Subscriber access provided by UNIV OF MISSISSIPPI
Article
An Orthogonal Injection Ion Funnel Interface Providing Enhanced Performance for Selected Reaction Monitoring-Triple Quadruple Mass Spectrometry Tsung-Chi Chen, Thomas L. Fillmore, Spencer A. Prost, Ronald J. Moore, Yehia M. Ibrahim, and Richard D. Smith Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.5b01482 • Publication Date (Web): 24 Jun 2015 Downloaded from http://pubs.acs.org on July 7, 2015
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a free service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are accessible to all readers and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
An Orthogonal Injection Ion Funnel Interface Providing Enhanced
2
Performance for Selected Reaction Monitoring-Triple Quadrupole Mass
3
Spectrometry
4 5 6
Tsung-Chi Chen, Thomas L. Fillmore, Spencer A. Prost, Ronald J. Moore, Yehia M. Ibrahim and Richard D. Smith
7 8 9
Biological Sciences Division, Pacific Northwest National Laboratory, Richland, WA 99352, USA
10 11 12 13 14
Prepared for Analytical Chemistry
15 16 17 18 19 20 21
Corresponding Authors: Yehia M. Ibrahim, E-mail:
[email protected] Richard D. Smith, Email:
[email protected] 22 23 24
1
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Abstract
2
The electrodynamic ion funnel facilitates efficient focusing and transfer of charged
3
particles in the higher pressure regions (e.g. ion source interfaces) of mass spectrometers,
4
and thus provides increased sensitivity. An “off-axis” ion funnel design has been
5
developed to reduce the source contamination and interferences from, e.g. ESI droplet
6
residue and other poorly focused neutral or charged particles with very high mass-to
7
charge ratios. In this study a dual ion funnel interface consisting of an orthogonal higher
8
pressure electrodynamic ion funnel (HPIF) and an ion funnel trap combined with a triple
9
quadrupole mass spectrometer was developed and characterized. An orthogonal ion
10
injection inlet and a repeller plate electrode was used to direct ions to an ion funnel HPIF
11
at 9-10 Torr pressure. Key factors for the HPIF performance characterized included the
12
effects of RF amplitude, the DC gradient, and operating pressure. Compared to the triple
13
quadrupole standard interface more than 4-fold improvement in the limit of detection for
14
the direct quantitative MS analysis of low abundance peptides was observed. The
15
sensitivity enhancement in liquid chromatography selected reaction monitoring (SRM)
16
analyses of low abundance peptides spiked into a highly complex mixture was also
17
compared with that obtained using a both commercial s-lens interface and a in-line dual
18
ion funnel interface.
2
ACS Paragon Plus Environment
Page 2 of 24
Page 3 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Analytical Chemistry
Introduction
2
The sensitivity of measurements using liquid chromatography with electrospray
3
ionization tandem mass spectrometry (LC/ESI/MS/MS) depends significantly upon the
4
overall ion utilization efficiency1,2, including both the effectiveness of the ionization
5
processes and the efficiency of ion transmission from the source to the detector. The ESI
6
efficiency for producing gas-phase ions is related to the solvent evaporation as well as
7
repeated droplet fission3 that can take place at atmospheric pressure but also in the lower
8
pressure regions, introducing both thermodynamic and kinetic constraints upon ion
9
production.2,3
10
Numerous approaches and interface designs for intermediate-pressure ion sampling
11
and transmission devices4-7 have been developed to enhance ion transfer from
12
sub-ambient pressure regions to the vacuum required for MS. The electrodynamic ion
13
funnel has been shown to be broadly effective for the capture and focus of ions over a
14
wide pressure range (30 Torr).5,8,9 Ion funnel designs generally utilize a stack of
15
ring electrodes with gradually decreasing inner diameters. Ions travelling through the ion
16
funnel are confined due to the radio frequency (RF) potentials of 180 degree phase
17
shifted on adjacent electrodes, typically in conjunction with an auxiliary direct current
18
(DC) field to drive ions through the ion funnel5,6 to focus ions though a conductance limit 3
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
to the subsequent stages of the mass spectrometer. While ion funnels effectively transfer
2
ions though sub-ambient pressure regions, they also have a modest focusing effect for
3
larger particles and droplets, especially if entrained in a strong axial gas flow. The
4
mixture of the ions and neutrals (e.g. from an electrosprayed solution) in a high collision
5
rate environment can lead to additional gas phase chemistry that can impact
6
measurements (e.g. by proton transfer), as well as performance degradation due to the
7
deposition on downstream ion optics. Problems become more pronounced as ions are
8
more effectively transferred from ESI sources though multiple inlets in order to increase
9
measurement sensitivity.10,11 Thus, the primary aims of “off-axis” ion introduction are to:
10
1) reduce the interface and related ion optics contamination, and 2) decrease detector
11
noise from excited and fast neutrals or very high m/z particles.12 The charged particles
12
originating from the shrinking droplets are influenced by both gas dynamics and electric
13
fields. Off-axis source concepts have been implemented, e.g. using ion funnels13,14,
14
S-lens,15,16 conjoined ion guides,17,18 and bent RF-only quadrupole ion guides at
15
intermediate pressure regions.12,19 The separation of the ions and neutrals within these
16
devices can be facilitated by additional off-axis electrodes to obstruct neutral species and
17
to produce fields that divert ions away from any directed gas flow.
18
In the present study, we introduce an orthogonal ion funnel-ion funnel trap 4
ACS Paragon Plus Environment
Page 4 of 24
Page 5 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
configuration on a triple quadrupole (QqQ) MS to improve robustness in conjunction
2
with sensitivity. Comparing to conventional off-axis interfaces, incomplete desolvated
3
droplets are unlikely to reach the exit orifice by the orthogonal ion funnel. While ions are
4
sharply turned by 90 degree away from the direction of gas flow, the large acceptance
5
area provided by the ion funnel maximizes the ion transmission efficiency. The reduced
6
directed gas flow from the source and effective elimination of neutral particles can
7
significantly improve both system robustness and detector signal-to-noise ratios. In this
8
work the characterization of the orthogonal ion funnel in a triple quadrupole MS was
9
evaluated and improved limits of detection (LODs) in both single stage MS and LC-SRM
10
modes were achieved.
5
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Instrumentation
2
The dual-stage ion funnel interface used in this work consists of a high pressure ion
3
funnel (HPIF) and an Ion Funnel Trap (IFT)20,21 as shown schematically in Figure 1a.
4
Ions generated in atmospheric pressure were introduced through either an inline (10 mm
5
offset from the instrument centerline) or an orthogonal injection capillary. Following the
6
HPIF is an IFT that can either trap or continuously transmit ions to the ion optics of the
7
next stage. For the experiments presented in this manuscript the IFT was operated in the
8
continuous (non-trapping) mode (i.e. functioning as a conventional ion funnel). As shown
9
in Figure 1b, a mechanical pump connected to the HPIF chamber pumps out neutral
10
species through an exhaust port and aligned with the expanding gas and charged particles
11
from the injection capillary. Stainless steel capillaries with 1 mm i.d. and 10 cm long for
12
both interfaces were chosen to increase the number of ions in conjunction with the HPIF
13
chamber. The pressures of HPIF and IFT chambers were in the range of 8-10 Torr and ~1
14
Torr, respectively, and a 2.5 mm i.d. conductance limit orifice followed the IFT allowing
15
efficient optimal transmission to the triple quadrupole MS (TSQ Vantage, Thermo Fisher
16
Scientific, San Jose, CA). The HPIF incorporated a repeller electrode that aided in
17
directing ions towards the IFT that incorporated a stack of 46 Tin coated copper ring
18
electrodes with 50.8 mm i.d. (i.e. the HPIF straight section in Figure 1) followed by a 6
ACS Paragon Plus Environment
Page 6 of 24
Page 7 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
stack of 72 ring electrodes transitioning from an initial i.d. of 50.8 mm and decreasing
2
gradually to a final electrode i.d. of 2.5 mm (i.e. the HPIF converging section). An RF
3
waveform of equal amplitude but opposite phase was applied to adjacent ring electrodes
4
in conjunction with a DC gradient to drive ion motion through the interface.
5 6
Experimental
7
Chemicals and Materials. For the initial tuning of the instrument, the triple
8
quadrupole calibration stock solution (25 µM polytyrosine-1,3,6 in 1:1 water/methanol
9
that contained 0.1% formic acid, Thermo Scientific Pierce, Rockford, IL) was used to
10
optimize the DC gradient and RF amplitude applied to the HPIF and IFT. For quantitative
11
analysis experiments, samples with different concentrations were prepared including: 1)
12
The mixture of 9 peptides (Supplementary Materials Table S1) purchased from
13
(Sigma-Aldrich, St. Louis, MO) was prepared at concentrations from 6 pmol/µL to 100
14
nmol/µL in an electrospray solution of 1:1 water/methanol that contained 0.1% acetic
15
acid by volume, and 2) synthetic peptides (Thermo Scientific, San Jose, CA) labeled with
16
13
17
carbonic anhydrase, bovine β-lactoglobulin, Escherichia coli β-galactosidase, and
18
Prostate-specific antigen). Heavy peptides were estimated to be >95% pure, were spiked
C/15N on C-terminal lysine and arginine residues for the four target proteins (bovine
7
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
into a tryptic digest of Shewanella oneidensis strain MR-1 to final concentrations ranging
2
from 1 fmol/µL to 7.8 amol/µL. Detailed information on the target peptides is given in
3
Supplementary Material Table S3.
4
LC-SRM Analysis. The prepared solution was analyzed with a nanoAcquity UPLC
5
system (Waters Corporation, Milford, MA) equipped with one analytical capillary
6
column, 100 µm i.d.×100 mm with a 1.7 µm C18 (stationary phase, BEH130, Waters
7
Corporation, Milford, MA), and an autosampler. Solvents used in the UPLC system
8
consisted of 0.1% formic acid in water (mobile phase A) and 0.1% formic acid in 90%
9
acetonitrile (mobile phase B). Two µL of the solution for each concentration were loaded
10
into the analytical column at a flow rate of 300 nL/min. The LC-SRM separation used a
11
binary gradient of 10-15% B in 4.5 min, 15-25% B in 17 min, 25-38.5% B in 11 min,
12
38.5-95% B in 1 min, and 95% B for 4 min.
13
Nano-ESI Source. A chemically etched emitter22 of 20 µm i.d. was directly
14
connected to a 75 µm i.d. capillary (Polymicro Technologies, Phoenix, AZ) for the MS
15
analysis or to a LC analytical column through a zero volume stainless steel union (Valco
16
Instrument Co. Inc., Houston, TX) for LC-SRM analysis. For the MS analysis, sample
17
solutions were directly infused using a syringe pump (Thermo Scientific, San Jose, CA).
18
Data Analysis. MS and MS2 data acquired using the TSQ Vantage were analyzed 8
ACS Paragon Plus Environment
Page 8 of 24
Page 9 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
using an in-house developed data extraction program. The most abundant peaks (MS) or
2
the three most abundant transitions (MS2) for each peptide (Tables 1 and 3) were used for
3
quantification analyses. In the MS quantitation the LOD was defined as the lowest
4
concentration point of target proteins at which the signal-to-noise ratio (S/N) of surrogate
5
peptides was >3 in a mass range of ±2 Da for the target peptides. In the LC-SRM
6
quantitation the tandem MS data from TSQ Vantage were analyzed using Skyline23
7
software to produce the peak area of extracted ion chromatograms (XIC) from multiple
8
transitions monitored for target peptides. The S/N for LC-SRM measurements were
9
calculated by the peak apex intensity over the highest background noise in a retention
10
time region of ± 30 seconds for each target peptide. All data were processed by the
11
program automatically and were later manually inspected to ensure correct peak detection
12
and accurate integration.
13 14 15
9
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Results and Discussion
2
Characterization of the Orthogonal Injection HPIF. The optimal tuning of optics
3
for orthogonal ion injection was expected to be different from the inline injection since
4
ions experience significantly different gas dynamics. The characterization of RF
5
amplitude was first performed to provide optimal radial confinement. During the
6
experiment the pressures of HPIF, IFT and analyzer chambers were maintained at 8, 1
7
and 3×10-6 Torr respectively. Figure 2a shows the Total Ion Current (TIC) for
8
singly-charged polytyrosine-1, 3, 6 as a function of the RF amplitude with frequency of 1
9
MHz in the orthogonal (orange squares) and inline injection (aqua diamonds) modes. The
10
ion signal in the inline mode reached a plateau at lower RF amplitude of 120 Vp-p, in
11
comparison to the orthogonal mode at 160 Vp-p, which indicates that a stronger RF
12
confinement in radial direction is required for orthogonal injection. The higher RF
13
confinement, presumably, is required to overcome the effect of gas dynamics in the
14
orthogonal direction and prevent ions from hitting the side of the HPIF. Figure 2b shows
15
the normalized selected ion chromatogram (SIC) measurements for three individual ions
16
as a function of the RF amplitude in the orthogonal mode. The intensity is normalized to
17
the highest value for each. The highest SICs for ions of m/z 182 (green squares), 508
18
(blue circles) and 997 (red triangles) were obtained with RF amplitudes of 160, 180 and
10
ACS Paragon Plus Environment
Page 10 of 24
Page 11 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
220 Vp-p respectively. These shifts of the curves with RF amplitude are ascribed to the
2
influence of low mass cut off resulting from the different RF trapping potentials. The
3
results indicate that RF >220 Vp-p is sufficient for efficient ion transmission over at least a
4
range of m/z 182 to m/z 997.
5
Figure 2c compares TIC measurements for the orthogonal (orange squares) and
6
inline (aqua diamonds) injection modes as a function of electrical field along the HPIF
7
straight section. The variation of the electrical field was implemented by changing the
8
voltages applied at the repeller and the HPIF entrance (and which then dictates the
9
potentials applied to the HPIF electrodes through its resistor chain). As can be observed
10
in Figure 2c ions, in both injection modes, are still transmitted through the straight region
11
with a negative electrical field gradient due to gas dynamic effects, but as expected fewer
12
ions were transmitted in the orthogonal injection mode. For instance at a repulsive
13
electric field of 30V/cm no ions were transmitted in the orthogonal injection mode while
14
for the inline capillary 70% of the maximum signal (for inline injection mode) is still
15
observable.
16
exiting the inlet capillary, which also can transmit undesirable charged droplets and
17
neutral particles. The ion currents increased with a positive electrical field gradient in
18
both modes. The highest observable current was at the gradient of 2.2 V/cm for inline
These results indicate that ions were accelerated by the expansion of gas
11
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
injection and 35.1 V/cm for orthogonal injection mode when the repeller voltage reached
2
to its maximum output at 380 V due to the limitations power supply. The higher field
3
strength for the optimal transmission in the orthogonal injection mode reflects the
4
differences in gas dynamics for the two arrangements. Contrary to the effect of DC
5
gradient observed in Figure 2c the repulsive DC field on the converging section show no
6
effect (Supplementary Materials Figure S1) on ion intensities in either modes, orthogonal
7
or inline, due to the reduced gas dynamics in the region far away from the inlet capillary.
8
Figure 2d shows the effect of the pressure on TIC measurements in the orthogonal
9
mode. The ion intensities were measured as the function of HPIF chamber pressure while
10
the IFT chamber pressure was maintained at 1 Torr and the RF parameters for both
11
funnels were fixed at 220 Vp-p/1 MHz for HPIF and 184 Vp-p/1.3 MHz for the IFT. The
12
pressures of both chambers were adjusted by choking the flow to the mechanical pumps.
13
MS results optimized at 8-12 Torr, corresponding to 95% or above the maximum
14
intensity achieved in the orthogonal injection mode. Insufficient RF focusing at higher
15
pressure (>12 Torr) resulted in a decrease in ion intensities in both injection modes (and
16
which can be mitigated by further increasing the RF amplitude).9
17
The sensitivity of the orthogonal HPIF-IFT interface was first evaluated using a
18
nine peptide mixture. A label-free MS quantitation of a peptide mixture (Supplementary 12
ACS Paragon Plus Environment
Page 12 of 24
Page 13 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
Materials Table S1) was performed prior to the LC-SRM experiment, and baselines were
2
established using both a commercial interface (s-lens) and an inline HPIF-IF interface on
3
the same triple quadrupole MS. During the experiment the capillary temperature was set
4
at 140 ⁰C for HPIF and 310 ⁰C for commercial interface (Figure 1a) and the resolution
5
setting on triple quadrupole is with peak width of 0.7 Da for Q1 and Q3. The sensitivities
6
of MS quantitation and LC-SRM were assessed based on the LOD and limit of
7
quantification (LOQ) values. Figure 3 shows three randomly selected examples of
8
calibration curves from Kemptide (Figure 3a and 3b), Renin (Figure 3c and 3d) and
9
Fibrinopeptide A (Figure 3e and 3f) by the orthogonal (red squares), inline HPIF-IF
10
(green circles) and commercial standard s-lens interface (blue triangles). The results in
11
Figure 3a, c and e show that the 4- to 32-fold improvements in LODs and overall higher
12
S/N was obtained by the ion funnel configurations (orthogonal and inline) in comparison
13
to the standard source. In Figure 3 b, d and f the results of intensity measurement indicate
14
that the best LODs with good linearity were obtained by the orthogonal injection in
15
comparison to the inline injection mode. The detailed measurement of all nine target
16
peptides in Supplementary Materials Table S2 shows that orthogonal injection HPIF
17
significantly improves S/N and coefficient of variation (CV) for the analytes less than
18
6.25 nmol levels. Overall the LOD and LOQ values obtained from each surrogate peptide 13
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
by orthogonal injection show sensitivities 2- to 8-fold better than inline injection HPIF
2
and 8- to 32-fold better than the commercial s-lens interface.
3
A further evaluation on the utility of the orthogonal injection was conducted using a
4
label-free LC-SRM analysis of 18 peptides (Supplementary Materials Table S3) from
5
four proteins prepared at equal concentrations ranging from 31.2 amol/µL to 2 fmol/µL
6
for the three interfaces. Figure 4 illustrates the calibration curves of selected 6 target
7
peptides out of the 18 peptides analyzed. The scan width for the QqQ was set at 0.002
8
m/z and 15 ms for the scan time in the SRM mode. The sensitivities of LC-SRM data
9
were assessed based on the LOD and the linearity of the calibration curves. The
10
integrated peak areas of three monitored transitions for each peptide were recorded as a
11
function of the amount loaded
12
detectable concentration from orthogonal injection HPIF are estimated to be 2-fold better
13
than using inline injection HPIF and 2- to 8-fold better than the s-lens interface for most
14
of the peptides. The summary of quantitative performance of the 18 peptides using three
15
interfaces is shown in Supplementary Materials Table S4. The better detection limit was
16
obtained in 13 out of 18 peptides by orthogonal, 6 by inline injection HPIF and 3 from
17
standard interface. For linearity, moderate improvements were shown using orthogonal
18
injection HPIF upon the LC-SRM which also attributed to reduced chemical background.
at the LC column. The results indicate that the lowest
14
ACS Paragon Plus Environment
Page 14 of 24
Page 15 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1 2
Conclusions
3
An orthogonal injection ion funnel interface has been designed, implemented,
4
characterized and evaluated in conjunction with a LC-SRM triple-quadrupole MS.
5
Orthogonal extraction of ions from the source gas flow path is shown to significantly
6
enhance the sensitivity for detection of analytes from peptide mixture and complex
7
biological samples. In comparison with the standard triple-quadrupole interface, the
8
significant improvement on S/N measurement and 4- to 32-fold enhancement of LOD for
9
single stage MS quantitation were achieved using a peptide mixture. For the LC-SRM
10
based protein quantification, the XICs of monitored peptides showed 2- to 8-fold
11
enhancement of the LOD with good linearity of measurements.
12 13
15
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Acknowledgments
2
The authors thank Dr. Tujin Shi, Dr. Ian Webb, Dr. Jonathan Cox and Dr. Jia Guo for
3
helpful discussions. The authors also thank Mr. Grant Fujimoto for the assistance for the
4
data processing programming. Portions of this research were supported by the NIH
5
National Cancer Institute (1R33CA155252) and General Medical Sciences Proteomics
6
Research Resource at Pacific Northwest National Laboratory (PNNL) (GM103493-12),
7
the Laboratory Directed Research and Development Program at PNNL, and the
8
Department of Energy Office of Biological and Environmental Research Genome
9
Sciences Program under the Pan-omics program. All the experiments were performed in
10
the Environmental Molecular Sciences Laboratory, a US Department of Energy (DOE)
11
national scientific user facility located at PNNL in Richland, Washington. PNNL is a
12
multiprogramming national laboratory operated by Battelle for the DOE under contract
13
DE-AC05- 76RLO01830.
14
16
ACS Paragon Plus Environment
Page 16 of 24
Page 17 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1 2
Analytical Chemistry
Supporting Information Available This information is available free of charge via the Internet at http://pubs.acs.org/
3
17
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1 2 3
References
4 5
Spectrometry 2015, 26, 55-62. (2) Marginean, I.; Page, J. S.; Tolmachev, A. V.; Tang, K.; Smith, R. D. Analytical
6 7 8
Chemistry 2010, 82, 9344-9349. (3) Kebarle, P.; Tang, L. Analytical Chemistry 1993, 65, 972A-986A. (4) Dodonov, A.; Kozlovsky, V.; Loboda, A.; Raznikov, V.; Sulimenkov, I.; Tolmachev, A.;
(1) Cox, J.; Marginean, I.; Smith, R.; Tang, K. Journal of The American Society for Mass
9 10
Kraft, A.; Wollnik, H. Rapid Commun. Mass Spectrom. 1997, 11, 1649-1656. (5) Kelly, R. T.; Tolmachev, A. V.; Page, J. S.; Tang, K.; Smith, R. D. Mass Spectrometry
11 12
Reviews 2010, 29, 294-312. (6) Gerlich, D.; Wiley: New York, 1992, pp pp. 1-176.
13 14
(7) Bruins, A. P. Mass Spectrometry Reviews 1991, 10, 53-77. (8) Shaffer, S. A.; Tang, K.; Anderson, G. A.; Prior, D. C.; Udseth, H. R.; Smith, R. D.
15 16
Rapid Commun. Mass Spectrom. 1997, 11, 1813-1817. (9) Ibrahim, Y.; Tang, K.; Tolmachev, A. V.; Shvartsburg, A. A.; Smith, R. D. Journal of
17 18 19 20 21 22
the American Society for Mass Spectrometry 2006, 17, 1299-1305. (10) Kim, T.; Udseth, H. R.; Smith, R. D. Analytical Chemistry 2000, 72, 5014-5019. (11) Kim, T.; Tang, K.; Udseth, H. R.; Smith, R. D. Analytical Chemistry 2001, 73, 4162-4170. (12) Whitehouse, C. M.; Welkie, D. G.; U.S. Patent US8507850 B2, 2013. (13) Tang, K.; Tolmachev, A. V.; Nikolaev, E.; Zhang, R.; Belov, M. E.; Udseth, H. R.;
23 24 25 26 27 28 29 30 31 32 33
Smith, R. D. Analytical Chemistry 2002, 74, 5431-5437. (14) Mordehai, A.; Werlich, M. H. U.S. Patent 8324565 B2, 2009. (15) Senko, M. W.; Kovtoun, V. V. U.S. Patent 7514673 B2, 2013. (16) Wouters, E. R.; Splendore, M.; Mullen, C.; Schwartz, J. C.; Senko, M. W.; Dunyach, J. J. In Presented at the 57th Conference of the American Society for Mass Spectrometry: Philadelphia, PA, 2009. (17) Giles, K. U.S. Patent 12/679,139, 2011. (18) Giles, K.; Wildgoose, J.; Williams, J.; Cecco, M. D. In Presented at the 61th Conference of the American Society for Mass Spectrometry: Minneapolis, Minnesota, 2013. (19) Fialkov, A. B.; Steiner, U.; Jones, L.; Amirav, A. International Journal of Mass
34 35
Spectrometry 2006, 251, 47-58. (20) Belov, M. E.; Prasad, S.; Prior, D. C.; Danielson, W. F.; Weitz, K.; Ibrahim, Y. M.;
36
Smith, R. D. Analytical Chemistry 2011, 83, 2162-2171. 18
ACS Paragon Plus Environment
Page 18 of 24
Page 19 of 24
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
(21) Ibrahim, Y.; Belov, M. E.; Tolmachev, A. V.; Prior, D. C.; Smith, R. D. Analytical
2 3
Chemistry 2007, 79, 7845-7852. (22) Kelly, R. T.; Page, J. S.; Luo, Q.; Moore, R. J.; Orton, D. J.; Tang, K.; Smith, R. D.
4 5
Analytical Chemistry 2006, 78, 7796-7801. (23) MacLean, B.; Tomazela, D. M.; Shulman, N.; Chambers, M.; Finney, G. L.; Frewen,
6 7
B.; Kern, R.; Tabb, D. L.; Liebler, D. C.; MacCoss, M. J. Bioinformatics 2010, 26, 966-968.
8 9
19
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Repeller HPIF
(a)
IFT
Page 20 of 24
Triple Quadrupole MS Q1
Q0
Q2 1.5 mtorr
Inline Orthogonal
8 torr
4x10-6 torr
1.1 torr Q0
s-Lens
4x10-6 torr
Q3
Detector
1
(b)
2 3 4 5 6 7 8
Figure 1. (a) Schematic diagram of the triple quadrupole MS with the standard source (s-lens), orthogonal and inline high pressure ion funnel and ion funnel trap, and the ion optical configuration for the each segment along HPIF and IFT. (b) Section view of the orthogonal injection HPIF-IFT atmospheric pressure interface. Inlet capillaries of 1 mm i.d. are used for in-line and orthogonal injection.
9 10 11
20
ACS Paragon Plus Environment
Page 21 of 24
(b) 3x10
8
2x10
8
100
Intensity (%)
Intensity
(a)
1x108 Orthogonal Inline
0 100
200
80 60 RF frequency: 1 MHz
40
m/z 182 m/z 508 m/z 997
20 100
300
(c)
200
300
RF Amplitude (VPP)
RF Amplitude (VPP)
(d)
3x108
95%
Intensity (%)
100
Intensity
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1x108
Orthogonal Inline
1 2 3 4 5 6 7 8
-40
-20
0
20
60 40 20
0 -60
80
8
40
10
12
14
16
18
Pressure (Torr)
E (V/cm)
Figure 2. (a) Comparison of total ion current as the function of RF amplitude in orthogonal and inline injection modes. (b) Comparison of ion intensities from ions of m/z 182, 508 and 997 as the function of RF amplitude varied from 60 to 300 Vp-p in orthogonal injection mode. (c) Measurement of ion transmission efficiency at HPIF straight section, and (d) Effect of pressure on ion transmission for the HPIF in orthogonal and inline injection mode.
21
ACS Paragon Plus Environment
Analytical Chemistry
(b)
200
Kemptide Orthogonal Inline sLens
100
0
(c)
0
100
Kemptide 2+, m/z 387 1x107
1x108
(d)
0 0
0
25
50
0
6x10
200
Renin 3+, m/z 587
3x106
3x107
(f) Fibrinopeptide
0 0
0
50
400
7 5x106
(e) 120
25
Orthogonal Inline
Renin
30
0
8
0
Intensity
s/n
60
50
2x10
3x107
Intensity
s/n
(a)
5x10
12
24
0
200
400
7
Fibrinopeptide A 2+, m/z 769
60
0
1 2 3 4 5
Intensity
3x106
s/n
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 22 of 24
0
50
100
2x106
3x107
0 0
0
12
24
0
Concentration (nmol/µL)
200
400
Concentration (nmol/µL)
Figure 3. Three selected calibration curves from the 9 peptide mixture for concentrations ranging from 200 pmol/µL to 400 nmol/µL using orthogonal injection HPIF, inline injection HPIF and standard triple quadrupole atmospheric pressure interfaces.
22
ACS Paragon Plus Environment
Page 23 of 24
EDLIAYLK
Intensity (a.u.)
3.5x108
1.8x108
0.0
2.5x108
0.5
Intensity (a.u.)
1.0
1.5
0.0
2.0
IVGGWECEK
0.0
1.0
1.5
2.0
LFTGHPETLEK
1.8x108
0.5
1.0
1.5
0.0
2.0
LWSAEIPNLYR
0.5
1.0
1.5
2.0
TGPNLHGLFGR
1.2x108
1.8x108
0.0
0.5
3.5x108
1.5x108
3.5x108
6.0x107
0.5
1.0
1.5
0.0
2.0
Concentration (fmol/µL)
0.5
1.0
1.5
2.0
Concentration (fmol/µL)
1 2 3 4 5 6
HGFLPR
5.0x108
Orthogonal Inline sLens
3.0x108
Intensity (a.u.)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
Figure 4. Typical calibration curves for 18 peptide mixed solution at concentration from 1.56 amol/µL to 2 fmol/µL using LC through orthogonal injection HPIF, inline injection HPIF and standard triple quadrupole atmospheric pressure interfaces to MS in SRM mode.
23
ACS Paragon Plus Environment
Analytical Chemistry
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
1
Table of Contents Graphics
2
24
ACS Paragon Plus Environment
Page 24 of 24