Joseph Jordan, J. Keith Grime, and Dan H. Waugh Department of Chemistry Pennsylvania State University University Park, Pa. 16802

C. David Miller, Herbert M. Cullis, and D. Lohr American Instrument Co. 8030 Georgia Avenue Silver Spring, Md. 20910

Enthalpimetric Analysis I

I

Titrimetric Moae

Calorimetric

Thermometric Enthalpy Titrations

Direct injection Enthalpimetry

Mode

on fast processes, both thermometric and direct enthalpimetric procedures can be adapted t o study rate-controlled processes. When the latter type of conditions prevails, the designations kinetic titrimetry and kinetic DIE are appropriately used. The kinetic approach to enthalpimetric analysis, in general, can be referred to as thermokinetic analysis (TKA). In all the above instances the amount of heat evolved (or absorbed), Q, in the reaction may be calculated from Joule heating calibration experiments. The calibration circuitry is described later. Fundamental Principles T h e principles involved in enthalpimetric analysis are based on fundamental correlations inherent in adiabatic calorimetry:

Q = -n,-AH

Kinetic Titrimetry

(1)

where Q denotes the total amount of heat evolved (or absorbed), n,, the number of moles of product formed, and AH the molar enthalpy of reaction. Furthermore, the temperature change, A T , observed in an adiabatic cell can be explicated as in Equation 2:

Kinetic DIE

Figure 1. Relevant n o m e n c l a t u r e

A T = Q/K In view of the multiplicity of terms used to describe this field of analytical chemistry, a brief discussion of the relevant nomenclature is warranted. T h e terms used in this report are explicated in Figure 1. T h e two main types of enthalpimetric analysis, thermometric enthalpy titrations ( T E T ) and direct injection enthalpimetry (DIE), can be delineated as titrimetric and calorimetric methods, respectively. In practical terms they can be differentiated by the mode of introduction of the reactants into the adiabatic cell. T E T experiments are characterized by the continuous addition of the titrant to the titrate under effectively adiabatic conditions. The total amount of heat evolved (if the reac-

tion is exothermic) or absorbed (if the reaction is endothermic) is monitored autgmatically using the unbalance potential of a Wheatstone bridge circuit, incorporating a temperature sensitive semiconductor (thermistor) as one arm of the bridge (vide infra). The virtual cessation of heat evolved (or absorbed) is used to locate the equivalence point of the titration. DIE, on the other hand, involves the injection of a single “shot” of a reactant into the solution under investigation, the reagent being in stoichiometric excess. T h e corresponding caloric “pulse”, which is directly proportional to the number of moles of analyte reacted, is recorded using identical circuitry to that employed with T E T . Although initially the emphasis was

(2)

where K denotes the effective heat capacity of the system. Combination of Equations 1 and 2 yields

-AH-n,

AT=-

(3)

K

The heat capacity of the system, K , will remain essentially constant if the change in volume of the solution is minimized. Moreover, if it is assumed that A T is caused entirely by the reaction and that AH = AHo = constant, the observed temperature change, A T , is obviously proportional to the number of moles of product formed, np (and hence of reactant consumed):

AT = constanten,

(4)

A typical T E T enthalpogram (a thermometric titration curve) is

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

427A

shown in Figure 2a. Curvature in the region of the equivalence point may be caused by less than quantitative conversion of reactants to products (equilibrium curvature) or by the slow attainment of titration equilibria (kinetic curvature). In general, prohibitive limitations may arise if the relevant equilibrium constant is unfavorable. Furthermore, conventional thermometric enthalpy titrations are normally restricted to reactions whose effective rate is fast compared to the rate of addition of titrant. An end-point correction procedure for slower rates has been described. Appropriate computations can be used for the thermometric determination of first order rate constants up to a limit of 600,000 M-I s-l ( 1 ) . In a recent communication (2),it has been suggested that the hitherto universally accepted linear relationship between A T and the volume of added titrant, hu, is valid only over a narrow range of volume if one takes into account that some heat transfer actually occurs between the ambient medium and the reaction cell which is necessarily not quite 100% adiabatic. It is recommended that a more practical representation of a thermometric enthalpy titration would be achieved by a plot of llhT vs. l f u . A typical direct injection enthalpogram is shown in Figure 2b. T h e technique is based on the measurement of the total heat evolved (or absorbed), Q, rather than of an end point. The application of T E T and DIE enthalpograms to the determination of an analyte is contingent upon the knowledge of certain fundamental quantities-the stoichiometry of the titration reaction with the titrimetric technique and the heat of reaction with the calorimetric approach. Fundamental thermodynamic information (the heat, AH, the equilibrium constant, K , and hence the free energy, AG, and entropy, AS,of a reaction) can be deduced from thermometric titration curves ( 3 ) . Practical Aspects The methodology in its present form has been the subject of several reviews (3-10). The salient features of the experimental apparatus are a reagent addition system, an adiabatic cell incorporating a temperature sensor, a calibration heater and a stirring device, and a temperature-measuring circuit and a recorder (Figure 3). Reagent Addition. Titrimetric Procedures. Reproducible, constant titrant addition is generally achieved using either a motor-driven syringe or in some instances a pulse-free peristaltic pump. Uniformity of titrant and titrate temperatures is routinely obtained by passing the tubing conveying the titrant to the adiabatic 428A

* ANALYTICAL CHEMISTRY, VOL.

dT dV

d2T dV2

0

Volu

rant added

d 4 dV3

" Votume of titrant added

Volume of titrant added

Figure 2. Typical enthalpograms. TET, thermometric enthalpy titration; DIE, direct injection enthalpimetry

cell through a constant temperature bath. The disadvantages of liquid titrants can be circumvented by the injection of gaseous reagents into liquid titrates (6, 14). The heat capacity of a gas is negligible when compared to that of a comparable volume of liquid; therefore, the temperature of gaseous titrants need not be regulated. Furthermore, changes in the heat capacity of the adiabatic cell and its contents will be small, thus minimizing a nonlinearizing effect on the resultant enthalpogram. The idea may be further extended to the thermometric determination of gaseous samples with gaseous titrants (11).The titrants, which can be diluted if necessary with an inert gas, are standardized by injection into a liquid absorbate and subsequent titration with an appropriate reagent. The fundamental requirements for a gas-phase titration are identical to those for liquid-phase experiments-a favorable equilibrium constant and relatively fast kinetics. Calorimetric Methods. For conventional liquid-phase direct injection enthalpimetry, the emphasis is on quasiinstantaneous injection of the reagent. As the injected reagent is usually in 48, NO. 4, APRIL 1976

excess and the heat capacity of the system is conveniently determined after each injection, precise volume measurements are not a prerequisite of this approach. Accordingly, a manually operated syringe is adequate. For sequential analysis, in which a series of samples is injected into an excess of reagent, gravimetrically calibrated syringes with Chaney adaptors are recommended. The same thermostating considerations as in T E T experiments obviously apply. Ingenious immersion pipets (where the titrate solution itself serves as thermostat for the titrant) have been used to obviate any temperature mismatch between reactants. Such a system is indeed the basis of a versatile direct injection enthalpimeter for multicomponent analysis ( 4 ) . Adiabatic Cell. Successful enthalpimetric determinations are contingent upon the measurement of small temperature changes. Consequently, it is important that extraneous heat effects in the reaction cell be minimized. A conventional silvered Dewar flask has a relatively large heat transfer coefficient. However, special thinwalled Dewar flasks have been de-

offers powerful problem-solving capability for HP GC/MS systems

A New, mass spec BASIC controls the spectrometer, produces graphics on the display terminal or plotter, and accesses data files in disc storage. In addition, the HP 5934A may be used asa general purpose computer system using BASIC, ALGOL, or FORTRAN IV with a proven HP disc operating system. Mounted in a 1 1 0 c m (43.5-inch) high cabinet are all Hewlett-Packard instruments composing the system: 21 MXComputerwith 16,000oroptional32,OOO word memory; 7900A Dual Disc Drive for storing programs and spectral data; and mass spectrometer interfaces. System control is primarily through a graphic display terminal which provides high resolution alphanumeric and graphic displays, accepts all ASCII keyboard inputs, and also provides two graphic cursors that the operator uses to "point" to a CRT screen location. A plotter also may be used as an additional output device. For more information, contact the nearest HP office. Or write to Hewlett-Packard, 1601 California Avenue, Palo Alto, California 94304.

Features include: rn free 19,000-spectra library new powerful computer rn enhanced BASIC and systems software rn simplified, more convenient operation rn lower cost than previous system These and other HP 5934A Data System features offer even greater savings in time and more operating convenience for sample identification than the outstanding HP 5933A Data System it replaces. HP's powerful new 5934A Data System controls the scan operations of a Hewlett-Packard mass spectrometer, stores spectral data on discs, then displays and tabulates these data at operator command. Each disc can store over 12,000 mass spectra per GC run, typically 1 0 hours of uninterrupted analysis. The system is easily expandable to 2 5 5 interchangeable cataloged discs, providing over 5 billion bits of data storage.

HEWLETT -h PACKARD

ui

Sales and service from 172 offices

In

65 countries 23601

CIRCLE 96 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

429A

pressed in calories per chart ordinate unit, Co t h e calibration factor effective before t h e start of t h e titration (when the volume is U O ) , and Cfinalthe calibration factor measured after the titration (volume = ufinaJ. T h e fraction of total titrant added required to reach the equivalence point is denoted by a: a

E

uep - L'O ('final

Figure 3. Diagram of apparatus and circuitry used in enthalpimetric analysis

signed (22,22). incorporating a n appropriate reaction cell which has a heat transfer modulus of 11 X min-' and an equilibration time of only 3 s. This cell now forms the basis of a sophisticated isothermal titrationmicrocalorimeter ( 1 2 ) (v.i.). Analytical data can be obtained by using simple styrofoam insulated reaction cells which maintain pseudo-adiabatic conditions for the short period of a titration. Thi:, method is not recommended for fundamental thermochemical investigations. At the heart of enthalpimetric analysis lies the temperature measurement system. Thermometric titrations were performed a s early as 1913 utilizing a conventional mercury thermometer as temperaturf. sensor (8). Modern interest in enthalpimetric analysis was engendered bq thta simple. inexpensive, rapidly respondling, automated instrumentation made possible by t h e use of thermistors as temperature transducers. Thermistors are semiconductor devices which have a large temperature coefficient of resistivity, on the order of -4% p m degree. Even though the thermisl or resistance, RT, varies mponential'lj with temperature:

RT == A exp R / ( C + T )

(5)

(where A, R , and C are material constants characteristic of a given thermistor). RT approximates a linear function of temperature over a small range ( T 5 0 . 1 " C ) . The changing resistance is usually monii ored with the aid of a Wheatstone bridge circuit, whose unbalance potential is proportional to AT (and to Q if K is constant) and plotted automatically on the ordinate scale of a recording dc millivoltmeter. T h e corresponding chart abscissa is a measure of volume and/or time. T o 430 A

avoid localized temperature and concentration gradients in the adiabatic reaction cell, efficient stirring is imperative. Calibration Heater. Generally, the caloric axis (ordinate) of enthalpograms is calibrated chemically or electrically, though this is not strictly necessary for a purely end-point determinative titrimetric use. Chemical calibration involves reliance on a n accurately known heat of reaction and requires stringent reproducibility of reaction volumes in successive experiments. This limitation is circumvented by electrical calibration with occasional verification against suitable thermochemical standards. T h e circuitry for electrical calibration by Joule heating appears as part of Figure 3. The calorimeter is calibrated by passing a constant current, i, through a calibration heater, RH (which is immersed in t h e solution), and a standard resistor, R q . T h e heat dissipated within the calorimeter may then be calculated from Equation 6:

i2RHt VSVHt ( 6 ) 4.184 4.184 Rr where V s and V Hare potential drops measured across the standard resistor and the calibration heater. Changes in the heat capacity of the system occur as titrant is added. Accordingly, calibrations are generally performed both before and after the experiment. T h e desired caloric quantity is the total reaction heat evolved (or absorbed) u p to the equivalence point. A weighted average of t h e preand post-titration calibrations is used:

Q (calories) = --

Ce, =

CO+ a ( C f i n a 1 - C O )

where C,, is the desired calibration factor a t the equivalence point, ex-

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

(7)

- UO

In the calorimetric mode of enthalpimetric analysis, u,, = ufinal and only one calibration is required. Temperature Measuring Circuits-dc. Conceptually, the simplest method for continuous measurement of t h e smaH resistance change of t h e temperature sensing thermistor is its incorporation as one arm of a dc Wheatstone bridge (Figure 3). In this configuration the bridge unbalance potential, Vunba121nce,may be related by Equation 8 to the resistances of the three unchanging arms, R1, R?, R:i, to the variable resistance of the thermistor-containing arm, RT, and to the applied bridge potential, V a p p l i e d : Vunhalance = VapIilied

x

The variation of the unbalance potential is, for all purposes herein described, linear over a small temperature range (AT 5 0.1 "C). An unavoidable feature of such bridge circuits is t h a t a finite current must pass through the thermistor which is, as a result, self-heated. T h e sensed temperature is therefore not strictly t h a t of t h e solution in which the device is immersed but rather is the temperature of the thermistor itself, the difference between the two being dependent upon t h e rates of self-heating and of heat dissipation a t the stirred solution interface. (Stirring itself imparts heat to the solution.) T o minimize this temperature differential, the applied bridge potential is either pulsed (v.i.) or minimized. Methods for bridge potential optimization have been described ( 1 3 ) .In ultrahigh sensitivity experiments, alinear ramp generator circuit may also be necessary. In these situations the temperature increase due to stirring and thermistor heating may be prohibitively large, resulting in a steeply sloping baseline. This problem can be obviated by the introduction of an equal but opposite electrical signal ( 3 ) .Low bridge voltages require that the entire bridge circuit including the thermistor leads he carefully shielded to avoid noise pickup. T h e small unbalance potential is amplified by a low noise dc amplifier, the output of which is usually fed into a recording dc millivoltmeter.

FREE CATALOG PCR’S NEW GENERALCATALOG G-9 is yours for the asking. Several hundred organics, new from PCR,and a broader rangeof NMR reagents, MASS SPEC standards and many DERIVATIZING AGENTS will solve your analytical needs.

P.O. Box 1466/Gainesville, Florida 32602 (904) 376-7522 CIRCLE 177 ON READER SERVICE CARD

FEATURING PLENARY LECTURES AWARD SYMPOSIA: Ellis R. L i p p i n c o t t Medal, Anachem Award, BenedettiPichler Award, and seven student awards. TECHNICAL SESSIONS: I n c l u d i n g Coblentz Symposium, Fourier T r a n s f o r m Group, Raman Technical Group, Sources of Excita. t i o n f o r Spectroscopy, Photodiode Array Detectors, N e w Developments in Atomic Absorption and Fluorescence Spectroscopy, L i q u i d and Gas Chroma. tography, Analytical Chemistry Related to Clinical Problems, Trace Element Analysis I n c l u d i n g Standards, Polymer Analysis, Computers in Analytical Chemistry, X-ray Spectrochemical Analysis, New Applications f o r Magnetic Resonance, Spectroscopy in Space Exploration, m a n y others. EXHIBITS PLANNED SOCIAL A N D CULTURAL EVENTS Featuring t h e bicentennial a n d international scope of t h e meeting. Transportation available. CONTINUING EDUCATION: i n c l u d i n g FACSS workshops, ACS s h o r t courses, SAS s h o r t courses (Modern Emission Spectroscopy and Fourier T r a n s f o r m Spectroscopy: IR and NMR), Tech. nical Films. Company sponsored workshops and seminars.

.

FOR FURTHER INFORMATION CONTACT: Dr. Edward G. Brame, Jr., General C h a i r m a n DuPont Experimental Station Wilmington, Delaware 19898 USA (302) 772-3787 Dr. Edward C. Dunlop, P r o g r a m C h a i r m a n

Federation of Analytical Chemist and Spectroscopy Societies CIRCLE 78

ON

READER SERVICE CARD

CIRCLE

26 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

431 A

Figure 4. Schematic diagram of pulsed dc (partial duty) Wheatstone bridge circuit

Differential Titrations. There have been numerous modifications of t h e standard dc amplified bridge circuit. Differential titrations involve a simultaneous blank titration in an identical adiabatic cell, where titrant is added t o t h e pure solvent. The temperature difference between the actual titration a n d the blank is recorded on t h e ordinate of corresponding differential enthalpograms. In this way, extraneous heat effects du'e to undesired simultaneous or seque:ntial reactions, heats of dilution, and temperature mismatch between titrant and titrate may be eliminated (10 ). Differential titrations are implemented by the replacement of bridge resistor R1 with a thermistor possessing characteristics equivalent t o those of RT, and its placement in t h e blank titration cell. By an examination of Equation 8, equivalent temperature changes of t h e two thermistors will cause :no change in t h e bridge unbalance potential. Derivative Titrations. (Figure 2, curves c, e, g.) In a n attempt t o improve thermometric end-points, first derivative, second derivative, and even third derivative plots of t h e bridge output signals have been obtained through t h e use of differentiating circuits. By the use of such derivative technicluesi, end-points may yield well-defined nonmonotonic pulses amenable to trigger digital threshold readouts. Resolution Enhancement. The ultimate resolution of the temperature measurement circuitry is limited by omnipresent noise sources (Johnson and pickup) and dependent upon t h e temperature coefficient of the thermistor and its heat dissipation. Theoretical treatments of the ultimate resolution of thermistors and thermistor bridges have very recently become available (13).'The practical resolution is 10 microdegrees. Investigations have shown t h a t a barium titanate ceramic device, commonly called a positive temperature 432 A

coefficient ( P T C ) device, has a temperature coefficient on the order of +16% per degree (15).This inherent fourfold sensitivity enhancement over thermistor technology has proved useful in high-sensitivity thermochemical studies. The problem of thermistor selfheating has been attacked via the development (27) of a pulsed (partial duty) dc Wheatstone bridge (Figure 4).The applied bridge potential is greater than t h a t used in conventional dc bridge circuitry ( a gain in sensitivity), but t h e concomitant temperature differential between the thermistor and the measured solution is minimized by pulsation of the bridge power. In practice, the bridge is powered for 10%of the total duty cycle of 50 ms. T h e basic duty cycle of t h e bridge is set by a master timing gener-

- 1 -

ator, t h e output of which is a rectangular wave train. This train is used t o periodically energize a dc power supply which in turn powers the thermistor bridge. T h e power dissipated in the thermistor bridge is ccintrolled by the output of the timing generator. After amplification of t h e bridge unbalance potential, the still pulsatile train is fed t o a sample-and-hold circuit which is triggered by the master timing pulse. The triggering is adjusted so t h a t the sampling time occurs in t h e middle of t h e duty cycle of t h e power supply, thus eliminating transients which occur a t either end of t h e master pulse. Thermistor Bridges-ac. Several ac-powered bridge circuits have been described in t h e literature during t h e past several years (3,16, 17).This may be related to t h e ready availability of highly stable ac amplifiers and to t h e improvement of signal-to-noise ratios using phase sensitive detectors. A schematic diagram of a typical ac thermistor bridge is illustrated in Figure 5 . However, the effectiveness of the ac system is somewhat limited by the introduction of stray capacitance. In our opinion, the advent of stable d c amplifiers and the application of good shielding techniques to dc bridge circuitry minimize the advantages of ac circuits. Other Thermistor Circuits. Some novel bridge designs have been proposed but remain generally unaccepted. These include linearized Wheatstone bridges and circuits utilizing operational amplifiers ( 3 ) ,which exhibit improved linearity and increased sen-

+ .

~~

Phase Sensitive Detector

Filter

Figure 5. Schematic diagram of ac thermistor bridge

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4 , APRIL 1976

OUTPUT

CIRCLE 1 2 8

ON

READER SERVICE CARD

Marine Chemistry in the Coastal Environment k o w availabie-a comprehensive volume cclntainiig tne most recent advances in t i l s new and increasiigly important fI~lC

ACS Symposium Series No. 18 Thomas M Church. Ed/toi A special symp o s w m sponsored b y the Middle Atlantic Reglon of the American Chemical Sooety

T i e collection represenx ar- indispersatiie source of informatioi for every mariie sc ertist Empnasis is not rrerely or1 describing coastal proolems but on srowlnc the potential n applying *he to3Is of -nodern oceanography and ctiemistl-y to solve these problems

Forty-one chapters cover six malo, ateas physical. organic, a n d tracer marine chemistry, estuarine geochemistry, hydrocarbons a n d metals in the estuarme environment, ocean dlsposal forum, applications a n d resources In niarine chemistry, a n d organic a n d b l o l o g m l marine chemistry

Hellma-the largest assortment of highest precision glass and quartz cells. Standard Flow-through . Constant-temperature Anaerobic . Special Designs Also available-ULTRAVIOLET LIGHT SOURCES Deuterium L a m p s . Mercury Vapor Lamps Hollow Cathode L a m p s . Power Supplies

7'0 Dages (1975)Clothbourd S35 75 (ISBN 0-8412-0300-8) LC 75-28151

Write lor literature Box 544 Borough Hall Station Jamaica, NewYork 11424 Phone (212) 544-9534

Zider from.

Special Issues Sales American Chemical Society 1'155 Sixteenth St., N.W. Washington, D.C. 20036

I CIRCLE 94 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1 9 7 6

433 A

sitivity and ease of grounding. These advantages notwithstanding, the simplicity and general lack of noise of the standard dc bridge circuit have insured its continued use for most enthalpimetric instrumentation. Quartz Resonator Thermometer. Because of the scarcity of temperature transducers possessing a high sensitivity to very small temperature changes, thermistors have remained the mainstay of enthalpimetric temperaturemeasuring devices. Recently, the quartz resonator thermometer (QRT) has been proposed as a new instrument for enthalpimetric analysis ( 3 ) . This device consists of a small quartz crystal incorporated as part of an oscillator circuit. T h e resonant frequency of this circuit (about 28.2 MHz f 1 kHz per degree) as monitored by a frequency counter, is a linear function of the temperature of the quartz crystal. The degree of resolution is dependent upon the length of the counting period. With highly sophisticated frequency counters, a resolution of a few microdegrees may be obtained with a counting time of 10 s. The merits of this system are twofold: T h e measurement is highly insensitive to external noise sources, and the output may be directly calibrated in terms of a standard temperature scale. In contradistinction, thermistor bridge circuitry previously discussed is capable only of recording outputs proportional to temperature change. The use of the Q R T for enthalpimetric analysis is en cumbered by several inherent disadvantages. These include extended counting periods necessary for the attainment of adequate resolution, the large physical size of the probe coupled with its considerable thermal mass, and inadequate chemical resistance of the probe enclosure.

OUT'PUT

I

I

1

I

Figure 6. Schematic diagram of flow enthalpimetry

as a change in T p of the product stream. A process stream can be assayed by monitoring the heat pulse obtained when a sample of reagent is injected into the stream a t periodic intervals. The shape of the resultant enthalpogram is dictated by sample size. A typical continuous flow enthalpogram is shown in Figure 2d. One of the latest and potentially significant advances in the development of enthalpimetric flow systems is the immobilization of a large stoichiometric excess of reagent (in the form of a n enzyme) on the walls of the reaction chamber ( 1 8 ) .As the analyte flows through, it is catalytically converted to products with zero order ki-

netics prevailing. The corresponding heat change is monitored in the usual manner. The enzyme (reagent), acting as a true catalyst, remains unchanged, thus permitting sequential analyses. In a similar vein, an enzyme thermistor based on the measurement of heat generated in an enzyme-coated PVC coil in intimate contact with a thermistor has recently been described (19). Peak E n t h a l p i m e t r y . A new method called peak enthalpimetry has been reported (20) in the context of contemporary requirements for sequential analysis with particular reference to clinical analysis. A typical peak enthalpimetric apparatus is schematized in Figure 7. A carrier liquid (an inert

Modifications of Enthalpimetric Analysis

Flow Systems. Continuous Flow E n t h a l p i m e t r y . Continuous flow enthalpimetry has proved to be a valuable modification of enthalpimetric analysis, particularly when utilized for the on-line analysis of industrial process streams ( 1 0 ) .T h e technique, based on the fundamental principles inherent in flow calorimetry, consists of passing two reactant solutions a t a constant rate through a mixing chamber (the reagent being in stoichiometric excess) and continuously monitoring the heat output of the product stream (Figure 6). The temperature of the product stream, T p ,is directly related to the temperatures of the sample and reagent streams, to the molarity of the sample, and to the ratio of the sample and reagent flow rates. Accordingly, any fluctuation of the sample concentration will manifest itself 434A

Figure 7. Schematic diagram of peak enthalpimetry apparatus

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

Now your JX Baker distributor can also sfin YOU thincrs he doesn't have! I

U

I

How can your Baker distributor ship you things he doesn't have? Easy. He now can immediately contact the Baker Super Service Center andin almost every instance-have the desired item shipped directly to you within 24 hours. It's as though your J. 7: Baker distributor has just added the largest reagent warehouse in the world to his backyard to serve you. (To super-serve you.) Now depend on your nearest Baker distributor for all of your laboratory reagent needs. And do you have our new 428-page Catalog 750 featuring thousands of Baker quality laboratory chemicals? If not, please write. Thanks.

' 4 4 4

+

~

And, of course, J. T. Baker's Super Service is also available for.. .

NEW PRODUCTS FOR ATOMIC ABSORPTION1 FLAME EMISSION SPECTROPHOTOMETRY. The J T Baker line contains many products for atomic absorption/flame emission spectrophotometry including solvents, standards, solutions, etc And this line keeps expanding with the addition of many new highquality products like an ultrapure ULTREX" Methyl iso-Butyl Ketone, suitable for trace metal analysis

m 4+

JXBaker

~

~

~

~

~

+4

W@

J T Baker Chemical Co.,222 Red School Lane Phillipsburg N J 08865 201 859-5411

~

44

~

~

I

CIRCLE 2 7 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY,

VOL.

48, NO. 4, APRIL 1976

435A

~

solvent) and a reagent solution are made to flow into a mixing cell from constant pressure reservoirs. The two separate streams merge and are homogenized by the vibration (60 Hz) of glass beads contained within the mixing cell. A discrete volume of sample is injected into the carrier stream, producing a cylindrical plug of sample so-

lution bounded by the solvent. As the sample is swept toward the mixing chamber, diffusional spreading occurs, resulting in an essentially Gaussian distribution profile as it reaches the reaction cell. The pseudo-Gaussian temperature profile of the product stream, measured in a differential mode, reflects the concentration pro-

file of the analyte entering the reaction cell. Moreover, the output from the thermistor bridge is a measure of the instantaneous concentration of the sample. Thus, an integration of the area bounded by the peak will be a measure of the amount of reactable material in the sample (Figure 2f). With the incorporation of automatic

Tabte I. Selected Applications of Enthalpimetric Analysis Analyte(4

Type of determination

Glucose, ATP Sodium pyruvate, NADH Hexokinase, ATP

Enzyme assay Substrate determination using enzyme catalysis

PHARMACEUTICAL

Alkaloids Sulfonamides

INDUSTRIAL

Petrochemical

Polyesters, polyethers

Fer t i l izer s Ceramics Bayer Process control Nuclear technotogy Minerals Bauxites Coal

INORGANIC

Complexometric

Anions

0 R G A N IC FUNCTIONAL GROUP

436A

[El (25)

Urease Phosphotungstic acid Standard antigens and haptens

Kinetic DIE TET

(281

Codeine phosphate, morphine sulfate, etc. Su If at hiazo le, sulfisoxazole

Silicotungstic acid

TET

KOH in propan-2-01

TET with catalyzed end-point

Water, oxygen, and low molecular weight alcohols in hydrocarbons GI ycerole-al kylene oxide polyethers, butane-1 ,&dioladipic acid polyesters PO:-, K+, NH;, NO; car bam ide

Triethylalum inum

Continuous Flow Enthalpimetry

Acetic anhydride

DIE

(33)

Several sequentially added reagents H*O, H+

Dl E

(341

Free acid in nuclear fuels

NaOH

T ET

Si.O,, AI,O,, Fe,O,, TiO,, and others Phenol

Various

DIE

KOH in propan-2-01

TET with catalyzed end-point

(6)

M n ( l l ) in presence of Ca(ll) I-, NO;, SO:-,and others

EDTA

Differential TET

(381

Various

Peak enthalpimetry

(39)

NH,, aliphatic amines

HCI (9) KOH

TET DIE

Antibodies {IgG, IgM)

NO:- in glass Na OH /A I ,OJN a, CO 3 ,

TET

(16)

(29,301

DIE TET

(37)

Mntlll in oresence of

zn(tlj

'

GASEOUS

Ga s/ga s Gasp iq u id

DIE

DIE kinetic

Ref

Kinetic DIE

Proteins Immunological

Method

Reagentk)

BIOCHEMICAL

cow so,

For an exhaustive review, see Ref. 10

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

Ultra-pure hydrogen is often in demand in the laboratory environment, where it may be required for flame ionization and/or thermal conductivity studies for gas chromatography. Ultra-pure hydrogen is also required for semiconductor processing and many other production-line activities. When your work demands a reliable source of hydrogen with a known output purity, Laboratory Data Control offers two alternatives: Direct, electrolytic dissociation of water into hydrogen and oxygen has become recognized as a very safe Hydrogen Generators Utilizlng of Dissociation Of n20 ultra-pure hydrogen for c h romat og raphic and other scientific p urposes . To fill this laboratory need, LDC continues to offer the Mark IV Hydrogen Generator. With a rated output capacity of 150 ml/min at standard temperature and pressure, this precision instrument delivers hydrogen with less than 10 parts per billion impurities. Discharge pressure is operator-selectable from 0 to 60 PSIG, while low stored volume (less than 2cc) eliminates explosion danger from the lab. Mark IV Hydrogen Generators may also be duplexed to

CIRCLE 126

ON

READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4, APRIL 1976

4 3 7 ~

injection and digital printout, t h e seriatim performance of many analyses is feasible. Catalimetric Enthalpy Titrations. Standard enthalpimetric analysis has sensitivity limits dictated by t h e heat liberated (or absorbed) during a reaction. For titrates present a t low concentration and those with small heats of reaction, a catalimetric technique has been developed (4,101. This methodology takes advantage of a secondary indicator reaction which occurs only after virtual completion of t h e analytically interesting primary reaction: T h e sequence is controlled by selecting an indicator reaction which has an appropriate equilibrium constant. A judiciously chosen catalimetric reagent which reacts with excess titrants is added. If the heat of reaction between the catalimetric reagent and titrant is large, t h e heat of the analysis reaction becomes unimportant per se, but t h e correct equivalence point is indicated by t h e commencement of heat evolution (or absorption) engendered by the secondary reaction. In a recent application an alkali-catalyzed polymerization has been used as t h e indicator reaction in the determination of acidic functions of sulfanilamide derivatives (and of sulfonamide formulations) by use of catalimetric enthalpy titrations (21). Isothermal Titration Calorimetry. Evident fundamental assumptions underlying the concept of enthalpimetric analysis are t h a t the relevant system is effectively adiabatic and t h a t AH is independent of temperature. If an efficient adiabatic cell is used, these approximations hold for most analytical measurements using virtually instantaneous reactions. For precise fundamental measurements and for reactions with slower kinetics, isothermal calorimetry has been utilized. T h a t method is beyond the scope of t h e present review. However, a technique combining titration calorimetry and isothermal calorimetry has been reported (4, 12,221, a description of which is warranted. T h e method closely parallels t h a t of thermometric enthalpy titrations, with the exception t h a t the temperature of the adiabatic calorimeter and its contents is held constant throughout t h e titration. This is achieved via the incorporation of a thermoelectric cooling device and a variable heater (Figure 8). T h e heat of reaction can be correlated to the powcr required to maintain isothermal conditions. Both endothermic and exothermic reactions have been studied in this manner a t temperatures held constant to f0.0002 O C and heat effects measured with a precision of &O 02 calories. T h e isothermal method, albeit very precise, is time consuming. a restriction that may 438A

Figure 8. Isothermal titration calorimeter (from ref. 22) Courtesy of the American Institute of Physics

limit its applicability from an analytical viewpoint. Applications

A brief summary of some selected applications is shown in Table I. T h e survey outlines a few innovative and topical applications of enthalpimetric investigations taken from recent reports. This list should not be regarded as an exhaustive review. More complete surveys of enthalpimetric determinations can be found elsewhere (6, I O , 23, 2 4 ) . As the varied nature of the survey suggests. enthalpimetric analysis has found acceptance in many fields. T h e high tolerance level of the technique to nonreactable impurities has resulted in numerous industrial and commercial applications where matrix problems might otherwise be prohibitive. Indeed, enthalpimetric flow methods have been particularly successful in continuous quality control of industrial process streams. An enthalpimetric approach to clinical analysis has been recently developed ( 2 6 ) .Its salient feature is reliance on a fundamental thermodynamic property, A H , in lieu of empirical standards. Specificity has been achieved by the utilization of t h e inherent selectivity found in enzymecatalyzed reactions. Moreover, t h e possibility exists for serum and, in

ANALYTICAL CHEMISTRY, VOL. 48, NO. 4 , APRIL 1976

some cases, whole blood analysis without prior deproteinization (25). Clearly, the volume of reports in the literature suggests that there is growing interest in the utilization of enthalpimetric methods for both analytical and fundamental studies. T h e versatility and simplicity of the technique, combined with a classic foundation in fundamental thermochemistry, make enthalpimetric analysis an inherently attractive methodological approach. References ( 1 ) P. M’. Carr and J . 45, 634 (1973).

.Jordan. Ancil. Chem..

(2) G . 0. Piloyan and Yu Y.Dolinina. Tn/ a n t a . 21. 955 (1974) ( 3 ) P. If-. Carr, C‘rit. Rei. Anal. (’hcm., 3,

491 (1972). J.Barthel. “Thermometric Titrations”. LVilev, Kew York. Y.Y., 1975. (5) I,. S.Bark and 5.lf.Bark, ”Thermometric Titrimetry”. Pergamon Press. Oxford, England, 1969. (6)L. S.Bark, P. Bate, and J. I(.Gfime. “Thermometric and Enthalpimetric Titrimetry”, in “Selected Annual Reviews of the Analytical Sciences”,Vol 11. L. S. Bark, Ed.. Society for Analytical Chemistry, London, England, 1972. (7) J . J . Christensen and K. hi. Izatt. ”Thermochemistry in Inorganic Solution Chemistry”, in “Physical hlethods in Advanced Inorganic Chemistry“. p 548, H.A.O. Hill and P. Day. Eds.. Interscience, h-ew York, K.Y., 1968. (8) ,J. .Jordan. “Thermometric Enthalpy Titrations”, in ”Treatise in Analytical (4)

Chemistry”, I. M. Kolthoff and P. J. Elvings, Eds., Part I, Vol 8, p 5175, Interscience, New York, N.Y., 1968. (9) H.J.V. Tyrrell and A. E. Beezer, “Thermometric Titrimetry”, Chapman and Hall, London, England, 1968. (10) G. A. Vaughan, “Thermometric and Enthalpimetric Titrimetry”, Van Nostrand Reinhold, London, England, 1973. (11) B. J. Duffield and D. N. Hume, Anal. Lett., 7,681 (1974). (12) J. J. Christensen, J. W. Gardner, D. J. Eatough, R. M. Izatt, P. J. Watts, and R. M. Hart, Reu. Sci. Instrum., 44, 481 (1973j . (13) L. D. Bowers and P. W. Carr, Thermochim. Acta, 10,129 (1974); 11,225 (1975). (14) P. G. Zambouin and J. Jordan, Anal. Chem., 41,437 (1969). (15) J. A. Lynch and J. Jordan, Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, Paper No. 54, Cleveland, Ohio, March 1974. (16) E. B. Smith and P. W. Carr, Anal Chem., 45, 1688 (1973). (17) R. L. Berger, W. S. Friauf, and H. E. Cascio, Clin. Chem., 20, 1009 (1974). (18) L. M. Canning, Jr., and P. W. Carr, Anal. Lett., 8,359 (1975). (19) K. Mosbach and B. Danielsson, Biochim. Biophys. Acta, 364,140 (1974). (20) A. C. Censullo, J. A. Lynch, D. H. Waugh, and J. Jordan, “Biochemical and Clinical Applications of Titration Calorimetry and Enthalpimetric Analysis”, in “Analytical Calorimetry”, R. S. Port and J. F. Johnson, Eds., Plenum Press, New York, N.Y., 1974. (21) E. J. Greenhow and L. E. Spencer, Anhl. Chem., 47, 1384 (1975). (22) J. J. Christensen, H. D. Johnston, and R. M. Izatt, Reu. Sei. Instrum., 39,1356 (1968). (23) L. D. Hansen, R. M. Izatt, and J. J. Christensen, “Applications of Thermometric Titrimetry to Analytical Chemistry” Treatise on Titrimetry”, Vol 11, J. Jordan, Ed., Marcel Dekker, New York, N.Y., 1974. (24) R. N. Heistand, Reu. Anal. Chem., 2, 193 (1975). (25) C. D. McGlothlin‘and J. Jordan, Anal. Chem., 47,786, 1479 (1975);Clin. Chem., 21, 741 (1975). (26) L. Donnovan, K. Barclay, K. Otto, and N. Jespersen, Therrnochim. Acta, 11,151 (1975). (27) D. Lohr, unpublished. (28) N. D. Jespersen, J . Am. Chem. S o c . , 97, 1662 (1975). (291 J. Jordan and N. D. Jespersen, Coll. Int. Cent. Nat. Rech. Sei., 201, 59 (1972). (30) J. A. Lynch, thesis, Pennsylvania State University, University Park, Pa., 1976. (31) L. S.Bark and J. K. Grime, Anal C h m Acta, 64,276 (1973). (32) E. J. Greenhow and L. E. Spencer, Anal. Chem., 47,1384 (1975). ( 3 3 ) I. I. Kaduji and J. H. Rees, Analyst (London),99,435 (1974). (34) I. Saj6 and B. Sipos, Talanta, 19,669 (1973). (35) K. Doering, Therrnochim. Acta, 8,485 (1974). (36) F. G. Bodewig,Anal. Chem., 46,454 (1974). ( 3 7 ) M. Mike, Hung. Sci. Instrum., 28,47 (1974). ( 3 8 ) Doi Kunio, Anal. Chim. Acta, 74,357 (1975). (39) R. Geyer, J. Birkahn, and G. Eppert, Chem. Technol., 12,775 (1974).

With Our NEW

SCAN R

you can turn this

into this

Our N E W Model 4101 Scan Recorder is a totally n e w k i n d of instrument. With your P O L A R O G K A P H for example, the 41 01 can baseline compensate your data and remove interfering peaks. With a S P E C T R O P H O l O h l E T E K , our 4101 can give you double-beam performance from a single-beam instrument. The 4101 removes the e f f e c t of light source/mono-

With its base price of $2,750*, shouldn’t you find out more about our N E W Model 4101? Write or call today:

chromator spectral

and

corrects your spectra.

m

Princeton Rppliid Resarch Corporation

P. 0. Box 2565,

Princeton, New Jersey 0 8 5 4 0 Phone: 609/452-21 11 Telex: 84.3409

In Europe-contact

Princeton Applied Research GmbH, 0 8 0 3 4 Unterpfaffenhofen, Waldstrasse 2, West Germany.

*Options are extra.

336

Circle 173 for additional information

6s

Now.. iable fluorescence detectbr for liquid chromatography

excitation UVIVIS monochromator and readily adjustable emission wavelength i n conjunction w i t h a very accessible 5 uI flow t h r o u g h cuvette makes detection at picogram levels possible. The 970 Spectrofluro m o n i t o r can be attached and operated in conjunction w i t h existing HPLC systems. W r i t e today for complete technical literature.

Research a t the Pennsylvania S t a t e University ( b y .J.J.. J.K.C.. and D.H.\V.) was supported by Grant GP38478X from the Karional Science Foundation.