Molecularly-imprinted polymers for compound-specific isotope analysis of polar organic micropollutants in aquatic environments Rani Bakkour, Jakov Bolotin, Börje Sellergren, and Thomas B. Hofstetter Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.8b00493 • Publication Date (Web): 20 May 2018 Downloaded from http://pubs.acs.org on May 20, 2018
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
Molecularly-imprinted polymers for compound-specific isotope analysis of polar organic micropollutants in aquatic environments Rani Bakkour,†,‡, ¶ Jakov Bolotin,† Börje Sellergren,§ and Thomas B. Hofstetter∗,†,‡ Eawag, Swiss Federal Institute of Aquatic Science and Technology, Dübendorf, Switzerland, Institute of Biogeochemistry and Pollutant Dynamics, ETH Zürich, Zürich, Switzerland, , and Department of Biomedical Sciences, Malmö University, Malmö, Sweden E-mail:
[email protected] ∗ To
whom correspondence should be addressed
† Eawag ‡ ETH
Zürich address: Technical University of Munich, Chair for Analytical Chemistry and Water Chemistry, München,
¶ present
Germany § Malmö University
1
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Abstract
1
2
Compound-specific isotope analysis (CSIA) of polar organic micropollutants in environ-
3
mental waters requires a processing of large sample volumes to obtain the required analyte
4
masses for analysis by gas chromatography / isotope-ratio mass spectrometry (GC/IRMS).
5
However, the accumulation of organic matter of unknown isotopic composition in standard en-
6
richment procedures currently compromises the accurate determination of isotope ratios. We
7
explored the use of molecularly-imprinted polymers (MIPs) for selective analyte enrichment
8
for 13 C/12 C and 15 N/14 N ratio measurements by GC/IRMS using 1H-benzotriazole, a typical
9
corrosion inhibitor in dishwashing detergents, as example of a widely detected polar organic
10
micropollutant. We developed procedures for the treatment of > 10 L of water samples, in
11
which custom-made MIPs enabled the selective cleanup of enriched analytes in organic sol-
12
vents obtained through conventional solid phase extractions. Hydrogen bonding interactions
13
between the traziole moiety of 1H-benzotriazole and the MIP were responsible for selective
14
interactions through an assessment of interaction enthalpies and 15 N isotope effects. The pro-
15
cedure was applied successfully without causing isotope fractionation to river water samples
16
as well as in- and effluents of wastewater treatment plants containing µg/L concentrations of
17
1H-benzotriazole and dissolved organic carbon (DOC) loads of up to 28 mg C/L. MIP-based
18
treatments offer new perspectives for CSIA of organic micropollutants through the reduction
19
of the DOC-to-micropollutant ratios.
2
ACS Paragon Plus Environment
Page 2 of 31
Page 3 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
20
Introduction
21
The analysis of stable isotope ratios in organic soil and water contaminants has become a widely used
22
approach to identify the sources of pollution and to identify (bio)degradation pathways. 1–7 Whereas
23
constant ratios of 13 C/12 C, 2 H/1 H, 15 N/14 N, and of other elements in a compound enable one to infer
24
precursor materials, synthesis routes, and formation pathways of pollutants, 8–14 changes of isotope
25
ratios lead to stable isotope fractionation patterns that reveal the (bio)chemical reaction by which a
26
pollutant is degraded. 15–21 However, due to the poor sensitivity of gas and liquid chromatography
27
used in combination with isotope-ratio mass spectrometry, 22–25 the applications of compound-
28
specific isotope analysis (CSIA) have largely focused on so-called legacy contaminants such as
29
halogenated solvents, nitroaromatic explosives, and fuel constituents 26–33 Those compounds are
30
often found in the high µg L−1 to mg L−1 concentration range and can be extracted from the
31
environmental matrices in straightforward procedures, for example, through transfer of the analytes
32
into the gas phase and enrichment onto solid sorbents. 32–37 Unfortunately, such procedures are
33
not necessarily applicable for CSIA of polar organic micropollutants of current interest 38 such as
34
pesticides, pharmaceuticals, consumer chemicals, and personal care products. Their sub-µg L−1
35
concentrations in natural and treated waters require the processing of large sample volumes greater
36
than 5 L by solid-phase extractions (SPE, e.g., 39 ) to obtain the necessary analyte mass for isotope-
37
ratio mass spectrometry. Moreover, the inherently poor selectivity of SPE-based procedures will
38
very likely lead to a co-enrichment of organic matter of unknown isotopic composition and thus
39
compromise accurate isotope ratio measurements.
40
One promising option to increase the selectivity of analyte enrichment procedures is the use of
41
molecularly-imprinted polymers (MIPs). MIPs are synthetic, typically custom-made materials ca-
42
pable of molecular recognition of a target analyte through specific intermolecular interactions. 40–44
43
Synthesis is carried out in the presence of the target analyte and involves the polymerization of
44
functional monomers that are selected based on their ability to interact with functional groups
45
of the target molecule. Therefore, the final materials exhibit a three-dimensional structure with
46
high specificity. 45 Even though the concept of solid-phase extraction with molecularly-imprinted 3
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 4 of 31
47
polymers (MISPE) has been used successfully for years, 40,46–48 it has never been applied in com-
48
bination with isotopic analyses. In fact, MISPE based on non-covalent interaction between analyte
49
and sorbent seems ideally suited for the treatment of SPE-extracts, in which polar organic micropol-
50
lutants are normally enriched from large volumes of water samples. 49 Only µL volumes of organic
51
solvents can be injected onto gas chromatographs coupled to isotope-ratio mass spectrometers.
52
The substantial analyte enrichment to the mg L−1 level causes concomitant enrichment of organic
53
matrix and, therefore, requires further sample treatment. Because those extracts typically consist
54
of organic solvents, polar organic micropollutants can be extracted based on H-bonding and ionic
55
interactions with the MIP without interferences of water.
56
The goal of this work was to explore new sample clean-up strategies based on molecularly-
57
imprinted polymers for the stable isotope analysis of polar organic micropollutants in aquatic
58
environments. Here, we focused our work on the analysis of
59
benzotriazole (1H-BT), a corrosion inhibitor, which is widely used in dishwashing detergents.
60
Like many polar organic micropollutants, 1H-BT is too persistent to be degraded in waste water
61
treatment plants (WWTP) and therefore found in µg L−1 concentrations in rivers as well as influents
62
and effluents of WWTPs containing very different organic matrices. 50–52 The specific objectives of
63
our work were to (i) propose sample treatment procedures for MISPE-based stable isotope analysis
64
through the synthesis and application of a custom-made MIP for 1H-BT. (ii) We characterized
65
the selectivity of our materials in a comparison with other polar organic micropollutants such as
66
triazines and substituted benzotriazoles and (iii) evaluated the potential of method-induced isotope
67
fractionation from incomplete analyte recovery. (iv) Finally, we illustrate how MISPE-based
68
procedures for CSIA can be applied in different matrices by analysing δ13 C and δ15 N of 1H-BT in
69
riverwater, influent and effluent of a WWTPs, as well as in dishwashing detergent taps.
4
ACS Paragon Plus Environment
13 C/12 C
and
15 N/14 N
ratios of 1H-
Page 5 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
70
Experimental section
71
A list of all chemicals including suppliers and purities is provided in the Supporting Information
72
(SI).
73
Safety considerations
74
Milling and sieving the synthesized polymers can generate dust. The processing of materials should
75
be carried out under the fumehood. Additionally, gloves, goggles and respiration mask are advised
76
to be worn. The radical generator, azobisisobutyronitrile, decompose at temperature ≥ 45 °C and
77
may explode if large amounts are exposed to high temperatures. Do not exceed temperature of 40
78
°C during its purification.
79
Synthesis of molecularly-imprinted polymers
80
Molecularly-imprinted polymers were synthesized following an adapted procedure of Chapuis
81
et al. 53 for terbutylazine. The polymerization mixture consisted of 119.1 mg (1 mmol) 1H-
82
benzotriazole (1H-BT) as template, 344.4 mg (4 mmol) methacrylic acid as a functional monomer,
83
3964 mg (20 mmol) ethylene glycol dimethacrylate as crosslinker, 39.4 mg (0.24 mmol) azobi-
84
sisobutyronitrile as radical generator, and 5.6 mL dichloromethane as porogene. We dissolved
85
the template in the functional monomer, cooled the solution to 0℃ for 15 minutes and added the
86
remaining reagents. The polymerization mixture was purged with N2 in a borosilicate glass tube
87
(OD = 18 mm, ID = 15 mm), sealed with a screw-cap, and placed in an ice bath for 30 minutes. The
88
same procedure was used for preparation of non-imprinted polymer (NIP) except for the absence
89
of the template 1H-BT.
90
Polymerization took place over 24 hours at 2 ± 0.2℃ by immersing the glass tubes in a merry-
91
go-round photoreactor (DEMA 125, Hans Mangels GmbH, Bornheim-Roisdorf, Germany; Figure
92
S1) filled with a sodium nitrate filter solution (0.15 M, cut-off of wavelength > 320 nm) and
93
centred with a medium pressure mercury lamp (150 W) surrounded by a borosilicate glass jacket. 54
5
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 31
94
After polymerization, the tubes were removed and crushed to recover the polymer monoliths.
95
The polymers were sequentially ground (Mixer Mill MM 301, Retsch GmbH, Germany) and
96
sieved through 100 µm sieve (Retsch GmbH, Germany). The particle fraction < 100µm was
97
sedimented in methanol to obtain particle size distribution between 20 and 100 µm. We removed
98
the template by ten repetitive cycles of Soxhlet extraction using mixtures of methanol and formic
99
acid (90/10 vol-%) until ≥ 98% of the 1H-benzotriazole was recovered. Nonimprinted polymers
100
were extrated in the identical manner. The identical procedure was applied to synthesize a MIP
101
with 5-,6-(CH3 )2 -benzotriazole as template for the quantitative analysis of imprinting efficiency.
102
Possible interferences of residual analytes on the accuracy of stable isotope analysis were evaluated
103
exemplarily for 1H-BT through (a) quantification of its concentration before and after the use of
104
MIPs with 1H-BT in organic solvent and (b) the comparison of
105
1H-BT before and after the use of MIPs with 1H-BT spiked river water samples.
106
Sample preparation procedures using molecularly-imprinted polymers
107
The multi-step procedure for the enrichment and purification of 1H-BT in (a) in- an effluent of
108
waste water treatment plants (WWTP), (b) spiked river water, and (c) dishwashing tabs is shown
109
schematically in Figure 1. Generally, sample preparation consisted of two solid phase extraction
110
steps (Fig. 1, B and D) with different pre- and post treatments.
111
Enrichment of 1H-BT from waste water plant influent and effluent
112
Samples from WWTP influent (5 L, sample no. 1 in Figure 1, step A) and effluent (10 L) were
113
filtered through 0.7 µm glass fiber filters (Whatman, GF/F 47mm) and acidified with HCl (32 %) to
114
pH 2.0. Filtered samples 2 containing 1H-BT were enriched by conventional solid phase extraction
115
using OASIS HLB cartridges (Waters, 6 g, 35 mL) in step B using a 12-port vacuum extraction
116
manifold (Supelco, Switzerland). Prior to sample loading, cartridges were conditioned with 100
117
mL hexane, 100 mL ethyl acetate, 100 mL methanol and 100 mL MiliQ water. Water samples
118
2 were percolated through the cartridges using a vacuum pump at a flow rate of ≤ 7.2 mL/min 6
ACS Paragon Plus Environment
13 C/12 C
and
15 N/14 N
ratios of
Page 7 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
WWTP influent/effluent
aqueous sample (1) filtration acidification
B
flow rate 1-CH3 -BT. The observed sequence
307
agrees with the notion that IF are highest for the template used for synthesis. 48 We verified this
308
outcome by assessing IF of the same analytes on another custom-made MIP, where 5-,6-(CH3 )2 -
309
BT was employed as a template instead of 1H-BT. The IF (CH3 )2 −BT -values are shown in Table 1
310
for the 5-,6-(CH3 )2 -BT-imprinted MIP and they are of similar magnitude as IF-values for 1H-
311
BT imprinted polymers. IF (CH3 )2 −BT decrease in the sequence 5-,6-(CH3 )2 -BT > 5-CH3 -BT >
312
1H-BT > 1-CH3 -BT confirming that the selectivity was highest for the template molecule. With
313
both MIPs, we did not observe selective interactions of 1-CH3 -BT. Absence of selectivity for N-
314
substituted benzotriazole suggests that the interactions of analytes with the polymer was based on
315
hydrogen bonding of the triazole moiety with the carboxylic acid functional group of the monomer
316
(methacrylic acid, Figure 3). Substitution at the 1H position on the triazole ring caused complete
317
loss of selectivity towards 1-(CH3 )-BT on both polymers.
318
Further evidence for the selectivity of the MIPs was obtained from the comparisons of the
319
apparent interaction enthalpies, ∆HMIP , 63 with various analytes, and interaction enthalpies of non-
320
imprinted polymers, ∆HNIP (Table 1). ∆HMIP were most negative, that is most favourable, for the
321
template molecule 1H-BT and became less negative for the substituted benzotriazoles. ∆HMIP -
322
values ranged between 7.0 and 16 kJ/mol which is in the lower range of interaction enthalpies for
323
hydrogen bonding (8-84 kJ/mol, 64,65 ). Note that interactions of benzotriazoles with non-imprinted
324
materials were also exothermic but the difference between MIP and NIP, ∆HMIP − ∆HNIP , revealed
325
the larger selectivity of the MIPs. ∆HMIP − ∆HNIP for benzotriazoles followed the same sequence
326
reported for IF-values. However, values approached zero for the remaining compounds showing
327
the lack of selective interactions of the MIP with benzothiazole, naphthalene, and atrazine.
328
Isotope fractionation from incomplete analyte extraction
329
We quantified the extent of C and N isotope fractionation from analyte-MIP interactions in a
330
separate experiment on MISPE cleanup step E (Figure 1). Table 2 shows mass fractions, mi , 18
ACS Paragon Plus Environment
Page 18 of 31
Page 19 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
(a)
(b) HO
HO
H N N
CH3
N
N
N
H O
O
O
N
H
O
O
O
Figure 3. (a) Hydrogen bonding between the template, 1H -BT, and the functional monomer, methacrylic acid, used for imprinting enables selective interactions. (b) Absence of selective binding due to N-methyl substitution in 1-CH3 -BT. 331
and the corresponding δ13 C and δ15 N-values of 1H-BT recovered in three sample fractions (Fi ).
332
The total 1H-BT recovery was 99%. F1 corresponds to the fraction of analyte collected after
333
breakthrough during the loading of the cartridge with 1H-BT in dichloromethane. This fraction
334
amounted to 41% of the total analyte mass. F2 is the fraction of analyte obtained through washing
335
of the cartridge with the identical solvent (42%). F3 reflects the remaining 16% of 1H-BT that
336
were recovered from the MIP through elution with methanol. The same type of experiment was
337
carried out with dichloromethane extracts of river water using five-fold higher MIP (500 mg). As a
338
consequence of higher retention capacity, no 1H-BT was detected in F1 , whereas all 1H-BT mass
339
was recovered in the two remaining fractions, F2 (74%, Table 2) and F3 (26%).
340
The C isotope signatures of 1H-BT in the presence and absence of organic matrix from river
341
water samples varied only slightly (±1h) in the three fractions. The extent of C isotope fractionation
342
pertinent to the MISPE cleanup step calculated with eq. 4, ∆13 C, was negligible regardless of the
343
sample matrix (−0.6 ± 1.4h and 0.1 ± 1.1h, Table 2). By contrast, δ15 N values of 1H-BT varied
344
substantially (up to 13h) between the three sample fractions F1 to F3 . Due to the almost complete
345
recovery of the analyte after the elution with methanol, MISPE induced N isotope fractionation, 19
ACS Paragon Plus Environment
Analytical Chemistry
εN = -5.0 ± 2.2 ‰
-18 -20
15
δ Ni,MIP (‰)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 20 of 31
-22 -24 -26 -28 1.0
0.8
0.6
0.4
0.2
0.0
c/c0
Figure 4. N isotope fractionation of 1H -BT through stepwise elution from the MIP. δ15 Ni MIP (±1σ ) stand for the calculated N isotope signatures of 1H -BT remaining on the MIP after elution of different 1H -BT fractions. c/c0 is the fraction of 1H -BT remaining associated with the MIP. The data was generated by using information from Table 2 and eqs. 6 and 7. The solid line was obtained through non-linear regression of eq. 5 with an N of −5.0 ± 2.2h.
346
∆15 N is, nevertheless, negligible within uncertainty in both sample matrices (−1.4 ± 1.3h and
347
−0.2 ± 1.1h).
348
These observations suggests that a non-stoichiometric recovery of the analyte after the MISPE
349
cleanup step would not cause systematic C isotope fractionation whereas N isotope signatures could
350
be biassed towards more positive δ15 N values. We tentatively quantified an operational N isotope
351
enrichment factor, N , from the small data set for the MISPE cleanup step E in absence of organic
352
matrix using eq. 5. Figure 4 shows the average N isotope signatures of 1H-BT that remained
353
associated with the MIP (δ15 Ni,MIP , eq. 7) after sequential loading step F1 and F2 , respectively
354
(blue and green areas). The most enriched fraction of 1H-BT, F3 , with a δ15 N of −18.1 ± 0.9h,
355
was eluted from the MIP through the change of solvent (red area in Figure 3). The N isotope
356
fractionation trend was quantified with an N of −5.0 ± 2.2h (eq. 5) and illustrates that the 15 N-
357
containing 1H-BT is retained preferentially by the MIP. This isotopic preference corresponds to
358
an apparent 15 N kinetic isotope effect of 1.0054 ± 0.0022 (eq. 8) under the chosen experimental
20
ACS Paragon Plus Environment
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
ACS Paragon Plus Environment
21
(h) c (h) c
∆13 C ∆15 N
F1
0.0 ± 1.3 0.0 ± 1.2
1.00 ± 0.03 0.41 ± 0.02 −26.1 ± 0.5 −25.4 ± 0.3 −26.9 ± 0.2 −28.3 ± 0.1
reference F3
−0.6 ± 1.4 −1.4 ± 1.3
0.42 ± 0.02 0.16 ± 0.03 −25.8 ± 0.4 −24.9 ± 0.1 −25.6 ± 0.2 −18.1 ± 0.9
no matrix F2 n.d n.d. n.d.
F1
0.1 ± 1.1 −0.2 ± 1.1
0.74 ± 0.02 −26.4 ± 0.2 −30.5 ± 0.3
0.26 ± 0.02 −25.6 ± 0.6 −17.5 ± 0.6
river water extract F2 F3
for F1 and F2 determined from 1H-BT in dichloromethane, F3 in methanol, n.d. = not detected due to larger MIP b uncertainties are ±1σ of triplicate measurements of concentrations or isotope mass used in MISPE cleanup step; c propagated uncertainties (95% confidence intervals). ratios;
a Data
(-) b (h) b (h) b
mi δ13 C δ15 N
Fi
Table 2. Sample fractions of 1H -BT, Fi collected during the MISPE cleanup step E (Figure 1): mass fractions of 1H -BT, mi , C and N isotope signatures, δ13 C and δ15 N, and extent of C and N isotope fractionation, ∆13 C and ∆15 N derived with eq. 4. a
Page 21 of 31 Analytical Chemistry
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 22 of 31
359
conditions. The observation of N isotope fractionation in the absence of any isotope fractionation
360
for C implies that the N atoms of 1H-BT were involved in the interactions with the MIP and
361
confirms the above conclusion that the selectivity of the MIP is indeed based on H-bonding with
362
the triazole moiety of the analyte (Figure 3). Finally, this
363
recoveries during the MISPE cleanup step of ≥ 81% are sufficient to limit such shifts of δ15 N to
364
≤ 1.1h (eq. 9), that is to the typical precision of N isotope ratio measurements by GC/C/IRMS. 5
365
Application in wastewater treatment plants
366
We applied the sample preparation procedures shown in Figure 1 to determine the C and N isotope
367
signatures of 1H-BT in the influent and effluent of a WWTP, one of the frequently encountered
368
matrices of the pollutant with DOC contents of several mg C/L. 51,52 The outcome was verified
369
through a comparison the δ13 C and δ15 N values with those from 1H-BT in dishwashing detergents
370
sold in villages that discharge their sewer into the studied WWTP.
15 N
isotope effect implies that 1H-BT
371
Figure 5 shows the δ13 C and δ15 N values of 1H-BT in the influent and effluent of the WWTP
372
as well as in dishwashing detergents sold by retailers in the study area. δ13 C and δ15 N values of
373
1H-BT in dishwashing detergents span from −27h to −23h and −11h to 1h, respecitvely. δ13 C
374
cover a similar range observed previously by Spahr et al. 55 with a simplified approach whereas our
375
analysis reveals an even wider distribution of δ15 N in 1H-BT. The wastewater samples (2 in Figure 1)
376
contained substantial amounts of DOC that amounted to 7.3 and 28 mg C/L for effluent and influent,
377
respectively. The substantial interferences on C isotope ratio measurements were reduced through
378
an additional liquid-liquid extraction during step F (Figure 1). δ15 N values of 1H-BT in influent
379
and effluent samples were identical within uncertainty (−4.5 ± 0.9h vs. −5.6 ± 1.3h) whereas
380
the difference of their δ13 C was only minimal (−26.8 ± 0.2h vs. −26.1 ± 0.4h). Even though a
381
quantitative analysis of the C and N isotope signatures in the WWTP and in dishwashing detergents
382
is beyond the scope of this work, data in Figure 5 reveals some consistent trends which confirm the
383
applicability of MISPE-based procedures for CSIA. Similar C and N isotope signatures of 1H-BT
384
in influent and effluent agree with the frequently made observation that this micropollutant is not 22
ACS Paragon Plus Environment
Page 23 of 31
Migros Denner Spar average
Coop Dorfladen Volg
2 0 -2
15
δ N (‰)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
-4 -6 -8 -10 Influent Effluent
-12 -28
-26
-24
13
δ C (‰)
Figure 5. C and N isotope signatures of 1H -BT in the influent and effluent of a WWTP (coloured triangles) as well as in various solid, liquid, and gel-like dishwashing detergents sold by retailers in the study area (grey symbols). The dashed lines correspond to the arithmetic mean of δ13 C and δ15 N of 1H -BT in detergent samples. Error bars represent ±1σ . 1H -BT influent and effluent concentrations were 4.3 and 3.8 µg/L, respectively.
385
degraded during the waste water treatment process. 51,52 Moreover, those values are also within the
386
range of δ13 C and δ15 N values found in dishwashing detergents that are sold in the study area and
387
they are within less than 1h of their average values.
23
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
388
Conclusion
389
Our work illustrates that sample treatment procedures which rely on the selective cleanup of
390
enriched organic micropollutants by molecularly-imprinted polymer offer new perspectives for
391
CSIA. Extractions of organic micropollutants in large volume samples by conventional SPE are
392
an essential step to enrich adequate amounts of analytes for isotope-ratio mass spectrometry but
393
they will inevitably lead to concomitant enrichment of organic matter. This matrix can be removed
394
selectively in organic extracts with MIPs. The application of molecularly-imprinted polymers
395
for CSIA of 1H-BT especially in influent and effluent samples from WWTPs revealed, however,
396
that despite good selectivity, organic matrix interferences may become too large thus making the
397
application of MISPE impracticable. Our analysis suggest that the C-normalized concentration
398
ratio of dissolved organic carbon (DOC) to micropollutant can be a useful parameter to assess the
399
feasibility of MISPE-based sample cleanup for CSIA, particularly for measurements of C isotope
400
ratios. In fact, the C-normalized DOC:1H-BT ratio decreased systematically as a consequence of
401
the series of treatment of WWTP samples. Whereas the DOC:1H-BT ratio in the aqueous WWTP
402
influent was 11000 (1 in Figure 1), it decreased to 860 in sample 3 after SPE and 340 in sample
403
5 after MISPE. In WWTP effluent, the DOC:1H-BT ratio decreased from 3200 in the aqueous
404
effluent (1) to 190 after SPE (3) and 76 (5) after MISPE. Whereas we implemented liquid-liquid
405
extraction in order to further reduce DOC:1H-BT ratios in WWTP samples, alternative approaches
406
such as preparative HPLC on sample 5 could also be applied. 39 A systematic and quantitative
407
assessment of the acceptable ratio of matrix:micropollutant is warranted to assess the feasibility
408
of MISPE sample preparation for CSIA more generally. Such efforts will enable the application
409
of stable isotope analyses for tracking sources and transformation processes of many contaminants
410
of emerging concerns including, for example, pharmaceuticals, pesticides, and water disinfection
411
byproducts. 11,12,16,17,39
24
ACS Paragon Plus Environment
Page 24 of 31
Page 25 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
412
Acknowledgement
413
This work was supported by the Swiss National Science Foundation (grant no. CRSII2-141805).
414
We thank Stefanie Jucker, Hans-Jürgen Schindler, Samuel Derrer, Marius Wiesmann, Alessandra
415
Sodero, Tony Merle, and Aileen Melsbach for experimental support.
416
Supporting Information Available
417
Description of field site and sampling campaign, additional preparation steps for dishwashing
418
detergents, derivation of interaction enthalpies, molecular structure of analytes. This material is
419
available free of charge via the Internet at http://pubs.acs.org/.
420
References
421
(1) Elsner, M.; Imfeld, G. Curr. Opin. Biotech. 2016, 41, 60–72.
422
(2) Hofstetter, T. B.; Berg, M. TrAC-Trends Anal. Chem. 2011, 30, 618 – 627.
423
(3) Hofstetter, T. B.; Bolotin, J.; Pati, S. G.; Skarpeli-Liati, M.; Spahr, S.; Wijker, R. S. CHIMIA
424
2014, 68, 788–792.
425
(4) Schmidt, T. C.; Jochmann, M. A. Annu. Rev. Anal. Chem. 2012, 5, 133–155.
426
(5) Elsner, M.; Jochmann, M.; Hofstetter, T.; Hunkeler, D.; Bernstein, A.; Schmidt, T.; Schim-
427
428
429
melmann, A. Anal. Bioanal. Chem. 2012, 403, 2471–2491. (6) Hofstetter, T. B.; Schwarzenbach, R. P.; Bernasconi, S. M. Environ. Sci. Technol. 2008, 42, 7737–7743.
430
(7) Thullner, M.; Centler, F.; Richnow, H. H.; Fischer, A. Org. Geochem. 2012, 42, 1440 – 1460.
431
(8) Ivdra, N.; Fischer, A.; Herrero-Martín, S.; Giunta, T.; Bonifacie, M.; Richnow, H.-H. Environ.
432
Sci. Technol. 2016, 51, 446–454.
25
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
433
434
435
436
437
438
(9) Gilevska, T.; Gehre, M.; Richnow, H.-H. Journal of Pharmaceutical and Biomedical Analysis 2015, 115, 410–417. (10) Kaown, D.; Shouakar-Stash, O.; Yang, J.; Hyun, Y.; Lee, K.-K. Ground Water 2014, 52, 875–885. (11) Spahr, S.; Bolotin, J.; Schleucher, J.; Ehlers, I.; von Gunten, U.; Hofstetter, T. B. Anal. Chem. 2015, 87, 2916–2924.
439
(12) Spahr, S.; von Gunten, U.; Hofstetter, T. B. Environ. Sci. Technol. 2017, 51, 13170–13179.
440
(13) Arnold, W. A.; Bolotin, J.; von Gunten, U.; Hofstetter, T. B. Environ. Sci. Technol. 2008, 42,
441
442
443
444
445
446
447
448
449
450
451
452
453
454
455
7778–7785. (14) Hunkeler, D.; Laier, T.; Breider, F.; Jacobsen, O. S. Environ. Sci. Technol. 2012, 46, 6096– 6101. (15) Wijker, R. S.; Bolotin, J.; Nishino, S. F.; Spain, J. C.; Hofstetter, T. B. Environ. Sci. Technol. 2013, 47, 6872–6883. (16) Maier, M. P.; Prasse, C.; Pati, S. G.; Nitsche, S.; Li, Z.; Radke, M.; Meyer, A. H.; Hofstetter, T. B.; Ternes, T. A.; Elsner, M. Environ. Sci. Technol. 2016, 50, 10933–10942. (17) Schürner, H. K. V.; Maier, M. P.; Eckert, D.; Brejcha, R.; Neumann, C.-C.; Stumpp, C.; Cirpka, O. A.; Elsner, M. Environ. Sci. Technol. 2016, 50, 5729–5739. (18) Badin, A.; Broholm, M. M.; Jacobsen, C. S.; Palau, J.; Dennis, P.; Hunkeler, D. J. Contam. Hydrol. 2016, 192, 1–19. (19) Rosell, M.; Gonzalez-Olmos, R.; Rohwerder, T.; Rusevova, K.; Georgi, A.; Kopinke, F.-D.; Richnow, H.-H. Environ. Sci. Technol. 2012, 46, 4757–4766. (20) Ratti, M.; Canonica, S.; McNeill, K.; Bolotin, J.; Hofstetter, T. B. Environ. Sci. Technol. 2015, 49, 9797–9806. 26
ACS Paragon Plus Environment
Page 26 of 31
Page 27 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
456
457
Analytical Chemistry
(21) Ratti, M.; Canonica, S.; McNeill, K.; Bolotin, J.; Hofstetter, T. B. Environ. Sci. Technol. 2015, 49, 12766–12773.
458
(22) Merritt, D. A.; Hayes, J. M. Anal. Chem. 1994, 66, 2336–2347.
459
(23) Krummen, M.; Hilkert, A. W.; Juchelka, D.; Duhr, A.; Schlüter, H.-J.; Pesch, R. Rapid
460
Commun. Mass Spectrom. 2004, 18, 2260–2266.
461
(24) Sessions, A. L. J. Sep. Sci. 2006, 29, 1946–61.
462
(25) Gilevska, T.; Gehre, M.; Richnow, H.-H. Anal. Chem. 2014, 86, 7252–7257.
463
(26) Passeport, E.; Landis, R.; Mundle, S. O. C.; Chu, K.; Mack, E. E.; Lutz, E. J.; Sherwood Lol-
464
465
466
467
468
lar, B. Environ. Sci. Technol. 2014, 48, 9582–9590. (27) Aelion, M. C.; Höhener, P.; Hunkeler, D.; Aravena, R. Environmental Isotopes in Biodegradation and Bioremediation; CRC Press: Boca Raton, 2010. (28) Jochmann, M. A.; Schmidt, T. C. Compound-specific Stable Isotope Analysis; The Royal Society of Chemistry, 2012.
469
(29) Horst, A.; Renpenning, J.; Richnow, H.-H.; Gehre, M. Anal. Chem. 2017, 89, 9131–9138.
470
(30) Heckel, B.; Rodríguez-Fernández, D.; Torrentó, C.; Meyer, A. H.; Palau, J.; Domènech, C.;
471
472
473
474
475
476
Rosell, M.; Soler, A.; Hunkeler, D.; Elsner, M. Anal. Chem. 2017, 89, 3411–3420. (31) Renpenning, J.; Kümmel, S.; Hitzfeld, K. L.; Schimmelmann, A.; Gehre, M. Anal. Chem. 2015, 87, 9443–9450. (32) Horst, A.; Lacrampe-Couloume, G.; Sherwood Lollar, B. Anal. Chem. 2015, 87, 10498– 10504. (33) Berg, M.; Bolotin, J.; Hofstetter, T. B. Anal. Chem. 2007, 79, 2386–93.
27
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
477
478
(34) Skarpeli-Liati, M.; Turgeon, A.; Garr, A. N.; Arnold, W. A.; Cramer, C. J.; Hofstetter, T. B. Anal. Chem. 2011, 83, 1641–1648.
479
(35) Zwank, L.; Berg, M.; Schmidt, T. C.; Haderlein, S. B. Anal. Chem. 2003, 75, 5575–5583.
480
(36) Passeport, E.; Landis, R.; Lacrampe-Couloume, G.; Lutz, E. J.; Mack, E. E.; West, K.;
481
Morgan, S.; Sherwood Lollar, B. Environ. Sci. Technol. 2016, 50, 12197–12204.
482
(37) Herrero-Martín, S.; Nijenhuis, I.; Richnow, H.-H.; Gehre, M. Anal. Chem. 2015, 87, 951–959.
483
(38) Schwarzenbach, R. P.; Escher, B. I.; Fenner, K.; Hofstetter, T. B.; Johnson, C. A.; von
484
485
486
Gunten, U.; Wehrli, B. Science 2006, 313, 1072–7. (39) Schreglmann, K.; Hoeche, M.; Steinbeiss, S.; Reinnicke, S.; Elsner, M. Anal. Bioanal. Chem. 2013, 405, 2857–2867.
487
(40) Martín-Esteban, A. TrAC Trends Anal. Chem. 2013, 45, 169–181.
488
(41) Sellergren, B. TrAC Trends Anal. Chem. 1999, 18, 164–174.
489
(42) Sellergren, B. TrAC Trends Anal. Chem. 1997, 16, 310–320.
490
(43) Sellergren, B. Angew. Chem. Int. Ed. 2000, 39, 1031–1037.
491
(44) Turiel, E.; Martín-Esteban, A. Anal. Chim. Acta 2010, 668, 87–99.
492
(45) Komiyama, M.; Takeuchi, T.; Mukawa, T.; Asanuma, H. Molecular Imprinting: From Fun-
493
494
495
damentals to Applications; Wiley-VCH Verlag GmbH & Co. KGaA, 2004. (46) Sun, Z.; Schuessler, W.; Sengl, M.; Niessner, R.; Knopp, D. Anal. Chim. Acta 2008, 620, 73–81.
496
(47) Stevenson, D. TrAC Trends Anal. Chem. 1999, 18, 154–158.
497
(48) Caro, E.; Marcé, R. M.; Borrull, F.; Cormack, P.; Sherrington, D. TrAC Trends Anal. Chem.
498
2006, 25, 143–154. 28
ACS Paragon Plus Environment
Page 28 of 31
Page 29 of 31 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
499
500
Analytical Chemistry
(49) Huntscha, S.; Singer, H. P.; McArdell, C. S.; Frank, C. E.; Hollender, J. J. Chromatogr. A 2012, 1268, 74–83.
501
(50) Hollender, J.; Zimmermann-Steffens, S. G.; Koepke, S.; Krauss, M.; McArdell, C. S.; Ort, C.;
502
Singer, H. P.; von Gunten, U.; Siegrist, H. Environ. Sci. Technol. 2009, 43, 7862–7869.
503
(51) Schymanski, E. L.; Singer, H. P.; Longrée, P.; Loos, M.; Ruff, M.; Stravs, M. A.; Ripollés Vi-
504
505
506
507
508
dal, C.; Hollender, J. Environ. Sci. Technol. 2014, 48, 1811–1818. (52) Huntscha, S.; Hofstetter, T. B.; Schymanski, E. L.; Spahr, S.; Hollender, J. Environ. Sci. Technol. 2014, 48, 4435–4443. (53) Chapuis, F.; Pichon, V.; Lanza, F.; Sellergren, B.; Hennion, M. C. J Chromatogr B 2004, 804, 93–101.
509
(54) Canonica, S.; Meunier, L.; von Gunten, U. Water Res. 2008, 42, 121–128.
510
(55) Spahr, S.; Huntscha, S.; Bolotin, J.; Maier, M. P.; Elsner, M.; Hollender, J.; Hofstetter, T. B.
511
512
513
514
515
516
517
Anal. Bioanal. Chem. 2013, 405, 2843–2856. (56) Schimmelmann, A.; Lewan, M. D.; Wintsch, R. P. Geochim. Cosmochim. Acta 1999, 63, 3751–3766. (57) Schimmelmann, A.; Albertino, A.; Sauer, P. E.; Qi, H.; Molinie, R.; Mesnard, F. Rapid Commun. Mass Spectrom. 2009, 23, 3513–3521. (58) Pati, S. G.; Kohler, H.-P. E.; Hofstetter, T. B. In Measurement and Analysis of Kinetic Isotope Effects; Harris, M. E., Anderson, V. E., Eds.; Academic Press, 2017; pp 292–329.
518
(59) Meier-Augenstein, W. J Chromatogr A 1999, 842, 351–371.
519
(60) Ricci, M. P.; Merritt, D. A.; Freeman, K. H.; Hayes, J. M. Organic Geochemistry 1994, 21,
520
561–571.
29
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
521
522
(61) Sherwood Lollar, B.; Hirschorn, S. K.; Chartrand, M. M. G.; Lacrampe-Couloume, G. Anal. Chem. 2007, 79, 3469–3475.
523
(62) Matsui, J.; Miyoshi, Y.; Doblhoff-Dier, O.; Takeuchi, T. Anal. Chem. 1995, 67, 4404–4408.
524
(63) Sellergren, B.; Shea, K. J. J. Chromatogr. A 1995, 690, 29–39.
525
(64) Laurence, C.; Berthelot, M. Perspectives in Drug Discovery and Design 2000, 18, 39–60.
526
(65) Yilmaz, E.; Schmidt, R. H.; Mosbach, K. In Molecularly Imprinted Materials; Yan, M.,
527
Ramström, O., Eds.; CRC Press, 2004; pp 25–57.
30
ACS Paragon Plus Environment
Page 30 of 31
Page 31 of 31
528
TOC art
m/z 44 (mV)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
pre-MISPE post-MISPE N
standard
N N H
600
800
1000
31
1200
ACS Paragon Plus Environment
time (sec)