Electromembrane extraction of unconjugated fluorescein isothiocyanate from solutions of labeled proteins prior to flow induced dispersion analysis Magnus Saed Restan, Morten Enghave Pedersen, Henrik Jensen, and Stig Pedersen-Bjergaard Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.9b00730 • Publication Date (Web): 30 Apr 2019 Downloaded from http://pubs.acs.org on April 30, 2019
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
2 3 4 5
Manuscript for Analytical Chemistry
Electromembrane extraction of unconjugated fluorescein isothiocyanate from solutions of labeled proteins prior to flow induced dispersion analysis
6 7 8 9 10 11 12
Magnus Saed Restana, Morten E. Pedersenb , Henrik Jensenb,c, Stig Pedersen-Bjergaarda,c aDepartment
of Pharmacy, University of Oslo, P.O. Box 1068 Blindern, 0316 Oslo, Norway ApS, c/o University of Copenhagen, Universitetsparken 2 2100 Copenhagen, Denmark cDepartment of Pharmacy, Faculty of Health and Medical Sciences, University of Copenhagen, Universitetsparken 2, 2100 Copenhagen, Denmark
bFIDA-Tech
13 14 15
Abstract
16
In this initial research on feasibility, removal of unconjugated fluorescein isothiocyanate (FITC)
17
after fluorescent labeling of human serum albumin (HSA) with electromembrane extraction (EME)
18
was investigated for the first time. A 100 µL solution of 0.1 mg/mL HSA was fluorescently labeled
19
with 0.01 mg/mL FITC in a molar ratio of 10:1 in an Eppendorf tube for 30 min under agitation
20
and absence of light. Then the labeled solution was transferred to 96-well EME with 3 µL 0.1 %
21
(w/w) Aliquat 336 in 1-octanol as the supported liquid membrane (SLM) and 200 µL 10 mM NaOH
22
as waste solution. EME was performed for 10 min with a voltage of 50 V, with the anode in the
23
waste solution and at 900 rpm agitation. Negatively charged and unconjugated FITC was extracted
24
electro-kinetically into the SLM and to the waste solution. Analysis of purified samples, by Taylor
25
dispersion analysis (TDA), showed a 92 % removal of unconjugated FITC (FITC clearance: 92 %,
26
RSD: 3 %), while 79 % of the HSA/FITC complex remained in the sample (Protein retention: 79 %,
27
RSD: 18 %). Conserved functionality of the HSA/FITC complex after EME was proven by a binding
28
affinity study with anti-HSA using flow induced dispersion analysis (FIDA). In this real sample, the
29
dissociation constant (Kd) and hydrodynamic radius of the complex were determined to be 0.8 µM
30
and 5.87 nm respectively, which was in concordance with previously reported values.
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 2 of 19
Introduction
31 32 33
Electromembrane extraction (EME) is a three-phase microextraction technique developed for
34
analytical applications, and was presented for the first time in 20061. The technique is a
35
combination of liquid phase microextraction (LPME) and electrophoresis, and has been reported
36
for extraction of chargeable analytes from aqueous samples such as biological fluids2-4,
37
environmental waters5,6, and beverages7. In EME, analytes are extracted through a supported
38
liquid membrane (SLM) and into aqueous acceptor phase. The SLM is an organic solvent
39
immobilized inside the pores of a thin polymeric membrane. The main driving force for mass
40
transfer in EME is an electrical field sustained across the SLM. Two electrodes connected to an
41
external power supply are introduced into the system, one in the sample and one in the acceptor
42
phase. The electrical field force charged analytes to migrate towards the electrode of opposite
43
polarity. For the extraction of basic (cationic) analytes, the cathode is placed in the acceptor phase
44
while the anode is placed in the sample. For acidic analytes, the direction of the electrical field is
45
reversed. Since the electrical field only affects charged compounds, pH of the sample and acceptor
46
phase has to be controlled to a level where the analytes of interest are fully charged (i.e two-three
47
pH units from their pKa). EME can typically be completed in 5-10 minutes1,8, and the adaptation of
48
EME to a 96-well format open for high-throughput operation9. In the current literature, EME has
49
been used for analytical extraction of target analytes from complex samples. Target analytes have
50
been transferred electro-kinetically from the original sample solution, through the SLM, and into
51
pure aqueous solution containing acid or base as acceptor phase. Subsequently the analytes in the
52
acceptor phase have been measured by chromatography, mass spectrometry, or related
53
instrumental techniques.
54
Recently, EME was utilized for purification for the first time. In one paper, EME was used to
55
remove and recycle template molecules after polymerization of molecularly imprinted polymers10.
56
In the second paper, EME was used for rapid and efficient salt removal from minute volumes of
57
saline samples11. From a conceptual point of view, these papers represented a new direction for
58
EME and new types of applications may be developed. To explore this potential, fundamental
59
research is requested, and in the present work we test for the first time EME for removal of
60
unconjugated fluorescent reagent from protein samples. Fluorescence labeling is a common way
61
to enhance the sensitivity and detectability of proteins and involves labeling of purified protein
62
with a fluorescent reagent and subsequently detection by fluorescence spectrophotometry12. To
63
ensure effective labeling of the protein, surplus fluorescent reagent is added to the protein
64
solution. However, in order to suppress the background level of fluorescence, the samples need to
65
be purified to remove unconjugated fluorescent reagent
13-15.
ACS Paragon Plus Environment
This is of general importance in
Page 3 of 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
66
experiments with fluorescence detection, including those conducted by flow induced dispersion
67
analysis (FIDA). In FIDA, the in-capillary diffusivity of an analyte is used to calculate the analyte
68
concentration16-18. FIDA can also be used in binding studies of antibodies to fluorescently labeled
69
protein, and removal of unconjugated fluorescent reagent is mandatory for accurate calculation
70
of binding constants through the change in diffusivity.
71
The most common fluorescent reagents are derivatives of fluorescein, witch have different
72
fluorescence, stability and protein binding properties. Fluorescein isothiocyanate (FITC) is one
73
example, and is frequently used19. Existing techniques for removal of unconjugated FITC include
74
gel filtration20, dialysis21, and purification columns for dye removal22,23. The mutual goal for these
75
techniques is (1) to remove as much as possible of unconjugated fluorescent reagent (high reagent
76
clearance) and (2) to avoid loss of labeled protein (high protein retention). The existing
77
techniques have proven successful in the separation of labeled protein and unconjugated
78
fluorescent reagent24,25 but can be time and labor intensive26, and with a loss of labeled protein26.
79
The current paper is the first report on EME removal of unconjugated fluorescent reagent from
80
protein solutions. The experimental work was fundamental and intended as an initial research on
81
feasibility. Although real samples were included in the current experiments, it should be
82
emphasized that more work is required before the concept can generally be applied as part of
83
experimental work with proteins. FITC was selected as model substance for the fluorescent
84
reagent. FITC is an ideal candidate for EME (Figure S1). The carboxylic group can be ionized in
85
neutral and alkaline solution, which is required for EME. In addition, mass transfer is favored by
86
the small size (389.4 Da) and hydrophobic character (log P 4.8) of FITC. The common proteins
87
cytochrome C, human serum albumin (HSA), and myoglobin were selected as model proteins. First
88
EME was optimized for removal of FITC (high FITC clearance), and for retention of protein in the
89
sample (high protein retention). Second, samples of FITC labeled HSA was processed with EME
90
and subjected to Taylor dispersion analysis (TDA)27-30, to further verify EME performance and to
91
document the stability of the HSA/FITC complex during EME. Finally, EME was combined FIDA
92
for a real sample application, where binding and affinity studies of HSA and anti-HSA antibody
93
were performed.
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Experimental
94 95
Chemicals and solutions
96
FITC,sodium tetraborate decahydrate, boric acid, 2-nitrophenyl octyl ether (NPOE), 1-octanol,
97
methanol, acetone, horse heart myoglobin, bovine heart cytochrome C, HSA, and monoclonal
98
antibody against HSA (mouse, IgG1) were purchased from Sigma Aldrich (St Louis, Missouri, USA).
99
Sodium chloride and sodium hydroxide (NaOH) were purchased from Merck (Darmstadt,
100
Germany). Aliquat 366 was purchased from Cognis Corporation (Cincinnati, Ohio, USA). Deionized
101
water was obtained from a Milli-Q water purification system (Molsheim, France)
102
A stock solution of 1 mg/mL FITC dissolved in acetone was prepared freshly each day and diluted
103
with either 50 mM borate buffer pH 9.20 or the reaction buffer (50 mM borate buffer pH 9.20 with
104
150 mM NaCl) to the working concentration of 0.01 mg/mL (2.57·10-5M). Solutions of each of the
105
proteins were made to a concentration of 0.1 mg/mL for the optimization studies. For the FITC
106
labelling reaction, 10 mg HSA was weighed and dissolved in 10 mL 50 mM borate buffer pH 9.20
107
providing a molar concentration of 1.50·10-5 M. All solutions were stored in darkness at < 8°C.
108
Electromembrane extraction (EME)
109
Equipment used for EME is illustrated in Supporting Information Figure S2. A laboratory built 96-
110
well stainless steel plate with 0.5 mL wells was used as compartment for the waste solutions and
111
as anode (96-well waste reservoir plate). A 96-well MultiScreen-IP filter plate with polyvinylidene
112
fluoride (PVDF) filter membranes with 0.45 µm pore size (Merck Millipore Ltd, Carrigtwohill,
113
Ireland) was used as SLM and compartment for the samples (96-well filter plate). The volume of
114
each well in the 96-well filter plate was 350 μL and this defined the maximum possible volume of
115
sample. A laboratory built aluminum plate with 96 rods tailor-made for the wells of the
116
MultiScreen plate, was used as cathode (96-electrode plate). A model ES 0300-0.45 (Delta
117
Elektronika BV, Zierikzee, The Netherlands) was used as power supply, and a Vibramax 100
118
Heidolph shaking board (Kellheim, Germany) was used to agitate the extraction system.
119
EME was performed according to the following procedure for each sample: a) 200 μL of 10 mM
120
NaOH was pipetted into the 96-well waste reservoir plate, b) 3 µL organic solvent was pipetted
121
on to the filter membrane of the 96-well filter plate, and c) 100 µL sample was pipetted into the
122
96-well filter plate. The 96-well waste reservoir plate and the 96-well filter plate were clamped,
123
and the 96-electrode plate was located on top. The rod electrodes of the electrode plate was then
124
in contact with the samples. The waste reservoir plate (anode) and the electrode plate (cathode)
125
were connected to the power supply. The extraction time varied from 1 to 10 minutes, the voltage
126
was between 1 and 100 V, and agitation was set to 900 rpm. For a couple of experiments, the
127
direction of the electrical field was reversed.
ACS Paragon Plus Environment
Page 4 of 19
Page 5 of 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
128
FITC clearance and protein retention
129
FITC clearance is the percentage of FITC removed from the sample after EME compared to the
130
non-extracted sample, as defined by equation (1). FITC clearance of 100% indicates complete
131
removal of FITC from the sample after EME. Protein retention is the percentage of protein left in
132
the sample after EME compared to the non-extracted sample, as defined by equation (2). Protein
133
retention of 100% indicates no protein loss after EME.
134
𝐹𝐼𝑇𝐶 𝑐𝑙𝑒𝑎𝑟𝑎𝑛𝑐𝑒 = 1 ― [𝐹𝐼𝑇𝐶 ]𝑠𝑎𝑚𝑝𝑙𝑒,𝑏𝑒𝑓𝑜𝑟𝑒 𝐸𝑀𝐸 ∙ 100%
135
𝑃𝑟𝑜𝑡𝑒𝑖𝑛 𝑟𝑒𝑡𝑒𝑛𝑡𝑖𝑜𝑛 =
136
FITC labeling protocol
137
The fluorescent labeling was performed on HSA and the reaction was done with a molar ratio of
138
10:1 FITC to HSA31. Calculation of the volume of 0.01 mg/mL FITC (VFITC(uL)) needed to obtain
139
molar ratio of 10:1, was done with the general formula:
140
𝑉𝐹𝐼𝑇𝐶(𝜇𝑙) = 𝑀𝑃𝑟𝑜𝑡𝑒𝑖𝑛(𝑔/𝑚𝑜𝑙) ∙ 1000 𝑚𝑔/𝑚𝑙 ∙ 𝐶𝐹𝐼𝑇𝐶 (𝑚𝑜𝑙/𝑙) ∙ 106 𝜇𝑙/𝑙
141
The HSA solution was prepared to a molar concentration of 1.50·10-5 M and set to a volume of 100
142
µL. The volume of 0.01 mg/mL FITC needed to obtain a molar ratio of 10:1 was 0.585 mL.
143
The labelling reaction was performed in a 1.5 mL low-bind Eppendorf tube (Eppendorf, Hamburg,
144
Germany). 100 µL of HSA solution was mixed with 585 µL of a freshly made 0.01 mg/mL FITC
145
solution (in reaction buffer) in the Eppendorf tube. The tube was wrapped in aluminium foil, for
146
light protection, and was agitated for 30 minutes at 900 rpm. After fluorescent labeling, 100 µL of
147
the solution was transferred to EME.
148
To avoid saturation of the fluorescence detector in the TDA and FIDA analysis, both the extracted
149
and un-extracted samples were diluted with 50 mM borate buffer pH 9.20 to a theoretical
150
concentration of 10 nM or 20 nM for the HSA/FITC –complex.
151
Flow injection analysis with UV-detection
152
In experiments reported in section 3.1 and 3.2, FITC and proteins were individually measured
153
using a 3000 Ultimate HPLC-UV (Thermo Fisher, Waltham, MA, USA) for automated flow injection
154
analysis with UV-detection. Thus, the HPLC column was replaced with PEEK tubing. Mobile phase
155
consisted of a 50 mM borate buffer pH 9.20 and the injection volume was 20 µL. The flow of mobile
156
phase was 0.1 mL/min. UV-detection was performed at 214 nm for Myoglobin and Cytochrome C,
(
[𝐹𝐼𝑇𝐶 ]𝑠𝑎𝑚𝑝𝑙𝑒, 𝑎𝑓𝑡𝑒𝑟 𝐸𝑀𝐸
(
[𝑃𝑟𝑜𝑡𝑒𝑖𝑛]𝑠𝑎𝑚𝑝𝑙𝑒,𝑎𝑓𝑡𝑒𝑟 𝐸𝑀𝐸
)
(1)
) ∙ 100%
(2)
[𝑃𝑟𝑜𝑡𝑒𝑖𝑛]𝑠𝑎𝑚𝑝𝑙𝑒,𝑏𝑒𝑓𝑜𝑟𝑒 𝐸𝑀𝐸
𝑉𝑃𝑟𝑜𝑡𝑒𝑖𝑛(𝑚𝑙) ∙ 𝐶𝑃𝑟𝑜𝑡𝑒𝑖𝑛(𝑚𝑔/𝑚𝑙)
𝑀𝑜𝑙𝑎𝑟 𝑟𝑎𝑡𝑖𝑜
ACS Paragon Plus Environment
(3)
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
157
216 nm for HSA, and 495 nm for FITC. The system was washed with a pure methanol solution
158
prior to each injection to avoid any carryover.
159
Taylor Dispersion Analysis conditions
160
The TDA experiments were conducted on a capillary electrophoresis instrument (Agilent 3D CE,
161
Agilent Technologies, Waldbronn, Germany) using light emitting diode fluorescence detection
162
(ZETALIF LED, Picometrics, Labège, France) with excitation wavelength 480 nm. A standard fused
163
silica capillary (inner diameter: 75 µm, outer diameter: 375 µm, length total: 100 cm, length to
164
detection window: 88 cm) was used. The capillary cassette was temperature-controlled to 25°C,
165
and the capillary inlet, outlet and sample vials were indirectly thermostated to 24°C via the lab
166
heating, ventilation and air conditioning system.
167
The capillary was rinsed and equilibrated prior to each sample analysis with 1 M NaOH at 1 bar
168
for 5 min followed by 50 mM Borate buffer pH 9.20 at 1 bar for 7 min. The HSA/FITC–complex
169
sample was injected at 50 mbar for 10 s (corresponding to 1 % of the capillary volume).
170
Subsequently, the injected sample was mobilized towards the detector with running buffer (50
171
mM borate buffer, pH 9.20) at 50 mbar for 20 min. All samples were analyzed in triplicate.Flow-
172
Induced Dispersion Analysis conditions
173
The FIDA experiments were conducted on a FIDAlyzer instrument (FIDA-Tech ApS, Copenhagen,
174
Denmark) using light emitting diode fluorescence detection (ZETALIF LED, Picometrics) with
175
excitation wavelength 480 nm. A fused silica capillary (inner diameter: 75 µm, outer diameter:
176
375 µm, length total: 100 cm, length to detection window: 90 cm) with an inner coating of
177
poly(ethylene glycol) was used. The capillary was temperature-controlled to 25°C inside the
178
FIDAlyzer instrument. Capillary inlet and samples were also temperature-controlled to 25°C.
179
The HSA/FITC–complex, purified via EME was diluted with 67 mM phosphate buffer pH 7.4 to a
180
fixed indicator concentration of 20 nM in varying analyte concentrations of anti-HSA antibody (0
181
– 5 µM). All samples were pre-incubated to reach equilibrium prior to analysis.
182
The capillary was flushed and equilibrated prior to each sample analysis with 67 mM phosphate
183
buffer at 1500 mbar for 5 min. The analyte sample (anti-HSA at varying concentrations) was
184
injected at 1500 mbar for 45 s, subsequently the indicator sample (HSA/FITC–complex, mixed
185
with anti-HSA at varying concentrations) was injected at 50 mbar for 10 s (39 nL, corresponding
186
to 1 % of the capillary volume). The injected indicator sample was then mobilized towards the
187
detection point with the analyte sample at 400 mbar for 180 s. All samples were analyzed in
188
triplicate.
189
ACS Paragon Plus Environment
Page 6 of 19
Page 7 of 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
190
TDA and FIDA data analysis
191
The Taylorgrams were processed using FIDA Data Analysis Software, version 01 (FIDA-Tech ApS,
192
Copenhagen, Denmark) in order to calculate the hydrodynamic radii of two species in one signal
193
with identical residence time, i.e. unconjugated FITC and HSA/FITC complex. Subsequently the
194
software was used for plotting the binding curve. The software utilized equations (S1), (S2), (S3)
195
and (S4) (Supporting Information)18.
ACS Paragon Plus Environment
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
196
Results and discussion
197 198
Concept and principle
199
The purpose of this fundamental research was to investigate the potential of EME for removal of
200
unconjugated fluorescent reagent after protein labeling. The workflow included (1) labeling of
201
protein with fluorescent reagent, (2) EME and (3) final measurement by TDA or flow-induced
202
dispersion analysis (FIDA). The EME device is presented in Figure S2, and consisted of (a) a 96-
203
well waste reservoir plate, (b) a 96-well filter plate, (c) a 96-electrode plate, (d) an agitator, and (e)
204
an external power supply. The three 96-plates (a,b,c) were sandwiched, and provided 96
205
individual extraction channels. This enabled simultaneous processing of up to 96 samples with
206
labeled protein (one sample per channel).
207
EME involved the following four steps: (1) loading the supported liquid membrane (SLM), (2)
208
sample loading, (3) waste solution loading, and (4) extraction. During SLM loading (1), 3 μL organic
209
solvent was pipetted on to a filter on the 96-well filter plate (b). This procedure was repeated for
210
the number of samples with labeled protein to be processed. The organic solvent rapidly diffused
211
into the pore volume of the filter and was immobilized by capillary forces, creating the SLM.
212
During sample loading (2), samples were pipetted above the loaded SLMs in the 96-well filter
213
plate (b). The samples contained labelled protein and fluorescent reagent. During waste solution
214
loading (3), a waste solution was pipetted into the 96-well waste reservoir (a) for each sample,
215
and the three plates (a,b,c) were clamped together and placed in the agitator (d). Finally, during
216
EME (4) the electrical field was applied by connecting the 96-well waste reservoir plate and the
217
96-electrode plate to the external power supply (5). The anode (positive potential) was located
218
in the waste solution while the cathode was in the samples.
219
FITC was selected as model fluorescent reagent. Three model proteins were selected to cover
220
different molecular sizes and isoelectric point, namely HSA (MW = 66.5 kDa, pI = 5.8)32, equine
221
heart myoglobin (MW = 17 kDa, pI = 6.7 (1) and 7.2(2))33 and bovine heart cytochrome C (MW =
222
12kDa, pI = 10.5)34.
223 224
Operational parameters for efficient FITC clearance and protein retention
225
Major operational EME parameters were considered to be (1) the chemical composition of the
226
SLM, (2) voltage, and (3) time. 1-octanol and NPOE are both commonly used as SLM solvents in
227
EME, mainly for the extraction of hydrophobic (log P > 1.5) acidic and basic drug compounds,
ACS Paragon Plus Environment
Page 8 of 19
Page 9 of 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
228
respectively8,35. In initial EME experiments, FITC clearance was investigated with 1-octanol and
229
NPOE. However, both 1-octanol and NPOE proved to be ineffective for extraction, and FITC
230
clearance was limited to 5-10 % and