E k c trochec;iical Detdction I

I

IN M I C R O C O L SEPARAT ~ t is difficult to find an area of analytical chemistry that has not been affected by the general movement toward the miniaturization of existing techniques; separation science has by no means been excluded from this trend. In particular, major advances toward miniaturization have been made in HPLC and in capillary zone electrophoresis (CZE) ; thus, it is now possible to analyze samples with volumes in the nanoliter to femtoliter range. With the development of low-volume separations comes the need to reassess

Andrew G. Ewing Jody M. Mesaros Peter F. Gavin The Pennsylvania State University 0003 - 2700/94/0366 -527A/$04.50/0

0 1994 American Chemical Society

Electrochemical detection is ideally suited for the miniaturization of HPLC and CZE detection strategies to accommodate minute amounts of analyte. Conventional detection methods designed for macroseparations cannot effectively be applied to extremely low volumes; therefore, the development of detection schemes for microseparations is challenging. The focus of this Report is on electrochemical detection schemes that have been used

successfully for microseparations resulting from the miniaturization of HPLC and CZE. In addition, an example of the utility of microseparations using electrochemical detection-single-cell analysis-is included. Microseparation techniques

HPLC is a ubiquitous analytical tool in which the separation of analytes is based on their differentialretention by a stationary phase. Typically, a stainless steel tube several millimeters in diameter is packed with particles coated with the stationary phase, and the sample to be separated is dissolved in a suitable solvent and passed through the tube by a pressure-driven flow. HPLC is useful for many analytical samples, but because of the inherent low efficiency resulting from the use of large-

Analytical Chemistry, Vol. 66, No. 9, May 1, 1994 527 A

bore tubes, complex samples are often difficult to resolve. During the late 1970s and early 198Os, HPLC became the target of the miniaturization trend in an attempt to gain higher column efficienciesand increased resolving power. Miniaturization would also decrease expenses associated with both the packing of large columns and the need for large volumes of solvent.The ability to sample from microenvironments with limited volumes was also an incentive for miniaturization. Thus the groundwork was laid for the development of micro LC and the use of small-bore glass or fusedsilica capillary columns instead of stainless steel tubes. The techniques that emerged from the use of capillaries are open-tubular liquid chromatography (OTLC) and microcolumn LC (1). OTLC has the higher efficiency potential of the two techniques because, as its name implies, no packing material is used and therefore no eddy diffusion, which would add to solute band dispersion, exists. An OTLC column is prepared by coating the inner wall of a 3-5Gpm i.d. capillarycolumn with the stationary phase. Alternatively, microcolumn LC involves a capillary column packed with particles that are chemically modified to yield the stationary phase. There are two types of packed capillaries:partially packed (typically 40-80-pm id.) and tightly packed (typically 40-200-pm i.d.). Partially packed columns can be made by packing a large-bore capillarywith particles and pulling the capillary on a glass drawing machine to decrease the internal diameter. The particles are then chemically modified to create the stationary phase. More often, however, a capillaryis tightly packed by using slurry packing methods. In this case, a small-bore capillaryis pressure packed with chemically modified particles that are suspended in an organic solvent. Separations in open-tubular and packed microcolumns occur by the same retention process as in HPLC. A complementary separation mechanism is used in CZE (2,3). In general, CZE separation is carried out in a solvent-filled glass or fused-silicacapillaryacross which a potential-as high as 40 kV is applied. The applied potential drives the separation

via two distinct phenomena. The first phenomenon, termed electroosmosis, produces a very strong flow toward the cathodic electrode caused by the interactions of positively charged solvent components with the capillaryinner wall (4).The second phenomenon, electrophoresis, is the movement of charged species in an electric field. When charged analytes are introduced into the capillary, they undergo electrophoresis based roughly on their charge-to-size ratios and separate into discrete bands.

Sample introduction is carried out at the anodic end of the capillaryin one of three ways: by applying the separation voltage for a few seconds to electromigrate a small plug of material, by raising the column entrance to induce gravity flow of the sample into the capillary, or by forcing the sample into the capillaryvia pressure. Generally, because of the electroosmotic flow, analytes migrate toward the cathodic end of the capillarywhether they are positively or negatively charged. Electroosmotic flow has a flat flow profile, unlike the laminar flow profile found in pressure-driven separation techniques, and this profile accounts in part for the very high separation efficienciesfound in CZE. These high efficienciesare often reduced, however, by convection currents caused by resistive heating in the capillary. Hjerten demonstrated increased s e p arating power of zone electrophoresis in 1967by rotating large-bore open capillaries (5).The problem of joule heating has been addressed more recently through the miniaturization of electrophoresis techniques. The use of 200-500-pm i.d. glass capillaries (6) and 200-pm i.d. Teflon tubes (7)

528 A Analytical Chemistry, Vol. 66, No. 9, May 1, 1994

increased the surface area-to-volume ratio of the electrophoresis column, leading to improved heat dissipation across the liquid - solid interface of the capillary. In the first use of capillarieswith internal diameters < 100 pm, efficienciesof > 400,000 theoretical plates were realized (8).Since that time, capillaries as small as 2-pm i.d. have been used (9),and CZE has become widely accepted as a high-efficiency microseparation technique. The development of microuolumn sep aration techniques allows quantitative analysis of volumelimited microenvironments such as the contents of single cells (3, 10, 11).This analyticalchallenge relies on the ability not only to sample extremely low volumes but also to detect analytes at low concentrations. The development of sensitive detection schemes is therefore critical to the successful application of microseparations. Early detection schemes for microsep arations were scaled-down versions of macroscale detectors. This approach was not very profitable, however, because macroscale detector designs were not easily adapted to low sample volumes. It was found that detectors were best made as a part of the microseparation column, UV-vis absorbance detection has been used extensively in the development of microseparation schemes, and it was the first type of detector to be placed on a column. This was accomplished by removing the protective polyimide coating from the silica microcolumn to provide a detection window near its end. Because of the small-diameter capillariesused, however, the pathlength of these detection cells is very short and yields concentration detection limits of only to lov6M. Also, problems involved in focusing and collecting light onto and from the capillaryhave been limiting factors in the use of UV-vis absorbance for microseparations. Despite these drawbacks, this detection technique is still commonly used, because it is applicable to a broad range of analytes. Advances toward more sensitive detection methods have been reported (12-1 7). Laser-induced fluorescence (LIF) is another detection scheme that has been applied to microseparations using the oncolumn format. In most cases, light from a laser is focused onto an on-column window and accounts for the increased sensi-

tivity of the technique. Although lasers offer a limited range of excitation wavelengths and samples need to be derivatized before analysis to make them fluorescently active, LIF detection is still attractive because low detection limits can be obtained. Attomole detection limits are routine, and zeptomole detection limits have been achieved by using a sheath flow chamber with optimized collection optics (18) and sensitive detectors with improved data collection methods (19). LIF detection has also been successfully applied to the analysis of single human erythrocytes using native fluorescence, derivatization,and indirect detection techniques (20,21). MS is another detection scheme common in macroscale separations that has been applied to microcolumn separations. This technique is attractive because of its general applicability and its ability to yield structural information. Coupling microcolumns to MS is not as straightforward as coupling to optical detection methods because the on-column scheme cannot be used. Therefore, the technology for the routine use of MS in microseparations is still being developed. Much progress has been made, as outlined in a recent review article on CE/MS, which reports mass detection limits in the attomole range (22). Detection limits in the attomole range also have been reported for microcolumn LC with MS detection (23,24).

ical detection are amperometry, conductimetry, and potentiometry. In the amperometric mode, compounds undergo oxidation or reduction reactions through the loss or gain, respectively, of electrons at the electrode surface. The electrical current arising from the electrons passed to or from the electrode is recorded and is proportional to the concentration of analyte present. Conductimetric and potentiometric methods measure the conductance or potential changes in the solution between two electrodes caused by the introduction or removal of charged species. Amperometry is the most easily implemented of the three modes of detection and is therefore the most commonly used. Coupling electrochemical detection cells to microcolumns was first attempted using the off-column format. However, the best detection schemes have involved on-

column detectors. The first on-column electrochemical detector for microcolumn LC was described by Manz and Simon in 1983 (25). An ion-selectiveelectrode 1pm in diameter was inserted into the end of a 25ym i.d. OTLC column to potentiometrically detect K+with detection limits in the picomole range. This same principle of on-column detection was used in 1984 by Knecht, Guthrie, and Jorgenson, who placed a 9-pm diameter carbon fiber electrode into the end of a 15-wi.d. OTLC column (26). Using amperometric detection, they achieved femtomole detection limits for ascorbic acid, catechol, and 4-methylcatechol. The development of electrochemical detection for CZE was not as straightforward as that for microcolumn LC because of the high potential fields used to effect the separation. In most cases, when an

I

Detection

capillary

Carbon fiber nlnctr0de

Electrochemical detection

Electrochemical detection is successful in macroscale separations because of its ease of implementation; ability to selectively detect and, in some cases, identify analytes; and high sensitivity.For these reasons it should also prove to be useful in microcolumn separations. In addition, electrochemical detection is a concentration-sensitivetechnique and therefore well suited for microseparations. As the volume of the microcolumn decreases, the mass detection limits should improve. The major drawback of electrochemical detection is the inherent selectivityof the technique, which generally limits analysis to easily oxidized or reduced species. However, this disadvantage can in some cases be circumvented, as will be discussed later. The three basic modes of electrochem-

Figure 1. Electrochemical detection schemes for CZE.

Analytical Chemistry, Vol. 66,No. 9, May I, 1994 529 A

electrode was placed in the capillary, noise arising from the high voltage a p plied across the capillary interfered with detection. Despite this problem, two basic schemes have been developed that are routinely used for electrochemical detection with CZE. The first design was implemented in 1987 (27).A porous glass coupler, shown in Figure la, was used to decouple the electrode from the separation potential. The design involves a fracture near the

end of the capillary that is covered with porous glass to facilitateion movement but not bulk flow. The separation voltage is therefore dropped across the capillary only up to the point of the fracture. A p proximately 1cm of capillary extends past the coupler and serves as the detection capillary. An electrode is manipulated into the detection capillary, which allows amperometry of analytes that are separated in the capillaryand then electroosmotically pumped into the detection capillary.

kigure 2. Representative electropherograms of compounds separated using (a) a porous glass coupler design and (b) optimized end column detection. (a) Compounds detected: A, serotonin; B, norepinephrine; C, epinephrine; D, 1-dihydroxyphenylalanine;E, 5 - hydroxyindoleaceticacid; F, homovanillic acid; G, dihydroxyphenylaceticacid; H, ascorbic acid. (Apparent volumes injected were 36 pL.) (b) Compounds detected (apparent volumes injected in parentheses): A, dopamine (47 pL); B, isoproterenol (40 pL); C, catechol (27 pL). (Adapted from References 9 and 31.) 530 A Analytical Chemistry, Vol. 66, No. 9, May 1, 1994

An electropherogram of several compounds, including catechols and ascorbic acid detected using the coupler design, is shown in Figure 2a. Detection limits as low as 0.5 amol have been reported for the separation of a standard solution of neurotransmitters by using a 5pm i.d. capillary (9). The implementation of the coupler design permitted the first use of oncolumn electrochemical detection for CZE. However, the porous glcss coupler is fragile and difticult to maintain. In 1992 O’Shea and co-workers (28)used a similar design with Nafion tubing instead of porous glass. This innovation provided an easier method for construction, resulting in a 100%success rate for larger capillaries. A carbon fiber electrode was again used to detect analytes amperometrically with detection limits in the low-attomole range. A coupler using Teflon heat-shrunk tubing, which appears to be easier to construct and to maintain than the original porous glass design, also has been reported (29).The use of couplers made electrochemical detection in CZE possible by terminating the separation voltage, and therefore noise currents, before detection. The second approach to successful electrochemical detection has been to eliminate the need for couplers through the use of capillarieswith small internal diameters (30).It appears that the potential field caused by the separation voltage decays very rapidly at the end of these capillaries,which allows electrodes to be manipulated up to the end of the column without the use of couplers (29,30).This technique has been demonstrated for the detection of several common neurotransmitters with a 1@pmdiameter carbon fiber electrode placed near the end of a c a p illary, as shown in Figure l b (30).Detection limits as low as 56 amol have been reported for catechol; however, these results are tempered by difficulties in placement of the electrode, which lead to low precision. Recently, a solution to the problem of end column electrode placement was developed (31).The end of a 5pm i.d. capillary is chemically etched to provide a conical entrance for electrode placement. A carbon fiber electrode is then easily inserted and held in the enlarged capillary exit, as shown in Figure IC.An electro-

pherogram of M catechol detected with this method is shown in Figure 2b. Detection limits of 14-19 amol have been achieved for standard samples of compounds commonly found in neurons. In addition to lower detection limits, more reproducible results have been obtained when compared with those obtained using the end column method. End column placement of an electrode has also been facilitated through the use of electrodes with diameters that are large relative to the internal diameter of the capillary (32).Detection limits comparable to those obtained using end column detection have been achieved using microelectrodes, and run-to-run reproducibility using the large electrodes has been high. The noise arising from the separation potential is an important considerationin electrochemicaldetector design for CZE separations. In a recent study by Lu and Cassidy (29),several experiments were performed to determine the nature of this background noise and to provide insights into better detector designs. Examination of end column detection reveals that the separationvoltage has a greater effect on noise when capillaries with large, rather than small, internal diameters are used. In addition, placement of the electrode is more critical for capillaries with larger internal diameters because of the slower decay of the electric field at the column end relative to small-bore capillaries. Coupler designs have been compared with end column detection and are necessary when the internal diameter of the capillary is > 25 pm (29). These conclusionsstress that careful electrode placement in end column detection configurationsis important for improving the detection limits of transition metals, inorganic ions, and catechols (33). Concentrationdetection limits of 5 x lo-’’ M have been reported for cate chol using a 25pm i.d. capillary and a l@umdiameter carbon fiber electrode. In addition, reported efficiencies for different electrode positions at the end of the capillary are highly dependent on electrode placement. The best efficiencies are obtained when the electrode is placed inside the capillary, apparently because convection currents and analyte diffusion outside the capillary are avoided.

Figure 3. Electropherogram of carbohydratesdetected using PAD. Sample concentrations: 1 x 1OP4 M for inositol and 2 x 10-4 M for others. Peak identification: 1, inositol; 2, sorbitol; 3, unknown; 4, maltose; 5, glucose; 6, rhamnose; 7, arabinose; 8 , fructose; and 9, xylose. (Adapted from Reference 38.)

The development of electrochemical detection for microseparationsoffers a highly sensitive method for the study of low-volume environments. However, the selectivity of the technique can limit the range of observable analytes. Much research has been conducted to expand the range of analytes compatible with electrochemical detection. This research has flourished with the development of different electrodes and detection schemes for CZE. Expanding the range of detection

Because of its sensitivity, amperometry is the electrochemicaldetection mode most commonly used in microseparations;however, because the potential of the electrode is held at one value, only those species that are easily oxidized or reduced at that potential are detected. To extend the detectable range of analytes, voltammetry can be used, This technique involves a p plying a triangular potential waveform to the electrode to oxidize and reduce analytes that are electroactive in the potential range used. If the waveform is applied quickly enough, real-time voltammetric analysis of compounds eluting from a microseparationcapillary can be obtained

(34).Fast-scan voltammetric analysis has the added advantage of analyte identification, as different compounds yield different characteristic vo1ta”ograms. In addition, co-eluting peaks can be resolved if the potential at which they are electroactive difters significantly. However, voltammetry suffers from mass detection limits approximately 100 times greater than those for amperometry. Another method that can be used to extend the range of analytes involves metal electrodes. Compounds not detectable at carbon fibers-such as alcohols, glycols, carbohydrates, sulfur compounds, amines, and amino acids-can interact with empty &orbitals on the surface of a platinum or a gold electrode to form an electrochemically active species. A drawback that has limited the use of metal electrodes, though, is the tendency of electrochemically generated products to accumulate on the electrode surface. This process, termed electrode fouling or poisoning, results in the loss of the electrode’s ability to detect analytes. Fouling has been addressed through the use of pulsed amperometric detection (PAD), a well-established technique for HPLC (35).This technique involves rapid, continuous application of a three-step po-

Analytical Chemistry, Vol. 66, No. 9, May l , 1994 531 A

tential waveform to the electrode.The first step is used to detect the analytes; the second step, at a more positive potential, serves to electrochemically clean the surface of the electrode for the next detection pulse; and the third step is at a potential that reactivates the electrode for subsequent detection. The first use of PAD in a flowing stream with a microelectrode was demonstrated in the detection of carbohydrates at a gold electrode after separation by CZE (36).This study also included the use of pulsed voltammetric analysis to gain qualitative information. Further studies have used PAD for the detection of carbohydrates and metal ions (37,38).A representative electropherogram of nine carbohydrates separated by CZE and detected with PAD is shown in Figure 3. The use of metal electrodes does not necessarily imply that PAD is required for successful detection. This was recently demonstrated for two types of metal electrodes. A copper electrode that was electrochemically pretreated to create a surface oxide layer was used to electrocatalytically detect carbohydrates separated with CZE (39).A single copper electrode was used for hundreds of separations during a two-week period with no deterioration in performance. A gold/mercury amalgam electrode was also used to selectively detect low levels of thiols through the catalytic oxidation of mercury -thiol species after CZE analysis (40). The use of metal electrodes has extended the range of observable analytes; however, each electrode discussed above is still generally applicable to a specific class of compounds. This is an advantage when selectivity is desired, but sometimes a broader class of compounds needs to be explored. Potentiometricor conductimetric analyses are alternative techniques that allow detection of more than one class of compounds. In the case of potentiometricanalysis, ion-selective electrodes can be used to detect both inorganic and organic ions at the same time. These electrodes were used to detect several common neurotransmitters in addition to alkali metals (41, 42).With conductimetricanalysis, the conductivity change is measured when an analyte band that has a mobility different from that of the solvent electro-

Figure 4. Chromatovoltammogram of a single nerve cell sampled from the land snail Helix aspersa. Peak identification: DA, dopamine; DHBA, 3,4-dihydroxybenzylamine; 5-HT, serotonin (two peaks result from oxidation of phenol and indole nitrogen at different potentials); TYR, tyrosine; TRP, tryptophan. (Adapted from Reference 50.)

Figure 5. Plot of norepinephrine versus epinephrine contained in individual bovine adrenomedullarycells. Cells containing an undetectable amount of norepinephrine were given a norepinephrinevalue of zero; the dashed line represents a catecholamine ratio of unity. (Adapted from Reference 51 .)

532 A Analytical Chemistry, Vol. 66, No. 9, May 1, 1994

lyte enters the electrochemicaldetection region. As in potentiometry, this technique has been used to detect both metal and organic ions (43,44).In addition, the recent development of suppressed conductivity detection for CZE has resulted in more sensitive detection of analytes when compared with conventional conductimetric detection (45,46). When the analyte of interest is not electroactive and cannot be directly detected by electrochemicalmeans, indirect detection can be used (47).This technique does not depend on oxidation or reduction of a specific class of compounds. By adding an electrophore to the separation solvent, nonelectroactive species can be detected as a dip in the background signal. Although this technique is useful for a broader range of solutes, there is a tradeoff with sensitivity. An alternative to the detection of nonelectroactive solutes by indirect detection is to use chemical derivatization to attach an electroactive group to the solute. This involves chemically reacting the sample prior to analysis with an electroactive tag that specifically derivatizes a target analyte. The most prevalent electroactive tag used is naphthalene-2,3-dicarboxaldehyde (NDA), which reacts with primary amines to form an electroactive compound. NDA derivatization methods have proven to be a valuable addition to electrochemicaldetection techniques, as demonstrated by the analysis of amines in biological samples with CZE and OTLC (20,48).

tents and physiology are considered. Pooling large cell populations for macroscale analysis is one technique that yields average values that are a composite of the attributes of all cells present. This scheme limits the information gained about cellular function when a heterogeneous population of cells is examined. For example, nowhere is the heterogeneity of cell types as complex as it is in the brain. Staining procedures can be used to determine cell contents, but the results are interpreted through visual inspection and are therefore subject to varying interpretations. Considering the problems faced in the investigation of individual cells, singlecell analysis via microseparation techniques becomes an attractive analytical tool. An example of single-cell analysis that led to new biological interpretations was the first whole-cell analysis of invertebrate neurons with OTLC (49).In this experiment, an individual neuron from the land snail HeZk aspersa was homogenized in a 500-nLmicrovial. The sample was centrifuged, and 1-100 nL of the supernatant was injected onto an OTLC column using a micropipette injector. Combined with this cell sampling and separation technol-

ogy, fast-scan voltammetric detection was used to determine neurotransmitter and amino acid levels in three different neurons (49,SU).Each of the three cells resulted in characteristic chromatovoltammograms, one of which is shown in Figure 4. According to this analysis, two neurons of Helh aspersa contained both dopamine and serotonin, and another contained neither neurotransmitter.All three neurons contained tyrosine and tryp tophan. These results refute Dale’s hypothesis that neurons contain only one type of neurotransmitter. In a separate report, 17 nonelectroactive amino acids in single nerve cells of Helh aspersa were determined amperometrically with OTLC after derivatization with NDA (48).The results from these experiments underline the importance of single-cell analysis. Single bovine adrenomedullarycells were recently analyzed using reversedphase microbore LC with amperometric detection (51). These cells are - 20 pm in diameter and are 10-20 times smaller than invertebrate cells. This analysis r e g resents an important step forward in singlecell sampling methodology. Norepinephrine and epinephrine concen-

Single-cell analysis

One of the more challengingareas of biological analysis is the separation of the contents of single cells. Microcolumn LC and CZE can be used to sample volumes as low as nanoliters to femtoliters; therefore, the methodology exists to explore these ultrasmall environments.Because of its selectivity and sensitivity, electrochemical detection is also well suited for studying single cells. Several biologically interesting species are electrochemically active, and the selectivity of electrochemical detection allows these species to be detected despite the complexity of biological samples. The importance of performing singlecell analysis becomes apparent when the traditional means of determining cell con-

rigure 6. CZE analysis of a whole single nerve cell from the pond snail Planorbis corneus. Peak identification: A and B are apparently dopamine (different retention times suggest storage in different vesicles); C, neutral species; D, uric acid; E, dihydroxyphenylacetic acid. (Adapted from Reference 55.)

Analytical Chemistry, Vol. 66, No. 9, May 1, 1994 533 A

trations were evaluated in individual cells to test the hypothesis that two distinct types of adrenomedullary cells exist: one containing mostly norepinephrine and the other containing mostly epinephrine. The concentrations of these components in 22 cells are shown in Figure 5. The results indicate that most adrenal cells contain either norepinephrine or epinephrine; four cells, however, contained significant levels of both hormones. The work discussed above involved manipulation of single cells in nanoliter microvials before the analysis. Single-cell sampling can also be performed directly in a separation capillary for analysis by CZE with electrochemical detection (52, 53).The sample injection procedure for cytoplasmic samples involves chemically etching the separation capillary at the sample introduction end to a fine tip of 6 - 10-pm outer diameter (9,53).The tip is inserted into the cell of interest, and a cytoplasmic sample is removed by applying the separation voltage for a few seconds to electroosmotically inject a small aliquot into the capillary. The capillary is then placed in a buffer solution, and separation is carried out. Cytoplasmic levels of the neurotransmitters dopamine and serotonin were determined amperometrically in two different neurons found in the pond snail Planorbis comeus (9,54).The technology demonstrated in this study established a method for directly sampling a single-cell cytoplasm into a capillary electrophoretic system. CZE in 25pm i.d. capillarieswas also used to sample single whole neurons in Planorbis comeus (53).In this experiment, electroosmotic flow was used to inject a cell into an etched capillary, and subsequent injection of nonphysiological buffer was used to lyse the cell membrane. This methodology has recently been used to test the hypothesis that neurotransmitters exist in two vesicular compartments in a nerve cell (55).Apparently, one compartment is for immediate use in neurotransmission and the other is for longer term storage. Changing the waiting period for cell lysis following injection into the end of the capillary results in a dramatic change in the electropherogram. Short lyse times result in two peaks for dopamine that can be attributed to the different resistances of the two vesicle types to lys-

ing (Figure 6). One of these peaks is abolished following treatment of the cell with the vesicle-depleting drug reserpine. The method shows promise as a means to probe vesicular concentrations of dopamine at the single-cell level. Future directions

In addition to aiding future analysis of single cells, the development of electrochemical detection for microseparations will undoubtedly provide improvements in detection limits and new applications.To increase sensitivity,techniques such as sinusoidal voltammetry can be used to achieve picomolar detection limits in flowing streams and are being applied to CZE (56).More biological techniques, such as coupling CZE to microdialysis to monitor in vivo release of amino acids (57),are on the horizon. The development of minia-

turized and chemically or biologically modified electrodes should extend the range of analytes observed using electrochemical detection (58).If capillarieswith submicrometer internal diameters become available, amperometry will be well suited for detection. Electrodes with total tip diameters of 400 nm have already been constructed and could be used with these capillaries (59). New microseparation methods should also benefit from the advantages of electrochemical detection. In our laboratories, we have developed a continuous microseparation technique based on CZE (60).This has been accomplished via sep aration in a channel structure that has an internal height and width of 48 pm and 2 cm, respectively.Sample introduction is carried out using conventional CZE to continuously sample material from mi-

Figure 7. Experimental setup for continuous electrophoresis in narrow channels. (a) Sample is placed in reservoir 1, and sampling capillary is attached to a stepper motor that moves it across the entrance of the channel. The channel is suspended across buffer reservoirs 2 and 3. (b) Representation of the separation process in the channel for three analytes with different mobilities. Left: the analytes first being deposited into the channel. Middle: the migration paths after the capillary has moved to the middle of the channel. Right: the capillary at the end of the channel entrance, where sample deposition is typically halted. (Adapted from Reference 60.)

534 A Analytical Chemistry, Vol. 66, No. 9, May I, 1994

croenvironments. The end of the CZE capillary is inserted into the entrance of the channel and is slowly moved across its width to allow continuous deposition of sampled material. Material migrates in the channel in straight paths and separates along the 5-cm length of the channel. The principles of this new separation mode are illustrated in Figure 7. Although LIF detection with fiber-optic arrays to deliver excitation light and collect fluorescence has been used to demon-

strate the separation method, we are currently implementing an electrochemical detection scheme for continuous channel separations. The detector consists of 100 gold microelectrodes chemically deposited across the channel plate at its exit. Each electrode is individually addressed to provide a response based on the position of an eluting analyte across the channel exit width. Figure 8 shows a schematic of the channel with a micrograph of the electrode array detector. The electro-

chemical detection array should provide improved detection limits and will allow further miniaturization of the electrophoresis channel internal height. In conclusion, electrochemical detection has emerged as one of the more powerful detection methods for microseparations because of its selectivity,sensitivity, and ability to detect low levels of many biologically relevant molecules without prior derivatization.The future of electrochemical detection appearsJo be bright, as evidenced by recent developments in the construction of new electrodes and the implementation of different detection schemes. In addition, the use of microseparations to investigate single cells has demonstrated the applicability of this technique in biochemical analyses. We acknowledge our co-workers whose efforts are referenced in this Report. We thank NSF, NIH, and the Office of Naval Research for financial support. We also thank Dave Lilienfeld, Garry Bordonaro, and the NSF-funded National Nanofabrication Facility at Cornel1 University for assistance in developing the microelectrode array.

References

Figure 8. Schematic of electrode array detector for continuous electrophoresisin narrow channels. Beige rectangular portion represents the top channel plate (quartz, 2.5 x 5.0 x 0.23 cm). Electroactive region is composed of 100 individually addressed gold microelectrodes (each 95 pm wide; 5-pm interelectrode spacing). The electrodes are fabricated directly onto the bottom channel plate (quartz, 7.5 x 7.5 x 0.23 cm) by standard microlithographic techniques. An individual address to each electrode is accomplished through the termination at the appropriate contact'pad. The inset micrograph shows a 10-electrode portion of the array. The white horizontal portions are electroactive regions; the black portions are quartz substrate.

(1) Novotny, M. Anal. Chem. 1988,60, 500 A-510 A. (2) Jorgenson, J. W.; Lukacs, K. D. Science 1983,222,266- 72. (3) Ewing, A. G.; Wallingford, R A.; Olefirowicz, T. M. Anal. Chem. 1989,61, 292 A- 303 A. (4) Rice, C. L.; Whitehead, RJ. Phys. Chem. 1965,69,4017 - 24. (5) Hjerten, S. Chromatogr.Rev. 1 9 6 7 , 9 , 122- 219. (6) Virtanen, R Acta Polytech. Scand. 1 9 7 4 , 123,l- 67. (7) Mikkers, F.E.P.; Everaerts, F. M.; Verheggen, T.P.E.M. J Chromatogr. 1979, 169,ll- 20. (8) Jorgenson, J. W.; Lukacs, K. D. Anal. Chem. 1981,53,1298-1302. (9) Olefirowicz,T. M.; Ewing, A. G. Anal. Chem. 1990,62,1872-76. (10) Kennedy, R T.; Oates, M. D.; Cooper, B. R; Nickerson, B.; Jorgenson, J. W. Science 1989,246,57- 63. (11) Ewing, A. G.J. Neurusci. Methods 1 9 9 3 , 48,215 - 24. (12) Yu, M.; Dovichi, N. J. Anal. Chem. 1989, 61,37 -40. (13) Tsuda, T.; Sweedler, J. V.; Zare, R N. Anal. Chem. 1990,62,2149- 52. (14) Chervet, J. P.; Van Soest, RE.J.; Ursem, M. J. Chromatogr. 1991,543,439 - 49. (15) Taylor, J. A; Yeung, E. S.J. Chromatogr. 1991,550,831-37. (16) Wang, T.; Aiken, J. H.; Huie, C. W.; Hartwick, R A Anal. Chem. 1991,63, 1372 -76. (17) Xue, Y.; Yeung, E. S. Anal. Chem. 1 9 9 3 , 65,1988-93. (18) Wu, S.; Dovichi, N. J. Chromatogr. 1989,

Analytical Chemistry, Vol. 66, No. 9, May 1, 1994 535 A

fleaoff? 480,141-55. (19) Sweedler, J. V.; Shear, J. B.; Fishman, H. A; Zare, R. N.; Scheller, R H. Anal. Chem. 1991,63,496-502. (20) Hogan, B. L.; Yeung, E. S. Anal. Chem. 1992,64,2841-45. (21) Lee, T. T.; Yeung, E. S. Anal. Chem. 1992,64,3045-51. (22) Smith, R N.; Wahl, J. H.; Goodlett, D. R; Hofstadler, S. A. Anal. Chem. 1993,65, 574 A- 584 A. (23) Lee, E. D.; Henion, J. D.; Covey, T. R

J. Microcol. Sep. 1 9 8 9 , 1 , 14- 18. (24) Moseley, M. A; Deterding, L. J.; de Wit, J.S.M.; Tomer, K. B.; Kennedy, R T.; Bragg, N.; Jorgenson, J. W. Anal. Chem. 1989,61,1577-84. (25) Manz, A.; Simon, W. J. Chromatogr. Sci. 1983,21,326 -30. (26) Knecht, L. A; Guthrie, E. J.; Jorgenson, J. W. Anal. Chem. 1984,56,479 - 82. (27) Wallingford, R A.; Ewing, A. G. Anal. Chem. 1987,59,1762 -66. (28) O'Shea, T. J.; Greenhagen, R D.; Lunte,

S. M.; Lunte, C. E.; Smyth, M. R; Radzik, D. M.; Watanabe, N.J. Chromatogr. 1992,593,305- 12. (29) Lu, W.; Cassidy, R M. Anal. Chem. 1994,66,200 -04. (30) Huang, X.; Zare, R N.; Sloss, S.; Ewing, A. G. Anal. Chem. 1991,63,189-92. (31) Sloss, S.; Ewing, A. G. Anal. Chem. 1 9 9 3 , 65,577 - 81. (32) Ye, J.; Baldwin, R P. Anal. Chem. 1 9 9 3 , 65,3525 - 27. (33) Lu, W.; Cassidy, R M.; Baranski, A. S. J. Chromatogr. 1993,640,433-40. (34) White, J. G.; St. Claire, R L., 111;Jorgenson, J. W. Anal. Chem. 1986,58, 293 - 98. (35) Johnson, D. C.; Lacourse, W. R Anal. Chem. 1990,62,589 A- 597 A. (36) O'Shea, T. J.; Lunte, S. M.; Lacourse, W. R Anal. Chem. 1993,65,948-51. (37) Lu, W.; Cassidy, R. M. Anal. Chem. 1993, 65,1649-53. (38) Lu, W.; Cassidy, R M. Anal. Chem. 1993,65,2878-81. (39) Colon, L. A; Dadoo, R.; Zare, R N. Anal. Chem. 1993,65,476-81. (40) O'Shea, T. J.; Lunte, S. M. Anal. Chem. 1993,65,247 - 50. (41) Haber, C.; Silvestri, I.; Roosli, S.; Simon, W. Chimia 1991,45,117-21. (42) Nann, A; Silvestri, I.; Simon, W. Anal. Chem. 1993,65,1662-67. (43) Huang, X.; Pang, T. J.; Gordon, M. J.; Zare, R N.J. Chromatogr. 1988,425, 385 - 90. (44) Huang, X.; Luckey, J. A; Gordon, M. J.; Zare, R. N. Anal. Chem. 1989,61, 766 - 70. (45) Dasgupta, P. K.; Bao, L. Anal. Chem. 1993,65,1003- 11. (46) Avdalovic, N.; Pohl, C. A.; Rocklin, R D.; Stillian, J. R. Anal. Chem. 1993,65, 1470- 75. (47) Olefirowicz,T. M.; Ewing, A. G. J. Chromatogr. 1990,499,713- 19. (48) Oates, M. D.; Cooper, B. R; Jorgenson, J. W. Anal. Chem. 1990,62,1573-77. (49) Kennedy, R T.; St. Claire, R L., III; White, J. G.; Jorgenson, J. W. Mikrochim. Acta 1 9 8 7 , 2 , 3 7-45. (50) Kennedy, R T.;Jorgenson, J. W. Anal. Chem. 1989,61,436-41. (51) Cooper, B. R; Jankowski, J. A.; Leszczyszyn, D. J.; Wightman, R M.; Jorgenson, J. W. Anal. Chem. 1992,64, 691- 94. (52) Wallingford, R A.; Ewing, A. G. Anal. Chem. 1988,60,1972-75. (53) Olefirowicz,T. M.; Ewing, A. G. Chimia 1991,45,106- 08. (54) Olefirowicz, T. M.; Ewing, A. G. J. Neurosci. Methods 1990,34,11- 15. (55) Kristensen, H. K.; Lau, Y. Y.; Ewing, A. G. J. Neurosci. Methods, in press. (56) Cullison, J. K.; Kuhr, W. G., unpublished work. (57) O'Shea, T. J.; Weber, P. L.; Ba"e1, B. P.; Lunte, C. E.; Lunte S. M.; Smyth, M. R J. Chromatogr. 1992, 608,189-95. (58) O'Shea, T. J.; Lunte, S. M. Anal. Chem. 1994,66,307- 11. (59) Strein, T. G.; Ewing, A. G. Anal. Chem. 1992,64,1368-73. (60) Mesaros, J. M.; Luo, G.; Roeraade, J.; Ewing, A. G. Anal. Chem. 1993,65, 3313-19. 4

I

CIRCLE 2 ON READER SERVICE CARD

536 A

Analytical Chemistry, Vol. 66,No. 9,May 7, 1994

~

I

Andrew G. Ewing is a professor of chemistry at Penn State (152 Davey Laboratory, University Park, PA 16802). He received his B.S. degreefiom St. Lawrence University (Ny)in 1979 and his Ph.D. in analytical chemistry fiom Indiana University in 1983. He then spent 13 months as a postdoctoral fellow at the University of North Carolina-Chapel Hill. He is a Camille and Henry Drey%s Teacher-Scholar. His research interests include CE, electrophoresis at ultrasmall electrodes, and the application of these methods to single-cell and neurochemical analysis.

- --.\-~

,

i

Jody M. Mesaros is a fourth-yeargraduate student at Penn State, where she is pursuing a Ph.D. in analytical chemistry. She received her B.S. degree in chemistryfiom Temple University in 1990. Her research interests include continuous electrophoresis in narrow channels and instrumental design and development.

...Is Yours Today With MACROMOLECULES

n

U

n t h e ever changing field of polymer chemistry, if you’re not reading MACROMOLECULES, you’re missing what your colleagues regard as the most authoritative, in-depth coverage anywhere. Each biweekly issue of MACROMOLECULES will bring you up-to-date, peer-reviewed coverage of all aspects of polymer chemistry including: A synthesis A polymerization mechanisms and kinetics A chemical reactions A solution characteristics A bulk properties of organic polymers, inorgan polymers and biopolymers For information you can trust to keep you on t h e cutting edge of t h e science of polymer chemistry, you simply cannot afford to do without a subscription to MACROMOLECULES.

7 US. I

CanaddMexico

I

Europe

I

Elsewhere

I

~~

ACS MEMBERS 1 YEAR

NONMEMBERS

1 YEAR

(Member rates are for personal use only. Journal subscriptions start January 1994. For nonmember subscriptions in Japan, contact Maruzen Company, Ltd.)

To place your order simply call 1-800-333-9511in the US.*,or send your order to: Peter F. Gavin is a second-year graduate student at Penn State, where he is pursuing his Ph.D. in analytical chemistry. He received his B.S. degree in chemistry from the State University of New York College at Fredonia in 1992. His research interests include the development of sensitive detection techniques for CE and microcolumn LC involving electrochemical or mass spectrometric techniques.

ACS IIIPUBUCATIONS h e n t i a l Resources the Chemical Sciences .for

American Chemical Society Member and Subscriber Services

P.O. Box 3337 Columbus, OH 43210 * Outside the U.S. dial (614) 447-3776,or you may fax your order to 614-447-3671.

MACROMOLECULES Editor: Field Winslow -AT&T Bell Laboratories Analytical Chemistry, Vol. 66, No. 9, May 1, 1994 537 A