Optimization and Application of Direct Infusion Nanoelectrospray

Feb 28, 2017 - *E.C.: E-mail: [email protected]., *E.H.: E-mail: ... polarities, enabling further data mining and relative quantification of...
0 downloads 3 Views 1MB Size
Subscriber access provided by Fudan University

Article

Optimization and application of direct infusion nanoelectrospray HRMS method for large-scale urinary metabolic phenotyping in molecular epidemiology Elena Chekmeneva, Goncalo dos Santos Correia, Queenie Chan, Anisha Wijeyesekera, Adrienne Tin, J. Hunter Young, Paul Elliott, Jeremy Kirk Nicholson, and Elaine Holmes J. Proteome Res., Just Accepted Manuscript • DOI: 10.1021/acs.jproteome.6b01003 • Publication Date (Web): 28 Feb 2017 Downloaded from http://pubs.acs.org on March 3, 2017

Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.

Journal of Proteome Research is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.

Page 1 of 35

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Journal of Proteome Research

Optimization and application of direct infusion nanoelectrospray HRMS method for large-scale urinary metabolic phenotyping in molecular epidemiology Elena Chekmeneva

a*

, Gonçalo dos Santos Correia

a

e

a

,

Queenie Chan

e

Wijeyesekera , Adrienne Tin , Jeffery Hunter Young , Paul Elliott

b,c

, Anisha

b,c,d

, Jeremy K. Nicholson

a,d

, and Elaine Holmes a,d*

a

Biomolecular Medicine, Department of Surgery and Cancer, Faculty of Medicine, Imperial

College London, South Kensington, London SW7 2AZ, UK b

Department of Epidemiology and Biostatistics, School of Public Health, Faculty of

Medicine, Imperial College London, London W2 1PG, UK c

MRC-PHE Centre for Environment and Health, Imperial College London, London W2 1PG,

UK d

MRC-NIHR National Phenome Centre, London SW7 2AZ, UK

e

Department of Medicine, John Hopkins University, School of Medicine, Baltimore, MD

* Corresponding authors: Dr Elena Chekmeneva (email: [email protected]) and Prof Elaine Holmes (e-mail: [email protected])

1 ACS Paragon Plus Environment

Journal of Proteome Research

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 2 of 35

Abstract Large scale metabolic profiling requires development of novel economical high-throughput analytical methods to facilitate characterization of systemic metabolic variation in population phenotypes. We report a fit-for-purpose direct infusion nanoelectrospray high resolution mass spectrometry (DI-nESI-HRMS) method with time-of-flight detection for rapid targeted parallel analysis of over forty urinary metabolites. The newly developed two-minute infusion method requires less than ten microliters of urine sample and generates high-resolution MS profiles in both positive and negative polarities enabling further data mining and relative quantification of hundreds of metabolites. Here we present optimisation of the DI-nESIHRMS method in a detailed step-by-step guide and provide a workflow with rigorous quality assessment for large-scale studies. We demonstrate for the first time the application of the method for urinary metabolic profiling in human epidemiological investigations. Implementation of the presented DI-nESI-HRMS method enabled cost efficient analysis of >10000 24-hour urine samples from the INTERMAP study in 12 weeks and >2200 spot urine samples from the ARIC study in less than 3 weeks with the required sensitivity and accuracy. We illustrate the application of the technique by characterizing the differences in metabolic phenotypes of the USA and Japanese population from the INTERMAP study.

Key words: Direct infusion mass spectrometry, metabolic profiling, molecular epidemiology, high-throughput analysis

1.

Introduction

Technological advancements including automated sample handling, data processing together with reduction in cost, have seen an increase in the application of metabolic profiling approaches in molecular epidemiology.1,2 In order to achieve sufficient power to discover statistically significant associations at the population level, these studies require consecutive analysis of tens or hundreds thousands of samples. Analyses on a molecular epidemiology scale can deliver solutions for metabolome-wide association (MWAS) studies, exploring the statistical relationships between disease risk factors and metabolite concentrations in large human sample cohorts.3 One strategy for achieving a higher sample throughput in metabolic profiling is to completely omit the chromatographic separation phase by using direct infusion mass spectrometry (DIMS), which involves continuous ionization of a static sample.4,5 This analytical technique has been used in untargeted ESI-MS profiling studies of plant, food, and environmental samples, as well as in animal and human biofluid and tissue samples for the 2 ACS Paragon Plus Environment

Page 3 of 35

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Journal of Proteome Research

last two decades.6 The first reports on the use of DIMS for ESI-MS profiling of crude fungal extracts date back to 1996-1997.7 Application of DIMS with nESI, coupled to a computational workflow for global metabolic profiling of various tissue extracts using ultrahigh mass accuracy and resolution Fourier Transform Ion Cyclotron Resonance (FT-ICR) mass spectrometry have been significantly developed and optimized in the last 5 years.8,9,10 In addition to operating in profiling mode for untargeted phenotyping of human plasma11 and urine,5 DIMS has been successfully applied to the quantification of serotonin related compounds in urine samples12 and free fatty acids in human serum.13 The comparison studies of the DIMS and LC-MS approaches for global metabolic profiling of human serum14 and targeted analysis of heat shock proteins in cell line models15 have shown comparable classification, prediction and quantification capabilities of both methods with a clear gain in analysis time when using DIMS. The main limitations of any DIMS method include the strong matrix effect and ion suppression due to the fact that all ionisable components of a sample are infused at the same time and competing for charge. These limitations can be minimized by employing nanoelectrospray ionization conveniently available in chip-based ionisation interfaces such as the Advion TriVersa NanoMate. The lower flow rates achieved with nanoelectrospray improve ionisation efficiencies for all analytes in a mixture. This is explained by the formation of reduced size charged droplets than in conventional electrospray, which leads to fewer droplet fusion events, less solvent evaporation prior to ion release into the gas phase and thus, a larger portion of analyte available for MS analysis.16 Despite advances in MS technology and instrument resolution, DIMS methods also suffer from an inability to distinguish structural isomers, which requires chromatographic separation and additional offline MS/MS experiments. The separation of isomeric compounds cannot be always attained in general LC-MS profiling methods. Another limitation of DIMS methods are the simultaneous presence of multiple types of MS signals for each metabolite in the spectrum in addition to its molecular ion, such as adducts and in-source fragments. This complicates the processing and interpretation of the data. Nevertheless, this phenomenon can deliver useful structural and biochemical information for biomarker discovery by using statistical correlation approaches.17, 18 The choice of mass analyser can affect the quality of the data obtained by DIMS. With ultrahigh mass accuracy and high resolution mass spectrometers such as Orbitrap and FT-ICR the masses obtained during the measurement are expected to be close to the theoretical mass of metabolites, which facilitates the prediction of ion elemental formulae. However, the increase 3 ACS Paragon Plus Environment

Journal of Proteome Research

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 4 of 35

in mass accuracy and resolution can be accompanied by the loss of sensitivity, scan speed and decrease in detector dynamic range, requiring specific data acquisition approaches to mitigate these effects such as previously reported spectral stitching technique.9 The hybrid quadrupole – time of flight (Q-TOF) instrument offers high resolution mode (up to 50.000) and flexible scan rates, and their costs are more accessible for many research laboratories compared to the costs of purchasing and maintaining ultra-high mass accuracy and high resolution mass spectrometers. Despite the aforementioned limitations, DI-nESI-HRMS methods offer an appealing option for rapid MS sample screening and parallel semi-quantitative analysis of large sample numbers and can be applied to large cohorts of biobanked samples. Although the targeted LC-MS/MS methods performed on triple quadrupole mass spectrometers are highly precise and accurate, they are usually focused on specific metabolic pathways, and the panels of analysed metabolites are limited.19,20 Wider coverage of metabolite space requires consecutive implementation of several targeted LC-MS/MS assays, which does not meet the increasing demands of time and cost-efficiency in metabolic phenotyping applied to large clinical and epidemiological sample sets. In addition, it can be economically inefficient and environmentally unfriendly due to the increased solvent consumption. It is evident that a decision must be taken to balance the number of measured analytes and selectivity of an assay against throughput and sensitivity. Recently, we have reported the development and application of a direct infusion nanoelectrospray high resolution mass spectrometry (DInESI-HRMS) method with time-of-flight detection for multiplexed parallel metabolic profiling of urine samples.21 As a proof of concept, this method was applied to a relatively small set of 101 urine samples, and was validated for eight targeted metabolites showing to be precise and accurate according to the standard requirements. We and others have shown that the DI-nESI-HRMS method using the NanoMate system, which requires minimal amounts of sample owing to the injectable volume being in the order of nL, ensures a stable instrument performance, high sensitivity, and is not compromised by contamination.13,11,21 In a single rapid analytical high resolution sweep the method provides wide metabolome coverage including amino acids, TCA cycle metabolites, fatty acids, acylcarnitines, gut microbial co-metabolites, and other, mostly polar, metabolites. The quantification capability provided by the current method allows characterization of population phenotypes, and selection of representative samples for further UPLC-MS (MS/MS) analysis. Moreover, DI-nESI-HRMS provides high resolution accurate mass global profiles of samples that can be further explored for discovery and semi-quantification of 4 ACS Paragon Plus Environment

Page 5 of 35

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Journal of Proteome Research

biomarkers, enabling interrogation of their association with metabolic phenotypes and clinical outcomes. The proposed DI-nESI-HRMS method for multiplexed parallel untargeted and targeted semiquantitative metabolic analysis of the urine samples is not an alternative to a conventional UPLC-MS (MS/MS) analysis but a valuable complementary tool in the research portfolio of large scale metabolic phenotyping laboratories.19,20,22 The main purpose of this work is to show the applicability of the optimised DI-nESI methodology with a Q-TOF mass analyser to measure multiple compounds in parallel in accordance with standard requirements of good precision, accuracy, stability, robustness and reproducibility on a scale of thousands of samples that is able to deliver MWAS solutions. In order to validate the DI-nESI-HRMS method, we applied it to two relatively large population cohorts, The INTERnational study of MAcro/micronutrients and blood Pressure (INTERMAP)23 and Atherosclerosis Risk in Communities (ARIC),24 which address hypertension and cardiovascular risk factors respectively. Based on the previous 1H NMR spectroscopic MWAS analysis of >10,000 24hour urine specimens from the INTERMAP study,3,25,26 we aimed to extend coverage of the blood pressure related metabolome and to quantify a panel of metabolites with varying biological functions. A sub-cohort of > 2200 spot urine samples from the ARIC study was used for targeted metabolic analysis to measure specific urinary markers of oxidative capacity and assess their role as predictors of incident type 2 diabetes. Analysis was based on 47 selected metabolites for quantification in each study tailored to the research question, with 16 metabolites common for the two studies allowing comparison of the performance of the method across different populations (Table 1 and Table S-1).

2.

Materials and Methods

2.1 Materials and preparation of standard solutions A description of the solvents and reference standards (Table S-1) is provided in Supporting Information (SI). The labelled and non-labelled standard stock solutions were prepared at a concentration of 1 mg/mL in methanol or methanol/water mixture if the standard was not soluble in methanol only. The multi-analyte mixture of labelled internal standards and the multi-analyte mixture of non-labelled calibrators were prepared at different levels of concentration by mixing stock solutions in the total volume of 10 mL. In this way, the multianalyte mixture of the non-labelled standards with different levels of concentration for each analyte was used as the most concentrated calibrator – Working Solution (WS) 1. The concentrations of the standards in the WS1 are shown as the upper limit of the linear range in 5 ACS Paragon Plus Environment

Journal of Proteome Research

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 6 of 35

Table 1, for the INTERMAP and ARIC studies, respectively. More working solutions (WS2WS6) used for generation of calibration curves and spiked validation quality control (QC) samples were prepared from WS1 by serial dilution. The concentration of each labelled internal standard was fixed and within the expected linear range for each metabolite.

2.2 Sample Preparation The dilution factor of urine samples was optimised for each study separately by assessing the effect of dilution on the intensities of endogenous metabolites and the added internal standards. Various reference and quality control materials were developed and used for method optimization and assessment of data quality during the experiments. Pooled Study Reference samples (SR) were prepared for both studies by pooling together 10 µL aliquots of all urine samples. The pooled SR samples were diluted by the same factor as urine samples from each study. The SR dilution series were prepared by diluting pooled SR by the factors of 1/10, 1/20, 1/50, 1/100 and were used to ensure that the detected MS features positively correlate with sample dilution. Urine sample preparation in INTERMAP study. Aliquots of 10 µL of thawed urine samples stratified by the four populations (USA, UK, Japan, and PRC) and two separate clinic visits (first-F and repeated-R) from the INTERMAP study were pipetted in randomized order into deep-well plates and diluted by a factor of 50. Aliquots of 10 µL of each urinary sample were mixed in order to obtain 8 pooled SR samples for each population and visit. The pooled SR samples were also diluted by a factor of 50. In ARIC study, the individual urine samples and pooled SR sample were prepared in the same way as for the INTERMAP study using a dilution factor of 20. Calibration series and QC samples were prepared by pipetting 50 µL of diluted pooled SR samples in a well-plate adding 25 µL of the multi-analyte mixture of labelled internal standards at fixed concentration and 25 µL of the corresponding calibrator solution (both prepared in methanol), and making up the total volume to 150 µL by adding methanol to maintain 1:2 water-methanol proportion. In this way, the urine samples were diluted in a well-plate by a factor of 150 (1/50 and 50 µL/150µL) for the INTERMAP study and by a factor of 60 (1/20 and 50 µL/150µL) for ARIC study, while the calibrators used for the calibration series and QC samples were diluted 6 times in a well-plate (25 µL/150µL). All these factors have been taken into account for the quantification of metabolites and

6 ACS Paragon Plus Environment

Page 7 of 35

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Journal of Proteome Research

calculations of the limits of quantification (LOQ). The description of the analyte-free matrix for external calibration is presented in SI. Each sample plate, in addition to the study samples, contained a calibration series, SR and dilution SR series, validation QC samples (two samples prepared at three different concentration levels), and blank solutions. The sample plates were sealed and subjected to 1 min ultrasonication followed by 10 min centrifugation at 1500g at 4 °C before MS analysis.

2.3 Validation experiments Linear ranges, inter- and intra-day accuracy and precision, long-term stability and method robustness were assessed for the DI-nESI-HRMS method using FDA guidelines27,28 for the validation of bioanalytical methods. A standard addition method in pooled SR samples was used by spiking the multi-analyte series of non-labelled standards into each calibrator solution and validation QC sample. The mixture of labelled analogues, used as internal standards, was spiked into each sample at a known fixed concentration. The MS quantification was performed using the ratio of intensities of the metabolite to the intensity of its internal standard. The response for each QC sample was compared to the generated calibration curve to give concentration values. The validation study was performed over separate days. On the first day, the well-plate contained the calibration series consisting of six points plus zero point (spiked with the mixture of labelled internal standards only), six replicates of validation QC samples prepared at three different concentration levels for each metabolite (low, medium and high) which were defined for each metabolite from the corresponding linear range (Table 1). Blanks (1:2 water:methanol) were injected after the calibration series and each QC series in order to test carry-over. On the second and third days, each plate contained a calibration series and one replicate of the QC series. The calibration series and QC samples in analyte-free matrix were prepared in the same way. The intra- and inter-day accuracy and precision were calculated as closeness of a measured value to a known value of spiked concentration (%) and coefficient of variance (CV%), respectively. Method LOQ was determined as 10 times the standard deviation of the y-intercept divided by the slope of the calibration curves obtained in analytefree matrix. Stability study was carried out by preparing three additional validation and QC series from the INTERMAP study to assess the stability of the samples. One of them was measured immediately while second one was measured after three freeze-thaw cycles, and the third one after long-term (three months) storage at -80 °C. The CV% values of measurements were 7 ACS Paragon Plus Environment

Journal of Proteome Research

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 8 of 35

evaluated. Additionally, the untargeted analysis of the full-scan data was done to assess possible changes in global urine profiles. The robustness of the method was assessed using the data from the spiked validation QC samples, dilution SR series and pooled SR samples measured during the analysis of 144 sample plates from INTERMAP study and 33 sample plates from ARIC study. The DI-nESIHRMS method robustness was also assessed from the analysis of a synthetic test mixture (TM). The details of these measurements are described in SI.

2.4 Nanoelectrospray MS system Chip-based nanoelectrospray infusion analysis was performed using the TriVersa NanoMate system (Advion BioSciences, Ithaca, New York) coupled to a Waters Synapt® G2-Si (Waters MS Technologies, UK) operating in high resolution continuum mode in negative and positive ion modes with automatic polarity switching. The MS experimental conditions are detailed in the Supporting Information.

2.5 INTERMAP and ARIC Study cohorts To exemplify the application of the DI-nESI-HRMS method, it was applied for the analysis of

two

independent

epidemiological

studies.

The

INTERnational

study

of

MAcro/micronutrients and blood Pressure (INTERMAP) is an international population-based cross-sectional study on relations of multiple dietary factors to hypertension among 4,680 men and women ages 40 to 59 years from 17 diverse population samples in China, Japan, United Kingdom and the United States.23,25 Data collected according to a common protocol include eight blood pressure measurements, four 24-h dietary recalls and two timed 24-h urine collections per person (>10.000 urine specimens in total). Borate preservative was added to the urine specimen bottles prior to collection.23 Institutional ethics committee approval was obtained for each site, and all participants gave written informed consent. Atherosclerosis Risk in Communities (ARIC) community-based prospective cohort addresses cardiovascular risk factors and aims to measure urinary markers of oxidative capacity and assess their role as predictors of incident type 2 diabetes. The participants, 15,792 predominantly black and white adults, aged 45-64 years at baseline, were from four U.S. communities and attended four scheduled visits during the period 1987-1998.24 Written informed consent was obtained from all participants and the study protocol was approved by institutional review boards at each clinical site. In this work we analysed a sub-cohort of > 2200 spot urine specimens collected on the fourth visit. 8 ACS Paragon Plus Environment

Page 9 of 35

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Journal of Proteome Research

2.6 Data pre-processing Basic data visualization and quality control was achieved using the MassLynx 4.1 software (Waters Corporation, UK). For the analysis, the raw data were converted to the mzML format with ProteoWizard software29 followed by processing using in-house scripts in Python 2.7.4 as described previously.21 These included averaging of the acquired scans, linear correction of the mass accuracy drift using the observed m/z values of internal standard compounds, interpolation of the data to a final resolution of 0.00056 m/z, peak picking using the SavitzkyGolay algorithm and local linear baseline estimation.30 For the untargeted analysis, a data matrix containing all the spectra in one ion mode interpolated on a common m/z scale was used. Calibration curves plotting, quantification and corresponding calculations were performed in Microsoft Excel. The full-scan spectra were pre-processed and normalised in MATLAB R2014a (MathworksTM). The pre-processing consisted in removal of background signals by subtracting the spectrum of a blank sample from each sample profile, and removal of the m/z features showing high variation in SR samples.

2.7 Data analysis The measured concentration of metabolites (µg/mL) in urine samples were normalised by creatinine concentration (mM, measured as well by DI-nESI-HRMS method in each sample) and used as µmol/mmol creatinine for further analyses. Differences in metabolite concentrations in USA and Japanese populations from INTERMAP study were investigated by linear regression analysis, using the default linear regression modelling tools (lm) in R.31 The dependent variable (initially expressed in metabolite concentration normalized by creatinine, µmol/mmol creatinine) was log-transformed with country (USA or Japan), gender (discrete variables), age and BMI (continuous variables) used as explanatory variables, without interactions (y ~ Population + Gender + Age + BMI). Principal Component Analysis (PCA) and Orthogonal Projection to Latent Structures– Discriminant Analysis (OPLS-DA)32 were performed in SIMCA 14 (Umetrics, Sweden) to further investigate metabolic phenotypic differences in global urine profiles of two populations. The discriminant features were annotated using additional MS/MS experiments, statistical spectroscopy (STOCSY17) and on-line databases (HMDB33 and Metlin34).

9 ACS Paragon Plus Environment

Journal of Proteome Research

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

3.

Page 10 of 35

Results and Discussion

DI-nESI-HRMS method optimization for large-scale metabolic profiling The requirements for validation of regulatory studies and clinical diagnostic assays are normally very strict.27,35,36 In accordance with the FDA guidelines,27 QC samples are incorporated into a study to demonstrate that the analytical method is sufficiently accurate, precise and sensitive enough to measure the actual concentration of a compound. The acceptable levels for accuracy and precision in these studies are set at 15 % by FDA guidelines (except for the LLOQ level QC where 20 % is acceptable).27 However, the method must be “fit-for-purpose”, and the validation criteria should be appropriate for the intended use and meet the predefined needs of the study.37 Multiplexed studies, targeting tens or hundreds of compounds in a single analysis, present a particular challenge, especially in large population studies with a high degree of within- and between-individual variability of urinary metabolites. Different levels of assays have been proposed, each of them obeying a certain level of rigor of validation. For biomarker studies, FDA guidelines are slightly more “relaxed”, following the 20 and 30% rule for accuracy and precision of QC samples at high and LLOQ levels, respectively, which is more appropriate for metabolic profiling studies.28,37 In the present study we used these acceptance criteria taking into account not only the biological variability of the metabolites in urine samples from different populations, but also the limitations of the DIMS method mentioned in the Introduction. Prior to starting the method validation work, we carried out several pre-validation studies in order to assess the stability and purity of the labelled and non-labelled standards (the details are described in SI), selectivity of the method, dilution factor for the urine samples from each study, matrix effect using analyte-free matrix and established the concentration ranges for each metabolite. In order to facilitate the view of the multi-step process, the workflow followed in this work for DI-nESI-HRMS method optimization and validation is depicted in Figure S-1 in Supporting Information. Selectivity. Method selectivity refers to its ability to measure each analyte in the presence of the others in the matrix. This issue needs special control using the DIMS methods as the chromatographic separation step is omitted. We assessed selectivity of DI-nESI-MS by

10 ACS Paragon Plus Environment

Page 11 of 35

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Journal of Proteome Research

measuring the Response Factor for labelled and non-labelled compounds in neat methanol. The serially diluted mixtures of standard stock solutions prepared at the same concentration at four different levels were infused into mass spectrometer. The ratios of the intensities of non-labelled standard to its labelled internal standard were calculated for each level of concentration. For those standards where the ratio value was different from 1 additional MS measurements using the mixtures with different number of the standards were carried out. The standards that interfered with the quantification of endogenous metabolites by having very close accurate mass or experiencing a strong matrix effect were excluded from further analysis. The remaining metabolites together with their biochemical function, internal standards and corresponding m/z values are listed in Table 1 and S-1. In addition, the signal purity of metabolites of interest were analysed by off-line MS/MS analysis of the pooled urine sample and compared to the MS/MS spectra obtained for the authentic standards in neat solvent. The MS/MS spectra obtained for the standards of glutamic acid and phenylacetylglutamine in neat methanol as well as in a spiked and nonspiked urine sample from the INTERMAP study are shown in Figure S-3. For some low level metabolites the MS/MS spectra in pooled urine sample showed the interference from other compounds with close accurate mass caused by resolution of the quadrupole and the absence of chromatographic separation. Dilution study. The assessment of the optimal dilution factor for urine samples is a key step in method optimisation for each sample cohort. We characterized the dilution factors for the urine samples from both INTERMAP and ARIC studies by assessing the effect of dilution (1/10, 1/20, 1/50, and 1/100) on the intensities of endogenous metabolites and the added internal standards. In DIMS, sample dilution improves ionization efficiency and reduces matrix suppression effect caused by the presence of inorganic salts. We found that the optimal dilution factor for the urine samples from the INTERMAP study was of 1/50, whereas for the urine samples from ARIC study it was 1/20. The INTERMAP urine samples were preserved by adding boric acid during the sample collection.23 The presence of boric acid in the urine specimens from the INTERMAP study seemed to enhance the ionization of analytes in DI-nESI-HRMS analysis. We calculated the ratio of instrument responses obtained at 1/20 and 1/50 dilutions for selected metabolites and their internal standards (Figure S-4) in the urine samples from the INTERMAP study. The relative increase of intensities of the labelled compounds was 100% in more diluted urine sample (1/50) while the relative decrease of intensities of endogenous metabolites was minimal (10000 INTERMAP samples was completed in 12 weeks in batches (sample plates) created for each population and clinic visit separately. In order to demonstrate that our method is robust for the long-term analysis we used the data obtained for different types of SR and QC samples from randomly selected sample plates (14 and 10) from the USA and Japanese populations (first visit) respectively. The calibration series, dilution SR series and validation QC samples from each selected sample plates were processed for targeted analysis for two populations. For the untargeted global profile analysis, only pooled SR sample from two populations were processed together in order to assess the analytical against biological variability. The inter-batch CV% values presented in Table S-8 obtained for each metabolite did not exceed 20% for the most part of quantified metabolites in validation QC samples. The largest CV% values were obtained for the most diluted SR samples 1/100 (Table S-8). However, for the majority of the metabolites the 30% cut-off level acceptable for biomarker assays was not exceeded. From the selected sample plates, the concentration values for six metabolites quantified in positive and negative ion modes in global QC samples are shown in Figure S-7. The dashed lines represent the mean concentration ± 2 standard deviation values. The untargeted analysis performed on the pooled SR samples from USA and Japan populations together allowed estimation of the biological and analytical variability. Using OPLS-DA fitted models to explore systematic differences in two populations, clear separation of the two populations was apparent as demonstrated by the cross-validated scores plots in Figure S-8 obtained for the full-scan global profile analysis in negative (A) and positive (B) ion modes. The model characteristics describe models with high predictive values (Q2Y = 0.908 and 0.921, respectively) indicating that the biological differences are higher compared to analytical variability which was less than 20% as shown from the targeted analysis. The precision of the targeted analysis was also assessed for the ARIC study using CV% values of the metabolites quantified in pooled SR samples and dilution SR series measured in all 33 sample plates (Table S-9). For the majority of metabolites in the pooled SR and 1/10 diluted SR samples the CV% values did not exceed 20% with only two of 26 metabolites 15 ACS Paragon Plus Environment

Journal of Proteome Research

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60

Page 16 of 35

having CV% of 20-30%. The higher CV% values were observed, as expected, for the most diluted SR sample 1/50. Additionally, the analytical error and system stability was assessed using the synthetic test mixture (TM) analysed using the same experimental protocol as urine samples. The details and results of this assessment are presented in Supporting Information and in Figure S-8. Briefly, using normalisation by the total intensity, CV% values for all analytes from the mixture measured over three days and after instrument maintenance did not exceed 20%.

Metabolic differences between population phenotypes – application of DI-nESI-HRMS The DI-nESI-HRMS method proposed herein provides a framework for high-throughput metabolic phenotyping of biological samples from large biobank sample collections. After thorough optimization and validation of the method, we applied it for the analysis of >10000 24-h collection urine samples from the INTERMAP study and >2200 spot urine samples from the ARIC study. With all the data we obtained in a in a time window in the order of weeks, we are now able to explore metabolic phenotypes of different populations and obtain systems-level information reflecting genetic (inheritable) and environmental (e.g. diet related) influences. As an example of DI-nESI-HRMS method application at the population level, we present the comparison of metabolic profiles of USA and Japanese population from the INTERMAP study. We used both semi-quantitative and global profile data, and applied univariate and multivariate statistical approaches to identify metabolic differences between the two populations. Metabolite concentrations obtained from the targeted analyses of 2075 urine samples from USA, first visit (1036 male and 1039 female) and 1119 samples from Japan, first visit (560 male and 559 female) were normalised by creatinine measured by DI-nESI-HRMS in each sample, and log-transformed prior to the calculation of a linear regression model with concentration (as µmol/mmol creatinine) as the dependent variable and population (USA, Japan), gender, age and BMI as the explanatory variables, which are also covariates of each other. The test comparing the levels of metabolites in two populations revealed significant differences for 32 out of 39 quantified metabolites (Table 2), p