Donnan Failure of Ion-Selective Electrodes with ... - ACS Publications

Nov 16, 2015 - to Donnan failure exhibited by hydrophilic ion exchange ... The absence of Donnan failure caused by hydrophobic co-ions may be of ...
0 downloads 0 Views 745KB Size
Subscriber access provided by NATIONAL UNIV OF SINGAPORE

Article

Donnan Failure of Ion-Selective Electrodes with Hydrophilic High-Capacity Ion-Exchanger Membranes Shogo Ogawara, Jesse L. Carey, Xu U. Zou, and Philippe Buhlmann ACS Sens., Just Accepted Manuscript • DOI: 10.1021/acssensors.5b00128 • Publication Date (Web): 16 Nov 2015 Downloaded from http://pubs.acs.org on November 22, 2015

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

ACS Sensors is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 27

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Sensors

Donnan Failure of Ion-Selective Electrodes with Hydrophilic High-Capacity Ion-Exchanger Membranes Shogo Ogawara, Jesse L. Carey III, Xu U. Zou, and Philippe Bühlmann*

Department of Chemistry, University of Minnesota, 207 Pleasant Street SE, Minneapolis, Minnesota 55455, United States

* [email protected]

1 ACS Paragon Plus Environment

ACS Sensors

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Abstract Hydrophilic ion-exchanger membranes with a high ion-exchange capacity find not only numerous applications in ion separations but have also interesting properties when used as sensing membranes of ion-selective electrodes (ISEs). As pointed out in the literature, the hydrophilic nature of these membranes may reduce biofouling of ISE sensing membranes as caused by electrically neutral, hydrophobic interferents. This work shows that hydrophilic high-capacity ion exchanger membranes are more resistant to Donnan failure (i.e., the limitation of the upper detection limit by co-ion transfer into the sensing membrane) than both hydrophobic ionophore-doped and hydrophobic ionophore-free ion-exchanger membranes. Nernstian responses of hydrophilic anion-exchanger membranes were found for anions as large as 2.0 nm, in spite of the crosslinked nature of the anion exchanger that was used. This shows that the high resistance of hydrophilic anion-exchangers to Donnan failure caused by cations such as tetrabutylammonium is not the result of size exclusion. For typical ions, the hydrophilicity of the anion exchanger does not play a decisive factor either. Instead, the excellent resistance to Donnan failure exhibited by hydrophilic ion exchange membranes is primarily caused by the high activity of exchangeable ions in the ion exchanger phase, which disfavors partitioning of ions of opposite charge (along with target ions) from samples into these sensing membranes. The absence of Donnan failure caused by hydrophobic co-ions may be of substantial benefit for measurements in biological samples.

Keywords: Ion-selective electrodes; Hydrophilic ion-exchange membrane; Donnan effects; Donnan failure; size exclusion

2 ACS Paragon Plus Environment

Page 2 of 27

Page 3 of 27

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Sensors

The majority of measurements with ion-selective electrodes (ISEs)1-6 are performed in their linear (Nernstian) response range. Deviations of the EMF response from linearity at the lower or upper detection limit are usually accompanied by other problems too, such as increased response times and lower selectivities. The deviation from linearity at the upper detection limit, which occurs when samples contain hydrophobic co-ions, is known as Donnan failure.7-14 The resistance of ISEs to Donnan failure is important, especially when ISEs are used in samples that contain organic hydrophobic co-ions, as it is often the case for samples of biological origin. Different approaches have been described in the literature to reduce interferences from electrically charged and neutral, lipophilic interferents. Fluorous ISE membranes15-17 were introduced because fluorous phases are both hydrophobic and lipophobic, and exhibit particularly high selectivities, in part because of the poor solvation of interferents by the fluorous membrane matrix.18-20 More recently, it was pointed out that ISEs with hydrophilic, polymeric ion-exchanger membranes suffer less from interferences by lipophilic ions of the same charge sign as the analyte than ISEs with conventional hydrophobic (i.e., lipophilic) sensing membranes.21 It was explained that the transfer of hydrophilic ions from an aqueous sample into a water-swelled hydrophilic ion-exchanger membrane does not require the dehydratization of these ions and, therefore, does not result in a selectivity for lipophilic over hydrophilic ions.21 The same can also be said about potentiometric sensors based on water-filled nanopores.10,22 which have ionic sites and ionophores covalently attached to the nanopore walls. From a fundamental point of view, Donnan failure is a well-understood phenomenon that can be interpreted in two complementary and equally valid ways. On the one hand, one may focus on the ion with a charge sign opposite to the analyte ion, typically referred to as the co-ion. In this view, the failure of co-ion exclusion from the sensing membrane, i.e., Donnan failure, is favored by any chemical interaction that lowers the chemical potential of the co-ion in the

3 ACS Paragon Plus Environment

ACS Sensors

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

sensing membrane, such as solvation and ion pair formation, and it is disfavored by the electrical potential difference across the sample/membrane interface. Using this explanation, the extent of Donnan failure can be discussed quantitatively on the basis of the electrochemical potential of the co-ion, which includes both thermodynamic and electrostatic components.9 The latter are directly correlated to the surface charge density,8,10 which has to reach a minimum value to permit permselectivity, e.g., of a nanopore,10,23,24 or result in the failure of a liquid junction based on a nanoporous glass plug.25 Alternatively, Donnan failure may be discussed in terms of partitioning of salts consisting of the analyte ion along with a co-ion between the aqueous sample and the sensing membrane. As long as the concentration of such salts in the sample is low, a Nernstian response to the target ion is observed, and the concentration of the co-ion in the sensing membrane is very low in comparison to the analyte ion. Donnan failure is reached when the transfer of the salt consisting of the analyte ion and a co-ion into the sensing has reached a level that the ratio of these two ions in the sensing membrane approaches their ratio in the salt9,26 (and, concomitantly, the analyte ion concentration in the sensing membrane starts to exceed substantially the ion exchanger capacity of the sensing membrane). While Donnan failure as a phenomenon is well understood, a comparison of the extent of Donnan failure of different types of ISE membranes has not been reported. We were interested in such a comparison, in particular in view of the promise of hydrophilic high-capacity ion-exchanger (HHCIE) membranes for measurements in biological samples, but also because of our recent interest in using ISEs with HHCIE membranes to monitor nanoparticle syntheses.27 For a better understanding of the advantages of ISEs with HHCIE membranes, we compare here the Donnan failure characteristics of a typical commercial HHCIE membrane with those of ionophore-free28 and ionophore-doped hydrophobic ISE membranes. The widely

4 ACS Paragon Plus Environment

Page 4 of 27

Page 5 of 27

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

ACS Sensors

used plasticized poly(vinylchloride) (PVC) was chosen in this work as the membrane matrix for the hydrophobic ISE membranes, which is why the ion-exchanger and ionophore-free membranes are referred to in the following as PVC/IE and PVC/ionophore membranes, respectively. The high hydrophilicity of the HHCIE membranes used in this work is highlighted by the fact that they take up 11.5 wt % water at T = 25 ℃, which is two orders of magnitude more than the water content of 0.1 wt % reported for plasticized PVC membranes.29 Undoubtedly, the this high hydrophilicity is to a large extent the result of the high ion exchange capacity of the HHCIE membranes, which in this study exceeded the ion exchange capacity of the ionophore-free ion exchanger membranes with a plasticized PVC matrix by a factor of 16. We also tested whether size exclusion has an effect on the performance of ISEs with the HHCIE membranes.

Experimental Section Materials All tetraalkylammonium salts were of >98%, purity and were used as received. Sample solutions were prepared using deionized and charcoal-treated water (0.18 MΩ cm specific resistance) obtained with a Milli-Q PLUS reagent grade water system (Millipore, Bedford, MA, USA). High molecular weight poly(vinyl chloride) (PVC), 2-nitrophenyl octyl ether (o-NPOE), tridodecylmethylammonium

chloride

(TDDMACl),

and

potassium

tetrakis(4-chlorophenyl)borate (KTpClPB) were purchased from Fluka (Buchs, Switzerland), 5,10,15,20-tetraphenyl-21H,23H-porphine manganese(III) chloride (Mn(III)TPPCl, Figure 1) from

Sigma

Aldrich

(St.

Louis,

MO,

USA),

and

disodium

4-hydroxy-3-{(4-sulfo-1-naphthalenyl)azo}-1-naphthalenesulfonate (carmoisine, >95% purity; see Figure 2) from Pfaltz & Bauer (W. Chester, PA, USA).

5 ACS Paragon Plus Environment

ACS Sensors

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Ion-selective membranes The anion exchange membranes sold by FuMA-Tech GmbH, Bietigheim-Bissingen, Germany) as “FAB anion exchange membranes”30 (a PEEK reinforced anion exchanger resin, 0.13 mm thick, sold in a form loaded with bromide as counter ion, ion exchange capacity >1.3 mol/g for a monovalent anion, electrical resistance