Wireless Acoustic-Surface Actuators for Miniaturized Endoscopes

Nov 17, 2017 - Max Planck Institute for Intelligent Systems, 70569 Stuttgart, Germany. ‡ Department of ..... Figures of the experimental setup; meth...
0 downloads 9 Views 917KB Size
Subscriber access provided by Gothenburg University Library

Article

Wireless Acoustic-Surface Actuators for Miniaturized Endoscopes Tian Qiu, Fabian Adams, Stefano Palagi, Kai Melde, Andrew G Mark, Ulrich Wetterauer, Arkadiusz Miernik, and Peer Fischer ACS Appl. Mater. Interfaces, Just Accepted Manuscript • DOI: 10.1021/acsami.7b12755 • Publication Date (Web): 17 Nov 2017 Downloaded from http://pubs.acs.org on November 19, 2017

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

ACS Applied Materials & Interfaces is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

Wireless Acoustic-Surface Actuators for Miniaturized Endoscopes Tian Qiu1,#, Fabian Adams1,2,#, Stefano Palagi1, Kai Melde1, Andrew Mark1, Ulrich Wetterauer2, Arkadiusz Miernik2, Peer Fischer1,3,* 1

Max Planck Institute for Intelligent Systems, Stuttgart, Germany 2

Department of Urology, University Medical Centre Freiburg 3

Institute of Physical Chemistry, University of Stuttgart # Equal contribution *Corresponding author: [email protected]

Abstract Endoscopy enables minimally invasive procedures in many medical fields, such as urology. However, current endoscopes are normally cable-driven, which limits their dexterity and makes them hard to miniaturize. Indeed current urological endoscopes have an outer diameter of about 3 mm and still only possess one bending degree of freedom. In this paper, we report a novel wireless actuation mechanism that increases the dexterity and that permits the miniaturization of a urological endoscope. The novel actuator consists of thin active surfaces that can be readily attached to any device and are wirelessly powered by ultrasound. The surfaces consist of two-dimensional arrays of micro-bubbles, which oscillate under ultrasound excitation and thereby generate an acoustic streaming force. Bubbles of different sizes are addressed by their unique resonance frequency, thus multiple degrees of freedom can readily be incorporated. Two active miniaturized devices (with a side length of around 1 mm) are demonstrated: a miniaturized mechanical arm that realizes two degrees of freedom, and a flexible endoscope prototype equipped with a camera at the tip. With the flexible endoscope, an active endoscopic examination is successfully performed in a rabbit bladder. This results show the potential medical applicability of surface actuators wirelessly powered by ultrasound penetrating through biological tissues.

Keywords active surface, wireless actuator, endoscopy, ultrasound, acoustic streaming

1 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 2 of 17

Introduction Flexible endoscopy is an important diagnostic tool in many medical fields including urology. Its major applications include the detection of hematuria, the diagnosis of bladder carcinoma and other pathologies, as well as the visualization of the upper urinary tract. Bladder cancer is the most common malignancy of the urinary tract 1, with annually 74,000 new cases in the USA and an estimated 16,000 deaths 2. Cystoscopic controls are necessary not only for the initial diagnose of bladder cancer but also for follow-up examinations. Patients afflicted with bladder cancer undergo mostly repeated endoscopies of the urinary bladder using flexible devices. Compared to rigid instruments, the tip of the flexible endoscope benefits the procedure due to larger coverage areas and reduced trauma. Unlike larger flexible endoscopes, such as colonoscopes, current flexible thin endoscopes (overall diameter ~3 mm), as are for instance used in urology, only have one bending section near the tip. Apart from the overall movements of the entire endoscope (push/pull and rotate by the operator), only one bending degree of freedom (DoF) can be realized in the tip itself. As they are cable-driven, they are hard to be miniaturized, for example with a sheath diameter of 8.7 Fr. (2.9 mm diameter) 3. Furthermore, two cables are required to drive one rotational DoF, and adding more DoFs or more bending sections will result in an even larger device diameter, as is the case for some colonoscopes 4, which are then so large that they will no longer be suitable for urological endoscopic applications. These disadvantages motivate the following technical developments: 1) Miniaturization, to reduce the diameter of the endoscope; 2) Multiple DoFs, to achieve two bending degrees-of freedom of the endoscope tip and more bending sections; 3) Wireless actuation, to remove the constraint of the cable. The overall goal is to achieve more space coverage with minimal damage to the human tissue and minimal auxiliary medical measures, e.g. avoid general anesthesia. The ultimate solution is a wireless micro-robot that can actively move through a lumen or swim through biological fluids

5-6

to perform medical inspection or treatment. In the quest to realize active endoscopes, different

actuation schemes have been investigated, such as shape memory alloy actuators 7, sliding concentric tubes pneumatic/hydraulic actuators

10

8-9

and

. These tethered approaches offer large actuation force and precise control of the

position, however, their structures are often complicated and they are rigid and difficult to be miniaturized. Wireless approaches, such as magnetics, have been also applied, e.g. to steer magnetic tips of catheters capsule endoscopes

12-14

11

or untethered

. However, large magnetic fields and/or gradients might be required to achieve the forces

needed for operation, which results in the need for cumbersome and expensive dedicated equipment. It is necessary

2 ACS Paragon Plus Environment

Page 3 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

to develop a new kind of actuator which has a simple structure and is powered wirelessly and that therefore can be miniaturized and integrated more efficiently. Ultrasound has been widely used in urology for imaging, tissue ablation, lithotripsy and drug delivery, and it has also been proposed as a means to transfer power wirelessly.15-16 It has recently been shown that ultrasound can also be shaped to form complex 3D patterns and this provides additional opportunities for medical ultrasound.

17

However,

the common approaches convert the mechanical vibrations to electricity using the piezoelectric effect and then utilize the electricity to power a device. This conversion process adds additional complexity and suffers from low efficiency. Recently, we developed an actuator that directly converts ultrasound power into mechanical work via acoustic streaming from a surface-array of micro-bubbles.

18-19

Acoustic streaming is a steady fluid flow driven by

acoustically driven oscillations, which can be achieved at the transducer surface or at the air-liquid interface. 20-21 It is well known that oscillating bubbles in fluids can cause resonant oscillations and thus acoustic streaming,22 and microfluidic pumps and mixers have been developed based on this phenomenon 23. It was shown by Dijkink et al. that oscillating bubbles in a tube can propel or power a rotary motor at centimeter scale propelled by one or a few bubbles have been reported by several groups

25-26

24

, and micro-swimmers

. However, the powering of a

miniaturized medical instrument based on acoustic active surfaces has not yet been reported. In this paper, we report the application of our wireless acoustic surface actuators, to a miniaturized, multi-DoFs, ultrasound-powered endoscope prototype. As illustrated in the schematic of Fig. 1, the device is actuated by acoustic streaming, which is generated from an active surface consisting of a two-dimensional array of micro-bubbles. Bubbles of different sizes are addressed with unique ultrasound frequency, thus multiple DoFs are achieved. Two miniaturized devices with side length around 1 mm were developed, i.e. a 2 DoFs miniaturized mechanical arm (mini-arm) and a flexible endoscope equipped with a camera. The endoscope is inserted into a rabbit bladder through the tool channel of a rigid cystoscope, and is successfully used for a cystoscopic examination.

Methods Fabrication of the active surface The active surface consists of a regular array of cylindrical micro-cavities with a carefully chosen size of the cavity. Photolithography was used to fabricate the surface as reported previously

18

3 ACS Paragon Plus Environment

. In short, negative photoresist (SU-8

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

2000, MicroChem, MA, US) was spin-coated on a silicon wafer (WS-650 spin coater, Laurel, PA, US), exposed on a mask aligner (MJB4, SUSS MicroTec, Garching, Germany) with a chromium mask (Delta Mask, Enschede, Netherlands), developed and hard baked. Then, the wafer was diced by an automatic dicing saw (DAD 321, DISCO, Tokyo, Japan) to a rectangular array. When a dried active surface is immersed under water, air is trapped in the cavities as bubbles, due to the small size of the cavity and the hydrophobicity of the SU-8 photoresist (~90° contact angle 27). The diameter of the bubble is the same as the cavity, and the length is normally shorter than the cavity, as shown in the inset of Fig. 2. The size of the bubble determines the resonance frequency of a single bubble. This corresponds to the frequency at which the maximal acoustic streaming of the active surface occurs, as reported in our previous study 18. Sodium dodecyl sulfate (SDS, 2.4 g/L) is added to the solution as a surfactant to lower the surface tension of water from 73 mN/m to 38 mN/m 28. Our previous study shows both theoretically and experimentally that reducing the surface tension increases the oscillation amplitude of the air-liquid interface, which substantially increases the acoustic streaming force (up to ~3.5 times) 28. Assembly of the miniaturized endoscope The mini-arm has a dimension of 1 mm × 1 mm in cross-section and is 5 mm long. It contains three sections with two hinges, which are mounted orthogonal to each other (Fig. 2(a) middle), such that the arm can achieve two DoF. The sections and the pins were made out of aluminum and brass, respectively. After assembly of the mini-arm, two active surfaces (1 mm × 4 mm) with different micro-cavity sizes, i.e. actuator 1: an array of 9600 “small” cavities, each 55 µm in depth, 10 µm in diameter, 10 µm in spacing; actuator 2: an array of 1520 “large” cavities, each 120 µm in depth, 30 µm in diameter, 20 µm in spacing, were attached to two orthogonal surfaces of the 1st section using cyanoacrylate (Superflex Gel, UHU, Bühl, Germany). The flexible arm was assembled and integrated with a Naneye 2D CMOS sensor (Awaiba, Yverdon-Les-Bains, Switzerland), which is 1 mm × 1 mm in cross-section and 1.85 mm long. The flat wire connected to the Naneye is a flexible ribbon of 4 wires with a total dimension of 0.72 mm in width and 0.188 mm in height, which also serves as an elastic sheet to provide the restoring force in the flexible arm design. An active surface (1 mm × 9 mm, with “large” cavities) is attached to one side of the flat wire using cyanoacrylate.

4 ACS Paragon Plus Environment

Page 4 of 17

Page 5 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

Ultrasound excitation set-up A customized ultrasound excitation set-up was made to test the endoscope. It is a glass tank of 50 × 50 × 50 mm3 with one piezoelectric ceramic disk (ϕ50 mm × 3 mm, PIC 181, PI Ceramic, Lederhose, Germany) attached to one side wall by epoxy glue. To increase the propulsive force and stabilize the bubbles, 2.4 g/L SDS surfactant was added in all fluids 28. The driving signal was sinusoidal wave generated from a function generator (33220A, Agilent technologies, CA, US), and amplified 30 times with a customized power amplifier and transformer. Two frequencies, i.e. 51.3 kHz and 115.3 kHz, were used to respectively address the arrays of “large” and “small” bubbles, respectively. For the wireless actuation of the endoscope in the rabbit bladder, an ultrasonic bath (2510E-DTH, Bransonic, CT, US) was used for higher power ultrasound of 100 W at 42 kHz ± 6%. Endoscope actuation in the rabbit bladder The cystoscopy with the ultrasound-powered endoscope is demonstrated in an isolated rabbit bladder (Oryctolagus cuniculus forma domestica). The organs were obtained from a local slaughterhouse. The careful representation of the urinary bladder in the retroperitonal convolute was followed by the preparation including ligation of both ureters and the removal of the bladder including the proximal urethra. To observe the movement of the endoscope inside the bladder, a rigid 8 Fr. cystoscope (Richard Wolf, Knittlingen, Germany) was used. The bladder was fixed to the cystoscope by a 4-0 Vicryl suture (Ethicon, Somerville, USA) and filled with ~100 mL SDS aqueous solution. The developed flexible endoscope was introduced by a 3 Fr. flexible gripper (Richard Wolf) through the tool channel of the cystoscope into the bladder, as shown in detail in Fig. S-1. The illumination and visualization were achieved by a 5132 Auto LP HLight Module and a 5520 1CCD Endocam Module (Richard Wolf) connected to the rigid cystoscope. The video from the cystoscope and the flexible miniaturized endoscope are recorded and shown as Supporting Videos.

Results and discussion Wireless actuation of the mini-arm

5 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

As shown in Fig. 3, the mini-arm consists of three sections connected in series and can achieve two rotational DoF. Two active surfaces with different micro-cavity sizes are attached to two orthogonal surfaces on the 1st section. The mini-arm is mechanically held under water by tweezers. The mini-arm contains no electrical components or connections. Instead, the mini-arm is powered completely wirelessly by ultrasound. The ultrasound that is used to excite the arm is generated by a piezoelectric transducer which can also be seen in Fig. 3, behind the arm at a distance of ~30 mm. Once the mini-arm is put under water, air is trapped in the micro-cavities of the active surfaces as individual bubbles, due to the surface hydrophobicity and small diameter of the cavities (see Methods). The size of the micro-bubbles is determined by the cavity size, which has a unique different acoustic resonant frequency. A detailed physical study of the frequency response of the bubbles is reported in a previous work 18. In short, the resonant frequencies of micron sized bubbles lie in the range of 50-200 kHz, and the smaller the bubble size, the higher is its resonant frequency. For the actuation of the mini-arm, an ultrasonic field of 115.3 kHz was first applied and it only excited the “small” bubbles on the actuator 1, which caused ultrasonic streaming and thus propelled the 1st section rotating to the left (image at 11 s in Fig. 3 (b)). When the ultrasound was switched off, the arm returned to the original position due to gravity. Similarly, when an ultrasound of 51.3 kHz was applied, only actuator 2 was activated, thus the 2nd section of the mini-arm rotated out of the plane (image at 24 s in Fig. 3 (b)). See also Supplementary Video 1 for the real time movement of the mini-arm. The reproducibility is demonstrated by switching each frequency on and off for two times (see video). Increasing the DoF in the tendon-driven endoscope requires adding at least two more driving wires, thus the endoscope structure is more rigid, complicated and the diameter is larger. The wireless driven active surface reported here solves this problem. The bandwidth response of each kind of actuator is about 10 kHz, thus in principle 18 independent actuators can be achieved within the 20 – 200 kHz frequency range. Whereas for cystoscopy 3 DoFs (two rotational DoFs  ,  and one translational DoF, see Fig. 1) are normally enough, for more complicated intraluminal anatomy, such as the upper urinary tract, more DoFs will be beneficial. If the hinges are arranged in orthogonal directions, three arm sections will realize three independent rotational DoFs. Since the back and forth motions of one DoF requires two actuators, six actuators in total will be needed. The redundant DoFs could be useful for applications other than bending the endoscope, for example, to open and close a gripper. More importantly, as all actuators are driven wirelessly, no additional wire is required for more DoFs, and the overall diameter of the

6 ACS Paragon Plus Environment

Page 6 of 17

Page 7 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

endoscope can remain very small. It is noteworthy that each active surface can be 100–200 µm thin, therefore equipping the endoscope tip with 2 DoFs (4 active surfaces) will increase its diameter of much less than half a millimeter. Moreover, even more active surfaces can be added at different positions along the articulated tip without making the diameter any larger. Notably, the streaming actuation force is independent of the direction of the incoming ultrasound, which can be applied from the most convenient position. It is also interesting to point out that the adoption of active surfaces is a scalable approach. The mini-arm is designed to be 1 mm × 1 mm in cross-section, because the smallest commercial camera has a cross-section of 1 mm × 1 mm. In other words, 1 mm is far from the size limit for the active surface, and the principle illustrated here is still valid for even smaller devices to micrometer range. Wireless actuation of the flexible endoscope Although the actuation scheme is wireless, there is currently no wireless camera small enough to fit through the tool channel of a cystoscope. Therefore a tethered camera such as the Naneye 2D sensor was used. The design of the flexible arm takes advantage of the connecting wires for the camera, and adopts it as an elastic sheet to provide the restoring force when the ultrasonic transducer is turned off. Fig. 4 (a) and (b) show the movement of the flexible endoscope, which consists of three parts: a miniaturized camera at the tip pointing forward, a flat wire connecting the camera and also acting as a recoil spring, and an active surface. The effective bending length of the wire is 22.5 mm, and the streaming active surface is attached next to the camera causing the deflection in one direction. The deflection is driven by an ultrasound transducer ~30 mm away, and is controlled by the driving voltage which shows a linear relationship to the output power in this range (Fig. 4(c)). The total length of the flexible arm is about 33 mm, which is realistic for cystoscopy in a human bladder. A normal human bladder capacity is ~400 ml 29, so considered as a sphere, a filled human bladder has an average radius of approximately 46 mm. This also means the miniaturized camera can be quite close to the bladder inner surface, thus providing a clearer image of the bladder wall. When the wire is thinner or softer, or the ultrasound power is increased, it is possible to realize even larger deflections. The flexible arm has a simpler design compared with the above mentioned mini-arm, i.e. no mechanical hinges are required and it can be fabricated more easily. There are ~3400 bubbles on an area of 1 mm × 9 mm surface, each generating a force up to ~120 nN (bubbles 120 µm in depth, 30 µm in diameter) 30. Therefore, the estimated output force generated by the active surface is ~0.4 mN. The force is smaller than those typically generated by tethered

7 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 8 of 17

approaches, but large enough to effectively steer a miniaturized camera, as demonstrated in the experiments. As the force is small and the flexible endoscope prototype is compliant, the actuator-camera setup should be safe for the urinary tract wall inspection and cause no damage to the tissue.31 Furthermore, such systems might be potentially introduced via urologic catheters such as ureteric catheters replacing conventional endoscopic inspection of the urinary tract by rigid, semi-rigid and flexible devices. Safety and stability of the ultrasonic streaming actuation Safety is one of the most important concerns of medical devices, especially for invasive procedures. The safety of the ultrasound powered acoustic streaming is discussed here. It is difficult to compare the values directly with current ultrasonic medical devices, because in diagnostic ultrasound, a higher frequency (2 MHz) coupled with pulse mode operation is used. The ultrasound used here is continuous wave with a frequency between 50 and 200 kHz. Two parameters are often used to represent the ultrasound intensity, i.e. the spatial peak-temporal average intensity  and the spatial peak-pulse average intensity  . Typical pressure amplitude in a B-scan is 1.68 MPa

32

, and the

recommended intensity limits set by the U.S. FDA for use in a peripheral vessel are 720 mW/cm2 for  and 190 W/cm2 for  33. A continuous wave means the duty cycle is 100%, therefore  =  =  in this case. Currently, there is no safety guideline for continuous ultrasound, however, a previous study shows that the tissue injury intensity threshold of animal tissues for continuous ultrasound is 470 W/cm2 34. Measured by a needle hydrophone, the pressure amplitudes of the ultrasound used to power the mini-arm and the flexible endoscope are both around 100-200 kPa, which is much lower than 1.68 MPa in the B-scan. The corresponding ultrasound intensity can be calculated by the following equation 35:

 =



where  is the spatial-peak ultrasound intensity, is the pressure amplitude, is the density of water (1000 kg/m3), is the speed of sound in water which is ~1520 m/s at 37 °C. If we take the highest pressure =200 kPa, then  =2.6 W/cm2. This value is higher than the FDA recommended value for  720 mW/cm2, which may lead to some thermal effects 32; but two orders of magnitude lower than the FDA recommended value for  190 W/cm2 and the reported tissue injury intensity threshold 470 W/cm2, so it is low enough that it is unlikely that the ultrasound used here will cause lasting mechanical damage or induce cavitation

32

. However, further studies on animal

8 ACS Paragon Plus Environment

Page 9 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

experiments are required to establish more carefully if and when one may expect thermal effects of continuous wave ultrasound on tissue. Another issue is the stability of the air bubbles. It is observed in the experiments that the air bubbles trapped in the active surface are stable under ultrasonic excitation for at least 30 min, and the streaming actuation can be excited repeatedly during this period. Under the microscope, we observed that the air bubbles will gradually dissolve in water and the bubble size will become smaller, thus the driving frequency needs to be adjusted and be increased accordingly. But the dissolution process is relatively slow compared to the operating time of a cystoscope (depending on the examiner normally 5-15 min for an examination). The dimension of the bubble only changes by about 0.8% per minute (see Fig. S-3 in the Supporting Information). The dimension change is slow enough that no frequency adjustments are needed during the typical operation time of the active surface (~5-15 minutes for a cystoscopy examination). Changing air to other inert gases that are insoluble in water, such as SF6, a commonly used contrast agent for ultrasound imaging 36, will result in even longer operating time. SDS is added to the solution as a surfactant to lower the surface tension, increase the oscillation amplitude of the air-liquid interface, and thus increase the acoustic streaming force (up to ~3.5 times) 28. SDS was chosen as it is readily available in the laboratory, but it is not supposed to be used in clinical applications. Many biocompatible surfactants, such as amino acid-based surfactants 37

, can instead be used as safe substitutes with comparable effects on the acoustic actuation. Particular attention to

safety and biocompatibility will be given in the future translational studies. Miniaturized endoscope for cystoscopy in a rabbit bladder The flexible endoscope was tested in a rabbit bladder ex vivo. The rigid cystoscope is used to hold and refill the bladder (Fig. 5), and visualize the movement of the flexible endoscope inside the bladder. However, in a real application the cystoscope is not necessary, as the flexible endoscope can be inserted into the bladder through the urethra alone (as illustrated in Fig. 1) or through a passive catheter. In clinical practice, cystoscopy often starts with an empty bladder without urine, then in order to obtain a clear view of the bladder wall, the operator fills the bladder with saline solution through the irrigation channel of the cystoscope. As there is no irrigation channel in our miniaturized endoscope prototype due to its small size, another thin catheter could be used to fill the bladder with saline solution before the examination by the acoustically actuated endoscope, which requires an aqueous environment to operate.

9 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 10 of 17

The miniaturized endoscope entered the bladder through the working channel of the endoscope, at first it was stationary pointing downwards (image at 0 s in Fig. 6(a)). When the ultrasound was applied, streaming from the active surface occurred and the flexible arm or the endoscope is bent to the right (3 s in Fig. 6(a)). The streamline is weakly visualized in the image, but clearer seen in the Supplementary Video S3. When the ultrasound was turned off, the flexible arm returned to the original position due to the restoring force of the flat wire (6 s in Fig. 6(a)). We found that the bubbles on the active surface were stable during the whole period of examination (~10 min) and contact with the bladder wall did not release the bubbles. From the miniaturized camera on the tip of the endoscope, another video was recorded (snapshots in Fig. 6(b)). When the flexible arm was pointing downwards, the structure of the fat tissue was seen as a marker and can be directly compared to the cystoscopic images in Fig. 6(a). When the flexible arm is bent, a blood vessel on the bladder sidewall can be visualized. This vessel can be seen from the outside of the bladder in Fig. 5 (b), but because of its position it could not be visualized with the cystoscope in Fig. 6(a). In the Supplementary Video S3, the two videos from the cystoscope (left) and miniaturized camera (right) are synchronized, and to demonstrate the reproducibility of the imaging procedure the ultrasound is switched on and off for two times. In the video, a certain level of shake of the camera is observed, which can be a problem in endoscopy. The reason is that the ultrasound power is not very stable in the experiment and the device is operated in an openloop control. The image stability can be improved by correcting for fluctuations in ultrasound power, e.g. with feedback from an in situ pressure sensor, or by sensing the bending angle of the endoscope. The miniaturized camera has a high frame rate over 60 frames per second, which generates redundant frames, thus the video can be digitally processed with anti-shaking algorithms to minimizing the shaking effect. The current image quality is still not comparable to the current endoscope, but with the fast development of microelectronic devices, we believe, the imaging quality of the miniaturized camera can be further improved and the miniaturized wireless device will be the trend in the future endoscopy. Although, due to smaller dimensions, a rabbit’s bladder was studied as a proof-of-concept, this work reports a general method to actuate a miniaturized endoscope in low viscosity bodily fluids. The miniaturization of the instrument leads to a significant reduction of surgical trauma and thus reducing unwanted complications for the patient. It not only can be used as a cystoscope, but may also be applied to the upper urinary tract and the kidney. Alternatively the flexible endoscope can be tested in a realistic phantom of the kidney and urinary tract

38

. It can

potentially simplify preparatory steps of surgery procedures, for example, a double-J catheter might be required to

10 ACS Paragon Plus Environment

Page 11 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

dilate the ureter for a period of time prior the surgical manipulation on the upper urinary tract via an endoscope. With the actuation scheme developed herein, an instrument with less than 1 mm diameter could potentially avoid unnecessary interventions. Consequently the severeness of urological endoscopic surgeries could be decreased resulting in shortened hospital stay of the patients and thereby contribute to a cost reduction.

Conclusion In this paper, we report the first thin-layer surface actuators that permit the wireless power and control of an endoscope. The wireless actuation is based on two-dimensional arrays of micro-bubbles that resonate with an external ultrasound, giving rise to frequency-selective, directional acoustic streaming. The endoscope has an overall side dimension of 1 mm, and fits through the tool channel of a cystoscope. A complete cystoscopy was successfully performed by this endoscope in a rabbit bladder. The novel actuation scheme is general and can therefore be further miniaturized. The method demonstrated herein thus holds potential as a powerful actuator to steer miniaturized devices for minimally-invasive in vivo inspections.

Acknowledgement The research was in part supported by the European Research Council under ERC Proof of Concept grant number 737530. F.A. acknowledges support through Ferdinand Eisenberger-Forschungsstipendium 2015, Nr.: AdF1/FE-15. We thank the Richard Wolf GmbH for the loan of endoscope systems.

Supporting Information Videos of the wireless actuation of the endoscope; Figures of the experimental set-up; Methods of the measurement of the bubble stability over time.

References 1. Cheung, G.; Sahai, A.; Billia, M.; Dasgupta, P.; Khan, M. S., Recent Advances in the Diagnosis and Treatment of Bladder Cancer. BMC Med. 2013, 11, 13. 2. Siegel, R. L.; Miller, K. D.; Jemal, A., Cancer Statistics. Ca-Cancer J Clin. 2015, 65, 5-29. 3. Babjuk, M.; Bohle, A.; Burger, M.; Capoun, O.; Cohen, D.; Comperat, E. M.; Hernandez, V.; Kaasinen, E.; Palou, J.; Roupret, M.; van Rhijn, B. W. G.; Shariat, S. F.; Soukup, V.; Sylvester, R. J.; Zigeuner, R., EAU Guidelines on Non-Muscle-invasive Urothelial Carcinoma of the Bladder: Update 2016. European Urology. 2017, 71, 447-461. 11 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 12 of 17

4. Olympus PCF-H190D Colonoscope Specifications. http://www.olympuseuropa.com/medical/en/medical_systems/products_services/product_details/product_details_98437.jsp. 5. Qiu, T.; Lee, T. C.; Mark, A. G.; Morozov, K. I.; Munster, R.; Mierka, O.; Turek, S.; Leshansky, A. M.; Fischer, P., Swimming by Reciprocal Motion at Low Reynolds Number. Nat. Commun. 2014, 5, 5119. 6. Palagi, S.; Mark, A. G.; Reigh, S. Y.; Melde, K.; Qiu, T.; Zeng, H.; Parmeggiani, C.; Martella, D.; Sanchez-Castillo, A.; Kapernaum, N.; Giesselmann, F.; Wiersma, D. S.; Lauga, E.; Fischer, P., Structured Light Enables Biomimetic Swimming and Versatile Locomotion of Photoresponsive Soft Microrobots. Nat. Mater. 2016, 15, 647-653. 7. Peirs, J.; Reynaerts, D.; Van Brussel, H., Design of a Shape Memory Actuated Endoscopic Tip. Sensor Actuat a-Phys. 1998, 70, 135-140. 8. Dupont, P. E.; Lock, J.; Itkowitz, B.; Butler, E., Design and Control of Concentric-Tube Robots. IEEE Trans. Robot. 2010, 26, 209-225. 9. Robert J. Webster, I.; Jones, B. A., Design and Kinematic Modeling of Constant Curvature Continuum Robots: A Review. The International Journal of Robotics Research. 2010, 29, 1661-1683. 10. Ashwin, K. P.; Jose, D. P.; Ghosal, A., Modeling and Analysis of a Flexible End-effector for Actuating Endoscopic Catheters. Proceedings of the Fourteenth International Federation for the Promotion of Mechanism and Machine Science World Congress. 2015, 329-336. 11. Chautems, C.; Nelson, B. J. In The Tethered Magnet: Force and 5-DOF Pose Control for Cardiac Ablation, IEEE International Conference on Robotics and Automation (ICRA), IEEE: 2017; pp 48374842. 12. Menciassi, A.; Valdastri, P.; Quaglia, C.; Buselli, E.; Dario, P., Wireless Steering Mechanism with Magnetic Actuation for an Endoscopic Capsule. IEEE Eng. Med. Bio. 2009, 1204-1207. 13. Carpi, F.; Kastelein, N.; Talcott, M.; Pappone, C., Magnetically Controllable Gastrointestinal Steering of Video Capsules. IEEE Trans. Bio-Med. Eng. 2011, 58, 231-234. 14. Yim, S.; Gultepe, E.; Gracias, D. H.; Sitti, M., Biopsy using a Magnetic Capsule Endoscope Carrying, Releasing, and Retrieving Untethered Microgrippers. IEEE Trans. Bio-Med. Eng. 2014, 61, 513-521. 15. Mazzilli, F.; Thoppay, P. E.; Praplan, V.; Dehollaini, C. In Ultrasound Energy Harvesting System for Deep Implanted-medical-devices (IMDs), IEEE International Symposium on Circuits and Systems (ISCAS), 20-23 May 2012; pp 2865-2868. 16. Charthad, J.; Weber, M. J.; Ting Chia, C.; Saadat, M.; Arbabian, A. In A mm-sized Implantable Device with Ultrasonic Energy Transfer and RF Data Uplink for High-power Applications, IEEE Proceedings of the Custom Integrated Circuits Conference (CICC), 15-17 Sept. 2014; pp 1-4. 17. Melde, K.; Mark, A. G.; Qiu, T.; Fischer, P., Holograms for Acoustics. Nature. 2016, 537, 518522. 18. Qiu, T.; Palagi, S.; Mark, A. G.; Melde, K.; Adams, F.; Fischer, P., Wireless Actuation with Functional Acoustic Surfaces. Appl. Phys. Lett. 2016, 109, 191602. 19. Qiu, T. Microdevices for Locomotion in Complex Biological Fluids. PhD thesis, École polytechnique fédérale de Lausanne EPFL, Lausanne, 2016. 20. Lighthill, S. J., Acoustic Streaming. Journal of Sound and Vibration. 1978, 61, 391-418. 21. Wiklund, M.; Green, R.; Ohlin, M., Acoustofluidics 14: Applications of Acoustic Streaming in Microfluidic Devices. Lab on a Chip. 2012, 12, 2438-2451. 22. Tho, P.; Manasseh, R.; Ooi, A., Cavitation Microstreaming Patterns in Single and Multiple Bubble Systems. J. Fluid Mech. 2007, 576, 191-233. 23. Hashmi, A.; Yu, G.; Reilly-Collette, M.; Heiman, G.; Xu, J., Oscillating Bubbles: a Versatile Tool for Lab on a Chip Applications. Lab on a Chip. 2012, 12, 4216-4227. 24. Dijkink, R. J.; van der Dennen, J. P.; Ohl, C. D.; Prosperetti, A., The 'Acoustic Scallop': a Bubblepowered Actuator. J. Micromech. Microeng. 2006, 16, 1653-1659. 25. Feng, J.; Cho, S. K. In Two-Dimensionally Steering Microswimmer Propelled by Oscillating Bubbles, 27th IEEE International Conference on Micro Electro Mechanical Systems (MEMS), 2014; pp 188-191. 12 ACS Paragon Plus Environment

Page 13 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

26. Ahmed, D.; Lu, M.; Nourhani, A.; Lammert, P. E.; Stratton, Z.; Muddana, H. S.; Crespi, V. H.; Huang, T. J., Selectively Manipulable Acoustic-powered Microswimmers. Sci. Rep. 2015, 5, 9744. 27. Kumar, V.; Sharma, N. N., Synthesis of Hydrophilic to Superhydrophobic SU8 Surfaces. J. Appl. Polym. Sci. 2015, 132, 41934. 28. Qiu, T.; Palagi, S.; Mark, A. G.; Melde, K.; Fischer, P. In Wireless Actuator Based on Ultrasonic Bubble Streaming, International Conference on Manipulation, Automation and Robotics at Small Scales (MARSS), 18-22 July 2016; pp 1-5. 29. Abrams, P., Urodynamics. Springer: London, 2006. 30. Qiu, T.; Palagi, S.; Mark, A. G.; Melde, K.; Adams, F.; Fischer, P., Active Acoustic Surfaces Enable the Propulsion of a Wireless Robot. Advanced Materials Interfaces. 2017, 4, 1700933. 31. Wood, R.; Walsh, C., Smaller, Softer, Safer, Smarter Robots. Sci. Transl. Med. 2013, 5, 210ed19. 32. Bushberg, J. T., The Eessential Physics of Medical Imaging. 3rd ed.; Lippincott Williams & Wilkins: Philadelphia, 2002. 33. FDA, Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers. 2008. 34. Wang, Y. N.; Simon, J. C.; Cunitz, B. W.; Starr, F. L.; Paun, M.; Liggitt, D. H.; Evan, A. P.; McAteer, J. A.; Liu, Z.; Dunmire, B.; Bailey, M. R., Focused Ultrasound to Displace Renal Calculi: Threshold for Tissue Injury. Journal of Therapeutic Ultrasound. 2014, 2, 5. 35. Rienstra, S. W.; Hirschberg, A., An Introduction to Acoustics. 2004. http://www.win.tue.nl/~sjoerdr/papers/boek.pdf. 36. Schneider, M., Characteristics of SonoVue™. Echocardiography. 1999, 16, 743-746. 37. Moran, M. C.; Pinazo, A.; Perez, L.; Clapes, P.; Angelet, M.; Garcia, M. T.; Vinardell, M. P.; Infante, M. R., "Green" Amino Acid-based Surfactants. Green Chem. 2004, 6, 233-240. 38. Adams, F.; Qiu, T.; Mark, A.; Fritz, B.; Kramer, L.; Schlager, D.; Wetterauer, U.; Miernik, A.; Fischer, P., Soft 3D-Printed Phantom of the Human Kidney with Collecting System. Ann. Biomed. Eng. 2017, 45, 963-972.

13 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 14 of 17

Figure 1. Schematic of the wireless actuated miniaturized cystoscope. (a) The miniaturized cystoscope performs bladder imaging with two rotational DoFs and one translational DoF. (b) An enlarged view of the cystoscope tip. The active surface, which is powered wirelessly by ultrasound, rotates the 1st section relative to the 2nd section. (c) An enlarged view of the active surface, which consists of an array of micro-bubbles. The oscillation of the interface between air and liquid causes liquid streaming, generates a propulsion force in the opposite direction and actuates the device.

Figure 2. Wirelessly actuated endoscope powered by ultrasonic active surface. (a) A picture comparing a commercial flexible urological endoscope with 2.9 mm in diameter (left), with the mini-arm (middle) and flexible endoscope prototype (right) developed in this work. (b) An enlarged microscopic image of the red dashed box in (a) shows the active surface under water, which consists of an array of micro-bubbles with controllable sizes. The inset is a zoomin of the bubble in the white dashed box, which is trapped in a cavity of 30 µm in diameter and 120 µm in height.

14 ACS Paragon Plus Environment

Page 15 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

Figure 3. Wireless actuation of the mini-arm. (a) Sketch of the miniaturized arm made of aluminum and assembled through brass hinges. Two actuators 1 and 2 are attached orthogonally to the 1st section, which actuates the arm with 2 rotational DoF. (b) Sequential snapshots of a video. At 9-11 s, 1st section is actuated with ultrasound at 115.3 kHz; and at 20-24 s, the 2nd section is actuated with ultrasound at 51.3 kHz. See also Supplementary Video S1.

Figure 4. Wireless actuation of the flexible endoscope prototype. (a) Sketch of the flexible arm. The active surface is attached to the flat wire, which serves as an elastic sheet to provide the restoring force. (b) A superimposed image showing the deflection of the flexible arm is controlled by the driving voltage of the ultrasonic transducer. See also Supplementary Video S2. (c) The deflection exhibits a linear relationship with the driving voltage, and reaches a maximum of 60° in the current set-up.

15 ACS Paragon Plus Environment

ACS Applied Materials & Interfaces

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 16 of 17

Figure 5. The set-up for cystoscopy in rabbit bladder. (a) A schematic shows the flexible arm entering the bladder through the tool channel of a rigid cystoscope, which is fixed to urethra. The bending of the flexible arm is powered by the streaming of the active surface, and its movement is viewed via the cystoscope. (b) A picture of the bladder fixed with the endoscope. (c) An enlarged picture of the flexible arm coming out of the tool channel of a cystoscope.

Figure 6. Cystoscopy in a rabbit bladder. (a) Sequential snapshots of a video recorded by the rigid cystoscope. At 0 s, the flexible arm is stationary and pointing downwards. At 3 s, ultrasound is turned on and the arm is bent to the right due to the streaming of the active surface. At 6 s, ultrasound is turned off, the flexible arm returns to the original position. (b) Sequential snapshots of a video recorded by the Naneye camera on the flexible arm. At 0 s, the flexible arm is pointing downwards, and the shape of the fat tissue is seen as a marker comparing with the cystoscopic images above. At 3 s, when the flexible arm bends, a blood vessel on the bladder sidewall is visualized, which can be seen from outside the bladder in Fig. 5 (b), but cannot be seen from the cystoscopic images in (a). See also Supplementary Video S3.

16 ACS Paragon Plus Environment

Page 17 of 17

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Applied Materials & Interfaces

ToC figure 70x70mm (300 x 300 DPI)

ACS Paragon Plus Environment