Systematic Evaluation of Solid-Phase Microextraction Coatings for

Feb 18, 2011 - (2) For these reasons, in vivo solid-phase microextraction (SPME) is an attractive ..... As a general trend, the extraction efficiency ...
0 downloads 0 Views 4MB Size
ARTICLE pubs.acs.org/ac

Systematic Evaluation of Solid-Phase Microextraction Coatings for Untargeted Metabolomic Profiling of Biological Fluids by Liquid Chromatography-Mass Spectrometry Dajana Vuckovic and Janusz Pawliszyn* Department of Chemistry, University of Waterloo, 200 University Avenue West, Waterloo, Canada, N2L 3G1

bS Supporting Information ABSTRACT: In this study, we propose for the first time the use of solid-phase microextraction (SPME) in combination with liquid chromatography-mass spectrometry for untargeted metabolomic profiling of biological fluids. To achieve this goal, we first systematically evaluated 42 different SPME coatings for the extraction of 36 metabolites from different chemical classes and of widely varying polarities (log P range of -7.9 to 7.4) in order to identify SPME coatings which are the most suitable for metabolomic studies and to improve the extraction of polar metabolites over the existing commercial SPME devices. Three types of SPME coatings (mixed-mode coatings, polar-enhanced polystyrene-divinylbenzene, and phenylboronic acid) performed the best for simultaneous extraction of both hydrophilic and hydrophobic metabolites at physiological conditions, thus making them suitable for untargeted metabolomic profiling applications. A rapid and simple SPME method was then developed with single-use biocompatible mixed-mode coating for the metabolomic profiling of human plasma in combination with liquid chromatography-high-resolution mass spectrometry on a benchtop Orbitrap system. This optimized SPME method was evaluated versus ultrafiltration and solvent precipitation in terms of metabolite coverage and method precision. SPME detected 1592-3320 features versus 2082-3245 features detected by solvent precipitation methods and 2093-2686 detected for ultrafiltration using the same pooled human plasma sample. Method precision of SPME ranged between 11% and 18% (expressed as median relative standard deviation (RSD) of n = 7 replicates) versus 8-19% for solvent precipitation and 20-22% for ultrafiltration. The results demonstrate that the proposed SPME methodology reduces ionization suppression, provides free concentration information for hydrophobic analytes which are not detected by ultrafiltration methods, and can improve metabolite coverage over existing methodologies.

Metabolomics can be defined as the comprehensive analytical approach to study all low molecular weight species (typically defined as