Automated In Vivo Nano-sensing of Breath-borne Protein Biomarkers Haoxuan Chen, Jing Li, Xiangyu Zhang, Xinyue Li, Maosheng Yao, and Gengfeng Zheng Nano Lett., Just Accepted Manuscript • DOI: 10.1021/acs.nanolett.8b01070 • Publication Date (Web): 11 Jul 2018 Downloaded from http://pubs.acs.org on July 12, 2018
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
1
Automated In Vivo Nano-sensing of Breath-borne Protein
2
Biomarkers
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37
Haoxuan Chen1, Jing Li1, Xiangyu Zhang1, Xinyue Li1, Maosheng Yao*,1 and Gengfeng Zheng*,2 1
State Key Joint Laboratory of Environmental Simulation and Pollution Control, College of Environmental Sciences and Engineering, Peking University, Beijing 100871, China 2 Laboratory of Advanced Materials, Department of Chemistry and State Key Laboratory of Medical Neurobiology, Fudan University, Shanghai 200438, China
Revision submitted to Nano Letters
* Corresponding Author: Maosheng Yao, Email:
[email protected], Ph: +86 01062767282 State Key Joint Laboratory of Environmental Simulation and Pollution Control, College of Environmental Sciences and Engineering, Peking University, Beijing 100871, China Gengfeng Zheng, Email:
[email protected] Laboratory of Advanced Materials, Department of Chemistry and Shanghai Key Laboratory of Molecular Catalysis and Innovative Materials, Fudan University, Shanghai 200032, China
June 21, 2018 Beijing, China
38 39
1
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
40
Page 2 of 37
Table of Contents Graphic
Non-invasive in vivo monitoring of physiological conditions onsingle singleprotein protein level level on IL-6 target response
Exhaled breath
Biomarkers
YYYYYYY Living rat
41
SiNW sensor device
Real-time signal
42 43 44 45
2
ACS Paragon Plus Environment
Page 3 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
46
ABSTRACT
47
Toxicology and bedside medical condition monitoring is often desired to be both
48
ultrasensitive and noninvasive. However, current biomarker analyses for these
49
purposes are mostly offline and fail to detect low marker quantities. Here, we report a
50
system called dLABer (detection of living animal’s exhaled breath biomarker) that
51
integrates living rats, breath sampling, microfluidics, and biosensors for the
52
automated tracking of breath-borne biomarkers. Our data show that dLABer could
53
selectively detect (online) and report differences (of up to 103-fold) in the levels of
54
inflammation agent interleukin-6 (IL-6) exhaled by rats injected with different
55
ambient particulate matter (PM). The dLABer system was further shown to have an
56
up to 104 higher signal-to-noise ratio than that of the enzyme-linked immunosorbent
57
assay (ELISA) when analyzing the same breath samples. In addition, both
58
blood-borne IL-6 levels analyzed via ELISA in rats injected with different PM
59
extracts and PM toxicity determined by a dithiothreitol (DTT) assay agreed well with
60
those determined by the dLABer system. Video recordings further verified that rats
61
exposed to PM with higher toxicity (according to a DTT assay and as revealed by
62
dLABer) appeared to be less physically active. All the data presented here suggest that
63
the dLABer system is capable of real-time, noninvasive monitoring of breath-borne
64
biomarkers with ultrasensitivity. The dLABer system is expected to revolutionize
65
pollutant health effect studies and bedside disease diagnosis as well as physiological
66
condition monitoring at the single-protein level.
67 68
Keywords: dLABer, Biomarker, Particulate matter, Toxicity, Biosensor, Real time, 3
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
69 70
Disease monitoring, Physiological condition
71
1. Introduction
Page 4 of 37
72
Epidemiological studies have shown that ambient air pollution is a leading
73
contributor to the global disease burden, e.g., exposure to particulate matter (PM) with
74
a diameter of 2.5 µm (PM2.5) is described to have resulted in 4.2 million deaths and
75
103.1 million disability-adjusted life-years (DALYs) in 2015 (1, 2). For every 10 µg
76
m−3 increase in PM10, the respiratory and cardiovascular daily mortality rates were
77
estimated by the World Health Organization (WHO) to be increased by 1.3% (95% CI:
78
0.5–2.09%) and 0.9% (95% CI: 0.5–1.3%), respectively (3, 4). However, these
79
epidemiological studies strongly rely on the metric of PM mass concentration, while
80
no direct correlation has been established between health outcomes and specific PM
81
components (5-8). Typically, for air pollution health studies, people first collect air
82
samples and store them for some time at low temperature and then conduct the health
83
effect or toxicity studies by exposing the collected air samples to cells or animals
84
(4,5,7-10). PM from different sources could have very different toxicity per unit of
85
mass due to different compositions (9). On the other hand, toxicology studies use cells
86
as subjects, and the results might not apply for humans (10). Although animal-based
87
studies are more representative, biomarker analysis using current methods takes a
88
long time, including taking of blood samples, and importantly, the results might not
89
represent the health state at the time of exposure because biomarker levels evolve
90
over time (11-13). Nonetheless, biomarker analysis is widely used in toxicological
91
studies (14). For example, blood samples from animals are usually taken after 4
ACS Paragon Plus Environment
Page 5 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
92
exposure, and then, relevant biomarkers are typically analyzed offline (15, 16). This
93
practice has caused significant problems when interpreting the in vivo health effects of
94
environmental pollutants.
95
Among these health studies, exhaled breath is increasingly being used as a
96
non-invasive biomarker method compared to blood sample for exposure analysis
97
(17-19). Exhaled breath contains a large number of slightly volatile (e.g., nitric oxide,
98
carbon monoxide, hydrocarbons) and nonvolatile compounds (e.g., cytokines, lipids,
99
adenosine, histamine) that can be used to assess the respiratory and systemic health
100
status (20-22). Among the breath-borne substances, fractional exhaled nitric oxide
101
(FeNO), 8-isoprostane, malondialdehyde, interleukins, and other compounds are
102
widely used as markers of oxidative stress and airway inflammation resulting from
103
occupational and daily exposure to air pollutants (23-27). Additionally, exhaled breath
104
has been studied for early diagnosis of lung diseases, such as the use of FeNO for
105
diagnosis and treatment of asthma (28). In many other studies, breath-borne proteins
106
and various cytokines, such as interleukins, TNF-α and leptin, were shown to be
107
associated with lung cancer and can possibly be used for early-stage disease screening
108
(29-31). Technically, breath-borne volatile organic compounds (VOCs) are in the gas
109
phase and can be directly analyzed via gas chromatography-mass spectrometry
110
(GC-MS). Many studies have addressed analysis of breath-borne VOCs from living
111
animal subjects using GC-MS or modified carbon nanotubes (32-34). However, with
112
regard to nonvolatile organic compounds (e.g., proteins, nucleic acids, and viruses),
113
current methods, such as ELISA, for measuring biomarkers take a long time, typically 5
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 37
114
up to several hours. Recently, the silicon nanowire field effect transistor (SiNW FET)
115
has emerged as a rapid and ultrasensitive tool in the nanoscience field and has already
116
been used to detect many biomolecules, including small molecules, proteins, nucleic
117
acids, viruses, and others (35-40). When detecting biological species, including
118
disease protein biomarkers, the SiNW FET was shown to have a sensitivity up to
119
several orders of magnitude greater than ELISA (41-44). In our previous work, we
120
developed a real-time airborne virus monitoring system (GREATpa) by integrating air
121
sampling, microfluidics and an SiNW sensor device (45). The system has been further
122
successfully applied to detect viruses in exhaled breath condensates with high
123
sensitivity (29 viruses/µl) and selectivity, and the results largely agreed with those
124
from a reverse transcription quantitative polymerase chain reaction (RT-qPCR)
125
analysis (46). Likewise, similar to virus detection, this same system provides an
126
outstanding opportunity for tracking in situ PM exposure health effects through
127
real-time monitoring of
biomarkers exhaled from animal or human subjects.
128
Here, we report a new system called dLABer (detection of living animal’s
129
exhaled breath biomarker) that integrates living animals, air sampling, and
130
microfluidics as well as a biosensor for the real-time detection of breath-borne
131
biomarkers. The system integrates technologies from various scientific fields, i.e.,
132
medical, toxicology, environmental, and nanotechnology fields. To test the system,
133
ambient PM collected from different countries were directly injected into rat blood
134
circulation using a previously reported protocol (47) from our laboratory. IL-6 levels
135
in exposed and control rat breath samples were then monitored in real time using the 6
ACS Paragon Plus Environment
Page 7 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
136
dLABer system. The IL-6 levels in collected breath samples were further measured
137
using a standard ELISA for verification and sensitivity comparisons. In addition, IL-6
138
levels in rat blood sera were also measured. As a confirmation step, the toxicities of
139
PM samples from different cities were also measured using dithiothreitol (DTT). This
140
work develops a frontier method that could potentially revolutionize pollutant health
141
effect studies as well as bedside breath-borne disease diagnosis and monitoring.
142 143
2. Materials and methods
144
2.1 Detection of living animal’s exhaled breath biomarker (dLABer) system
145
In this work, we developed an online PM toxicity analysis system named dLABer,
146
as shown in Figure 1 and Supporting Information S1 (video). The system is composed
147
of three major parts: exhaled breath biomarker collection from living rats, sample
148
delivery, and a biomarker monitoring and analysis module based on a commercial
149
nanobiosensor. For exhaled breath biomarker collection, we constructed a sampler
150
based on the impinging principle that operated at 1 liter breath of air per minute. As
151
observed in Figures 1 and S1, the rat cage was made in-house with only one air inlet
152
(for supplying fresh air to the rats) and one air outlet (for sampling the biomarkers
153
exhaled by the rats). The air was driven by a pump for unidirectional flow in the cage,
154
and biomarkers exhaled by the rats were carried by the air flow and finally absorbed
155
into the collection liquid phosphate-buffered saline (PBS) in a polypropylene conical
156
tube. The conical tube was modified with an outlet in the bottom that allowed the
157
peristaltic pump to automatically transport the sample to the nanobiosensor. The 7
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 8 of 37
158
sample transport flow rate was set to 15 µl per minute and can be adjusted according
159
to experimental need. The initial volume of PBS was 10 ml, and the liquid was
160
automatically replenished by the peristaltic pump at the same speed as sample
161
delivery to maintain a constant volume in the tube during the experiment. For
162
biomarker monitoring and analysis, the commercialized SiNW FET Nanocard (Vista
163
Therapeutics, Inc., USA) was used as the nanobiosensor. Before use, the SiNW was
164
decorated in our laboratory with a rat-specific antibody obtained from an ELISA kit
165
(Becton, Dickinson and Company). When the target antigen bound to the antibody on
166
the nanowire, the conductance of the transistor changed proportionally to the antigen
167
quantity. The detection signal was then programmed using a commercial system
168
(Vista Therapeutics, Inc., USA). Detailed pictures of the biosensor chips used are
169
shown in Figure 2.
170
Exhaled breath Collection
Peristaltic pump
Detection signal
Rat cage
NanoBiosensor Signal amplifier PBS
Air
Computer Pump Biomarker collector
sample
171
Nanobiosensor
8
ACS Paragon Plus Environment
Page 9 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
172
Fig. 1. Experimental sketch and photo of the dLABer system that integrates living rats,
173
breath sampling, microfluidics and a commercial FET-based biosensor for the
174
real-time detection of breath-borne biomarkers; the airborne biomarkers were directly
175
exhaled by the exposed rats (a dLABer system video is provided in Supporting
176
Information S1).
177 178
2.2 PM sample collection and suspension preparation
179
In this work, we applied the dLABer system to real-time analysis of the toxicity
180
of ambient PM collected from different countries by monitoring the IL-6 level in the
181
exhaled breath of rats exposed to the PM via injection. PM samples from different
182
cities (Cities A, B, C and D; these four cities are from four different continents) were
183
collected through automobile air conditioning filters as described in our previous
184
work (48). Detailed analysis of the PM toxicity in the rat model will be reported
185
elsewhere. In this work, the PM samples and the rats were mainly used for testing the
186
dLABer system that was developed. A PM suspension was made by vigorously
187
vortexing 2 mg of dust per ml deionized (DI) water for 20 min at a vortex rate of 3200
188
rpm (Vortex Genie-2, Scientific Industries Co., Ltd.). The extract solution was then
189
prepared by filtering the PM suspension through a nylon filter with an average pore
190
size of 0.45 µm.
191 192 193
2.3 Rat breeding and PM injection The jugular vascular catheterization (JVC) rat model described in our previous 9
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 10 of 37
194
work (47) was used to test the dLABer system for the real-time monitoring of
195
breath-borne biomarkers. Thirty male 10-week-old Wistar rats weighing 200–240 g
196
who had undergone a jugular vein catheterization operation were purchased from
197
Beijing Vital River Laboratory Animal Technology Co., Ltd. A flexible sterile catheter
198
was embedded into the jugular vein with approximately 1 centimeter of catheter out
199
of the skin and fixed onto the back of the rat with staples. All the rats were kept in an
200
animal care facility under a natural 12 h light/12 h dark cycle and were fed a normal
201
chow diet. After 1 week of acclimation, the rats were randomly divided into 5 groups
202
(6 rats in each group) for exposure to PM from four different countries or normal
203
saline (NS) as a control. All the groups were injected with the same volume ( 1 ml) of
204
extract suspension or NS. After injection, one rat from each group was placed in the
205
dLABer system for online exhaled breath biomarker IL-6 analysis, and blood samples
206
were taken (1 ml) 1 hour after injection. The blood serum was separated immediately
207
by centrifugation at 3000 rpm for 4 min, and the serum sample was kept at -20 °C for
208
further analysis. PM extract injection and blood sampling were performed through the
209
embedded catheter using sterile syringes with 23G flat-end needles. All rats were
210
monitored by recording cameras, and the video records were used for analysis of
211
animal behavior in response to different levels of PM exposure. All animal
212
experiments were approved by the Institutional Review Board of Peking University,
213
and relevant experiments were performed in accordance with ethical standards
214
(approval # LA2017204).
215 10
ACS Paragon Plus Environment
Page 11 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
216
2.4 Real-time monitoring of breath-borne biomarkers from PM-injected rats
217
Before performing experiments, the SiNW sensor (Nanocard) was activated and
218
functionalized in our laboratory by binding an IL-6 antibody to the surface of the
219
nanowires via a two-step procedure described previously (37). First, we used a 1%
220
ethanol/ddH2O solution containing 10% v:v 3-(trimethoxysilyl)propyl aldehyde to
221
react with SiOH groups for 40 min at room temperature; thereby, aldehyde groups that
222
were used for the next step, i.e., the antibody linkage, were attached to the nanowire
223
surface. The nanowires were washed with isopropanol at 3 µl/min for 20 min and
224
dried under a gentle flow of nitrogen gas for 3-5 min until the entire surface appeared
225
to be dry. Next, the Nanocards were heated overnight (approximately 16 hours) at
226
40 °C under vacuum. Then, rat IL-6 antibody (Becton, Dickinson and Company) was
227
diluted to 100 µg/ml in 10 mM Na-PBS at pH 8.4 containing 4 mM sodium
228
cyanoborohydride and pulled by a peristaltic pump over the nanowires at 3 µl/min for
229
five hours. Unreacted aldehyde groups on the surface were passivated by reaction
230
with ethanolamine at 6 µl/ml in 10 mM Na-phosphate buffer, pH 8.4, containing 4
231
mM sodium cyanoborohydride at 3 µl/min for two hours. Finally, Na-phosphate
232
buffer was pulled over the nanowires for 45 min at 3 µl/min. A portable Vista
233
NanoBioSensor unit and NanoBioSensor software (Vista Therapeutics, Inc., USA)
234
were used for conductance signal processing. Recombinant rat IL-6 was serial diluted
235
with PBS and used as a standard. As shown in Figs. 1 and S1, the exhaled breath from
236
the rats was continuously collected using a homemade liquid sampler and transported
237
to the biosensor for real-time breath-borne biomarker analysis. 11
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 12 of 37
238 239
2.5 Measurements of IL-6 using an enzyme-linked immunosorbent assay
240
In addition, for quality control, the IL-6 levels in exhaled breath samples and
241
blood samples from the same rats were analyzed using traditional ELISA kits for rat
242
IL-6 detection (Becton, Dickinson and Company) according to the manufacturer’s
243
instructions. Briefly, solid-phase IL-6 antibody was prepared by coating microliter
244
plate wells with purified rat IL-6 antibody. IL-6 standards, diluted blood serum or
245
exhaled breath samples, detection antibodies (biotinylated), and horseradish
246
peroxidase (HRP) reagent were added to coated wells step by step to form
247
antibody-antigen-antibody-enzyme complexes. Then, 3,3’,5,5’-tetramethylbenzidine
248
(TMB) substrate solution was added after the wells were washed. The TMB substrate
249
turned blue after catalysis by HRP. The reaction was terminated by addition of a
250
sulfuric acid solution, and the suspension finally turned yellow. The color change was
251
measured spectrophotometrically at a wavelength of 450 nm, and the biomarker
252
concentrations in the samples were then determined by comparing the optical density
253
(OD) of the samples to a standard curve. In the ELISA, recombinant rat IL-6 was
254
serially diluted with diluent from the assay kit and used as a standard, and a diluent
255
assay and blood samples collected from rats injected with NS were used as controls.
256
The ELISA results were further compared with those from the dLABer system.
257 258 259
2.6 PM toxicity analysis using a DTT assay A DTT assay was used to analyze the oxidative potential of PM extract samples 12
ACS Paragon Plus Environment
Page 13 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
260
from different cities. Briefly, redox-active compounds catalyze the reduction of
261
oxygen to superoxide by DTT, which is oxidized to disulfide. The remaining thiol is
262
allowed to react with 5,5′-dithiobis-2-nitrobenzoic acid (DTNB), generating mixed
263
disulfides and 5-mercapto-2-nitrobenzoic acid (TNBA), which is then measured based
264
on its absorption at 412 nm. The measured ROS generation potentials are expressed as
265
the normalized index of oxidant generation (NIOG) according to (49). The DTT
266
analysis results for different PM samples were compared with those for breath and
267
blood samples analyzed by the dLABer system and by ELISA. In addition, the
268
water-soluble metal elements in PM extract solutions (i.e., Fe, Cu, Zn, Pb, Ni, V, Cr,
269
Mn, Co, Mo, and Cd) were analyzed via inductively coupled plasma mass
270
spectrometry (ICP-MS, Aurora M90, Bruker, Inc., Billerica, MA, USA). The PM
271
extract samples for ICP-MS analysis were diluted 10:1 with Milli-Q water. To analyze
272
PM toxicity, the bacterial community in PM samples from four cities was investigated
273
by high-throughput sequencing. Briefly, 1.5 ml of each PM suspension was prepared
274
for 16S rDNA bacterial amplicon sequencing using the Illumina Miseq 2×300 bp
275
sequencing platform (Sangon Biotech, Inc., Shanghai, China). Microbial DNA was
276
extracted from bacterial colony samples using an E.Z.N.A. Soil DNA Kit (Omega
277
Bio-tek, Norcross, GA, U.S.) according to the manufacturer's protocol. The V3–V4
278
region of the bacterial 16S ribosomal RNA genes was amplified by polymerase chain
279
reaction (94 °C for 3 min; followed by 5 cycles at 94 °C for 30 s, 45 °C for 30 s, and
280
62 °C for 30 s; 20 cycles at 94 °C for 20 s and 55 °C for 20 s; 72 °C for 30 s; and a
281
final
extension
at
72
°C
for
5
min)
using
the
primers
341
F 13
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 14 of 37
282
5′-(CCCTACACGACGCTCTTCCGATCTG(barcode)CCTACGGGNGGCWGCAG)-
283
3′
284
5′-(GACTGGAGTTCCTTGGCACCCGAGAATTCCAGACTACHVGGGTATCTAA
285
TCC)-3′. PCRs were performed in a 30 µL mixture containing 15 µL of 2× Taq master
286
mix, 1 µL of each primer (10 µM), and 10-20 ng of template DNA. The second
287
polymerase chain reaction round was carried out in the same 30 µl mixture (95 °C for
288
3 min; followed by 5 cycles at 94 °C for 20 s and 55 °C for 20 s; 72 °C for 30 s; and a
289
final extension at 72 °C for 5 min). After purification using the Agencourt AMPure
290
XP system (Beckman Instruments, Inc., USA) and quantification using a Qubit2.0
291
DNA detection kit (Life Technologies, Inc., USA), a mixture of amplicons was used
292
for sequencing on an Illumina platform performed by Sangon Biotech, Inc., Shanghai,
293
China.
and
805R
294 295
3. Statistical analysis
296
In this study, the SiNW FET conductance level data for all samples were not
297
normally distributed; thus, the differences were analyzed via Kruskal-Wallis one-way
298
analysis. Pairwise multiple comparisons (Dunn's method) were performed to analyze
299
differences between all pairs of samples. In addition, the IL-6 concentrations in the
300
blood sera of rats in different groups were analyzed via ELISA. Despite the
301
occurrence of blood clotting in the catheters, at least 5 blood serum samples were
302
obtained for all groups. Differences between groups were analyzed using an
303
independent sample t-test. Differences between NIOG values of PM from different 14
ACS Paragon Plus Environment
Page 15 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
304
cities were analyzed using one-way ANOVA. The dLABer, ELISA and DTT analysis
305
results were compared to further validate the accuracy and reliability of the dLABer
306
system. All statistical tests were performed with the statistical component of
307
SigmaPlot 12.5 software (Systat Software, Inc.), and a p value less than 0.05 indicated
308
a statistically significant difference at a confidence level of 95%.
309 310
4. Results and discussion
311
4.1 Calibration of the dLABer system using IL-6 biomarker standards
312
Recombinant rat IL-6 was serial diluted 10 times with PBS and used as a standard
313
for evaluating the performance of the nanobiosensor. PBS and standard samples of
314
different concentrations, ranging from 5×100 to 5×104 pg/ml, were delivered
315
continuously and successively via a microfluidic channel to the Nanocard at a flow
316
rate of 20 µl/min, which was controlled by the peristaltic pump. Fifteen SiNW FET
317
circuits on the Nanocard conducted electricity, and the conductivity of the circuits
318
changed in response to the amount of IL-6 attached to the SiNW FET surfaces. The
319
SiNW FET conductance data were monitored and recorded in real time using signal
320
collecting and amplifying equipment (NanoBioSensor and its software). The
321
conductance vs time data of two good sensor circuits are shown in Figure 2 (A). When
322
PBS flowed through the Nanocard, the conductance levels of SiNW FET remained at
323
approximately 9.1×10-8 S for sensor #1 and 6.5×10-8 S for sensor #2. The conductance
324
levels decreased rapidly when the Nanocard was held without liquid flow. The
325
decrease was observed again when the system switched between two standard 15
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 16 of 37
326
samples with different IL-6 concentrations. The conductance levels of SiNW FET
327
increased to 8.4×10-8 S for sensor #1 and 6.3×10-8 S for sensor #2 when detecting
328
IL-6 standard samples of 5 pg/ml, which was comparable to the levels of PBS. The
329
small difference between the 5 pg/ml IL-6 standard sample and PBS might be caused
330
by the different ionic strengths of the two solutions, because the recombinant rat IL-6
331
was reconstituted with DI water according to the manufacturer’s instructions before
332
being diluted in PBS. The difference between PBS and the standard sample of 5 pg/ml
333
was examined and found to be significant, with a p value < 0.05. As observed in
334
Figure 2 (A), IL-6 concentration levels of 5, 50, 500, 5000, and 50000 pg/ml
335
corresponded to average conductance levels of 8.4, 9.4, 9.7, 10.1, and 12.4 × 10-8 S
336
for sensor 1# and 6.3, 6.8, 7.0, 7.3, and 8.7 ×10-8 S for sensor #2. As observed in
337
Figure 2 (A), a 10-fold increase in IL-6 concentration resulted in an ~3-20% increase
338
in the conductance level. This quantitative relationship could vary with sensor chips
339
due to variations in fabrication and functionalization, and thus, the Nanocard should
340
be calibrated with standards and negative controls before use. The linear regression
341
curves for the calibration results are shown in Figure 2 (B). The regression curves
342
were shown to have reasonable linearity, with R2 values of 0.8744 and 0.8691 for
343
sensor #1 and sensor #2, respectively, indicating that the dLABer system can be used
344
for quantitative determination. As shown in Figure 2 (B), when the 50000 pg/ml IL-6
345
standard sample was delivered to the Nanocard, the conductance level increased to a
346
very high level, which was different from the conductance change induced by other
347
standard samples (5-5000 pg/ml). This might be partially due to the excessive ionic 16
ACS Paragon Plus Environment
Page 17 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
348
strength of the higher concentration standards. Overall, the calibration results showed
349
that the dLABer system has good linearity with respect to biomarker concentration
350
and the corresponding electrical conductance signal. (C)
(A)
Nanocard IL-6 Conc. pg/ml
5×104
Inlet 5×101
PBS
5×102
5×103
5×100
Outlet
BreathIL-6 analysis using Breath-borne ILantibodyantibody-coated FET (B)
(D)
With rats PBS
Exhaled breath sample PBS
Without rats PM+PBS 24 μg/ml
Selectivity test against particulate matter (PM PM) )in the air
351 352
Fig. 2. (A) Detection of rat IL-6 standards using the dLABer system. Two parallel
353
conductance signals from two independent SiNW sensors (Vista Therapeutics, Inc.,
354
USA) were selected and are shown in the figure. The corresponding IL-6
355
concentration (logarithm-transformed) in every standard sample is labeled above the
356
curve in the figure (5×100~5×104 pg/ml); (B) Linear regression curves of normalized
357
conductance and IL-6 concentration (corresponding to the two conductance curves,
358
drawn from the detection results of rat IL-6 standards); (C) Detailed photos of the
359
commercialized SiNW sensor along with the flow tubes (Vista Therapeutics, Inc.,
360
USA); (D) Selectivity test against particulate matter (PM) in the air: detection of IL-6 17
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 18 of 37
361
in exhaled breath samples and PBS buffer solution, which was used to continuously
362
collect PM in the air. An exhaled breath sample was collected from 6 healthy rats
363
inside a cage kept in a lab on the Peking University campus during hazy days for 6
364
hours. The PM concentration in PBS buffer solution was approximately 24 µg/ml
365
(calculated as follows: 6 h × 6 L/min (air sampling rate) × 110 µg/m3 (average
366
ambient PM concentration during the experiments) / 10 ml (volume of PBS buffer)).
367
Two independent SiNW sensor conductance signals are shown in the figure.
368 369
As shown in Figure 2 (D), the particulate matter collected from the air did not
370
interfere with nanowire sensing. As shown in this work (described below), the PM in
371
the air also contained many different types of bacteria; and previous studies have also
372
shown that ambient PM contains many different types of viruses (50) and allergens
373
(proteins), such as Der p 1 and Der f 1 (up to 1282 ng/m3) and endotoxin (up to 83.6
374
ng/m3) (51,52). Despite the complex composition of air samples, the conductance
375
signals from the PM, shown in Figure 2 (D), were comparable to those from PBS
376
buffer without the presence of rats. Here, our main application of the technology is on
377
ambient air, thus the non-interferences from the ambient pollutants such as proteins
378
and endotoxin, and microbial particles are very critical. As shown in Figure 2 (A), the
379
sensor responded well to different levels of IL-6. Our ELISA data using the same
380
coating antibody also corresponded well with the dLABer system data (described
381
below). Overall, the system demonstrated outstanding selectivity against many
382
ambient components. 18
ACS Paragon Plus Environment
Page 19 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
383 384
4.2 Detection of IL-6 in exhaled breath samples
385
After calibration of the nanobiosensor using standards, the dLABer system was
386
used for real-time IL-6 detection in the exhaled breath of rats injected with NS and
387
PM from different cities (A, B, C, and D), as shown in Supporting Information S1.
388
Here, for stable analysis and validation, exhaled breath was first sampled using the
389
dLABer system for 1 hour, and IL-6 in exhaled breath was collected into
390
approximately 10 ml of PBS in collection tubes. After the collection process, the
391
samples were delivered to the nanobiosensor for detection. As shown in Figure 3 (A),
392
a Nanocard was first exposed to PBS, which served as the negative control. The
393
conductance levels were 10.4 ×10-8 S for sensor #1 and 7.3 × 10-8 S for sensor #2,
394
which were higher than the levels observed when PBS was introduced into the
395
Nanocard before calibration. This is largely due to residual attachment of IL-6 from
396
the high-concentration standard sample. Although PBS served as a cleaning solution
397
to detach IL-6 from the SiNW surface, the process required a long time to allow the
398
sensor to return to its original state. Additionally, the affinity between antibody and
399
antigen (IL-6) was an important factor affecting the direction of the reversible
400
reaction. The antibody used has a high K value (chemical equilibrium constant),
401
which indicates easy association and difficult dissociation of antibody and antigen
402
(53). Antibodies with proper affinity could be helpful for improving the reusability of
403
the Nanocard in future applications. A few minutes after the conductance level of PBS
404
became stable, the exhalation samples from rats injected with PM from different cities 19
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 20 of 37
405
were delivered to the Nanocard. As observed in Figure 3 (A), the conductance levels
406
of these samples varied significantly, indicating different IL-6 levels in exhaled breath
407
induced by PM from different cities. The conductance levels were converted into the
408
IL-6 concentration in exhaled breath samples by calculation via the standard curve
409
shown in Figure 2 (B). As shown in Figure 3 (B), rats injected with NS exhibited the
410
lowest IL-6 production. As presented in Figure 3 (B), the signal-to-noise ratios (city
411
PM vs NS rat injection) of breath-borne IL-6 measured with the dLABer system
412
ranged from ~10 for City B to ~104 for City D.
20
ACS Paragon Plus Environment
Page 21 of 37
(A)
-8
Conductance, 10 S
16
City D City A
14
City C City B
12
PBS
NS
10
PBS
8
Sensor #1 Sensor #2
6 4 0
500
1000 1500 2000 2500 3000 3500 4000 4500 5000
Time, s
(B) 1e+12
Concentration of IL-6, pg/ml
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
Sensor #1 Sensor #2
1e+11 1e+10 1e+9 1e+8 1e+7 1e+6 1e+5 1e+4 1e+3 1e+2
NS
City A
City B
City C
City D
413 414
Fig. 3. (A) Detection of IL-6 in exhaled breath samples collected from rats injected
415
with normal saline (NS) and PM from different cities (Cities A-D). The exhaled
416
breath samples were collected 1 h after injection. Two independent SiNW sensor
417
conductance signals are shown in (B); IL-6 concentrations in exhaled breath samples 21
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 22 of 37
418
that correspond to the conductance results are shown in (A) and were calculated using
419
the linear regression curves of IL-6 standards presented in Figure 1.
420 421
In our previous in vivo study of the toxic effects of PM, the inflammatory marker
422
IL-6 was shown to increase remarkably in serum after injection of PM compared with
423
negative controls injected with NS, indicating that a rapid inflammatory response was
424
induced by PM exposure (47). Other studies have also reported associations between
425
short-term exposure to air pollution and an acute increase in IL-6 (54, 55). As
426
observed in Figure 3 (A), when the exhaled breath sample labeled “City A” flowed
427
into the Nanocard, the conductance level immediately increased to 13 and 11 (x 10-8 S)
428
for approximately 7-8 min for sensor #1 and sensor #2, respectively. This conductance
429
fluctuation might be caused by the continuous flow and the fact that there are always
430
some antigens associating with or falling off the antibodies on the SiNW. Rats in each
431
group were injected with the same dose of PM extract (1 ml of 2 mg/ml PM
432
suspension), but the IL-6 levels were different in the exhaled breath of rats in different
433
groups, indicating that different levels of inflammatory responses were induced and
434
the toxicities of PM from different countries are not the same, due to the different
435
components and sources. The PM from City D induced the highest level of IL-6
436
production in the exhaled breath samples (highest conductance level), followed by
437
Cities A, C, and B. These results, along with those reported in Supporting Information
438
S1, suggest that the dLABer system can detect IL-6 exhaled by rats in real time and
439
thus can distinguish different PM toxicities. The two independent sensors generated 22
ACS Paragon Plus Environment
Page 23 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
440
consistent results, and thus, the results showed good repeatability.
441 442
Consistent with the above results, as shown in Figure 4 (A) and Supporting
443
Information S2 (videos) taken 1 h after injection, rats exposed to PM from Cities A, C,
444
and D appeared to be drowsy. They remained in the corner of the cage and did not
445
move after the PM injection, while the rats in the NS (control) and City B groups
446
remained relatively active. They moved around the cage and remained alert or curious
447
to external interruptions (e.g., knocking on the cage). Similar behavioral discrepancies
448
were observed between groups of rats injected with high and low doses of PM in our
449
previous study (47), suggesting that the PM injected into the blood circulation system
450
might have caused acute health effects, possibly including nerve injury, in the rats
451
within a short time. The results of this work suggest that the effects of PM from
452
different cities differ due to the PM components and other differences. Overall, the
453
dLABer system can be used for the real-time tracking of breath-borne biomarkers in
454
rats exposed (here, via injection) to PM samples.
455 456
23
ACS Paragon Plus Environment
Nano Letters
(A)
(B) City B
City A
200
IL-6 in exhaled breath sample
150
Concentration of IL-6, pg/ml
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 24 of 37
NS
100 50 NS
City A City B City C City D
(C) 300
IL-6 in blood serum
200
100
City C
City D
0 City A
457
City C
City D
458
Fig. 4. (A) Video snapshot of rats in groups injected with PM from different cities.
459
Rats in the NS and City B groups appeared to be more active than rats in the City A, C,
460
and D groups, which produced high IL-6 levels in exhaled breath. All related videos
461
are provided in Supporting Information S1. (B) IL-6 levels in the exhaled breath
462
samples from the same rats (analyzed using the dLABer system) assessed with a
463
traditional ELISA ; (C) IL-6 levels in the blood serum of rats injected with normal
464
saline (NS) and PM from different cities measured with a traditional ELISA. The
465
values represent the average concentration of 5 or 6 rats in one group, and error bars
466
indicate standard deviations.
467 468
To further validate our method and results obtained using the dLABer system, the
469
IL-6 levels in collected exhaled breath samples (the same samples that were analyzed
470
using dLABer) and blood sera from rats were analyzed via a standard ELISA. As
471
shown in Figure 4 (B), rats injected with PM from City D produced the highest levels
472
of IL-6, followed by Cities A, C, and B. In contrast, rats injected with NS exhibited 24
ACS Paragon Plus Environment
Page 25 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
473
the lowest IL-6 levels. These results were consistent with those obtained using the
474
dLABer system. As observed in Figure 3, the dLABer system had a signal-to-noise
475
ratio of 10-104, while the ELISA, shown in Figure 4, had a ratio of ~ 1.5-4 when
476
assessing the same breath samples. In addition to the exhaled breath samples, blood
477
samples were taken 1 h after intravenous PM injection, and IL-6 levels were then
478
measured using via ELISA. The average IL-6 levels in the blood samples (after 5-fold
479
dilution) from NS control group and City B group rats were below the detection limit
480
of the ELISA kit; the average IL-6 levels and standard deviations in other groups are
481
shown in Figure 4 (C). As shown in Figure 4 (C), the average IL-6 concentration in
482
blood serum from rats in the City D group was approximately 220 pg/ml, which was
483
higher than that in serum from rats in the City A and City C groups and,
484
unsurprisingly, higher than that in NS and City B group rat sera. In general, these
485
results matched the results obtained with the dLABer system. The PM extracts were
486
intravenously injected through a catheter and induced IL-6 release as a result of
487
systemic inflammation; the IL-6 concentration can increase rapidly in the blood, as
488
revealed in our previous work (47). The results suggested that the IL-6 in blood
489
circulation can be exhaled through pulmonary blood exchange. Upon IL-6
490
accumulation during sampling, IL-6 can be detected using our dLABer system in real
491
time, as shown in S1. In addition to the real-time breath-borne IL-6 monitoring
492
capability, the dLABer system was found to be more sensitive (up to 104 times higher
493
signal-to-noise ratio) than ELISAs when detecting small differences between samples,
494
such the difference between the City B and City C groups. In this work, we collected 25
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 26 of 37
495
airborne IL-6 from a confined space occupied only by rats, as shown in Supporting
496
Information S1 (video); thus, the biomarker would be collected into the liquid even if
497
it came from a different rat-related source. Here, we did not specifically differentiate
498
between potential IL-6 sources, but the IL-6 likely primarily comes from exhaled
499
breath, with a negligible contribution from sweat (56). In theory, the exhaled breath
500
samples were the same temperature as living rats, i.e., 37 °C, but in this work, they
501
were collected into DI water that had a temperature of approximately 20 °C.
502
Accordingly, the temperatures were the same for all breath samples collected and
503
analyzed and the sensors in this work. In the future, experiments could be designed to
504
study the effects of different temperature gradients on sensor responses when
505
analyzing the same breath samples.
506 507
Here, the ELISA blood-borne biomarker results agreed with the breath-borne
508
dLABer system results, showing that PM from different cities induced different levels
509
of IL-6 production and thus presented different toxicities. PM is a complex and
510
heterogeneous mixture, and its composition varies greatly with different locations
511
because the emission sources differ (4). Among PM contents, metals are generally
512
studied and have been shown in many studies to cause adverse health effects via ROS
513
generation (57). The metal concentrations in normal saline and PM extracts from
514
different cities in this work are shown in Figure 5 (A). Except for lead (Pb), all metal
515
concentrations in the PM extracts were found to be significantly higher than those in
516
normal saline. Previous studies noted that metals bound to PM2.5 could be the 26
ACS Paragon Plus Environment
Page 27 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
517
toxicants responsible for ROS generation (induction of oxidative stress) and
518
inflammatory injury (15, 57-59). In addition, one study showed that oxidant radical
519
generation and cytokine production (IL-6 and TNF-α) in airways were significantly
520
increased after administration of metal-rich ambient PM2.5 into contralateral lung
521
segments of healthy volunteers (60). The PM from City A and City D had relatively
522
higher Fe, Zn, Mo, and Co levels, while PM from City B had relatively high Mg, Cu,
523
V, and Ni levels. These differences in metal levels might cause different degrees of
524
toxicity, while the contribution of specific elements to toxicity and their possible
525
synergistic mechanisms need to be further investigated. In another study, PM2.5
526
samples collected from a traffic site containing higher concentrations of metals, such
527
as Fe, Cu, Ni, and Mn, e.g., were shown to exhibit higher oxidative ability using a
528
DTT assay (61). The NIOGs of PM from different cities determined by the DTT assay
529
are shown in Figure 5 (B). As observed in the figure, there are significant differences
530
between the NIOGs of the PM from four cities (p < 0.05). The PM from City D
531
yielded the highest NIOG, followed by Cities C and A, and the PM from City B
532
exhibited the lowest NIOG. The difference in metal components is shown in Figure 5
533
(A) and might partially explain the different PM toxicities; however, many chemical
534
and biological materials are present in PM, and their overall contribution to PM
535
toxicity are still not well understood. For example, the biological components, known
536
as bioaerosols, have been found to induce ROS production, and thus, the PM toxicity
537
might be modulated (62). As shown in Figure 5 (C), the bacterial communities in the
538
PM samples collected here varied greatly among the four cities. Among the top 10 27
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 28 of 37
539
bacteria phyla (listed in Supporting Information S3), the proportion of gram-negative
540
bacteria in City A-D samples was 57.15, 54.07, 68.19 and 62.59, respectively.
541
Endotoxins released by these gram-negative bacteria might also contribute to the total
542
toxicity of PM (63, 64).
543
The NIOG values provided here refer to the ability of the PM to induce oxidative
544
stress and damage, and this index might be a product of numerous factors (65). The
545
proinflammatory behavior of IL-6 in response to toxic compounds and its positive
546
relationship with the level of ROS generation have been well addressed previously
547
(59). Nonetheless, aside from the fact that the PM from City C exhibited a higher
548
NIOG than expected, the DTT assay results for other cities corresponded well with
549
the breath-borne IL-6 results obtained using the dLABer and ELISA methods.
550
28
ACS Paragon Plus Environment
Page 29 of 37
.025
1000 Normal saline City A City B City C City D
100
10
1
.020
NIIOGNIOG (toxicity)
concentration, ng/ml
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
.015
.010
.005
0.000
.1 Fe Mg Cu Zn Ti Mn V
Cr Mo Co Ni Pb
City A
City B
City C
City D
(C)
551 552
Fig. 5. (A) The measured metal levels in PM suspensions used for injection
553
determined by ICP-MS; (B) Normalized index of oxidant generation (NIOG)
554
determined by the DTT assay for (PM) air samples collected from different cities. The
555
values represent averages of at least 5 samples from each city, and error bars indicate
556
the standard deviation; (C) Culturable bacterial species distribution at the genus level
557
in PM samples from different cities. Different colors in the plot represent different
558
species names corresponding to the text on the right-hand side, and the length of the
559
color block represents the relative abundance of the species. The top 15 bacterial
560
phyla in PM samples from four cities are listed in Supporting Information S3.
561 29
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 30 of 37
562
Many epidemiological studies have already provided strong evidence that PM
563
pollution is responsible for a variety of diseases, including respiratory disease,
564
cardiovascular illness, and related morbidity and mortality (1). Among these studies,
565
animal model-based toxicology studies have been used to establish cause and effect
566
relationships for specific PM components by precise control of exposure variables.
567
The method typically includes PM exposure (inhalation or tracheal instillation) and
568
subsequent biological analyses, such as biomarker and histopathological analysis.
569
Biomarkers in blood and urine are usually used in both epidemiological and
570
toxicological studies for exposure analysis or pathological analysis. Overall, these
571
studies have led to a good mechanistic understanding of PM health effects. However,
572
the results are often impacted by delayed analysis of in situ responses to exposure in
573
human or animal subjects. Biomarkers in living systems are likely to evolve over time;
574
thus, current methods might miss the intermediate health endpoints resulting from
575
pollutant exposure. Compared with other bio-fluids, exhaled breath analysis is
576
noninvasive and more convenient for human subjects Therefore, it has received
577
significant attention in not only environmental health studies but also in other
578
biological and medical studies, such as drug monitoring, pharmacokinetics research
579
(66, 67), and diagnosis of lung disease (21, 68). For example, one study showed that
580
compared with healthy children (IL-4 = 35.7±6.2 pg/ml), children with asthma
581
problems had higher IL-4 levels (53.7±4.2 pg/ml) in exhaled breath condensate, and
582
the IL-4 levels decreased to (37.5±5.6 pg/ml) after steroid treatment (69). In other
583
studies, the IL-6 level increased in the exhaled breath condensate of patients with lung 30
ACS Paragon Plus Environment
Page 31 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
584
diseases, such as obstructive sleep apnea (OSA) (70), chronic obstructive pulmonary
585
disease (COPD) (71) and lung cancer (31). Currently, most biomarker analyses, such
586
as those of interleukin in blood, urine, and exhaled breath, are performed through an
587
ELISA. However, ELISAs are time consuming, offline, and not able to track
588
biomarker dynamics in real time. In addition, their sensitivity is relatively low
589
compared with emerging nanotechnology, e.g., field-effect-transistor (FET)-based
590
methods (43). Here, we developed a system called dLABer that integrates living
591
animals, breath sampling, microfluidics, and a FET-biosensor for the real-time
592
monitoring of breath-borne biomarkers. The system is able to overcome both the
593
offline and sensitivity limitations of the current methods. Our results showed that the
594
dLABer system is able to monitor breath-borne IL-6 from rats in real time and
595
provide direct in situ evidence of the health effects of PM exposure. Here, the PM
596
exposure level and rats were selected solely to test the developed dLABer system.
597
Undoubtedly, the system could also be used to monitor breath-borne biomarkers
598
exhaled by humans in various scenarios, such as epidemiological studies. Benefiting
599
from its real-time monitoring capability and high sensitivity, the dLABer system can
600
be further developed for evaluation of therapeutic response, pharmacokinetics studies
601
and even bedside monitoring of biomarker-based disease status in the future.
602 603
Conclusions
604
Here, we developed a novel system that allows us to track in real-time
605
breath-borne biomarkers from living subjects. Our work bridges the nanotechnology 31
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 32 of 37
606
field with the environmental, medical and other engineering fields in solving air
607
pollution problems that are often thought to be extremely difficult if not impossible.
608
To test the feasibility of the system, rats injected with PM from different cities were
609
used. The dLABer system results were further validated via ELISA of the breath
610
samples collected. In addition, analysis via ELISA of the blood samples collected
611
from rats injected with different PM samples (of the same mass) and the PM toxicity
612
analysis using the DTT method generally agreed well with the dLABer system results.
613
All the ELISA and DTT assay data and the rat behavior video recordings suggest that
614
the dLABer system is capable of tracking breath-borne biomarkers in real time, with
615
ultrasensitivity, as indicated by the 104 higher signal-to-noise ratio than that of the
616
ELISA. Breath-borne biomarkers (proteins), to the best of our knowledge, have not
617
been previously analyzed using nanotechnology, such as SiNW analysis. Thus far, in
618
the air pollution health effect field, no one has taken advantage of the latest
619
cutting-edge nanotechnology for analyzing associated biomarkers, especially for
620
online analysis. Biomarker levels can evolve over time; accordingly, current methods
621
(sample collection and storage for some time, followed by analysis via ELISA or
622
another technique) could miss intermediate health conditions or outcomes due to air
623
pollution exposure. Our work develops an integrated system that can noninvasively
624
monitor the health effects of air pollution in real time or online and, importantly, is
625
able to capture the dynamics of the effect of air pollution on health. Here, rats and PM
626
exposure were selected to test the dLABer system, but in the future, the system could
627
be used along with other types of sensors to detect biomarkers in humans in various 32
ACS Paragon Plus Environment
Page 33 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
628
scenarios. The system can be used to immediately monitor breath-borne biomarkers in
629
humans for disease status monitoring or tracking the effectiveness of medications at
630
the bedside. The dLABer system developed here holds great promise for
631
revolutionizing pollutant health effect studies and bedside disease diagnosis, as well
632
as physiological condition monitoring at the single-protein level.
633 634
Associated Content
635
Supporting information
636
S1. Video of the dLABer system showing real-time breath-borne biomarker
637
monitoring
638
S2. Video clip of rats 1 hour after injection with PM from different countries
639
S3. Top 10 bacteria phyla in PM samples from four cities identified using
640
high-throughput sequencing
641
Acknowledgments
642
This study was supported by the NSFC Distinguished Young Scholars Fund
643
Awarded to M. Yao (21725701), the National Natural Science Foundation of China
644
(91543126, 21611130103, 21477003, and 41121004), and the Ministry of Science and
645
Technology (2016YFC0207102, 2015CB553401, and 2015DFG92040). A patent has
646
been filed for the dLABer system, and the authors declare no competing interests.
647 648 649 650 651 652
References 1.
Cohen AJ, et al. (2017) Estimates and 25-year trends of the global burden of disease attributable to ambient air pollution: an analysis of data from the Global Burden of Diseases Study 2015. The Lancet 389(10082):1907-1918. 33
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
653 654 655 656 657 658 659 660 661 662 663 664 665 666 667 668 669 670 671 672 673 674 675 676 677 678 679 680 681 682 683 684 685 686 687 688 689 690 691 692 693 694 695 696
2.
Page 34 of 37
Forouzanfar MH, et al. (2016) Global, regional, and national comparative risk assessment of 79 behavioural, environmental and occupational, and metabolic risks or clusters of risks, 1990-2013;2015: a systematic analysis for the Global Burden of Disease Study 2015. The Lancet 388(10053):1659-1724.
3.
WHO (2006) WHO air quality guidelines for particulate matter, ozone, nitrogen dioxide and sulfur dioxide. Global update 2005. Summary of risk assessment.
(World Health
Organization, http://apps.who.int/iris/bitstream/10665/69477/1/WHO_SDE_PHE_OEH_06.02_eng.pdf). 4.
Heal MR, Kumar P, & Harrison RM (2012) Particles, air quality, policy and health. Chem. Soc. Rev. 41(19):6606-6630.
5.
Kelly FJ & Fussell JC (2012) Size, source and chemical composition as determinants of toxicity attributable to ambient particulate matter. Atmos. Environ. 60:504-526.
6.
Rückerl R, Schneider A, Breitner S, Cyrys J, & Peters A (2011) Health effects of particulate air
7.
Cassee FR, Héroux M-E, Gerlofs-Nijland ME, & Kelly FJ (2013) Particulate matter beyond mass:
pollution: a review of epidemiological evidence. Inhal. Toxicol. 23(10):555-592. recent health evidence on the role of fractions, chemical constituents and sources of emission. Inhal. Toxicol. 25(14):802-812. 8.
Lelieveld J & Pöschl U (2017) Chemists can help to solve the air-pollution health crisis. Nature 551(7680):291-293.
9.
Steenhof M, et al. (2011) In vitro toxicity of particulate matter (PM) collected at different sites in the Netherlands is associated with PM composition, size fraction and oxidative potential the RAPTES project. Part. Fibre Toxicol. 8(1):26.
10.
Longhin E, et al. (2013) Cell cycle alterations induced by urban PM2. 5 in bronchial epithelial cells: characterization of the process and possible mechanisms involved. Part. Fibre Toxicol. 10(1):63.
11.
Helleday R, et al. (2006) Exploring the Time Dependence of Serum Clara Cell Protein as a Biomarker of Pulmonary Injury in Humans. Chest 130(3):672-675.
12.
Blomberg A, et al. (2003) Clara cell protein as a biomarker for ozone-induced lung injury in
13.
Pleil JD (2016) Breath biomarkers in toxicology. Arch. Toxicol. 90(11):2669-2682.
14.
Suhaimi NF & Jalaludin J (2015) Biomarker as a Research Tool in Linking Exposure to Air
15.
Pardo M, Shafer MM, Rudich A, Schauer JJ, & Rudich Y (2015) Single Exposure to near
humans. Eur. Respir. J. 22(6):883-888.
Particles and Respiratory Health. BioMed Research International 2015:10. Roadway Particulate Matter Leads to Confined Inflammatory and Defense Responses: Possible Role of Metals. Environ. Sci. Technol. 49(14):8777-8785. 16.
Beck-Speier I, Karg E, Behrendt H, Stoeger T, & Alessandrini F (2012) Ultrafine particles affect the balance of endogenous pro- and anti-inflammatory lipid mediators in the lung: in-vitro and in-vivo studies. Part. Fibre Toxicol. 9(1):27.
17.
Grob NM, Aytekin M, & Dweik RA (2008) Biomarkers in exhaled breath condensate: a review of collection, processing and analysis. J. Breath Res. 2(3):037004.
18.
Vereb H, Dietrich AM, Alfeeli B, & Agah M (2011) The Possibilities Will Take Your Breath Away: Breath
Analysis
for
Assessing
Environmental
Exposure.
Environ.
Sci.
Technol.
45(19):8167-8175. 19.
Cao W & Duan Y (2006) Breath Analysis: Potential for Clinical Diagnosis and Exposure 34
ACS Paragon Plus Environment
Page 35 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
697 698 699 700 701 702 703 704 705 706 707 708 709 710 711 712 713 714 715 716 717 718 719 720 721 722 723 724 725 726 727 728 729 730 731 732 733 734 735 736 737 738 739 740
Assessment. Clin. Chem. 52(5):800-811. 20.
Horváth I, Hunt J, & Barnes PJ (2005) Exhaled breath condensate: methodological recommendations and unresolved questions. Eur. Respir. J. 26(3):523-548.
21.
Horváth I, et al. (2017) A european respiratory society technical standard: Exhaled biomarkers
22.
Kim KH, Jahan SA, & Kabir E (2012) A review of breath analysis for diagnosis of human health.
in lung disease. Eur. Respir. J. 49(4). TrAC Trends in Analytical Chemistry 33:1-8. 23.
De Prins S, et al. (2014) Airway oxidative stress and inflammation markers in exhaled breath from children are linked with exposure to black carbon. Environ. Int. 73:440-446.
24.
Romieu I, et al. (2008) Exhaled breath malondialdehyde as a marker of effect of exposure to
25.
Lehtonen H, et al. (2007) Increased alveolar nitric oxide concentration and high levels of
air pollution in children with asthma. J. Allergy Clin. Immunol. 121(4):903-909.e906. leukotriene B4 and 8-isoprostane in exhaled breath condensate in patients with asbestosis. Thorax 62(7):602-607. 26.
Huang W, et al. (2012) Inflammatory and Oxidative Stress Responses of Healthy Young Adults to Changes in Air Quality during the Beijing Olympics. Am. J. Respir. Crit. Care Med. 186(11):1150-1159.
27.
Pleil JD (2008) Role of Exhaled Breath Biomarkers in Environmental Health Science. Journal of Toxicology and Environmental Health, Part B 11(8):613-629.
28.
Dweik RA, et al. (2011) An Official ATS Clinical Practice Guideline: Interpretation of Exhaled Nitric Oxide Levels (FeNO) for Clinical Applications. Am. J. Respir. Crit. Care Med. 184(5):602-615.
29.
Chan HP, Lewis C, & Thomas PS (2009) Exhaled breath analysis: Novel approach for early detection of lung cancer. Lung Cancer 63(2):164-168.
30.
Sarah AH, et al. (2016) Exhaled breath condensate for lung cancer protein analysis: a review
31.
Horváth I, Lázár Z, Gyulai N, Kollai M, & Losonczy G (2009) Exhaled biomarkers in lung cancer.
of methods and biomarkers. J. Breath Res. 10(3):034001. Eur. Respir. J. 34(1):261-275. 32.
Bayn
A, et al. (2013) Detection of Volatile Organic Compounds in
Brucella
abortus-Seropositive Bison. Anal. Chem. 85(22):11146-11152. 33.
Peled N, et al. (2012) Detection of volatile organic compounds in cattle naturally infected with
34.
Haick H, et al. (2009) Sniffing Chronic Renal Failure in Rat Model by an Array of Random
Mycobacterium bovis. Sensors Actuators B: Chem. 171-172:588-594. Networks of Single-Walled Carbon Nanotubes. ACS Nano 3(5):1258-1266. 35.
Gao Z, et al. (2007) Silicon Nanowire Arrays for Label-Free Detection of DNA. Anal. Chem. 79(9):3291-3297.
36.
Patolsky F, Zheng G, & Lieber CM (2006) Fabrication of silicon nanowire devices for ultrasensitive, label-free, real-time detection of biological and chemical species. Nat. Protoc. 1:1711-1724.
37.
Patolsky F, et al. (2004) Electrical detection of single viruses. Proc. Natl. Acad. Sci. U. S. A.
38.
Li B-R, et al. (2013) An Ultrasensitive Nanowire-Transistor Biosensor for Detecting Dopamine
101(39):14017-14022. Release from Living PC12 Cells under Hypoxic Stimulation. J. Am. Chem. Soc. 135(43):16034-16037. 35
ACS Paragon Plus Environment
Nano Letters 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
741 742 743 744 745 746 747 748 749 750 751 752 753 754 755 756 757 758 759 760 761 762 763 764 765 766 767 768 769 770 771 772 773 774 775 776 777 778 779 780 781 782 783 784
Page 36 of 37
39.
McAlpine MC, et al. (2008) Peptide−Nanowire Hybrid Materials for SelecRve Sensing of Small
40.
Ohno Y, Maehashi K, & Matsumoto K (2010) Label-Free Biosensors Based on
Molecules. J. Am. Chem. Soc. 130(29):9583-9589. Aptamer-Modified Graphene Field-Effect Transistors. J. Am. Chem. Soc. 132(51):18012-18013. 41.
Cui Y, Wei Q, Park H, & Lieber CM (2001) Nanowire Nanosensors for Highly Sensitive and Selective Detection of Biological and Chemical Species. Science 293(5533):1289-1292.
42.
Zheng G, Patolsky F, Cui Y, Wang WU, & Lieber CM (2005) Multiplexed electrical detection of
43.
Zhang A & Lieber CM (2016) Nano-Bioelectronics. Chem. Rev. 116(1):215-257.
44.
Stern E, et al. (2009) Label-free biomarker detection from whole blood. Nature
45.
Shen F, et al. (2011) Integrating Silicon Nanowire Field Effect Transistor, Microfluidics and Air
cancer markers with nanowire sensor arrays. Nat. Biotechnol. 23:1294.
Nanotechnology 5:138. Sampling Techniques For Real-Time Monitoring Biological Aerosols. Environ. Sci. Technol. 45(17):7473-7480. 46.
Shen F, et al. (2012) Rapid flu diagnosis using silicon nanowire sensor. Nano Lett. 12(7):3722-3730.
47.
Zhang X, Kang J, Chen H, Yao M, & Wang J (2018) PM2.5 Meets Blood: In vivo Damages and Immune Defense. Aerosol and Air Quality Research 18(2):456-470.
48.
Li J, et al. (2013) Characterization of biological aerosol exposure risks from automobile air
49.
Li Q, Wyatt A, & Kamens RM (2009) Oxidant generation and toxicity enhancement of
conditioning system. Environ. Sci. Technol. 47(18):10660. aged-diesel exhaust. Atmos. Environ. 43(5):1037-1042. 50.
Cao C, et al (2014) Inhalable microorganisms in beijing’s PM2.5 and PM10 pollutants during a severe smog event. Environ. Sci. Technol 48:1499-1507.
51.
Yao M, et al (2009) A comparison of airborne and dust-borne allergens and toxins collected from home, office and outdoor environments both in New Haven, United States and Nanjing, China. Aerobiologia 25:183-192.
52.
Yao M, et al (2009)
Comparison of Electrostatic Collection and Liquid Impinging Methods
when Collecting Airborne House Dust Allergens, Endotoxin and (1,3)-β-D-Glucans. J Aerosol Sci. 40: 492-502 53.
Braden BC, et al. (1995) Protein motion and lock and key complementarity in
54.
Thompson AM, et al. (2010) Baseline repeated measures from controlled human exposure
antigen-antibody reactions. Pharm. Acta Helv. 69(4):225-230. studies: associations between ambient air pollution exposure and the systemic inflammatory biomarkers IL-6 and fibrinogen. Environ. Health Perspect. 118(1):120-124. 55.
Nordenhall C, et al. (2000) Airway inflammation following exposure to diesel exhaust: a study of time kinetics using induced sputum. Eur. Respir. J. 15(6):1046-1051.
56.
Faulkner S, et al. (2014) The detection and measurement of interleukin-6 in venous and capillary blood samples, and in sweat collected at rest and during exercise. Eur. J. Appl. Physiol.
57.
114 (6): 1207–1216.
Chen LC & Lippmann M (2009) Effects of Metals within Ambient Air Particulate Matter (PM) 36
ACS Paragon Plus Environment
Page 37 of 37 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Nano Letters
785 786 787 788 789 790 791 792 793 794 795 796 797 798 799 800 801 802 803 804 805 806 807 808 809 810 811 812 813 814 815 816 817 818 819
on Human Health. Inhal. Toxicol. 21(1):1-31. 58.
Carter JD, Ghio AJ, Samet JM, & Devlin RB (1997) Cytokine Production by Human Airway Epithelial Cells after Exposure to an Air Pollution Particle Is Metal-Dependent. Toxicol. Appl. Pharmacol. 146(2):180-188.
59.
Thompson AMS, et al. (2010) Baseline repeated measures from controlled human exposure studies: Associations between ambient air pollution exposure and the systemic inflammatory biomarkers IL-6 and fibrinogen. Environ. Health Perspect. 118(1):120-124.
60.
Schaumann F, et al. (2004) Metal-rich Ambient Particles (Particulate Matter2.5) Cause Airway Inflammation in Healthy Subjects. Am. J. Respir. Crit. Care Med. 170(8):898-903.
61.
Fujitani Y, Furuyama A, Tanabe K, & Hirano S (2017) Comparison of Oxidative Abilities of PM 2.5 Collected at Traffic and Residential Sites in Japan. Contribution of Transition Metals and Primary and Secondary Aerosols. Aerosol and Air Quality Research 17(2):574-587.
62.
Samake A, et al. (2017) The unexpected role of bioaerosols in the Oxidative Potential of PM. Sci. Rep. 7(1):10978.
63.
Long CM, et al. (2001) A pilot investigation of the relative toxicity of indoor and outdoor fine particles: In vitro effects of endotoxin and other particulate properties. Environ. Health Perspect. 109(10):1019-1026.
64.
Zhang Y, Gaekwad J, Wolfert MA, & Boons G-J (2007) Modulation of Innate Immune
65.
Sameenoi Y, et al. (2012) Microfluidic Electrochemical Sensor for On-Line Monitoring of
Responses with Synthetic Lipid A Derivatives. J. Am. Chem. Soc. 129(16):5200-5216. Aerosol Oxidative Activity. J. Am. Chem. Soc. 134(25):10562-10568. 66.
Li X, et al. (2015) Drug Pharmacokinetics Determined by Real-Time Analysis of Mouse Breath. Angewandte Chemie International Edition 54(27):7815-7818.
67.
Khoubnasabjafari M, Rahimpour E, & Jouyban A (2018) Exhaled breath condensate as an
68.
Hunt J (2002) Exhaled breath condensate: An evolving tool for noninvasive evaluation of lung
alternative sample for drug monitoring. Bioanalysis 10(2):61-64. disease. J. Allergy Clin. Immunol. 110(1):28-34. 69.
Shahid SK, Kharitonov SA, Wilson NM, Bush A, & Barnes PJ (2002) Increased Interleukin-4 and Decreased Interferon-γ in Exhaled Breath Condensate of Children with Asthma. Am. J. Respir. Crit. Care Med. 165(9):1290-1293.
70.
Carpagnano GE, et al. (2002) Increased 8-Isoprostane and Interleukin-6 in Breath Condensate of Obstructive Sleep Apnea Patients. Chest 122(4):1162-1167.
71.
Bucchioni E, Kharitonov SA, Allegra L, & Barnes PJ (2003) High levels of interleukin-6 in the exhaled breath condensate of patients with COPD. Respir. Med. 97(12):1299-1302.
820 821
37