Miniaturized thermal-assisted purge-and-trap technique coupling with surface-enhanced Raman scattering for trace analysis of complex samples Zhengyi Chen, Gongke Li, and Zhuomin Zhang Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.7b02912 • Publication Date (Web): 10 Aug 2017 Downloaded from http://pubs.acs.org on August 11, 2017
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 26
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1 2 3
Miniaturized
thermal-assisted
purge-and-trap
technique
4
coupling with surface-enhanced Raman scattering for trace
5
analysis of complex samples
6 7 8
Zhengyi Chen, Gongke Li*, Zhuomin Zhang*
9 10
School of Chemistry, Sun Yat-sen University, Guangzhou 510275, China
11 12 13 14
* Corresponding authors: Gongke Li, Zhuomin Zhang
15
Tel.: +86-20-84110922
16
Fax.: +86-20-84115107
17
E. mail :
[email protected] (Gongke Li)
18
[email protected] (Zhuomin Zhang)
19
1 `
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
20
ABSTRACT
21
It still remains a great challenge for quantification of trace analytes in complex
22
samples by surface-enhanced Raman scattering (SERS) technique due to potential matrix
23
influence or weak SERS responses of analytes. In this work, a miniaturized thermal-
24
assisted purge-and-trap (MTAPT) device was designed and developed to eliminate matrix
25
influence coupled with derivatization method before SERS analysis. The design of
26
MTAPT chamber was optimized based on quantitative calculation of its dead volume by
27
computational fluid dynamics simulation. The small straight chamber was selected as an
28
optimized design with a recovery of 96.1% for formaldehyde. The practical feasibility of
29
MTAPT was validated based on four real analytical applications including phenthiol in
30
industrial water, formaldehyde in flour, sulfion in waste water and methanol in industrial
31
alcohol. The results showed that SERS responses of all analytes dramatically increased
32
by eliminating sample matrices after MTAPT process. Phenthiol, formaldehyde, sulfion
33
and methanol in real samples could be accurately quantified with recoveries of 80.9%-
34
110.0%, and the analytical results were validated by corresponding standard methods.
35
The time consumption of MTAPT-SERS for real sample analysis including sample
36
preparation and determination was within 16 min. It is highly expected that the
37
combination of MTAPT technique with portable SERS instrument can greatly expand the
38
range of SERS analysis. The proposed MTAPT-SERS method has high potential for on-
39
site analysis of complex samples.
40 41
Keywords: Miniaturized thermal-assisted, Purge-and-trap, SERS, Trace analysis,
42
Complex sample.
2 `
ACS Paragon Plus Environment
Page 2 of 26
Page 3 of 26
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
43
Analytical Chemistry
Introduction
44
In recent years, SERS technique for rapid and on-site analysis of real samples has
45
aroused great attention1-3, because it can provide more delicate information of target
46
molecular fingerprints with potential single-molecule level of sensitivity4,
47
analysis techniques. However, SERS analysis becomes impractical when being applied in
48
complex samples, since SERS amplification of Raman signals highly depends on noble
49
metal nanoparticles such as Au, Ag or their composites, which are susceptible to the
50
matrix of real samples and finally influence the accuracy of SERS analysis6-8. To achieve
51
accurate analysis of real samples by SERS, it is crucial to eliminate sample matrices via
52
efficient sample preparation
53
attaching on the surface of SERS substrates to generate severe interference signals12.
9-11
5
than other
in order to avoid large amounts of matrix molecules
54
Recently, many efforts have gone to reduce matrix interference and enrich target
55
analytes during SERS analysis of real samples by coupling of traditional sample
56
preparation or separation methods such as liquid-liquid extraction13, solid-phase (micro-)
57
extraction SP(M)E10, magnetic extraction3, headspace microextraction14 or HPLC15, thin
58
layer chromatography9, capillary electrophoresis16 and ion-pair chromatography (IC)17 .
59
Most of these sample preparation methods are based on liquid-liquid or solid-liquid
60
extraction-separation process and usually require large amounts of organic solvent and
61
sample solutions with long proceeding time, which usually results in potential
62
interference to targets and relatively low extraction efficiency. The ideal sample
63
preparation technique coupling to rapid SERS analysis should efficiently eliminate the
64
complex sample matrices using few organic solvents with short proceeding time and
65
portable instrument scale18. Sample preparation technique, based on liquid-gas or solid-
3 `
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
66
gas extraction-separation process, such as purge-and-trap (P&T) and SPME, would be
67
potentially suitable for rapid SERS analysis because they avoid using large volumes of
68
organic solvent19 and any further clean-up post-treatment20. Thus, it is expected that the
69
development of liquid-gas or solid-gas sample preparation methods would benefit the
70
enhancement of analytical selectivity and accuracy by SERS. Chen et al.17 developed a
71
thin-film headspace microextraction for SERS analysis of trace sulfur dioxide in wine
72
samples. This method utilized headspace microextraction process to achieve the efficient
73
separation of target volatile SO2 and matrix remaining in the sample solution, which
74
resulted in the selective enrichment of SO2 and great enhancement of SRES response
75
simultaneously. However, up to now there are few works focusing on the development of
76
efficient and portable sample preparation technique coupling with rapid SERS analysis
77
for complex samples 11. Moreover, most sample preparation methods coupled with SERS
78
were designed for specific volatile compounds within a limited range. Few works are
79
devoted to the combination of portable but efficient sample preparation techniques with
80
SERS ananlysis for real sample analysis.
81
P&T technique is a traditional liquid-gas separation technique with the great
82
potential to a suitable sample preparation technique for rapid SERS analysis because
83
P&T can efficiently separate the trace analytes from complex sample matrices and has
84
been widely applied for the enrichment and analysis of voltile targets from various
85
samples20. However, most of commercial P&T devices are adapted to the large-scale
86
analytical instruments such as gas chromatography (GC) and GC-mass spectrometry
87
(GC-MS). Generally, P&T device includes a sample vessel, a thermal-assisted desorption
88
unit and a trapping unit19. Conventional sample vessels are frit purge tubes or common
4 `
ACS Paragon Plus Environment
Page 4 of 26
Page 5 of 26
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
89
frit bottle21. Common thermal-assisted desorption unit and trapping unit is based on the
90
manifold heater22 and Tenax-GC trap23, respectively. Thus, commercial P&T devices are
91
so bulky that they are not suitable for rapid on-site SERS analysis. To overcome these
92
problems, a miniaturized thermal-assisted purge-and-trap (MTAPT) device is developed
93
in this work. The main challenge is how to realize the miniaturization of separation
94
system in MTAPT but remain its high transferring efficiency at the same time. On one
95
hand, it is crucial to design and develop the efficient micro-heater with battery-voltage for
96
MTAPT. Recently, as an important unit, micro-heater using battery-voltage can heat to
97
300℃ within a short time and has been applied for volatiles monitoring24 or sample
98
preparation25 due to its high heating efficiency. An MTAPT device with high heating
99
efficiency was designed by using low voltage of micro heater in our work. On the other
100
hand, in order to remain high transferring efficiency, it is essential to quantitatively
101
investigate the dead volume in the chamber for the optimization of MTAPT design.
102
However, until now there are no studies focusing on the investigation of the quantitative
103
method for dead volume. Simulation strategies such as computational fluid dynamics
104
(CFD) analysis are ideal methods for quantitative calculation of many physical
105
parameters of the gas in a specific region26, such as pressure and velocity. Benefit from
106
the velocity distribution function of CFD, the dead volume, which is minimal magnitude
107
of velocity, could be vividly present during the design of the chamber’s geometry of
108
MTAPT. Compared with the traditional P&T device, MTAPT presents not only
109
portability but also various functions due to its superior liquid-gas transformation
110
performance and efficient heating rate.
111
The targets after treatment with MTAPT are volatile and most of them possess weak
5 `
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 6 of 26
112
SERS response due to their small Raman cross sections. However, SERS is usually
113
suitable for analysis of targets containing highly polarizable aromatic groups6, which are
114
large Raman cross sections and yield the strong SERS response
115
SERS analysis, derivatization is an efficient strategy to enhance Raman cross section by
116
modification of polarizable aromatic groups28,
117
derivatization strategy is limited to those chemical active volatile compounds, such as
118
hydrazine6, formaldehyde7, 30, 31 and acetone8. Therefore, for some chemically inert or
119
non-volatile compounds, derivatization strategy for SERS activated becomes unavailable.
120
MTAPT is a good way to solve this problem, since it can transfer some non-volatile
121
compounds to volatile or chemically active targets.
29
27
. To realize MTAPT-
. However, the applicability of
122
In this work, an MTAPT device with high heating efficiency was designed
123
containing a micro-heater and specific heating chamber. The miniaturization process was
124
optimized in detail based on the CFD simulation. And then four application strategies of
125
MTAPT were developed based on various targets with sharply different properties
126
including strong SERS response, easily SERS activated and volatile, non-volatile and
127
difficultly SERS activated compounds. The application of MTAPT-SERS was illustrated
128
by four targets including phenthiol, formaldehyde, sulfion, and methanol. MTAPT-SERS
129
technique was actually applied for the rapid analysis of these four trace targets in real
130
complex samples.
131
Experimental
132
Chemicals. 3-Methyl-2-benzothiazolinone hydrazone hydrochloride monohydrate (98%),
133
N,N-dimethyl-p-phenylenediamine
134
dinitrophenylhydrazine (98%), trifluoroacetic acid (99.5%, TFA) and acetaldehyde
sodium
sulfide
nonahydrate
6 `
ACS Paragon Plus Environment
(99.99%),
4-
Page 7 of 26
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
135
(99.5%) were purchased from Aladdin Chemistry Co. Ltd (Shanghai, China).
136
NH4Fe(SO4)2·12H2O was obtained from Tianjin Chemical Reagent Plant (Tianjin, China).
137
Au/SiO2 colloids (2.94×10-4 mol/L) having pinholes consisting of Au NP cores with a
138
diameter of 55 nm around and silica shell thickness of 1-2 nm and Au colloids were
139
supplied by the groups of Prof. Tian in Xiamen University (Xiamen, China)32.
140
Instruments. A portable DeltaNu (Laramie, WY) battery-powered Raman spectrometer
141
(model Inspector Raman, diode laser excitation wavelength λex=785 nm) with a liquid-
142
N2-cooled CCD detector (Model Spec-10:400B, Roper Scientic, Trenton, NJ) was used
143
for SERS analysis. The LTQ-Orbitrap (Thermo-Fisher, San Jose, CA, USA) was used for
144
identifying the target analyte and by-product during derivazation. All chromatographic
145
analyses for comparison study were performed by a Shimadzu LC-20AB (Shimadzu,
146
Japan) which consisted of binary pumps (LC-20AT), an injection valve (Rheodyne model
147
7725i) equipped with a 20 µL injection loop, a UV-Vis wavelength detector, a system
148
controller (SPD-20A) and an acquisition data software of LC-Solution. Chromatographic
149
separation was carried out on a Diamonsil C18 column (250 mm × 4.6 mm i.d., 5 µm
150
particle size) from Dikma (Beijing, China), and a 10 mm ×4.6 mm i.d. C18 security guard
151
column from Elite (Dalian, China). Current source (ZHAOXIN KXN-3002D DC, China)
152
was purchased from local market which could precisely mediate the voltage from 0.0-
153
36.0 V.
154
Development of MTAPT technique. The MTAPT system mainly included four parts: a
155
carrier gas pump, a vapour chamber, a heating vessel and an adoptor (Scheme 1). The
156
carrier gas was pumped by a commercial mini-aquarium pump or mini-N2 cylinder. The
157
vapour chamber was consisted of a sample injection pore for injecting sample and a
7 `
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
158
heater entrance pore for loading vessel. Silicone gasket was plugged in the sample
159
injection pore and the heater entrance pore for preventing the leakage of volatile targets.
160
A pore on the adoptor was used for gas venting. The heating vessel was well designed
161
because it played an important part during the miniaturization. Considering its portability,
162
a commercial metal ceramics heater with the size of 1.0 cm × 1.0 cm was employed to
163
miniaturize the vapour chamber in MTAPT as much as possible. Considering heating
164
efficiency and safety, the resistance and the max voltage of micro-heater was set at 3.0 Ω
165
and 17 V, respectively. The micro-heater could heat to 120℃ at 6.0 V simply by use of
166
several batteries. The vessel was made of anodic aluminum oxide owing to its much
167
higher resistance to corrosion and chemical inertness than the other common metals such
168
as copper and iron.
169
Computational fluid dynamics analysis of chamber geometry. The effect of the
170
chamber geometry on fluid flow conditions and further sampling efficiency were
171
investigated by computational fluid dynamics (CFD) analysis. Four different chambers
172
consisting of one sample inlet, gas inlet and outlet were designed for obtaining optimal
173
chamber geometry. The dimensions of these four chambers were illustrated in Figure S1.
174
The Navier–Stokes equation was established for the calculation using a commercial
175
software package of ANSYS 14.0/fluent CFD. The modelling was made based on three-
176
dimensional steady and incompressible flow. The modelled fluid was air with the density
177
of 1.29 kg/m3 and the realizable k-є model (RKE) was used for turbulence model.
178
Application strategies of MTAPT-SERS for real sample analysis. MTAPT-SERS
179
strategies were developed and applied for analysis of four selected targets including
180
phenthiol, formaldehyde, sulfion and methanol in corresponding complex samples in
8 `
ACS Paragon Plus Environment
Page 8 of 26
Page 9 of 26
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
181
order to validate the feasibility of the method. The corresponding strategy was present in
182
Scheme 1S. The specific sample preparation and analysis protocols were presented in the
183
experimental section of “Supplementary Material”. After being treated by MTAPT, 200
184
µL of sample solutions were mixed with 100 µL of corresponding SERS substrate
185
followed by SERS detection in the specific acquisition integration time and power. SERS
186
acquisition integration time for detection of phenthiol, formaldehyde, sulfion and menthol
187
was 5s, 1 s, 1 s and 3 s, respectively. SERS powers for detection of phenthiol,
188
formaldehyde, sulfion and menthol were set at high power, medium high power, high
189
power and high power.
190
Results and discussion
191
The Design of MTAPT Device. The presence of dead volumes would make the
192
components remain in the sampling chamber, especially in gas recirculation
193
illustrate the effect of the miniaturization of chamber on dead volume, CFD simulation
194
was carried out for four different kinds of vapour chambers which were circuitous (Type
195
a, Figure S1a), straight but big (Type b, Figure S1 b), straight with long heater vessel
196
entrance (Type c, Figure S1c) and small with short heater vessel entrance (Type d,
197
Figure S1d) respectively. The resulting velocity vectors (Figure 1) showed the gas
198
recirculation regions that occurred in the chambers but varied with different dead
199
volumes. Especially, it was clearly seen there was gas recirculation region in the inlet of
200
circuitous chamber (Figure 1a) in comparison with other straight chambers (Figure 1b-d).
201
For further study, the velocity magnitude distribution of these four chambers was
202
quantitatively counted based on their CFD simulations data, and their volumes were
203
precisely calculated based on their geometry. Dead volume was quantified by the peak
9 `
ACS Paragon Plus Environment
3233
. To
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 10 of 26
204
value (vp) and the distribution percent of velocity magnitude below 0.005 m/s (v