Fluorometric sniff-cam (gas-imaging system) utilizing alcohol dehydrogenase for imaging concentration distribution of acetaldehyde in breath and transdermal vapor after drinking Kenta Iitani, Toshiyuki Sato, Munire Naisierding, Yuuki Hayakawa, Koji Toma, Takahiro Arakawa, and Kohji Mitsubayashi Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.7b04474 • Publication Date (Web): 23 Jan 2018 Downloaded from http://pubs.acs.org on January 24, 2018
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Fluorometric sniff-cam (gas-imaging system) utilizing alcohol dehydrogenase for imaging
2
concentration distribution of acetaldehyde in breath and transdermal vapor after drinking
3 4
Kenta Iitania, Toshiyuki Satoa, Munire Naisierdinga, Yuuki Hayakawaa, Koji Tomab, Takahiro
5
Arakawab and Kohji Mitsubayashia,b,*
6 7
a
8
5-45 Yushima, Bunkyo-ku, Tokyo 113-8510, Japan
9
b
Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, 1-
Department of Biomedical Devices and Instrumentation, Institute of Biomaterials and
10
Bioengineering, Tokyo Medical and Dental University,2-3-10 Kanda-Surugadai, Chiyoda-ku,
11
Tokyo 101-0062, Japan
12 13
* Corresponding author. Tel.: +81 3 5280 8091, Fax: +81 3 5280 8094
14
E-mail:
[email protected] 15 16 17
1 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 2 of 30
18
Abstract
19
Understanding concentration distributions, release sites and release dynamics of VOCs from
20
the human is expected to lead to methods for noninvasive disease screening and assessment of
21
metabolisms. In this study, we developed a visualization system (sniff-cam) that enabled to
22
identify a spatiotemporal change of gaseous acetaldehyde (AcH) in real-time. AcH sniff-cam
23
was composed of a camera, a UV-LED array sheet, and an alcohol dehydrogenase (ADH)-
24
immobilized mesh. A reverse reaction of ADH was employed for detection of gaseous AcH
25
where a relationship between fluorescence intensity from nicotinamide adenine dinucleotide
26
and the concentration of AcH was inverse proportion; thus, the concentration distribution of
27
AcH was measured by detecting the fluorescence decrease. Moreover, image differentiation
28
method that calculated a fluorescence change rate was employed to visualize a real-time
29
change in the concentration distribution of AcH. A dynamic range of the sniff-cam was 0.1–10
30
ppm that encompassed breath AcH concentrations after drinking. Finally, the sniff-cam
31
achieved to visualize concentration distribution of AcH in breath and skin gas. A clear
32
difference of breath AcH concentration was observed between aldehyde dehydrogenase type 2
33
active and inactive subjects, which was attributed to metabolic capacities of AcH. AcH in skin
34
gas showed a similar time course of AcH concentration to the breath, and a variety of release
35
concentration distribution. Using different NADH-dependent dehydrogenases in the sniff-cam
36
could lead to a versatile method for noninvasive disease screening by acquiring
37
spatiotemporal information of various VOCs in breath or skin gas.
38 human volatilome,
39
Keywords:
40
dehydrogenase, image differential
fluorescence, spatiotemporal, acetaldehyde, alcohol
41 42
Introduction
2 ACS Paragon Plus Environment
Page 3 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
43
In the conventional and current diagnosis of diseases, the blood test has usually been 1,2
. At the same time, there has been growing interest in non-invasive
44
commonly utilized
45
diagnosis methods aiming at exhaled volatilome such as volatile organic compounds (VOCs)
46
in breath or skin gas as the alternative or auxiliary in next-generation medicines
47
VOCs are originally contained in blood as a result of specific diseases and metabolism, and
48
then they are emanated directly from skin or breath after the exchange in the lungs 7,8.
3–6
. These
49
Acetaldehyde (AcH), an intermediate of alcohol metabolism, is produced by ADH-
50
mediated reaction in the liver after ingesting alcohol 9, and it has toxicity to DNA 10,11. For
51
example, it was reported that AcH causes the mutational DNA damage in hematopoietic stem
52
cells. 12 In this study, we used AcH as a model analyte since the contained in blood is released
53
through breathing with a partition coefficient of breath : blood = 1 : 109
54
Skin gas is constantly released and can be easily collected without a mental and physical
55
burden; therefore, skin gas is considered to be more suitable for monitoring of VOCs than
56
exhaled breath. However, measurement of VOCs in skin gas is challenging since their
57
concentrations are usually much lower than the breath, e.g., ppb or sub-ppb level 15; a highly
58
sensitive system is strongly required for the detection. Conventionally, skin gas sample is
59
collected by a hermetically container such as a sample bag and measured by analytical
60
devices such as a gas chromatograph (GC) and a mass spectrometer (MS). For example,
61
Tsuda et al. and Turner et al. measured ethanol, acetone and etc. in skin gas that were released
62
from a forearm, hand, and finger and collected by a sample bag
63
ionization detector or selected ion flow tube MS. Mochalski et al. developed a room-type skin
64
gas collection system which a subject was in, and collected only emanated skin gas by
65
discharging exhaled breath out of the room; the VOCs contained in skin gas was measured
66
using a proton transfer reaction MS
67
drawbacks including complicated processes of measurement and requirement of highly skilled
16,17
13
and skin gas
14
.
using GC and flame
18
. Despite the advantages of these methods, some
3 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 4 of 30
68
personnel made them unsuitable for simple disease screening and metabolism evaluation.
69
Besides, it is difficult to investigate release sites and dynamics in detail because these
70
methods measured the total amount of released VOCs and are unable to focus on a spot. It is
71
known that VOCs’ concentration in exhaled breath and skin gas shows temporal fluctuation
72
19–21
73
important to deepen the understanding of a correlation between skin gas and physiological
74
conditions: (i) Direct emanation from blood through the dermis and the epidermis, (ii)
75
vaporization of sweat secreted from glands in a skin and (iii) production by metabolism of
76
resident bacteria on a skin surface 22–25.
, and elucidating following potential release pathways for each VOC is considered to be
77
Up to now, we have introduced a fiber optic gas sensor “bio-sniffer” that showed high
78
sensitivity and selectivity due to the substrate specificity of an enzyme, and it allowed
79
continuous measurement of VOCs 26–30. Also, applying the bio-sniffer’s technique to imaging,
80
a biofluorometric gas imaging system (sniff-cam) was developed
81
AcH sniff-cam employing a reverse reaction of ADH
82
accurately display a temporal change of AcH concentration by measuring the fluorescence
83
intensity of β-nicotinamide adenine dinucleotide (NADH) because the fluorescence
84
reached a plateau after initial decrease and did not return to a baseline although the gas
85
injection was stopped. This was attributed to measurement principle in which consumed
86
NADH has not filled up again throughout the measurement. In this study, we further advanced
87
the AcH sniff-cam to improve the responses and to enable visualize a spatiotemporal change
88
of AcH concentration in real-time. The response was improved by employing an image
89
differential method that utilized a slope of fluorescence intensity change instead. Since the
90
slope kept changing while the fluorescence intensity increased and approached plateau, the
91
slope output showed a clear peak and returned to a baseline when the gas injection was
31
. Recently, we reported
32
. However, it was difficult to
4 ACS Paragon Plus Environment
Page 5 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
92
stopped. Afterwards, the advanced AcH sniff-cam was applied to real-time visualization of
93
spatiotemporal change of AcH contained in exhaled breath and skin gas after drinking.
94 95
Experimental
96
Materials and reagents
97
Tris-HCl buffer solution (Tris, pH6.5, 0.1 M) that was used in all of the
98
experiment was made by 2-amino-2-hydroxymethyl-1,3-propanediol (99.9%, Cat. No.
99
013-16385, Wako), hydrochloric acid (35%, Cat. No. 083-03485, Wako) and ultra-pure
100
water that was generated by a Mill-Q purification system. Reagents for fabrication of
101
an ADH-immobilized mesh were ADH (EC 1.1.1.1, 369 units/mg solid, Cat. No.
102
A7011, from Saccharomyces cerevisiae, Sigma-Aldrich, USA), phosphate buffer (PB,
103
pH6.5, 0.1 M), bovine serum albumin (BSA, Cat. No. 306-13383, Wako, Japan) and
104
glutaraldehyde (GA, 25%, Cat. No. 079-00533, Wako Pure Chemical Wako, Japan). A
105
cotton mesh (100% cotton, the thickness of 1 mm, the interval of 1 mm) that was used
106
as a substrate for enzyme immobilization was purchased from Ohki Healthcare
107
Holdings (Japan). Reduced form of β-nicotinamide adenine dinucleotide (NADH, Cat.
108
No. 44326000) was purchased from Oriental Yeast (Japan). Standard gaseous AcH was
109
generated using a permeation tube (Cat. No. P-92-1, Gastec, Japan) with a standard gas
110
generator (Cat. No. PD-1B-2, Gastec, Japan). For selectivity assessment, some VOCs
111
including ethanol (99.5 %, Cat. No. 14033-00. Wako), methanol (99.8%, Cat. No. 21926-95,
112
nacalai tesque), 2-propanol (99.7, Cat. No. 166-04836, Wako), acetone (99.5%, Cat. No.
113
016-00346, Wako) were generated by a diffusion tube (No.3200, Gastec, Japan), and the other
114
such as methyl mercaptan that was done by a permeation tube (Cat. No. P-71-5, Gastec,
115
Japan). An alcohol beverage (concentration of 20%, distilled liquor) for breath and skin gas
116
experiments was purchased at a local store.
5 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 30
117 118
Construction of the AcH sniff-cam
119
Figure 1a shows a visualization principle of gaseous AcH distribution employed
120
in an AcH sniff-cam. AcH was detected as a fluorescence decrease from NADH
121
consumption which occurs through an ADH-mediated reverse reaction. In our previous
122
work, we investigated the reduction/oxidation reaction of ADH can be controlled by
123
pH condition. Also, unwanted interfere between forward and reverse reaction could not
124
occur under optimal conditions.33 NADH has an autofluorescence property that
125
exhibits absorption of excitation light at 340 nm wavelength and emission of
126
fluorescence at 490 nm wavelength. Since NADH and AcH concentrations are
127
correlated, AcH concentration can be determined by detecting fluorescence decrease of
128
NADH. The AcH sniff-cam was composed of a UV-LED array (λ of 340 nm, 9×9
129
array, 340X081SFN, Dowa, Japan), an ADH-immobilized mesh and a high-sensitive
130
camera (HEED-HARP, Pioneer, Japan) (Figure 1b). The UV-LED array and the
131
camera were aligned in the same optical, and 5 mm and 110 mm apart from the ADH-
132
immobilized mesh, respectively. The UV-LED array was connected to a power source
133
(KX-100H, Takasago, Japan) that supplied constant current 540 mA to the LED, and
134
enabled to excite NADH uniformly in the ADH-immobilized mesh. The gain
135
parameter of the high-sensitive camera was fixed at 15 in all of the experiments, and
136
video signal that was conforming to National Television System Committee (NTSC)
137
was out from composite connector of a camera controller. An analog video signal was
138
converted to a digital video format by a video encoder (Intensity shuttle USB3.0, Black
139
Magic Design, Australia). Two bandpass filters placed in front of the UV-LED array
140
(BPFex, λ of 340 ± 42.5 nm, Edmund Optics, USA) and the high-sensitive camera
141
(BPFfl, λ of 490 ± 10 nm, Asahi Spectra, Japan) excluded unwanted light to improve a 6 ACS Paragon Plus Environment
Page 7 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
142
signal to noise ratio. The ADH-immobilized mesh was fabricated by the following
143
process. First, 0.65 mg of ADH and 1.5 mg of BSA were dissolved in PB (200 µL, 0.1
144
M at pH 6.5). Then, 200 µL of the mixture was dropped and spread uniformly on a 20
145
× 20 mm2 cotton mesh substrate, and then the mesh was left in a refrigerator for 60
146
minutes. Subsequently, 10-fold diluted GA solution (32 µL, 2.5 v/v% in PB, 0.1 M at
147
pH 7.0) was dropped onto the mesh for cross-linking and left in the refrigerator for 90
148
minutes. Finally, the ADH-immobilized mesh was rinsed with 300 µL of Tris and then
149
used immediately. Standard gaseous AcH was generated by a standard gas generator
150
with a permeation tube. Clean carrier air that was compressed and filtered by a
151
compressor and air filters was supplied to the standard gas generator through
152
polytetrafluoroethylene tubes (F-8006-017, the outer diameter of 6 mm, the inner
153
diameter of 4 mm, FLON industry, Japan) with excellent chemical and adsorption
154
resistance. Flow rates of generated standard gaseous AcH and clean carrier air were
155
controlled at 40 mL/min by mass flow controllers (MFC, RK1250, KOFLOC, Japan),
156
and gas was introduced to the AcH-immobilized mesh. A glass gas outlet was placed
157
on a backside of the ADH-immobilized mesh. Incidentally, it is reported that the
158
environmental temperature affects enzyme activity and fluorescence intensity of
159
NADH34. Thus, environmental temperature was kept at 25 ± 1.5 ºC of all experiment
160
using an air conditioner. Thus, the temperature did not affect the enzyme activity and
161
fluorescence intensity of NADH.
162 163
(Figure 1 comes here)
164 165
Differential method for visualization of spatiotemporal changes of AcH
7 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 8 of 30
166
In the previous report, visualization of gaseous AcH spatial distribution was achieved
167
using AcH sniff-cam that employed a reverse reaction of ADH. However, there was a problem
168
in temporal resolution. After a decrease of fluorescence intensity on the ADH-immobilized
169
mesh, the intensity reached a plateau and did not recover to the initial value even though
170
injection of AcH was stopped. This was because there was no NADH supply once it was
171
consumed. This problem made it difficult for previous AcH sniff-cam to be used for the study
172
of gaseous AcH release dynamics. In order to solve this problem, an image differential
173
method that utilized slope of fluorescence intensity change instead of the intensity itself was
174
adopted. Application of the differential method to a video taken by the AcH sniff-cam was as
175
follows. First, an original video that showed fluorescence decrease as a result of NADH
176
consumption was converted to one in which florescence increased by subtracting a subsequent
177
frame from the first frame of the video. This process is described by the following equation: = − ( > 0),
Eq.1
178
wherei i is a frame index number; FI and FD are fluorescence increasing and decreasing
179
images, respectively. Next, a simple moving average (SMA) filter described in equation 2 was
180
used to remove high-frequency noise from FI that was caused by an image sensor of the
181
camera. 1 =
()
( ≥ ),
Eq.2
182
where n is the number of the frames in the SMA filter; FIk is a kth frame in the fluorescence
183
increasing video; SFI is smoothed fluorescence increasing images. Then, a differential video
184
that was composed of slopes of fluorescence intensity changes was created by differential
185
method described in equation 3. =
− (×∆)) ∆ = ( ∆ = 2, !"# = 30, > 60), ∆ ∆
Eq.3
8 ACS Paragon Plus Environment
Page 9 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
186
where i is a frame index number; fps is frame per s that was of 30; ∆t is a time step that was of
187
2 sec; SFI is a smoothed fluorescence increasing image; DI is a differential image.
188
Finally, high-frequency noise remaining in the DI was removed by the SMA filter. The
189
SMA filter had a trade-off relationship between noise reduction performance and response
190
time. When the n in the SMA filter increased, noises were decreasing but response time
191
became longer. Thus, an influence of the n to noise and response time was investigated with
192
various conditions (n = 0, 10, 30, 100, 300, and 600 frames) of the SMA filter. In the
193
experiment, 10 ppm of gaseous AcH was applied to the ADH-immobilized mesh wetted by
194
NADH solution (500 µM, in 300 µL of Tris at pH 6.5, 0.1 M). After creating DI, that was
195
smoothed by various conditions of SMA filters. Then, noise and response time at each
196
condition were evaluated.
197
In order to confirm the applicability of the AcH sniff-cam to the measurement of
198
breath and skin gas, sensitivity and selectivity of the AcH sniff-cam were investigated. Each
199
experiment was performed using 500 µM NADH dissolved in 0.1 M Tris at pH 6.5. For
200
evaluation of the sensitivity of the AcH sniff-cam, various concentrations of gaseous AcH
201
generated by the standard gas generator were applied to the ADH-immobilized mesh at a flow
202
rate of 40 mL/min. In the experiment, initially clean carrier air was applied for 20 s, and then
203
gaseous AcH was applied for 20 s. Finally, clean carrier air was applied again for 80 s. For
204
investigation of selectivity, typical VOCs produced by the standard gas generator were
205
applied to the AcH sniff-cam in the same procedure with the above-mentioned sensitivity
206
study, and concentrations of the gas samples were set to be those after drinking: AcH (5 ppm),
207
ethanol (100 ppm), methanol (0.1 ppm), 2-propanol (0.1 ppm), acetone (0.6 ppm), methyl
208
mercaptan (0.007 ppm), and mixture of AcH (5 ppm) and ethanol (100 ppm) were used 35,36.
9 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 10 of 30
209
Note that all of the numerical data of mean intensity and differential value were
210
calculated at the region of interest (ROI, 80 pixels × 80 pixels) that was set at the center of
211
frames in each video.
212 213
Visualization of spatiotemporal change of the AcH concentration in breath
214
Visualization of AcH in exhaled breath after drinking was taken place with subjects
215
with a different phenotype of aldehyde dehydrogenase type-2 (ALDH2). A subject with lower
216
activity of ALDH2 (ALDH2[-]) cannot catalyze AcH efficiently. In general, blood AcH
217
concentration of ALDH2[-] subject becomes higher after drinking than that of the subject with
218
high activity of ALDH2 (ALDH2[+]), thus breath AcH concentration is also expected to be
219
higher than the ALDH2[+] 37–39. These breath and skin gas experiments were conducted based
220
on the authorization of the Human Investigations Committee of Institute of Biomaterials and
221
Bioengineering, Tokyo Medical and Dental University (authorization code of 2012-6) that
222
acted up to Declaration of Helsinki. The subjects had been prohibited to smoke, take medicine
223
and alcohols since 72 hours before the experiment. The most important of the experiment with
224
a drinking was unified concentration of blood ethanol in all subject. Concentration of ethanol
225
in blood fluctuates by volume of blood that almost correlates to body weight. Therefore, each
226
subject who had been fasting for 4 hours administrated alcoholic beverage with an ethanol
227
concentration of 0.4 g/kg body weight within 15 minutes by refer to previous research.
228
the experiment, exhaled breath was directory applied to the ADH-immobilized mesh that was
229
soaked with 500 µM NADH solution (in 300 µL, 0.1 M of Tris at pH 6.5) through a breath
230
flow controller 41 that made a flow rate of the breath constant (40 mL/min). The measurement
231
was taken place at -15, 0, 15, 30, 45, 60, 90, 120, and 180 minutes after drinking. In order to
232
validate results by the AcH sniff-cam, the identical breath sample collected in a sample bag
233
(Cat. No. 2-0081-03, As One, Japan) was measured simultaneously by a gas detection tube as
40
In
10 ACS Paragon Plus Environment
Page 11 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
234
a standard comparison. The concentration of breath AcH measured by the AcH sniff-cam was
235
calculated using a calibration curve, and its spatial distribution was visualized using the image
236
differential method.
237 238
Visualization of distribution of the AcH released from palm skin
239
For skin gas visualization, a palm was chosen due to convenience for the
240
measurement. Taking an average size of palm into consideration, a larger (90 × 90 mm2)
241
ADH-immobilized mesh than that used in breath AcH visualization was prepared. Preparation
242
of this larger mesh was the same as the one for the breath experiment except for a point where
243
a total amount of chemicals was increased proportionally to the size. Briefly, a mixture of
244
13.17 mg of ADH, 20.0 mg of BSA and 1800 µL of PB was dropped and spread uniformly
245
onto the cotton mesh, then left for 60 minutes in a refrigerator. Afterwards, ADH was cross-
246
linked by 648 µL of GA in the refrigerator for 90 minutes. The mesh was then rinsed by 5000
247
µL of Tris at pH 6.5 and wetted by 500 µM NADH solution (in 0.1 M of Tris at pH 6.5, 6075
248
µL), followed by being placed in front of the camera. Figure 1d shows a skin gas injector with
249
a thickness of 3 mm was composed of a polymethylmethacrylate (PMMA) frame and a
250
honeycomb spacer for adequate imaging of spatiotemporal change of skin gas. Because it was
251
used to keep the constant distance between the ADH-immobilized mesh and the palm skin
252
contour surface. Imaging of transdermal AcH after drinking was performed by the following
253
steps. First, a background image without skin gas was recorded for 60 s, and then skin gas
254
was applied to the ADH-immobilized mesh by attaching a subject’s palm on the spacer for 20
255
s. Afterwards, resultant fluorescence on the mesh was recorded by the camera for 60 s. An
256
image of transdermal AcH distribution was made by subtracting the background image from
257
the skin gas. The measurement was taken place at -15, 30, 60, and 90 minutes after drinking,
11 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 12 of 30
258
and alcohol administration conditions were same with the above-mentioned breath
259
experiment.
260 261
Results and discussion
262
Responsivity, quantitativity, and selectivity of the AcH sniff-cam
263
Figure S-1a and S-1b show time courses of the mean intensity of the ROI and
264
differential value. An SMA filter was applied to the intensity of the ROI at different frame
265
numbers (n = 0, 4, 8, 16 and 32), and the differential values were obtained from the filtered
266
intensity. The result indicated that the response time was almost the same among the frame
267
number of 0–32. The relationship between noise and the number of frames in the SMA filter
268
is shown in Figure S-1c. The standard deviation of the baseline (time from -20 to 0 s) was
269
defined as a noise. The noises of both mean intensities of the ROI and differential value
270
decreased as increasing the number of frames. However, the performance of SMA’s high-
271
frequency noise reduction was saturated over 16 frames; thus, we used 16 frames for the SMA
272
filter to the mean intensity of the ROI in subsequent experiments. We further investigate the
273
influence of the frame number of the SMA filter on the differential value. As Figure S-2a
274
shows, the time course of the differential values becomes smoother as increasing the frame
275
numbers. Figure S-2b summarizes response time and noise as a function of the number of
276
frames. When 300 frames were used in the SMA filter, significantly small background noise
277
was obtained as well as a good response time; therefore, we decided to use 300 frames in the
278
SMA filter that was applied to the differential value.
279
Figure 2a shows fluorescence and differential images 0, 20 and 120 s after applying 10
280
ppm standard gaseous AcH to the ADH-immobilized mesh. Originally the fluorescence image
281
at 0 second was uniformly bright image due to the presence of NADH on the mesh. The
282
images in Figure 2a are the ones converted by equation 1. As a result of image differential
12 ACS Paragon Plus Environment
Page 13 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
283
method, observation of both spatial and temporal change of AcH concentration on the ADH-
284
immobilized mesh surface was achieved (supplemental movie 1). Figure 2b shows temporal
285
changes of mean intensity of the ROI and differential value. The ROI intensity has reached a
286
plateau after the initial increase and did not recover to the baseline even after AcH application
287
was stopped. In contrast, the differential value exhibited a clear peak 20 s after applying AcH
288
and returned to the baseline immediately after stopping AcH gas. Mean intensity of the ROI
289
between 90–100 s after applying the AcH was defined as a signal output (∆I); the peak of the
290
differential value was defined as a signal output of the differential value (∆D). The response
291
time of mean intensity of the ROI T90 was defined as a time which took for signal to reach
292
90% of ∆I; the response time of differential value TR was the time to reach peak maximum of
293
∆D. It was revealed that the response time was improved by 15 s (T90 = 35 s and TR = 20 s for
294
10 ppm AcH) using the image differential method. Changing of fluorescence intensity
295
continued for a while even after stopping AcH gas since a little amount of AcH remained on
296
the ADH-immobilized mesh. In contrast, a reaction speed was decreasing immediately
297
because that of ADH was sensitive to the concentration of AcH. The differential value
298
indicated the reaction speed of ADH reverse reaction. Therefore, the image differential
299
method enabled to improve response time by accurately detect changing of concentration on
300
the ADH-immobilized mesh. Note that a speed of enzymatic reaction is proportional to
301
substrate concentration only in the first-order reaction region. Thus, an improvement of
302
reaction time may not be occured for high concentration of AcH.. However, this need not be
303
taken into account in our experiment because AcH concentration in breath and skin gas are
304
lower than 10 ppm.
305 306
(Figure 2 comes here)
307 13 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 14 of 30
308
Figure 3a shows temporal changes of a differential value to various concentrations of
309
standard gaseous AcH, a calibration curve and differential images that were obtained 20 s
310
after applying 1, 5 and 10 ppm AcH. Each plot in the calibration curve was taken from the
311
peak of each differential value. As a result, the AcH sniff-cam showed a wide dynamic range
312
(0.1–10 ppm) that encompasses a concentration range of breath AcH after drinking (0.4–9.0
313
ppm). The ADH-immobilized mesh was designed for disposable by concerning sanitary of the
314
system because the sniff-cam supposed to use in medical fields. Thus, note that each plot of
315
the calibration curve in figure 3a was measured by a different ADH-immobilized mesh that
316
was prepared at the same time. The coefficient of variation of ∆D among three deferent ADH-
317
immobilized mesh was 1.5% as shown in figure S-3. Plots were fitted by the following
318
equation with a correlation coefficient of 0.997:
∆ = 0.02 × [()* (""+)]- + 0.07 × [()*(""+)] + 0.08
Eq.4
319
Figure 3b displays relative outputs ∆D to investigate selectivity of the AcH sniff-cam
320
to gaseous AcH. The relative output was obtained by normalizing the output for each VOC by
321
that for 5 ppm AcH. It was observed that the AcH sniff-cam showed output signals only from
322
samples containing AcH, which validated a high selectivity of the AcH sniff-cam due to a
323
high substrate specificity of the enzyme.
324
(Figure 3 comes here)
325 326
Visualization of breath AcH
327
Figure 4a shows differential images of 0, 20 and 120 s after applying breath samples
328
from ALDH2[+] and ALDH2[-] subjects 30 minutes after drinking. The spatiotemporal
329
change of breath AcH could be visualized similarly to standard gaseous AcH. Figure 4b
330
displays time course of those differential values. Both peaked at about the same timing and
331
returned to the baseline, and breath AcH concentration of the ALDH2[-] subject was about 3-
14 ACS Paragon Plus Environment
Page 15 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
332
fold higher (8.4 ppm) than that of the ALDH2[+] (2.8 ppm). Stagnation of recovering to initial
333
value in differentiation value at around 10 s after differentiation value reaching peak was
334
observed unlike a result of standard gaseous AcH that was shown in Figure 2. This
335
phenomenon seems originated of humidity of exhaled breath. When the breath was loaded in
336
PTFE tube, the breath was cooled immediately, and then amount of saturated water vapor was
337
decreased, finally condensation was caused on wall of the PTFE tube. After that when
338
applying the breath was stopped and starting the applying dry carrier air, AcH remaining in
339
the PTFE tube as condensation was applying to ADH-immobilized mesh again. Therefore, the
340
stagnation of decrease of differentiation value was occurred. The differential analysis was
341
very sensitive to concentration change of AcH on ADH-immobilized mesh surface. As a
342
result, unexpected influence of humidity in breath was realized. Note that this little disorder
343
on response caused by the humidity of breath did not influenced quantifying the AcH in the
344
breath as described in following discussion.
345 346
(Figure 4 comes here)
347 348
Time courses of breath AcH concentrations for ALDH2[+] and ALDH2[-] are shown
349
in Figure 5. In the graphs, vertical bar graphs represent values of the standard comparison by
350
AcH detector tubes. The AcH sniff-cam revealed that AcH concentration in exhaled breath
351
initially increased and peaked 30 minutes after drinking, and then gradually decreased with
352
time. It was reported that the concentration of AcH in blood showed a maximum 15–30
353
minutes after drinking
354
concentration was related to ethanol metabolism, and it reflected AcH blood concentration.
355
Obtained peak breath AcH concentrations for ALDH2[+] and ALDH2[-] subjects were 2.75 ±
356
0.39 ppm and 8.64 ± 0.32 ppm, respectively; these values were in good agreement with
42
. Therefore, it was considered that the change of breath AcH
15 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 16 of 30
13,43
. Also, AcH
357
reported results (ALDH2[+], 1.2 ± 1.0 ppm; ALDH2[-], 6.9 ± 2.5 ppm)
358
concentrations measured by detector tubes showed similar values and the same trend to the
359
AcH sniff-cam. These results and evidence demonstrated the sniff-cam’s capability of
360
accurate and quantitative visualization of the spatiotemporal distribution of breath AcH.
361 362
(Figure 5 comes here)
363 364
Distribution of AcH released from a palm skin
365
Figure 6a shows top and 30-degree tilted views of transdermal AcH distribution
366
released from a palm skin. Fluorescence intensity changed when applying skin gas, and this
367
image was created by merging pictures of a hand and a fluorescence distribution profile. From
368
the image, higher output was observed at around the metacarpal phalangeal joint of a second
369
finger and the thenar eminence than the other parts probably due to different emanation rate
370
over a palm skin as observed in our previous results of skin emanated ethanol that was
371
obtained by chemiluminescent sniff-cam
372
visualizing a concentration distribution of ethanol emanated from palm skin 31. More clearly
373
distribution image compared to previous one was obtained in this study. The difference was
374
seemingly made by an improvement of skin gas injector shown in figure 1d. Previously, skin
375
gas was applied to an enzyme immobilized mesh only using PMMA frame that had a 90 × 90
376
mm2 square hole. There was a possibility of touching an enzyme immobilized mesh when the
377
skin gas injector with a square hole was used; thus, skin surface had to fix in the air. Such a
378
situation, dilution, and agitation of skin gas could occur easily. In contrast, a honeycomb
379
spacer that allows fitting with skin surface at all position certainly without the possibility of
380
touching the ADH-immobilized mesh was newly added in this study. Thus, skin gas could be
381
applied to ADH-immobilized mesh directly without dilution and agitation. As the result, clear
44
. Fluorometric ethanol sniff-cam also achieved
16 ACS Paragon Plus Environment
Page 17 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
382
distribution image might have been obtained because of skin gas that reaching to the ADH-
383
immobilized mesh more increased than the previous system. However, the current AcH sniff-
384
cam cannot measure skin gas in real-time because a hand blocked the excitation light from the
385
UV-LED array when subject applied skin gas to the ADH-immobilized mesh. Therefore, the
386
image differential method could not be applied to the AcH sniff-cam. Figure 6b shows a time
387
course of mean intensity difference of the ROI ∆I from the 15 minutes before drinking. The
388
temporal change of ∆I showed a similar trend to the breath concentration. The concentration
389
of skin gas AcH was calculated by calibration curve based on difference of fluorescence
390
intensity between before and after exposure of AcH ∆I
391
exhaled breath. The results suggest that the measurement conditions of the calibration curve
392
were not fit to the measurement conditions of the skin gas. For example, standard gaseous
393
AcH has the flow-rate of 40 mL/min although flow-rate of skin gas was not constant.
32
showed higher than the one in the
394
Besides, parameters for formation of concentration distribution of AcH from palm skin
395
are still unclear. Perspiration is closely related to the release pathway of skin gas, and it is
396
reported that there is much number of sweat gland on palm skin than the other part of the
397
body.
398
occurs by measuring several biological information such as sweat rate, skin temperature and
399
concentration distribution of skin gas, simultaneously in the future.
45
Therefore, we expect to investigate mechanism in which concentration distribution
400 401
(Figure 6 comes here)
402 403
Conclusions
404
In this research, we developed the AcH sniff-cam using a reverse reaction of ADH and
405
an image differential method that allowed spatiotemporal visualization of gaseous AcH
406
concentration. After the system construction, the image differential method that calculated 17 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 18 of 30
407
reaction speed of the reverse reaction was developed in order to improve temporal response of
408
the system. Characterization of the AcH sniff-cam revealed that the AcH sniff-cam had a
409
suitable dynamic range (100–10000 ppb) and selectivity for measurement of AcH in breath
410
and skin gas. Also, the image differential method allowed to improve temporal response by 15
411
s. Finally, spatiotemporal changes of AcH concentration in exhaled breath and distribution of
412
AcH in skin gas were visualized by the AcH sniff-cam. Breath AcH concentration peaked 30
413
minutes after drinking, and it was 3-fold higher for ALDH2[-] subject than that for the
414
ALDH2[+]. The difference in AcH release rate was also clearly visualized over a palm, which
415
also peaked 30 minutes after drinking. These results demonstrated that the sniff-cam holds a
416
huge potential for deployment in simple and noninvasive metabolism evaluation and disease
417
screening by acquiring spatiotemporal information of various VOCs in breath and skin gas.
418 419
Acknowledgement
420
This work was supported by the JSPS KAKENHI Grant Numbers JP17H01759,
421
JP16J09604, and JP15H04013, the Japan Science and Technology Agency (JST), and the
422
Ministry of Education, Culture, Sports, Science and Technology (MEXT) Special Funds for
423
Education and Research “Advanced Research Program in Neo-Biology”.
424 425
Supporting Information. Brief statement in nonsentence format listing the contents of the
426
material supplied as Supporting Information.
427
171208-AcH-vis-breath-skingas.docx Graphs about optimization of image analysis
428
parameter, reproducibility of the ADH-immobilized mesh
429
supplemental movie 1.avi Video of spatiotemporal change of AcH on the ADH-immobilized
430
mesh
431
18 ACS Paragon Plus Environment
Page 19 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
432
References
433
(1)
Peeling, R. W.; Artsob, H.; Pelegrino, J. L.; Buchy, P.; Cardosa, M. J.; Devi, S.; Enria,
434
D. A.; Farrar, J.; Gubler, D. J.; Guzman, M. G.; Halstead, S. B.; Hunsperger, E.; Kliks,
435
S.; Margolis, H. S.; Nathanson, C. M.; Nguyen, V. C.; Rizzo, N.; Vázquez, S.; Yoksan,
436
S. Nat. Rev. Microbiol. 2010, 8, S30–S37.
437
(2)
Castro, C.; Gourley, M. J. Allergy Clin. Immunol. 2010, 125, S238–S247.
438
(3)
Gallagher, M.; Wysocki, C. J.; Leyden, J. J.; Spielman, A. I.; Sun, X.; Preti, G. Br. J. Dermatol. 2008, 159, 780–791.
439 440
(4)
Shirasu, M.; Touhara, K. J. Biochem. 2011, 150, 257–266.
441
(5)
Kippenberger, S.; Havlíček, J.; Bernd, A.; Thaçi, D.; Kaufmann, R.; Meissner, M. Exp. Dermatol. 2012, 21, 655–659.
442 443
(6)
van Schooten, F. J. J. Breath Res. 2012, 6, 27108.
444 445
Boots, A. W.; van Berkel, J. J. B. N.; Dallinga, J. W.; Smolinska, A.; Wouters, E. F.;
(7)
Miekisch, W.; Schubert, J. K.; Noeldge-Schomburg, G. F. . Clin. Chim. Acta 2004, 347, 25–39.
446 447
(8)
Kaneko, T.; Wang, P. Y.; Sato, A. Occup. Environ. Med. 1994, 51, 68–72.
448
(9)
Cederbaum, A. I. Clin. Liver Dis. 2012, 16, 667–685.
449
(10)
Brooks, P. J.; Theruvathu, J. A. Alcohol 2005, 35, 187–193.
450
(11)
Salaspuro, M. J. Dig. Dis. 2011, 12, 51–59.
451
(12)
Garaycoechea, J. I.; Crossan, G. P.; Langevin, F.; Mulderrig, L.; Louzada, S.; Yang, F.;
452
Guilbaud, G.; Park, N.; Roerink, S.; Nik-Zainal, S.; Stratton, M. R.; Patel, K. J. Nature
453
2018, 553, 171–177.
454
(13) Fukunaga, T.; Yamamoto, K.; Adachi, J.; Ueno, Y.; Mizoi, Y. Japanese J. alcohol Stud. drug Depend. 1989, 24, 405–416.
455 456
(14)
Sekine, Y.; Toyooka, S.; Watts, S. F. J. Chromatogr. B 2007, 859, 201–207.
19 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
457
(15)
(16)
(17)
(18)
(19)
Agapiou, A.; Amann, A.; Mochalski, P.; Statheropoulos, M.; Thomas, C. L. P. TrAC Trends Anal. Chem. 2015, 66, 158–175.
466 467
Mochalski, P.; Unterkofler, K.; Hinterhuber, H.; Amann, A. Anal. Chem. 2014, 86, 3915–3923.
464 465
Turner, C.; Parekh, B.; Walton, C.; Spanel, P.; Smith, D.; Evans, M.; Patrik, S.; Smith, D.; Evans, M. Rapid Commun. Mass Spectrom. 2008, 22, 526–532.
462 463
Tsuda, T.; Ohkuwa, T.; Itoh, H. In Gas Biology Research in Clinical Practice; KARGER: Basel, 2011; pp 125–132.
460 461
Mochalski, P.; King, J.; Unterkofler, K.; Hinterhuber, H.; Amann, A. J. Chromatogr. B 2014, 959, 62–70.
458 459
Page 20 of 30
(20)
Curran, A. M.; Rabin, S. I.; Prada, P. a.; Furton, K. G. J. Chem. Ecol. 2005, 31, 1607– 1619.
468 469
(21)
Fenske, J. D.; Paulson, S. E. J. Air Waste Manage. Assoc. 1999, 49, 594–598.
470
(22)
Schmidt, F. M.; Vaittinen, O.; Metsälä, M.; Lehto, M.; Forsblom, C.; Groop, P.-H.; Halonen, L. J. Breath Res. 2013, 7, 17109.
471 472
(23)
Kamei, T.; Tsuda, T.; Mibu, Y.; Kitagawa, S.; Wada, H.; Naitoh, K.; Nakashima, K. Anal. Chim. Acta 1998, 365, 259–266.
473 474
(24)
Hara, T.; Matsui, H.; Shimizu, H. PLoS One 2014, 9, e111833.
475
(25)
Haze, S.; Gozu, Y.; Nakamura, S.; Kohno, Y.; Sawano, K.; Ohta, H.; Yamazaki, K. J. Invest. Dermatol. 2001, 116, 520–524.
476 477
(26)
Mitsubayashi, K. Sensors Mater. 2015, 27, 1123–1130.
478 479 480
Suzuki, Y.; Ye, M.; Miyajima, K.; Arakawa, T.; Sawada, S.-I.; Kudo, H.; Akiyoshi, K.;
(27)
Toma, K.; Miyajima, K.; Sawada, S.; Arakawa, T.; Kudo, H.; Akiyoshi, K.; Mitsubayashi, K. Sensors Mater. 2016, 28, 1265.
20 ACS Paragon Plus Environment
Page 21 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
481
Analytical Chemistry
(28)
Mitsubayashi, K. Sensors Mater. 2016, 28, 1295.
482 483
(29)
(30)
(31)
(32)
(33)
(34)
(35)
Phillips, M.; Herrera, J.; Krishnan, S.; Zain, M.; Greenberg, J.; Cataneo, R. N. J. Chromatogr. B Biomed. Sci. Appl. 1999, 729, 75–88.
496 497
Chien, P.-J.; Ye, M.; Suzuki, T.; Toma, K.; Arakawa, T.; Iwasaki, Y.; Mitsubayashi, K. Talanta 2016, 159, 418–424.
494 495
Iitani, K.; Chien, P.-J.; Suzuki, T.; Toma, K.; Arakawa, T.; Iwasaki, Y.; Mitsubayashi, K. ACS Sensors 2017, 2, 940–946.
492 493
Iitani, K.; Sato, T.; Naisierding, M.; Hayakawa, Y.; Toma, K.; Arakawa, T.; Mitsubayashi, K. Analyst 2017, 142, 3830–3836.
490 491
Arakawa, T.; Sato, T.; Iitani, K.; Toma, K.; Mitsubayashi, K. Anal. Chem. 2017, 89, 4495–4501.
488 489
Chien, P.-J.; Suzuki, T.; Tsujii, M.; Ye, M.; Toma, K.; Arakawa, T.; Iwasaki, Y.; Mitsubayashi, K. Biosens. Bioelectron. 2017, 91, 341–346.
486 487
Ye, M.; Chien, P.-J.; Toma, K.; Arakawa, T.; Mitsubayashi, K. Biosens. Bioelectron. 2015, 73, 208–213.
484 485
Saito, H.; Hashimoto, Y.; Minamide, T.; Kon, T.; Toma, K.; Arakawa, T.;
(36)
Navaneethan, U.; Parsi, M. A.; Gutierrez, N. G.; Bhatt, A.; Venkatesh, P. G. K.;
498
Lourdusamy, D.; Grove, D.; Hammel, J. P.; Jang, S.; Sanaka, M. R.; Stevens, T.; Vargo,
499
J. J.; Dweik, R. A. Gastrointest. Endosc. 2014, 80, 1038–1045.
500
(37)
Yokoyama, A.; Omori, T.; Yokoyama, T. Keio J. Med. 2010, 59, 115–130.
501
(38)
Jones, A. W. Alcohol Alcohol 1995, 30, 271–285.
502
(39)
Jones, A. W. J. Anal. Toxicol. 1986, 10, 98–101.
503
(40)
Norberg, A.; Gabrielsson, J.; Jones, A. W.; Hahn, R. G. Br. J. Clin. Pharmacol. 2000,
504
49, 399–408.
21 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
505
(41)
Page 22 of 30
Arakawa, T.; Wang, X.; Kajiro, T.; Miyajima, K.; Takeuchi, S.; Kudo, H.; Yano, K.; Mitsubayashi, K. Sensors Actuators B Chem. 2013, 186, 27–33.
506 507
(42)
Jones, A.; Neiman, J.; Hillbom, M. Br. J. Clin. Pharmacol. 1988, 25, 213–221.
508
(43)
Mizoi, Y.; Hishida, S.; Ijiri, I.; Maruyama, J.; Asakura, S.; Kijima, T.; Okada, T.; Adachi, J. Alcohol. Clin. Exp. Res. 1980, 4, 354–360.
509 510
(44)
Bioelectron. 2015, 67, 570–575.
511 512
Arakawa, T.; Kita, K.; Wang, X.; Miyajima, K.; Toma, K.; Mitsubayashi, K. Biosens.
(45)
Wilke, K.; Martin, A.; Terstegen, L.; Biel, S. S. Int. J. Cosmet. Sci. 2007, 29, 169–179.
513 514
22 ACS Paragon Plus Environment
Page 23 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
515
Analytical Chemistry
Graphics and captions
516 517
Figure 1. (a) Detection principle of AcH sniff-cam that employs a reverse reaction of ADH.
518
NADH absorbs UV light (λ = 340 nm) and emits fluorescence (λ = 490 nm). (b) Schematic
519
illustration of the experimental setup for characterization of the AcH sniff-cam. In
520
measurement of the exhaled breath, breath flow controller shown in (c) was used instead of
521
gas generator and MFCAcH in (b). In order to keep spaces between a palm skin surface and the
522
ADH-immobilized mesh constant, (d) PMMA frame and honeycomb spacer were used for
523
visualization of transdermal AcH.
524 23 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 24 of 30
525 526
Figure 2. (a) Fluorescence (upper) and differential images (lower) obtained by AcH sniff-cam
527
0, 20 and 120 s after applying 10 ppm AcH. The differential images were created by applying
528
an image differential method to the fluorescence images. (b) Comparison of the signal
529
responses of fluorescence intensity and differential value. The average of the mean intensity
530
of the ROI between 90–100 s after applying the AcH was defined as a signal output (∆I); the
531
peak of differential value was defined as a signal output (∆D).
532
24 ACS Paragon Plus Environment
Page 25 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
533 534
Figure 3. (a) Calibration curve of the AcH sniff-cam to gaseous AcH. The dynamic range was
535
0.1–10 ppm with the correlation coefficient of 0.997. The inset shows time course of the
536
differential value with various concentration of AcH (0.1, 0.5, 1, 3, 5, 8 and 10 ppm). The
537
upper images are differential images 20 s after applying 1, 5 and 10 ppm of AcH. (b) The
538
selectivity of the AcH sniff-cam against typical chemical components in breath. The output of
539
each sample was normalized by that of AcH.
540 541
25 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 26 of 30
542 543
Figure 4. (a)Differential images 0, 20 and 120 s after applying exhaled breath of ALDH2[+]
544
and ALDH2[-] subject 30 minutes after drinking. (b) Time course of the corresponding
545
differential values.
546
26 ACS Paragon Plus Environment
Page 27 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
547 548
Figure 5. Time course of concentration of breath AcH after drinking obtained from (a)
549
ALDH2[+] subjects and (b) ALDH2[-] subjects. As a comparison, the concentration was
550
measured by gas detection tubes simultaneously (vertical bar).
551
27 ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 28 of 30
552 553
Figure 6. (a) Distribution of transdermal AcH over a palm of ALDH2[+] subject 30 minutes
554
after drinking. (b) Time course of fluorescence intensity ∆I, differences from the intensity 15
555
minutes before drinking.
556
28 ACS Paragon Plus Environment
Page 29 of 30 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
557
Analytical Chemistry
for TOC only ADH
acetaldehyde
CH3CHO +
NADH
Ex. 340 nm
+
H+
pH6.5
ethanol
CH3CH2OH + NAD+
pH10.0 Fl. 490 nm
distribution of acetaldehyde UV-LED array
ADH-immobilized mesh
camera BPF
558
29 ACS Paragon Plus Environment
ADH
Analytical Chemistry
acetaldehyde
1 3 2 3 4 Ex. 340 nm 5 6 7 8 UV-LED array 9 10 11 12 13 14 15 16 17 18 19 20 ADH-immobilized mesh 21 22 23
CH CHO +
NADH
+
H+
pH6.5 pH10.0
Page 30 of 30
ethanol
CH3CH2OH + NAD+
Fl. 490 nm
distribution of acetaldehyde
camera BPF
ACS Paragon Plus Environment