A simple method for high-sensitivity determination of cosmogenic 35S in snow and water samples collected from remote regions Mang Lin, Kun Wang, Shichang Kang, and Mark H. Thiemens Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.6b05066 • Publication Date (Web): 03 Mar 2017 Downloaded from http://pubs.acs.org on March 14, 2017
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
A simple method for high-sensitivity determination of cosmogenic 35S in snow and
2
water samples collected from remote regions
3 4
Mang Lin1*, Kun Wang2,4,1, Shichang Kang2,3 and Mark H. Thiemens1
5 6
1. Department of Chemistry and Biochemistry, University of California San Diego, La Jolla,
7
California 92093, USA
8
2. State Key Laboratory of Cryosphere Sciences, Northwest Institute of Eco-Environment and
9
Resources, Chinese Academy of Sciences (CAS), Lanzhou 730000, China
10
3. CAS Center for Excellence in Tibetan Plateau Earth Sciences, Beijing 100101, China
11
4. University of CAS, Beijing 100049, China
12 13
* Email:
[email protected] 14
1
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 2 of 28
15
Abstract
16
Cosmogenic 35S is useful in understanding a wide variety of chemical and physical processes in
17
the atmosphere, the hydrosphere and the cryosphere. The 87.4-day half-life and the ubiquity of
18
sulfur in natural environments renders it an ideal tracer of many phenomena. Measurements of
19
35
20
of sample (>20 L) due to their high analytical activity background and low counting efficiency.
21
Here, we present a new set of snow/water sample collecting and handling procedures for high-
22
sensitivity determination of cosmogenic
23
Laboratory experiments using diluted
24
minute) showed a
25
from the true value. Using this method, we successfully measured
26
sample collected from a glacier on the Tibetan Plateau to be 47±7 mBq/L. Based on 35S activities
27
in 9 natural samples measured in this study, a first proof-of-concept approximation for age
28
determinations and source attributions was presented. This new method will provide a powerful
29
tool in studying 35S in small volumes of snow and water samples, especially those from remote
30
but climatically important regions such as the polar regions and the Tibetan Plateau and
31
Himalayas. The measurements are particularly important as the radioactive sulfur provides an
32
actual clock of glacial melting processes. With the growing rate of glacial loss, the need for
33
measurements from remote locations becomes all the more important.
S in snow and water samples are scarce as existing analytical methods require a large volume
35
35
35
S using a low-level liquid scintillation spectrometer.
S standards (with activities of 20 L) is required. Such large volumes of water are difficult to obtain in the field,
57
especially in the remote but climatically important region such as the polar regions and the
58
Tibetean Plateau and Himalayas. For example, the Tibetean Plateau and Himalayas region,
59
which is also widely known as the third pole because it contains the largest ice store outside the
60
polar regions, provides fresh water supply to nearly 40% of the world’s populution.17-18 The
3
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 4 of 28
61
glacier retreat and snow melting in this region play an crucial role in the hydrological cycle in
62
Asia but are not well understood because the harsh environment restricts comprehensive field-
63
based measurements. Therefore, a high-sensitive analytical method for measuring
64
volumes of water sample is required to assist in such quantification.
35
S in small
65
Recently, Brothers et al.19 modified the sample preparation method for the low-level liquid
66
scintillation counting technique and suggested to convert sulfur samples to aqueous
67
Na2SO4/H2SO4. It was shown that counting aqueous Na2SO4/H2SO4 generated a high counting
68
efficiency (up to ~95%) and a low background activity (~0.9 counts per minute, cpm). This
69
optimizition made 35S measurements in small sizes of samples possible. Even though this method
70
has led to a growing number of
71
insights into the vertical and horizontal transport processes in the atmosphere,20-24 the utility of
72
35
73
preparation method for natural water. Although our previous studies successfully measured
74
in small volumes (~3 L) of glacier snow and runoff samples collected from the Tibetan Plateau
75
and Himalayas using ion exchange resins,25-26 the reliability of the sample handling method was
76
not unambigously demonstrated and reported in detail, which resulted in uncertainties in data
77
interpretation and restricted further comprehensive studies. In this study, a series of control
78
experiments were conducted to evaluate the accuracy of previous measurements25-26 and develop
79
a new, simple and economical method for collections of snow, ice and meltwater samples and
80
the highly sensitive and accurate analysis of comosgenic 35S activity. The 35S activities in natural
81
samples collected from a glacier on the Tibetan Plateau were determined using our newly
82
developed method.
35
S measurements in aerosol samples, which provide new
S in hydrologic studies remains limited, mainly due to a lack of sample collection and
83
4
ACS Paragon Plus Environment
35
S
Page 5 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
84
Analytical Chemistry
Experimental Sections
85
Ultra-Low-Level Liquid Scintillation Spectrometer and Counting Efficiency. In this study, an
86
ultra-low-level liquid scintillation spectrometer (Quantulus 1220, PerkinElmer) was used to
87
quantify cosmogenic
88
Shape Analyzer and the Pulse Amplitude Comparator) on the Quantulus 1220 followed the
89
default setting for
90
decay of 35S (EMax = 167 keV) and 14C (EMax = 156 keV).5, 19 The 35S counting window was set as
91
channel 1-450 to minimize the possible influence of
92
sample was counted for 12 times (2-h counting for each cycle) and the average was reported.
93
Potential outliers, if any, were identified by the Dixon’s Q-Test (99% confidence level) and
94
rejected. In all measurements made in this study, no outlier was identified. Using the
95
propagation-of-error method of the standard environmental counting statistics,27 the net counting
96
error (95% confidence level) can be calculated as follows:
14
35
S activity. The setting of two software selectable parameters (the Pulse
C because of the significant overlap between the energy spectra of the β-
= 1.96 × (
14
C (with peak at channel ~485). Each
+ )
97
where S and B represent sample and background counting rates, respectively, and tS and tB
98
represent elapsed counting times at which sample and background counting rates were measured,
99
respectively. In this study, the long elapsed counting time (24-h = 2-h × 12) yields a small error
100
(e.g., 0.09 cpm for a sample with a counting rate of 2.0 cpm). Given the standard environmental
101
counting statistics also recommends to divide the total counting time into several equal cycles to
102
check variability of counting rates,27 we carefully checked results for 12 counting cycles (see
103
examples in Figure 1) and found that the standard variation of 12 counting rates (e.g., 0.16 cpm
104
for a sample with an average counting rate of 2.0 cpm) was greater than the net counting error
5
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 28
105
estimated previously. Therefore, the larger error (i.e., one standard deviation for 12 counting
106
cycles) was reported in this study. This error was further propagated in subsequent arithmetic
107
calculations (e.g. corrections for the background activity and the decay time).27-28
108
Standard or sample was prepared in aqueous form (Na2SO4/H2SO4) instead of solid form 133
Ra,
228
(BaSO4) to eliminate the potential contamination of other radionuclides (e.g.
110
and their daughter products) in BaCl2 reagents.8,
111
quantitatively transfer to a 20-mL plastic scintillation vial (Fisherbrand) and mixed with 10 mL
112
of scintillation gel (Insta-Gel Plus, PerkinElmer). Plastic scintillation vials had a lower
113
background activity (~0.15 cpm) than traditional glass vials which contain
114
The mixed cocktail was shaken and vibrated vigorously prior to loading into the ultra-low-level
115
liquid scintillation spectrometer.
19
Ba,
226
109
Ra
The pure Na2SO4/H2SO4 solution was
40
K (~0.45 cpm).19
116
The background and counting efficiency for various volumes of Na2SO4/H2SO4 samples are
117
summarized in Table 1. Similar to our previous study,19 the counting efficiency of this method
118
was greater than 90% if 1 mL of aqueous solution was used. However, such small volume of
119
solution was potentially problematic in natural sample analysis because as little as ~0.1 M of
120
SO42- can impede the formation of a stable microemulsion and cause phase instability due to the
121
divalent charge and relative size of divalent anions.8, 29 In addition, if the aqueous solution was
122
not completely clean, trace amount of organics from natural samples may also lead to color
123
quench problems in such a small volume of solution.29 Table 1 shows that the counting
124
efficiency dropped down to 79.0±0.4% when 2 mL of aqueous solution were used, but the
125
further losses of counting efficiency due to using larger sample volume (5 mL) were marginal.
126
Therefore, we diluted the solution to a larger volume (5 mL) to eliminate the potential problems
127
of phase instability and color quench. It should be noted that the sample should be counted as
6
ACS Paragon Plus Environment
Page 7 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
128
soon as possible after the sample-gel mixture was made, because we found that the counting
129
efficiency for 5 mL of aqueous solution dropped down to 2 mmol of sulfate
152
was added to cocktail. Based on the fitted polynomial curve (Figure 2b), an upper limit of 1.5
153
mmol of sulfate is recommended for this method. If the analytical goal is to measure samples
154
with low 35S SA, e.g. groundwater and seawater, converting sulfate to BaSO4 is required and a
155
method recently developed by Uriostegui et al.8 is recommended.
156
Standard Preparation. The
35
S activities in environmental samples, which depend on the
157
sample size and storage time, are usually less than 5 dpm. A H235SO4 standard (PerkinElmer),
158
which was stored separately in a locker at the Radiation Division of the Environment, Health and
159
Safety Department (EH&S) at the University of California San Diego (UCSD), was diluted to
160
laboratory sub-standards with comparable activities (< 5 dpm) in the EH&S. These sub-standards
161
were used to develop the sample extraction and purification method for
162
sensitive, accurate and economical way. To prevent any possible cross-contamination in
163
experiments, these sub-standards were stored in a different laboratory (instead of our wet
164
chemistry laboratory).
35
S analysis in a
165
Concentrating SO42- in the Field Using Anion Exchange Resins. Previous methods using
166
Amberlite IRA 400 anion exchange resins for large volumes of samples (>20 L) require the
167
acidification of samples using concentrated hydrochloric acid (HCl) and the exchange efficiency
168
enhancement using a pump or an industrial spinner.7-8 Such complicated concentration processes
169
are conventionally performed in the laboratory. In this study, because the aqueous
170
Na2SO4/H2SO4 method has a lower background activity and higher counting efficiency (Table
171
1), only ~3 L of water is required. We utilized a Bio-Rad AG1-X8 anion exchange resin
172
(analytical grade, 100-200 mesh, chloride form) to concentration sulfate, which adsorption
173
efficiency and recovery percentage for anions in fresh water samples is greater than 98%.30 In
8
ACS Paragon Plus Environment
Page 9 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
174
this study, the focus is on the recovery percentage of 35S. This concentration method is relatively
175
simple and straight forward and easily done in the field. It eliminates transporting large volumes
176
of water (~3 L per sample) from the field to the laboratory. This method is especially convenient
177
and economical for collecting samples from remote regions. Before using, the resin (3-4 mL)
178
was filled into a plastic column (Bio-Rad; i.d. 0.8 cm, length 4 cm) and rinsed by 5 mL deionized
179
(DI) water (18 MΩ·cm) for three times. Sulfate was collected by gravity dripping samples (or
180
laboratory sub-standards with known
181
capacity of the AG1-X8 anion exchange resin is 1.2 meq/mL, and therefore 3-4 mL of resin can
182
handle most snow and melted water samples.
35
S activities) through the resin column. The exchange
183
Elution of SO42- from Anion Exchange Resins. Sulfate trapped in resins was eluted by gravity
184
dripping 15 mL of 1 M hydrobromic acid (HBr) through the column in 5 mL increments. To
185
quantitatively collect all samples trapped in resins, 15 mL of DI water was dripped through the
186
column in 5 mL increments and collected after elution with HBr. Positive air pressure was
187
applied to inject residual eluent/water using a clean 60-mL syringe. The 30 mL of eluent
188
containing SO42-, NO3-, Cl- and Br- was collected by a 50-mL plastic centrifuge tube. After
189
stripping, excess Br-/Cl- (~15 mmol) must be removed because such high concentrations of ions
190
cannot be handled by the scintillation gel.
191
Purification Using Ag2O and Preparation of Aqueous SO42 for Counting. Two sets of
192
experiments were performed to remove excess Br-/Cl- ions in two different ways. In this set of
193
experiments (set 1), excess Br-/Cl- were precipitated as AgBr/AgCl by slowly adding Ag2O (~2
194
grams) in successive increments of ~0.4 grams. The pH of the solution was tested by pH paper to
195
ensure that Br-/Cl- completed precipitation and the solution turned neutral (pH≈7). The
196
AgBr/AgCl slurry was removed by a disposable vacuum-driven filtration system (with 0.22 um
9
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 10 of 28
197
Millipore Express Plus Membrane, Steriflip, Millipore) and the filtrate was collected by another
198
clean 50-mL centrifuge tube. An additional ~10 mL of DI water was used to wash any trace
199
amount of sample left in the original 50-mL centrifuge tube and the filter membrane into the new
200
50-mL centrifuge tube in successive increments of ~5 mL, bringing the sample volume to ~40
201
mL. This large volume of solution must be concentrated for further processing. It is noted that
202
H2SO4 reacts with Ag2O to form Ag2SO4,31 which is minimally soluble in water (0.83 g / 100 mL
203
at 25°C). To minimize the possible losses of
204
sample, 5 mL of 1000 ppm Na2SO4 solution containing no
205
solution was evaporated and dried in a clean oven (80°C) overnight. After drying, the sample
206
was re-dissolved by adding 2 mL of DI water to the 50-mL centrifuge tube and was subjected to
207
sonication for 1 hour.
35
S during handling small volume (~1 mL) of 35
S was added to the solution. The
208
The subsequent cleaning procedure followed the protocol developed by Brothers et al.19: the 2
209
mL of solution was passed through a polyvinylpyrrolidone (PVP) column and a Ag-cartridge
210
(Dionex OnGuard II) to remove organics and any remaining Cl- (or Br-), respectively, and was
211
collected by a 15-mL centrifuge tube. An additional ~8 mL of DI water was used to wash trace
212
amount of sample left in the 50-mL centrifuge tube, the PVP column and the Ag-cartridge into
213
the 15-mL centrifuge tube in successive increments of ~2 mL. The 15-mL centrifuge tube
214
containing ~10 mL solution was put in a freeze-dryer overnight. After freeze-drying, the sulfate
215
sample (white crystalline solid) was re-dissolved again by added 1 mL of DI water and was
216
sonicated for 1 hour. If any remaining dirty residue was observed in this step, ~2 mL of
217
hydrogen peroxide (H2O2) was added to remove organics and PVP-cleaning and freeze-drying
218
should be repeated. Following this cleaning procedure, the sample was ready to transfer to a
219
plastic scintillation vial for preparing the sample-gel mixture. To quantitatively transfer, the
10
ACS Paragon Plus Environment
Page 11 of 28
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
220
“empty” 15-mL centrifuge tube was washed with 1 mL DI water and transferred into the
221
scintillation vial and this procedure was repeated for four times. The sample-gel mixture was
222
prepared and sent to the Quantulus 1220 following the procedure described in the previous
223
section (Ultra-Low-Level Liquid Scintillation Spectrometer and Counting Efficiency).
224
Our results show that the recovery percentage of
35
S was ~90% in this set of experiments
225
(Table 2). It is noted that the sulfur carrier (Na2SO4) is important because without adding Na2SO4
226
the recovery percentage was less than 50% (i.e., the measured 35S was more than -50% deviation
227
from the true value) (set 0) (Table 2). Since our previous pilot study25 did not notice the
228
importance of the sulfur carrier, the measured
229
processing) may be underestimated. Another concern is the cost of Ag2O (Fisher Scientific,
230
current price: ~900 US$ for 100 grams). Previous studies used 15 mL of 3 M HCl as eluent to
231
strip NO3- and SO42- samples from AG1-X8 resins, which required ~6.5 grams of Ag2O (~60
232
US$) to remove excess Cl- for each sample.30 Even though 15 mL of 1 M HBr was used as the
233
eluent in this study, which can significantly cut down the required amount of Ag2O without
234
reducing the recovery percentage of anion because the relative selectivity of Br- for AG1-X8
235
resin is more than 2 times greater than Cl- (AG1, AG MP-1 and AG2 Strong Anion Exchange
236
Resin
237
rad.com/webroot/web/pdf/lsr/literature/9114_AG_1.pdf), the cost for each sample remained high
238
(~18 US$). Therefore, we developed another set of experiments to remove Br-/Cl- without using
239
Ag2O.
35
Instruction
S activities (without the sulfur carrier during
Manual:
http://www.bio-
240
Improved Purification Method Using H2O2. In this set of experiments (set 2), 5 mL of 30%
241
H2O2 was added to eluent to oxidize HBr/HCl to Br2/Cl2. The color of solution immediately
242
changing from colorless to yellow characterizes the formation of Br2. The H2O2 can also help to
11
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 12 of 28
243
remove organics in the solution. Following adding H2O2, 5 mL of 1000 ppm Na2SO4 containing
244
no
245
vapor pressure: