Gas Bubble Releaser for Use in Dumas Nitrogen Determination

Note: In lieu of an abstract, this is the article's first page. Click to increase image size Free first page. View: PDF | PDF w/ Links. Citing Article...
0 downloads 0 Views 151KB Size
January, 1946

ANALYTICAL EDITION

Make the test solution ammoniacal, warm, acidify with acetic acid, and then add the reagent solution. This procedure is not 50 sensitive as those already outlined. The nonspecific phenol reaction of tannic acid with dilute sodium acetate-ferric chloride solution is not so sensitive as the test described here. As a test-tube reaction the identification limit is 25 micrograms; the limiting concentration is 1 to 40,000. The corresponding figures for the drop reaction on a spot plate are 1 microgram and 1 to 50,000. Hence the new test is three times as sensitive as the ferric-phenol reaction when carried out in a test tube, and just as sensitive as the latter when spot test techniques are used. DETECTION O F N A T U R A L T A N N I N S AND DIFFERENTIATION F R O M SYNTHETIC T A N S

The test solution (0.5%) is prepared from the solid specimen. One milliliter of the filtered liquid is carried through one of the procedures as outlined. When testing extracts of tanning agents, test portions are prepared by diluting the specimen 1 to 10 and 1 to 100, and 1-ml, portions are used. The behavior of all solutions toward ammonia and acetic acid should be determined before adding the reagent. If a precipitate appears on adding acetic acid to the warm ammoniacal solution, it should be removed, and the test tube procedure carried out on the filtrate after again making the solution ammoniacal. The test with Fe(a,a’-dip)aSO4 was tried on a variety of tanning materials. The following natural tannins gave a positive reaction: gallotannin; tara powder; gambier; myrobalan; quebracho; and extracts of wattle, mangle, sumac, and fustic. I n marked contrast the following synthetic tanning agents gave a negative reaction: orotan N, syntan A, syntan S,mertanol 7 L, maxyntan, and tanasol. These findings indicate that this test is suitable for distinguishing commercial natural from synthetic tanning agents. Similar differentiating tests were also made on’finished, mostly colored, leathers. About 0.5 gram of the leather was cut into tiny bits and boiled in 2 ml. of ammonia water for 2 minutes. The clear filtrate was carried through the test procedure. Preliminary tests with ammonia and acetic acid were made in all

Gas

Bubble Releaser

63

cases. This step is indispensable when testing leathers colored with coal-tar dyes, because the addition of the reagent to the ammoniacal solutions of many acid dyes results in a precipitate that does not dissolve in acetic acid. If the preliminary test produces a precipitate, it is filtered off, and the tannin test is made on the clear filtrate after i t is again made ammoniacal. I n 19 out of 20 cases, the test revealed the nature (vegetable or synthetic) of the tanning agent. The single exception was a leather that had been tanned with a sulfited quebracho extract; it gave a negative response. Further studies will be necessary to determine whether the test fails with leathers tanned with strongly sulfited tannin extracts. A trial with a technical sulfited quebracho extract showed that it still gives a distinct tannin reaction a t a dilution of 1 to 1000. DETECTION OF T A N N I N S IN BEVERAGES

Three milliliters of the samples were taken for the tests. Positive reactions were given by 4 varieties of red wine (tart, sweet): 3 varieties of white wine (tart, sweet); and water extracts of tea, mat6, and guarana. Negative responses were obtained with Cinzano (Italian origin); beer (light, dark); and water extracts of coffee (raw, roasted). ACKNOWLEDGMENT

The samples of leather, and of the natural and synthetic tanning agents, were donated by Cortume Carioca, Rio de Janeiro. The authors express their gratitude to this firm for its cooperation. LITERATURE CITED

(1) Feigl, F., and Miranda, L. I., IND.ESG. CHEM.,ANAL.ED., 16, 141 (1944). (2) Nierenstein, M., Section on “Tannins”, “Allen’s Commercial Organic .4nalysis”, Vol. 5 , p. 5, Philadelphia, P. Blakiston’s Son & Co., 1927. (3) Powell, A. R., and Schoeller, W. R., Analyst, 50, 485 (1925); Z. anorg. Chem., 151, 221 (1926). (4) Yoe, J. H., and Sarver, L. A., “Organic Analytical Reagents”. pp. 130,256, New York, John Wiley & Sons, 1941.

for Use in Dumas Nitrogen Determination Asotometers RENATO POMATTI, The Texas Co., Beacon, N. Y.

A

RATHER common and annoying occurrence in running a Dumas nitrogen determination is the sticking of gas bubbles to the surface of the mercury in t,he azotometer. This has been attributed to too narrow an opening of the gas inlet of the nitrometer, too short a distance between the gas inlet and the level of the mercury, excessive greasing of the gas inlet stopcock (S), and the use of perfectly pure clean mercury when the azotometer is firjt filled ( 4 ) . FlaschentriiLer 1,f) and Keygand ( 5 ) say that this difficulty can be overcome by the addition of powdered copper oxide to the surface of the mc’rcury. Sichols ( 2 ) suggests the use of mercurous oxide for the same purpose. S o n e of these measures, which are prevent,ivc in character, has been found completely reliable. A simple direct method for releasing gas bubbles already sticking to the mercury has been used effectively in this laboratory for over a year.

A piece of steel or iron wire about 1 em. long, slightly curved, is placed in the azotometer so that i t rests on the surface of the mercury. Khenever gas bubbles are to be released, the piece of wire is swept over the mercury by means of a small permanenttype magnet held outside the azotometer near the level of the mercury and opposite the wire. The wire is attracted to the mag-

net and, in passing over the surface of the mercury, releases the bubbles. The wire can thus be moved back and forth simply by placing the magnet in the proper position opposite it. To facilitate this opbration two magnets can be used. These are held on opposite sides of the azotometer and brought near the azotometer alternately. I n addition, this device can be used to break up small bubbles a t the potassium hydroxide solution-gas interface. This is done by attracting the mire to the magnet and moving it up slowly to the interface. By moving the wire up and down through the interface, the gas bubbles are broken. Large gas bubbles stuck or moving up very slowly in the graduated portion of the azot,ometer can be released also by this means. LITERATURE CITED

(1) Flaschefiriiger, B., Miikrdchemie, 8, 1 (1930). (2) Nichols, &I. L., I N DENG. . CHEM.,A w a ~ED., . 5, 149 (1933). (3) Niederl, J. B., and Kiederl, V.,“Micromethods of Quantitative

Organic Analysis”, 2nd ed., p. 98, Sew York, John Wiley & Sons, 1942. (4) Pregl-Fyleman, “Quantitative Organic Microanalysis”, 2nd English ed., p. 97, Philadelphia, P. Blakiston’s Son & Co., 1930. (5) Weygand, C., “Quantitative analytische Mikromethoden der organischen Chemie in vergleichender Darstellung”, pp. 24-8, Leipaig, Akademische Verlagsgesellschaft, 1931.