The Utility and Potential Artifacts by Zeeman Cold Vapor Atomic Absorption Analysis of Mercury Stable Isotopes Xia Lu, Jiangtao Zhao, Xujun Liang, Lijie Zhang, Yu-Rong Liu, Xiangping Yin, Xiangkai Li, and Baohua Gu Environ. Sci. Technol. Lett., Just Accepted Manuscript • DOI: 10.1021/acs.estlett.9b00067 • Publication Date (Web): 07 Feb 2019 Downloaded from http://pubs.acs.org on February 12, 2019
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 16
Environmental Science & Technology Letters
1
The Utility and Potential Artifacts by Zeeman Cold Vapor Atomic Absorption Analysis
2
of Mercury Stable Isotopes
3 4
Xia Lu,1,2,#,* Jiangtao Zhao,3# Xujun Liang,2,# Lijie Zhang,2 Yurong Liu,2 Xiangping Yin,2 Xiangkai Li,1
5
Baohua Gu2,4,*
6 7 8 9
1Key
Laboratory of Cell Activities and Stress Adaptations, School of Life Sciences, Lanzhou University,
Lanzhou 730000, China 2Environmental
10
3School
11
4Department
12
Sciences Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831, USA
of Nuclear Science and Technology, Lanzhou University, Lanzhou 730000, China of Biosystems Engineering and Soil Science, University of Tennessee, Knoxville, TN
37996, USA
13 14
# These
authors contributed equally.
15 16 17 18 19
1 ACS Paragon Plus Environment
Environmental Science & Technology Letters
20
ABSTRACT
21
Zeeman cold vapor atomic absorption spectrometry (CVAAS) has been widely used for
22
environmental mercury (Hg) detection and quantification for decades, but little is known about its utility
23
and potential artifacts in analyzing Hg with varying isotope compositions. We show that each Hg isotope
24
responds differently by CVAAS analysis, with 200Hg and 202Hg isotopes exhibiting ~10 times greater
25
signal intensities than 198Hg and 201Hg isotopes. However, all Hg isotopes show a linear correlation
26
between Hg concentrations and the signal intensity, validated by both measurements and theoretical
27
simulations. Zeeman CVAAS could thus offer a convenient, inexpensive tool for determining Hg
28
isotopes, particularly in using single- or dual-labeled Hg isotopes for tracing Hg biogeochemical
29
transformations, such as partitioning, ion exchange, sorption-desorption, and methylation-demethylation
30
in environmental matrixes. We also caution that care must be taken when CVAAS is used for quantifying
31
Hg in samples with changing isotope compositions to avoid measurement errors.
32
2 ACS Paragon Plus Environment
Page 2 of 16
Page 3 of 16
34 35
Environmental Science & Technology Letters
INTRODUCTION Cold vapor atomic absorption spectrometry (CVAAS) is widely used for mercury (Hg) detection
36
and quantification in environmental samples because of its high sensitivity, capable of detecting sub-
37
picogram (pg) quantities of Hg.1-5 It is also a preferred choice for field and air monitoring applications
38
because of its affordability and portability.1,2,6 However, a common issue with CVAAS is potential
39
interferences resulting from molecular absorption of gaseous impurities, such as benzene. To resolve this
40
issue, Zeeman background correction is used to improve the sensitivity, selectivity, and measurement
41
accuracy.7,8 This is done by splitting the Hg resonance line (λ=254 nm) into three polarized Zeeman
42
components (π, σ-, σ+), although the photodetector detects only the radiation of two σ components (see
43
Supporting Information (SI) “Zeeman CVAAS and theory” for details). The σ components are separated
44
by the polarization modulator, and the spectral shift of the σ components is less than 0.1 nm or equivalent
45
to a frequency of ~22 GHz (due to the hyperfine structure of Hg), which is much smaller than the widths
46
of molecular absorption bands and/or scattering spectra of most interfering molecules. When the light
47
passes through the sample chamber, the signal intensity is defined by the difference between the
48
intensities of the two components of the sample.7 Consequently, the background absorption from any
49
interferences in σ- and σ+ is always the same and thus canceled. However, the absorption line profiles of
50
most Hg isotopes fall within the σ- component, except 199Hg, which falls mostly in the σ+ component (SI,
51
Zeeman CVAAS and theory, Figure S1). The difference between σ- and σ+ intensities thus changes with
52
changing Hg isotope compositions and their vapor concentrations, thereby providing a tool for
53
quantifying Hg or Hg isotope concentrations.
54
Mercury has six stable isotopes with high natural abundances ranging from 6.9% to 29.9%.
55
Enriched Hg isotopes are often used as tracers for investigating Hg fate and transformations in
56
environmental systems, such as transport, sorption-desorption, methylation-demethylation, precipitation-
57
dissolution, biological uptake, and ion exchange reactions.9-13 However, to date few studies have taken the
58
advantage of Zeeman CVAAS for quantifying Hg isotopes,8 a technique that has been largely overlooked
3 ACS Paragon Plus Environment
Environmental Science & Technology Letters
59
in both laboratory and field investigations. Equally important is to realize that, if the Hg isotope effect is
60
not considered, CVAAS analysis could potentially lead to large measurement errors if Hg isotope
61
compositions in the reactant and the product change over the course of reactions. Although cold vapor
62
atomic fluorescence spectrometry (CVAFS) is also widely used because of its low cost, it does not
63
discriminate Hg isotopes as CVAFS measures the broad atomic fluorescence of Hg.14,15
64
The primary objective of this study was to demonstrate potential artifacts and the practical utility
65
of Zeeman CVAAS for determining Hg isotopes in tracing Hg biogeochemical transformations, such as
66
partitioning and concurrent Hg sorption-desorption reactions, in environmental matrixes. We provide
67
theoretical basis for quantifying individual Hg isotopes or their mixtures and compare the advantages and
68
disadvantages of using CVAAS for Hg isotope analyses. We suggest that CVAAS could offer a
69
convenient, inexpensive tool for quantifying Hg isotopes without using specialized equipment, such as
70
inductive coupled plasma–mass spectrometer (ICP-MS).9,12,13
71
MATERIALS AND METHODS
72
Standard solution of the natural Hg (1 µg mL-1 in 2% HNO3) was purchased from Brooks Rand
73
Laboratories (Seattle, WA). Enriched Hg stable isotopes (198Hg, 200Hg, 201Hg, and 202Hg) were supplied by
74
Oak Ridge National Laboratory, Oak Ridge, Tennessee, and the purity of these isotopes is certified as
75
follows: 198Hg (92.78%), 200Hg (96.41%), 201Hg (96.17%), and 202Hg (95.86%).11 All impurity isotopes are
76
listed in Supporting Information (SI Table S1). These isotopes were received as HgO or HgCl2 solids, and
77
their stock solutions were prepared by dissolving HgO or HgCl2 in 1–3% HCl and diluting to 1 mg/L in
78
1% HCl. The concentration of each Hg isotope was validated by ICP-MS analysis (Elan DRC-e,
79
PerkinElmer Inc.). The isotope and natural Hg working solutions at 10 µg/L were preserved in 0.5%
80
BrCl, and an aliquot (in the mass range of 200–1200 pg) was used and analyzed by CVAAS (Lumex RA-
81
915+, Ohio Lumex Co.), as described previously.3-5,16,17 The peak intensity as a function of the Hg mass
82
or concentration was plotted by linear fitting using the OriginPro software (OriginLab).5
4 ACS Paragon Plus Environment
Page 4 of 16
Page 5 of 16
Environmental Science & Technology Letters
83
Standard additions of Hg isotopes were performed to evaluate the utility and artifacts by CVAAS
84
analyses in an environmental water sample obtained from the contaminated East Fork Poplar Creek
85
(EFPC) in Oak Ridge, Tennessee, as described previously.18, 19 The filtered water contained ~3 mg/L
86
dissolved organic matter and 25 ng/L Hg at pH 7.8. Following the addition of a Hg isotope (~85 ng/L),
87
the total Hg (THg) concentration was determined after samples were oxidized in BrCl (5% v/v)
88
overnight.3,4,20,21 Additionally, sequential extraction was used to compare the partitioning and
89
immobilization of the freshly added 200Hg (as a tracer) with previously deposited old Hg in a
90
contaminated EFPC soil.22,23 The extractable pools of Hg included (1) deionized water, (2) 0.01M
91
HCl/0.1M acetic acid, (3) 1M KOH, (4) 12M nitric acid, and (5) aqua regia, and were used as a proxy of
92
Hg bioavailability in soil (see SI “Hg sequential extraction”).22,23 Furthermore, concurrent Hg adsorption
93
and desorption on mineral hematite were conducted using dual-labeled Hg isotopes (e.g., 200Hg and
94
198Hg),
95
a small aliquot (40 µL) of 198Hg (15 µg/L, final concentration). Hematite particles were then removed by
96
filtration (0.2 m), and Hg concentrations in supernatant solutions before and after 198Hg addition were
97
determined. The amounts of 198Hg and 200Hg adsorbed (or 200Hg desorbed) were calculated by the
98
difference between the amounts added and that found in the supernatant solution. The average
99
concentrations of duplicate samples were reported, and error bars represent the deviation between the two
100 101
in which 200Hg (17 µg/L) was equilibrated with hematite first for 24 h, followed by the addition of
replicates (X1 and X2), defined as the absolute value of (X1–X2)/2. The CVAAS method detection limit (MDL) was determined for four available Hg isotopes
102
(198Hg, 200Hg, 201Hg, and 202Hg) in both deionized water and the simulated water consisting of 1 mM
103
NaHCO3, 2 mM CaCl2, 2 mM MgCl2, and 1 mM ZnCl2 as interfering ions.24 No significant differences
104
were observed between samples prepared in deionized water or the simulated water (SI Figure S2) since
105
all Hg was converted to gaseous elemental Hg(0) by Sn(II) reduction and ultimately detected by CVAAS.
106
Therefore, MDLs were computed as an average (from deionized water and the simulated water). They
107
were 8.08±1.78, 3.90±0.73, 13.54±1.96, and 3.13±0.12 pg for 198Hg, 200Hg, 201Hg, and 202Hg, respectively.
5 ACS Paragon Plus Environment
Environmental Science & Technology Letters
108
Page 6 of 16
RESULTS AND DISCUSSION 250
Total Hg (ng/L)
200
CVAAS (no correction) ICP-MS
(a)
150
100
50
0
109
EFPC water
+ 198Hg
+ 200Hg
+ 201Hg
+ 202Hg
20
CVAAS (no correction) ICP-MS
Hg (g/g)
15
(b) 10
5
0 110 111 112 113 114 115 116
er s Wat
le /HCl 1 M) olub a c id c i OH ( t e K c A
(1 HNO 3
2 M)
a re Aqu
gia
Figure 1. Comparisons between Zeeman CVAAS and ICP-MS for determining Hg concentrations in environmental matrixes. (a) EFPC creek water spiked with or without Hg stable isotopes of 198Hg, 200Hg, 201Hg, and 202Hg (85 ng/L each), and (b) sequential extraction of Hg in a contaminated EFPC soil with freshly added 200Hg as a tracer (see SI, Hg sequential extraction for details).22, 23 Without correction of the Hg isotope effect, CVAAS analysis results in large measurement errors. 6 ACS Paragon Plus Environment
Page 7 of 16
117 118
Environmental Science & Technology Letters
“Artifacts” by Zeeman CVAAS analysis Potential artifacts associated with Zeeman CVAAS analyses of Hg isotopes in environmental
119
matrixes were first realized during measurements of Hg concentrations in the contaminated EFPC water
120
by standard addition of Hg isotopes of 198Hg, 200Hg, 201Hg, or 202Hg (85 ng/L each). Using standard
121
calibration of the ambient natural Hg, we found that the measured THg concentrations after additions of
122
201Hg, 198Hg, 202Hg,
123
the anticipated THg should be ~110 ng/L (analyzed by ICP-MS) (as the creek water contained ~25 ng/L
124
Hg). The results thus indicate that, for samples spiked with 198Hg and 201Hg, CVAAS substantially
125
underestimated the Hg concentration by 64.4% and 53.8%, respectively, but overestimated the Hg
126
concentration by 85.5% and 107.5% in samples spiked with 202Hg, and 200Hg, respectively.
127
and 200Hg were 39.2, 50.8, 204.1, and 228.2 ng/L, respectively (Figure 1a), whereas
Similarly, in a comparative study of sequential extraction of the previously deposited Hg and the
128
freshly added 200Hg (as a tracer) in a contaminated EFPC soil (SI, Hg sequential extraction),22 we found
129
that CVAAS substantially overestimated the four extractable fractions of Hg when compared to the
130
amounts analyzed by ICP-MS (Figure 1b). The measured THg concentrations by CVAAS in fractions of
131
water, acetic acid/HCl, KOH, HNO3, and aqua regia were 0.01, 1.28, 6.17, 6.42 and 14.35 g/g,
132
respectively, compared to the ICP-MS measured concentrations of 0.01, 0.29, 3.41, 3.46, and 10.75 g/g,
133
respectively (Figure 1b). In total, CVAAS overestimated the extractable Hg by about 56%. The result
134
again signifies serious measurement artifacts by CVAAS when the Hg isotope is used as a tracer and the
135
isotope effect is not accounted for (described below).
136
Theory and utility of Zeeman CVAAS in Hg isotope analysis
137
To provide theoretical basis for the observed CVAAS measurement errors (Figure 1), we
138
analyzed four Hg isotopes in hand, including 198Hg, 200Hg, 201Hg, and 202Hg, and performed simulations of
139
CVAAS responses to all six Hg isotopes (SI, Zeeman CVAAS and Theory). Results indicate that each Hg
140
isotope gave vastly different signal responses by CVAAS (Figure 2a). With the same amount of Hg added
141
to solution (200–1200 pg), the measured peak intensities of 200Hg and 202Hg were ~10 times higher than
7 ACS Paragon Plus Environment
Page 8 of 16
142
those of 201Hg and 198Hg, although they all showed a linear correlation with the added Hg mass. The
143
measured slopes of the signal intensity against the added Hg mass were 83.41±0.36, 73.65±0.40,
144
9.46±0.07, and 7.21±0.10 counts/pg for 200Hg, 202Hg, 198Hg, and 201Hg, respectively. Obviously, these
145
large differences in CVAAS responses to Hg isotopes explained large discrepancies observed when Hg
146
isotope tracers were used in the standard addition and sequential extraction experiments (Figure 1), in
147
which Hg isotope effects were not accounted for. However, when corrected using the calibration curve (or
148
the slope) of each Hg isotope shown in Figure 2a, the measured Hg concentrations by CVAAS matched
149
well with those determined by ICP-MS (SI Figure S3). 12 Experiment 198 Hg 200 Hg 201 Hg 202 Hg
Peak intensity (104 counts)
10 8 6
Simulation 198 Hg 200 Hg 201 Hg 202 Hg 199 Hg 204 Hg
200
Hg
202
Hg
(a)
199
Hg
4 204
Hg
2
198
Hg
201
Hg
0 0
200
400
600
800
1000
1200
Total Hg (pg) 12 Experiment Natural Hg 200 Hg:198Hg=3:7 200 Hg:198Hg=7:3
10
Peak intensity (104 counts)
1.
Environmental Science & Technology Letters
Simulation Natrual Hg
8 200
Hg:198Hg=7:3
(b)
6
Natural Hg
4 200
Hg:198Hg=3:7
2 0 0
150
200
400
600
800
Total Hg (pg)
8 ACS Paragon Plus Environment
1000
1200
Page 9 of 16
151 152 153
Environmental Science & Technology Letters
Figure 2. Zeeman CVAAS determination of (a) 198Hg, 200Hg, 201Hg, and 202Hg isotopes (symbols) and the simulation results (solid lines), and (b) mixed isotopes of 200Hg and 198Hg, and the natural Hg.
154
Why Hg isotopes respond differently by CVAAS is attributed to the absorption profiles of the
155
hyperfine structures of Hg which overlap with Zeeman - and + components of the light source profile
156
(SI, Zeeman CVAAS and Theory; Figure S1). As described earlier, the absorption line profiles of most
157
Hg isotopes fall within the σ- component, except 199Hg and part of 201Hg, which fall within the σ+
158
component. The difference between signal intensities of σ- and σ+ components changes with changing Hg
159
isotope compositions and concentrations and is defined by Eq. 1 (SI, Zeeman CVAAS and Theory).
160
Using Eq. 1, we can simulate the signal response to each individual Hg isotope, and the simulation results
161
for 198Hg, 199Hg, 200Hg, 201Hg, 202Hg, and 204Hg were plotted in Figure 2a as solid lines. The calculated
162
slopes are also listed in SI Table S2, in which Columns 4 and 5 are the adjusted slopes, corrected for
163
impurities present in the four measured Hg isotopes. The simulation results showed an excellent
164
agreement with the measured values (symbols) of four Hg isotopes (198Hg, 200Hg, 201Hg, 202Hg). Data for
165
199Hg
166
isotopes, simulation results gave a linear relationship between the signal intensity and the Hg
167
concentration (Figure 2a), although the slope varied with different Hg isotopes due to their differences in
168
hyperfine structures.
169
and 204Hg were not available because of their limited source availability. However, for all Hg
Both experiments and simulations (Figure 2a) clearly show the usefulness of CVAAS for
170
quantitative analyses of Hg isotopes because each Hg isotope gives a linear correlation and a distinct
171
slope, which is a constant for a given instrument with preset conditions. Importantly, when the two Hg
172
isotopes were mixed in solution, we found that the measured slope was additive, equal to the sum of the
173
products of the slope and the mass fraction of each isotope in the mixture. For example, at the 200Hg to
174
198Hg
175
(=0.3×83.41 + 0.7×9.46), which is in excellent agreement with the measured slope of 31.92 (Figure 2b,
176
dashed line and symbol). Similarly, at the 200Hg to 198Hg ratio of 7:3, the measured slope of the mixture
atomic ratio of 3:7, the slope of the total Hg (THg = 200Hg + 198Hg) was calculated to be 31.65
9 ACS Paragon Plus Environment
Environmental Science & Technology Letters
177
was 61.23, consistent with the calculated slope of 61.23 (Figure 2b). We also measured the instrument
178
response to the natural Hg and observed a slope of 36.24±0.17 (Figure 2b). However, this slope was
179
lower than the expected additive value of the four measured Hg isotopes (SI Table S2) and their
180
respective natural abundances (i.e., 198Hg, 9.97%; 200Hg, 23.10%; 201Hg, 13.18%; 202Hg, 29.86%). The
181
expected slope of the natural Hg should be at least 41.45 (by accounting the four measured isotopes), if all
182
Hg isotopes would obey the additive rule described above. The main cause of this deviation is attributed
183
to the 199Hg hyperfine structure overlapping mostly with the + component, and its signal intensity is
184
therefore subtractive when mixed with other Hg isotopes (SI, Eq. 1). This result thus explained why the
185
measured slope of the natural Hg was lower than the value expected from the summation of the four
186
measured isotopes (198Hg, 200Hg, 201Hg, 202Hg). Therefore, except for 199Hg, the signal intensity of any of
187
the mixed isotopes of 198Hg, 200Hg, 201Hg, 202Hg, and 204Hg equals to the summation of each individual Hg
188
isotope, confirmed by the measured values of any binary mixtures of 198Hg and 200Hg at varying atomic
189
ratios (Figure 2b; SI Figure S4a) or a mixed sample containing 198Hg, 200Hg, 201Hg, and 202Hg (SI Figure
190
S4b). These results thus indicate the robustness of using CVAAS as a tool for Hg isotope analysis.
191
Environmental applications of CVAAS for Hg isotope tracer studies: Pros and Cons
192
Enriched Hg isotopes are often used in tracing Hg biogeochemical transformations in both
193
laboratory and field investigations,9-13 such as standard addition, sorption and desorption, methylation and
194
demethylation, and ion exchange reactions in environmental systems. As illustrated in Figure 1b,
195
sequential extraction was used to assess Hg species distributions between the freshly added 200Hg tracer
196
(after 24 h) and the old Hg in a contaminated EFPC soil.22, 23 The freshly added 200Hg was determined
197
after subtracting the old Hg in the untreated soil (without 200Hg). Based on the measured 200Hg calibration
198
curve (Figure 2a, Table S2), we found that the distribution of 200Hg in extractable pools resembled the old
199
Hg (Figure 3a), in which negligible amounts of 200Hg were found in water- and acetic acid/HCl-
200
extractable pools (presumably more bioavailable than other Hg pools). Most of the added 200Hg seems
201
immobilized in the pools associated with soil organic matter (1M KOH) and minerals (12M HNO3 and
202
aqua regia) (Figure 3a). These results confirm previous observations that freshly added Hg is often rapidly 10 ACS Paragon Plus Environment
Page 10 of 16
Page 11 of 16
Environmental Science & Technology Letters
203
immobilized and becomes unavailable for biological uptake,10, 25 although detailed discussion of the
204
mechanisms of Hg immobilization in soil is beyond the scope of this work. 100 Water soluble Acetic acid/HCl KOH (1 M) HNO3 (12 M) Aqua regia
Hg distribution (%)
80 60
(a) 40 20 0
205
(b)
Hg in solution
12 198
10 8 6 4
200
Hg
(200Hg+198Hg)
Hg added
After sorption
200
Peak intensity (105 counts/mL)
14
New 200Hg
"Old" Hg in soil
Hg added
200
Hg
2 0
120
No hematite
(c)
Hematite
Hg adsorbed
Hg adsorbed (mg/g)
100 80 200
60
198
Hg Hg
40 20 0
206 207 208 209 210 211 212 213
No hematite
Hematite
Figure 3. (a) Comparisons between the new, freshly added 200Hg (as a tracer) and old Hg species distributions in a contaminated EFPC soil by sequential extraction.22, 23 (b, c) Determination of 200Hg sorption and then desorption by 198Hg on hematite (0.1 g/L) at pH 6.5. The initially added 200Hg was 17 µg/L. After equilibration with hematite, approximately an equal molar concentration of 198Hg (~15 µg/L) was added. Plot (b) is the measured CVAAS peak intensities of 200Hg and (200Hg +198Hg) in solution before and after their adsorption on hematite, and (c) is the amount of Hg adsorbed on hematite. 11 ACS Paragon Plus Environment
Environmental Science & Technology Letters
214
Perhaps the most important application of CVAAS is the use of dual-labeled Hg isotopes (e.g.,
215
200Hg
216
Hg adsorption and desorption reactions, as illustrated in Figure 3b,c. Because of a large difference (~10×)
217
in CVAAS responses to 200Hg and 198Hg (Figure 2a, Table S2), any desorption or displacement of one Hg
218
isotope by the other could be readily detected. Results show that, after pre-equilibration of 200Hg with
219
hematite for 24 h, the CVAAS intensity in the supernatant decreased from 12.51×105 (Initial) to 5.67×105
220
counts/mL (Figure 3b, 2nd column). However, a substantially increased intensity (10.3×105 counts/mL)
221
was observed (Figure 3b, 3rd column) following the addition of 198Hg. This intensity was much higher
222
than that expected (7.6×105 counts/mL, Figure 3b, 4th column) by the summation of 200Hg (assuming no
223
desorption) and the total added 198Hg in solution (assuming no adsorption of 198Hg). Evidently, a large
224
fraction of the pre-adsorbed 200Hg was desorbed by 198Hg, resulting in a substantially increased intensity
225
because CVAAS response of 200Hg is ~10 times higher than that of 198Hg (SI Table S2). Since the
226
supernatant solution now contained both 200Hg and 198Hg, the measured total intensity cannot be directly
227
converted to the Hg concentration unless the THg concentration or the ratio of 200Hg to 198Hg in solution
228
is known. In this case, THg is quantified using techniques such as CVAFS, as described earlier, so that
229
200Hg
230
and 198Hg adsorbed on hematite determined (Figure 3c). Interestingly, we found that the addition of 198Hg
231
to the hematite suspension that had been pre-equilibrated with 200Hg resulted in desorption or exchange of
232
200Hg
233
and 198Hg) to trace simultaneous Hg transformations in environmental systems, such as concurrent
and 198Hg in solution can be calculated using the slopes given in Table S2, and the amounts of 200Hg
by 198Hg and led to approximately an equal distribution of 200Hg and 198Hg on hematite. Together, these two case studies (Figure 3) demonstrate the practical utility of Zeeman CVAAS
234
for accurate determination of single or dual-labeled Hg isotopes in studies such as tracing Hg fate and
235
transport, sorption-desorption, and transformation reactions in environmental systems, as reported
236
previously.9-13 Another major advantage of using Zeeman CVAAS is its low detection limit, field
237
portability, and wide availability in many Hg labs because of its low cost (~10× less expensive than a
238
typical ICP-MS). It could also be used in real-time analysis and monitoring. A disadvantage is that
239
CVAAS measures the total signal intensity and cannot resolve individual Hg isotopes when mixed. 12 ACS Paragon Plus Environment
Page 12 of 16
Page 13 of 16
Environmental Science & Technology Letters
240
Fortunately, most tracer studies use only one or two Hg isotopes (Figure 3). In the latter case, only a
241
secondary measurement is necessary to determine the THg by techniques such as CVAFS and ICP-MS.
242
However, regardless of its application, realizing the utility and potential artifacts of CVAAS in Hg
243
isotope analysis is critically important, as it is among the most widely used technique for Hg
244
quantification and monitoring in both laboratory and field investigations.
245 246
ASSOCIATED CONTENT
247
Supporting Information
248
All data needed to evaluate the conclusions in the paper are present in the paper and/or the Supporting
249
Information, which is available free of charge via the Internet at http://pubs.acs.org.
250 251
AUTHOR INFORMATION
252
Corresponding author:
253
Email:
[email protected];
[email protected] 254
Notes:
255
The authors declare no competing financial interest.
256 257
ACKNOWLEDGMENTS
258
This research was sponsored by the Office of Biological and Environmental Research within the Office of
259
Science of the U.S. Department of Energy (DOE), as part of the Mercury Science Focus Area project at
260
the Oak Ridge National Laboratory (ORNL) and by the National Natural Science Foundation of China
261
(41807477). The Department of Energy will provide public access to these results of federally sponsored
262
research in accordance with the DOE Public Access Plan (http://energy.gov/downloads/doe-public-
263
access-plan). ORNL is managed by UT-Battelle, LLC under Contract No. DE-AC05-00OR22725 with
264
DOE.
265
13 ACS Paragon Plus Environment
Environmental Science & Technology Letters
266
REFERENCES
267 268
1.
Hadeishi, T., Isotope-shift Zeeman effect for trace-element detection - Application of atomic physics to environmental problems. Appl. Phys. Lett. 1972, 21 (9), 438-440.
269 270
2.
Brooks, S. C.; Southworth, G. R., History of mercury use and environmental contamination at the Oak Ridge Y-12 Plant. Environ. Pollut. 2011, 159 (1), 219-228.
271 272 273
3.
Gu, B.; Bian, Y.; Miller, C. L.; Dong, W.; Jiang, X.; Liang, L., Mercury reduction and complexation by natural organic matter in anoxic environments. Proc. Natl. Acad. Sci. USA 2011, 108 (4), 14791483.
274 275
4.
Lu, X.; Gu, W. Y.; Zhao, L.; Ul Haque, M. F.; DiSpirito, A. A.; Semrau, J. D.; Gu, B., Methylmercury uptake and degradation by methanotrophs. Science Adv. 2017, 3, e1700041.
276 277 278
5.
Lu, X.; Johs, A.; Zhao, L.; Wang, L.; Pierce, E. M.; Gu, B., Nanomolar copper enhances mercury methylation by Desulfovibrio desulfuricans ND132. Environ. Sci. Technol. Lett. 2018, 5 (6), 372376.
279 280
6.
Panichev, N. A.; Panicheva, S. E., Determination of total mercury in fish and sea products by direct thermal decomposition atomic absorption spectrometry. Food Chem. 2015, 166, 432-441.
281 282
7.
Sholupov, S. E.; Ganeyev, A. A., Zeeman atomic-absorption spectrometry using high-frequency modulated light polarization. Spectrochim Acta B 1995, 50 (10), 1227-1236.
283 284
8.
Ganeyev, A. A.; Sholupov, S. E., New Zeeman atomic-absorption spectroscopy approach for mercury isotope analysis. Spectrochim Acta B 1992, 47 (11), 1325-1338.
285 286 287
9.
Jiang, P.; Li, Y. B.; Liu, G. L.; Yang, G. D.; Lagos, L.; Yin, Y. G.; Gu, B. H.; Jiang, G. B.; Cai, Y., Evaluating the role of re-adsorption of dissolved Hg2+ during cinnabar dissolution using isotope tracer technique. J. Haz. Mater. 2016, 317, 466-475.
288 289 290 291 292 293
10. Harris, R. C.; Rudd, J. W. M.; Amyot, M.; Babiarz, C. L.; Beaty, K. G.; Blanchfield, P. J.; Bodaly, R. A.; Branfireun, B. A.; Gilmour, C. C.; Graydon, J. A.; Heyes, A.; Hintelmann, H.; Hurley, J. P.; Kelly, C. A.; Krabbenhoft, D. P.; Lindberg, S. E.; Mason, R. P.; Paterson, M. J.; Podemski, C. L.; Robinson, A.; Sandilands, K. A.; Southworth, G. R.; Louis, V. L. S.; Tate, M. T., Whole-ecosystem study shows rapid fish-mercury response to changes in mercury deposition. Proc. Natl. Acad. Sci. USA 2007, 104 (42), 16586-16591.
294 295 296
11. Olsen, T. A.; Brandt, C. C.; Brooks, S. C., Periphyton biofilms influence net methylmercury production in an industrially contaminated system. Environ. Sci. Technol. 2016, 50 (20), 1084310850.
297 298 299
12. Jonsson, S.; Skyllberg, U.; Nilsson, M. B.; Lundberg, E.; Andersson, A.; Bjorn, E., Differentiated availability of geochemical mercury pools controls methylmercury levels in estuarine sediment and biota. Nature Comm. 2014, 5, 4624.
300 301 302
13. Jonsson, S.; Skyllberg, U.; Nilsson, M. B.; Westlund, P. O.; Shchukarev, A.; Lundberg, E.; Bjorn, E., Mercury methylation rates for geochemically relevant hg-II species in sediments. Environ. Sci. Technol. 2012, 46 (21), 11653-11659.
14 ACS Paragon Plus Environment
Page 14 of 16
Page 15 of 16
Environmental Science & Technology Letters
303 304 305
14. Bloom, N.; Fitzgerald, W. F., Determination of volatile mercury species at the picogram level by low-temperature gas-chromatography with cold-vapor atomic fluorescence detection. Anal Chim Acta 1988, 208 (1-2), 151-161.
306 307 308
15. Bramanti, E.; D'Ulivo, A.; Lampugnani, L.; Zamboni, R.; Raspi, G., Application of mercury cold vapor atomic fluorescence spectrometry to the characterization of mercury-accessible -SH groups in native proteins. Anal Biochem 1999, 274 (2), 163-173.
309 310 311
16. Liu, Y.; Lu, X.; Zhao, L.; An, J.; He, J. Z.; Pierce, E. M.; Johs, A.; Gu, B., Effects of cellular sorption on mercury bioavailability and methylmercury production by Desulfovibrio desulfuricans ND132. Environ. Sci. Technol. 2016, 50, 13335-13341.
312 313 314
17. Zhao, L.; Chen, H.; Lu, X.; Lin, H.; Christensen, G. A.; Pierce, E. M.; GU, B., Contrasting effects of dissolved organic matter on mercury methylation by G. sulfurreducens PCA and D. desulfuricans ND132. Environ. Sci. Technol. 2017, 51, 10468–10475.
315 316 317
18. Dong, W.; Liang, L.; Brooks, S. C.; Southworth, G.; Gu, B., Roles of dissolved organic matter in the speciation of mercury and methylmercury in a contaminated ecosystem in Oak Ridge, Tennessee. Environ. Chem. 2010, 7 (1), 94–102.
318 319 320
19. Miller, C.; Southworth, G.; Brooks, S. C.; Liang, L.; Gu, B., Kinetic controls on the complexation between mercury and dissolved organic matter in a contaminated environment. Environ. Sci. Technol. 2009, 43, 8548-8553.
321 322 323
20. Hu, H.; Lin, H.; Zheng, W.; Tomanicek, S. J.; Johs, A.; Feng, X.; Elias, D. A.; Liang, L.; Gu, B., Oxidation and methylation of dissolved elemental mercury by anaerobic bacteria. Nature Geosci. 2013, 6 (9), 751-754.
324 325 326
21. Lin, H.; Morrell-Falvey, J. L.; Rao, B.; Liang, L.; Gu, B., Coupled mercury-cell sorption, reduction, and oxidation affecting methylmercury production by Geobacter sulfurreducens PCA. Environ. Sci. Technol. 2014, 48 (20), 11969-11976.
327 328 329
22. Miller, C. L.; Watson, D. B.; Lester, B. P.; Lowe, K. A.; Pierce, E. M.; Liang, L. Y., Characterization of soils from an industrial complex contaminated with elemental mercury. Environ. Res. 2013, 125, 20-29.
330 331 332
23. Bloom, N. S.; Preus, E.; Katon, J.; Hiltner, M., Selective extractions to assess the biogeochemically relevant fractionation of inorganic mercury in sediments and soils. Analytica Chimica Acta 2003, 479 (2), 233-248.
333 334 335
24. Dong, W.; Liang, L.; Brooks, S.; Southworth, G.; Gu, B., Roles of dissolved organic matter in the speciation of mercury and methylmercury in a contaminated ecosystem in Oak Ridge, Tennessee. Environ. Chem. 2010, 7 (1), 94-102.
336 337 338 339
25. Hintelmann, H.; Harris, R.; Heyes, A.; Hurley, J. P.; Kelly, C. A.; Krabbenhoft, D. P.; Lindberg, S.; Rudd, J. W. M.; Scott, K. J.; St Louis, V. L., Reactivity and mobility of new and old mercury deposition in a Boreal forest ecosystem during the first year of the METAALICUS study. Environ. Sci. Technol. 2002, 36 (23), 5034-5040.
340
15 ACS Paragon Plus Environment
Environmental Science & Technology Letters
12 Experiment Natural Hg 198 Hg 200 Hg 201 Hg 202 Hg
8
4
Peak intensity (10 counts)
10
Simulation Natural Hg 198 Hg 200 Hg 201 Hg 202 Hg 199 Hg 204 Hg
200
Hg
202
Hg
6 199
Hg
4
Natural
204
Hg
2
198
Hg
201
0
0
200
400
600
800
1000
Added Hg (pg)
ACS Paragon Plus Environment
Hg
1200
Page 16 of 16