A Simple Absorption Pipet

Two distinct pressure systems are maintained with a ... pressure system, pressure is maintained ... of a small simple magnet (old type of telephone be...
2 downloads 0 Views 177KB Size
September15,1934

INDUSTRIAL AND ENGINEERING CHEMISTRY

APPARATUS Two distinct pressure systems are maintained with a pressure difference of from 4 t o 30 mm. These are in intermittent communication with one another by way of a valve operated by a magnet, controlled in turn by a HershbergHuntress (2) sulfuric acid manostat a n d t h e r m i o n i c relay. I n the lowpressure system, pressure is maintained s-6 only approximately constant b y means of a mercury manostat relay, and may be from 4 to 30 mm. lower than the pressure in the distillation system. The flask containing the valve which is in FIGURE 2 direct communication with the distillation apparatus represents the other system, in which the pressure is accurately controlled. The valve-operating mechanism enclosed in flask F-1 consists of a small simple magnet (old type of telephone bell-ringing magnet), the movable part (hinged lid) of which is covered with a strip of smooth rubber. The magnet and a long 0.125-inch (0.32-cm.) tube, t, the end of which has a capillary opening, c, are securely fastened to a base as indicated. The movable part of the magnet effects intermittent closure of the ca illary. The sulfuric acid manostat and thermionic regy, M and TR, are essentially as described by Hershberg and Huntress (B), a mercury-sealed ground-glass joint (Figure 2) being substituted for the ground-steel joint used in their apparatus. The mercury manostat (Hg-M) is similar to the sulfuric acid manostat in construction and provides for a wide range of pressures. It is rendered less sensitive than the ordinary U-type mercury manostat by making bulb b relatively large. The manometer, M-2, is a simple U-type made from Pyrexglass tubing at least 2 cm. in inside diameter, to compensate for the high viscosity of the acid (d). Sulfuric acid, of the concentration used in the sulfuric acid manostat (sp. gr. 1.71), may be used as a manometer liquid, the apparatus being properly trapped to prevent diffusion of gases or vapors to or from the acid. Melville (3) mentions the use of sulfuric acid for such a purpose. After filling, it was carefully boiled out by gentle heating, the pressure having been reduced to 1 to 2 mm. of mercury; 7.9 mm. difference in acid level corresponded to 1 mm. pressure of mercury. Flasks F-1 and F-2 are used to eliminate possible pressure fluctuations when the valve opens. It was noticed that with a pressure difference of 5 to 6 mm. in the two systems and with flask F-1 of 2-liter capacity, there was a movement of about 1 mm. in the sulfuric acid levels on opening add closing the valve. Substituting a 5-liter flask eliminated this difficulty entirely, a pressure difference of 32 mm. of mercury in the two sysFlask %-2 may be of any convenient size; 2-, 5-, and 12flasks and a 20-gallon (75.7-liter) tank gave equally good results, the only difference being somewhat more fre uent starting and stopping of the vacuum pump with the smJler vessels. T I is a trap filled with quicklime. Other traps may be used at TzrFa,etc. The entire apparatus with the exception of the pump was made portable as in the case of the Hershberg-Huntress scheme (I), by mounting i t on an instrument board with a fixed base, connections with distillation apparatus being made with rubber tubing at XI,Xz,and XI. The general principle of dual pressure control described above can be applied to some advantage even with the pump (mechanical or water-aspirator type) continuously operated. When so used the mercury manostat, Hg-M, and pump relay are eliminated. Leakage or an inadequate condensing system would manifest itself by too frequent operation of the thermionic relay, T R , and valve. OPERATION For use with a fractionating column the following sequence has been found desirable and permits the taking of fractions during the distillation without disturbing the pressure: To evacuate the system the manostat stopcock, 5-6, is opened and stopcocks 8-1, 5-3,and S-4 are turned so as to connect F-1 with F-2, and the by-pass between receivers R, and Rz is opened

387

through S-2, so that the entire system is now connected directly with F-2 and the vacuum pump. When the required pressure is reached, 8-2 is closed and 5-3 is turned so as to connect RI and Rz. At the same time the sulfuric acid manostat is adjusted so that capillary tip c of tube t inside F-1 is closed by the rubber pad as discussed above and the manostat stopcock is closed. 5-1 is turned so as to connect F-1 with R1. Evacuation of F-2 is continued until the desired pressure difference between the two systems is reached, as observed on manometers M-1 and M-3, when the mercury manostat, Hg-M, is adjusted. T o remove a fraction of the distillate from the receiver RS, the following procedure is recommended:

5-3is turned to shut off R1 from Rz, and 5-4 is opened to the atmosphere. Through 5-3, Ra is brought to atmospheric pressure, and the fraction is run out through 8-5. During this operation, the pressure in the distillation apparatus as a whole is being maintained constant. After the fraction has been taken, the pressure in Rz is reduced to that which prevails in the still, in two stages as follows: Rz is connected with F-2 through 8-4, and evacuation continued until the pressure recorded on M-3 has practically reached that of M , and 8-4is promptly turned off. Final adjustment of the pressure in Rz is accomplished by establishing communication between it and F-1 through 8-1, after which all stopcocks are turned to their original positions. SUMMARY

I n the pressure-control assembly here described, a dual pressure system is maintained with the two systems about 4 to 30 mm. apart. One vessel, in continuous communication with the distillation system, constitutes the higher pressure system, the pressure being maintained constant b y means of a valve operated by a magnet controlled b y a HershbergHuntress sulfuric acid manostat and thermionic relay. Another vessel, in continuous communication with the pump but in intermittent communication with the first vessel, constitutes the lower pressure system, the pressure being maintained only approximately constant. The pump operates only occasionally and is controlled in turn by a mercury manostat and relay. The sulfuric acid U-type manometer described is about eight times as sensitive as the ordinary mercury manometer.

LITERATURE CITED (1) Hershberg and Huntress, IND. ENG. CHEM.,Anal. Ed., 5, 144 (1933). (2) Ibid., 5,344(1933). (3) Melville, H.J., J . Chem. SOC.,1931,2509. (4) Palkin, S.,IND.ENG.CHEM.,25,95 (1933). R ~ C ~ I May T ~ D 18, 1934.

A Simple Absorption Pipet ELDONA. MEANESAND EDWARD L. NEWMAN Wichita Testing Laboratories, Wichita, Kans.

T

HE accompanying dia-

gram illustrates a carbon dioxide absorption tube for organic combustion analysis. Besides being simple in construction, it has the added advantages of being in one c o m p a c t u n i t , easy to fill, easily hung on the balance, and much more rugged than the Liebig or Geissler bulbs. V e r y s a t i s f a c t o r y results have been obtained from the use of this pipet. RECJDIYED July 7, 1934.

SOOP

LIME TUBE

*

I I