Interfacing Microfluidics with Negative Stain ... - ACS Publications

Institute of Molecular Biophysics, Florida State University, 91 Chieftain Way, Tallahassee, Florida 32306, United States. Anal. Chem. , 2016, 88 (1), ...
0 downloads 0 Views 1MB Size
Subscriber access provided by ORTA DOGU TEKNIK UNIVERSITESI KUTUPHANESI

Technical Note

Interfacing Microfluidics with Negative Stain Transmission Electron Microscopy Nikita Mukhitov, John Michael Spear, Scott Stagg, and Michael G. Roper Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.5b03884 • Publication Date (Web): 07 Dec 2015 Downloaded from http://pubs.acs.org on December 16, 2015

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 22

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

1 2

Interfacing Microfluidics with Negative Stain

3

Transmission Electron Microscopy

4

Nikita Mukhitov §, John M. Spear §, Scott M. Stagg, Michael G. Roper*

5 6 7

*To whom correspondence should be addressed:

8

Michael G. Roper, Florida State University, 95 Chieftain Way, Tallahassee, FL, 32306

9

E-mail: [email protected]

10 11

§

Contributed to the work equally

12 13 14 15 16

1

ACS Paragon Plus Environment

Analytical Chemistry

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 2 of 22

17

Abstract

18

A microfluidic platform is presented for preparing negatively stained grids for use in

19

transmission electron microscopy (EM). The microfluidic device is composed of glass etched

20

with readily fabricated features that facilitate the extraction of the grid post-staining and

21

maintains the integrity of the sample. Utilization of this device simultaneously reduced

22

environmental contamination on the grids and improved the homogeneity of the heavy metal

23

stain needed to enhance visualization of biological specimens as compared to conventionally

24

prepared EM grids. This easy-to-use EM grid preparation device provides the basis for future

25

developments of systems with more integrated features, which will allow for high throughput

26

and dynamic structural biology studies.

27

2

ACS Paragon Plus Environment

Page 3 of 22

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

28

Analytical Chemistry

Introduction

29

Cryogenic electron microscopy (cryoEM) is quickly becoming a routine method in the

30

determination of near atomic, high-resolution structures of biological molecules. However, for

31

most samples before cryoEM data can be collected, the sample quality and heterogeneity

32

must be first characterized using negative staining.1 The conventional workflow for negative

33

staining of biological molecules consists of manual application of sample, removal of excess

34

solution, and finally the addition of a heavy metal solution (for example, uranyl acetate,

35

ammonium molybdate, or osmium tetroxide) (Figure 1A).

36

contrast of the images by forming a strongly scattering shell around the biological sample,

37

which gives biological molecules a light appearance on a dark background. It also

38

encapsulates the sample, preserving the molecular structure during dehydration before the

39

sample is inserted into the EM column. Because of these many roles, the thickness and

40

uniformity of the stain is crucial for image quality and distinction of structural features.2

The stain acts to increase the

41

Conventionally, EM grids are prepared by hand and, as such, variability is introduced

42

due to user-to-user differences. The variability of the staining can have large effects on the

43

final stained sample, ultimately hindering the resolution, image processing, and data analysis.1-

44

5

45

manual sample handling and are thus prone to variations in the final result.2,3

While several reports have described procedures for optimized negative staining, all require

46

Over the last two decades, microfluidics has shown its potential in offering reproducible,

47

small volume, fluid control for the design of integrated lab-on-a-chip systems.6,7 This has

48

resulted in the development of platforms that integrate multiple handling steps and minimize

49

user-contributed variability. While microfluidics has previously been interfaced with other

50

analytical platforms, it has only rarely been coupled with EM. In many cases, specialty devices

51

were built for time-resolved cryoEM8-9 and for negative staining10-11; however, these required

52

specialty instrumentation10-11 or fabrication methods8-9. 3

ACS Paragon Plus Environment

Analytical Chemistry

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 4 of 22

53

The objective of this work was to develop a platform that could alleviate the variability

54

associated with manual staining of EM grids. To achieve this, we developed a microfluidic

55

system that enclosed the grid in a chamber where the delivery of the sample and drying was

56

performed in a controlled fashion (Figure 1B). Images acquired from this system indicated

57

reproducible stain thickness and image quality. We expect that this initial system will be

58

further enhanced with the available suite of microfluidic tools. For example, subsequent

59

iterations of this design with valves12, timers13-15, or other microfluidic features16,17 can be used

60

for structural biology studies on dynamic systems or in high throughput applications.

61 62

Experimental Methods

63

Reagents. Nitric acid, hydrogen peroxide and hydrofluoric acid were obtained from VWR

64

(Radnor, PA). Sodium hydroxide and ethanol were obtained from Sigma Aldrich (St. Louis,

65

MO).

66

(Morrisville, PA). For all solutions, ultrapure deionized water was used (Barnstead

67

International, Inc., Dubuque, IA).

(Tridecafluoro-1,1,2,2-tetrahydrooctyl)

trichlorosilane

was

obtained

from

Gelest

68 69

Microfluidic Fabrication. The fabrication workflow for the production of borosilicate glass

70

microfluidic devices is shown in Figure 2. The extraction divot was etched to a depth of 45

71

µmusing conventional methods.18 The channels and grid chamber were then fabricated using

72

a second etch to a depth of 50 µm. The same process was repeated for another piece of

73

glass, producing the mirror image of the design features to serve as the bottom complement.

74

All dimensions of the channels were verified using a Mitutoyo SJ-411 Surface Profiler

75

(Kanagawa, Japan). Fluid access holes were drilled with a 1.1 mm diamond-tipped drill bit

4

ACS Paragon Plus Environment

Page 5 of 22

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

76

(Crystalite Corp., Lewis Center, OH). The finished top slide was then fitted with a reservoir

77

(Idex, Lake Forest, IL) using epoxy.

78

For surface modification, the glass was cleaned by submerging in 5 M NaOH for 10

79

minutes. The surface was rinsed with water and dried with N2. Subsequently, the slides were

80

oxidized in a plasma cleaner (Harrick, Ithaca, NY) for 2 minutes. Immediately after, the slides

81

were placed in a vacuum desiccator and (tridecafluoro-1,1,2,2-tetrahydrooctyl) trichlorosilane

82

was deposited using a method previously described.19 Afterwards, the slides were rinsed with

83

water, dried with N2, and stored in clean petri dishes at room temperature until use.

84 85

Kvβ2.1 Expression and Purification. Full length rat Kvβ2.1 (GenBankTM accession number:

86

CAA54142) was expressed and purified as previously reported.20 Briefly, the PCR verified

87

pQE70 expression vector was transformed into BL21-DE3 cells using standard molecular

88

techniques and Kvβ2.1 purified using the Talon Co2+ His-Pur resin (BD Biosciences, East

89

Rutherford, NJ). Eluted protein was purified using a Superose 6 10/300 column (GE Life

90

Sciences, Pittsburgh, PA) using an ÄKTA FPLC (GE Life Sciences). Purified protein was

91

concentrated to 1.9 mg/ml using a 5000 molecular weight cut off spin concentrator (Thermo,

92

Waltham, MA) and stored at -80°C until further use. The final storage buffer contained the

93

following: 20 mM Tris pH 8.0, 150 mM KCl, 1 mM 2-mercaptoethanol, and 10% (w:v) glycerol.

94

All reagents used were purchased from Amresco (VWR, Radnor, PA).

95

Prior to staining, Kvβ2.1 was thawed on ice and centrifuged at 16,000X g for 20 minutes

96

at 4°C to remove any aggregated or precipitated materials. The protein was diluted to 0.095

97

mg/ml using dilution buffer (20 mM Tris pH 8.0, 150 mM KCl, 1 mM 2-mercaptoethanol) prior to

98

use in microfluidic or conventional sample preparation.

99

5

ACS Paragon Plus Environment

Analytical Chemistry

1 2 100 3 4 101 5 6 7 102 8 9 103 10 11 104 12 13 14105 15 16 106 17 18 19107 20 21108 22 23 109 24 25 26110 27 28111 29 30 31112 32 33113 34 35 114 36 37 38115 39 40116 41 42 43117 44 45118 46 47 48119 49 50120 51 52121 53 54 55122 56 57123 58 59 60

Page 6 of 22

Conventional Sample Preparation. For conventional preparation of negative stain EM grids, 3 µl of diluted Kvβ2.1 was applied to continuous carbon coated GC400 copper mesh grids (Electron Microscopy Sciences, Hatfield, PA) that were rendered hydrophilic by glow discharge in 3:1 (v:v) Ar/O using a Solarus 950 advanced plasma system (Gatan, Pleasanton, CA). Grids were hand blotted using filter paper (Ted Pella, Redding, CA), then immediately stained using 3 µl of 2% (w:v) uranyl acetate (Electron Microscopy Sciences). Excess stain was then hand blotted using filter paper (Ted Pella) and the grids were air dried and stored at room temperature in a sealed desiccator before imaging.

Microfluidic Sample Preparation. The bottom glass microfluidic slide was placed into an aluminum manifold built in-house (Figure 3). A carbon coated, copper grid was rendered hydrophilic in the same manner as described above and placed into the device chamber. Eight drops (~20 µL each) of dilution buffer were distributed around the non-etched parts of the glass slides. The top slide was aligned using the manifold and lowered to the bottom slide, displacing the buffer and creating a thin sealing layer. The top of the manifold was attached and screwed down to seal the device. Sample (20 µL) was loaded in the inlet and a vacuum was applied at the outlet to fill the chamber with sample.

Alternatively, sample could be

delivered to the inlet using a syringe with appropriate fitting. The vacuum was removed and sample was left in the chamber for 10 seconds after which 50 µL of uranyl acetate stain was loaded into the inlet and carried through with vacuum. After 10 seconds, compressed air was blown into the inlet and used to dry the grid. The air also purged the thin film of buffer between the slides, enabling the device to be opened and the grid extracted via the divot with a pair of forceps. The grids were stored at room temperature in a sealed desiccator before imaging.

6

ACS Paragon Plus Environment

Page 7 of 22

1 2 124 3 4 125 5 6 7 126 8 9 127 10 11128 12 13 14129 15 16130 17 18 131 19 20 21132 22 23133 24 25 26134 27 28135 29 30136 31 32 33137 34 35 36138 37 38139 39 40140 41 42 43141 44 45142 46 47 143 48 49 50144 51 52145 53 54 146 55 56 57147 58 59148 60

Analytical Chemistry

Electron Microscopy and Particle Classification. EM micrographs of Kvβ2.1 were collected on a CM-120 BioTwin (Phillips, Hillsboro, OR) operating at 120 keV using a Tem-Cam F224 slow scan CCD camera (Tietz, Freising, Germany) with a magnification that resulted in a nominal pixel size of 2.88 angstroms per pixel at the specimen level. EM micrographs were uploaded to the Appion processing suite.21 Kvβ2.1 particles were picked in a semi-automatic fashion using the template picker FindEM.22

Two dimensional (2D) class averages were

generated using the maximum likelihood alignment algorithm within the Xmipp package.23 3D volumes of Kvβ2.1 were generated using the Frealign reconstruction package.24 6038 particles were used for the final reconstruction for both the device-prepared and manually-prepared samples. Both reconstructions refined to 21 Å resolution as assessed by the FSC0.143 criterion.25 The Kvβ2.1 crystal structure (PDB ID: 1ZSX), back-filtered to 50 Å, was used for the initial model. Kvβ2.1 volumes were visualized using Chimera.26

Results and Discussion The method developed achieved reproducibility in the staining process by integrating all the sample preparation steps into a single device that housed the EM grid.

Further

advantages brought to light by this device were: (1) it could be operated by hand, (2) the device was reusable, and (3) the system is amenable for future integration with on-chip valves and plumbing. The device consisted of a chamber with support barriers around the edges that held an EM grid in place during delivery of sample and stain. The support barriers were used to prevent the grid from sliding during the application of the sample. It was found that in the absence of the support barriers, the grid slid in the chamber, tearing the thin layer of carbon film, ruining the sample.

The extraction divot also permitted easy grid extraction upon

disassembly of the device. On occasion, the grid would stick to the top glass slide again 7

ACS Paragon Plus Environment

Analytical Chemistry

1 2 149 3 4 150 5 6 7 151 8 9 152 10 11153 12 13 14154 15 16155 17 18 156 19 20 21157 22 23158 24 25 26159 27 28160 29 30161 31 32 33162 34 35163 36 37164 38 39 40165 41 42166 43 44 167 45 46 47168 48 49169 50 51 52170 53 54171 55 56172 57 58 59173 60

Page 8 of 22

ruining the carbon film and grid. Having a chamber and divot etched also on the top slide alleviated this problem. A challenge encountered was how to seal the device to ensure the sample or stain would not wick out of the channels, but would still allow the device to be opened afterwards and allow removal of the grid. Reversible sealing of the microfluidic device was achieved by derivatizing the glass surface with a fluorosilane to render the glass both fluorophilic and hydrophobic. We initially attempted to seal the device with a fluorous liquid between the slides13,19, but large fibril-like structures were observed after EM images were acquired. By using a few drops of aqueous buffer between the glass surfaces in conjunction with an aluminum manifold to hold the pieces together (Figure 3), the seal was sufficient to prevent the sample from wicking between the slides as well as allowing a syringe pump to drive solution through the device, while still enabling the device to be opened after staining. In Figure 4, the general quality of the grids prepared by hand (top row) and the microfluidic device (bottom row) were compared using low magnification images.

In both

cases, five grids were prepared consecutively and a 3x3 square area from each grid is shown. The first benefit that was observed with the use of the microfluidic device was improvements to the cleanliness of the grids. Large uneven stain features, seen as dark black regions on these images, was commonplace in the conventional method, but was mostly absent from the grids made using the microfluidic system (Figure 4). A further zoomed-out view (80 X) of two grids is shown in the Electronic Supplementary Information (ESI), Figure S1. Variability was also observed with respect to the staining. The right-most image of the top row of Figure 4 shows striations of different stain thicknesses in the grids that were prepared conventionally. The thickness of the stain will dictate the contrast and signal to noise ratio of the images. The homogeneity of the images shown in Figure 4 was quantified by measuring the RSD of the gray scale values as described in the ESI. The RSD from the grids prepared by hand was 8

ACS Paragon Plus Environment

Page 9 of 22

1 2 174 3 4 175 5 6 7 176 8 9 177 10 11178 12 13 14179 15 16180 17 18 181 19 20 21182 22 23183 24 25 26184 27 28185 29 30186 31 32 33187 34 35188 36 37189 38 39 40190 41 42191 43 44 192 45 46 47193 48 49194 50 51 52195 53 54196 55 56197 57 58 59198 60

Analytical Chemistry

24% (Figure S2A), while the RSD from those prepared by the device was 2% (Figure S2B). It should be pointed out that the conventionally made grids were still usable at higher magnifications by zooming in on regions that did not contain these artifacts. However, the contamination and stain variations consumed functional space on the grid and may make image processing and specimen classification more difficult. At higher magnifications, the quality of the images acquired following preparation in the device was equal, if not better, to those prepared conventionally.

A magnification series

comparing both preparations is shown in Figure 5 (top row, conventional preparation; bottom row, microfluidic preparation).

As a test sample, Kvβ2.1 was used as a model biological

specimen. Kvβ2.1 is a 320 kDa cytosolic complex that directly interacts with voltage-gated potassium channels.20 It was chosen for these studies due to its relatively small size as smaller specimens often are more difficult to observe in negative stain. Again, stain artifacts were evident in the low magnification image from the grid prepared by hand (Figure 5A, left image), and uneven staining was observed in the 28,000 X magnification image (Figure 5A, middle). Nonetheless, by selectively choosing the location for higher magnification images, stained Kvβ2.1 particles can be discerned (Figure 5A, right). When using the device for grid preparation (Figure 5B), the low magnification images showed negligible contamination and a consistent stain thickness. As the magnification increased, the Kvβ2.1 particles are seen as monodisperse and evenly stained yielding high contrast particles with consistent staining from image to image. To illustrate the viability of this approach for structural biology, 2D class averages and 3D reconstructions were made from the Kvβ2.1 negative stain images produced by hand or with the device (Figure 6). Reference-free 2D class averages from device-prepared sample were qualitatively more feature rich and showed more unique orientations (Figure 6C) than the averages from the manually prepared sample (Figure 6A). This observation was consistent 9

ACS Paragon Plus Environment

Analytical Chemistry

1 2 199 3 4 200 5 6 7 201 8 9 202 10 11203 12 13 14204 15 16205 17 18 206 19 20 21207 22 23208 24 25 26209 27 28210 29 30211 31 32 33212 34 35 36213 37 38214 39 40215 41 42 43216 44 45217 46 47 218 48 49 50219 51 52220 53 54 55221 56 57222 58 59223 60

Page 10 of 22

with 3D reconstructions of the Kvβ2.1 particles (Figure 6B and 6D). Both device-prepared and manually-prepared samples refined to 21 Å resolution, but the contours of the device-prepared sample better matched a fitted Kvβ2.1 crystal structure with only 2,320 atoms outside of the EM density (Figure 6D) as opposed to 2,904 atoms in the manually-prepared sample (Figure 6B) when both reconstructions enclosed the same total volume. Together, these analyses revealed that the data collected from manual preparation were of lower quality and showed fewer features in the 3D reconstruction (Figure 6A and 6B) as compared to the data collected from the microfluidic system (Figures 6C and 6D). One possible interpretation for the better results from the device-prepared sample is that the microfluidic chamber may have protected Kvβ2.1 from flattening during the dehydration step of grid preparation. These results validate our device-based approach and provide a foundation for future integration with other microfluidic features.

Conclusion A simple microfluidic system is presented that allows reproducible staining of EM grids with the added benefit of reduced stain artifacts. High magnification images from grids prepared by the microfluidic system showed similar image qualities as those prepared by hand. With this methodology for housing the grid, opportunities abound for more integrated systems using elastomeric materials for incorporation of valving and other microfluidic features.

For example, this system can subsequently be complemented with gradient

generators or multi-analyte perfusion and reaction timers to study both multi-variable interactions as well as reaction kinetics. This proof of principle paves the way for future added layers of complexity that can be used to uniquely investigate structural biology dynamics.

Supporting Information 10

ACS Paragon Plus Environment

Page 11 of 22

1 2 224 3 4 225 5 6 7 226 8 9 227 10 11228 12 13 14229 15 16 17230 18 19231 20 21 232 22 23 24233 25 26234 27 28 29235 30 31 236 32 33 34237 35 36238 37 38 39239 40 41240 42 43241 44 45 46242 47 48243 49 50244 51 52 53245 54 55246 56 57 58 59 60

Analytical Chemistry

The Supporting Information is available free of charge on the ACS Publications website. Low magnification (80 X) images of a grid prepared in the conventional manner and by the microfluidic device; description of the method used to measure the gray scale levels of Figure 4; surface plots of the gray scale levels of Figure 4.

Acknowledgements This work was supported in part by a grant from the National Institutes of Health (R01 GM108753) to SMS, and a Developing Scholar Award from Florida State University (MGR). The authors declare the following competing financial interest(s): Authors NM, JMS, SMS, and MGR are coauthors on an invention disclosure related to portions of the work reported herein.

References: 1. Ohi, M.; Li, Y.; Cheng, Y.; Walz, T. Biol. Proceed. 2004, 6, 23-24. 2. Rames, M.; Yu, Y.; Ren, G. J. Vis. Exp. 2014, 90, e51087. 3. Booth, D. S.; Avila-Sakar, A.; Cheng, Y. J. Vis. Exp. 2011, 58, e3227. 4. Melchior, V.; Hollingshead, C. J.; Cahoon, M. E. J. Cell Biol. 1979, 81, 665-671. 5. Sander, B.; Golas, M. M. Microsc. Res. Tech. 2011, 74, 642-663. 6. Squires, T. M.; Quake, S. R. Rev. Mod. Phys. 2005, 77, 977-1026. 7. Easley, C. J.; Karlinsey, J. M.; Bienvenue, J. M.; Legendre, L. A.; Roper, M. G.; Feldman, S. H.; Hughes, M. A.; Helett, E. L.; Merkel, T. J.; Ferrance, J. P.; Landers, J. P. Proc. Natl. Acad. USA 2006, 103, 19272-19277. 8. Lu, Z.; Shaikh, T. R.; Barnard, D.; Meng, X.; Mohamed, H.; Yassin, A. S.; Mannella, C. A.; Agrawal, R. K.; Lu, T. M.; Wagenknecht, T. J. Struct. Biol. 2009, 168, 388-395.

11

ACS Paragon Plus Environment

Analytical Chemistry

1 2 247 3 4 248 5 6 7 249 8 9 250 10 11251 12 13 14252 15 16253 17 18 254 19 20 21255 22 23256 24 25 26257 27 28258 29 30259 31 32 33260 34 35261 36 37262 38 39 40263 41 42264 43 44 265 45 46 47266 48 49267 50 51 52268 53 54269 55 56270 57 58 59 60

Page 12 of 22

9. Lu, Z.; Barnard, D.; Shaikh, T. R.; Meng, X.; Mannella, C. A.; Yassin, A. S.; Agrawal, R. K.; Wagenknecht, T.; Lu, T. M. J. Micromech. Microeng. 2014, 20, 115001. 10. Kemmerling, S.; Ziegler, J.; Schweighauser, G.; Arnold, S. A.; Giss, D.; Müller, S. A.; Ringler, P.; Goldie, K. N.; Goedecke, N.; Hierlemann, A.; Stahlberg, H.; Engel, A.; Braun, T. J. Struct. Biol. 2012, 177, 128-134. 11. Giss, D.; Kemmerling, S.; Dandey, V.; Stahlberg, H.; Braun, T. Anal. Chem. 2014, 86, 4680-4687. 12. Unger, M. A.; Chou, H. P.; Thorsen, T.; Scherer, A.; Quake, S. R. Science 2000, 288, 113-116. 13. Shen, F.; Du, W.; Davydova, E. K.; Karymov, M. A.; Pandey, J.; Ismagilov, R. F. Anal. Chem. 2010, 82, 4606-4612. 14. Cho, H.; Kim, H. Y.; Kang, J. Y.; Kim, T. S. J. Colloid Interf. Sci. 2007, 306, 379-385. 15. Pompano, R. R.; Platt, C. E.; Karymov, M. A.; Ismagiolv, R. F. Langmuir 2012, 28, 1931-1941. 16. Zhao, B.; Moore, J. S.; Beebe, D. J.; Science 2001, 291, 1023-1026. 17. Dertinger, S. K. W.; Chiu, D. T.; Jeon, N. L.; Whitesides, G. M. Anal. Chem. 2001, 73, 1240-1246. 18. Mukhitov, N; Yi, L; Schrell, A. M.; Roper, M. G. Anal. Chem. 2014, 1367, 154-160. 19. Du, W.; Li, L.; Nichols, K. P.; Ismagilov, R. F. Lab Chip 2009, 9, 2286-2292. 20. Weng, J.; Cao, Y.; Moss, N.; Zhou, M. J. Biol. Chem. 2006, 22, 15194-15200. 21. Lander, G. C.; Stagg, S. M.; Voss, N. R.; Cheng, A.; Fellmann, D.; Pulokas, J.; Yoshioka, C.; Irving, C.; Mulder, A.; Lau, P.; Lyumkis, D.; Potter, C. S.; Carragher, B. J. Struct. Biol. 2009, 166, 95-102. 22. Roseman, A. M. J. Struct. Biol. 2004, 145, 91-99.

12

ACS Paragon Plus Environment

Page 13 of 22

1 2 271 3 4 272 5 6 7 273 8 9 274 10 11275 12 13 14276 15 16277 17 18 278 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Analytical Chemistry

23. Scheres, S. H.; Valle, M.; Nuñez, R.; Sorzano, C. O.; Marabini, R.; Herman, G. T.; Carazo J. M. J. Mol. Biol. 2005, 348, 139-149. 24. Grigorieff, N. J. Struct. Biol. 2007, 157, 117-125. 25. Rosenthal, P. B.; Henderson, R. J. Mol. Biol. 2003, 333, 721-745. 26. Pettersen, E. F.; Goddard, T. D.; Huang, C. C.; Couch, G. S.; Greenblatt, D. M.; Meng, E. C.; Ferrin, T. E. J. Comput. Chem. 2004, 25, 1605-1612.

13

ACS Paragon Plus Environment

Analytical Chemistry

1 2 279 3 4 5 280 6 7 281 8 9 282 10 11 12283 13 14284 15 16 285 17 18 19286 20 21287 22 23 24288 25 26289 27 28290 29 30 31291 32 33292 34 35 293 36 37 38294 39 40295 41 42 296 43 44 45297 46 47298 48 49 50299 51 52300 53 54301 55 56 57302 58 59303 60

Page 14 of 22

Figure Captions

Figure 1. Workflow for the preparation of negative stain grid for EM. [A] The procedure for conventional sample preparation by hand is shown.

A series of five steps is used to

produce the final stained sample, but the large number of steps involving manual handling leads to variable qualities in the final sample.

Additionally, the grid is exposed to the

environment and is prone to contamination. [B] The procedure using a microfluidic platform for the sample preparation is shown. Upon loading the grid in the chamber, the sample is applied and stained within the device, minimizing variations in preparation steps. Because the grid is preserved in the device throughout the preparation stage, there is less chance for environmental contamination.

Figure 2. Device fabrication. [A] First the extraction divot was etched into the glass with subsequent etching of the grid chambers and channels. The divot allowed for an efficient means to extract the grids post-staining. [B] The top and side view of the final, fabricated device is shown. The use of the support barriers kept the grid in place during delivery of solutions.

Figure 3. Completed view of device. A final image of the EM negative stain preparation device is shown.

A metal manifold was used to align and clamp the device together.

Reservoirs permitted incorporation of pumps for automated delivery of reagents. A US quarter is shown on the left for scale.

Figure 4. Macroscopic stain features. Low magnification (130 X) images of macroscopic stain features with and without the sample prep device. Each panel represents a separately 14

ACS Paragon Plus Environment

Page 15 of 22

1 2 304 3 4 305 5 6 7 306 8 9 307 10 11308 12 13 14309 15 16310 17 18 311 19 20 21312 22 23313 24 25 26314 27 28315 29 30316 31 32 33317 34 35318 36 37319 38 39 40320 41 42321 43 44 322 45 46 47323 48 49324 50 51 52325 53 54326 55 56327 57 58 59328 60

Analytical Chemistry

prepared grid. (Top) When prepared by hand, contamination with particulates and debris was noted. The three images on the right show striations in the stain thickness. (Bottom) When prepared with the microfluidic device, the grid was predominantly clean and only infrequent particulates were observed. Stain thickness was even throughout the grid as determined by the homogeneous image contrast.

Figure 5. Magnification series of Kvβ2.1 particles. High magnification images of samples prepared by hand [A] and with the microfluidic device [B]. The final stained images at 45,000 X magnification are comparable, with a slightly improved stain coverage when using the microfluidic device. These images are representative of the five grids that were prepared by hand and the device.

Figure 6. Comparison of 2D class averages and 3D reconstructions from TEM images acquired using manual preparation and the microfluidic device. [A] Reference-free 2D class averages from images collected on a manually prepared grid. [B] 3D reconstruction of Kvβ2.1 from the manually prepared grid. The Kvβ2.1 crystal structure (Blue, ribbon representation, PDB ID: 1ZSX) was fitted to the EM density map for comparison purposes. [C] Reference-free 2D class averages from images collected on a grid prepared within the microfluidic device. [D] 3D reconstruction from the device-prepared grid also fitted with the Kvβ2.1 crystal structure. The surface contours (gray) for both reconstructions were adjusted so that both reconstructions enclosed the same volume.

15

ACS Paragon Plus Environment

Analytical Chemistry

1 2 329 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 330 29 30 31331 32 33 332 34 35 36333 37 38 39334 40 41 42335 43 44 45336 46 47337 48 49 50338 51 52 53339 54 55 56340 57 58341 59 60

Figure 1.

16

ACS Paragon Plus Environment

Page 16 of 22

Page 17 of 22

1 2 342 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36343 37 38 344 39 40 41345 42 43 44346 45 46 47347 48 49 348 50 51 52349 53 54 55350 56 57 58351 59 60

Analytical Chemistry

Figure 2.

17

ACS Paragon Plus Environment

Analytical Chemistry

1 2 352 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 353 27 28 29354 30 31 32355 33 34 356 35 36 37357 38 39 40358 41 42 43359 44 45 46360 47 48361 49 50 51362 52 53 54363 55 56 57364 58 59 60

Figure 3.

18

ACS Paragon Plus Environment

Page 18 of 22

Page 19 of 22

1 2 365 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 366 19 20 21367 22 23 24368 25 26 369 27 28 29370 30 31 32371 33 34 35372 36 37 373 38 39 40374 41 42 43375 44 45 46376 47 48 49377 50 51378 52 53 54379 55 56 57380 58 59 60

Analytical Chemistry

Figure 4.

19

ACS Paragon Plus Environment

Analytical Chemistry

1 2 381 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 382 27 28 29383 30 31 32384 33 34 385 35 36 37386 38 39 40387 41 42 43388 44 45 46389 47 48390 49 50 51391 52 53 54392 55 56 57393 58 59 60

Figure 5.

20

ACS Paragon Plus Environment

Page 20 of 22

Page 21 of 22

1 2 394 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27395 28 29 30396 31 32 397 33 34 35398 36 37 38399 39 40 41400 42 43 401 44 45 46402 47 48 49403 50 51 52404 53 54 55405 56 57406 58 59 60

Analytical Chemistry

Figure 6.

21

ACS Paragon Plus Environment

Analytical Chemistry

1 2 407 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22408 23 24409 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

For TOC Only.

22

ACS Paragon Plus Environment

Page 22 of 22