Miniature Cylindrical Ion Trap Mass Spectrometer - Analytical

Custom-designed electronic components include the RF scanning and ... and lens power supplies, as well as a custom-software application. Direct leak ...
3 downloads 0 Views 114KB Size
Anal. Chem. 2002, 74, 6145-6153

Miniature Cylindrical Ion Trap Mass Spectrometer Garth E. Patterson,‡ Andrew J. Guymon, Leah S. Riter, Mike Everly, Jens Griep-Raming,† Brian C. Laughlin, Zheng Ouyang, and R. Graham Cooks*

Chemistry Department, Purdue University, West Lafayette, Indiana 47907-1393

A miniature cylindrical ion trap mass spectrometer is described, and preliminary data are presented. Functionality and performance of laboratory-scale instruments have been maintained to the extent possible in this battery-operated mass spectrometer. Capabilities include tandem mass spectrometry experiments. Custom-designed electronic components include the RF scanning and amplification system, data acquisition components, and lens power supplies, as well as a custom-software application. Direct leak and membrane introduction inlet systems are used for sample introduction. A mass/charge range of ∼250 Th with unit mass resolution has been demonstrated.

Field-portable analytical instrumentation is of increasing importance.1-3 In situ analysis is required to preserve chemical information and save time, particularly for applications, such as chemical warfare agent detection and toxic pollutant measurement. Developing field-portable instrumentation requires miniaturization of all the individual components of the instrument. The first step in the process of building a miniature mass spectrometer involves the miniaturization of the mass analyzer itself. A recent review covers miniaturization of a variety of mass analyzer types.4 A cylindrical ion trap (CIT) mass analyzer was chosen for the effort described here. This analyzer can readily be fabricated in an appropriately small size; it provides tandem MS capabilities, gives good sensitivity, and is capable of trapping ions covering an adequate mass range for field applications.5 Moreover, this mass analyzer has other characteristics that further facilitate the miniaturization process. For instance, the amplitude of the applied RF trapping potential (and the size of the amplifier used to generate the trapping signal) is directly related to the size of the ion trap, making smaller traps more attractive. Ion trap mass analyzers can also be operated at relatively higher pressure than other mass analyzers, thus easing the requirements for the vacuum system. † Current address: Thermo Finnigan, MAT GmbH, Barkhausenstrasse 2, 28197 Bremen, Germany. ‡ Current address: Griffin Analytical Technologies, Inc. West Lafayette, IN 47906. (1) Lopez-Avila, V.; Hill, H. H. Anal. Chem. 1997, 69, R289-R305. (2) Richardson, S. D. Anal. Chem. 2002, 74, 2719; Anal. Chem. 2002, 74, 2719-2742. (3) Ketola, R. A.; Kotiaho, T.; Cisper, M. E.; Allen, T. M. J. Mass Spectrom. 2002, 37, 457-476. (4) Badman, E. R.; Cooks, R. G. J. Mass Spectrom. 2000, 35, 659-671. (5) Wells, J. M.; Badman, E. R.; Cooks, R. G. Anal. Chem. 1998, 70, 438444.

10.1021/ac020494d CCC: $22.00 Published on Web 11/09/2002

© 2002 American Chemical Society

Ion traps with cylindrical geometry have been described previously, both in the form of Paul (quadrupole) ion traps,5-8 and as ion cyclotron resonance (ICR) traps.9,10 Mass analysis via the mass-selective instability scan in a CIT mass analyzer (Paul type) was first described by Wells et al. in 1998.5 The performance of this analyzer is similar to that of the commercially available hyperbolic ion trap. Sequential product scans (one of the MS3 experiments) were demonstrated, as was baseline resolution of adjacent ion signals; the mass range was comparable to that available using an ion trap with hyperbolic electrodes operated under the same experimental conditions. The cylindrical geometry is physically simple to construct in a range of sizes, and the experimental data taken so far5-8 demonstrate that it generates a field that is an adequate approximation to the quadrupolar field required for mass analysis. The performance of an ion trap mass spectrometer depends strongly on the analyzer electrode geometry. Each of the dimensions of the cylindrical ion trap used in this miniaturization effort was chosen to provide the highest performance in a small ion trap volume. For a detailed description of the relationship of instrument performance to electrode geometry, see Badman et al.8 Briefly, the end-cap electrode spacing was chosen so as to add a small positive octapolar field in order to compensate for the local field defects associated with the 1.0-mm radius holes in the end-cap electrodes. Resonance ejection using a frequency that corresponds to the octapolar resonance results in fast ion ejection and produces narrow peaks.11 The electrode dimensions were also chosen to provide a mass/charge range of ∼250 Thomson (Th; units of mass/charge12) when using an applied RF of 2 MHz and 1000 V0-P. This choice of CIT size represents a compromise between smaller sizes that would increase relative field faults due to machining errors and larger analyzers, which would add to the power consumption and weight of the overall instrument. A trapping amplitude and frequency were chosen to provide adequate (6) Langmuir, D. B.; Langmuir, R. V.; Shelton, H.; Wuerker, R. F. Containment Device U.S. Patent 3,065,640, 1962. (7) Kornienko, O.; Reilly, P. T. A.; Whitten, W. B.; Ramsey, J. M. Rapid Commun. Mass Spectrom. 1999, 13, 50-53. (8) Badman, E. R.; Johnson, R. C.; Plass, W. R.; Cooks, R. G. Anal. Chem. 1998, 70, 4896-4901. (9) Gabrielse, G.; Mackintosh, F. C. Int. J. Mass Spectrom. Ion Processes 1984, 57, 1-17. (10) Easterling, M. L.; Mize, T. H.; Amster, I. J. Int. J. Mass Spectrom. Ion Processes 1997, 169, 387-400. (11) Franzen, J.; Gabling, R. H.; Schubert, M.; Wang, Y. Nonlinear Ion Traps; March, R. E., Todd, J. F. J., Eds.; CRC Press: Boca Raton, FL, 1995; Vol. 1, pp 49-167. (12) Cooks, R. G.; Rockwood, A. L. Rapid Commun. Mass Spectrom. 1991, 5, 93.

Analytical Chemistry, Vol. 74, No. 24, December 15, 2002 6145

Table 1. Major Components of the Miniature Mass Spectrometer (Version 5.0) item

manufacturer

turbo pump diaphragm pump vacuum manifold CPU timing card waveform generation DC supplies multiplier current amplifier RF amplifier filament

Pfeiffer Vacuum KNF Neuberger Finnigan ASUS National Instruments National Instruments (X2) EMCO K&M Keithley Custom SIS

purpose primary vacuum roughing pump on-board processor timing sequence RF modulation, end-cap waveforms lens supplies signal amplification signal amplification electron generation

defining characteristic high flow, small size small size, low power consumption readily available (modified) off-the-shelf high resolution high sample rate small size readily available, high gain off-the-shelf item small size, low power consumption off-the-shelf item

Figure 2. Schematic representation of the major components of the miniature mass spectrometer (verison 5.0) depicting their interconnections Figure 1. Schematic representation showing the ion source, ion optics, mass analyzer, and detector. The filament, mounting hardware, and gate electrode are based on those used in the Finnigan ITS-40 instrument.

ion trapping. Ions are trapped in a pseudopotential well, D, the depth of which (eq. eq 1)

D)

Vq 8

(1)

depends on the Mathieu stability parameter, q, and V, the amplitude of the applied RF. Under typical conditions in this device, it is under 10 volts. Instrument Description. The description of the instrument is divided into subsections on the basis of the relationship of each component to the overall miniaturization effort. Custom-written software is used to control the instrument and acquire data, and this is also described here. The instrument was designed in separate modules to facilitate its continued development. This modular approach is of particular importance for the ion source, because the timing electronics and power supplies required by each type of ion source may be different. An electron impact ionization source is described here, but the system also includes the capability to control an atmospheric sampling glow discharge ionization source, to be fitted in the future. Since building and characterizing the instrument, a new much smaller version (16 vs 55 kg) has also been built and partially characterized. This paper describes the original form (version 5), 6146

Analytical Chemistry, Vol. 74, No. 24, December 15, 2002

unless a specific reference is made to the new smaller version 7 instrument. Table 1 includes all of the major components in the miniature mass spectrometer, along with the component manufacturer, purpose, and defining characteristics. Each of these components is described in the following text. Figure 2 is a schematic representation of the interconnectivity of the major components. It should be noted that only a limited discussion on the development and miniaturization of the cylindrical ion trap mass analyzer is given here, since a more thorough description has appeared elsewhere.8 The mass analyzer accounts for