SP-ICP-MS Analysis of Metallic Nanoparticles in Environmental Samples with Large Dissolved Analyte Fractions Dina M. Schwertfeger, Jessica R. Velicogna, Alexander H. Jesmer, Richard P. Scroggins, and Juliska I. Princz Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.6b02716 • Publication Date (Web): 15 Sep 2016 Downloaded from http://pubs.acs.org on September 15, 2016
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Title: SP-ICP-MS Analysis of Metallic Nanoparticles in Environmental Samples
2
with Large Dissolved Analyte Fractions
3
Authors: D. M. Schwertfeger, Jessica R. Velicogna, Alexander H. Jesmer, Richard P. Scroggins,
4
Juliska I. Princz*
5
Affiliations: Biological Assessment and Standardization, Environment and Climate Change
6
Canada, 335 River Road South, Ottawa, Ontario K1V 1C7 Canada
7
* To whom correspondence may be addressed (
[email protected])
8 9
Juliska Princz, Ph.D.
10
Soil Toxicology Laboratory, Biological Assessment and Standardization
11
Environment and Climate Change Canada
12
335 River Road
13
Ottawa (Ontario) K1V 1C7
14
Email:
[email protected] 15
Telephone:
613-990-9544
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 2 of 23
2 16
Abstract
17
There is an increasing interest to use SP-ICP-MS to help quantify exposure to engineered
18
nanoparticles, and their transformation products, released into the environment. Hindering the
19
use of this analytical technique for environmental samples is the presence of high levels of
20
dissolved analyte which impedes resolution of the particle signal from the dissolved. While
21
sample dilution is often necessary to achieve the low analyte concentrations necessary for SP-
22
ICP-MS analysis, and to reduce the occurrence of matrix effects on the analyte signal, it is used
23
here to also reduce the dissolved signal relative to the particulate, while maintaining a matrix
24
chemistry that promotes particle stability. We propose a simple, systematic dilution series
25
approach where by the first dilution is used to quantify the dissolved analyte, the second is used
26
to optimize the particle signal, and the third is used as an analytical quality control. Using simple
27
suspensions of well characterized Au and Ag nanoparticles spiked with the dissolved analyte
28
form, as well as suspensions of complex environmental media (i.e., extracts from soils
29
previously contaminated with engineered silver nanoparticles), we show how this dilution series
30
technique improves resolution of the particle signal which in turn improves the accuracy of
31
particle counts, quantification of particulate mass and determination of particle size. The
32
technique proposed here is meant to offer a systematic and reproducible approach to the SP-ICP-
33
MS analysis of environmental samples and improve the quality and consistency of data
34
generated from this relatively new analytical tool.
35
36
INTRODUCTION
37
In order to evaluate the risk of engineered nanoparticles (ENPs) released into the environment, it
38
is necessary to perform quantitative assessments of both the exposure and effects to non-target
ACS Paragon Plus Environment
Page 3 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
3 39
organisms in environmental media.1,2 Single-particle inductively-coupled plasma mass
40
spectroscopy (SP-ICP-MS) has been identified as a valuable tool in this regard, given its ability
41
to detect and quantify the elemental composition of nanoparticles in suspensions at
42
environmentally relevant concentrations (i.e., ng L-1).3 However, SP-ICP-MS is not without its
43
shortcomings. One difficulty that arises in SP-ICP-MS, particularly for the analysis of
44
environmental media, occurs when dissolved analyte concentrations are large relative to the
45
analyte mass in the nanoparticles, which can lead to erroneous particle information.4
46
For SP-ICP-MS, samples must be sufficiently dilute such that, ideally, one particle at a time is
47
ionized in the plasma, which generates a burst of ions that enter the mass spectrometer and result
48
in a spike of the intensity signal, often referred to as pulse intensity. In contrast, the dissolved
49
analyte will produce a constant, continuous signal because of the steady stream of ions entering
50
the plasma. The intensity of a particle is essentially the pulse intensity after subtracting the
51
intensity of the dissolved or background signal, which can then be related to the mass of the
52
particle via calibration with well characterized nanoparticles, or via dissolved analyte calibration.
53
5
54
fraction of the analyte within the particle are known. Particle concentration (i.e., particles per
55
mL) is determined by counting the number of pulses detected and accounting for operational
56
factors that affect the rate of ions reaching the detector, such as sample flow rate and transport
57
efficiency (i.e., the fraction of nebulized particles that reach the plasma). An overview of SP-
58
ICP-MS theory and related calculations are provided in Laborda et al.,6 while an evaluation of
59
transport efficiency calculations is provided in Pace et al.5
Particle size is calculated from its mass if the analyte density, particle geometry and mass
60
One of the key aspects in processing SP-ICP-MS data is distinguishing the pulse intensity of
61
the particle from the dissolved signal. The intensity threshold used to discriminate the two
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 4 of 23
4 62
analyte forms is often determined using an iterative computation.4 An approximate threshold
63
limit is established, above which, events will be removed (e.g., events with intensities above the
64
mean intensity + 3δ). This initial calculation usually removes the majority of particle events. Of
65
the remaining events, the mean intensity is re-calculated and the procedure is repeated until
66
convergence (i.e., no more events can be removed). The final intensity threshold can be
67
converted to analyte mass, or particle volume, to determine the detection limits for particulate
68
mass or particle size, respectively. Some suppliers of SP-ICP-MS instruments are now supplying
69
software which performs these computations automatically for each sample.
70
A difficulty in the computation used to resolve particle intensity from the dissolved signal
71
occurs when there is a high level of dissolved analyte. When this occurs, the dissolved signal
72
overlaps the particle signal, resulting in the computation of a threshold intensity that is larger
73
than the intensities of some (or potentially all) the particles in a sample. For example, when an
74
aqueous suspension of Ag nanoparticles (60 nm nominal size) was spiked with increasing
75
concentrations of dissolved Ag, the number of particles detected decreased and the particle size
76
detection limit increased.4
77
There are many instances where environmental samples are expected to contain high levels of
78
the dissolved analyte. This may occur if the analyte of interest naturally occurs in the
79
environment in the form of dissolved ionic or complexed species (i.e., geologic or pedogenic in
80
origin), or if ENPs released into the environment undergo weathering and transformation to
81
dissolved species during their translocation to, or residence in, the media being tested.
82
Few researchers have developed reliable techniques to address the problem of analyzing
83
environmental samples with large dissolved analyte fractions. Hadioui and colleagues4
84
developed a technique whereby samples were first passed through an exchange column to
ACS Paragon Plus Environment
Page 5 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
5 85
remove much of the dissolved Ag+ species from samples prior to introduction into the ICP-MS.
86
This technique resulted in more accurate sizing and counting of the nanoparticles in their
87
suspensions and was further applied to the analysis of ZnO nanoparticles in wastewater.7 While
88
effective, the procedure adds a considerable amount of time to the analysis. This may be a
89
reasonable trade-off for suspensions containing NPs that are highly unstable and readily dissolve,
90
but may not be practical for all scenarios.
91
In this study, we investigate a sample dilution technique, which is ideally suited to the
92
analysis of suspensions that are relatively stable, such as soil and biosolids extracts. The
93
technique involves diluting samples in order to reduce the dissolved signal, thereby increasing
94
the ability to resolve the particle signal. While sample dilution for SP-ICP-MS is not a novel
95
concept, it is often done haphazardly from one sample to the next with the focus being to reduce
96
the occurrence of analytical coincidence (i.e., two or more particles entering the plasma at once).
97
Unlike standard ICP-MS analysis for which well-established technical guidance have been
98
published (e.g., EPA 6020A, 200.8), SP-ICP-MS is still in its infancy and lacks supportive
99
technical guidance and implementation of consistent quality controls, particularly for the analysis
100
of complex environmental matrices. We propose the use of a simple dilution series wherein each
101
successful dilution achieves a specific goal: (1) the capture of a dissolved analyte reading above
102
the limit of quantitation (LoQ) for the dissolved signal, (2) the capture of an optimized particle
103
signal, and (3) a quality control check that the particle analysis is truly optimized and free of
104
coincidence and matrix effects.
105
For this study, Au and Ag nanomaterials were selected. Citrate-stabilized Au nanoparticles
106
are highly stable and thus allowed us to explore the dilution technique without any interfering
107
effects of particle dissolution. Silver nanoparticles are one of the most widely used engineered
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 23
6 108
nanomaterials and their release from consumer products into the wastewater stream and
109
subsequent accumulation in sewage sludge represent a major translocation pathway.8 In North
110
America, processed sewage sludge (i.e., biosolids) is applied to agricultural lands representing an
111
important ecotoxicological exposure scenario. As such, governing agencies have initiated
112
assessments on the fate and effects of ENPs released into the terrestrial environment.9 The
113
overall goal of this study is to help further the development of a systematic, standardized
114
approach to SP-ICP-MS analysis of environmental samples (e.g., soil and biosolids) which is
115
expected to improve the quality of data generated by this relatively new analytical technique.
116 117
EXPERIMENTAL SECTION
118
Chemical. All Au and Ag nanoparticles used in this study were commercially purchased and
119
are listed in Table 1, along with suppliers’ specifications. Verification of these specifications was
120
carried out using dynamic light scattering (DLS), transmission electron microscopy (TEM) with
121
energy-dispersive X-ray spectroscopy (EDS), and SP-ICP-MS. All purchased suspensions were
122
placed in an ultrasonic bath (Branson 3510, 40 kHz, 130 watts, 6 L capacity) for at least 20 min
123
prior to analysis. In the case of the Ag nano-powder, a 15 g L-1 dispersion was prepared using
124
NanopureTM water (Barnstead 2-cartridge water filtration system, > 18.3 MΩ·cm, TOC < 10 µg
125
L-1) and placed in the ultrasonic bath for 40 min. The DLS analysis was conducted on ca. 20 mg
126
L-1 dilutions of initial suspensions using a MalvernTM Nano ZS Zetasizer with NIBS® optics
127
arranged at 175° for backscatter detection. Data was analyzed using cumulants analysis.10
128
Imaging of the same dispersion was conducted using an FEITM Tecnai G2 F20 TEM (FEI,
129
Hillsboro OR) at the Carleton University Nano Imaging Facility (Ottawa, Ontario). Details of the
ACS Paragon Plus Environment
Page 7 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
7 130
SP-ICP-MS analysis are provided in the relevant section below. The Au and Ag nanomaterials
131
are hereafter referred to by their nominal sizes.
132
Test Samples. Two types of samples were analyzed in this study: (1) suspensions of gold
133
nanoparticles (AuNP) or silver nanoparticles (AgNP) prepared in NanopureTM water, and (2) soil
134
extracts prepared from AgNP-contaminated soils. The nanoparticle suspensions were prepared
135
with 30 nm AuNP (NIST 8012), 60 nm AuNP (8013) or 30 nm AgNP (Ted Pella). For each
136
material used, a set of suspensions were prepared for which the particle concentration was kept
137
constant at approximately 105 particles mL-1, while increasing the dissolved Au or Ag
138
concentrations. Stock solutions of dissolved Au and Ag were prepared with an Au standard
139
solution (High-Purity Standards, 1000 µg mL-1 in 2% HCl) and silver nitrate (Sigma Aldrich
140
Reagent Plus®), respectively. Dissolved analyte concentrations ranged from 0.1 to 10 µg L-1 for
141
the suspensions containing the 30 nm particles, and 0.1 to 40 µg L-1 for suspensions containing
142
the 60 nm particles. Dilutions of test samples were always prepared fresh using NanopureTM
143
water and analyzed immediately. The soil extracts analyzed in this study were used to
144
demonstrate the dilution series approach to SP-ICP-MS analysis on real environmental samples
145
and were taken from a soil previously amended with AgNP-contaminated biosolids. Details of
146
the method development of the soil extraction technique are not critical to the understanding of
147
how to optimize the SP-ICP-MS analysis presented here in, and are outside the scope of this
148
study; however, further details have been included in a manuscript being prepared by the authors.
149
Soil preparation and the soil extraction procedures are briefly described as follows: An aliquot of
150
a 15 g L-1 suspension of the 40 nm AgNP (Nano-composix) was mixed into a slurry of municipal
151
biosolids, aged 3 d in sealed containers, then incorporated into an agricultural field soil (sandy
152
loam, pH 5.8) at a rate of 1 g biosolids per 100 g soil (dry weights) to obtain various Ag
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 8 of 23
8 153
concentrations. The soil samples used in this study included a soil sample spiked with 833 µg g-1
154
Ag taken from an earthworm assay (nAgEA), a soil sample spiked with 833 µg g-1 Ag taken
155
from a microbial assay (nAgMicr), and a soil sample spiked with 145 µg g-1 Ag taken from a
156
collembolan assay (nAgFC). Soils were extracted with tetrasodiumpyrophosphate (2.5 mM
157
TSPP) using a soil-to-reagent ratio of 1:10 (g:mL) and mixed on a rotary mixer for 30 min
158
followed by ultrasonication. Suspensions were allowed to settle overnight to allow large particles
159
to settle out. Sample suspensions were initially diluted 100x and all subsequent dilutions were
160
maintained with a 0.025 mM TSPP matrix.
161
SP-ICP-MS Instrumentation. SP-ICP-MS analysis was performed using a PerkinElmer
162
NexION 300D ICP-MS operating in single particle mode using the SyngistixTM Nano-
163
Application module (PerkinElmer Inc., Waltham, MA). Instrumental and data acquisition
164
parameters of the analysis are provided in Table 2. Time resolved signals were acquired using
165
fast-scanning mode, for which settling times have been reduced to negligible. Due to the fast
166
data acquisition rate, dwell times as low as 10 µs can be achieved. Previous research has shown
167
that dwell times of 50 to 100 µs are optimal for our instrumental set-up and for the analysis of
168
nanoparticles similar to ours.11 Transport efficiency was determined by the particle size method5
169
using the 60 nm NIST Au standard for reference; all calculations were performed within the
170
SyngistixTM module. Dissolved and particle calibrations were performed for
171
analytes, with dissolved calibrations ranging from 0.1 to 10 µg L-1. Dissolved Au and Ag
172
calibration solutions were prepared in 0.5% HCl and 1% HNO3 matrices, respectively. Gold
173
particle calibrations were performed using NIST Au 30 and 60 nm suspensions while for the Ag
174
particle calibrations, Ted Pella Ag 30 and 60 nm suspensions were used. For particle
175
calibrations, the intensity histograms were fitted to a log normal curve from which the most
ACS Paragon Plus Environment
197
Au and
107
Ag
Page 9 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
9 176
frequent area intensity was derived and related to particle mass (i.e., size). The dissolved and
177
particle calibrations produced comparable particle size data with mean particle intensities
178
derived from the dissolved calibration being on average 9% lower. The particle calibration
179
method was selected to process the results of this study, as the un-acidified matrix was more
180
consistent with that of the samples. SP-ICP-MS analysis requires that samples are sufficiently
181
dilute such that analytical coincidence is avoided. Based on our operating parameters, we
182
estimated that coincidence could be avoided if sample particle concentrations were ≤ 1.2 x 105
183
mL-1; therefore target particle concentrations for calibration suspensions were approximately 105
184
particles mL-1. The method detection limit (MDL) and LoQ for the dissolved analysis were
185
calculated as the mean reading of six method blanks (i.e., 0.025 mM TSPP solution) plus 3δ and
186
10δ, respectively.
187
Data Processing. Initial data processing was performed within the SyngistixTM Nano-
188
application module. Assuming a particle event in the plasma is approximately 350 µs,11 the 50 µs
189
dwell time used allows for multiple readings per particle. An algorithm is used to identify when a
190
succession of readings comprise a particle event from which peak area is calculated. In this way,
191
particle intensity is related to peak area. The determination of the intensity threshold used to
192
discriminate whether a signal is part of a particle event, or the dissolved (i.e., background) signal
193
is also performed within the SyngistixTM module using an iterative “mean + 3δ” computation,
194
similar to that described by Hadioui et al.4 The software records the number of particle events
195
and produces a frequency distribution based on particle intensity (i.e., peak area). Particle
196
intensity was related to particle mass (and thus size) using a particle calibration curve derived
197
from analyzing the 30 and 60 nm nanoparticle suspensions (Au or Ag) based on the assumption
198
that Au and Ag particles were spherical and pure Au0 or Ag0 with a density of 19.32 and 10.49 g
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 10 of 23
10 199
cm-3, respectively. Particle size distributions (PSD) were fit to a Gaussian smoothing function
200
from which mean and mode particle sizes were derived. It must be noted that while the particle
201
geometry and mass fraction may be known for well characterized particles (i.e., test samples
202
prepared from purchased suspensions), this is usually not the case for particles in environmental
203
media, for which a plasma event may include transformation products of various composition, or
204
naturally occurring particles. Therefore, for particles detected in environmental media, only the
205
mass of the analyte detected in a discrete particle (i.e., particulate Ag) is detectable, the size of
206
the actual particle is unknown. However, for the analysis of particulate Ag in our soil extracts,
207
we processed the data using a default spherical shape and mass fraction of 1, in order to calculate
208
an equivalent mono-metal particle size, denoted with the subscript “eqv” (e.g., PSDeqv), indicative,
209
not of actual particle sizes, but reference sizes used to aid in the understanding of the amount of
210
mass detected.
211
RESULTS AND DISCUSSION
212
Particle sizes of the Au and Ag nanomaterials were verified by DLS (Table S1, Figure S1,
213
Supporting Information) as well as by TEM/EDS and SP-ICP-MS (selected data, Figures S2 and
214
S3, Supporting Information). Measured particle sizes generally fell within the suppliers’ size
215
specifications, except for the Nanocomposix 40 nm Ag particles which showed a minor fraction
216
of particles smaller than 19 nm.
217
To investigate the effects of the SP-ICP-MS particle signal when samples contain large
218
concentrations of the dissolved metal, nanoparticle suspensions containing ca. 105 particles mL-1
219
were spiked with increasing amounts of dissolved metal and analyzed. Figure S4 (Supporting
220
Information) shows a portion of the time-resolved Au signal during analysis of the 30 nm
ACS Paragon Plus Environment
Page 11 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
11 221
suspensions with and without dissolved Au added. When analyzing the suspension spiked with
222
the dissolved Au, the steady dissolved signal drowns out most of the particle signals, resulting in
223
fewer particles detected of relatively large intensities, likely agglomerates or artifacts of
224
coincidence. Particle intensities were converted to particle size for which frequency histograms
225
were derived. Figure 1 shows the histograms generated from analyzing the 30 nm Au suspension
226
with increasing amounts of dissolved Au. Adding 0.5 µg L-1 dissolved Au had little effect on the
227
size distribution, while the addition of 1, 2 and 3 µg L-1 dissolved Au significantly lowered the
228
number of detectable particles while skewing particle sizes to increasingly larger sizes. This
229
trend is clearly illustrated with both 30 nm and 60 nm Au suspensions, in Figure 2, which shows
230
that at low concentrations of dissolved Au, the detected particle sizes and counts were accurate
231
and constant, but above a certain threshold, increasing concentrations of dissolved Au led to
232
erroneous increases to particle size and decreases in the number of particles detected. For the 30
233
nm Au suspensions, this threshold of dissolved Au is between 0.5 and 1.0 µg L-1, while for the
234
60 nm Au suspensions, it was between 10 and 20 µg L-1. When suspensions containing the
235
highest dissolved Au additions tested, i.e., 30 nm + 3 µg L-1 dissolved Au and 60 nm + 40 µg L-1
236
dissolved Au, were diluted such that the dissolved concentrations were below these thresholds,
237
particle size and particle counts were measured more accurately (Figure 2). The reason for this
238
improvement is demonstrated by comparing the particle histograms of the 30 nm + 3 µg L-1
239
dissolved Au sample before, and after dilution (Figures 1e and 1f). For the diluted sample, the
240
dissolved signal was sufficiently reduced to allow for the detection of the particles, thereby
241
resolving the sample’s particle size histogram. A similar trend was observed for suspensions
242
made from the 30 nm Ag nanoparticles for which a dissolved Ag threshold between 0.5 and 1.0
243
µg L-1 was also observed. (Figure S5, Supporting Information). These data show that the
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 12 of 23
12 244
dissolved analyte had a similar effect in magnitude on particle data for these similarly-sized
245
particles, despite being different elements.
246
When analyzing environmental samples, the particle sizes as well as particulate and dissolved
247
concentrations are generally unknown. Because the dissolved analyte fraction is an important
248
parameter for environmental assessments, we began with obtaining dissolved analyte readings
249
within the calibration range and above the LoQ of 0.1 μg L-1, diluting the samples if necessary.
250
For the three environmental samples analyzed in this study, only the nAgMicr extract required an
251
initial dilution (i.e., 100x) to achieve a valid dissolved reading (Table 3). With knowledge of the
252
dissolved fraction, the next analysis involved diluting the sample to achieve a dissolved analyte
253
concentration near the MDL of 0.05 μg L-1, thereby maximizing the ratio of particle signal /
254
background (dissolved) signal. This “optimal” dilution also needed to provide a particle
255
concentration below the (coincidence) criteria of 105 particles mL-1. When applying this dilution
256
technique to the analysis of soil extracts taken from ecotoxicological assays, a third dilution that
257
was 2x the optimal dilution proved advantageous as a quality control. If the dilution selected for
258
the “optimal” analysis was truly optimal, then diluting it 2x should yield a similar particle
259
intensity (i.e., size) with half the number of particles counted. For example, analysis of the soil
260
extract from the earthworm assay (nAgEA) showed 5.05 µg L-1 dissolved Ag and generated an
261
equivalent particle size distribution (PSDeqv) with a mode of 68.2 nm (Table 3). The optimal
262
dilution for the analysis of this sample was 100x, which reduced the dissolved signal to near the
263
MDL and provided a particle count < 105 particles mL-1, resulting in a PSDeqv with a mode of
264
37.4 nm (Figure 3a). When this optimal dilution was further diluted 2x, the resulting PSDeqv
265
showed a similar mode, but approximately half the number of particles, as expected, thus
ACS Paragon Plus Environment
Page 13 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
13 266
providing confidence that the analysis performed on the “optimal” dilution selected was indeed
267
not significantly impacted by excess dissolved analyte, analytical coincidence, nor matrix effects.
268
For the analysis of the soil extract from the microbial assay, the initial 100x dilution captured
269
a dissolved reading of 0.48 µg L-1 (Table 3). While we would not expect this dissolved
270
concentration to impact the particle signal (see Figure S5, Supporting Information), the particle
271
concentration detected at this dilution exceeded 105 particles mL-1, thus resulting in a high
272
probability for analytical coincidence. An optimal dilution of 2500x was determined which
273
shifted the PSDeqv left (Figure 3), resulting in the smaller mode of 32.7 nm and a greater overall
274
particle concentration, evidence that analytical coincidence likely affected the analysis of the
275
100x dilution. A further 2x dilution reduced the number of detectable particles to ca. 700 (data
276
not shown). However, we have found that when the event sample size drops much below
277
approximately 1000 events, there is greater variability among SP-ICP-MS data from analytical
278
replicates. Extending the sample time to 120 s allowed for a larger sample size of ca. 1400
279
events, which allowed for more reproducible data, and yielded a PSDeqv which generally
280
overlapped that from the optimal dilution, as would be expected given the operational
281
parameters. This technique of increasing sample time to increase the number of particle events
282
can be particularly useful when analyzing a sample with a large ratio of dissolved analyte /
283
particle counts, for which the technique of diluting the sample to reduce the dissolved signal
284
yields a sample with a very low particle count.
285
Similar to the previous samples, dilution of nAgFC shifted the initial PSDeqv left (Figure 3c),
286
resulting in a lower mode of 40.6 nm. However, a further 2x dilution (with the extended sample
287
time of 120 s), revealed a bimodal PSDeqv. This sample was re-run several times to verify this
288
bimodal pattern and data from duplicate analyses were combined. Given that over 30% more
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 14 of 23
14 289
particulate Ag mass was detected from the analysis of the most dilute sample compared to the
290
next more concentrated sample (Table 3), it is likely that the detection of the bimodal distribution
291
was due to an improved signal optimization (i.e., lower particle DL), rather than from dissolution
292
of the 40 nm particles. Each of the two distributions was fit to a log-normal curve separately to
293
determine their equivalent mode particle sizes which were 28.7 and 43.4 nm. While other
294
analytical techniques are required to further identify/characterize particles in these two distinct
295
size classes (e.g., Ag2S, AgNP with surface precipitate), the example here illustrates the
296
usefulness of checking the data acquired from what was believed to be the “optimal” dilution by
297
analyzing the sample with a further 2x dilution.
298
The most obvious drawback to analyzing environmental samples using a dilution series is the
299
potential for particle dissolution. While research has shown that AgNPs in distilled water likely
300
undergo fairly rapid dissolution in the short-term (i.e., < 24 h),12 AgNPs contained in more
301
complex matrices, particularly in the presence of natural organic matter, are more stable,13,14 with
302
humic substances likely providing both steric and charge stabilization. Mitrano et al.14 showed
303
that the presence of ligands, such as chloride and sulfide, also had a stabilizing effect on AgNPs,
304
and suggested the formation of relatively insoluble halide-Ag or sulfide-Ag precipitates on the
305
particle surface. Thus environmental samples containing ENPs, and/or their transformation
306
products may very well be stabilized by their solution chemistry enough to carry out dilution and
307
analysis in the very short-term (i.e., minutes) without significant changes to analyte mass of the
308
particulates. It would be useful to perform dilution-stability tests on samples of various complex
309
matrices to gain information regarding the rate of change that dilution may have on particle
310
stability. The dilution approach shown here may be more suited for soil extracts or wastewaters
311
which generally have a large humic component. In the case of soil analysis, all extractions (e.g.,
ACS Paragon Plus Environment
Page 15 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
15 312
soil pore water) and preparatory procedures are by necessity operationally defined, thus, it would
313
be advantageous to utilize an extraction technique that is amenable to SP-ICP-MS analysis. We
314
have found that the alkaline and dispersing attributes of TSPP result in large particle/dissolved
315
concentration ratios and stable extracts. For some environmental samples, it may be possible to
316
inhibit, or retard particle dissolution by maintaining the ionic strength of the environmental
317
sample during the dilution, (i.e., using a dilute Na solution) and/or adding a dispersant or
318
surfactant which may offer some steric stability. In our case, we maintained the TSPP
319
concentration of 0.025 mM in all dilutions.
320
Conclusions
321
For SP-ICP-MS analysis of environmental samples affected by engineered nanoparticle
322
contamination, quantification of both the dissolved and nano-particulate forms of the metal
323
analyte are of interest; however, it is doubtful that one dilution/analysis can optimally capture
324
both signals. As we have shown, a large dissolved signal can drown-out the particle signal
325
leading to erroneous sizing and particle counts. The degree to which an analysis will be
326
negatively impacted by presence of the dissolved analyte depends on both the particle size and
327
dissolved analyte concentration. The approach demonstrated here relies on analyzing samples in
328
a dilution series such that (1) an analysis captures the dissolved signal within the calibrated range
329
and above the LoQ, (2) an analysis captures an optimal particle signal, and (3) a third analysis is
330
used to verify the optimal particle data. An optimal particle signal can be achieved by diluting
331
the sample such that the dissolved signal is at or below the MDL and the particle concentration is
332
< 105 particles mL-1, to minimize the probability of analytical coincidence. Sampling time may
333
be extended to ensure sufficient number of events (i.e., > 1000) to achieve a suitable level of
334
statistical confidence when generating the particle frequency histograms and for ensuring
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 16 of 23
16 335
sufficient reproducibility. For the verification analysis, a 2x dilution of the optimal sample
336
dilution is recommended. The verification analysis ensures that interferences from the dissolved
337
signal, sample matrix effects and analytical coincidence have been sufficiently minimized. While
338
humic substances contained in environmental samples are likely to help stabilize particles and
339
buffer any rapid changes to particles that may be caused by the dilution process, further research
340
on how to preserve and stabilize environmental samples for SP-ICP-MS analysis is required. The
341
dilution technique demonstrated here offers a systematic and reproducible approach to the
342
analysis of environmental samples which is anticipated to improve the quality of data generated
343
by this relatively new analytical tool.
344 345 346 347
Associated Content
348
Additional information as noted in the text.
349
Acknowledgements
350
The research was funded by Environment and Climate Change Canada through the Chemicals
351
Management Plan.
352
References
353 354 355 356 357 358 359 360 361 362 363
(1) von der Kammer, F.; Ferguson, P. L.; Holden, P. A.; Masion, A.; Rogers, K. R.; Klaine, S. J.; Koelmans, A. A.; Horne, N.; Unrine, J. M. Environ. Toxicol. Chem. 2012, 31, 32-49. (2) Montano, M. D.; Ranville, J. F.; Lowry, G. V.; Lowry, G. V.; Von Der Kammer, F.; Blue, J. Environ. Chem. 2014, 11, 351-366. (3) Hassellöv, M.; Readman, J. W.; Ranville, J. F.; Tiede, K. Ecotoxicol. 2008, 17, 344-361. (4) Hadioui, M.; Peyrot, C.; Wilkinson, K. J. Anal. Chem. 2014, 86, 4668-4674. (5) Pace, H. E.; Rogers, N. J.; Jarolimek, C.; Coleman, V. A.; Higgins, C. P.; Ranville, J. F. Anal. Chem. 2011, 83, 9361-9369. (6) Laborda, F.; Jimâenez-Lamana, J.; Bolea, E.; Castillo, J. R. J. Anal. At. Spectrom. 2011, 26, 1362-1371. (7) Hadioui, M.; Merdzan, V.; Wilkinson, K. J. Environ. Sci. Technol. 2015, 49, 6141-6148.
Supporting Information
ACS Paragon Plus Environment
Page 17 of 23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
17 364 365 366 367 368 369 370 371 372 373 374 375 376 377
(8) Gottschalk, F.; Nowack, B. JEM 2011, 13, 1145-1155. (9) Regulatory Cooperation Council (RCC) Nanotechnology Initiative, Final Report on Work Element 2, Priority Setting: Development of a joint nanomaterial classification scheme.: Available at: Government of Canada NanoPortal (http://nanoportal.gc.ca), Accessed May 28, 2012. (10) Koppel, D. E. J. Chem. Phys. 1972, 57, 4814. (11) Hineman, A.; Stephan, C. J. Anal. At. Spectrom. 2014, 29, 1252. (12) Liu, J.; Hurt, R. H. Environ. Sci. Technol. 2010, 44, 2169-2175. (13) Cumberland, S. A.; Lead, J. R. J.Chrom. A 2009, 1216, 9099-9105. (14) Mitrano, D. M.; Ranville, J. F.; Bednar, A.; Kazor, K.; Hering, A. S.; Higgins, C. P. Environ. Sci.: Nano 2014, 1.
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 18 of 23
18 378 379 380 381
382 383 384 385 386 387
TABLES & GRAPHS Table 1. Supplier Information for the Nanomaterials Used Metal
Nominal size
Supplier
Product Name
Surface Coating
Form
Supplier’s size specification
Au
30 nm
NIST*
Citrate
60 nm
NIST
Ag
30 nm
Ted Pella
Ag
60 nm
Ted Pella
Ag
40 nm
Nano-composix
Aqueous dispersion Aqueous dispersion Aqueous dispersion Aqueous dispersion Powder
27.6 ± 2.1 nm
Au
Reference Material 8012 Reference Material 8013 PELCO NanoXact Silver PELCO NanoXact Silver Redispersible silver nanospheres
Citrate Citrate Citrate PVP, 88% by mass
* National Institute of Standards and Technology
Table 2. SP-ICP-MS Instrumental and Analytical Details Parameter Group ICP-MS component
Parameter
Parameter Value
ICP-MS model
Spray chamber
Perkin Elmer NexION 300D Low pressure PFA nebulizer Cyclonic, glass
Neb gas flow
1.04 L/min
Plasma RF power
1600 W
Sample flow rate
0.270 mL/min
Analyte monitored
107
Sample introduction Transport efficiency Instrument dwell time Sample analysis time Readings per sample
Manual, with a 60 s rinse and 20 s read delay 9.2% to 9.8%
Nebulizer
ICP-MS settings
Single Particle Parameter
Ag or 197Au
50 µs 60 s (or more) ~ 1.2 x 106 (or more)
388 389
ACS Paragon Plus Environment
56.0 ± 0.5 nm 32.7 ± 4.8 nm 62 ± 7 nm 38.6 ± 9.8 nm
Page 19 of 23
19 390 350
(a)
300
Frequency
Frequency
250 200 150 100 50 0 24
391
37
44
49
54
35
58
400 350 300 250 200 150 100 50 0
(b)
24
61
37
44
49
54
10
(c)
30
58
(d)
8
Frequency
Frequency
25 20 15 10 5
4 2
24 37 44 49 54 58 61 64 67
24 37 44 49 54 58 61 64 67 69
392
6
0
0
3
140
(e)
(f)
120 100
2
Frequency
Frequency
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
80 60 40 20
0
0 24 37 44 49 54 58 61 64 67 69 72 74 76
393 394 395 396 397
Diameter (nm)
24
37
44
49
54
58
Diameter (nm)
Figure 1. Particle size histograms for the 30 nm Au particle suspension with (a) no dissolved Au added, (b) 0.5 µg L-1 dissolved Au, (c) 1 µg L-1 dissolved Au, (d) 2 µg L-1 dissolved Au, (e) 3 µg L-1 dissolved Au, and (f) 3 µg L-1 dissolved Au following 5x dilution and extending sample analysis time to 120 s.
ACS Paragon Plus Environment
Analytical Chemistry
20 30 nm 30 nm (5x dilute)
90
60 nm 60 nm (10x dilute)
mean size (nm)
80 70 60 50 40 30 20 1.0
2.0
3.0
4.0
5.0
log [dissolved Au] (ng L-1)
398 1000000 100000
particles mL-1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 20 of 23
10000 1000 100 1.0
399 400 401 402 403 404 405 406 407 408 409
2.0
3.0
4.0
5.0
log [dissolved Au] (ng L-1)
Figure 2. Mean particle size (nm) and number of particles (mL-1) measured in 30 nm and 60 nm Au particle suspensions (nominal particle concentrations = 105 particles mL-1) spiked with increasing concentrations of dissolved Au, shown as log concentration (ng L-1) along the x-axis. Included are the results for the 30 nm + 3 µg L-1 dissolved Au sample which was diluted 5x (×), and results for the 60 nm + 40 µg L-1 dissolved Au sample which was diluted 10x (+). Note: The dilution factors have been applied to the particle counts so that values shown represent the final particle concentrations for the original (undiluted) samples.
ACS Paragon Plus Environment
Page 21 of 23
21 nAgEA (1x) nAgEA (50x) nAgEA (100x)
60 50
Frequency
40 30 20 10 0 0
20
40
60
80
100
120
140
410 nAgMicro (100x) nAgMicro (2500x) nAgMicro (5000x)
140 120
Frequency
100 80 60 40 20 0 0
20
40
60
80
100
120
411 nAgFC (1x) nAgFC (100x) nAgFC (200x) - 1st curve nAgFC (200x) - 2nd curve
60 50 40
Frequency
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
30 20 10 0 0
412 413 414 415 416 417 418 419
20
40
60
80
100
120
140
Diameter (nm)
Figure 3. Equivalent particle size distributions (PSDeqv) for particulate Ag detected in soil extracts using a dilution series technique. Soil samples were taken from: an earthworm assay (nAgEA), a microbial assay (nAgMicro), and a collembolan assay (nAgFC). Extract dilutions are shown in brackets. Note: analysis time for nAgMicro (5000x) and nAgFC (200x) was extended from 60 s to 120 s.
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 22 of 23
22 420 421 422 423
Table 3. Analytical information and SP-ICP-MS results for soil extracts analyzed for 107Ag using a dilution series. Sample time for analyses was 60 s, except for nAgMicr (5000x) and nAgFC (200x) which were analyzed for 120 s. Note: DF = dilution factor; part. = particle
Sample
Dissolv ed Ag (µg L-1)
Mode sizeeqv (nm)
# peaks detecte d
Final* part. counts log [part.] (mL-1)
1
5.05
68.2
8763
5.5
1.03
100
0.06
37.4
2542
6.9
6.77
200
< 0.05
36.6
1227
6.9
6.08
100
0.48
39.8
13055
7.7
41.5
2500