Was the Suspect in Contact with the Victim? An Instrumental Methods

Apr 1, 2003 - Based on their experimental data, the students are asked to state the likelihood that the suspect was in contact with the victim...
0 downloads 0 Views 52KB Size
In the Laboratory

Was the Suspect in Contact with the Victim?

W

An Instrumental Methods Experiment for the Analysis of Single Fibers Using FT-IR Microscopy Sharin Bender and Sheri J. Lillard* Department of Chemistry, University of California, Riverside, CA 92521; *[email protected]

Forensic science coupled with instrumental analysis is a logical and exciting strategy for teaching upper-level analytical chemistry (1). L. J. Kaplan’s course (Williams College, Williamstown, Massachusetts) for nonscience majors, Chemistry and Crime: From Sherlock Holmes to Modern Forensic Science (2), has demonstrated the effectiveness of this approach for lower-level students in the lecture and the laboratory. Brewer et al. (University of South Carolina) have found that their course on forensic analytical chemistry increased student interest and participation in the chemistry lecture (3). Our goal, in revitalizing our instrumental methods laboratory, is to add forensic-based experiments as a way to provide interesting examples in order to sustain interest and motivation in the upper-level laboratory. Forensic-based experiments can be easy to implement into existing laboratory formats using available instruments and usually require only minor modifications. For example, we have incorporated a new experiment into our instrumental methods course in which blood alcohol content was measured with headspace GC and the students were asked to determine whether the “suspect” had been driving under the influence (4). In addition, these experiments can be performed at all levels, as demonstrated by Elderd et al. in an introductory experiment for general chemistry students in which arson accelerants were identified using GC (5). The forensic approach requires the students to understand the implications involved with incorrect determinations (i.e., someone could be sent to jail), thus encouraging careful analytical procedures and cautious interpretations of the results. Our primary goal was to trigger student enthusiasm for the laboratory (3–5). We found that our students enjoyed applying their measurements to important forensic-based conclusions, instead of simply calculating a number. IR spectroscopy is a well-established technique that is routinely used by the FBI for polymer fiber analysis (6). The student is given a fictitious scenario of a murder case and asked to follow a procedure that is similar to the standard method used by many forensic scientists in an attempt to prove a suspect innocent or guilty. The objective is to collect and compare IR spectra of various fibers for the purpose of determining whether there is a connection between an alleged murderer and a homicide victim. The student uses an Equinox 55/S FT-IR spectrometer with a Bruker A590 microscope to analyze the fibers. Control fibers from known sources are given to the student from which spectra are taken and physical observations (e.g., color, texture) are made. Then, observations and IR spectra of fibers obtained from the victim’s clothing at the crime scene are analyzed, and both physical and spectral matches are determined by comparison with the standard fibers. A spectral match is based on peak intensity and frequency associated with functional

groups in a window of 4000 to 600 cm᎑1. From these data, the student assesses the likelihood that there was contact between the victim and the suspect. This experiment is not overly time-consuming, with preparation of the fibers taking approximately 30 min. The introduction to the instrumentation and the spectral measurements take approximately 2–3 h, depending on how many students are in a group (typically 2–3) and how smoothly their experiments are run. Thus, the entire experiment can be completed within a 4-h laboratory period. Two points to be made in this laboratory experiment are the importance of good, clean analytical technique and the critical understanding of the phrase “beyond a reasonable doubt”. When given an actual scenario, the imagination kicks in and we begin to realize that all these crimes we hear about on the news every night could happen to someone we know or maybe even ourselves. Careful and accurate measurements in the chemistry laboratory are imperative in both scientific and legal procedures that take place in determining whether someone is innocent or guilty. Contamination and mislabeling can cause misrepresentation of data and could put an innocent person in jail or set a guilty person free. The phrase “beyond a reasonable doubt” has been a dominating force within the American justice system. Attributing actual data to the prosecution or release of a suspect emphasizes the intricate relation between instrumental analysis and forensic science and how this pertains to everyday life. Experimental Procedure and Equipment Fibers were collected from a purple sweater (65% rayon, 35% nylon), a black sweater (100% wool), a black sweatshirt (50% cotton, 50% polyester), and a blue carpet (unknown composition). Gloves were worn to avoid contamination. After collection, the fibers were placed in a paper bag until used. Standard glass microscope slides were prepared by drilling a 0.25-in. hole through the center. A single fiber was isolated and mounted over the hole in the glass slide using forceps and tape. Care was taken to ensure the fiber was as flat and taut as possible to avoid interference fringes. An Equinox 55/S FT-IR spectrometer equipped with a Bruker A590 microscope and a mercury–cadmium–telluride detector was used. Spectra were taken through a 36× objective corresponding to 720× total magnification. Spectra were collected with Bruker OPUS spectroscopic software Version 2.0 over a frequency window of 4000–600 cm᎑1. Percent transmission was obtained using signal averaging over 32 scans with a resolution of 4.0 cm᎑1. A 16 K Fourier transform was collected and a Blackman Harris 3-Term apodization function with a zero filling factor of 2 was performed. The 2nd derivative method with 25 smoothing points was used for peak picking.

JChemEd.chem.wisc.edu • Vol. 80 No. 4 April 2003 • Journal of Chemical Education

437

In the Laboratory

Figure 1. IR spectrum of a single fiber obtained from a black sweatshirt composed of 50% cotton and 50% polyester. Conditions as described in experimental procedure.

Figure 2. IR spectrum of a single fiber obtained from a purple sweater composed of 65% rayon, 35% nylon. Conditions as described in experimental procedure.

Table 1. IR Peak Data from a Black Sweatshirt Fiber (Figure 1)

Table 2. IR Peak Data from a Purple Sweater Fiber (Figure 2)

Wavenumber/cm᎑1

Abs. Intensity

Rel. Intensity

Wavenumber/cm᎑1

Abs. Intensity

Rel. Intensity

3852

0.801

0.011

3851

0.840

0.019

3743

0.802

0.009

3742

0.839

0.014

3628

0.775

0.032

3626

0.791

0.048

0.010

3357

0.803

0.015

0.684

0.146

3363

0.787

2939

0.665

0.145

2937

2242

0.616

0.197

2869

0.760

0.012

1733

0.580

0.219

2358

0.736

0.062

1635

0.739

0.021

2243

0.636

0.178

1585

0.721

0.045

1733

0.612

0.153

1531

0.754

0.014

1633

0.722

0.027

1454

0.602

0.168

1585

0.702

0.043

1363

0.674

0.067

1539

0.722

0.016

1248

0.678

0.022

1454

0.606

0.213

1172

0.657

0.108

1362

0.643

0.051

902

0.740

0.018

1170

0.638

0.119

765

0.718

0.012

1072

0.701

0.053

694

0.717

0.024

667

0.654

0.011

Hazards Care must be taken when using liquid nitrogen—appropriate gloves, lab coat, and eye protection should be worn when filling the detector. No other significant hazards are present with this experiment. Results Spectra of two different fibers are shown in Figures 1 and 2, with corresponding peak tables shown in Tables 1 and 2, respectively. Functional groups were identified by matching frequencies and intensities of peaks to those of literature values. Spectra are representative of functional groups, such as alkanes or alcohols, that construct compounds that make up the fiber. The peaks are not representative of the individual compounds, such as acrylic or cotton, therefore, similar 438

peaks can be observed for fibers composed of entirely different materials. For example, the black sweatshirt composed of 50% cotton and 50% polyester contains a triple bond associated with a nitrile or an alkyne, represented by the peak at 2242 cm᎑1 with a relative intensity of 0.197, as shown in Figure 1. Likewise, the purple sweater composed of 65% rayon, 35% nylon also contains a triple bond associated with a nitrile or an alkyne, represented by the peak at 2243 cm᎑1 of 0.178 relative intensity, as can be seen in Figure 2. It is important to distinguish these as having similar functional groups although they are composed of very different compounds. Therefore, matching fibers entails interpreting the entire spectrum, not just a few peaks. Figures 3 and 4 (and corresponding Tables 3 and 4) show spectra and peak tables for the two remaining fibers, the black sweater composed of 100% wool and the blue carpet of unknown composition, respectively.

Journal of Chemical Education • Vol. 80 No. 4 April 2003 • JChemEd.chem.wisc.edu

In the Laboratory

Figure 3. IR spectrum of a single fiber obtained from a black sweater composed of 100% wool. Conditions as described in experimental procedure.

Figure 4. IR spectrum of a single fiber obtained from a blue carpet of unknown composition. Conditions as described in experimental procedure.

Table 3. IR Peak Data from a Black Sweater Fiber (Figure 3)

Table 4. IR Peak Data from a Blue Carpet Fiber (Figure 4)

Wavenumber/cm᎑1

Abs. Intensity

Rel. Intensity

Wavenumber/cm᎑1

Abs. Intensity

Rel. Intensity

3749

0.878

0.010

3302

0.442

0.287

3628

0.870

0.011

3081

0.504

0.067

3565

0.871

0.006

2931

0.447

0.152

3430

0.860

0.021

2860

0.458

0.031

3345

0.871

0.008

2362

0.679

0.030

3054

0.857

0.014

1637

0.494

0.247

2969

0.804

0.080

1559

0.503

0.037

2907

0.832

0.020

1474

0.515

0.037

2502

0.871

0.008

1373

0.533

0.042

2360

0.846

0.035

1282

0.548

0.036

2107

0.869

0.010

1200

0.569

0.030

1959

0.855

0.020

1146

0.611

0.027

1717

0.729

0.124

1043

0.653

0.036

1615

0.835

0.014

935

0.615

0.094

1578

0.814

0.035

669

0.607

0.070

1506

0.764

0.075

1456

0.774

0.042

1410

0.731

0.114

1371

0.757

0.013

1339

0.751

0.017

1307

0.741

0.041

1240

0.745

0.033

1092

0.731

0.076

1019

0.737

0.061

972

0.770

0.045

873

0.745

0.076

793

0.781

0.029

727

0.714

0.094

668

0.783

0.023

Conclusions Infrared spectroscopy has been incorporated into routine analysis used by the FBI for fiber comparison. In this experiment, students used FT-IR microscopy with a mercury–

cadmium–telluride detector to analyze and compare fibers that were (supposedly) collected at a homicide scene. The main goal is to provide hands-on experience in forensic analysis to upper-level chemistry students. The importance of proper analytical technique is portrayed by associating the experiment to a fictitious, yet realistic scenario. If the students were to imagine that someone’s life actually depended on the results of their experiment, the severity of proper technique becomes a life-threatening issue, not just a good grade. The students voiced their enthusiasm and provided creative conclusions on the fate of the victim and suspect. Creativity has been shown to increase both participation and the desire to learn about analytical and writing techniques. Also, the students were asked to state whether or not the suspect could be linked beyond a reasonable doubt to the victim. This caused the students to think on a realistic level about how instrumental analysis is incorporated into judiciary relations.

JChemEd.chem.wisc.edu • Vol. 80 No. 4 April 2003 • Journal of Chemical Education

439

In the Laboratory

Many students are concerned with what kind of career a degree in chemistry can provide. Since this course is designed for upper-level chemistry students, many are preparing to graduate and enter the workplace. This experiment presents an opportunity to explore employment options while maintaining a learning environment. Acknowledgment The authors thank Dan Borchardt for his assistance with the FT-IR instrumentation. W

Supplemental Material

Instructions for the students and notes for the instructor are available in this issue of JCE Online.

440

Literature Cited 1. Gaenssien, R. E.; Kubic, T. A.; Deslo, P. J.; Lee, H. C. J. Chem. Educ. 1985, 62, 1058–1060. 2. Kaplan, L. J. http://otis.cc.williams.edu/Chemistry/faculty.html (accessed Jan 2003). 3. Brewer, W. E.; Lambert, S. J.; Morgan, S. L.; Goode, S. R. ChemConf ’98, On-Line Conference on Chemical Education 1998, http://www.infor m.umd.edu/EdRes/Topic/Chemistr y/ ChemConference/ChemConf98/forensic/ChemConf.htm (accessed Jan 2003). 4. Zabzdyr, J. L.; Lillard, S. J. J. Chem. Educ. 2001, 78, 1225–1227. 5. Elderd, D. M.; Kildahl, N. K.; Berka, L. H. J. Chem. Educ. 1996, 73, 675–677. 6. Forensic Science Communications. April 1999, vol 1, no 1. http:// www.fbi.gov/programs/lab/fsc/backissu/april1999/houcktoc.htm (accessed Jan 2003).

Journal of Chemical Education • Vol. 80 No. 4 April 2003 • JChemEd.chem.wisc.edu