Flow cytometry and sorting - American Chemical Society

P.O. Box5507 L-452. Livermore, Calif. 94550. Instrumentation. FlowCytometry and Sorting. The ability to rapidly obtainquan- titative information on ce...
3 downloads 0 Views 15MB Size
Dan Pinkei

Instrumentation

Lawrence Liverme National Laboratory Biomedical Sciences Division University of California P.O. Box 5507 L-452 Livermore. Calif. 94550

Flow Cytometry and Sorting The ability to rapidly obtain quantitative information on cells or other particles of biological interest is of major importance to medicine and biology. Classification and sorting of individual chromosome types, detection of malignant cells by determining their protein and DNA content, and searching a large population of cells for rare mutants require rapid automated techniques. The development of flow cytometry, in which a stream of suspended particles is made to flow on a controlled trajectory past optical or electrical sensors, has been motivated by these needs. Work over the last 15 years has

resulted in instruments capable of measuring several parameters of each particle at a rate of thousands of particles per second, with the resulting data sets stored and analyzed by computer. In addition, the computer can make decisions based on the measurements rapidly enough to allow sorting of desired particles from the general population for further study. Measurements of such parameters as fluorescence (both intrinsic and of cellcomponent-specific stains), fluorescence depolarization, phosphorescence, light scatter, absorbance, and electrical conductivity and capacitance are possible.

Flow cytometry has found widespread application, with several hundred installations in use worldwide. Three commercial manufacturers supply general-purpose flow cytometers and sorters, and several others build dedicated clinical instruments. Many laboratories choose to build their own for specific applications or exploration of advanced techniques. This article describes the.basic technology and capability of the field, drawing examples from work being done at Lawrence Livermore National Laboratory. Much more detail is contained in References 1,2,and 3. Reference 4 is a comprehensive source of information

I

t

\

Sheath Fluid

Pulse

i

-

2.B 30

a1

Pulse Area rPulse Width

Flow Chamber I ,

,

Time

,Optical Detectors

' k a a

z"

AA Signal Property

Figure 1. Flow cytometric Instrumentation Vibratinglhe flow dumber caus8s bIiquM jed to break up into regular drops. Elenlcal charging of the appqxiate drqos allows swting 01 individualpanicles based on b i r measured parameters W03-2700/82/035 1-503AS01.OO/O Q 1982 American Chemical Society

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3, MARCH 1982

503A

INA Content

Figure 2. The cell cycle (a) M A cartent d a call doubles befae It divides. (b) Distributionof ONA mn(enl d a Wpulatlon of cells m m l y spaced though the division cycle. (c) Corn puter anaiwls of flow cytomeb’lc measurementsallows quantlflcetlm of VNI IraCfion of cells in the various stases of the cycle. The dashed line separams VNI ConblbuIlon of Sphase cells from VIoae of the other phases

in this area, and its more than 70 contributors indicate the scope of the international effort. Many of the articles related to new developments in flow instrumentation and applications are published in the journal Cytometry.

lnstrumeniatlon To accurately measure suspended particles, the flow must be controlled so that each particle follows nearly the same trajectory through the analysis sensors. This is accomplished by establishing a sheath of nonturbulent fluid centered around a small-diameter sample injection tube (Figure 1). The sample stream exits from the injection tube, the stream diameter decreasing further as the sample passes through the nozzle of the flow chamber. In typical instruments, the sheath exit orifice is on the order of 200 pm in diameter while the sample stream (core) diameter is several micrometers. The use of “hydrodynamic focusing” avoids the problems of clogging and velocity gradients across the sample stream that would occur if control of the particle trajedaries were attempted using an orifice of the core diameter dimension. Fluid velocity in the orifice is typically 10 m/s, attained by applying a fraction of an atmosphere pressure. The Reynolds number of the flow is low enough so that it is laminar and extremely stable. The flow field orients elongated objects with their long axis parallel to the direction of flow, and by special shaping of the orifice and sample injection tube, even the orientation of flat cells about the flow axis can be controlled. It has been found that cells can with504A

stand passage through the instrument and be viable for further study. CW lasers with outputs ranging from milliwatts to a few watts in a single line are usually used for cell illumination, although mercury lamps also work well. To maximize illumination of the particles, the light source is focused on the sample stream. Focus across the core (perpendicular to the flow direction) is limited by the need .to have uniform illumination over the range of particle trajectories allowed by the sample stream diameter. Focus along the stream is frequently much tighter, to shorten the time of illumination or to extract one-dimensional morphologic information by scanning individual particles as they pass through the beam. Collection optics focused on the laser beam-sample stream intersection point pick up the signals and transmit them to photodetectors. Photomultiplier tubes are universally used for fluorescence detection, and photodiodes are suitable for the larger signals of forward scattering. Illustrated in the upper right of Figure 1is the hypothetical output of a detector as a function of time for the passage of a single object through the beam, which typically takes 10-6 s. The dimensions have been chosen such that the length of the object, shown oriented as discussed above, is larger than the laser beam. If internal optical properties of the cells can be neglected, the time course of the fluorescence intensity is given by the convolution of the laser beam intensity profile with the distribution of stain along the particle. Characteristics of the pulse that might be measured are

ANALYTICAL CHEMISTRY, VOL. 54. NO. 3, MARCH 1982

area, width, height, and details of the profile. The desired pulse parameter is digitized and measured by a multichannel analyzer. The data is displayed as a histogram that shows the number of particles vs. the value of the measured parameter. Two different-sized particles are shown in the sample stream in Figure 1.Each population is responsible for one of the peaks in the resulting histogram. Individual objects can be sorted based on the measurements. Mechanically, sorting is usually accomplished by vibrating the flow chamber with a piezoelectric crystal while allowing the fluid to exit into the air. The vibration causes the fluid jet to break into regular droplets as illustrated in Figure 1. Typical droplet formation frequencies are in the tens of kHz range with jet diameters under 100 pm. If, on the basis of the cytometric measurements, a cell meets preset criteria, a voltage is applied to the sheath (which is a conducting fluid such as saline) with the appropriate delay and duration so that the drop containing the desired cell is electrically charged when it breaks off. Several hundred microseconds are available for the electronics to make the sorting decision. The droplets pass between a pair of charged plates, and those with a charge are deflected into the collection vessel. The more sophisticated instruments can measure several properties of each particle. The use of a conducting sheath fluid allows the effect of the presence of a cell on the ac or dc conductivity of the flow chamber orifice to be measured. Developed by Wallace

New programmable UV-100detector makes the 5000 LC even better

@varian

5000 LIOUID CHROMATOGRAPH

The chromatogram shows how the new bmldt-m UV-100 detector permit^ a~tomat~c lime p,ogmmmmg 01 wavelength and range to improve an analysis The UV IW 1s controlled through the interactwe CRTand keyboard of the Model 5oW Liquid Chromatograph

Rellabie. fully integrated. and simple to workwith. theVarian Model 5000 is widely recognized as the best liquid chromatographyou can buy. Now, the best LC is even better. We've built-ina fullyautomated.variable wavelength detector, the sensitive. fast new UV-100. Now, everything in the 5000 system is automated and it's ail programmed and controlled through the same heipful interactive CRT and keyboard. Time program wavelength selection, detector au1ozero and range The UV-100 allows you to select as many wavelengths as you need to optimize detection of all components in a run. v,a simple keyboard entries, you program any wavelengths you want, in increments as small as 1 nm. from 190nm to 370 nm. at exactly the timeS you want. youget optlmum detection 01 each component as it elutes from your LC Your wavelength entries are part of the complete LC program stored in memory and displayed on the CRT Over 100 unattended waveiength changes are possible for overnight operation or automated methods development. In addition, to improve the accuracy of your data and the quality of your report, you can attenuate peaks and rezero the baseline at any point in a ruii and as often as you like. Also, through program linking, you can change the 3 time constants up to 9 times during an analysis. Patented' flow cell and selectable time constants for best detection We designed the UV-100for conventional as well as high speed LC applications. With 15 to 30cm analytical columns at normal fiow rates, the 4.5 microliter fiow cell and 0.5 sec time constant give you

optimum peak heights and resolution. For higher throughput with shorter columns at higher flow rates, the 50 millisecond time constant preserves peak fidelity. The patented new flow cell maximizes light throughput and minimizes flowiR.1. effectsthrough controlled optical imaging.

Integrated to do mom in less space We saved you important bench space by incorporating the optics. electronics. and microprocessor control of the UV-100 in the M y integrated Model 5CCO In a single unit. that takes only 31 inches of bench space. YOU have uniquely-reliable single-pump ternary gradient capability. programmable c o b n n temperature control. manual Or automated sample injection. and the fully programmable uv-100 detector. One keyboard controls all LC functions. from a single clock, to keep all eiementsof the system in Perfect synchrony. The CRT monitor lets you know the immediate status of all system operating parameters and guides you in making new entries. Noother LC system does so much in so linle space. Further,each Modei 5000 comeswith full support from Varian. training courses, iocal service and applications ~4' assistance. 1 For full details on the fuily automatic Model 5000 circie Reader Service I Number241. To have a Varian Representative contact you circle Reader Service Number242. ~

'Patent pending.

v

The UV~lWis buin-in fheModelSWD.

varian FO~immcdiatcauiatan~contrct:.Florham Park. NJ (201)822-3700 ParkRldge. lL(312)825-7772. Hou~ton.TX(713) 783~1800* LorA1105,CA(4,5)968~8141 Orwrile 811 Hansen Way, Palo Alto. CA94303. In Europe. SfeinhausersfraSse. CH-6300 Zug.SwQeriand.

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3, MARCH 1982

505A

Major shortcomings of most LC columns currently available have included inadequate reproducibility of retention, selectivity and performance (efficiency, permeability, peak symmetry), plus poor chemical and mechanical stability (short lifetime). NOW! ULTRASPHERE'" columns represent a series of advances in modern column technology which significantly reduces these problems. Advanced Packing The Reversed Phase ODS (C-18), Octyl (C-8) and IP (Ion Pairing) columns feature a maximum coverage, endcapped bonded si Iica-based material. These columns are packed with 5 ,um spherical particles having extremely uniform particle and pore size distributions for high efficiencies and lower pressure drop.

Improved Retention and Selectivity Based on the carbon content and surface area, the ULTRASPHERE columns possess a maximum surface coverage of the bonded moiety (C-18 and C-8). The remaining accessible silanols are endcapped with a second silanization to minimize the undesirable influence of unreacted accessible silanol groups on solute retention. The absence of micropores (as found in some silicas) also minimizes unbonded silanols. This ensures good peak shape with excellent reproducibility in column-to-column (batch-tobatch) retention and selectivity. Extended Column Lifetime Column stability is further enhanced by a new proprietary packing process. The result is a column with high chemical and mechanical stability

which resists troublesome headspace formation that will degrade column performance. We believe these columns represent state-of-the-art technology and should enable you to attain the best possible results from your liquid chromatograph. To ensure maximum column performance, each and every column is individually packed and tested. Sample test mixture, test chromatogram and conditions are supplied with each column. For more information, write for Beckman Catalog 7358. Or ask your local Sales Representative. Beckman Instruments Inc., Scientific Instruments Division, 1780 Fourth Street, Berkeley, CA 94710

EECKMAN CIRCLE 30 ON R t A D E n SERVlCt CARD

A New Generation of Columns

for liquid Chromatography

Beckman reinventsthe liquid chromatograph. First as Altex, and now as part of Beckman, we have led the way in HPLC pump technology. Seven years ago we introduced the innovative Model 100, a dual piston, electronic pulse corrected pump-still considered the standard of the industry. This was followed by the cost-effective 11OA-the world’s f i t high performance single piston rapid refill pump.

Now, 10,000 pumps later, Beckman introduces the Model 112with the p-Flowm Solvent Delivery System, combining the best of both technologies into the most advanced solvent delivery system available. The Model 112 was “reinventd to meet industry needs demanding a liquid chromatograph with precision and accuracy plus trouble

mechanical parts, the 112 provides greater reliability.

freeoperation. Built right into the high-pressure piston chamber is a newp-Flow””transducer. Regard-

L’ ~

,

--]

less of solvent changes, p-Flow System automatically adjusts the compressibility compensation to produce flow rates accurate down to .01 ml/min.

We reinvented the drive system using direct torque conversion -no gears! With fewer

,

Tot@ new 340 Series utilizing the p-Flow Solvent Delivery System, is designed for high resolution gradient separations. Combined with the 160165 Detectors you receive unsurpassed HPLC performance.

E

Write for brochure #7396 for more information on t h e Model 112 Solvent Delivery System and 340 Systems HPLC, 1716 Fourth St., Berkeley, CA 94710 o r call your Beckman representative.

3

CIRCLE 31 ON READER SERVICE CAR0

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3. M A R C H 1982

507A

Can Your GC Analyze It Can

Pyroprobe: The Pyroprobe expands the capabilities of your GC, GUMS or GC/IR by allowing you to analyze most solids including synthetic polymers, geo-polymers and biopolymers. And that's not all. A Series 120 Pyroprobe can perform: 0 Analytical Pyrolysis 0 Headspace Analysis of Volatile Organics 0 Simulated Combustion Studies 0 Simulated Thermal Analysis

Pyroprobe;.. The Solid Solution

Coulter, the dc conductivity measurement provides information on cell volume. Cytometers with an electrical sensor plus one or more excitation beams spaced along the flow path have been developed. Since the flow is stable, signals from the same particle can be associated by their fixed separations in time. Collection, manipulation, analysis, and display of this multiparameter data require computer systems. To facilitate sorting, dedicated microprocessors are coming into widespread use. Sample preparation is often a major task. If the study involves solid tissue, dispersal of the cells into a dilute single particle suspension without altering the characteristics of interest is required. While in some applications size information obtained electrically or by light scatter is sufficient, most of the time fluorescent stains (probes) are employed. Probes specific for DNA, RNA, proteins, and membranes are available, and fluorescently tagged antibodies for specific cellular components are frequently used. Dyes sensitive to pH or membrane potential give information on the functional state of the cells. Changes in the local environment of dyes in the membrane and cytoplasm can be monitored by fluorescence depolarization measurements. Energy transfer between different fluorescent molecules has been used to obtain information on the mutual distribution of their binding sites. In what follows I will describe some applications of this technology to measurements on cells and chromosomes drawn from work at Lawrence Livermore National Laboratory. I wish to emphasize that work in these and many other areas is under way in a large number of other laboratories.

Cell Cycle Kinetics

Chemical Data Systems, Inc. 7000 Limestone Road Oxford, PA 19363 Phone 215-932-3636 Telex 83-5308 >IRCLE 36 ON READER SERVICE CARE

508 A

ANALYTICAL CHEMISTRY, VOI

Cell reproduction follows a series of distinct stages whose detailed characterization is of significance to both fundamental biology and clinical medicine. Flow cytometry has made major contributions to these studies since probes for various biochemical constituents, such as DNA and proteins, which vary throughout the division cycle, allow monitoring of the stage of each cell. Much of this work is directed toward understanding and improving cancer chemotherapeutic agents, which may interrupt the cell cycle or have maximum effectiveness a t specific points. One cell constituent that can he used to monitor the progress of a cell through the cell cycle is DNA. The DNA content of a cell doubles before division so that each of the daughter cells has a complete copy of the genetic information. In Figure 2a, DNA content is plotted as a function of time NO. 3, MARCH 1982

DNA

Distribution

~ v c l eD

Figure 3. Computer simulation of a synchronized cell population The distribution 01 cells mrWgh lhe cell cycle is reflected in the DNA COntent distribution

during one cycle of division. The initial period with l x DNA content is called Growthl (GI). During this period, the biochemistry appropriate for beginning DNA synthesis is carried out. During the DNA synthesis (S) phase, the DNA content continuously increases. After the completion of the doubling, the phase called Growth2 (Gz)is reached, where preparation for cell division, labeled M for mitosis, is made. Two daughter cells in the GI phase of the cell cycle result from the division. The distribution of DNA content of a population of cells that are randomly distributed in their cycles is shown in Figure 2b. If the cells are stained fluorescently for DNA and measured with a flow cytometer, a distribution such as Figure 2c results. The features of the fluorescence distribution are not as sharp as that of the actual DNA distribution due to slight variability in stain uptake, nonspecific staining of cellular components besides the DNA, and instrumental factors. The analysis of these fluorescence histograms using computer models allows determination of the fraction of cells in the various compartments of the cell cycle. Dyes such as Hoechst 33342 do not interfere with cell viability, so living cells can he sorted based on their DNA content and then grown in culture.

Good GC Does Not HaveTo Be Expensive Let Shimadzu’s 26 years of expeAence provide you with economical solutions to your problems. 4 C-R2A The microcomputer data processor features 2 channel operationfor GCand LC. Simpleone keystrokeoperation is expanded by Basic language program capability. tion. Combined with GCSA offers to

GC-8Ab

c

4 GC-QA Thecompact microprocessor controlled system fea-

--

C-Rl 6 b The single channel data processor with wide builtin printer plotter, operates equal to or better than any comparable processor available. The easy to operate system is the companion to upgrade your GC from manual data reduction.

and self diagnostics. It has the highest ratio performance/cost available.

1

I

4GC-uim2 Smallest GC available, but has the performance of most systems 10 times its size. Modular design invitesse lective automation (sample,data and temperature).

I ....

selectable Automation S k economical advantage. Compare beiore you make your next purchase. ay for more information on these and other Shimadzu instruments. SHIMADZU SCIENTIFIC INSTRUMENTS, INC. 9t,7-HR~dBranshRoad.Columbla.Md.2t045.U S.A Phone (301)997-1227Telex 87959

SHIMADZU (EUROPA) GmbH Iwo DOosBldolfl.JohBnnBsWever Str.1 F R.Gerrnsny Phone (021 11314061 Telex 06566639

SHIMADZU CORPORATION INTERNATIONAL MARKETING DIV. ShinjdwMitsuiBuiMmp. 1-1. NIshi$hinlYku.2-ehorne. ShmiUmU.mu. Tokyo lW, Japan. Phone Tokyo 03-346-5bll Telex: 0232-3291SHMDT J.

CIRCLE 189 ON READER SERVICE CARD

I

I

only

Scott! "

1 2

could have saved himsell some frostbitten fingers had he been able to order Scott gases while pursuing his cryogenic researches. Much more fortunate is today's chemist Heneedonlypickup his copy of the Scott Specialty Gases Catalog to locate any one of 180 pure gases and gas mixtures in an infinite number 01

concentrations. And if one doesn't want ta manhandle large, unwieldy cylinders, there are Scottye gases in disposable containers 17 pure gases and 99 stock gaseous mixtures, some with as many a s 9 components.

You or your colleagues have Drobablv been buvinocalibration gases from Scott: If you've been too busy analyzing unknowns, though, to analyze the benefitsol buying your carrier gases from Scott too, you may bepassing up a good thing. Scotts' reputation for quality and service is unexcelled for calibratior and carrier gases. Whether yourfield issingle-oi multi-digit Kelvin, you will finc the 160-page Scott Specialt) Gases Catalog a useful tool Request your personal copy today.

Scott Specialty G ~ e s

-

a divlion 01 Scott Environmental

Technolwy Inc

-

Plumrteedrllb. PA 18949 215:7604.961 S.n Bemrdlno, CA924ll -714:-7-2571 Troy, MI W84 313: Y1929SO

CIRCLE 188 ON READER SERVICE CAR1

510A

Relative Fluorescence

Figure 4. Flow cytometric analysis of Chinese hamster chromosomes Above each peak in the flwxeacence distribution is a drawing of !he conespanding chomo6ome type

The action of drugs that perturb the cycling of cells can be monitored flow eytometrically by following changes in DNA distribution after drug administration. Figure 3 is a computer simulation of three such serial distributions. Initially all of the cells are concentrated in GIphase, as might exist after the administration of a drug that halts their progress through the cell cycle at this point. This state of the cell population yields a single peak in the DNA distribution. When the block is removed the cells resume cycling, but not all at exactly the m e rate. By 2.9 h many of the cells have progressed into S phase, which is indicated by the presence of cells with higher DNA content. After several passes through the cycle, the dispersion in cell cycle times will randomize the distribution of the cells. Drug therapy for cancer is complicated by, among other things, the fact that cancer cells differ only slightly from normal cells. Thus, it is difficult to find biochemical differences that can be exploited to give preferential killing of a tumor in the midst of normal cells. Differences in the cell cycle offer an alternate possibility. This approach is based on the fact that some drugs kill by interfering with a cellular process at a particular part of the cycle, S phase for example. If such an S-phase-specific drug were administered to an unsynchronized cell population characterized by the DNA content distribution of Figure 2, less than half of the cells would be killed. However, if it were given 5.7 h after the release of the block in Figure 3, most of the cells would die. If a mixed population of two kinds of cells with different cycling rates, tumor and normal for

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3, MARCH 1982

example, are blocked and then released at the same time, there may he a later time at which the ratio of the killing of the tumor to normal cells will be maximized. Complicated studies of drug action on normal and tumor cells in whole animals, in which flow cytometric techniques play a major role in monitoring the effect on the cells, are now under way to improve our understanding of this area.

Chromosomes The DNA of higher life forms is packaged in compact units called chromosomes. Normal humans have two each of 22 different chromosomes (one from the mother and one from the father) and either two X chromosomes (female) or an X and a Y (male). It is the chromosomes that duplicate during the S phase of the cell cycle, and by the M phase they can be visualized (Figures 4 and 6), in general, as pairs of parallel rods joined at a narrow region called the centromere. During cell division they separate, and one goes to each of the daughter cells. Figure 4 shows the fluorescence distribution of chromosomes isolated from Chinese hamster cells and stained for their DNA content (5,6). Since each chromosome type contains a specific amount of DNA, the flow measurements result in a series of peaks, each representing a single chro mosome type if the resolution is high enough. Peak widths below a 2% coefficient of variation are ohtainable on chromosomes with g of DNA. Major applications of flow measurements on chromosomes include the sorting of purified single chromosome fractions for biochemical and gene mapping (7)studies, and comparison

Chromatograph ic Adsorbents for TLC and CC.

-

PARTICLE SIZE SIGNIFICANCE IN TLC I

I

I * >LOW I OPTIMUM I POOR RESOLUTION SPEED^ SPEED &

I RESOLUTIPN I

I

I

I dODE 710 -

-+----

CONVENTIONAL TLC SILICA

- --.j

\

\ '\

\ I

'\\

-

.-

.-

-"

L .

-v

V L

PARTICLE SIZE-MICRONS

For the best separation efficiency,you can't beat DavidBchromatographic adsorbents for 'TLC and CC. Narrow particle size range is thereason. Each of our ten Davisil grades consists ofparticles that are precisely sized - not too small because that would slow separation, and not too large because that would weaken resolution. With the right Davisil grade, you'll get optimum performance and consistency., .on the

bench, in pilot plant replication, and for commercialization.WhLat's more,Davisil is produced ri ht here in America, by the world leader in si ica gel technology and manufacturing. So you can count on Davison for the quantities you need, and the consistent quality that chromatography demands. For a broad selection of narrow- range adsorbents, talk taus.

f

W. R. Orace MCo., Davison Chemical Div., Dept. APR-CA, P.O. Box2117, Balto. MD21203. (301) 659-9023. Distrlbutor inquiries invited. CIRCLE 89 ON READER SERVICE CARD

14. I!

Figure 5. Dual parameter analysis of human chromosomes Pmpmive view and mnespondingmtour plot 01 me tiuwe~~ence distribution of the smiier human dromo8omes, numbers 9 to 22 and Y

of the distributions obtained from different individuals to understand normal variations and those that may be characteristic of genetic diseases. While most Chinese hamster chromosomes can be resolved, this is not the case with human chromosomes, many of which are too similar in DNA content. However, DNA is not a homogeneous molecule, and some of its more subtle properties can he exploited to improve the discrimination. The genetic information in DNA is encoded in the sequence of its constituent bases, A (adenine), T (thymine), G (guanine), and C (cytosine), which always occur in pairs, A associated with T and G with C. The ratio of A-T to G-C paira in each chromosome is close to 1.5, hut slight differences occur. These differences can he detected by using two dyes, one of which preferentially binds to A-T and the other to G-C. If a single excitation wavelength is used, separation of the signals from the two dyes must rely on differences in the emission spectra. There is usually overlap in the emissions, 80 some light from each of the dyes must he fil tered out to discriminate between the two signals. This is undesirahle, since in many cases the measurement precision is limited by photon statistics. It is better to use dyes with differing excitation bands, which requires two excitation sources that intersect the sample stream at different points. An instrument of this type is used for the following measurements. Figure 5 shows two presentations of the same human chromosome histogram (8).The chromosomes have been double stained with Hoechst 33258 512A

(A-T binding preference), excited with 351 363 nm wavelength light, and chromomycin A-3 (G-C binding preference), which is excited at 458 nm. One illustration of the value of the double stain procedure is the resolution of the peaks corresponding to chromosomes 14,15,16, and 17. They appear at the same chromomycin brightness, so this stain alone would not resolve them. However, they are separated along the Hoechst axis. With Hoechst alone, chromosome peaks 16 and 18 would not he distinguished. These measurements also yield information on the dye-DNA interaction and some details of the base sequences of the chromosomes.

+

The differences in ratios of the two fluorescence signals for each chromosome are greater than the differences in their AT-GC ratios and thus enhance the separation of the peaks (9). Two factors contribute to this enhancement. First it appears that Hoechst binds preferentially to three sequential A-T base pairs while chromomycin has its highest affinity for G-C triplets. The probability of finding such binding sites changes nonlinearly with changes in the average base composition of the UNA, and this en. aggerates the changes in the two fluorescence signals. In addition, the dye concentration on the chromosomes is sufficiently high 80 that some Hoechst (mntinued on page 517 A )

Flgure 6. Slit scanning of chromosomes photographa and colresponding tiow scans of two chromomes: nwmai (rigM) and sbnOrmal (iefl)

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3. MARCH 1982

Precision weighing the better way! ... The Torbal Friction Free Torsion Band Balance with the All American Balance

for accuracy, speed, convenience and reliability. Select a TORBALO precision balance for virtually any purpose, and save a great deal of time and expense when you do it. With TORBAL’s low maintenance balances, you’ll also save money with years of trouble free performance. All TORBAL balances feature rugged torsion-band suspension making them far more reliable and durable than knife-edge balances. With torsion band fulcra there are no knife edges, no metal-to-metal wear, and no bearings to accumulate dust or dirt. And the weighing mechanism is not affected by corrosion or abrasive dusts in the atmosphere either.

TORBAL Principle The fundamental parts of the EA-1 weighing mechanism are shown below. Instead of the pointer and fixed index as used in a conventional balance, the EA-1 uses a sensor called an The E-core, through an amplifier, senses the rotation of the beam and produces a meter deflection proportional to this deflection. All the weighing operations on the EA-1 are performed to return the meter needle to the original point of equilibrium or zero.

TORBAL Model EA-1 Shown is the TORBAL EA-1 analytical balance with electronic speed and the highest standards of accuracy, operating ease and retained sensitivity. Capacity: 160 g Precision: 0.03 mg 8-10 g dial-in-tare, extended tare range to 208 grams.

TUKERL

THE TORSION BALANCE COMPANY

CIRCLE 205 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3. MARCH 1982 * 513,.

7

in your laboratory!

SPECTROPHOTOMETRY

Ne~Everythng You Need invideo AtomicAbsorption from Sargent-Welch's new SP9 Video AA System.

ANEW

STAR

L

-

0 Video

display of complete flame and furnace cookbooks for all elements

0 Transient

I

peak display with post-run rescaling plus Video display of Calibration graph, furnace status and stored programs, basic information, page index, ramp information and much more.

Roador inquiry l o . 14

The amalgam of video technology and research performanceoptics in the PU8800 sets new standards of formation display and operating ease for UV/VIS. Choosing whether to displayresultsin STATUS, GRAPHlCSor DATAscreenmodes isaboutthemost difficult part of the PU8800.The indecisive can copy each screen display in turn on the integral printer. The inexperiencedcan choose from a list of 10 stored program titles (not codes) and follow foolproof prompts. The suspicious can watch the SelfTest program calibratethe wavelength, checkabsorbance and search for optical, mechanicalor electrical faults. The selfish use the keyboard lock to ke pleasure)of the PU8800

I

Circle the appropriate inquiry number for lurther information about Sargent-Welch products. Inquiry No.

W MOO Video UVnlS rpectnphltcmeter SP9 Video Msystem Olhnr UVIIIS inrbunenta PU WOO M rystm IR spoctrophotoneters

81 82 83 84

15

Roador lnqulrf H

The Sargent-Welch PU 8800 bringing a refreshing new light to UVfllS. SARGENT-WELCH SCIENTIFIC COMPANY

7300 NORTH LINDER AVENUE, P.O. BOX 1026, SKOKIE, ILLINOIS 60077 Tel 312-677 I

Anahe,im, Baton Rouge, Birmingham, Cincinnati, Cleveland, Dallas, Denver, Detroit, Skokie, Springfield NJ.Washington D.C. Inquiries outside USA contact Pye Unicam Ltd, York Street, Cambridge, England CB1 2PX.

0600

Whatman Glass Microfibre Filters: Excellence. Whatman Glass Microfibre Filters represent near oerfection in a laboratorv deDth filter. Thev are hiahlv efficient, retaining particles as small as 07 .m ; in liqid;. They're fast; flow rates are high and remain high because loading capacity is greater than cellulosic or membrane filters. And pure; binder-free and made entirely from the highest grades of borosilicate glass using only demineralized water. They are virtually unaffected by humidity, will not age chemically, and will store indefinitely under normal conditions. L

.

Whatman Glass Microfibre Filters are resistant to chemical attack (except by hydrofluoric acid and strong alkalies) and are able to retain their physical integrity in conditions which would seriously affect the performance of other filter media. Thev can be used to temDeratures of 500°C. A full range of'grades is available,'providing

the optimum combination of retention efficiency, flow rate and loadina CaDacitv. I

.

Applications are virtually limitless including, as examples, liquid clarification, air and water pollution analyses, precipitation of proteins and nucleic acids, cell harvesting, RIA. . . even replacement of cellulose filters in many applications where time and higher performance are important. Whatman Glass Microfibre Filters. From the people who first introduced them. Excellent in every way.

-

At laboratory supply dealers everywhere,

Whatman Laboratory Products InC. Whatman Paper Division 9 Bridewell Place, Clifton, New Jersey 07014 Tel: (201) 773-5800

MWhatmun See us at FASEB Exhibit Booth No. J201522. CIRCLE 239 ON READER SERVICE CARD

PAPER DIVISION

clean-up technique is simpler and faster. Rheodyneslechnical Note 2 tells how to prepare samples with greater speed and lower cost than purification steps using liquid/liquid partition column chromatography or commerc i al se paratio n c a t r i d ges The rapid clean-up technique uses Rheodynes Model 7125 Syringe Loading Sample lnjecbr with a short colurnn iri the sainple loop It allows complex sampIEs to be injected directly into the chromatographic system This technique can be used to Eliminate interferences from complex matrices Concentrate samples to improve detection limits Shorten analysis time by trapping late-eluting Jeaks

Send for Tech Note #2 The simplified procedures are fully described in this well-illustrated 4-page technical note IContact Rheodyne Inc FO Box 996 Cotati Calif 94928 U !; 4 Phone (707)664-905Ci

and chromomycin molecules are close enough for the Hoechst excited state to transfer its energy to the chromomycin before it can be radiated. Variations in energy transfer efficiency among chromosome types also enhance observed fluorescence differences. The measurements just discussed dealt with properties characteristic of the whole chromosome. Other techniques allow measurement of some aspects of chromosome morphology. Using a narrowly focused excitation beam, the distribution of dye along a chromosome can be scanned as it flows past and the pulse profile analyzed ( 1 0 , l l ) .Measurements with scanning microscope systems have shown that there is less DNA content per unit length at the centromere than at other points. Figure 6 shows the profiles of two chromosomes obtained in flow, and the centromeric dips are clearly visible. The time for the scan is about 1 ps, approximately 50 points 20 ns apart being sampled for a typical 10-pm-long chromosome. The centromere can be resolved in chromosomes as small as 3.5 pm in length. On the right is a normal chromosome with a single centromere. On the left is a chromosome from a cell exposed to X-rays. An error was made in rejoining breaks caused by the radiation, and the result is a dicentric chromosome with two centromeres. Chromosome aberrations, like the dicentrics, can be used as Diological dosimeters, since their frequency of occurrence depends on the dose received. This is particularly useful in cases of inadvertent human exposure, where no other dosimeters are present. Searching for aberrations visually is time-consuming, since they are rare for low exposures. The high analysis rates possible in flow may find application in speeding this process. Conclusion Among the additional applications of flow cytometry are blood analysis, bacterial classification, tumor classification and diagnosis, monitoring of cancer therapy, and immunological studies. All of these exploit the high speed and accuracy of the technique, which allow collection of information on a sufficient number of individual particles to describe the distribution of characteristics in a population with good statistical precision. Many require the ability to sort subpopulations for further study or use. Applications outside of biology are waiting to be explored. References (1) Herzenberg, L,. A.; Sweet, R. G.; Herzenberg, L. A. Scz Am 1976,234 (3),

-

t

c x

\

I

r i c \ u;

108. \

(continued on page 519 A )

CIRCLE 186 ON njI4DER SERVICE CARD

Journal of Automatic Chemistry is a prime source of information on all facets of the use and practice of automation in chemistry., containing not only in-depth discussion of instrumentation but also articles dealing with such topics as management economics and the general philosophy of automation. A lively news section completes a highly informative quarterly which is essential reading for those seeking to obtain the maximum benefit from automation in chemistry.

To quote from a recent review in Nature: ‘The standard of papers is usually good, even excellent, and the “virtual” A 4 format allows good use to be made of illustrations. The generally high quality of the articles read in the sample issues suggests that this journal has so far done an excellent job and should be scanned on a regular basis by all those who are concerned with automation.’ Subscriptions to Volume 4 (1982) cost g30.00 in the UK and SF16 1.OO elsewhere, payable in the local currency equivalent of the Swiss Franc rate. This includes the cost of postage and packing. Further information on the journal, sample copies, and details of a highly attractive advertising package, are available from the publisher.

4 John Street, London WClN 2ET, UK Tel: 01-405 223719 Telex: 858540 ~~

CIRCLE 206 ON READER SERVICE CARD

ANALYTICAL CHEMIST1

VOL. 54, NO. 3, MARCH 1982

e

517A

Winter 1982 ACS

NEWTITLES Nuclear and Chemical Dating Techniques

Roger Adams: Scientist and Statesman

This 24-chapter book reviews the latest developments and state of the art of physicochemical dating methods together with biogeochemical and archaeological applications. Chapters are grou ed into three sections: cosmochemical and geoclemical applications; natural archives: interpreting the isotopic and chemical record: and chemical evolution extinction and archaeology. ACS Symposium S h e s 176, 5’16 pages, 1982, Cloth, ISBN 0-8412-0669-4. US & Canada $48.95 Export $58.95

A comprehensive biographical sketch of Roger Adams’s, whose work was important in the development of American chemistry and chemical education. Adam’s early years, education, and career achievements in academia, industry, research, and government are described. His contributions to Illinois chemistry in particular, and the education of chemists are expounded. His services to the American Chemical Society and other professional organizations are also discussed. 250 pages, 1981, Cloth, ISBN 0-8412-0598-1. US & Canada $13.95 Export

Inorganic Reactions in Organized Media

$16.95

This 14-chapter volume is probably the largest single collection of papers dealing k i t h reactions involving micelles and microemulsions. Topics include the transfer from the monomeric staie to the inverse micellar structure, cage and magnetic isotope effects of micellization, light-induced electron-transport processes in micellar systems, photoprocesses in synthetic vesicles, and hotoinduced reactions at the surface of colloifal oxides. ACS Symposium Series 177, 254 pages, 1982, Cloth, ISBN 0-8412-0670-8. US & Canada $31.95 Export $38.95

Biochemical Responses Induced by Herbicides

Supercomputers in Chemistry This 15-chapter volume examines the applications of supercomputers to several areas of chemistry, including quantum chemistry, chemical kinetics, molecular dynamics, polymer chemistry, and organic synthesis. Detailed discussions of the adaptation of auantum chemistrv codes for the Cray--l vector computer are pre- . sented. The large-scale computational reauirements of quantGm chemistry, as well as &ose of chemical dynamics and quantum optics, are also dealt with. ACS Symposium Series 173, 278 pages, 1981, Cloth, ISBN 0-8412-0666-X. US & Canada $31.95 Export $38.95

Metal Bonding and Interactions in High Temperature Systems The latest research on the following four key areas: interactions of alkali metal atoms and dimer bonding, bonding in small metallic clusters, bonding and interactions of the alkali hydrides and halides and mtal atom-water interactions, and applications of alkali metal and alkali compound va ors. These groupings are timely, and their corngination offers an intertwining of basic and applied research efforts. ACS Symposium Series 179, 1982, Cloth, ISBN 0-8412-0689-9. US & Canada $54.95 Export

This 13-chapter book presents state-of-the-art reports on mechanisms and modes of action throu h which herbicides affect plant growth at the su%cellular and molecular level. Several chapters concentrate on interferences with photoinduced electron transport by herbicides. ACS Symposium Series 181, 277 pages, 1982, Cloth, ISBN 0-8412-0699-6. US & Canada $35.95 Export $43.95

Polymer Science Overview: A Tribute To Herman F. Mark The story of Herman Mark’s life is told in seven chapters, four being personal reminiscences by long-time associates. Three chapters cover the histor of polymer science. Other topics covered includk polymer education, the interfacial macromolecule, polymers in the future, enzyme engineering, olymeric monolayers and liposomes, free-radicaypolymerization, and current uses for block copolymers. ACS Symposium Series 175, 368 pages, 1981, Cloth, ISBN 0-8412-0668-6. US & Canada $35.95 Export $43.95

Short-Lived Radionuclides in Chemistry and Biology These 28 chapters are organized into five sections

- two dealing with chemical reaction studies,

two focusing on radiobiochemistry, and one addressing microbiological and environmental techni ues. The emphasis is on experiments in w h i c h k e investigator must produce the tracer on site when it is to be used. Advances in Chemistry Series 197, 547 pages, 1981, Cloth, ISBN 0-8412-0603-1. US & Canada $72.95 Export $87.95

An overview of both the fundamental roperties and the technological aspects is proviied in this comprehensive, 19-chapter volume. Five chapters detail the thermodynamic, mechanical, and transport properties of perfluorosulfonated polymers. Spectroscopic, structural, and modeling studies are presented i n the next group of seven cha ters A plications of thesemembranes are descrged in &e final, six-chapter section. ACS Symposium Series 180, 1982, Cloth, ISBN 0-8412-0698-8. US & Canada $48.95 Export $58.95

Catalytic Aspects of Metal Phosphine Complexes Twenty-four chapters focus on P-31 NMR and the nature of the metal-phosphorus bond, the chemistry of bulk phosphine ligands, polydentate phosphines, txe reactivitv of transition-metal phosphorus compounds, and asymmetric synthesis This book covers a wide varietv of topics ranging from platinum-tin homogeneous h i drogenation catalysts and Cobaloximes containing phosphorus-donor ligands to binuclear orrranoulatinum comulexes as catalvtic intermedi a k a i d recent advances in polyp6osphine synthesis. Advances in Chemistry Series 196, 1982, Cloth, ISBN 0-8412-0601-5. US & Canada $69.95 Export $83.95

518A

This 29-chapter volume focuses on the current state of the art of Massbauer spectroscopy and its diverse applications. Topics include the use of molecular orbital calculations to study chemical bonding i n halogen-containing compounds and the electronic properties of intermetallic, inorganic, and organometallic compounds; conversion electron Mb’ssbauer spectroscopy; phase analvsis: and environmental. analvtical. and biological applications. Advances’in Chemistry Series 194, 642 ages 1981 Cloth ISBN 0-8412-0593-0. I%& banad; $69.95 Export $83.95

Chemistry of Asphaltenes This 14-chapter volume constitutes a comprehensive treatment of the chemistry of asphaltenes. Investigations with several t pes of asphaltenes, including those from petroikum, coal liquids, and tar sand bitumens, are reported. Oxidative, Lewis acid assisted, and hydfopyrolitic degradation: chromatographic and acid-base separation; and mass spectrometric and infrared analysis of asphaltenes are detailed. Advances in Chemistry Series 195 272 ages, 1981, Cloth ISBN 0-8412-0592-2. U S & 8anada $39.95 Expdrt $47.95

N-Nitroso Compounds This volume presents the current status of nitrosamine research. Twenty-six chapters are organized into three sections: chemistry and metabolism, chemistry of formation and blocking, and analysis and occurrence. Specific instances of the reduction of nitrosamines in contaminated products are presented as well as basic research. ACS Symposium Series 174,416 pages, 1981, Cloth, ISBN 0-8412-0667-8. US & Canada $39.95 Export $47.95

Catalysis Under Transient Conditions Thirteen innovative manuscripts that greatly augment the literature in the field. Chapters present specific details on three-way automotive catalysts, carbon monoxide oxidation and chemisorption, nitric oxide reduction, nitrous oxide decomposition, vapor phase electrocatalytic ethylene epoxidation, and methanol oxidation. ACS Symposium Series 178, 308 pages, 1982, Cloth, ISBN 0-8412-0688-0. US & Canada $24.95 Export $29.95

CALL TOLL FREE 800-424-6747 Charge to your VISA or Master Charge Card

$65.95

Perfluorinated Ionomer Membranes

Mb’ssbauer Spectroscopy and Its Chemical Applications

r l l l l l l l l l l l l l l l l l l l l l l l -

I American Chemical Society, 1155 Sixteenth Street, N.W. Washington, D.C. 20036 I Title I I

i i

0 Payment Enclosed [Free postage and handling)

California residents add 6% sales tax

0 MasterCard’Interbank Code:

I Account Number: II Signature 0 Purchase Order Enclosed I

Name

ANALYTICAL CHEMISTRY, VOL. 54, NO. 3, MARCH 1982

~

Total Payment -~

I

P I’IS.4

Price

Expiration Date:

P.O. Number

(2) norm, Y . K.; Wheeleas, L.L. Science 1977,198,149.

(3) AmdMovan, D. J.; Jonn, T. M. In “Annual Review of Biophysics and Bioengineering”; Mullins, L. J.; Hagins, W. A,; Newton, C.; Weber, G.,Eds.; Annual Review Inc.: Palo Alto, Calif,, 1978;

Vnl . ~ .7... (4) Melamed, M. R.; Mullaney, P. F.;Men-

FREE

‘fi

Products for Analysis 1982

delsahn, M. L., Eds. “Flow Cytornetry and S o r t i n p John Wiley and Sons: New York, Chic eater, Brishane, and Toronto, 1979. (5) Gray, J. W.; Carrano, A. V.; Steinmetz, L. L.; Van Dilla, M. A,; Moore, D. H. 11; Mayall, B. H.; Mendelsohn, M. L. Proc. Nat. Acod. Sei. USA 1975.72 (4). 1231. (6) Sillar, R.; Young, B. J. Histochem. Cy-

tochem. 1981,29,74.

(7) Lebo, R. V.; Carrano, A. V.; B u r k h a h

Schultz,K.;Dozy.A. M.;Yu,L.-C.;Kan, Y. W. Proc. Not. Acad. Sei. USA 1979,

Send today for yourcopy

76 (11). 5804. (8) Gray, J. W.;Langlois, R.; Carrano, A.

V.; Burkhart-Schultz. K.; Van Dilla, M. A. Chromosoma 1979,73,9. (9) Langlois, R.; Carrano, A. V.; Gray, J. W.; Van Dilla, M. A. Chromosoma 1980,77,229. (10) Gray, J. W.; Peters, D.;MerriII,

J.T.; Martin, R.; Van Dilla, M. A. J. Histochem. Cytochem. 1979.27 (I),441.

(11) Cram, L. S.;ArndMovin, D. J.; Grim-

wade, B.; Jovin, T. M. J. Histoehem. Cv-

toehem. 1979,27,445.

Our most complete catalog yet! 150 pages of comprehensive information on our: test kits, analytical and process instruments, lab equipment, glassware, reagents, agricultural products and literature.

Work performed under the auspice of the US. Department of Enegy by the Lawrence Livermope National Laboratoryunder contract number W-7405-ENG-48.

J

I

Dan Pinkel received his BA degree in physics from the University of Michigan, and MS and PhD degrees in solid-state physics from the Universit y of California at San Diego. After a postdoc at UCLA in medical physics, where he first came into contact with flow cytometry, he went to his present position in the Cytophysics Section of the Biomedical Sciences Division at Lawrence Livermore National Laboratory. Current research interests include instrumentation development and high-precision measurements on mammalian sperm.

l r

Laboratory, Process, and Flue gas operations

0 Fast nert a ess IR detec- 0 A new technology seals tor rystem ena winaows to tne detec0 Eoc, c-,rent orive tor tor ce A ow5 htgher

range can De changed m minutes ~y exchanging sample cell and oetectoi cnopper d sc. no airect detector temperatdres ana 0 D gota or ana og o spay coop ng permits corroalw~gas of s m a 0 Chopper frequency 250 appllcatlona. Hz. 0 Thermoslaling of defector For detalled informa0 Sensilivilv stabilization. not maulred. New micro tlon. write or ~. . call. Takmari .. ... . . Speciallybrogrammed flow detector system very P.O.’Eox 37202, Cincinnati, tolerant lo ambient temOH 45222/(800) 543-44611 chopper produces pulses lor artomat c caloDrat on at perat-re cnanges Less InOho cal’co ecr 1513) tne rate of 3 600 times per than 0 1’- of range per K 761-0633 Te ex No mmte. 0 Measded component ana 21-4221 ~

~~

ANALYTICAL CHEMISTRY, VOL. 54. NO. 3. MARCH 1982

-

519A