A green and fast laser fusion technique for bulk silicate rock analysis by laser ablation ICP-MS Chenxi Zhang, Zhaochu Hu, Wen Zhang, Yongsheng Liu, Keqing Zong, Ming Li, Haihong Chen, and Shenghong Hu Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.6b02471 • Publication Date (Web): 15 Sep 2016 Downloaded from http://pubs.acs.org on September 15, 2016
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
A green and fast laser fusion technique for bulk silicate rock analysis
2
by laser ablation ICP-MS
3
Chenxi Zhang,† Zhaochu Hu,*,†, Wen Zhang†, Yongsheng Liu†, Keqing Zong†, Ming
4
Li†, Haihong Chen†, Shenghong Hu†
5
†
6
Geosciences, Wuhan 430074, China
7 8
‡
State Key Laboratory of Geological Processes and Mineral Resources, China University of
‡
The Beijing SHRIMP Center, Institute of Geology Chinese Academy of Geological Sciences,
Beijing 102206, P.R. China
9 10 11 12 13 14 15
*Author to whom correspondence should be sent. E-mail:
[email protected] Tel.: +86 27 61055600, Fax: +86 27 67885096
16 17 18 19 20 21 22
Submitted to Analytical Chemistry
23 1
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
24
For Table of Contents Only
25 26 27
2
ACS Paragon Plus Environment
Page 2 of 24
Page 3 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
28
ABSTRACT:
29
Sample preparation of whole-rock powders is the major limitation for their
30
accurate and precise elemental analysis by laser ablation ICP-MS. In this study, a
31
green, efficient, and simplified fusion technique using a high energy infrared laser
32
was developed for major and trace elemental analysis. Fusion takes only tens of
33
milliseconds for each sample. Compared to the pressed pellet sample preparation, the
34
analytical precision of the developed laser fusion technique is higher by an order of
35
magnitude for most elements in granodiorite GSP-2. Analytical results obtained for
36
five USGS reference materials (ranging from mafic to intermediate to felsic) using the
37
laser fusion technique generally agree with recommended values with discrepancies
38
of less than 10% for most elements. However, high losses (20–70%) of highly volatile
39
elements (Zn and Pb) and the transition metal Cu are observed. The achieved
40
precision is within 5% for major elements and within 15% for most trace elements.
41
Direct laser fusion of rock powders is a green and notably simple method to obtain
42
homogeneous samples, which will significantly accelerate the application of laser
43
ablation ICP-MS for whole-rock sample analysis.
44 45
3
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
46
Page 4 of 24
INTRODUCTION
47
Accurate determination of major and trace element concentrations and isotopic
48
ratios in geological samples is a prerequisite for most geological investigations.
49
Dissolution is usually required for whole-rock elemental and isotope ratio analyses
50
using modern instrumental techniques, such as inductively coupled plasma mass
51
spectrometry (ICP-MS) and multi-collector ICP-MS. However, dissolution is tedious
52
and time-consuming and is thus the limiting factor for high sample throughput,
53
especially for geological sample analysis.1, 2 Many hazardous digestion solvents (e.g.,
54
HF, HNO3, HCl) are used in analytical geochemistry laboratories, which may cause
55
risk to the operator and result in pollution of the environment. Laser
56
ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS) provides
57
several advantages,3,
58
sample preparation, low blanks, and high sample throughput. This method has thus
59
become a popular technique, not only for in situ microanalysis6-8 but also for bulk
60
analysis of geological materials.9-19 However, LA-ICP-MS is a micro-sampling
61
technique, and the need for preparation of stable, homogeneous, and mechanically
62
resistant targets prior to the analysis of whole-rock powdered samples has been a
63
major drawback hindering accurate and precise measurements using this technique.
64
Three sample preparation methods are generally used for this purpose: (a) pressed
65
powder pellets,12,
66
lithium-borate fusion glasses.14, 16, 29, 30
67
4
including the ability for multi-elemental analysis, simple
18, 20-23
(b) flux-free fusion glasses,11,
15, 17, 19, 24-28
and (c)
Pressed powder pellet is the first technique used in sample preparation for 4
ACS Paragon Plus Environment
Page 5 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
68
whole-rock LA-ICP-MS analysis and is usually used for X-ray fluorescence (XRF)
69
analysis. Several binders have been used for enhancing the mechanical resistance of
70
pellets and analytical sensitivity, including PVA,12,
71
powder,33 and vanillic acid;34 however, the ablation yields still vary greatly between
72
individual spots. The analytical precision (RSD) of this method is not sufficiently high
73
(approximately 10-20%) due to the inhomogeneity of the pressed pellets and the
74
dilution of trace elements by the binder. Grain-size has been recognized as an
75
important factor for producing homogeneous and cohesive undiluted pressed pellets.18,
76
22, 23, 35, 36
77
powder tablets without the addition of a binder by applying wet-milling protocols in
78
an aqueous suspension, using a high power planetary ball mill and agate tools. They
79
found that the precision was in the same range of 55%). Homogeneity in such samples can only be obtained by applying high melting
95
temperatures (1700-1800℃) and long fusion times (60-120 sec or higher), resulting in
96
the preferential loss of highly volatile elements (e.g., Pb, Zn, and Cs) and some
97
transition metal elements (e.g., Cr, Ni, and Cu) by alloying with the metal heater. An
98
alternative approach to realize fast fusion and sample homogeneity is the dilution of
99
high SiO2 concentrations by the addition of high-purity MgO.27,
37
However, this
100
introduces other problems, such as contamination and a tedious sample preparation
101
process. In addition, the molten glass has to be quenched immediately to avoid the
102
formation of mineral crystals. In spite of this, crystallization during quenching is
103
sometimes unavoidable, especially for Fe- and Ni-rich samples (olivine crystals
104
formed upon quenching in komatiitic15).
105
Another technique, lithium-borate fusion glass formation, reduces the
106
temperature required for fusion of the rock powder by mixing it with a flux agent
107
(generally LiBO2 or LiB4O7).9, 14, 29, 38 Consequently, the loss of volatile elements is
108
suppressed, as has been extensively used for XRF analysis for years. This method is
109
suitable for a wide variety of bulk compositions including mafic, intermediate, and
110
silicic rocks. Yu et al.30 fused seventeen reference materials, with SiO2 contents
111
ranging from 50% to 77%, with a lithium borate flux (sample: flux =1:3). Agreement 6
ACS Paragon Plus Environment
Page 7 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
112
of the LA-ICP-MS results is less than 10% relative with published reference values
113
and the analytical precision based on replicate analyses typically has RSD better than
114
5% when using a matrix-matched calibration strategy. However, the sample is diluted
115
because of the addition of a large amount of flux agent (the flux to sample ratio
116
commonly being 3-5:1), leading to increased limits of detection.30, 39 Additionally, the
117
flux introduces inevitable contamination into the samples (e.g., La and Ce)
118
the ICP-MS instrument (e.g., Li and B), due to which a longer rinsing time is required
119
between samples.30
16, 40
and
120
In this study, a green, cost-efficient, and simplified direct fusion technique was
121
established to prepare homogenous fused glasses for routine LA-ICP-MS whole-rock
122
analysis. Briefly, we used a high energy infrared laser to achieve rapid melting of
123
samples and instant cooling of glasses in tens of milliseconds, bypassing the use of
124
any chemical reagent or melting container. To test the melting capabilities of the
125
infrared laser for silicate rocks (especially for zircon-bearing granitic rocks and other
126
felsic samples), both major and trace elemental contents of five powdered
127
international
128
basalt-andesite-rhyolite) were determined simultaneously by LA-ICP-MS after fusion.
129
The reliability of this technique is demonstrated by the satisfactory accuracy and
130
precision of LA-ICP-MS data for most elements.
131
EXPERIMENTS
reference
materials
(spanning
the
compositional
range
132
Geological Materials. A series of RMs ranging from mafic (basalts BHVO-2
133
and BCR-2) to intermediate (andesite AGV-2 and granodiorite GSP-2) to felsic 7
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
134
(rhyolite RGM-2) composition were analyzed to assess the performance of the method
135
for which reliable compilation values are available from the geochemical GeoReM
136
database (http://georem.mpch-mainz.gwdg.de).41,
137
difficult because of their high melt viscosities, which hinders rapid homogenization of
138
the melt during fusion. In addition, it is also difficult to completely dissolve the
139
frequently present refractory accessory minerals, such as zircon, in felsic rock.
140
Therefore, the rhyolite RMs USGS RGM-2 and granodiorite RM USGS GSP-2,
141
whose SiO2 contents and refractory mineral contents are sufficiently high, were
142
selected to optimize and evaluate the new fusion technique.
42
Fusion of felsic rocks is very
143
Sample Preparation Procedure. Prior to sample fusion, the powdered reference
144
material was prepared into pressed pellets using a hand operated hydraulic pressure
145
pelleter (TP40, Herzog, Germany). Approximately 0.5 g powder sample, without
146
further grinding (200 mesh), was compressed at a pressure of 240 MPa for 5 min with
147
the help of a circular plastic resin ring (sample holder) to retain the form of the
148
powdered sample (Figure 1a). The pressed pellet was then fused at ambient
149
temperature and pressure. A high energy infrared laser (JHM-1GY200E, Wuhan
150
Chutian Industry Laser Equipment Co., LTD, China), consisting of optical maser,
151
laser power supply, cooling system, laser target designate system, light guide focusing
152
system, computer-control system and workbench, was used as the heat producer to
153
achieve complete fusion, which is widely used for metal welding in industry (Figure
154
S1). The pressed pellet was put under the laser focus position by 20 cm (Figure 2),
155
high-power and long-pulse mode of the laser was set during sample fusion, with a 8
ACS Paragon Plus Environment
Page 8 of 24
Page 9 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
156
repetition rate of 2 Hz. Under defocus and relatively lower repetition rate conditions
157
of infrared laser was used to ensure sample melting instead of ablation. Details of the
158
apparatus and operating conditions are listed in Table S1. Figure 1b shows the
159
momentary state of molten glass (GSP-2). After several pulses, the top layer
160
(approximately 1 mm thick) of the sample was fused, and natural cooling immediately
161
embedded in the pellet. A visual representation of this procedure is shown in Video S1.
162
As shown in Figures 1c and 1d, GSP-2 glasses contained more bubbles than BHVO-2
163
glasses because the discharge of bubbles was hindered by the high viscosity of the
164
glass melt. Nonetheless, a large amount of pure glass was still available.
165
Instrumentation. Experiments were conducted on an Agilent 7500a ICP-MS
166
(Agilent Technology, Tokyo, Japan) in combination with a 193 nm ArF excimer LA
167
system (GeoLas 2005, Lambda Physik, Göttingen, Germany) owned by the State Key
168
Laboratory of Geological Processes and Mineral Resources, China University of
169
Geosciences (Wuhan). The 193 nm excimer laser is installed with an optical
170
configuration that produces a fairly flat-topped lateral energy distribution, leading to
171
pan-shaped ablation pits on the sample. Helium was chosen as the ablation cell gas as
172
it has been found to consistently enhance the signal 2 folds compared to argon gas
173
with the 193 nm excimer laser.43 The carrier gas flows were optimized by ablating
174
NIST SRM 610 to obtain maximum signal intensity for U+, while keeping the ThO/Th
175
ratio