Subscriber access provided by Bethel University
Article
DYNAMIC CONTROL OF GAS CHROMATOGRAPHIC SELECTIVITY DURING THE ANALYSIS OF ORGANIC BASES Ernest Darko, and Kevin B. Thurbide Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.9b00703 • Publication Date (Web): 01 May 2019 Downloaded from http://pubs.acs.org on May 1, 2019
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46
Analytical Chemistry
DYNAMIC CONTROL OF GAS CHROMATOGRAPHIC SELECTIVITY DURING THE ANALYSIS OF ORGANIC BASES by Ernest Darko and Kevin B. Thurbide* Department of Chemistry, University of Calgary, 2500 University Drive, NW, Calgary, Alberta T2N 1N4, Canada.
Submitted for publication as a Research Article in: Analytical Chemistry
*Corresponding Author Phone: (403) 220-5370 Fax: (403) 289-9488 E-mail:
[email protected] ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 2 of 25 2
47
ABSTRACT
48
A novel method for controlling selectivity during the gas chromatographic (GC) analysis
49
of organic bases is presented. The technique employs tandem stainless steel capillary columns,
50
each coated with a pH adjusted water stationary phase. The first is a 0.5 m trap column coated
51
with a pH 2.2 phase, while the second is an 11 m analytical column coated with a pH 11.4 phase.
52
The first column traps basic analytes from injected samples, while the remaining components
53
continue to elute and separate. Then, upon injection of a volatile aqueous ammonia solution, the
54
basic analytes are released as desired to the analytical column where they are separated and
55
analyzed. Separations are quite reproducible and demonstrate an average RSD of 1.2% for
56
analyte retention times in consecutive trials. Using this approach, the retention of such analytes
57
can be readily controlled and they can be held in the system for periods of up to 1 hour without
58
significant erosion of peak shape. As such, it can provide considerable control over analyte
59
selectivity and resolution compared to conventional separations. Further, by employing a third
60
conventional GC column to the series, both traditional hydrocarbon and enhanced organic base
61
separations can be performed. The method is applied to the analysis of complex mixtures, such
62
as gasoline, and much less matrix interference is observed as a result. The findings indicate that
63
this approach could be a useful alternative for analyzing such samples.
64 65 66 67
Keywords: gas chromatography; water stationary phase; trap column; selectivity; organic bases
ACS Paragon Plus Environment
Page 3 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 3
68
INTRODUCTION
69
Gas chromatography (GC) continues to be a very widely used and reliable tool in
70
analytical separations, partly due to its exceptional sensitivity and performance in the analysis of
71
complex mixtures 1,2. Accordingly, GC is routinely employed for the analytical needs of several
72
sectors including the oil and gas, food and beverage, and pharmaceutical industries, among
73
others
74
technology over the years and they continue to evolve.
3-7.
Owing to its broad applicability, many advances have been made in GC separation
75
One area that holds great importance in this regard is developing and improving GC
76
column selectivity. Indeed, the benefits of ‘tuning’ the selectivity of a GC analysis to suit
77
particular sample needs have long been noted 8, since such efforts can strongly impact the
78
resolution and speed achieved by a method and greatly facilitate the analysis of complex
79
mixtures. While sample preparation and detection can assist in enhancing selectivity, one of the
80
most common routes for this is to alter the separation column. For instance, adjusting column
81
selectivity can be achievable by several means 8, including varying the analysis temperature 9-11,
82
altering the columns employed 12-16, and creating new stationary phases 17-20. Of note on the latter
83
front, novel coatings based on ionic liquids
84
benzothiadiazoles
85
have also been some reports demonstrating that humidified carrier gases can cause a water layer
86
to adhere to the surface of stationary phase particles and provide a means of separation
87
These efforts successfully showed that water could uniquely alter analyte retention in such
88
packed column GC systems, using either pure or higher ionic strength coatings on the particles.
89
However, the main challenges included relatively low column efficiencies and variable analyte
23
21,
metal-organic frameworks
22,
and dithienyl
are being explored for GC use. Previously, in packed column GC, there
ACS Paragon Plus Environment
24-26.
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 4 of 25 4
90
retention due to evaporation of the stationary phase 27. Related explorations in capillary column
91
GC are scarce, if any, perhaps due to the fused silica degradation that occurs in heated water 28,29.
92
Recently, we introduced a novel method of using water as a stationary phase in capillary 30.
93
GC
With this technique, a water coating is established on the inner wall of a narrow bore
94
stainless steel tube and forms a stationary phase that can be successfully utilized for GC
95
separations. The method can directly analyze various compounds in both aqueous and organic
96
samples, and demonstrates good efficiency and reliable performance over a range of conditions.
97
Further, the water stationary phase offers unique selectivity for polar over non-polar solutes
98
These GC attributes were also similar to our earlier work using pure water as a stationary phase
99
in capillary supercritical fluid chromatography 31.
30
100
Ionizable analytes such as organic bases are particularly interesting in this GC system
101
since they often largely exist in a charged state in the neutral pH water stationary phase 32,33. As
102
such, they are heavily partitioned into the phase and very difficult to elute/observe by this
103
method. In contrast to this, when the water stationary phase pH is suitably raised enough to
104
suppress analyte ionization, then organic bases can be readily and directly analyzed by this
105
approach with good sensitivity and peak shape
106
such bases is often hampered by unfavorable column active site interactions, which lead to poor
107
peak shapes and require laborious, error-prone derivatization steps to help overcome this 34,35.
108
32.
This is very useful since the GC analysis of
Certainly the GC analysis of organic bases is of great concern in many important areas 36,37
and pharmaceutical production
38-40.
109
like oil and gas development
110
involved are often complex and can interfere with analyses, many efforts are being made to
111
improve the separation of organic bases, including using novel coatings such as graphitized
112
carbon or porous polymer packings, and invoking various derivatization and column deactivation
ACS Paragon Plus Environment
Since the matrices
Page 5 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 5
113
schemes 41,42. Therefore, the continued exploration of new ways to analyze such bases in GC by
114
more direct, simple, and highly selective means is of great interest.
115
Here, we describe a novel method for analyzing organic bases in GC, which involves two
116
columns placed in series, each coated with a water stationary phase of different pH. The first
117
column is at a low pH to selectively trap organic bases from other injected sample components,
118
while the second column is at a high pH to provide good analytical separation of the bases
119
without prior derivatization. By raising the pH of the first trap column in-situ through injecting a
120
volatile base, analytes are released on demand to the second analytical column for separation and
121
their selectivity over other components is hence controlled externally. The general method
122
operating features, separation characteristics, and application to complex mixtures is presented.
123 124
EXPERIMENTAL
125
Instrumentation and Operation
126
An HP 5890-Series II GC instrument (Agilent, Palo Alto, CA, USA) equipped with a
127
flame ionization detector (FID) was employed. High purity Helium (Praxair, Calgary, Canada)
128
was used as the carrier gas and was saturated with water vapour using an ISCO model 100DX
129
syringe pump (Teledyne ISCO, Lincoln, NE, USA) that supplied water through a valco zero
130
dead volume tee union (Vici-Valco, Houston, TX, USA). The outlet of the tee union was
131
connected to a 1 m stainless steel (SS) pre-heating coil (1/16” O.D. x 250 µm I.D.;
132
Chromatographic Specialties, Brockville, ON, CAN) and both were kept inside the oven. The
133
coil outlet led into the injector. This arrangement helped to prevent evaporation of the water
134
stationary phase downstream during separations and provided stable analyte retention times as a
135
result
136
dependent on carrier gas flow and pressure, temperature had the greatest impact in this regard.
32.
While the amount of hydrating water required to maintain the stationary phase was
ACS Paragon Plus Environment
Analytical Chemistry 6
137
For instance, at 70 oC, 1 µL/min of water was sufficient to maintain the phase, and every 10 oC
138
increase required another 1 µL/min. Above 100 oC, this increase required near 9 µL/min to
139
stabilize the system. As well, due to the minor flow of water vapor through the column, FID
140
response is little impacted and the detector still provides reasonable performance. Of note, the
141
system yields an LOD of 12 pg C/s (S/N=2) and an LOQ of 60 pg C/s (S/N=10). Further, it
142
yields a strongly linear response (R2 value of 0.9997) over the 3 orders of magnitude investigated
143
up to µg quantitites of analyte injected, with no signs of signal saturation. Figure 1 illustrates
144
some response profiles obtained for different injected amounts of analyte in the water stationary
145
phase system.
146 147 148 149 150 151
FID Response (mV)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 25
30000
15000
1500
20000
10000
1000
10000
5000
500
152 0
153
0
2
5
8
2
5
8
0 2
5
8
Time (min) 154 155 156
Figure 1: Demonstration of FID response using the GC water stationary phase system. The panels (L-R) are peaks obtained for 91, 45, and 4.5 ng of hexanoic acid injected on-column.
157
Two columns (a trap and an analytical column) were placed in tandem to one another.
158
The trap column (1/16” O.D. x 250 µm I.D. x 0.5 m; Chromatographic Specialties) was
159
connected to the injector and its outlet was joined to the analytical column (1/16” O.D. x 250 µm
160
I.D. x 11 m; Chromatographic Specialties) via a zero dead volume union (Vici-Valco, Houston,
ACS Paragon Plus Environment
Page 7 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 7
161
TX, USA). Unless stated otherwise, about 24 cm/s of carrier gas was typically used with a 5:1
162
split ratio. The analytical column outlet was joined to a SS capillary restrictor (620 µm O.D. x 75
163
µm I.D. x 0.5 m) via a zero dead volume union (Vici-Valco). This restrictor provided system
164
backpressure and promoted phase stability at higher temperatures. It was positioned inside the
165
FID jet at the burner surface where effluent was deposited directly into the detector flame. An
166
injector/detector temperature of around 220 oC was maintained during the experiments. High
167
purity Hydrogen and Air (Praxair) were used to support the detector flame at respective flows of
168
90 and 350 mL/min. In some experiments, a conventional DB-5 GC column (95% methyl / 5%
169
phenyl polysiloxane stationary phase; 250 µm I.D. x 30 m; 0.25 µm thick) was also used.
170
Stationary Phase Preparation
171
The stationary phase preparation used here has been described previously 32,33. Briefly, a
172
1 M base stock solution of sodium hydroxide prepared in HPLC-grade water was used to adjust
173
the pH of the water supply to be employed as a stationary phase in the separation column.
174
Similarly, a 0.02 M stock solution of sulfamic acid was prepared in HPLC-grade water and used
175
to adjust the pH of the trap column. A calibrated pH meter was employed to measure the pH of
176
the stirred solutions as stock was added. These pH-adjusted solutions were then used
177
immediately to coat the SS capillary columns as described previously
178
columns were mounted inside the GC oven for separation use.
179
Chemicals and Reagents
30.
Once finished, the
180
HPLC-grade water (Honeywell Burdick & Jackson, Muskegon, USA) was used for
181
column coating solutions. Analytes, including ethylamine (70% in water; Arkema Inc, Canada),
182
butylamine, pentylamine, hexylamine, heptylamine and octylamine (all 98%; Sigma-Aldrich,
183
Canada) were used to prepare standard solutions (10 μg/μL) in heptane (>99%; Sigma-Aldrich).
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 8 of 25 8
184
A mixture of hexanol, ethanol, butanol, propanol (at 6 μg/μL), acetone, methanol, cyclopentanol
185
(at 9 μg/μL), propylamine, butylamine, pentylamine, heptylamine (at 10 μg/μL), 1,4-dioxane,
186
and verbenol (at 14 μg/μL) was prepared in heptane (all analytes ≥ 99% except hexanol (≥ 98%)
187
and verbenol (≥ 97%); all Sigma−Aldrich). A solution (10 μg/μL) of propylamine, butylamine,
188
pentylamine, and heptylamine was also prepared in automotive fuel purchased from a local
189
vendor. A standard solution of butylamine, pentylamine (at 10 µg/µL), 1,4-dioxane, and ethanol
190
(at 1 µg/µL) was prepared in heptane. All other details are described in text.
191 192
RESULTS AND DISCUSSION
193
General Operating Characteristics
194
Initial experiments were pursued to establish if basic analytes could be sufficiently ionized
195
and secured on the ‘trap’ column as a means of manipulating retention behavior. Preliminary
196
attempts at this employed pure water as the stationary phase. However, similar to previous
197
observations
198
eluted quite slowly with poor peak shape. Subsequently an acidic water stationary phase was
199
investigated for this purpose. A non-volatile acid was required for this purpose since it needed to
200
remain in the trap and maintain the pH therein. While earlier work found that sulfuric and
201
phosphoric acids could be useful for this, they were also found to erode the SS tubing33.
202
Subsequently, sulfamic acid was identified as an ideal choice without detrimental side effects in
203
the system33. Incorporating the latter information, it was found that when a pH 2.2 sulfamic acid
204
phase was instead used here, basic analytes were well retained on the trap column as a result of
205
their increased ionization and were not observed to elute at all. Thus, acidic phases were used for
206
trapping.
32,
basic analytes were often not held static enough at neutral pH and frequently
ACS Paragon Plus Environment
Page 9 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 9
207
Next it was examined if the trapped basic analytes could then be released in-situ by
208
introducing a base into the system. Earlier work found that sodium hydroxide was very non-
209
volatile and remained in the column over time, whereas ammonia solution readily volatilized and
210
was ejected from the system in a short period32. Therefore injection of aqueous ammonia was
211
employed here for releasing basic analytes since it was volatile enough to readily mobilize
212
through the system and also yielded a low FID response relative to organic compounds. The
213
primary aim of this approach was to dynamically neutralize the ionized analytes on the acidic
214
trap column and enable their subsequent elution. However, since the acidic phase itself also
215
consumed the injected ammonia, the process had to be optimized with respect to trap column
216
length, ammonia concentration, and the injected volume. Further, it should also be noted that
217
while attempts were made to trap, release and separate analytes using only the single (i.e. trap)
218
column, it was found that analytes did not separate well in this format. Thus, a tandem dual
219
column system was adopted where analytes were first trapped on a short acidic phase coated
220
column, released by subsequent injection of aqueous ammonia, and then separated on a final
221
longer basic phase coated analytical column.
222
The main parameters that were explored for optimization in this regard were ammonia
223
concentrations of 1 to 10 M, injected base volumes of 1 to 9 µL, and trap column lengths of 0.5
224
to 1 m. In general, the main challenge with this setup was found to be that if the trap column
225
was too long, or if the amount of base used for release was too low, then all of the analyte could
226
not be mobilized in one step and separations were incomplete. For example, using a 1 M
227
ammonia solution frequently failed to quantitatively recover the analyte from the trap column in
228
one step and so 10 M concentrations were explored. With a 2 m trap column, 1, 3, and 9 µL
229
injections of 10 M ammonia solution were found to recover about 75, 80, and 83% of the analyte
ACS Paragon Plus Environment
Analytical Chemistry 10
230
respectively in one step. However, in the latter case, the volume was so large that it generated an
231
FID signal that could interfere in analyses. As such, 3 µL of 10 M ammonia solution was further
232
explored. It was found that when using this to release analytes, recoveries increased about 20%
233
when moving from a 2 to a 0.5 m trap column. Therefore, ultimately, a 0.5 m long trap column
234
and 3 µL of 10 M ammonia was found to be optimum in most cases here for releasing trapped
235
solutes. Using this approach, little to no residual organic base remained in the system and 99% of
236
the injected analyte was normally recovered in this single step. Alternatively, for certain complex
237
samples, a 1 m trap was optimal and employed where noted.
238
Figure 2 demonstrates the dual column system under optimum conditions with a
239
chromatogram for the injection of pentylamine and butylamine. Without using a trapping
240
column, these analytes would normally appear between about 3 and 5 minutes. However, as can
241
be seen with the trapping system used in figure 2, after the heptane solvent appears upon
242
injection of the sample, the analytes do not elute in this time frame since they are trapped on the
243 244 245 246 247 248 249 250 251 252 253 254 255 256 257
FID Response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 10 of 25
0
5
Time (min)
10
15
Figure 2: Separation of pentylamine and butylamine (in order of elution) in heptane on dual column system (0.5 m trap with pH 2.2 sulfamic acid coating and 11 m analytical column with pH 11.4 NaOH coating). The analytes were released (indicated by the arrow) at 5 minutes with 3 µL of 10 M aqueous NH3. Column temperature is 100 oC and the carrier gas velocity is 24 cm/s.
ACS Paragon Plus Environment
Page 11 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 11
258
first acidic phase column. Subsequently though, when they are released after injection of NH3
259
solution at the 5 minute mark, they appear well resolved and with good peak shape near 10
260
minutes. Therefore, the system can be dynamically used to delay and control analyte elution.
261
It should be noted that consecutive injections could be performed reproducibly if the acidic
262
trap column phase was replenished in between runs. This was accomplished here with the
263
injection of 3µL of 1 M formic acid after each trial. Note that formic acid was chosen for this
264
since it was fairly volatile and could be readily mobilized through the system to re-acidify the
265
trap. In contrast to this, sulfamic acid was primarily used to coat the trap column since it was
266
non-volatile and would stably remain in place to hold the trap at an appropriate pH level during
267
trials33. It was found that this adequately acidified the trap column for reuse and the system
268
provided comparable performance in each run as a result. For example, in terms of
269
reproducibility, consecutive trials of the separation shown in figure 2 yielded analyte retention
270
times and peak areas with average RSD values of 1.2% and 0.8% respectively. As such, the
271
optimum operating conditions above were used going forward in this study to further
272
characterize the method.
273 274
Effect of Time on Analyte Elution In order to get a better idea of the capacity of the dual column system to hold analytes, a
275
study was conducted to examine the effect of time between trap and release of the analyte. The
276
typical results obtained are demonstrated below in figure 3. The chromatograms shown are from
277
a series of experiments where the injection of an amine pair was performed followed by their
278
release at consecutively later points in time for each trial. As seen in figure 3, the analytes can be
279
readily trapped for significant periods of time within the considerable range investigated. For
280
instance, no significant peak distortion or degradation of the separation is observed for trapping
281
periods investigated up to 1 hour. However, it should be noted that the peak resolution appeared
ACS Paragon Plus Environment
Analytical Chemistry 12
283 284
FID Response
282 10 min
0
10
20
285
30 40 Time (min)
50
60
70
60
70
60
70
287
FID Response
286 20 min
288 0
10
20
30 40 Time (min)
50
289
291
FID Response
290
292
30 min
0
10
20
30 40 Time (min)
50
294 295
FID Response
293
40 min
0
296
10
20
30 40 Time (min)
50
60
70
297 298 299 300 301 302 303 304 305 306 307
FID Response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 12 of 25
60 min
0
10
20
30 40 Time (min)
50
60
70
Figure 3: Chromatograms of pentylamine and butylamine (in order of elution) on dual column system with analytes released as a function of time. Analytes were released (indicated by arrow) at 10, 20, 30, 40, and 60 minutes. Other conditions are as in figure 2.
ACS Paragon Plus Environment
Page 13 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 13
308
to decrease slightly after this time. None the less, this demonstrates that elution of such analytes
309
can be delayed on-column for significant periods of time. As such, this can potentially allow for
310
adjustable selectivity of these organic bases to be achieved, which can in turn facilitate resolving
311
them from other sample components as desired.
312
Altering Selectivity and Resolution
313
In order to investigate this further, various separations were carried out with organic bases
314
among other analytes. The first of these was the separation of a sample containing amines,
315
ethanol, and 1,4–dioxane using a trapping and non-trapping water stationary phase system. The
316
results are shown in figure 4 and, as can be observed, the analytes readily elute on the single
317
column system with pentylamine and butylamine appearing at 2.1 and 2.7 minutes respectively,
318
while ethanol and 1,4-dioxane are retained for 6.3 and 8 minutes respectively (figure 4A).
319
However, when using the dual column system (figure 4B), the retention of the amines can be
320
increased such that they elute at much different times. Of note, the analytes now elute at 13.1 and
321
13.6 minutes respectively. Conversely, the retention of ethanol and 1,4-dioxane is largely
322
unaffected in the dual column system apart from a minor increase in elution time owing to the
323
extra 0.5 m trap column length employed. As a result, the formal selectivity realized changes
324
dramatically. Specifically, the selectivity obtained in figure 4A for 1,4-dioxane/pentylamine is
325
about 5.4, whereas in the dual column system of figure 4B it is 0.7 (or formally 1.4 for the new
326
reversed pentylamine/1,4-dioxane elution order of the pair as shown in the figure). Therefore,
327
using this approach it is possible to reverse the selectivity observed for certain pairs of analytes.
328 329 330
ACS Paragon Plus Environment
Analytical Chemistry 14
331 332
334
2
A
FID Response
333
1
3
0
335
α4,1 = 5.4
4
5
10
15 Time (min)
20
25
30
336 337 338 339 340
1 2 FID Response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 14 of 25
α1,4 = 1.4
3 4 0
5
B
10
15 Time (min)
20
25
30
341 342 343 344 345 346 347
Figure 4: Chromatograms of 1) pentylamine, 2) butylamine, 3) ethanol and 4) 1,4-dioxane on A) single analytical column (11 m, pH 11.4 NaOH coating) and B) dual column system (0.5 m trap with pH 2.2 sulfamic acid coating and 11 m analytical column with pH 11.4 NaOH coating). Analytes in B are released at the 12 minute mark. Other conditions are as in figure 2.
348
Inherent to this, the resolution between analytes can also be significantly altered. Figure 5
349
shows an example of this with the comparative separations of acetone, butylamine and
350
pentylamine on both a single analytical column and the dual column system. As shown in figure
351
5A, pentylamine appears first in the chromatogram on the single analytical column, while
352
acetone and butylamine co-elute and almost entirely overlap due to similar retention times in this
353
system. As a result, the resolution obtained for this analyte pair is only about 0.3. However,
354
when the same mixture is analyzed on the dual column trap system (figure 5B), acetone again
355
appears near the 5 minute mark as before while the two bases are held on the acidic trap
ACS Paragon Plus Environment
Page 15 of 25
15
356
358
FID Response
357
1
359
2, 3
0
5
360 361 362 363
A R2,3 = 0.3
10
15 Time (min)
20
1
3
15 Time (min)
20
25
30
B
2
FID Response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
R2,3 = 13.4
0
5
10
25
30
364 365 366 367 368 369 370 371 372
Figure 5: Chromatograms of 1) pentylamine, 2) acetone and 3) butylamine on A) single analytical column (11 m, pH 11.4 NaOH coating) and B) dual column system (0.5 m trap with pH 2.2 sulfamic acid coating and 11 m analytical column with pH 11.4 NaOH coating). Analytes in B are released at the 17 minute mark. Column temperature is 80 oC. Other conditions are as in figure 2.
373
column until they are released for analysis on the basic analytical column after 17 minutes.
374
Consequently, a near 40 fold increase in resolution to a value of 13.4 is achieved for the
375
overlapping analyte pair. Therefore the ability to significantly control basic analyte retention in
376
this system could potentially facilitate analyzing such analytes within more complex mixtures.
377
Application to Complex Mixtures
378
In order to further examine the potential analytical utility of the dual column system,
379
more complex mixtures were next studied. The first was a 13 component mixture of oxygenated
380
and basic analytes that was again analyzed on the single analytical column and then compared to
381
that of the dual column trap system. The typical results are demonstrated in figure 6. Similar to
ACS Paragon Plus Environment
Analytical Chemistry 16
382
384
FID Response
383
2 1
4 6 35
A 7
8
9 10
11
385
0
13 12
i
5
10 Time (min)
15
20
386 387 388 389
B
FID Response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 16 of 25
3 5
7 i
390
0
8 9
i i
5
10
11 12
2
4 6
1
13
10 Time (min)
15
20
391 392 393 394 395 396 397 398 399 400 401 402
Figure 6: Chromatograms of an analyte mixture on A) single analytical column (11 m, pH 11.4 NaOH coating) and B) dual column system (1 m trap with pH 2.2 sulfamic acid coating and 11 m analytical column with pH 11.4 NaOH coating). Analytes in B are released at the 17 minute mark. Analytes in order of elution are 1) heptylamine, 2) pentylamine, 3) hexanol, 4) butylamine, 5) acetone, 6) propylamine, 7) cyclopentanol, 8) butanol, 9) propanol, 10) ethanol, 11) 1,4dioxane, 12) verbenol, and 13) methanol. Unknown impurities are denoted by ‘i’. Other conditions as in figure 2.
403
the results shown in figure 5A, on the single analytical column the basic analytes elute amidst
404
various oxygenated compounds with significant interference and poor resolution in some cases
405
(figure 6A). However, when employing the dual column system (figure 6B), the oxygenates elute
406
similarly while the basic analytes are effectively held on the trap column until they are later
407
released at a more opportune time in the chromatogram. When this is then done at the 17 minute
408
mark, the analytes readily appear well resolved and with sharp prominent peak shape.
ACS Paragon Plus Environment
Page 17 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 17
409
Additionally, a few unknown impurities were also revealed as a result of modifying the elution
410
time of the interfering amines to an alternate retention window. Therefore the system can
411
potentially facilitate sample analyses in such circumstances by providing a means of changing
412
resolution as desired. It should be noted that the shift in retention time that can be seen for the
413
oxygenates in figure 6B is ascribed to the extra retention of these analytes on the 1 m trapping
414
column in place for those trials. This is because the trapping column is also coated with a water
415
stationary phase and these analytes continue to partition within it while traversing the trap
416
column. Thus, the analyte velocities established in figure 6A can be extended to an extra 1 m
417
length of travel in figure 6B. In doing so, the predicted retention times for figure 6B differ from
418
the actual retention times by an average of only 3%. Still, further experimentation would be
419
required to establish if other interactions could also make minor contributions to this as well. As
420
such, the slight delay in retention appears to be primarily a function of the extra column length.
421
To examine if analyte resolution can be controlled further, the system was applied to the
422
analysis of a commercial gasoline sample containing amines. Analyses of this nature are
423
routinely performed using GC and are important since such compounds can inhibit the quality of
424
petroleum products 43,44. Figure 7A demonstrates the analysis of this sample using a conventional
425
GC column. As seen, the sample is quite complex and numerous hydrocarbons are present as
426
evidenced from the large number of peaks eluting in the first 20 minutes. Further, the target
427
amines present in the sample elute in the first 10 minutes, greatly overlapping with these
428
hydrocarbons, and are obscured from being clearly observed as a result. Figure 7B then shows
429
the analysis of the same sample on the dual column trap system. As observed, a much clearer
430
chromatogram is produced. Of note, all of the hydrocarbon matrix components are now
431
unretained in the water stationary phase and rapidly elute in the void volume as a result.
ACS Paragon Plus Environment
Analytical Chemistry 18
433 434
FID Response
432
A
1-4
435
0
10
20
30 Time (min)
40
50
60
50
60
436 437
439
FID Response
438
3
440
4
B
2 1
0
10
20
30 Time (min)
40
441 442 443 444 445
2 3 4
FID Response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 18 of 25
C
0
10
20
30 Time (min)
40
1
50
60
446 447 448 449 450 451 452 453 454 455 456 457
Figure 7: Separations of a gasoline sample containing 1) heptylamine, 2) pentylamine, 3) butylamine, and 4) propylamine on A) conventional DB-5 column (80 oC, 50 cm/s carrier gas), B) dual column system (1 m trap with pH 2.2 sulfamic acid coating and 11 m analytical column with pH 11.4 NaOH coating; 100 oC, 18 cm/s carrier gas, analytes released at 30 minutes) and C) tandem system with DB-5 column from A joined with dual column from B (80 oC for 30 minutes then increased to 100 oC, 13 cm/s carrier gas, analytes released at 45 minutes). Other conditions as in figure 2.
ACS Paragon Plus Environment
Page 19 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 19
458
Conversely, the amines are retained until they are released from the trap column at 30 minutes
459
and then appear with sharp prominent peaks for each thereafter. Therefore, these features of the
460
dual column system can greatly increase analyte selectivity and simplify the analysis of such
461
samples. In particular, it can remove matrix interference and allow direct analysis of such bases
462
without the need for lengthy derivitization steps that are often required in order to attain
463
appropriate peak shapes for basic analytes 35.
464
Granted, while the results in figure 7B are advantageous for analyzing bases alone in
465
such samples, it can be seen that information regarding the hydrocarbon component of the
466
sample is simultaneously diminished as a result of their complete lack of retention in the
467
system.Indeed, in certain applications it may be desirable to maintain the ability to monitor the
468
hydrocarbon and organic base constituents in the same run. Therefore, in a final set of
469
experiments, it was examined if placing a conventional GC column upstream before the acidic
470
trap column could help facilitate this. Figure 7C shows the analysis of the same sample with a
471
conventional non-polar GC column ahead of the dual column trap system. As seen, when using
472
this now triple column system, the hydrocarbon portion of the sample is reasonably retained and
473
separated on the first column and this is preserved when passing through the remaining two
474
water-based columns since the non-polar analytes have no retention therein. Conversely, the
475
amines are held static upon leaving the first conventional column and entering the acidic trap
476
column. There, they are retained until the interfering matrix components are finished eluting and
477
then released at the 45 minute mark where they appear without being obscured by the matrix.
478
Therefore, this approach can afford the necessary resolution between the bases and the gasoline
479
matrix, and also support the separation of the matrix itself. Accordingly, it could be beneficial for
480
such analyses. It should also be mentioned here that for simplicity in these trials, the same input
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 20 of 25 20
481
carrier gas pressure was applied in each case of figure 7 and the resulting carrier gas velocity was
482
subsequently used. As such, in figure 7A, the conventional 30 m column yielded a higher
483
velocity than the 12 m trap system in figure 7B, which also employed a linear restrictor at the
484
outlet. Accordingly, when combining these in figure 7C, the flow was further reduced. With
485
regard to the DB-5 column used, it is also useful to point out that in using it for such analyses
486
over the course of a year with multiple injections of concentrated ammonia and formic acid, no
487
deterioration in its performance was noted and separations before and after such exposure
488
appeared unchanged.
489
It is worth noting that while many of the separations in this work were carried out at 80 to
490
100 oC, this was done to maximize speed while maintaining analyte resolution of the test bases
491
utilized. None the less, with adequate hydration the water stationary phase has been shown to be
492
stable to temperatures investigated up to 160 oC 32. Further, since analyte boiling point does not
493
correlate strongly with retention on this phase, many larger non-volatile solutes can still be
494
successfully eluted within this operating window 32. As well, while not required here, it should
495
also be pointed out that temperature programming with this phase is also readily achievable and
496
does not display any adverse effects on separations 30, 31. Finally, the system is interesting in this
497
regard for triple column applications since target base analytes are trapped initially while the
498
conventional GC column separation takes place. In this way, temperatures can be readily
499
manipulated for the primary column, while analytes remain held in the trap. Then, after the initial
500
separation occurs, the trap and analytical water phase columns can be further manipulated at
501
different temperatures since the release solvent serves to ‘re-inject’ the analyte portion of the
502
sample. Thus, the system can also afford certain flexibility in optimizing temperatures.
503
ACS Paragon Plus Environment
Page 21 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 21
504
CONCLUSION
505
A tandem GC column system employing pH-adjusted water stationary phases has been
506
studied and characterized for use in the analytical separation of organic bases. This technique
507
involves the use of a short trap column that can be coated with an acidic stationary phase
508
followed by a long analytical column coated with a basic phase, and employs the differential
509
partitioning behaviour of analytes between the two. The study showed the ability to dynamically
510
and externally control analyte selectivity in GC. As a result, organic bases were reproducibly
511
analyzed with greatly improved selectivity and resolution from other components. Finally, the
512
dual column trap system described here enabled the analysis of organic bases in a gasoline
513
matrix with high selectivity, and the additional separation of the hydrocarbon components when
514
coupled with a conventional GC column (i.e. triple column system). Therefore the versatility of
515
the technique is interesting and potentially beneficial for such analyses. A similar exploration for
516
the analysis of organic acids is being investigated.
517 518 519 520
ACKNOWLEDGEMENTS The authors are grateful to the Natural Sciences and Engineering Research Council of Canada for a Discovery Grant in support of this research.
521 522 523
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 22 of 25 22
524 525 526 527 528 529 530 531 532 533 534 535 536 537 538 539 540 541 542 543 544 545 546 547 548
REFERENCES
549
[11] Benická, E.; Krupčík, J.; Répka, D.; Sandra, P. J. Chromatogr. 1992, 33, 463-466.
550 551 552 553 554 555 556 557 558 559 560 561 562 563 564 565 566 567
[12] Baycan-Keller, R.; Oehme, M. J. J. Chromatogr. A 1999, 837, 201-210.
[1] Cazes, J. Ewing’s Analytical Instrumentation Handbook; Marcel Dekker: New York, 2005. [2] Veen, R.F. Principles and Practice of Bioanalysis, 2nd ed.; CRC Press - Taylor & Francis Group: Florida, 2008. [3] James, A. T.; Martin, A. J. P. J. Biochem. 1952, 50, 679-690. [4] Liska, I.; Slobodnik, J. J. Chromatogr. A 1996, 733, 235-258. [5] Kidwell, D. A.; Kidwell, J. D.; Shinohara, F.; Harper, C.; Roarty, K.; Bernadt, K.; McCaulley, R. A.; Smith, F. P. Forensic Sci. Int. 2003, 133, 63-78. [6] Conway, J. M.; Birnbaum, A. K.; Marino, S. E.; Cloyd, J. C.; Remmel, R. P. Biomed. Chromatogr. 2012, 26, 1071-1076. [7] Scott, S. A.; Douglas, G. S.; Uhler, A. D.; McCarthy, K. J.; Emsbo-Mattingly, S. D. J. Environ. Claim. 2005, 17, 71-88. [8] Freeman, R. R.; Kukla, D. J. Chromatogr. Sci. 1986, 24, 392-395. [9] Ettre, L. S. Introduction to Open Tubular Columns; Perkin-Elmer Corp: Norwalk, 1978. [10] Marriott, P. J.; Kinghorn, R. M. Anal. Sci. 1998, 14, 651-659.
[13] Smith, H.; Sacks, R. Anal. Chem. 1997, 69, 5159-5164. [14] Habram, M.; Slemr, J.; Welsch, T. J. High Resolut. Chromatogr. 1998, 21, 209-214. [15] Sharif, K. M.; Kulsing, C.; Marriott, P. J. Anal. Chem. 2016, 88, 9087-9094. [16] Krupcík, J.; Spánik, I.; Benická, E.; Zabka, M.; Welsch,T.; Armstrong, D. W. J. Chromatogr. Sci. 2002, 40, 483–488. [17] Matzke, C. M.; Kottensetette, R. J.; Casalnuovo, S. A.; Frye-Mason, G. C.; Hudson, M. L.; Sasaki, D. Y.; Manginell, R. P.; Wong, C. C. Proc. SPIE-Int. Soc. Opt. Eng. 1998, 3511, 261268. [18] Greally, B. R.; Nickless, G.; Simmonds, P. G.; Woodward, M.; de Zeeuw, J. J. Chromatogr. A 1998, 810, 119-130.
ACS Paragon Plus Environment
Page 23 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry 23
568 569 570 571 572 573 574 575 576 577 578 579 580 581 582 583 584 585 586 587 588 589 590 591 592 593 594 595 596 597 598 599 600 601 602 603 604 605 606 607 608 609 610 611 612 613
[19] Robson, M. M.; Bartle, K. D.; Myers, P. J. Microcol. Sep. 1998, 10, 115-123. [20] Dorman, F. L.; Schettler, P. D.; English, C. M.; Patwardhan, D. V. Anal. Chem. 2002, 74, 2133-2138. [21] Pacholec, F.; Butler, H. T.; Poole, C. F. Anal. Chem. 1982, 54, 1938-1941 [22] Chen, B.; Liang, C.; Yang, J.; Contreras, D. S.; Clancy, Y. L.; Lobkovsky, E. B.; Yaghi, O. M.; Dai, S. Angew. Chem. Int. Ed. 2006, 45, 1390–1393. [23] Sun, T.; Tian, L.; Li, J. M.; Qi, M. L.; Fu, R. N.; Huang, X. B. J. Chromatogr. A 2013, 1321, 109-118. [24] Phifer, L. H.; Plummer, H. K. Anal. Chem. 1966, 38, 1652-1656. [25] Karger, B. L.; Hartkopf, A.; Pasmanter, H. J. J. Chromatogr. Sci. 1969, 7, 315-317. [26] Karger, B. L.; Hartkopf, A. Anal. Chem. 1968, 40, 215-217. [27] Berezkin, V. G. Russ. Chem. Bull. 1999, 48, 1807-1816. [28] Karásek, P.; Planeta, J.; Roth, M. Anal. Chem. 2013, 85, 327-333. [29] Pollock, R. A.; Gor, G. Y.; Walsh, B. R.; Fry, J.; Ghampson, I. T.; Melnichenko, Y. B.; Kaiser, H.; DeSisto, W. J.; Wheeler, M. C.; Frederick, B. G. J. Phys. Chem. C 2012, 116, 2280222814. [30] Gallant, J. A.; Thurbide, K. B. J. Chromatogr. A 2014, 1359, 247-254. [31] Fogwill, M.; Thurbide K. B. Anal. Chem. 2010, 82, 10060-10067. [32] Darko, E.; Thurbide, K. B. J. Chromatogr. A 2016, 1465, 184-189. [33] Darko, E.; Thurbide, K. B. Chromatographia. 2017, 80, 1225–1232. [34] Kataoka, H. J. Chromatogr. A 1996, 733, 19-34. [35] Ferreira, A. M. C.; Laespada, M. E. F.; Pavon, J. L. P.; Cordero, B. M. J. Chromatogr. A 2013, 1296, 70-83. [36] Li, N.; Chen, C.; Wang, B.; Li, S. J.; Yang, C. H.; Chen, X. B. Appl. Petrochem. Res. 2015, 5, 285-295. [37] Grimmer, G.; Naujack, K. W. Fresenius Z. Anal. Chem. 1985, 321, 27-31. [38] Ternes, T. A. TrAC, Trends Anal. Chem. 2001, 20, 419-434.
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 24 of 25 24
614 615 616 617 618 619 620 621 622 623 624 625 626 627 628 629 630 631
[39] Dugal, R.; Masse, R.; Sanchez, G.; Bertrand, M. J. J. Anal. Toxicol. 1980, 4, 1-12. [40] Basheer, C.; Lee, J.; Pedersen-Bjergaard, S.; Rasmussen, K. E.; Lee, H. K. J. Chromatogr. A 2010, 1217, 6661- 6667. [41] Breitbach, Z. S.; Weatherly, C. A.; Woods, R. M.; Xu, C.; Vale, G.; Berthod, A.; Armstrong, D. W. J. Sep. Sci. 2014, 37, 558-565. [42] Kataoka, H.; Yamamoto, S.; Narimatsu, S. Part III Amines: Gas Chromatography, Encyclopedia of Separation Science; Academic Press: Massachusetts, 2000. [43] Altgelt, K. H. Composition and Analysis of Heavy Petroleum Fractions; 1st ed.; CRC Press - Taylor & Francis Group: New York, 1993. [44] Thornson, J. S.; Green, J. B.; McWilliams, T. B.; Yu, S. K. - T. J. High Resolut. Chromatogr. 1994, 17, 415-426.
ACS Paragon Plus Environment
Page 25 of 25 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
254x190mm (96 x 96 DPI)
ACS Paragon Plus Environment