Theory of Plate Height in Gas Chromatography - Analytical Chemistry

Richard. Kieselbach. Anal. Chem. , 1960, 32 (7), pp 880–881. DOI: 10.1021/ac60163a049. Publication Date: June ... C. Gaget , D. Morel , J. Serpinet...
1 downloads 0 Views 260KB Size
Gas Chromatography of Solid Organic Compounds SIR: Although gas chromatography has rapidly become a significant tool in all areas of organic-analytical chemistry, a major gap still exists in the general applicability of the method. Solid organic compounds generally have such high boiling points t h a t thermal decomposition of both t h e sample and the column liquid becomes a problem at the high temperatures usually necessary for operation. To minimize these difficulties, conditions have been adjusted so t h a t gas chromatograms may be obtained at temperatures as much as 250" C. below the boiling point of a component. The method depends on the fact that even a t these low temperatures many compounds have a vapor pressure sufficient for column operation. Retention times have been greatly decreased by reducing the amount of liquid substrate to 0.05 t o 0.2%, instead of the 20 to 30% usually used. Such small amounts of liquid phase could not be used with conventional solid supports such as Celite and firebrick, because of strong surface absorption effects TI hich Fere eliminated by the use of glass microbeads (200-micron diameter) as support for the liquid phase. It was also necessary, because of the low vapor pressure of the compounds, to reduce the sample size to a few tenths of a

milligram t o prevent saturation of the gas phase. A flash vaporizer was used for introduction of the sample and B-as operated a t 375" C. Samples were injected as solutions in a volatile solvent. A chromatogram of n-octacosane (m.p. 61" C., b.p. 412" C.) was obtained using a 2-meter, 0.05% siliconoil column. The retention time was 21 minutes at 200" C. Anthracene (m.p. 216" C., b.p. 354" C.) was eluted from a similar column a t a temperature of 100" C. The method seems to be applicable to a wide variety of organic molecules, as evidenced by the fact that chromatograms have been obtained with retention times less than 20 minutes for the following compounds a t the column temperatures noted: chrysene (m.p. 250" C., b.p. 448" C.) a t 190" C.; anthraquinone (m.p. 286" C., b.p. 380" C.) a t 175" C.; p-dibromobenzene (m.p. 87" C., b.p. 218" C.) a t 103" C.; octadecanol (m.p. 59" C., b.p. 210" C. a t 15 mm.) at 200" C.; 1,3diphenylbenzene (m.p. 86" C., b.p. 363" C.) a t 200" C.; methyl oleate (b.p. 216" C. a t 20 mm.) a t 148" C. The number of theoretical plates in such columns is below those reported for conventional chromatographic columns. However, other factors, especially temperature, are important and greatly

enhance the separation of similar compounds. For example, a mixture of isomeric eicosanes (n-eicosane, m.p. 37" C., b.p. 205" C. a t 15 mm.) could be resolved a t a temperature of 127" C., while n-octadecane was eluted with the air peak, and a-dodecane was not eluted after 90 minutes. Operation of gas chromatography columns a t low temperatures has obvious advantages. It is possible to n ork ITith higher boiling materials without thermal decomposition; thermal conductivity detectors have higher sensitivity; and a much wider choice of liquid substrate is possible. Over 500 chromatograms have been obtained using this technique, with results which would be difficult, if not impossible by conventional methods.

Bristol Laboratories Syracuse, N. Y.

C. HISHTA J. P. MESSERLY R. F. RESCHEB

Cornel1 University D. H. FREDERICKS Baker Laboratory W. D . COOKE Ithaca, K.Y. RECEIVEDfor review March 18, 1960. Accepted April 5, 1960,

Theory of Plate Height in Gas Chromatography SIR: The van Deemter equation (1) has been generally accepted as an expression for the relation between the height of an equivalent theoretical plate (HETP) and the physical parameters in a packed gas chromatographic column. Jones (3) developed an equation which differed from that of van Deemter primarily in the relative importance assigned to the roles of liquid and gaseous diffusion processes. Recent work has shown that neither of these equations is correct and t h a t the experimental data are accurately represented by another equation, also developed by Jones ($), which reconciles the differences b e h e e n the two earlier equations. This new equation clarifies much of the confusing and often contradictory information previously published on this subject. The new equation, integrated to allow for column pressure drop, is: 880

ANALYTICAL CHEMISTRY

The first three terms of Equation 1 are those of the familiar van Deemter equation. If Sa, is zero, as for a n inert sample, or very large, the C term becomes insignificant. If the van Deemter equation were correct, HETP nould approach A as a limit with increasing gas velocity in those cases. However, HETP goes through a minimum for all samples when the data are appropriately corrected for the effects of volumes external to the column and of the detector time constant. The coefficients in Equation 1 were evaluated by the method of least squares

from data obtained a-ith samples of air (Sa== 0), butane = 2), and cyclohexane (Sa== 31) on a 1-meter column of 10% Don. Corning 200 silicone oil, 50-centistoke viscosity, on 100- to 120mesh Chromosorb. Table I presents the results. This set of coefficients s h o w d a n excellent fit to the data from all three samples. The contribution of each term to the total plate height at optimum velocity is also shown in Table I. None of the terms can be neglected and the contribution of each of the last three terms mill become dominant for some particular value of Saz. Additional data and a full discussion of this work will be submitted to this journal for future publication. NOMENCLATURE

A

=

velocity-independent constant, of random flow pattern in a packed column, cm.

B

= coefficient of axial or longitudinal diffusion in gas phase of resistance t o mass transfer in liquid phase coefficient of resistance to mass transfer in moving gas phase diffusion coefficient of sample into carrier gas a t column outlet pressure, sq. em. see.-' diffusion coefficient of sample into column liquid phase, sq. cm. see.-' effective liquid film thickness, em. effective moving-gas diffusional path length, cm. effective stagnantcgas diffusional path length, cm. coefficient of relatively stagnant gas within column packing

D

=

Do0 =

DC dl

=

=

dm = d.

=

E

=

fi=

1°F (a)2

average HETP = em. = effective length of column packing, cm. pi = column inlrt absolute pressure, atm. P o = column outlet absolute pressure, atm. sa,= relative separation of sample peak from air peak =

L

t.

I.

Table

c = coefficient

A , Cm. 0.008

Results of Evaluation of Coefficients

c

B

1.16

di 2 Sec. Dl, 0.00566

Sq. Cm.

D dm',

E ds2, Sq. Cm.

0.000182

0.000576

Contribution of Individual Terms to

Sample Air Butane Cyclohexane

A 0.008

0.008 0.008

B 0.026 0.021 0.016

elution time of air, see. width of peak, measured between base fine intercepts of tangents to inflection points of peak, sec. u. = carrier gas velocity a t outlet of column packing, em. see.-'

t, t,

= =

LITERATURE CITED

(1) Deemter, J. J. van, Zuiderweg, F. J., Klinkenberg, A., Chem. Eng.Sci. 5, 271

(1956).

C 0

0.014 0.003

R Optimum, Cm. D E 0 0.005 0.013

0.026 0.003 0.000

(2) Jones, W. L., Gas Chromatography Symposium, U. S. Public Health Service, Cincinnati, Ohio, Feb. 21, 1957. (3) Jones, W. L., Southwide Chemical Conference, ACS-Instrument Society of America, Memphis, Tenn., Dec. 6, 1956. RICHARD KIESELBACH Engineering Research Laboratory Engineering Department E. I. du Pont de Semours & Co., Inc. Wilmington, Del. RECEIVEDfor review April 4, 1960. Accepted April 18,1960.

Determination of Primary Nitroparaffins by the Nitrous Acid Reaction SIR: The determination of nitroparaffins and particularly of primary nitroparaffins was reviewed recently (3). -4 spectrophotometric method for determining microgram quantities of primary nitroparaffins by coupling with diazotized sulfanilic acid was described ( 3 ) . I n the course of investigating the gas-phase products of olefins and nitrogen oxides ( I ) , the reactions of some representative members of various classes of nitroparaffins with nitrous acid also were measured. This type of reaction goes back to the classical red, white, and blue reaction of Meper and Locher (4). The blue reaction of nitrous acid with secondary nitroparaffins was investigated as a colorimetric procedure (8). I n the present work the nitrous acid reaction with primary nitroparaffins is of immediate interest, but the reactions of secondary nitroparaffins, dinitroparaffins, and a variety of nitroalcohols also are measured. EXPERIMENTAL

Nitro Compounds. T h e sources and purities of t h e nitroalkanes used were discussed previously (3). The 2nitro-1-butanol and 2,2-dinitropropane

were obtained from K & K Laboratories, Jamaica, N. Y. The 2-nitro-2-methyl1-propanol, 2-nitro-2-methyl-1,3-propanediol, 2-nitro-2-ethyl-l,3-propant diol, and tris(hydroxymethy1)nitromethane were provided by Commercial Solvents Corp. Procedure. To 2 ml. of a n aqueous solution of t h e sample containing t h e nitroalkane, a d d a b o u t 0.1 gram (one pellet) of potassium hydroxide t o ensure solution; then a d d 0.25 gram of potassium nitrite and 0.6 ml. of 1 8 N sulfuric acid. Shake this solution with 10 ml. of diethyl ether and separate t h e layers, saving t h e ether layer. Add 3 ml. of water and 4 or 5 drops of saturated sodium hydroxide solution to the ether layer, shake, and transfer the aqueous layer to the absorption cell. Read the absorbance a t 330 mp 5 minutes after mixing. RESULTS AND DISCUSSION

The final alkaline aqueous extract of 1-nitropropane has a n absorption band a t 330 mp. This absorption results from the formation of sodium propylnitrolate. Sodium ethylnitrolate has been reported to have a n absorption a t 325 m l with a molar absorptivity of about 7500 mole-' liter em.-' (6).

I n the present work, the absorbance of the propylnitrolate anion was found to be linear with concentration of l-nitropropane from 2 to a t least 50 y per rnl. The absorptivity was 0.054 =t0.006 y-' ml. em.-', corresponding to a molar absorptivity of about 6300 mole-' liter em.-'. Sodium propylnitrolate decomposes slowly with time as do the other alkylnitrolates (6). Because the decrease in absorbance amounts to 10% in the first

Table 1. Absorptivities of a Number of Aliphatic Nitro Compounds

Absorptivity, a, Y-'

Compounds 1-Kitropropane 2-Kitropropane 2,2-Dinitropropane 2-Nitro-1-butanol 2-Nitro-2-methyl-1-propanol 2-Xitro-2-methyl- 1,3propanediol

bll. Cm.-'

diol Tris(hydroxymethy1)nitromethane

0.04

2-Nitro-2-ethyl-1,3-propane-

VOL. 32, NO. 7, JUNE 1960

0.054

0.008 0.00

0.035 0.002 0.05

0.025

881