, C. C.; Chan, K.; Wang, Y. C.; Lin, J. T.; Chang, C. M.; Chen, C. S. Development of a novel bead-based 96-well filtration plate competitive immunoassay for the detection of Gentamycin. Biosens. Bioelectron. 2013, 49, 126-132. [12] Sun, F.; Liu, L.; Xu, H. K. C. Development of ELISA for melamine detection in milk powder. Food Agric. Immunol. 2013, 24, 79-86. [13] Du, H.; Yang, W.; Xing, W. Su, Y.; Cheng, J. Parallel detection and quantification using nine immunoassays in a protein microarray for drug from serum samples. Biomed. Microdevices 2005, 7, 143-146. [14] Du, H.; Wu, M.; Yuan, G. Sun, Y.; Lu, Y.; Zhao, S.; Du, Q.; Wang, J.; Yang, S.; Pan, M.; Lu, Y.; Wang, S.; Cheng, J. Development of miniaturized competitive immunoassays on a protein chip as a screening tool for drugs. Clin. Chem. 2005, 51, 368-375. [15] Chen, A.; Wang, G.; Cao, Q.; Wang, Y.; Zhang, Z.; Sun, Y.; Wang, H.; Xu, C.; Zhou, Q.; Mitchelson, K. R.; Cheng, J. Development of an antibody hapten-chip system for detecting the residues of multiple antibiotic drugs. J. Forensic Sci. 2009, 54, 953-960. [16] Zuo, P.; Ye, B. C. Small molecule microarrays for drug residue detection in foodstuffs. J. Agric. 23 / 25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408 409 410 411 412
Food Chem. 2006, 54, 6978-6983. [17] Zuo, P.; Zhang, Y.; Liu, J. Ye, B. C. Determination of β-adrenergic agonists by hapten microarray. Talanta 2010, 82, 61-66. [18] Noroozi, Z.; Kido, H.; Peytavi, R.; Nakajima-Sasaki, R.; Jasinskas, A.; Micic, M.; Felgner, P. L.; Madou, M. J. A multiplexed immunoassay system based upon reciprocating centrifugal microfluidics. Rev. Sci. Instrum. 2011, 82, 64303. [19] Knecht, B. G.; Strasser, A.; Dietrich, R.; Märtlbauer, E.; Niessner, R.; Weller, M. G. Automated microarray system for the simultaneous detection of antibiotics in milk. Anal. Chem. 2004, 76, 646654. [20] Guo, H.; Zhou, X.; Zhang, Y.; Song, B.; Zhang, J.; Shi, H. Highly sensitive and simultaneous detection of melamine and aflatoxin M1 in milk products by multiplexed planar waveguide fluorescence immunosensor (MPWFI). Food Chem. 2016, 197(Pt A), 359-366. [21] Zhu, Y.; Chen, Y.; Meng, X. Wang, J.; Lu, Y.; Cheng, J. Comprehensive study of the flow control strategy in a wirelessly charged centrifugal microfluidic platform with two rotation axes. Anal. Chem. 2017, 89, 9315-9321. [22] Wang, Y.; Liu, N.; Ning, B. Simultaneous and rapid detection of six different mycotoxins using an immunochip. Biosens. Bioelectron. 2012, 34, 44-45. [23] Karageorgou, E.; Armeni, M.; Moschou, I. Ultrasound-assisted dispersive extraction for the high pressure liquid chromatographic determination of tetracyclines residues in milk with diode array detection. Food Chem. 2014, 150, 328-334. [24] Wang, G.; Zhang, H.; Liu, J. A receptor-based chemiluminescence enzyme linked immunosorbent assay for determination of tetracyclines in milk. Anal. Biochem. 2019, 564, 40-46.
24 / 25
ACS Paragon Plus Environment
Page 24 of 25
Page 25 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
413
Analytical Chemistry
For Table of Contents Only
414