Single Particle Inductively Coupled Plasma Mass Spectrometry: A Powerful Tool for Nanoanalysis Single particle inductively coupled plasma mass spectrometry is an emergent ICPMS method for detecting, characterizing, and quantifying nanoparticles. Although the number of applications reported to date is limited, the relatively simple instrumental requirements, the low number concentration detection levels attainable, and the possibility to detect both the presence of dissolved and particulate forms of an element make this methodology very promising in the nanoscience related areas. Francisco Laborda,* Eduardo Bolea, and Javier Jiménez-Lamana Group of Analytical Spectroscopy and Sensors, Institute of Environmental Sciences, University of Zaragoza, 50009 Zaragoza, Spain necessary for monitoring the presence of nanomaterials in environmental and biological media, assessing their potential impact and supporting regulations.17−19 In this context, analytical chemistry is facing new challenges by regarding nanomaterials as analytes and not just as samples. In comparison with conventional analytes, the metrology of nanomaterials involves not only their detection and quantification but also their physicochemical characterization. One of the first challenges arises from the solid-state properties of nanomaterials. In addition to their chemical composition, a number of physical properties can be used for their characterization, which includes size, shape, surface charge, or surface area. Concentration levels are another critical challenge, which are especially significant when analyzing samples at environmentally relevant concentrations, where mass concentration detection limits below parts per billion should be achieved. Finally, theses challenges have to be multiplied by the number of nanomaterials and the potential samples to be analyzed, F. Laborda which include from raw nanomaterials and industrial and consumer products to environmental and biological samples containing these nanomaterials. ingle particle inductively coupled plasma mass spectromSingle particle ICPMS can be considered one of the etry (SP-ICPMS) takes advantage of the well established innovative and emerging analytical approaches demanded by elemental technique of ICPMS but performing measurements nanostakeholders. Although it is mainly used with suspensions on a “particle by particle” basis. Single particle analysis using of nanoparticles (NP), which are nanomaterials with all three 1 ICP optical emission spectrometry was first reported in 1986. external dimensions in the size range from 1 to 100 nm, it can The methodology was initially adopted for analysis of aerosol be also applied to colloids and microparticles up to several 1−5 6 and suspensions of microparticles and airborne particles micrometers as well as other nanomaterials with external 7 and cells being implemented in ICPMS instruments from 1993 8 dimensions in these ranges, like nanoplates and nanofibers. to improve the attainable sensitivity of optical emission. Single particle ICPMS is able to provide information about Afterward, the feasibility of SP-ICPMS for analysis of colloidal the elemental chemical composition of noncarbon nanomateriand microparticle suspensions was demonstrated by Degueldre 9−13 als (carbon based nanomaterials are excluded due to the More recently, the rapid increase et al. in a series of papers. intrinsic low sensitivity of this element in ICPMS) as well as in the development, production, and use of engineered their number concentration, size, and the number size nanomaterials has led to renew the interest on single particle distribution. Because its dynamic range can be extended up ICPMS as an alternative to other available techniques for to the micrometers region, polydispersed systems as well as detection, determination, and characterization of nanomateriaggregation or agglomeration processes may be studied. In als.14−16 addition, dissolved forms of the constituent elements of the Whereas nanomaterials are playing an increasing role in nanoparticles can be detected and determined.15 Finally, the many fields, the knowledge about their potential impact on low detection limits of ICPMS make single particle-ICPMS a human health and the environment as well as the development of regulations and legislation for their control is being outpaced. It is recognized that innovative analytical approaches are Published: December 5, 2013

S

© 2013 American Chemical Society

2270

dx.doi.org/10.1021/ac402980q | Anal. Chem. 2014, 86, 2270−2278

Analytical Chemistry

Feature

at high acquisition frequencies. By using fast data acquisition systems (greater than 104−105 Hz), detailed information about the transient signal produced by each NP can be obtained (Figure 1b). However, commercial ICPMS instruments with quadrupole or single collector sector field mass spectrometers typically do not attain these high frequencies, and signals are measured as individual pulses with acquisition periods of one or more milliseconds, as it is shown in Figure 1c,d. These time scans are recorded during several seconds or minutes, and they consist of a number of spikes above a steady baseline, as it is shown in Figure 2a. Whereas each spike is due to the pack of ions from a NP, the baseline is due to the background or the presence of dissolved forms of the element measured. Raw time scans can be processed by plotting the signal intensity vs the signal intensity frequency, obtaining histograms as shown in Figure 2b, where the first distribution is due to the background and/or the presence of dissolved forms of the element measured and the second to the NPs.

suitable method for the analysis of environmental samples at environmentally relevant concentrations.



ICPMS AND NANOPARTICLES ICPMS is considered one of the most versatile elemental techniques, providing rapid multielement analysis and low detection limits for a large range of samples on a routine basis. Although samples can be analyzed in any aggregation state, liquid is the most common. Liquid samples are introduced into the ICPMS instrument by using a nebulization system, consisting of a nebulizer and a spray chamber, which produces an aerosol of polydisperse droplets. Once the droplets are into the plasma, solvent evaporates, forming solid particles, which in turn are vaporized and their elements atomized and ionized. Ions are extracted through the interface into the mass spectrometer, where they are separated according to their mass/charge ratio and detected. Alternatively, suspensions can be introduced by using monodisperse droplet generators and in-house built introduction systems.20−23 Soluble forms of an element are distributed homogenously within a solution, even at very low concentrations. Thus the mass of element entering the plasma per unit of time and traveling to the detector as ions can be considered constant, producing a “steady” signal during the reading period. By contrast, if the sample contains NPs, the element is no longer distributed homogenously, being present as discrete groups of atoms. Then, if a sufficiently dilute suspension of NPs is nebulized into the plasma, a single pack of ions will be generated when each NP is vaporized, atomized, and ionized, which results in a transient signal of less than about 0.5 ms length.24 Figure 1 shows the time-resolved signals of an element



BASIC PRINCIPLES The basic assumption behind SP-ICPMS is that each recorded pulse represents a single NP. If this assumption is true then the frequency of the pulses is directly related to the number concentration of NPs and the intensity of each pulse is proportional to the mass of element, in fact to the number of atoms, in each detected NP. Theoretical basis of single particle detection applied to ICPMS were outlined by Degueldre et al.9−13 for NP suspensions continuously introduced through conventional nebulization systems. The relationship between the signal R (ions counted per time unit) and the mass concentration of a solution of an element M (CM), which is nebulized into an ICPMS, can be expressed as25 R = K introKICPMSKMC M

(1)

Kintr (= ηnebQsam) represents the contribution from the sample introduction system, through the nebulization efficiency (ηneb) and the sample uptake rate (Qsam). KICPMS is the detection efficiency, which represents the ratio of the number of ions detected versus the number of atoms introduced into the ICP and involves the processes of ionization, sampling through the ICPMS interface, as well as the transmission through the mass spectrometer. KM (= ANAv/MM) includes the contribution from the element measured, where A is the atomic abundance of the isotope considered, NAv the Avogadro number, and MM the atomic mass of M. For suspensions of spherical, solid, and pure NPs, the elemental mass concentration CMNP is expressed as M C NP =

Figure 1. Time resolved ICPMS signals from a solution (a) and a nanoparticle suspension (b, c, and d) of the same element. Frequency of data acquisition: (a and b) × 100, (c) × 3, and (d) × 1 (not in scale).

3 4 ⎛d⎞ πρ⎜ ⎟ XNPNNP 3 ⎝2⎠

(2)

where d is the NP diameter, ρ the density of the NPs, XNP the mass fraction of the element in the NP (equal to 1 for a metallic NP), and NNP the NP number concentration. Equation 1 can be adapted for the nebulization of this NP suspension as

being introduced into the ICPMS in dissolved form (part a) or as NPs (part b). From a quantitative point of view, particles with diameters below ∼1 μm are completely vaporized in the time they spent in a plasma under standard conditions,24 allowing the same sensitivity for suspensions or solutions of the same mass concentration. Although NP suspensions behave and can be managed in ICPMS like conventional solutions, the measurement of individual NPs implies to record signals with respect to time

R = K introKICPMSKMKNPNNP

(3)

where KNP (= 4/3πρ(d/2)3XNP) includes the properties of the NPs. Single particle ICPMS involves the use of suspensions sufficiently diluted in order to detect just one NP per reading. Under such conditions, the flux of NPs reaching the plasma 2271

dx.doi.org/10.1021/ac402980q | Anal. Chem. 2014, 86, 2270−2278

Analytical Chemistry

Feature

Figure 2. (a) Time scan of a nanoparticle suspension containing dissolved forms of the element contained in the nanoparticle. (b) Pulse intensity frequency histogram of data from part a.

number concentration below 108 L−1 must be used to reduce the occurrence of two-NP events,15 although this concentration depends in last instance on the nebulization system used, through its nebulization efficiency and the sample uptake rate (eq 4).

(QNP) and hence the frequency of NPs detected ( f NP) is given by the first contribution of eq 3, where fNP = Q NP = ηnebQ samNNP

(4)

When measurements are acquired in time-resolved mode, each reading lasts for a period equal to the dwell time (tdwell). Under such conditions, and if just one NP is detected during a single reading (ηnebQsamNNPtdwell = 1), rNP (= KICPMSKMKNP) represents the total counts per reading and NP, which can be expressed with respect to the diameter of the NP as rNP =

1 πρXNPKICPMSKMd3 6



DETERMINATION OF NANOPARTICLE SIZE In single particle analysis, the detection of NPs depends on two factors: (i) their number concentration, which should be high enough to allow counting a minimum number of events, and (ii) the size or the element mass per NP, which should be large enough to generate a pulse of ions detectable by the spectrometer. If there is no limitation in the number of NPs, their detection is associated to the capability of identifying the pulses from the NPs over the baseline produced by the continuous background in a time scan like shown in Figure 2a. The smallest pulse height that can be distinguished from the background determines the smallest detectable NP mass that can be related to the smallest detectable NP size. Alternatively, NP pulses can be identified in a frequency histogram (Figure 2b) as events located at higher intensities than the background distribution. In any case, size detection limits (LODsize) can be established by25

(5)

or to the mass of M per NP (mNP): rNP = KICPMSKMmNP

(6)

Equations 4−6 summarize the fundamentals behind single particle ICPMS. Quantitative determinations of NP number concentration are based on the linear relationship between the frequency of NP events and the number concentration (eq 4); whereas, the signal intensity of the NP events is proportional to the mass of analyte per NP (eq 6) or to the third power of the NP diameter for solid, spherical, and pure NPs (eq 5), allowing the determination of analyte mass per NP and size distributions, respectively.



PRACTICAL ASPECTS When a NP suspension is introduced by conventional nebulization, NPs arrive in the plasma randomly and their flux must be low enough in order to obtain separate NP signals, as it is shown in Figure 1b. If acquisition frequencies of 1000 Hz or lower are used (dwell times of 1 ms or longer) the flux must be even lower to avoid the measurement of two or more NPs within a dwell time, as it is depicted in the second event of Figure 1d. Although the use of low dwell times (e.g., 1 ms) can be considered an useful option to prevent this problem, the risk of measuring a fraction of the whole signal is increased. These NPs partially detected are indistinguishable from the entire signal produced by a smaller NP, as it is shown in Figure 1c (NPs 4 and 5 starting from the left). The probability of more than one NP being measured within a dwell time can be estimated by Poisson statistics.24 Because this probability decreases with decreasing the dwell time and/or the flux of NPs, compromise conditions can be selected which guarantee the counting of an adequate number of one-NP events. In practice, dwell times between 3 and 10 ms are commonly used. Under such acquisition conditions, the NP

LODsize

⎛ ⎞1/3 18σB =⎜ ⎟ ⎝ πρXNPKICPMSKM ⎠

(7)

when a 3σ criterion is used, where σB is the standard deviation of the continuous background. Equation 7 shows that LODsize depends basically on the detection efficiency. Thus, the improvement in the ionization conditions for elements with low ionization potentials as well as the increase on sampling or transmission of ions through more efficient instrumental designs can reduce the current size detection limits. Table 1 summarizes reported size and mass detection limits for metal and oxide based NPs. Metal NPs over ∼20 nm can be detected by SP-ICPMS, whereas the sizes increase for oxides, depending on their stoichiometry. With respect to mass, limits of detection are about tens of attograms per NP. On the other hand, the upper size limit for particle characterization in SP-ICPMS is limited by the selective removal of large particles in the spray chamber26 or their incomplete vaporization in the plasma.24 In any case, 1−5 μm are the maximum sizes recommended, which are sufficient for 2272

dx.doi.org/10.1021/ac402980q | Anal. Chem. 2014, 86, 2270−2278

Analytical Chemistry

Feature

Table 1. Size and Element Mass Detection Limits Reported for Selected Nanoparticles Determined by SP-ICPMS nanoparticle

LODsize

LODmass

ref

U Al2O3 ZrO2 ThO2 TiO2 FeOOH (goethite)

18 nm