A Method for the Determination of Iodide in Dried Blood Spots from Newborns by High Performance Liquid Chromatography Tandem Mass Spectrometry Un-Jung Kim, and Kurunthachalam Kannan Anal. Chem., Just Accepted Manuscript • DOI: 10.1021/acs.analchem.7b04827 • Publication Date (Web): 07 Feb 2018 Downloaded from http://pubs.acs.org on February 8, 2018
Just Accepted “Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are posted online prior to technical editing, formatting for publication and author proofing. The American Chemical Society provides “Just Accepted” as a service to the research community to expedite the dissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscripts appear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have been fully peer reviewed, but should not be considered the official version of record. They are citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offered to authors. Therefore, the “Just Accepted” Web site may not include all articles that will be published in the journal. After a manuscript is technically edited and formatted, it will be removed from the “Just Accepted” Web site and published as an ASAP article. Note that technical editing may introduce minor changes to the manuscript text and/or graphics which could affect content, and all legal disclaimers and ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errors or consequences arising from the use of information contained in these “Just Accepted” manuscripts.
Analytical Chemistry is published by the American Chemical Society. 1155 Sixteenth Street N.W., Washington, DC 20036 Published by American Chemical Society. Copyright © American Chemical Society. However, no copyright claim is made to original U.S. Government works, or works produced by employees of any Commonwealth realm Crown government in the course of their duties.
Page 1 of 24 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
1
A Method for the Determination of Iodide in Dried
2
Blood Spots from Newborns by High Performance
3
Liquid Chromatography Tandem Mass Spectrometry
4
Un-Jung Kima and Kurunthachalam Kannana,b*
5
a
Wadsworth Center, New York State Department of Health, and Department of Environmental
6
Health Sciences, School of Public Health, State University of New York at Albany, Empire State
7
Plaza, P.O. Box 509, Albany, New York 12201-0509, United States
8 9
b
Biochemistry Department, Faculty of Science and Experimental Biochemistry Unit, King Fahd Medical Research Center, King Abdulaziz University, Jeddah 21589, Saudi Arabia
10 11 12
*Corresponding author:
[email protected] 13
For submission to: Analytical Chemistry
1 Environment ACS Paragon Plus
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
14
ABSTRACT
15
Dried blood spots (DBS), collected for newborn screening programs in the United States, have
16
been used to screen for congenital metabolic diseases in newborns for over 50 years. DBS
17
provide an easy and inexpensive way to collect and store peripheral blood specimens, and
18
present an excellent resource for studies on the assessment of chemical exposures in newborns.
19
In this study, a selective and sensitive method was developed for the analysis of iodide in DBS
20
by high performance liquid chromatography electrospray tandem mass spectrometry. Accuracy,
21
inter- and intra-day precision, matrix effects, and detection limits of the method were
22
determined. Further validation of the method was accomplished by concurrent analysis of whole
23
blood and fortified blood spotted on a Whatman 903 filter card. A significant positive correlation
24
was found between measured concentrations of iodide in venous whole blood and the same
25
blood spotted as DBS. The method limit of detection was 0.15 ng/mL iodide. The method was
26
further validated by the analysis of a whole blood sample certified for iodide levels (proficiency
27
testing sample) by spotting on a filter card. Twenty DBS samples collected from newborns in
28
New York State were analyzed to demonstrate the applicability of the method. The measured
29
concentrations of iodide in whole blood of newborns from New York State ranged between
30
100 µg/L, and no more than 20% of the population
46
should have values
146
127 [127I-]) and 18O-labeled-perchlorate (107 [Cl18O4-] > 89 [Cl18O3-]; because the molecular ion
147
m/z 127 does not fragment further, the precursor ion selected for Q1 was also selected as the
7 Environment ACS Paragon Plus
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
148
product ion for Q3). Nitrogen was used both as curtain and collision gas. The optimized source
149
parameters were as follows: curtain gas = 20 psi, collision gas = 12 psi, nebulizing gas = 52.5
150
psi, heater gas = 55 psi, ionization voltage = - 1500 V, and source heater temperature = 500 ˚C.
151
Detailed MRM parameters are shown in the supporting information (Table S1).
152
Quality Assurance/Quality Control (QA/QC). Ten-point matrix-matched calibration
153
standards (0.5–200 ng/mL for I- and 50 ng/mL for Cl18O4-) were prepared, and the instrumental
154
responses were compared with a solvent-based calibration standard to correct for matrix effects.
155
As a check for background levels of contamination, equipment blank (EB), laboratory reagent
156
blank (LRB), procedural blank (PB) and two types of matrix blanks (MB) were analyzed. The
157
accuracy and precision of the method were examined by the analysis of iodide-fortified DBS
158
(i.e., in-house DBS) samples. The EB consisted of injection of reagent water by LC/MS/MS (i.e.
159
without any extraction or purification step). The LRB was the reagent water that was injected
160
between samples to verify that there was no carry-over of iodide between injections. The PB was
161
reagent water that was processed similarly to samples by passing through the entire analytical
162
procedure. The MBs were unspotted areas from Whatman 903 filter paper and whole blood
163
samples (used in fortification), which were analyzed similarly to those of samples by passing
164
them through the entire analytical procedure; the MB-1 was (unspiked) whole blood used in the
165
preparation of in-house DBS and the MB-2 was the unspotted area of DBS filter card. This
166
analysis was performed to determine the background levels of iodide in DBS filter card. Matrix
167
spike (MS) samples were prepared at three different concentrations of iodide: low (MS-L; 5
168
ng/mL), medium (MS-M; 40 ng/mL) and high (MS-H; 100 ng/mL). These concentrations
169
covered the range of expected levels found in human blood.10 Two whole blood samples certified
8 Environment ACS Paragon Plus
Page 8 of 24
Page 9 of 24 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
170
by the Trace Element PT Program of Wadsworth Center for total iodine were spiked onto DBS
171
cards and analyzed as quality control samples (QC-A and QC-B).
172 173
RESULTS AND DISCUSSION
174
Chromatographic Separation. Four different HPLC columns, namely the Dionex AS-21
175
anion exchange, the Acclaim Trinity P1 (Thermo Scientific), the Agilent Zorbax Sb-Aq, and the
176
Luna HILIC column (Phenomenex) (see supporting information) were tested for their ability to
177
resolve iodide peak from other matrix interferences with various combinations of mobile phases
178
and gradient programs.30-33 Although the intensity of the iodide signal from solvent and blood
179
extracts was higher with the HILIC column than the anion exchange column, the sensitivity and
180
selectivity of iodide in both whole blood and DBS extract were higher with the anion exchange
181
column. Figure 1 shows chromatograms of iodide (100 ng/mL) injected onto the anion exchange
182
column with the optimized mobile phase gradient and MS/MS parameters in reagent water,
183
blood serum, whole blood, and whole blood spotted in-house DBS specimen. In reagent water
184
(Fig. 1-A), iodide eluted faster (8.5 min for iodide, 11.5 min for Cl18O4-) than that from other
185
sample matrixes (11 – 13 min for iodide and 15.5 – 17.5 min for Cl18O4-). The retention time of
186
iodide from the in-house DBS specimen was reproducible within a 10% standard deviation.
187
9 Environment ACS Paragon Plus
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
188 189
Fig. 1. Typical HPLC-ESI-MS/MS MRM chromatograms of iodide (blue, 100 ng/mL, 2 µL
190
injection volume) and
191
blood serum (B), human whole blood (C), DBS filter card (D), and whole blood spotted in-house
192
DBS specimen (E).
18
O-labeled perchlorate (red, 50 ng/mL) in reagent water (A), human
193 194
Limits of Detection and Quantification. The limit of detection (LOD) was defined as the
195
statistically calculated minimum concentration of iodide at the 99% confidence interval. The
196
LOD was calculated by analyzing a matrix spike sample (DBS) fortified at 1 ng iodide/mL (MS-
197
L) seven times, followed by the calculation of mean recovery and the standard deviation
198
multiplied by the Student’s t-value (3.143). The calculated LOD for iodide in DBS sample
199
extracts was 0.15 ng/mL. The lower limit of quantification (LLOQ) was three times the LOD
200
value, which matched the lowest concentration of the calibration curve at 0.5 ng/mL. A
10 Environment ACS Paragon Plus
Page 10 of 24
Page 11 of 24 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
201
chromatogram showing the signal of iodide at 0.5 ng/mL is presented in the supporting
202
information (Fig. S1). The signal-to-noise (S/N) ratio of iodide at 0.5 ng/mL was between 8 and
203
10.
204 205
Background Contamination. A major issue associated with the use of DBS for trace level
206
quantitative analysis of environmental chemicals is background levels of contamination.21,22,29,34-
207
36
208
contamination by iodide in laboratory supplies and apparatus used in extraction and purification
209
of DBS samples. Trace levels of iodide (below 0.997. In comparison to iodide
232
standard prepared in reagent water, the calibration curves prepared with matrix-matched DBS
233
cards had a slope 2.5 times lower, suggesting the existence of ion suppression from DBS-based
234
sample matrix (Fig. 3). Thus, the quantification of iodide in DBS was based on the slope of the
235
matrix-matched calibration curve, which would eliminate the bias introduced by ion suppression
236
of the sample matrix.
I are radioactive, the isotopic dilution method of quantification is not practical.
18
O-labeled perchlorate was also spiked into samples to
237
The reagent water injected between calibration standards showed no carryover of iodide
238
between samples. A mid-point calibration standard (40 ng/mL) was injected after every ten h to
239
monitor for drift in instrumental sensitivity, and the observed coefficient of variation (CV) was
240
less than 15% for the standard injected throughout the analysis.
241
12 Environment ACS Paragon Plus
Page 13 of 24 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
242 243
Fig. 3. Slopes of responses of iodide from matrix-matched calibration curves (solid line)
244
prepared using dried blood spots and those prepared in reagent water (dotted line).
245
perchlorate was used as the internal standard.
18
O-labeled
246 247
Recovery (Accuracy).
Percent recoveries of iodide spiked onto DBS filter card were
248
determined as the measured concentrations divided by the nominal spiked concentrations at 5
249
ng/mL (MS-L; n=5), 40 ng/mL (MS-M; n=5) and 100 ng/mL (MS-H; n=5). These concentrations
250
cover the range of those found in human blood.10 The mean (RSD) recovery of iodide spiked into
251
sample matrix was 97.7 (6.2) % (Table 1).
252
The accuracy of the analytical method was further validated by spiking whole blood samples
253
certified for iodine levels (two samples certified at different levels) on DBS filter cards and
254
passing them through the entire procedure (QC-A and QC-B; certified values are 53.8 and 40.5
255
ng/mL, respectively) (Table 1). Repeated analysis (n=5 each) of two certified reference samples
256
spiked on the DBS cards showed a recovery of 99-102% with an RSD of 8.74%.
257
13 Environment ACS Paragon Plus
Analytical Chemistry 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Page 14 of 24
258
Table 1. Recoveries of iodide from dried blood spots (DBS) spiked at different concentrations
259
(in percent, mean ± SD) and those of certified reference whole blood samples spiked on DBS
260
filter card. Overall MS-L
MS-M
MS-H
QC-A
QC-B
RSD mean (%)
(n=5)
(n=5)
(n=5)
(n=5)
(n=5)
(%) (n=25)
Nominal concentration
5
40
100
53.8
40.5
95.3±5.96
96.4±5.35
95.6±4.74
102±8.74
98.7±6.10
(ng/mL) Iodide recovery 97.7
6.18
(%, mean±SD)
261 262
(Note. MS-L: matrix spike (low), MS-M: matrix spike (medium), MS-H: matrix spike (high), QC: quality control samples with certified values for iodide)
263 264
Precision.
Intra- and inter-day variations in responses of iodide were assessed by the
265
calculation of precision as percent relative standard deviation [% RSD] and percent relative error
266
[% RE]) from replicate analysis at three different levels of matrix spiked samples (MS-L, MS-M,
267
MS-H and QC-A) (Table 2). RSD (%) was calculated as the standard deviation of observed
268
iodide concentration from each matrix spike divided by the mean measured concentration,
269
whereas RE (9%) was calculated as the absolute difference between measured and nominal
270
concentration of iodide (i.e., |measured conc. – nominal conc.|) divided by the nominal
271
concentration of iodide in each matrix spike. Intra-day variations were calculated on a single day,
272
and inter-day variations were calculated for five separate days. Overall, the inter-day variations
273
were below 12% RSD and below 5% RE.
14 Environment ACS Paragon Plus
Page 15 of 24 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Analytical Chemistry
274
Previous studies have shown that hematocrit levels of whole blood can influence the diffusion
275
of blood on the DBS filter card, which can contribute to intra-disc variability if only a portion of
276
the DBS filter card is used in the analysis.15,37,38 Although our original goal was to use the entire
277
16-mm disc for the analysis (which would alleviate filter paper effect or intra-disc variability
278
introduced by punching a portion of the disc), because validation samples were available to
279
calculate intra-disc variability, we analyzed (n=5 each) 4-mm punches of spiked DBS specimen
280
(QC-A and QC-B) by taking samples from the central and peripheral area of the blood spot, as
281
well as the full circle (Table S2). No significant intra-disc variability in iodide concentration was
282
observed with RSD values