Interferometry in the seventies - Analytical Chemistry (ACS Publications)

Jun 1, 1974 - Peter R. Griffiths. Anal. Chem. , 1974, 46 (7), pp 645A–654a. DOI: 10.1021/ac60343a027. Publication Date: June 1974. ACS Legacy Archiv...
0 downloads 0 Views 2MB Size
Interferometry in the Seventies Peter R . Griffiths Chemistry Department Ohio U n i v e r s i t y A t h e n s , Ohio 4 5 7 0 1

In a recent editorial in this journal, Dr. Laitinen ( I ) cited infrared spectroscopy to illustrate the seven stages in the life of an analytical instrument. The seventh stage was defined as a “period of senescence” in which “other methods of greater speed, economy, convenience, sensitivity, selectivity, etc., surpass the method under consideration.” In the case of infrared spectroscopy, an eighth stage has emerged, which might perhaps be defined as the rejuvenation stage. While it is certainly true that newer instrumentation is surpassing the sensitivity and selectivity of conventional infrared spectrophotometers for many applications, the development of mid infrared Fourier transform spectroscopy has enabled measurements to be made a t far greater speed and/or sensitivity than is possible with a grating monochromator. In one of the first articles written for this feature some five years ago, Low (2) gave a brief description of the theory of infrared Fourier transform spectroscopy (FTS) and discussed the instrumentation then available together with several applications which had been carried out with FTS. At the end of this paper, Low summarized developments in the field of FTS which were taking place a t that time. Certainly the most im-

portant of these innovations was the development of a computerized highresolution Fourier transform spectrometer by a group a t Block Engineering in Cambridge, Mass. Shortly after the publication of Low’s article, this spectrometer became commercially available, and this and other similar Fourier transform spectrometers are now being used for a multitude of different applications in many laboratories across the world. This article will discuss the reasons behind the development of the Michelson interferometer for mid infrared spectroscopy and describe a few of the more interesting applications for which they are being used. Some knowledge of the theory of infrared FTS will be assumed, and several descriptions of this theory have been published in this and other journals.

Early Interferometers The first two interferometers used for mid infrared FTS were both designed for astronomical spectroscopy even though they were conceptually very different. The first type was developed by a French group led by Pierre and Janine Connes ( 3 )for the purpose of extremely high-resolution measurements of weak sources. This interferometer was rather sophisticated and (along with its successors) has

been used to obtain some of the highest resolution infrared spectra ever measured on any kind of infrared spectrometer ( 4 ) .This type of instrument is quite specialized in nature and is overdesigned for normal chemical spectroscopy in which a resolution higher than 0.1 cm- is rarely called for. The second type of interferometer was developed by Mertz ( 5 ) ,and has a much simpler design than Connes’ interferometer. Its most important feature is its high scan speed; the moving mirror of Mertz’ interferometer is translated a t a sufficiently high velocity that mid infrared radiation is modulated a t audio frequencies. The theory of this and other rapid-scanning interferometers has been described in several articles (2, 6, T ) , the most recent of which is by Griffiths et al. ( 7 ) .This interferometer yielded spectra of rather low resolution (20 cm- I), so that it did not meet with the general acceptance of the chemical community. even though many spectra were measured which had hitherto been difficult to measure with a conventional grating monochromator. The signal-to-noise ratio ( S / N ) of interferograms measured with a rapidscanning interferometer can be increased by coherently adding interfer-

A N A L Y T I C A L C H E M I S T R Y , VOL. 46, N O . 7, J U N E 1974

645A

Michelson interferometer Courtesy of Division of Physical Sciences, Smithsonian Institution

ogram data points digitized a t equal intervals of optical path difference. However, the constant frequency clock which was used on the Mertz interferometer meant that the interferogram was digitized a t equal intervals of time; only if the moving mirror of the interferometer was translated a t an absolutely constant velocity would the interferogram be digitized a t equal increments of optical path difference. The drive system of these early interferometers was based on a relatively simple transducer (similar to those used on loudspeakers) with no means of accurately controlling the mirror velocity, so that coherent addition could only be maintained over rather short optical path differences. For higher resolution spectroscopy, where much larger optical path differences are required, the scan speed produced by this type of drive is not sufficiently constant to permit true signal averaging, where the S / N increases with the square root of the number of scans. Fortunately, several parallel developments in fields unrelated to FTS allowed the successful development of relatively inexpensive Michelson interferometers which would yield spectra of a sufficiently high resolution to satisfy all but the most demanding of chemical spectroscopists.

Innovations in Interferometry Two problems had to be solved to allow high-resolution data to be measured. The first involved designing a drive of such precision that the moving mirror remains in precisely the same plane throughout the entire scan, with a tilt of less than half of the shortest wavelength in the spectrum being measured. This goal was initially achieved by the use of air bearing drives, although oil bearings have since been used in certain in646A

struments. The latest commercially available high-resolution interferometer (made by EOCOM Corp., 19722 Jamboree Road, Irvine, Calif. 92664) achieves an exceptionally tilt-free drive by the use of two symmetrical air bearings, whose center of resistance is the point a t which the drive force is applied, thus providing a dynamic mechanical drive stability. The optical throughput of an interferometer is given by the product of the solid angle of the beam and its cross-sectional area a t a focus. The solid angle of the beam passing through the interferometer is determined solely by the resolution and the highest frequency in the spectrum, so that for any measurement in which this solid angle has been optimized, the optical throughput is determined by the area of the mirrors in the interferometer; the larger the area of the mirrors, the higher is the optical throughput, and hence the higher is the S / N of the spectrum which can be measured. In practice it is more difficult to drive large mirrors with little or no tilt than small ones. Thus, until recently the mirrors on interferometers designed to measure highresolution (0.1 cm-') spectra have been small, generally of 1-in. diameter. The drive system developed by EOCOM has enabled 0.1 cm- resolution spectra to be measured with interferometer mirrors up to 3 in. in diameter, so that the optical throughput is almost 10 times greater than that of earlier interferometers. The second problem which had to be overcome was the development of a method of digitizing the interferogram a t equal intervals of mirror trauel rather than equal intervals of time. In some early Michelson interferometers designed for far infrared spectroscopy, a MoirB fringe device had been successfully used for this

A N A L Y T I C A L C H E M I S T R Y , VOL. 46,

NO. 7,

J U N E 1974

purpose, but this method was not found to be applicable to mid infrared spectroscopy by use of rapidscan interferometers. It was actually the recent availability of inexpensive helium-neon lasers which allowed the problem to be overcome. When monochromatic radiation, such as that from a He-Ne laser, is passed through a scanning Michelson interferometer, the beam is modulated sinusoidally, owing to successive conditions of constructive and destructive interference which occur a t wavelength intervals of optical path difference. By monitoring a second (reference) interferogram from a He-Ne laser a t the same time as the main (signal) interferogram is being measured, the signal interferogram can be digitized a t equal intervals each time that the reference interferogram reaches a certain value. Since the reference interferogram is usually ac coupled, the signal interferogram is most easily digitized a t every zero crossing of the reference interferogram. Although this technique allows the signal interferogram to be digitized a t equal intervals of mirror travel, it does not ensure that the first data point is collected a t the same mirror position for every scan during the signal averaging process. To overcome this problem, a beam of white visible light is also passed through the interferometer, which gives rise to an interferogram similar in form to the signal interferogram from a broadband infrared source, but modulated over a shorter optical path difference because of the shorter wavelength of visible light compared with infrared radiation. If data collection is initiated when the signal from this white light interferogram reaches a certain value, vi, coherent addition of the signal interferogram can indeed be

Model 496 interferometer Courtesy of Digilab Inc Cambridge Mass

achieved. Thus, in a modern Fourier transform spectrometer, three interferograms are actually measured: one from the infrared signal being studied, one from the monochromatic laser beam, and one from the white light reference beam. In some instruments, the same beamsplitter is used for all three beams, with small areas at the top and bottom of the signal beamsplitter being coated with different materials to allow the laser and white light beams to be modulated and reach their respective detectors (Figure 1). In others, a separate reference interferometer is used, the moving mirror of which is attached to (and therefore driven a t the same speed as) the moving mirror of the signal interferometer (Figure 2); Figure 3 shows the three interferograms measured with this type of interferometer. That the position of the peak of the white light interferogram can be displaced from that of the signal interferogram means that data points can be collected on both sides of zero path difference of the signal interferogram, leading to a somewhat more simple phase correction routine in the computation of the spectrum. For interferograms measured with the type of interferometer shown in Figure 1, the peaks of the white light and signal interferograms both occur simultaneously.

Detectors for FTS One further factor was critical to the successful measurement of mid infrared interferograms by use of these interferometers. For incandescent sources of the tvpe normall! used in conventional mid infrared spectrophotometer ,. the S/N of the interferograms can be very high. and if the speed of the moving mirror is slow, the S / N may be greater than the dynamic range of the analog-to-

digital convertor (ADC). Reduction of the S / N of the interferogram to the level where the noise can be sampled by the ADC may be achieved by increasing the speed of the moving mirror. However, if commonly used thermal detectors are used to measure these interferograms, the highest modulation frequencies in the signal are so high that the response time of the detector is too long to permit them to be measured. Only the development of sensitive pyroelectric bolometers with very short response times, and in particular the triglycine sulfate (TGS) detector, has permitted wide range mid infrared spectra to be easily measured. The TGS detector has approximately the same detectivity, D*, as the various thermal detectors commonly used in grating spectrophotometers. However, TGS does have several undesirable properties, in that it cannot be heated above its Curie temperature (40”C),and it reacts with atmospheric water vapor so that it must always be sealed behind an infrared transmitting window. Since no readily obtainable material transmits both mid infrared and far infrared radiation, different detectors have to be used for mid and far infrared spectroscopy. Other pyroelectric materials, in particular strontium barium niobate, have been developed which overcome these problems, but as yet their sensitivity is so much lower than TGS that they have not been used for spectroscopic measurements. Sensitive photoconductive detectors have been applied to FTS, but they must be used under conditions where the S/Nof the corresponding interferogram measured with a TGS detector is very low, so that the dynamic range problems discussed earlier in this section are avoided. Al-

though the useful frequency range of photodetectors is always less than that of pyroelectric detectors, a combination of indium antimonide and mercury cadmium telluride detectors (both of which operate at the temperature of liquid nitrogen) has been used to cover the mid infrared spectrum above 800 cm- a t a sensitivity which is a t least 10 times greater than is able to be attained with a TGS detector.

Data Systems In addition to its optics. a modern Fourier transform spectrometer requires a data system for signal averaging, performing the fast Fourier transform ( F I T ) and all subsequent data processing. Once again relatively recent technical developments, this time in the field of minicomputers, have given an extremely powerful and versatile data handling capability to Fourier transform spectrometers. By adding a disc memory to a minicomputer with 4K words of core, interferograms with 32K data points have been signal averaged and transformed in just 3 min to yield complete mid infrared spectra a t better than 0.5 cm- 1 resolution. For lower resolution spectra, comparable in resolution to those measured on a small grating spectrophotometer, the computation time is only a few seconds. All data systems designed for use with rapid-scanning interferometers possess the capability of signal averaging interferograms and performing the FFT; however, the amount of data manipulation which can be carried out on digital spectra already stored in the data system of the spectrometer certainly adds to the appeal of FTS to spectroscopists. By far the most versatile and popular data system for infrared FTS has been developed by Digilab, Inc. (237 Putnam

A N A L Y T I C A L C H E M I S T R Y , VOL. 4 6 , N O . 7 , JUNE 1 9 7 4

647A

I

He-Ne LASER

I

*

MOVING MlRROR

INFRARE SOURCE

WHITE LIGHT1

FOR WHITE LIGHT

I I

Ge PHOTODIODE (RED LIGHT DETECTOR)

I

.t

Figure 1. Schematic representation of interferometer which has beamsplitter for signal, laser reference, and white light interferograms in same plane

INFRARED DETECTOR

FIXED MIRROR

\

**

\

\

INFRARED SOURCE

\

REFERENCE INTERFEROMETER

-1

. L +.-._.___._ I

I

*

WHITE LIGHT SOURCE

I .+.+ J

Figure 2. Schematic representation of interferometer whose reference interferometers are not in same plane as signal interferometer

INFRARED DETECTOR

i

- -b -

He-NeLASER Ge PHOTODIODE

0

Si PHOTODIODE

Figure 3. Typical interferograms from ( A ) signal interferometer, ( B ) white light interferometer, and (C) laser reference interferometer Zero path differences for signal and white light interferograms can occur at different time for interferometer of type shown In Figure 2 . but they must occur simultaneously for interferometer of type shown in Figure 1

648A

A N A L Y T I C A L C H E M I S T R Y , V O L . 46, N O . 7, J U N E 1974

Frankly, we were amazed that the new Model 55 detector for Perkin-Elmer’sModel 1220 Liquid Chromatograph made sucha hit at the Cleveland meeting...

After all, the Model 55 wasn’t the f irst vari a ble-waveleng t h UV-Visi ble detector t o be offered for l i q u i d chromatography, nor yet the second. It was the fifth, to be exact. Some had been on the market for over one year.

And that turned out to be precisely the point. We demonstrated the 55 in operation with actual solvents, gradients, and samples. The most enthusiastic watchers, the ones who kept coming back bringing different friends, were those who had already used one of the four others. Those were the people who could really appreciate the difference. “Make it do that again,” they requested, and we ran the trace in Figure 1, which happens to be a bit of baseline at a wavelength of 210 nanometers, solvent water, and 0.02 absorbance units full scale, “But it gets Fig. 1. noisier, does i t not,” t h e y .p--.-..+a’ -, I guessed, 0.002A “when the solvent actually I flows?” “The solvent is flowing now at 2 ml/minute,” we said. “The design of the transfer optics to the-8 flow-through cell is so ingenious that the system doesn’t see flow noise at all. Nor does it see changes in the refractive index of the solvent.”

-

-L-l

t

I ‘‘Can7you do gradients with this thing?” they asked. “Gradients are the name of the game,” we said, with accuracy if not originality. To show what we meant, we proceeded to draw a pretty poor one, as in Figure 2. It shows a 12-minute separation of antioxidants on a Sil-X-I column, with the solvent changing from hexane to chloroform, measured at 254 nm. At 254 nm, chloroform absorbs more than hexane, so the baseline shifts. Not only is the result less than beautiful, but a couT. ple of small bands, w h i c h should be there, are lost. To improve things, the first thing you do is seek a wavelength where your samples absorb and your solvents do not. That’s often pretty easy to do, and any variable-wavelength detector can do it. For this example, we chose 280 nm. However, t h e t w o solvents have different refractive indexes. With detectors less cunningly designed than the Perkin-Eimer Model 55, that would still be enough to upset your baseline. With the 55, though, we got a picture-book run (Figure 3), with the baseline pasted dead flat to the horizontal, and the two little lost peaks CIRCLE 184, ON READER SERVICE CARD

reappearing so bravely as to bring a lump to the throat. “Amazing,” said some people; those were the people who had the experience to know how bad things could have been. “How much is this magic box?” “A mere $3920, in the US.,” we said. “Can it be used with other instruments besides the PerkinElmer Model 1220 Liquid Chromatograph?” we were asked. “Yes,” we answered with a smile. “It can also be used with Perkin-Elmer’s low-cost Models 1240 and 1250 L i q u i d C h r o m a t o graphs.” “Very interesting.” some questioners persisted. “What we meant is, can the Model 55 be used to dress up the performance of presently existing chromatographs made by other manufacturers?” There was a long pause. “Yes,” we finally said in a barely audible voice. Instrument Division, Perkin-Elmer Corporation, Main Avenue, Norwalk, Conn. 06856.

PERKIN-ELMER

Ave., Cambridge, Mass. 02139). Some features on this system are relatively routine, such as the capability of plotting spectra linear in absorbance or replotting them under ordinate or abscissa scale expansion; other routines are of a more specialized nature. The following paragraphs describe a few applications for which this data system has been used where the data handling capability is critical to the success of the experiment. One routine which has often been found to be extremely useful is the capability of storing a single-beam spectrum for later use as the reference for a ratio-recorded spectrum. With this feature, strong bands in the reference spectrum which would be difficult to compensate with a double-beam spectrometer can often be completely eliminated from the spectrum. For example, when studies of dilute solutions are made with longpath cells, it is convenient to use the same cell to hold the sample solution that is used to hold the reference solvent so that exactly the same path length is involved for each measurement. In another application, when spectra of materials under high pressure are measured by a cell with diamond windows, the absorption bands of diamond are always seen in spectra measured on a double-beam spectrophotometer. However, if the spectrum of the empty cell is measured and stored in the computer memory before the sample is prepared, the diamond bands can be eliminated from the transmittance spectrum of the sample. Sometimes it is not possible to compensate for absorption bands found in the reference spectrum in this manner. In cases of this type, programs have been written to allow these bands to be compensated digitally. The spectrum of each material whose absorption bands are to be compensated is measured and stored linear in absorbance, as is the spectrum of the sample. The absorbance spectrum of each material to be compensated is multiplied by a constant (greater or less than unity) chosen so that the bands are exactly compensated when the reference spectrum is subtracted from the sample spectrum. The use of this program can be illustrated by showing how bands owing to atmospheric H20 and COz can be compensated in long-path spectroscopic measurements of trace organic air pollutants before and after these programs were written. Measurements of this type made by FTS have been described by Hanst et al. (81, who employed a 400-m foldedpath gas cell to hold the samples of contaminated and clean air. In his first study, Hanst measured the sam650A

ple spectrum first and then had to carefully adjust the partial pressure of H2O and C02 in clean air so t h a t their bands were precisely compensated. Finding the correct partial pressure of H2O and C02 to exactly compensate the bands in the sample often took a lot of time and effort; however, with the programs described previously, the same effect can be achieved much more quickly. The linear absorbance spectrum of the a t mospheric species is measured (by taking the ratio of the spectrum of the cell filled with clean air and that of the evacuated cell) and stored. The linear absorbance spectrum of the contaminated sample is then measured in the same manner. The first spectrum may be multiplied by a constant chosen so that when the scaled reference spectrum is subtracted from the sample spectrum, all trace of the bands owing to H20 and COz is removed, leaving only the bands owing to the contaminants. Many other applications of the scaled subtraction routines could also be listed. This type of program is not only useful for linear absorbance spectra, but it can also be used for infrared emission spectroscopy. The background radiation often has to be subtracted out from an emission spectrum to observe or quantify the spectral features of interest. The background spectrum can usually be measured before or after the measurement of the source, but it is quite rare that this spectrum does not have to be scaled up or down before subtraction so that background peaks are precisely compensated. One of the best examples of the use of a data system for infrared FTS is seen in its application to the on-line identification of materials separated by gas chromatography (GC-TR) (7, 9). Effluent gas from the chromatograph is passed via a heated transfer line to a heated light-pipe gas cell. The signal from the GC detector is monitored, and when the monitor indicates that a peak is passing through the light-pipe, interferograms are signal averaged until it is through. At this point the interferogram is stored in the disc memory of the data system. Interferograms of each GC peak are measured in this way and stored in successive memory arrays; a t the end of the chromatogram, they are sequentially recalled and each spectrum is computed and plotted. If just one peak in the chromatogram is of interest, it can be automatically trapped in the light-pipe when the sample is a t its maximum concentration so that more extensive signal averaging can be performed to increase the S/Nabove that of the “on-thefly” method. Programs have also been written for stopping the flow of car-

A N A L Y T I C A L C H E M I S T R Y , V O L . 46, NO. 7, J U N E 1974

Table I. High-Resolution FTS Instrumentation Manufacturer

Model no.

Max resolution, cm - 1

Digilab Digilab

496 296 7201

0.1 0.5 0.06

1 2

0.5

1

EOCOM

Barnes Engineering 1 F 3

Mirror diam, in.

Up t o 3

rier gas through the GC column while the sample is trapped in the lightpipe (interrupted elution) so that no peaks are missed during signal averaging of the trapped sample.

Instrumentation There are now several commercial sources of Fourier transform spectrometers for mid infrared spectroscopy, and it is believed that this list will be increased in the near future. Manufacturers making interferometers capable of measuring high-resolution spectra (0.5 cm- or better) are listed in Table I. Of these companies, Digilab is the only one to supply a data system a t the present time, although it is understood that EOCOM will shortly be able to supply one also. The scan speed of all of these instruments is such that it usually takes less than 1 sec to measure a low-resolution (8 cm- spectrum and less than 10 sec to measure a high-resolution (0.5 cm- spectrum. Two companies, Digilab and Barnes Engineering (30 Commerce Road, Stamford, Conn. 06902), have modified versions of their instruments so that very rapid scan rates (up to 100 scans/sec) can be achieved for low-resolution spectroscopy. Another manufacturer. Willey Corp. (P.O. Box 670, Melbourne, Fla. 32901), has developed a lower resolution slow scanning interferometer for the purpose of measuring the total and diffuse reflectance spectra of samples. This instrument uses a rather different method of measuring the interferograms, in which interferogram data points of light reflected from the sample, from a 100%reflector and from a 0% reflector (blackbody), are collected during a single scan (Figure 4). The sample is held at the surface of an integrating sphere, and the 100% reflector and the detector are held a t other positions on the surface. For measurements made by use of an integrating sphere, a large detector gives the best results since the signal is effectively averaged over the entire surface, and Willey (10) has found that spectra of the highest S/N are measured with a

Dual wavelength single- and dual-channel capability

Altex means UV detection for liquid chromatography If you believe you must compromise when choosing a UV detector for liquid chromatography, then let us demonstrate the new Altex Model 151 or 152. These instruments have the sensitivity, features and cell geometry to fit your applications.

You get dual wavelength capability at 254 nm and 280 nm. Both wavelengths are present at all times in the flow cell and the effluent can be monitored at each wavelength individually or simultaneously. Both detectors feature 0.005 O.D. sensitivity and double beam optics. Analytical, preparative and biochemical flow cells are interchangeable for maximum flexibility. In addition, cell inlets and outlets are made with quick-connect zero dead volume Altex fittings to minimize leaks and bubble formation, We'll make it easy for you to compare these instruments with any UV detectors on the market Just contact us for a demonstration in your lab.

CIRCLE 4 ON READER SERVICE CARD A N A L Y T I C A L CHEMISTRY, V O L . 46, NO. 7, JUNE 1974

651 A

REFERENCE

Figure 4. Optical diagram for Wiiley Model 318 total reflectance spectrometer ( A ) Path for laser reference beam: since this is slow-scanning interferometer, no white light reference is needed. ( B ) Path for sample interferogram. (C) How radiation reaches 100% reflective reference. Between open (sample) auadrant of chopper and reflective (reference) quadrant are two black (0% reflective) quadrants

large (5 x 5 mm) mercury cadmium telluride detector. The latest spectra measured with this instrument show a great improvement over earlier spectra measured with a small TGS detector. This instrument yields spectra of high photometric accuracy, even when highly scattering samples are being measured. Applications of FTS In a recent article by Jakobsen and Griffiths ( I I ) , the theoretical advantages of FTS over grating spectroscopy have been quantitatively appraised. The main conclusion reached by these authors was that the principal benefit of FTS for chemical spectroscopy is derived through Fellgett’s (multiplex) advantage, whereas Jac652A

quinot’s (throughput) advantage is almost negligible. Quantitatively, Fellgett’s advantage is that the S / N of a spectrum measured by FTS is a t i m e s higher than the same spectrum measured on a scanning monochromator in the same time, where M is the number of resolution elements in the spectrum (i.e., the frequency range/resolution). In the light of this study, three types of measurements were cited in which FTS should show its largest advantage over dispersion spectroscopy: 0 Applications where a large number of resolution elements are to be measured, i.e., high-resolution spectroscopy over a wide spectral range 0 Applications where only a limited time is available for the measurement since, whereas Fellgett’s advan-

A N A L Y T I C A L C H E M I S T R Y , V O L . 46, N O . 7, JUNE 1974

tage is given by t m f o r the S/Nof spectra measured in equal times, it takes M times longer to measure a spectrum a t a certain S / N with a dispersive spectrometer than with an interferometer 0 Applications where, even though Fellgett’s advantage may be relatively small, the signal that is being measured is so weak that an unacceptably long time is required to measure the spectrum with a conventional spectrometer. In this case, even a small time saving can be important. An examination of the more recent papers where spectra have been measured by FTS shows that the measurements generally fall into one or more of these three categories. For example, in the studies of trace levels of air pollutants made by Hanst et al. (8)which were mentioned earlier, spectra were measured a t fairly high resolution to increase the width of the spectral “windows” between atmospheric absorption bands, to increase the specificity of the method, and to increase the absorbance of sharp features (Q branches) in the spectra of light molecules. GC-IR presents a perfect example of an application falling into the second category. In this case, the measurement time is often limited to less than 10 sec by the narrow half width of most GC peaks. In GC-IR the computation and plotting stages are performed after the completion of the chromatogram; and even though they are time-consuming operations, they neither affect the measurement time of the interferogram nor the S/Nof the spectrum for any peak. The largest number of reported FTS measurements fall into the final category. For example, interferometry is certainly the method of choice of most far infrared spectroscopists; here, the number of resolution elements is rather small, but the energy from far infrared sources is very low. An increasingly large number of measurements of the infrared emission spectra of materials a t fairly low temperature is being reported (12-14). Occasionally, these studies are made at fairly high resolution; this author believes that high-resolution infrared emission spectroscopy may well become the method of choice for monitoring pollutants emerging from smokestacks. A typical high-resolution emission spectrum of NzO is shown in Figure 5 which demonstrates the three main advantages of FTS-high resolution, short measurement time, and high sensitivity. Other examples of studies in which the energy is limited include infrared ultramicrosampling (15, 16),studies of adsorbed species on scattering substrates (27, l a ) , and studies of materials under high pressure by use of

-

EM Reagents"

Products for High Performance Liquid Chromatography

Perisorb" Phthalic acid esters Column packing: Perisorb' A

Figure 5. High-resolution emission spectrum of N20 ( A ) Full-range spectrum of 10-cm path length of nitrous oxide at 16OoCand pressure of 19 m m H g , measured on Digilab FTS.14 spectrometer at 0.5 cm-' resolution: measurement time, 10 min. Base line of this spectrum is not at zero owing to low-level emission from hot KCI windows and (possibly)brass walls of cell; hence, absorption features of atmospheric water vapor (between 1900 and 1300 cm-') and carbon dioxide (around 670 c m - ' ) can also be observed in this spectrum, ( B ) Expanded plot of above spectrum between 1360 and 1200 cm-' to demonstrate resolution of this measurement

I

1

I

I

0

10

20

30

Iacl

Perisorb' -these column packing materials have a chromatographically inert glass core with a coating of adsorptive silicon dioxide approximately one micron thick. Perisorb' A : adsorption-active layer of silicon dioxide Perisorb' RP: chemically modified support with a hydrolysis-stable, hydrophobic layer. Perisorb KAT, shell of strong-acid cation exchanger. Particle size 30-40 pm. '

LiChrosorb'" Column packing materials LiChrosorb" * totally porous HPLC column packings. LiChrosorbr" S I 60, LiChrosorb'M SI 100, LiChrosorbr"' Alox T, LiChrosoi b"' SI 60 silanized, mean particle sizes: 5 pm, 10 pm, 30 um

*A registered t r a d e name of E M Laboratories, I n c .

EM Laboratories, Inc. S

S

O

C

of~

~

E. Merck, Darmstadt, Germany

500 Executive Boulevard Elmsford, N e w York 10523 Phone 9 1 4 / 5 9 2 - 4 6 6 0 CIRCLE 66 ON READER SERVICE CARD 654A

of infrared spectroscopy as a n analytical technique.

The Future of FTS

References

It is now generally accepted that Fourier transform spectrometers produce much higher quality infrared spectra than conventional grating or prism spectrometers;indeed, the principal factor working against their more general acceptance is the price tag. Although the first low-resolution computerized mid infrared Fourier transform spectrometer for less than $50,000 has just been announced by Willey Corp., all instruments with a resolution of 0.5 cm- or better sell for more than $70,000. The next two years are almost certainly going to see a reduction in the cost of Fourier transform spectrometers designed for the analytical chemist. (This trend has come about in the last few years for pulsed NMR spectrometers.) When these instruments become commercially available, infrared FTS will be used for routine analysis in a n increasing number of laboratories. Ultimately, we may be sure that something better will replace FTS; the development of tunable lasers at the present time suggests that they might be used to measure high-quality absorption spectra within the next five years. Right now, however, interferometry is causing the rejuvenation

(1) H. A. Laitinen,Anal. Chem., 45,2305 (1973). (2) M. J. D. Low, ibid.,41 ( 6 ) , 97A (1969). (3) J. Pinard, J. Phys., 28 (C2),136 (1967). (4) G. Guelachvili and J . - P . Maillard, Aspen Int. Conf. on Fourier Spectrosc. 1970, G. A. Vanasse, A. T. Stair, and D. J . Baker, Eds., AFCRL-71-0019,p 151, 1970. (5) L. Mertz, Astron. J., 70,548 (1965). (6) M . J. D. Low, Naturuissenschaften, 57,280 (1970). (7) P . R. Griffiths, C. T. Foskett, and R. Curbelo, Appl. Spectrosc. Reu., 6, 31 (1972). ( 8 ) P . L. Hanst, A. S. Lefohn, and B. W. Gay, Appl. Spectrosc., 27,188 (1973). (9) K. L. Kizer, Amer. Lab., p 40 (June 1973). (10) R. R. Willey, personal communication, 1973. (11) R. J. Jakobsen and P. R. Griffiths. Appl. Spectrosc., in press (1974). (12) P. R. Griffiths, ibid.,26,73 (1972). (13) J. B. Bates and G. E. Boyd, ibid., 27, 204 (1973). (14) H . W. Prengle, C. A. Morgan, C.-S. Fang, L.-K. Huan P. Campani, and W. W. Wu. Enciron. 8 i i Technol.. 7. 417 (1973). (15) S. S. T. King, J. Agr. Food Chem., 21,526 (1973). (16) F. Block and P . R. Griffiths, Appl. Spectrosc., 27,431 (1973). (17) M. J. D. Low, A. J . Goodsel, and E. Takezawa, Enuiron. Sci. Technol., 5 , 1191 (1971). (18) M. J. D. Low, A. J. Goodsel, and N . Takezawa, bid., 6,268 (1972).

t

t

Perisorb

B

cells with small diamond windows and diffuse reflectance spectroscopy.

A N A L Y T I C A L CHEMISTRY, VOL

16, NO. 7, JUNE 1974