Subscriber access provided by READING UNIV
Article
Comprehensive Ion Mobility Calibration: Poly(ethylene oxide) Polymer Calibrants and General Strategies Jean R. N. Haler, Christopher Kune, Philippe Massonnet, Clothilde Comby-Zerbino, Jan Jordens, Maarten Honing, Ynze Mengerink, Johann Far, and Edwin De Pauw Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.7b02564 • Publication Date (Web): 24 Oct 2017 Downloaded from http://pubs.acs.org on October 30, 2017
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Comprehensive Ion Mobility Calibration: Poly(ethylene oxide)
2
Polymer Calibrants and General Strategies
3
Jean R. N. Haler1, Christopher Kune1, Philippe Massonnet1, Clothilde Comby-Zerbino2,
4
Jan Jordens3, Maarten Honing3, Ynze Mengerink3, Johann Far1, Edwin De Pauw1
5
1
Mass Spectrometry Laboratory, University of Liège, Quartier Agora, Allée du Six Aout
6 7
11, B-4000 Liège, Belgium 2
Institut Lumière Matière, Université de Lyon, Université Lyon 1, CNRS, 69100
8
Villeurbanne, France 3
9
DSM Resolve, Geleen, The Netherlands
10 11
*Corresponding author email address:
[email protected] 12 13
Abstract
14
Ion Mobility (IM) is now a well-established and fast analytical technique. The IM
15
hardware is constantly being improved, especially in terms of the resolving power. The
16
Drift Tube (DTIMS), the Traveling Wave (TWIMS), and the Trapped Ion Mobility
17
Spectrometry (TIMS) coupled to mass spectrometry are used to determine the Collision
18
Cross-Sections (CCS) of ions. In analytical chemistry, the CCS is approached as a
19
descriptor for ion identification and it is also used in physical chemistry for 3D structure
20
elucidation with computational chemistry support. The CCS is a physical descriptor
21
extracted from the reduced mobility (K0) measurements obtainable only from the
22
DTIMS. TWIMS and TIMS routinely require a calibration procedure to convert
23
measured physical quantities (drift time for TWIMS and elution voltage for TIMS) into
1 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
24
CCS values. This calibration is a critical step to allow inter-instrument comparisons. The
25
previous calibrating substances lead to large prediction bands and introduced rather large
26
uncertainties during the CCS determination.
27
In this paper, we introduce a new IM calibrant (CCS and K0) using singly charged
28
sodium adducts of poly(ethylene oxide) monomethyl ether (CH3O-PEO-H) for positive
29
ionization in both helium and nitrogen as drift gas. These singly charged calibrating ions
30
make it possible to determine the CCS/K0 of ions having higher charge states. The fitted
31
calibration plots exhibit larger coverage with less data scattering and significantly
32
improved prediction bands and uncertainties. The reasons for the improved CCS/K0
33
accuracy, advantages and limitations of the calibration procedures are also discussed. A
34
generalized IM calibration strategy is suggested.
35 36
Introduction
37
Ion Mobility spectrometry (IM) provides a fast electrophoretic separation of ions in the
38
gas phase. The interest of analytical sciences to couple IM separation, Mass Spectrometry
39
(MS; IM-MS) and conventional separation techniques1 (such as Liquid Chromatography,
40
LC-MS) has peaked during recent years1–11. IM-MS coupling improves the peak
41
capacity12 and raises the confidence level in the identification of compounds and in their
42
structural elucidations.
43
During a low field IM separation, the measurable quantities are either drift times (dt) or
44
elution voltages. They depend on the properties of the ions (charge, mass and shape)13
45
and on the buffer gas (pressure, temperature, nature), but they also strongly depend on the
46
IM setup and the experimental conditions. In order to compare results between different
2 /32
ACS Paragon Plus Environment
Page 2 of 32
Page 3 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
47
(assumed) low field IM-MS instruments, the raw experimental values have to be
48
converted into physical descriptors of the ions such as the Collision Cross-Section (CCS)
49
which should be independent of the IM hardware and the experimental conditions, as
50
long as the temperature or the drift gas are not changed. In practice, only Drift Tube (DT)
51
IM setups allow to directly extract these shape-related descriptors, meaning the Collision
52
Cross-Section or the reduced mobility (K0) from drift time measurements14. On other IM
53
setups (non-DT setups), a calibration procedure is usually required to convert
54
experimental values into ideally instrument-independent physical quantities of the ions,
55
i.e. CCS or K0 values15,16. Such calibrations then use reference values obtained from DT-
56
based instruments.
57
Different IM-fitted instruments such as the Traveling Wave IM (T-Wave), the DT IM,
58
the Differential IM (DMS) and the Trapped IM (TIMS) setups are commercially
59
available. Their hardware performance boundaries are constantly being pushed further17–
60
25
61
of ion cloud densities and drift voltages on resolving power are as well being optimized
62
in homemade drift tubes26–28. The improvements in resolving power of IM instruments
63
lead to better separations of closely eluted compounds and to smaller drift time
64
dispersions (i.e. a smaller IM peak width) with regards to the peak apex. While the IM
65
measured quantities gain in precision and in accuracy, the calibration processes should
66
avoid introducing uncertainties on CCS and K0 values during the calibration procedure.
67
Calibrations thus need to rely on highly reproducible and robust calibrating ions, for
68
which the CCS and K0 values are insensitive to reasonable changes in experimental
69
conditions.
, especially in terms of resolving power. Fundamental parameters such as the influence
3 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
70
Experimental IM data are related to a three-dimensional shape-dependent property of the
71
ions, reduced to a two-dimensional value. Given that IM measurements convert
72
measurable quantities (drift times, elution voltages) into CCS or K0 values related to the
73
shape of the ions, each calibrating ion needs to have one unique shape which is retained
74
in all IM(-MS) instruments and experimental conditions.
75
Here we suggest synthetic homopolymers, using poly(ethylene oxide) monomethyl ether
76
(CH3O-PEO-H) for proof of concept, as calibrating ions for CCS and K0 measurements in
77
positive ion mode with commercial instruments using N2 as a drift gas. For sake of
78
completeness, the DT-obtained values in He are also provided. The advantages of using
79
synthetic homopolymers are discussed. Their known and constant shapes when
80
increasing the polymer chain length (excluding the zones of structural rearrangements29)
81
stand in contrast to peptide or protein structures which often show multiple
82
conformations upon ion heating30,31. We show that IM calibrations using synthetic
83
homopolymers result in accurate CCS-deduced values for each benchmarked instrument
84
and decreased errors (uncertainties) on the calibrated measurements. The calibration is
85
applied to CCS and K0 measurements on the widely used Synapt G2 HDMS T-Wave
86
(Waters, UK) and the recently developed high-resolution TIMS (Bruker Daltonics, USA)
87
instruments.
88
Experimental
89
All presented data and plots compare measurements taken at room temperature in N2 as
90
drift gas. Drift Tube-obtained CCS and K0 values in N2 and in He used as drift gases are
91
given in the dedicated table in the Supporting Information (Table SI1).
92
Sample Preparation
4 /32
ACS Paragon Plus Environment
Page 4 of 32
Page 5 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
93
Poly(ethylene oxide) monomethyl ether (CH3O-PEO-H) 750 g/mol (ref. 202495) and
94
2000 g/mol (ref. 202509) samples were bought from Sigma-Aldrich. The samples were
95
diluted to 5 × 10-6 M in pure methanol (Biosolve) spiked with 10-5 M Na+ cations (NaCl
96
salt prepared in pure methanol) for injection. The calibrating substances used in this study
97
are contained in the 750 g/mol sample.
98
Ion Mobility-Mass Spectrometry: Drift Tube IM (DT)
99
The samples were infused at a flow rate of 150 µL/h into a homemade Ion Mobility-Mass
100
Spectrometer fitted with an Electrospray Ionization source (ESI). The instrument is a
101
Drift Tube-modified Maxis Impact (Bruker, Germany) Quadrupole Time-of-Flight mass
102
spectrometer (Q-ToF). The setup is described in detail elsewhere14. In brief, it contains
103
two 79 cm-long drift tube separated by a dual-stage ion funnel assembly. Both
104
extremities of both drift tubes are fitted with grids, precisely defining the drift distance.
105
After the IM separation, the ions pass through another dual funnel assembly which is
106
followed by a stacked-ring radio-frequency ion guide before entering the original transfer
107
optics of the Maxis Impact Q-ToF. The drift time measurements leading to the CCS and
108
K0 calculations reported in this article were obtained using the second drift region only.
109
The He and N2 DT reference values provided in Table SI1 were taken as 3 replica
110
measurements on 3 different days by 3 different operators. The first and last
111
measurements were spaced by 10 months. For more details on the experimental
112
parameters and on the CCS and K0 calculations, see the Supporting Information. Data
113
processing was performed using Excel 2011 and IgorPro 6.34A.
114
Ion Mobility-Mass Spectrometry: Trapped IM (TIMS)
5 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
115
For details on the TIMS-TOF (Bruker Daltonics, USA) parameters, see the Supporting
116
Information. The interpretation of the TIMS data was performed using a beta version of
117
Bruker’s DataAnalysis 5-0 software and the UIMF_Viewer software. Data processing
118
was performed using Excel 2011 and IgorPro 6.34A.
119
Ion Mobility-Mass Spectrometry: Traveling Wave IM (T-Wave)
120
For details on the T-Wave SYNAPT G2 HDMS (Waters, UK) parameters, see the
121
Supporting Information. The interpretation of the T-Wave data was performed using
122
Waters’ MassLynx 4.1 software. ATD peaks were fitted using PeakFit v.4.11 to
123
determine accurate drift times. Data processing was performed using Excel 2011 and
124
IgorPro 6.34A.
125
Calibration procedure: T-Wave
126
Ruotolo and coworkers described the Collision Cross-Section (CCS) calibration
127
procedure for Traveling-Wave-based instruments (T-Wave)15. The detailed protocol with
128
all necessary equations can be found in the Supporting Information.
129
In short, equation 1 represents the calibration curve fit. The reduced DT-measured CCS
130
values (Ω’; charge and reduced mass correction, see equation SI3) are plotted as a
131
function of the corrected drift times (dt”; see equation SI2), measured on the T-Wave
132
instrument. The two fit parameters a and b depend on the experimental conditions.
133
Ω' = a ⋅ ( dt '')
134
, Ω’ is the DT-obtained reduced CCS (see equation SI3), dt” is the corrected T-Wave-
135
obtained drift time (see equation SI2) and a and b are the fitting parameters of the
136
calibration equation depending on the experimental conditions.
137
Calibration procedure: TIMS
b
(1)
6 /32
ACS Paragon Plus Environment
Page 6 of 32
Page 7 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
138
The TIMS calibration procedure was detailed by the groups of Fernandez-Lima16 and
139
Park32. The detailed protocol with all necessary equations can be found in the Supporting
140
Information.
141
In short, equation 2 represents the calibration curve fit. The DT-measured reduced
142
mobility K0 values are plotted as a function of the inverse elution voltage (1/Ve; see
143
equation SI5 and equation SI6), obtained from the TIMS measurement. The two fit
144
parameters a and b depend on the experimental conditions (equation 2). Using equation 3
145
(derived from the Mason-Schamp equation13), the CCS values can be recalculated from
146
the K0 values.
147
1 K0 = a + b ⋅ Ve
148
, K0 is the reduced mobility, a and b are the fitting parameters of the calibration equation
149
depending on the experimental conditions, Ve is the elution voltage.
Ω = c⋅ 150
z 1 ⋅ µT K 0
(2)
(3)
151
, Ω is the Collision Cross-Section (CCS), c is a constant derived from the Mason-Schamp
152
equation (c = 18500 at 305 K), z is the charge of the ion, µ is the reduced mass of the drift
153
gas and the ion, T is the temperature (assumed to be 305 K) and K0 is the reduced
154
mobility.
155
Results and Discussion
156
The calibration process is meant to convert low field (small E/N) IM setup-dependent
157
measured drift times (or its equivalent) into invariant CCS or K0 values (if not changing
158
the temperature or the drift gas). Once the IM calibration is performed, the plot of the
159
CCS as a function of the mass (or mass-to-charge ratio; m/z) for a series of homologous 7 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
160
molecules should be identical for all IM setups. Such plots constitute the repeatability
161
tests (quality control) of the robustness of the calibration. Discrepancies could be
162
assigned to poor calibration fits originating from a temperature-dependent three-
163
dimensional structure of the calibrating substances, a strongly temperature-dependent
164
interaction between the drift gas and the calibrating substances, from a too small number
165
of calibrating points and/or from intrinsic errors in the calibration procedure due to
166
poorly defined ion trajectories. Before presenting the experimental results, the classical
167
(established) calibrating substances of the two non-DT IM setups used in this paper are
168
presented (T-Wave and TIMS).
169
Classical calibrating substances.
170
Biomolecules. The choice of the calibrating ions was up to now guided by equation 4. If
171
the CCS values of biomolecules are plotted as a function of their mass and fit with a pow
172
parameter of ~0.66, then their shapes were postulated to be globular in the gas phase15.
173
Different charge state ions were thus mixed in the calibration plots. However, when
174
fitting polymer-cation CCS evolutions, pow parameters near 0.66 can be found for non-
175
globular shapes. Figure SI1 illustrates such fits for the [PEO + 1Na+]1+ and the [PEO +
176
3Na+]3+ charge states. The globular shapes of [PEO + 1Na+]1+ (no Coulomb repulsion;
177
‘Fit 1+’ in Figure SI1) correlate with pow=0.61 whereas the elongated shapes33 of [PEO
178
+ 3Na+]3+ fit with pow=0.65 (‘Fit 3+’ in Figure SI1). The 0.66 criterion is thus not
179
sufficient to attribute shapes to ions.
180
Ω = A ⋅ mass pow
181
, Ω is the CCS, A and pow are two fitting parameters and mass represents either the mass
182
of the ion or the Degree of Polymerization (DP) of polymer ions.
(4)
8 /32
ACS Paragon Plus Environment
Page 8 of 32
Page 9 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
183
Additionally, Waters’ T-Wave and Bruker’s TIMS instruments are mainly operated using
184
N2 as drift gas. In contrast, most of the DT-obtained reference values of calibrating
185
substances are published mainly for He as drift gas. The range of calibrating substances
186
available in N2 is drastically reduced34,35.
187
Moreover, multiple conformation peaks are often observed in the ATD for one given
188
mass-to-charge ratio (m/z) of biomolecules36–39. Each of these peaks would have to be
189
annotated in literature (DT reference measurements) in order to be able to correlate the
190
reference peaks to the observed peaks in the IM setup needing calibration. However, such
191
multi-peak annotations are barely the case in literature.
192
Above all, in addition, these multi-peak ATDs obtained from the DT (reference)
193
instruments would have to be reproduced on the other IM-MS setups. Otherwise, the
194
assignment of each drift time peak is almost unachievable with a high level of
195
confidence. Even if performing such drift time assignments, the question of the
196
unambiguously invariant shape attributions to each conformation would nevertheless still
197
be unsolved. Indeed, ion temperature25,40–42 and ion dynamics43 studies showed that ion
198
history (ionization source, ion optics…) in different IM setups can differ from instrument
199
to instrument, making multi-peak ATD reproducibility for biomolecules nearly
200
impossible44.
201
Homopolypeptides such as poly(alanine), poly(glycine)34 and poly(proline)45,46 also
202
suffer from multiple conformations depending on the charge state and ion
203
temperature30,31,47.
9 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
204
All in all, the uncertainty of reproducible and invariant ATDs or CCS values (i.e. shapes)
205
of biomolecule-based calibrating substances or even of homopolypeptides could lead to
206
skewed IM values, misinterpretations and low-quality calibration fits (Figure 1.a.).
207
Figure 1
208
Agilent Tune Mix. On the TIMS instrument, the usual calibrating substances originate
209
from the Agilent Tune Mix48. They act as both a mass and a mobility calibrating
210
substance in one sample injection. Nevertheless, the limited number of the Agilent Tune
211
Mix48 calibrating points provides a non-negligible downside due to the operational mode
212
of the TIMS itself. Even when using long TIMS voltage ramps, only a few calibrating
213
ions elute during a TIMS experiment (Figure 2.a.). However, large TIMS ramps prevent
214
attaining the advertised high resolving power32. In order to improve the resolving power
215
and thus the separation of given ions, different TIMS parameters can be changed and
216
optimized. Rather rarely, the gas pressure in the TIMS cell is adjusted. Regularly, the
217
scan rates of the voltage ramp can be changed, but the TIMS ramp itself can as well be
218
changed, meaning shortened. If the elution voltage window is thus shortened, not enough
219
calibrating points of the Agilent Tune Mix can be extracted from the data. This affects the
220
calibration curve accuracy or may even prevent any calibration fit.
221
Figure 2
222
Proposing new calibrating substances: Synthetic homopolymers.
223
Using synthetic homopolymers as calibrating substances presents several advantages
224
compared to biomolecules and to the Agilent Tune Mix. First, synthetic homopolymers
225
such as PEO (poly(ethylene oxide) monomethyl ether) are low-priced, storable at room
10 /32
ACS Paragon Plus Environment
Page 10 of 32
Page 11 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
226
temperature, and easy to handle compared to biomolecules. Second, many synthetic
227
polymers such as PEO are not subjected to sample degradation or alteration over time.
228
Also, the large mass dispersities of polymer samples lead to a large mobility dispersity.
229
This results in the large coverage of dt”, Ω’, 1/Ve and K0 in the calibration plots by the
230
homopolymer calibrating substances (equation 1 and equation 2) whatever the chosen
231
instrument settings. Polymer samples will yield one-shot easy-to-handle calibrations (e.g.
232
CH3O-PEO-H 750 g/mol sample). This one-shot calibration then reduces the analysis
233
time for any commercially available ion mobility-mass spectrometer (T-Wave, TIMS…).
234
Finally, as we will discuss now, single-cation adducts such as sodium adducts of PEO
235
have an easily predictable shape (spherical shape) which was additionally established to
236
be temperature independent49,50. Their CCS also monotonically depends on the degree of
237
polymerization. These features enable easier troubleshooting detection and last but not
238
least a straightforward interpolation and extrapolation of the calibration curves.
239
Robustness and reproducibility. PEO does not have strong intramolecular interactions,
240
contrary to structured peptides or DNA. PEO-cation complexes provide single and unique
241
ATD peaks for all considered charge states and polymer chain lengths (see Figure 3.a.),
242
as opposed to other polymer ions being much stronger influenced by Coulomb
243
repsulsions51. Using PEO calibrating ions hence seems robust and reproducible, tackling
244
the issue of multi-peak ATD assignments of many biomolecules or synthetic
245
homopolypeptides.
246
Figure 3
247
Moreover, PEO polymer-cation complexes have robust shapes and do not exhibit any
248
change in conformations or any appearance of new stable shapes upon ion heating49,50.
11 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
249
Conformational changes crossing large energy barriers and trapping ions in new
250
geometries do not occur. The assignments of the PEO IM peaks therefore provide
251
unambiguous CCS or K0 values. Such synthetic homopolymer calibrations should then
252
even be usable throughout a large range of experimental temperatures in new non-DT-
253
based Variable Temperature IM-MS setups25 (VT-IM-MS), limiting the requirement of
254
recalibration of the IM cells for each effective temperature.
255
Calibration plots, improved error bars and accuracy. Table SI1 provides the DT-
256
obtained CCS and K0 values of the [PEO + 1Na+]1+ calibrating substances measured in
257
N2 and He at room temperature. The replica measurements were performed on 3 separate
258
days by 3 different operators. The first and last measurements were spaced in time by 10
259
months. Relative Standard Deviations (RSD), 95% confidence intervals (2SD) and the
260
exact number of replica measurements (n) are provided for each calibrating ion, as well
261
as the theoretical masses of the [PEO + 1Na+]1+ complexes, i.e. the theoretical m/z values.
262
Figure 1.a. and Figure 1.b. represent the CCS calibration curves of a T-Wave instrument
263
(according to equation 1). Figure 1.a. uses the published N2 tryptic digest of BSA
264
values34 whereas Figure 1.b. uses our new DT-obtained [PEO + 1Na+]1+ calibration
265
values. Figure 2.a. and Figure 2.b. represent the K0 calibration curves of a TIMS
266
instrument (according to equation 2). Figure 2.a. uses the published N2 Agilent Tune Mix
267
substance values16 whereas Figure 2.b. uses our new DT-obtained [PEO + 1Na+]1+
268
calibration values. The four plots contain the calibration curve fits described in equation 1
269
and equation 2, the 95% confidence bands and the 95% prediction bands of the fit curve.
270
The comparisons of Figure 1.a. and Figure 1.b. as well as of Figure 2.a. and Figure 2.b.
271
reveal major improvements with respect to the prediction bands, i.e. the greatest
12 /32
ACS Paragon Plus Environment
Page 12 of 32
Page 13 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
272
estimated interval of uncertainty on a calibrated value (‘worst case’ error), when using
273
PEO as calibrating substance.
274
The PEO calibrants [PEO + 1Na+]1+ also increased the number of calibrating points and
275
extends the ranges of the calibration plot axes, especially compared to the tryptic digest
276
of BSA calibration (T-Wave). Compared to the Agilent Tune Mix calibration (TIMS), the
277
synthetic homopolymer calibration improves the correlation coefficient (R²) especially in
278
cases of narrow TIMS ramp voltage conditions.
279
In order to further illustrate the enhanced quality of the calibration fits, i.e. the
280
improvement in the IM error bars (or interval of uncertainty) yielded only by the
281
calibration process, Figure 1.c. and Figure 2.c. plot the CCS and K0 errors calculated
282
from the 95% prediction bands of both the established calibrations and the new synthetic
283
homopolymer calibration. The evolution of these error bars are illustrated on the drift
284
times (or elution voltages) measured for PEO complexes bearing 1 to 4 sodium cations
285
and covering a mass range from ~0 to 4000 Da.
286
Figure 1.c. and Figure 2.c show that the synthetic homopolymer calibration yields
287
constant errors throughout the whole drift time (or elution voltage) and mass range. In the
288
case of the T-Wave, the tryptic digest of BSA calibration yields errors of up to 55 Å2
289
whereas the PEO calibration only yields errors of up to 15 Å2 (up to 73% decrease in
290
error for e.g. m/z ~3500, z = 4+, CCS ~1200 Å2). Moreover, both calibration errors
291
increase when the charge states of the calibrated analytes increase. For the [PEO +
292
1Na+]1+ calibration, however, this error is still constant within each investigated charge
293
state throughout the whole mass range. In the case of the TIMS, the Agilent Tune Mix
294
almost yields a constant error throughout all the different charge states, reaching error
13 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
295
values up to 0.050 cm2V-1s-1 while the PEO calibration reaches a constant average error
296
value of around 0.008 cm2V-1s-1 (84% decrease in error for e.g. m/z ~1500, z = 1+, K0
297
~0.49 cm2V-1s-1), identical for all investigated charge states.
298
In order to establish the accuracy of the PEO-calibrated values, the resulting CCS (T-
299
Wave and TIMS) or K0 (TIMS) values are compared to the reference DT-obtained values
300
in Figure 4.a. and Figure 5.a., and in the Supporting Information (Figure SI5 and Figure
301
SI6). The TIMS comparisons also include the values obtained through the established
302
calibrating substances (Agilent Tune Mix). In order to better visualize these differences
303
between the DT reference values and the calibrated CCS or K0 values, percentage
304
deviations of the calibrated values are calculated using equation 5. These plots are shown
305
in Figure 4.b. and Figure 5.b.
306
%( CCS or K 0 ) =
DT − Calibrated × 100 DT
(5)
307
, DT is the Drift Tube-obtained CCS or K0 value and Calibrated is the TIMS or T-Wave
308
CCS or K0 value after calibration.
309
Figure 4
310
Figure 5
311
Figure 4.a. exhibits that the TIMS and T-Wave CCS experimental values of the BSA
312
peptides are in better agreement with the DT-obtained reference values when using the
313
PEO calibration. According to Figure 4.b., the PEO-calibrated CCS values for 1+ and 2+
314
peptides are slightly more accurate than the Agilent Tune Mix-calibrated TIMS CCS
315
values. Figure 5.a. exhibits an overall better TIMS K0 accuracy for the different charge
316
states of PEO-sodium complexes using the PEO calibration instead of the Agilent Tune
317
Mix calibration. Figure 5.b. more clearly leads to the same conclusion, especially
14 /32
ACS Paragon Plus Environment
Page 14 of 32
Page 15 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
318
considering the PEO ions carrying 4 charges, and keeping in mind that the [PEO +
319
1Na+]1+ complexes’ deviation for the PEO calibration should not be taken into account
320
because they constitute the calibrating ions (reference). Figure SI3 and Figure SI4 in the
321
Supporting Information show the same plots for CCS values of PEO complexes and K0
322
values of the tryptic digest of BSA peptides. Figure SI5 and Figure SI6 show the
323
correlations between the CCS or K0 values from a T-Wave or a TIMS after the
324
calibration procedure (i.e. using [PEO + 1Na+]1+, a peptide digest from BSA with
325
trypsin34, or the Agilent Tune Mix16 as calibrating points) and the CCS or K0 values
326
obtained from the drift tube instrument used as reference. The slope of the linear
327
regressions and the statistical evaluators (correlation coefficient (R2), residual sum of
328
squares (RSS), F statistic (F-stat) and p-value) provided in these plots lead to the same
329
conclusions as Figure 4, Figure 5, Figure SI3 and Figure SI4, i.e. an overall improvement
330
of the CCS or K0 value accuracy. An important conclusion of those figures is that the
331
calibrating process based on solely the singly charged PEO complexes [PEO + 1Na+]1+ as
332
calibrants yields more accurate CCS (or K0) values even for the CCS or K0 determination
333
of multiply charged ions (e.g. PEO carrying 2 to 4 sodium cations).
334
All in all, improvements in accuracy and in the interval of uncertainty of the calibrated
335
values are obtained when using the synthetic homopolymer complexes ([PEO + 1Na+]1+)
336
as calibrating substances compared to the established calibrating substances.
337
Discussion on the origins of the improved calibration using synthetic homopolymers.
338
The improved calibration plots can be explained by several different considerations. As
339
already mentioned earlier, there are the robust shapes due to the temperature
340
independence of synthetic homopolymers.
15 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
341
Then, the choice of the singly charged polymer ions bears further significant
342
improvements because all the calibrating points share the same shape. Figure SI7 and
343
Figure SI8 illustrate the reproducibility and insensitivity of the [PEO + 1Na+]1+ calibrants
344
when calibrating in different T-Wave wave parameters. Lastly, the choice itself of
345
synthetic homopolymer ions used as calibrating substances plays an important role in
346
enhancing the calibration plots by mimicking the IM behavior of the ions to be measured.
347
Choosing the singly charged calibrating substances. The choice of using only one given
348
charge state over the whole mass range, namely the 1+ charge state [PEO + 1Na+]1+,
349
prevents any shape bias during the calibration. Indeed, 1+ complexes have intrinsically
350
spherical shapes due to the dominant entropic contribution in the absence of strong
351
intramolecular interactions and of Coulomb repulsion (i.e. maximized charge screening
352
of the cation). This eliminates potential biases caused e.g. by concavities and harmonizes
353
the specific interactions of all the [PEO + 1Na+]1+ calibrants with the drift gas particles as
354
revealed by Siems and coworkers52. Indeed, these authors improved the description of
355
hard sphere interactions for the Mason-Schamp equation and identified increasingly
356
disruptive CCS calibration inaccuracies due to differing interaction potentials with the
357
drift gas.
358
Furthermore, the singly charged homopolymers cover larger ranges of the calibration plot
359
axes. This increased number of calibrating points constitutes a significant advantage for
360
TIMS calibrations. Indeed, the TIMS voltage ramp settings dictate the window of
361
measurable K0 ranges. The numerous available calibrating points from [PEO + 1Na+]1+
362
will allow to perform calibrations whatever the ramp settings.
16 /32
ACS Paragon Plus Environment
Page 16 of 32
Page 17 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
363
Finally, being able to cover large ranges of the calibration plots by using only the [PEO +
364
1Na+]1+ calibrant ions, avoids the requirement of higher charge state ions to extend the
365
CCS range, which can introduce biases during the calibration (see Table SI2).
366
Identifying uncorrected instrumental biases. Even if calibration procedures should be
367
entirely independent of the instrumental setups (mathematical data transformations),
368
empirical evidence of inaccurate calibrated values depending on the calibrating
369
substances is mounting in literature53,54. Such observations result in the conclusions that
370
analyte-homologous calibrating substances should be used for calibration.
371
Hines and coworkers53 analyzed different calibrating substances for yielding reproducible
372
and accurate calibrated CCS values of phospholipids. They found that only lipid
373
calibrating substances were able to yield satisfying accuracies when CCS determination
374
of lipids was performed. Such issues were also observed when analyzing other
375
biomolecules. Because of their inherent sequence differences, inherent shape differences
376
translate into inherent potential energy surface (CCS) dispersities55. These CCS
377
dispersities are reflected in calibration plot dispersities. Despite the introduction of new
378
calibrating substances, calibrating substances which are too different from the chemical
379
nature of the analytes could yield inaccurate calibrated values (affected trueness). Owing
380
to the identical building blocks of given synthetic homopolymers, no CCS dispersities are
381
observed for synthetic homopolymer complexes in restrained mass ranges (see Figure
382
5.a. 2+ ions: DP 13-30, 31-42, 43-49). This results in greatly decreased dispersities on the
383
calibration plots, yielding greatly decreased error bands. The 95% prediction band fits
384
yielded by the PEO calibration are indeed well below (almost up to one order of
17 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
385
magnitude) the ones obtained for the established calibrating substances (Figure 1.c. and
386
Figure 2.c.).
387
This issue of CCS dispersities being reflected in the calibration plots has its origins in
388
flawed drift time (or equivalent) corrections during the calibration process (equation SI2
389
and equation SI6). If the drift time measurements were performed exactly between the
390
beginning and the end of the ion mobility cells and if the ions did not pass through any
391
different pressure/vacuum interfaces or ion optics devices, then one set of calibrating ions
392
would yield satisfying accurate results for all analytes (peptides, lipids, carbohydrates…)
393
as soon as the calibrating ion shapes are not affected by e.g. heating effects. Nevertheless,
394
the drift time measurements are only stopped later in the instruments. The ions have
395
undergone several ion optics and pressure changes. These stages in the IM-MS setups
396
induce in fact additional, presumably non-constant, and uncontrolled IM separations due
397
to variable drag forces. These variable drag forces depend on the different CCS values of
398
the ions at given kinetic energies. The kinetic energies are provided by the acceleration
399
voltages of ion optics, directly related to the mass of the ions. These additional,
400
interfering IM separations are not taken into account when correcting the drift times or
401
elution voltages for calibration (see equation SI2 and equation SI6). These performed
402
drift time (or equivalent) corrections are indeed purely mathematical (equation SI2) or
403
constant (equation SI6), meaning that they do not consider the interfering drifts according
404
to the CCS/mass ratio which occur in the ion optics after the IM cells. The interfering
405
CCS/mass ratio reflects a density parameter of the ions.
406
The corrections (t0) to the drift times of a DT setup are extracted experimentally when
407
performing the linear regressions for CCS determinations. The plots of the drift time
18 /32
ACS Paragon Plus Environment
Page 18 of 32
Page 19 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
408
evolutions of PEO, as well as the experimental DT corrections (t0) and the
409
mathematically calculated T-Wave drift time corrections (t0; equation SI2) are
410
represented in Figure 3. The difference between the DT and T-Wave t0 evolutions
411
represented in Figure 3.b. highlights the interfering and uncontrolled IM separations
412
occurring within the IM-MS instruments. As it is clearly observed for the [PEO + 2Na+]2+
413
complexes, the drift time evolution (Figure 3.a.) is very similar to their DT t0 evolution
414
(Figure 3.b.). However, the T-Wave t0 evolution (Figure 3.b.) of [PEO + 2Na+]2+ does
415
not reflect its drift time evolution. The interfering drifts according to the CCS/mass ratios
416
are hence not taken into account by the mathematical corrections (t0) in the T-Wave
417
calibration procedure.
418
The magnitude of the interfering drifts can nevertheless be estimated through the quality
419
control plot of a CCS calibration, namely the CCS versus mass plots. Indeed May and
420
coworkers9 established categorizations of CCS spaces as a function of the mass for
421
different analyte families such as lipids, peptides and carbohydrates. These uncontrolled
422
interfering IM separations will increasingly differ for analytes and calibrating substances
423
in increasingly differing CCS spaces. Indeed, isobaric ions can have different volumes,
424
resulting in different CCS/mass ratios, i.e. different density parameters of the ions.
425
Detailed discussions on the effect of the ion density parameter during ion mobility
426
experiments will be the focus of a future paper. Regarding the calibration process, it
427
affects the trueness of the drift times (or elution voltages).
428
Regrouping different biomolecules (enzymatic digests of proteins, proteins or other
429
biomolecules) in the same calibration plots (see Figure 1.a.) therefore prevents producing
430
high quality calibration plots, especially when using even several charge states of
19 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
431
biomolecules or polymers because it further increases the inevitable shape differences
432
incorporated in the IM calibration process54.
433
Remedying uncorrected instrumental biases: General concept of IM calibrations and
434
prospects. In general, singly charged synthetic homopolymer complexes constitute a
435
known, external and independent reference of known spherical shapes. All synthetic
436
homopolymers which do not exhibit multiple conformations or conformational changes
437
dependent on the experimental conditions, will lead to robust and reproducible calibrating
438
substances with high quality calibration fits.
439
Figure SI9 illustrates the close correlation of the CCS spaces (i.e. comparable density
440
parameters) of peptides and of PEO-sodium complexes. The PEO calibrating ions appear
441
to fulfill the calibration requirements for accurate peptide CCS determinations9,56. The
442
correlation of their CCS evolutions as a function of the mass can be described by their fit
443
parameters A from equation 4. The choice of the appropriate calibrating substance for
444
different analytes should thus be based on similar fit parameter A values between the
445
calibrants and the ions of interest.
446
As a general concept, the analyte CCS value has to be probed by an iterative calibration
447
procedure or by selecting tabulated ‘parameter A’ values of calibrants (see equation 4).
448
Indeed, after initial calibration, the CCS or K0 values can to be reevaluated using the
449
appropriate calibrants having the parameter A that matches the parameter A of the ions
450
being measured. This should thus improve the accuracy and the trueness of the measured
451
CCSs. Future efforts in IM calibrations should also go into instrumental developments
452
where the drift time measurements will have to be restricted to solely the IM cells,
453
eliminating the interfering drifts in ion optics, or to calibration databases of singly
20 /32
ACS Paragon Plus Environment
Page 20 of 32
Page 21 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
454
charged synthetic homopolymers sampling the different parameter A values (CCS
455
spaces), allowing an (automated) iterative calibration procedure.
456 457
Conclusions
458
We proposed a new calibrating substance for Ion Mobility calibrations based on synthetic
459
homopolymers, namely poly(ethylene oxide) monomethyl ether (CH3O-PEO-H). We
460
provided DT-obtained CCS and K0 calibrating points ([PEO + 1Na+]1+) in N2 and He
461
obtained from an electrospray ionization source operated in positive ion mode. The
462
provided values are average values of 3 replica measurements taken on 3 different days
463
by 3 different operators, with 10 months between first and last measurements. We tested
464
the IM calibrations on a T-Wave and a TIMS instruments.
465
The presented IM calibration strategy was based on singly charged synthetic
466
homopolymer [PEO + 1Na+]1+ complexes leading to unambiguously known, spherically
467
shaped calibrating substances. These singly charged calibrating ions are able to provided
468
improvements on prediction bands and uncertainties during the CCS determination of
469
singly charged and multiply charged ions. [PEO + 1Na+]1+ calibrating ions allowed a
470
controlled calibration procedure of the measured values into CCS or K0 values, that
471
ideally should be independent of the density parameter of the calibrating ions.
472
The increased number of calibrating substances is spread over a large range of the
473
calibration plot axes. This avoids the requirement of higher charged species as calibrating
474
substances which increase the error bars on the calibrated values. Additionally, in the
475
case of the TIMS setup, the calibration procedure can now be performed regardless of the
476
TIMS voltage ramps to improve the instrumental resolving power.
21 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
477
Given the little CCS or K0 value dispersity of synthetic homopolymers, the calibration fit
478
quality improved when using synthetic homopolymer calibrants and yields largely
479
decreased 95% prediction bands. This translates into diminished CCS or K0 errors due to
480
the calibration process, i.e. the interval of uncertainty on the calibrated values. This
481
allows for the first time to confidently extract CCS values extrapolated beyond the
482
calibration fit range.
483
Consequently, the focus of IM analyses on improved higher resolution commercial
484
instruments can now shift towards measuring the actual CCS or K0 errors, meaning the
485
peak width of an IM measurement57.
486
The more general concept of calibrating substance choices in real-world context was
487
discussed as well. In reality, the calibration plots are not fully independent of the nature
488
of the calibrating substances. Singly charged PEO homopolymers constitute appropriate
489
calibrating substances for peptide analyses, providing improved accuracy and trueness of
490
the measured CCS values because their density parameters (CCS/mass ratios) matched
491
very well. The choice of the appropriate singly charged synthetic homopolymer to be
492
used as calibrating substance should in fact be based on the density parameter which can
493
be extracted by fitting CCS versus mass plots and will be the subject of a future paper.
494
The general concept of an iterative calibration procedure based on singly charged
495
homopolymer calibrating substances, sampling the density parameters with e.g. pre-
496
established tabulated density values, was suggested.
497 498
Acknowledgments
22 /32
ACS Paragon Plus Environment
Page 22 of 32
Page 23 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
499
The authors thank the F.R.S.-FNRS for the financial support (Jean R. N. Haler and
500
Philippe Massonnet are F.R.I.A. doctorate fellows). Prof. Philippe Dugourd, Dr. Fabien
501
Chirot and Dr. Luke MacAleese (Université de Lyon) are thanked for the access to the
502
drift tube instrument and for the helpful discussions on this paper. Bruker is
503
acknowledged for their TIMS instrument and software support.
504 a.
b.
c.
505 506
Figure 1: 1.a. and 1.b. represent the Collision Cross-Section (CCS) calibration curves
507
generated for a T-Wave IM setup (equation 1). The reduced N2 CCS (Ω’) values are
508
plotted as a function of the corrected drift time (dt”) values measured in N2 on the T-
509
Wave instrument. Figure 1.a. uses the tryptic digest values of BSA described in
510
literature34 to calculate Ω’ values. Figure 1.b. uses DT-measured values of [PEO +
23 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
511
1Na+]1+ as Ω’ reference values. In both plots, the calibration fits are represented in black,
512
the 95% confidence bands are shown in blue and the 95% prediction bands are
513
represented in red. The calibration equations with their respective fit parameters and their
514
95% prediction band-deduced errors on the coefficients are depicted in each figure.
515
Figure 1.c. plots the largest expected CCS errors as a function of the Degree of
516
Polymerization (DP, number of polymerized monomer units in the polymer chain) of
517
PEO complexes (1 to 4 sodium cations). The largest expected errors are calculated from
518
the 95% prediction bands yielded from the tryptic digest of BSA calibration (in red) and
519
the [PEO + 1Na+]1+ calibration (in red).
520 a.
b.
c.
521
24 /32
ACS Paragon Plus Environment
Page 24 of 32
Page 25 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
522
Figure 2: 2.a. and 2.b. represent the reduced mobility (K0) calibration curves generated
523
for a TIMS setup (equation 2). The N2 reduced mobility (K0) values are plotted as a
524
function of the inverse of the elution voltage (1/Ve) values measured in N2 on the TIMS
525
instrument. Figure 2.a. uses the Agilent Tune Mix substances16 to calculate K0 values.
526
Figure 2.b. uses DT-measured values of [PEO + 1Na+]1+ as K0 reference values. In both
527
plots, the calibration fits are represented in black, the 95% confidence bands are shown in
528
blue and the 95% prediction bands are represented in red. The calibration equations with
529
their respective fit parameters and their 95% prediction band-deduced errors on the
530
coefficients are depicted in each figure. Figure 2.c. plots the largest expected K0 errors as
531
a function of the Degree of Polymerization (DP, number of polymerized monomer units
532
in the polymer chain) of PEO complexes (1 to 4 sodium cations). The largest expected
533
errors are calculated from the 95% prediction bands yielded from the Agilent Tune Mix
534
calibration (in red) and the [PEO + 1Na+]1+ calibration (in red).
535 a.
b.
536 537
Figure 3: 3.a. represents the DT N2 drift time evolutions of PEO-sodium complexes
538
bearing 1 to 4 cations for several polymer chain lengths (~DP 8 to 70). Figure 3.b.
25 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
539
represents the comparison of the drift time corrections (t0) of a DT (black markers) and a
540
T-Wave (red markers) IM setup. The DT t0 values are obtained experimentally whereas
541
the T-Wave t0 values are obtained mathematically (see equation SI2).
542 a.
b.
543 544
Figure 4: 4.a. represents N2 Collision Cross-Sections (CCS) of peptides provided by a
545
tryptic digest of BSA plotted as a function of the mass. The DT-obtained reference values
546
are represented in gray, the TIMS PEO-calibrated values in red and the Agilent Tune
547
Mix-calibrated16 values in blue. The T-Wave values calibrated by the PEO complexes are
548
represented in green. Figure 4.b. plots the relative percentage of the difference between
549
the DT CCS reference values and the TIMS calibrated CCS values (equation 5; PEO
550
calibration in red and Agilent Tune Mix calibration16 in blue). The dotted lines represent
551
the average percentages for the different charge states for each of the calibrations (color-
552
coded).
553
26 /32
ACS Paragon Plus Environment
Page 26 of 32
Page 27 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
a.
b.
554 555
Figure 5: 5.a. represents N2 reduced mobilities (K0) of PEO-sodium complexes bearing 1
556
to 4 cations plotted as a function of the Degree of Polymerization (DP, number of
557
polymerized monomer units in the polymer chain). The DT-obtained reference values are
558
represented in gray, the TIMS PEO-calibrated values in red and the Agilent Tune Mix-
559
calibrated16 values in blue. Figure 5.b. plots the relative percentage of the difference
560
between the DT K0 reference values and the TIMS calibrated K0 values (equation 5; PEO
561
calibration in red and Agilent Tune Mix calibration16 in blue). The dotted lines represent
562
the average percentages for the different charge states for each of the calibrations (color-
563
coded).
564 565 566
For TOC only CCS error
K0 error
cm2V-1s-1
Å2
Other Calibrations Polymer IM Calibration DP
DP
567 568
27 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
569
References
570
(1)
571 572
11246–11254. (2)
573 574
(3)
(4)
(5)
(6)
(7)
(8)
(9)
591
May, J. C.; Goodwin, C. R.; McLean, J. A. Curr. Opin. Biotechnol. 2015, 31, 117– 121.
(10)
589 590
Massonnet, P.; Upert, G.; Smargiasso, N.; Gilles, N.; Quinton, L.; De Pauw, E. Anal. Chem. 2015, 87, 5240–5246.
587 588
Wyttenbach, T.; Pierson, N. A.; Clemmer, D. E.; Bowers, M. T. Annu. Rev. Phys. Chem. 2014, 65, 175–196.
585 586
Bleiholder, C.; Dupuis, N. F.; Wyttenbach, T.; Bowers, M. T. Nat. Chem. 2011, 3, 172–177.
583 584
Bush, M. F.; Hall, Z.; Giles, K.; Hoyes, J.; Robinson, C. V.; Ruotolo, B. T. Anal. Chem. 2010, 82, 9557–9565.
581 582
Kurulugama, R. T.; Valentine, S. J.; Sowell, R. A.; Clemmer, D. E. J. Proteomics 2008, 71, 318–331.
579 580
Dwivedi, P.; Bendiak, B.; Clowers, B. H.; Hill, H. H. J. Am. Soc. Mass Spectrom. 2007, 18, 1163–1175.
577 578
Dugourd, P.; Hudgins, R. R.; Clemmer, D. E.; Jarrold, M. F. Rev. Sci. Instrum. 1997, 68, 1122.
575 576
Far, J.; Delvaux, C.; Kune, C.; Eppe, G.; de Pauw, E. Anal. Chem. 2014, 86,
Czerwinska, I.; Far, J.; Kune, C.; Larriba-Andaluz, C.; Delaude, L.; De Pauw, E. Dalt. Trans. 2016, 45, 6361–6370.
(11)
Massonnet, P.; Haler, J. R. N.; Upert, G.; Degueldre, M.; Morsa, D.; Smargiasso, N.; Mourier, G.; Gilles, N.; Quinton, L.; De Pauw, E. J. Am. Soc. Mass Spectrom.
28 /32
ACS Paragon Plus Environment
Page 28 of 32
Page 29 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
592
2016, 27, 1637–1646.
593
(12)
Lapthorn, C.; Pullen, F.; Chowdhry, B. Z. Mass Spectrom. Rev. 2013, 32, 43–71.
594
(13)
Mason, E. A.; McDaniel, E. W. Transport Properties of Ions in Gases; Wiley:
595 596
New York, 1988. (14)
597 598
Dagany, X.; MacAleese, L.; Dugourd, P. Rev. Sci. Instrum. 2015, 86, 94101. (15)
599 600
(16)
(17)
(18)
(19)
Kurulugama, R. T.; Darland, E.; Kuhlmann, F.; Stafford, G.; Fjeldsted, J. Analyst 2015, 140, 6834–6844.
(20)
609 610
Giles, K.; Williams, J. P.; Campuzano, I. Rapid Commun. Mass Spectrom. 2011, 25, 1559–1566.
607 608
Giles, K.; Pringle, S. D.; Worthington, K. R.; Little, D.; Wildgoose, J. L.; Bateman, R. H. Rapid Commun. Mass Spectrom. 2004, 18, 2401–2414.
605 606
Hernandez, D. R.; Debord, J. D.; Ridgeway, M. E.; Kaplan, D. a; Park, M. a; Fernandez-Lima, F. Analyst 2014, 139, 1913–1921.
603 604
Ruotolo, B. T.; Benesch, J. L. P.; Sandercock, A. M.; Hyung, S.-J.; Robinson, C. V. Nat. Protoc. 2008, 3, 1139–1152.
601 602
Simon, A.-L.; Chirot, F.; Min Choi, C.; Clavier, C.; Barbaire, M.; Maurelli, J.;
Cumeras, R.; Figueras, E.; Davis, C. E.; Baumbach, J. I.; Gràcia, I. Analyst 2015, 140, 1376–1390.
(21)
611
Michelmann, K.; Silveira, J. A.; Ridgeway, M. E.; Park, M. A. J. Am. Soc. Mass Spectrom. 2015, 26, 14–24.
612
(22)
Borsdorf, H.; Eiceman, G. A. Appl. Spectrosc. Rev. 2006, 41.
613
(23)
Ridgeway, M. E.; Wolff, J. J.; Silveira, J. A.; Lin, C.; Costello, C. E.; Park, M. A.
614
Int. J. Ion Mobil. Spectrom. 2016, 1–9.
29 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
615
(24)
616
Silveira, J. A.; Danielson, W.; Ridgeway, M. E.; Park, M. A. Int. J. Ion Mobil. Spectrom. 2016, 1–8.
617
(25)
Ujma, J.; Giles, K.; Morris, M.; Barran, P. E. Anal. Chem. 2016, 88, 9469–9478.
618
(26)
Kanu, A. B.; Gribb, M. M.; Hill, H. H. Anal. Chem. 2008, 80, 6610–6619.
619
(27)
Levin, M.; Krisilov, A.; Zon, B.; Eiceman, G. Int. J. Ion Mobil. Spectrom. 2014,
620
17, 73–77.
621
(28)
Kirk, A. T.; Zimmermann, S. Int. J. Ion Mobil. Spectrom. 2015, 18, 129–135.
622
(29)
Larriba, C.; Fernandez de la Mora, J. J. Phys. Chem. B 2012, 116, 593–598.
623
(30)
Hudgins, R. R.; Mao, Y.; Ratner, M. a; Jarrold, M. F. Biophys. J. 1999, 76, 1591–
624
1597.
625
(31)
Counterman, A. E.; Clemmer, D. E. J. Phys. Chem. B 2003, 107, 2111–2117.
626
(32)
Silveira, J. A.; Ridgeway, M. E.; Park, M. A. Anal. Chem. 2014, 86, 5624–5627.
627
(33)
Trimpin, S.; Plasencia, M.; Isailovic, D.; Clemmer, D. E. Anal. Chem. 2007, 79,
628 629
7965–7974. (34)
630 631
Bush, M. F.; Campuzano, I. D. G.; Robinson, C. V. Anal. Chem. 2012, 84, 7124– 7130.
(35)
632
Forsythe, J. G.; Petrov, A. S.; Walker, C. A.; Allen, S. J.; Pellissier, J. S.; Bush, M. F.; Hud, N. V; Fernández, F. M. Analyst 2015, 140, 6853–6861.
633
(36)
Liu, F. C.; Kirk, S. R.; Bleiholder, C. Analyst 2016, 141, 3722–3730.
634
(37)
Valentine, S. J.; Counterman, A. E.; Clemmer, D. E. J. Am. Soc. Mass Spectrom.
635 636 637
1997, 8, 954–961. (38)
Valentine, S. J.; Anderson, J. G.; Ellington, A. D.; Clemmer, D. E. J. Phys. Chem. B 1997, 101, 3891–3900.
30 /32
ACS Paragon Plus Environment
Page 30 of 32
Page 31 of 32
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
638
(39)
Shelimov, K. B.; Jarrold, M. F. J. Am. Chem. Soc. 1997, 119, 2987–2994.
639
(40)
Morsa, D.; Gabelica, V.; De Pauw, E. Anal. Chem. 2011, 83, 5775–5782.
640
(41)
Shvartsburg, A. A.; Li, F.; Tang, K.; Smith, R. D. 2007.
641
(42)
Chen, S.-H.; Russell, D. H. J. Am. Soc. Mass Spectrom. 2015, 26, 1433–1443.
642
(43)
Poyer, S.; Comby-Zerbino, C.; Choi, C. M.; MacAleese, L.; Deo, C.; Bogliotti, N.;
643 644
Xie, J.; Salpin, J.-Y.; Dugourd, P.; Chirot, F. Anal. Chem. 2017, 89, 4230–4237. (44)
645
Sun, Y.; Vahidi, S.; Sowole, M. A.; Konermann, L. J. Am. Soc. Mass Spectrom. 2016, 27, 31–40.
646
(45)
Counterman, A.; Clemmer, D. J. Phys. Chem. B 2004, 108, 4885–4898.
647
(46)
Shi, L.; Holliday, A. E.; Glover, M. S.; Ewing, M. A.; Russell, D. H.; Clemmer, D.
648
E. J. Am. Soc. Mass Spectrom. 2016, 27, 22–30.
649
(47)
Counterman, A. E.; Clemmer, D. E. J. Phys. Chem. B 2002, 106, 12045–12051.
650
(48)
Flanagan, J. M. Mass spectrometry calibration using homogeneously substituted
651 652
fluorinated triazatriphosphorines. US 5872357 A, 1999. (49)
653 654
Mass Spectrom. 1997, 8, 275–282. (50)
655 656
Wyttenbach, T.; Helden, G.; Batka, J. J.; Carlat, D.; Bowers, M. T. J. Am. Soc.
von Helden, G.; Wyttenbach, T.; Bowers, M. T. Int. J. Mass Spectrom. Ion Process. 1995, 146–147, 349–364.
(51)
657
Haler, J. R. N.; Far, J.; Aqil, A.; Claereboudt, J.; Tomczyk, N.; Giles, K.; Jérôme, C.; De Pauw, E. J. Am. Soc. Mass Spectrom. 2017, 1–8.
658
(52)
Siems, W. F.; Viehland, L. A.; Hill, H. H. Analyst 2016, 141, 6396–6407.
659
(53)
Hines, K. M.; May, J. C.; McLean, J. A.; Xu, L. Anal. Chem. 2016, 88, 7329–
660
7336.
31 /32
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
661
(54)
662 663
88, 5879–5884. (55)
664 665
Allison, T. M.; Landreh, M.; Benesch, J. L. P.; Robinson, C. V. Anal. Chem. 2016,
Henderson, S. C.; Li, J.; Counterman, A. E.; Clemmer, D. E. J. Phys. Chem. B 1999, 103, 8780–8785.
(56)
May, J. C.; Goodwin, C. R.; Lareau, N. M.; Leaptrot, K. L.; Morris, C. B.;
666
Kurulugama, R. T.; Mordehai, A.; Klein, C.; Barry, W.; Darland, E.; Overney, G.;
667
Imatani, K.; Stafford, G. C.; Fjeldsted, J. C.; McLean, J. A. Anal. Chem. 2014, 86,
668
2117–2116.
669
(57)
Kune, C.; Far, J.; De Pauw, E. Anal. Chem. 2016, 88, 11639–11646.
670
32 /32
ACS Paragon Plus Environment
Page 32 of 32