Measuring polymer film thickness by interference patterns

Measuring Polymer Film Thickness by Interference Patterns ... film is a useful applications experimentfor undergraduate students. ... CT, 1974, p. 3-1...
0 downloads 0 Views 331KB Size
Measuring Polymer Film Thickness by Interference Panerns The experiment reported by Allpress, Cowell, and Herd' that measures both the thickness and the vinyl acetate content of packaging film js a useful applications experiment for undergraduate students. The authors noted that the fdm thickness measured using interference fringe patterns of infrared radiation differed from that measured using a micrometer. The reported ratio of the spectroscopidly determined thickness to the micrometer-measured thickness is 1.3-1.5. This is probably very elme to the refractive index of the polymer film. In fact, a more precise equation relating thickness ( d )and interferenee patterns should he usedz

d=

rn Zn(i7, - F2)

where rn is the number of peak maxima between measured interference peaks; n is the refractive index; and i7 is the wavenumber. When chwsing two adjacent peak maxima, m = 1, and the equation becomes 1

.

d=-

Zn(A3 where d is in centimeters. An incorrect equation far this relationship was given in the ieported experiment. I have found that using the refractive index of the monomer or an average of the comonomer refractive indices obtained from standard tables aives a reasonable approximation to the film thickness for the polymer in question and compares reliably with that obtained using a micrometer K. N., Cowell, B. J., and Herd, A.'c., J. CHEM.EDUC.,58,741 (1981). Hannah, R. W., and Swinehart, J. S., "Experiments in Techniques of Infrared Spectroscopy," Perkin-Elmer, Nonvalk, CT, 1974, p. 3-1. Benjamin F. Plummer Trinity University Sen Antonio, TX 78284 1 Allpress,

Volume 61

Number 5

May 1984

439