Online HPLC-ESI-HRMS Method for the Analysis and Comparison of Different Dissolved Organic Matter Samples Claudia Patriarca, Jonas Bergquist, Per J. R. Sjöberg, Lars Tranvik, and Jeffrey Alistair Hawkes Environ. Sci. Technol., Just Accepted Manuscript • DOI: 10.1021/acs.est.7b04508 • Publication Date (Web): 15 Dec 2017 Downloaded from http://pubs.acs.org on December 20, 2017
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Environmental Science & Technology is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 24
Environmental Science & Technology
1
Online HPLC-ESI-HRMS Method for the Analysis
2
and Comparison of Different Dissolved Organic
3
Matter Samples
4
Claudia Patriarcaa,*, Jonas Bergquista, Per J. R. Sjöberga, Lars Tranvikb, Jeffrey A. Hawkesa
5
a
6
Sweden
7
b
8
ABSTRACT: Natural dissolved organic matter (DOM) is an ultra-complex mixture that is
9
essential to global carbon cycling, but is poorly understood because of its complexity. The
10
most powerful tool for the DOM characterization is high-resolution mass spectrometry
11
(HRMS) generally combined to direct infusion (DI) as sample introduction. Liquid
12
chromatography (LC) represents a compelling alternative to DI; however state-of-the-art
13
techniques involve only offline LC-HRMS approaches, which have important logistical
14
drawbacks that make DOM analysis more challenging. This study introduces a new method
15
based on online coupling of liquid chromatography to high resolution mass spectrometry, able
16
to overcome the disadvantages of usual approaches. It is characterized by high reproducibility
17
(%Bray-Curtis dissimilarity among replicates ≈ 2.5%), it reduces transient complexity and
18
contaminant interferences, thus increasing the signal to noise ratio (S/N), leading to the
19
identification of an overall larger number of formulas in the mixture. Moreover, the
20
application of an in silico fractionation prior to the statistical analysis allows an easy, flexible,
Department of Chemistry – BMC, Analytical Chemistry, Uppsala University, Uppsala,
Department of Ecology and Genetics, Limnology, Uppsala University, Uppsala, Sweden
1 ACS Paragon Plus Environment
Environmental Science & Technology
Page 2 of 24
21
fast and detailed comparison of DOM samples from a variety of sources with a single
22
chromatographic run.
23 24
INTRODUCTION
25
Dissolved organic matter (DOM) is one of the most heterogeneous mixtures in the
26
environment1. Despite its crucial role in the global carbon cycle, it is still poorly characterized
27
due to its extreme complexity2–6.
28
Advances in mass spectrometry (MS) technology have allowed resolution of thousands of
29
individual formulas from DOM6,7. High field strength (>7T) Fourier transform ion cyclotron
30
resonance (FTICR-) MS associated with direct infusion (DI) sample introduction is the most
31
popular technique for DOM analysis1,3,8–16, and Orbitrap technology is beginning to be
32
utilized for the same purpose17–19. The homogeneity of a natural DOM sample in direct
33
infusion experiments leads to a ionization suppression of many compounds at the same time,
34
and the absence of separation means that compounds having the same chemical formula but
35
different chemical structure are overlaid in the mass spectrum, therefore making it impossible
36
to discriminate which compound is responsible for the signal4,6.
37
Chromatography and other separation techniques facilitate the fractionation of compounds,
38
improving our ability to examine the nature of DOM5,8,9,20,21. Unfortunately, most separation
39
methods applied to DOM analysis are time consuming, as they involve fractionation and
40
collection steps with further sample treatment prior to mass spectrometric analysis. Generally,
2 ACS Paragon Plus Environment
Page 3 of 24
Environmental Science & Technology
41
researchers have aimed to isolate individual compounds from the complex natural mixture,
42
making it necessary to apply multiple separations on the material20,22. Most online, MS
43
coupled chromatographic studies of DOM to date have used low resolution mass spectrometry
44
without the ability to separate the numerous extant isobaric molecular masses at each nominal
45
mass. Moreover, few studies to date have considered using a broad chromatographic
46
separation to simply add a polarity dimension to the data before using the usual statistical
47
techniques for interpreting the data. This study describes the first online high resolution
48
method that we are aware of, and is an easy, fast, reproducible and flexible method that can be
49
adopted for deeper investigation of aquatic DOM by in silico fractionation of the data before
50
interpretation by multivariate statistics and data visualization. We applied the method to the
51
study of two terrestrial and one marine sample to demonstrate the new insights and
52
capabilities that it provides.
53
MATERIALS AND METHODS
54
Chemicals and samples
55
All reagents were of analytical or high-performance liquid chromatographic grade. Three
56
isomers with molecular formula C16H18O10 and deprotonated mass 369.08272, were selected
57
as model compounds. In this study, three samples from different natural sources were
58
considered. The first was a marine sample from the North Pacific Ocean, the second was
59
taken from a brown water lake in Sweden, and the last sample was from the reference material
60
Suwannee River Fulvic Acid (SRFA). Marine and lake samples were solid phase extracted
61
with PPL cartridges as described elsewhere23, while the reference material (SRFA) powder
62
was simply diluted before the analysis. More information about chemicals, samples location
63
and preparation are described in details in the Supporting Information.
64
Sample analysis
3 ACS Paragon Plus Environment
Environmental Science & Technology
Page 4 of 24
65
The chromatographic system was an Agilent 1100 system, equipped with a binary pump and a
66
100x2 mL well-plate autosampler. The chromatographic separation was conducted using an
67
Agilent PLRP-S poly(styrene/divinylbenzene) column (1.0x150 mm, 3 µm bead size, 100Å
68
pore size) fitted with a pre-column filter (0.5 µm, Supelco Column Saver).
69
The mass spectrometer was a LTQ-Velos-Pro Orbitrap MS (Thermo Scientific, Germany)
70
equipped with an electrospray ionization source (ESI) that was operated in negative mode.
71
In order to test the analytical reproducibility, each sample was injected three times and a
72
blank, consisting of initial mobile phase, was injected between samples to monitor and
73
minimise carry-over (Figure S1). 20 µL sample was injected and a three step gradient (Table
74
S1) consisting of A) 18 MΏ deionized water + 0.1% formic acid and B) acetonitrile were
75
used. The initial flow rate was set to 100 µL min-1 to minimize the processing time. After 2.1
76
minutes the flow rate was decreased to 50 µL min-1 until the end of the run. Directly after the
77
injection, the acetonitrile percentage was increased from 5 to 20% over 2 min and maintained
78
constant for 10 minutes followed by an increase to 40% at 13 min and held isocratic until 22
79
min. In the last step, to achieve the complete elution of the most strongly retained material,
80
the acetonitrile percentage was increased up to 90% and kept steady for 10 min. Lastly, the
81
initial mobile phase composition was restored and maintained for 10 minutes until the next
82
injection. The abovementioned gradient was delayed by 6 min (200 µL) due to the system
83
dead volume. Sample breakthrough and retention was tested at different loading volumes
84
using the model compounds.
85
The mass analyzer was externally calibrated for mass accuracy on the day of analysis using
86
the manufacturer’s guidelines and negative calibration solution. The ESI source and the
87
instrument ion optics settings were optimized as follows: ESI spray voltage: -3.0 kV; sheath
88
gas flow rate: 28; capillary temperature: 275 ºC; S-lens RF level: 68.7%. The ion optics was
89
tuned to maximize the signal of the base peak in SRFA. 4 ACS Paragon Plus Environment
Page 5 of 24
Environmental Science & Technology
90
The automatic gain control setting was used to trap 1 x 106 ions in the Orbitrap, with a
91
maximum accumulation time of 50 msec. Spectra were recorded between 150-1000 m/z at
92
resolution setting 100000 (m/∆m50%) and each spectrum was internally calibrated in lock mass
93
mode using a three expected compounds (369.0827, 269.0667 and 425.1089m/z), providing
94
suitable accuracy and precision (