Modern thermogravimetry - American Chemical Society

Norwalk, Conn. 06856. Modem. Thermogiauimetry. Although early activity in the devel- opment of the thermogravimetric method began as long ago as 1903 ...
1 downloads 0 Views 25MB Size
In

-.

- ~ -

Charles M. Earnest Perkin-Elmer Corporation

Norwalk. Conn. 06856

Although early activity in the development of the thermogravimetric method began as long ago as 1903 ( I , Z ) , thermogravimetry (TG) was slow coming into use as an analytical tool in some countries; serious work did not begin in the US.until the 1950s. The man who had the greatest impact on modern analytical thermogravimetry was C. Duval, a Frenchman who studied more than loo0 analytical precipitates using the first commercial thermogravimetric system, the Chevnard thennobalance. Recent years have seen major advances in this analytical technique through improved furnace technology, microcomputer control, and microcomputer data reduction. Today thermogravimetry is an important analytical tool in the various branches of chemistry, geology, fuel science, ceramics, and material science. TG is a technique in which the mass of a substance is monitored as a function of temperature or time as the sample specimen is subjected to a controlled temperature program in a controlled atmosphere. Thermogravimetry can he considered the modern-day frontier of the much older and wellestablished technique of gravimetry. TG and gravimetry differ in that in TG the sample mass is continuously monitored as it is heated; in gravimetry, the sample is weighed after extended periods of isothermal heating. Conventional TG uses much smaller samples (milligram quantities) than those encountered in gravimetry (gram quantities). In TG, only those events associated with a change in mass will be ohserved. Therefore, other thermal anal0003-2700184/A351-1471~1.5010 @ 1984 American Chemical Society

ysis techniques are necessary to study those physical phenomena, such as melting, crystallization, and glass transitions, that do not lead to a change in mass. Differential thermal analysis (DTA) is a technique in which the temperature difference (AT) between a sample and reference material subjected to a controlled temperature program is measured as a function of temperature. A closely related technique, differential scanning calorimetry (DSC), measures the difference in the rate of heat flow to (or from) the sample and reference material as a function of either temperature or time. In thermomechanical analysis (TMA), either the dimension or deformation of a substance under nonoscillatory load is measured as a function of temperature. Dynamic mechanical analysis (DMA), or dynamic thermomechanometry, measures the dynamic modulus and/or damping of a substance under oscillatory load as a function of temperature. The fundamental definitions for all of the thermal analysis techniques mentioned here state that the measurement is made while the specimen is subjected to a controlled temperature program. One might be misled hy this commonality to think that in thermogravimetry the sample temperature is always being elevated in a continuous fashion at a regular programmed rate. Although this is often the case, many important thermogravimetric procedures involve isothermal monitoring of the mass loss from a material. Thus, when one or more isothermal holds are involved in the total heating program, the mass loss is often monitored vs. time. One ANALYTICAL CHEMISTRY, VOL. 56, NO. 13, NOVEMBER 1984

r L 1471 A

Flgure 1. TG thermal curve with extrapolated onset temperature assignment for the decomposition of polytetraflwroethylene can see that it is with the isothermal thermogravimetric procedures that the closest kinship to conventional gravimetry lies. The technique of thermogravimetry is often referred to as thermograuimetric analysis or TGA. The name thermograuimetry and its abbreviated symbol TG were given to this technique by a nomenclature committee of the International Confederation for Thermal Analysis (ICTA) (3.5). This committee has made an effort to remove the term analysis from as many thermal methods as possible. The A in DTA remains because DTA is such an old and established technique. One reason the removal of the A from TGA has not been accepted by some is the possible confusion of TG with T8,a glass transition commonly studied by the thermal analysis techniques of DSC, TMA, and DMA. This has caused problems in computerized literature searches in which the computer does not distinguish between TG and Tg.This committee also eliminated the use of thermogram in favor of thermal curue to refer to the graphical presentation of thermal analysis data. Figure 1shows a typical TG thermal curve that gives the weight percent of the original sample vs. temperature for a polytetrafluoroethylene (F'TFE) specimen that was heated in dynamic nitrogen at 20 'C/min. The two major items of information one obtains from such a curve are the extrapolated onset temperature and the percentage weight loss assignments. The data handling shown in Figure 1 describes the determination of the T0,,=t (606.7 "C) for the decomposition of this F'TFE specimen. Since this 1472A

thermal curve was obtained in a dynamic inert atmosphere, the Tonetis a measure of the thermal stability of the PTFE. One also notes that all of the original F'TFE specimen is converted to volatile decomposition products. Organic polymers exhibiting this behavior may he quantified in the presence of thermally stable additives or fdlers, such as carbon black and glass, by such pyrolytic separation.

lnstrumentatlon To perform experimental thermogravimetry, the instrument must be capable of both simultaneous heating

and weighing. Therefore, the instmment used for performing thermogravimetric measurements is often referred to as a thermobalance. This term originated with the early work of Honda (6),who described an apparatus with which be obtained TG curves for MnSOqH20, CaC08, and CrC03. The major instrumentation requirements for performing thermogravimetry include a sensitive recording microbalance; a furnace and an appropriate enclosure for heating the sample specimen; a temperature programmer, heater control circuitry, and associated electronics; a pneumatic system for dynamic purging of the furnace/ sample chamber; a data acquisition system; a purge gas switching device; and first-derivative capability. The latter two items are optional: The purge gas switching capability is needed only for those applications in which the purge gas is changed during the experiment, and the fust-derivative computer is often eliminated when computerized data-handling systems are used. Many commercial TG systems are available. The major differences in these are in the furnaces (size, design, and positioning), degree of microcomputer control of the hardware, and in the capabilities of the data acquisition systems offered with the thermal analysis system. There is also some variation in the maximum balance sensitivity used in these systems. Figure 2 shows a TG thermal curve obtained for the thermal decomposition of nylon 6,6 at a constant heating rate of 20 Wmin. In many cases, the first-derivative trace of the weight lost profile is presented along with the TG

Figure z. IU ana UIU rnermai curves tor me aecomposition ot nylon 6,6 in 6ynam-

ic nitrogen atmosptwe

ANALYTICAL CHEMISTRY, VOL. 56, NO. 13, NOVEMBER 1984

Those small LC columns need a better filter. s. The Rheodyne 7315 for lumns And the Rheodyne ort high-speed columns. They’re identical except for a tiny internal filter element. The accompanying chromatograms show the dramatic improvement in resoluhon that occurs when a new low-dispersion filter is used in place of a conventional filter. For more about these new column inlet filters-plus our Tech Note 6 describing the experimental effect of filters on resolution-contact Rheodyne, Inc, PO Box 996, Cotati, California 94928, U S A Phone (707)664-9050.

I

The new generation chi A.

The CAMAG TLC S U is capable of: scanning tt layer chromatograms by sorbance or by fluorescence scanning electropherograms or similar objects in transmitted light . scanning drcular/anticircular chromatograms . automatic recording of spectra . dualwavelength scanning

- .-- .-

.

Ask for

our 12-page full-

CH -4132 MunenzlSwazeriand

For funher information contact

-1

APPLIED

New Hdnouei County lndurtrial Airpark Route6. Box 55. Wilmington NC28405 8001334-8395 in NC9191763-4536 C A M A G ha$ agents 8n almost a l l counfrier. (n U 5 A joinfly represented by AAl and E M Sciente

1474A

---.- . --.-

Temperature("C)

I Flgun 3. Thermal C U N e (a) and derivative itwrmai curve (b) for selected Curie point standards (alumel, 163 'C; nldtel. 354 O C ; perkalloy, 596 O C ; and iron. 780 'C)

thermal curve. This first-derivative trace corresponds to the technique of derivative thermogravimetry (DTG), the resulting DTG thermal curve representing the rate of mass loss vs. temperature. The DTG ordinate has units of mass per unit time (mglmin). The DTG thermal curve may be generated simultaneously with the TG thermal curve by electronic means. This may be achieved by a simple R-C circuit or by the more modern approach of an operational amplifier with active filtering by field effect transistors to eliminate or minimize noise. When a computer is used as the data-handling device for thermogravimetric studies, the DTG curve may be generated, after the TG data are stored or received, by a mathematical algorithm

ANALYTICAL CHEMISTRY, VOL. 56, NO. 13, NOVEMBER 1984

in the software. The TG and DTG thermal curves presented in Figure 2 for nylon 6.6 represent a hard-copy printout from a computerized thermogravimetric system. Thermal curves obtained from thermogravimetry offer what is poasibly the most accurate ordinate scale (weight or weight percent) of all thermal analysis techniques. The temperature axis, on the other hand, is often less well defied. There are several problems that complicate the &bration of the TG temperature axis. First, since the thermocouple is generally close to but not in contact with the sample during analysis, the actual temperature of the sample is not being measured. The composition and flow rate of

Elemenkid d y s b made Convenient, Rapid, 4 Nondestructive Simultaneous, Quantitative, Productive, Sensitive, Easy, and Affordable.. . If you haven't considered energy dispersive x-ray fluorescence (EDXRF) hecause it was too cumhersome. different, or espensive. call Tracor Xray and find out ahout the Series to00 XRF System and how it compares to Ad, ICP and other elemental analysis methods - including other XRF systems.

The Tracor Xray t000 XRF System nil1 change the way you think ahout SRF. And it just may clrauge the w y you do elemental analysis. At Tracor Xray. EDXRF has come dowi to earth.

I

Tracor Xray 345 E. Middlelicld Anad Mountain View. CA 94043 (4151 9674350

ANALYTICAL CHEMISTRY, VOL. 56. NO. 13, NOVEMBER 1984

I

-

147.5,.

the Maaic of AIItech l O N-PAKD’” GC Columns offers

* An end to fragile packed glass columns * Improved efficiencies * Faster analysis times * Extremely inert packing * Bonded phases * Simple installation * convenience Conventional GC * Tested and guaranteed columns

Alltech 0.53 mm fused silica open tubular NON-PAKD columns offer the best of both worlds of capillary columns and pack columns. Now many analyses done by packed GC columns N-PAKD columns

r

urnbe 1 ON READER SERVICE CARD

1476A

ANALYTICAL CHEMISTRY. VOL. 56, NO. 13, NOVEMBER 1984

the dynamic gaseous atmosphere surrounding the TG sample also influence the sample temperature. The composition of the purge gas is chosen to satisfy the need of the application or analysis. For those analyses made in inert atmospheres, such as pyrolysis, either nitrogen or argon is generally used. Characterizations or analyses requiring oxidative environments are performed in dynamic air or dynamic oxygen atmospheres. As was pointed out years ago by Gray (7),an additional factor that can influence the true sample temperature in thermogravimetry is the effect of radiation. As temperatures in excess of 500 OC are reached in a thermogravimetric apparatus, radiation progressively hecomes the dominant mechanism for heat transfer. When using an open-cup-type TG sample pan at these temperatures, radiation to and from the surface of the sample can become a significant factor. Thus, a black sample specimen such as coal can be at a different temperature than a white sample such as calcium carbonate or calcium oxide, all other conditions being the same. The hest approach to calibrating the actual sample pan temperature has been through the use of ferromagnetic materials that exhibit known Curie point transitions. In this case, small pieces of the ferromagnetic metal (or metal alloy) are placed in the TG sample pan, and the actual mass of materials is either electronically tared or subtracted using a computer software algorithm. A small permanent magnet is then placed either above or helow the TG sample pan. The vertical component of the magnetic field causes the balance to read either more or less depending on the position of the magnet. When the magnet is placed helow the TG sample pan, the magnetic force on the sample acts as an equivalent magnetic mass on the balance beam to increase the apparent sample weight. When heated, the magnetic domains in ferromagnetic materials become disoriented and transform to the paramagnetic state at cbarachristic temperatures (Curie points). At these temperatures, the TG thermal curve will show an apparent weight loss due to the loss of magnetic mass. Figure 3 shows the thermal curves obtained by the author using several Curie point standards. Since the Curie point transitions are reversible, the same standards may he reheated numerous times for the purpose of temperature calibration. Furthermore, in cases where one wants to minimize the effects of radiation, the standards may he buried in a specimen of the sample material for calibration purposes. This type of calibration procedure is

Union Giken apparatus achieves a dead time as short as f i millisecond for measurement of fast reactions. G a s pressure is directly applied to the surface of the reactants in each reservoir, mixing and flow being controlled from the end of the system by a high-response stop valve. The RA-401 STOPPED FLOW SPECTROMETER is the ideal match for fast reactions.

" l " C

*.,&*_

' $ 7

*I

I,..._ "

L

-* .(1

. . , I I

.,l"

i i S l l . ,

,rl

The RA451 DATA PROCESSOR generates a simulated reaction cuwe from the measured results and automatically calculates the rate constant.

Figure 4. Block diagram for a microcomputer-controlled thermal analysis system

ideally suited for TG instruments that are constructed with amicrofurnace inside the furnace tube, since a small horseshoe magnet may be placed around the n m w furnace tube. However, for those systems using larger furnaces external to the furnace tube, the positioning of the permanent magnet becomes much more awkward. An alternate approach, which uses a fusible link of a melting-point standard and seems to show some promise for temperature calibration of those instruments using external furnaces, has recently been published (6.9).

Microcomputer controllers The firat step in the computerization of thermal analysis instrumenta-

tion was the introduction of microcomputer-based controllers in 1978. These controllers made possible multistep heating programs, automatic purge gas switching, methods storage and recall, and completely unattended operation. The precision of the power delivery to the furnace has also been greatly enhanced by the controllers. One such system includes an automatic temperature calibration routine, in which the ppgrammer temperature and the actual thermocouple temperature are matched at three points along the total heating ramp. Thii has been extended to a nine-point matching program in one of the more recent commercial systems. Figure 4 shows a hasic block di-

I

.Dl"".

0,s "rn

The RA415 RAPID SCAN AlTACH. MENT monitors the change in aborp tion spectrum during the reaction.

Other attachments are available foi T-jump, P-jump, fluorescence and flash photolysis.

I

ATAOO BUSSAN CO., LTD, 7 4 3 5Ch0mM1rnb.thi. Minau-ku, To*+. 108 Japal) Tela: 8411 Ceble: TAGOA TOKYO Phon-?OJI 4324741

Figure 5. Miuocomputer-wntroiled TG proximate analysis of a bituminous coal specimen

CIRCLE 3 ON READER SERViCE CARD

1478A

ANALYTICAL CA-EMISTRY. VOL. 56,

No.

13. NOVEMBER I984

BAKERBOND Chiral PhaseTM Products Your Best Choice for Racemic Separations

I I

What can anyone offer that is new and exciting in a quarter meter type monochromator? PTl's model 01-001 monochromator offers you five distinct major benefits unmatched by any competitive model: low price, variable slits that are standard, versatile optical configurztion, full range of accessories and a fourteen day trial period to convince yourself asto thevalidityof theclaims made by the company. Price In the market today, high quality quarter meter monochromators range in price from $2,500.00 to $3,500.00. Then there are expensive options to make the instruments conform to your needs. PTl's model 01-001 costs $ 1,975, complete with virtually everything that you need. Variable Slits Normally, monochromators are fitted with one or two fixed slits. Additional slits cost upward of $100.00. Variable slits are always optional and usually cost in excess of $ 700.00. The PTI modell 01-001 comes complete with two micrometer adjustable slits including in the basic price. Versatile Optical Configuration Virtually all monochromators available today, are only usable in one optical configuration.Since monochromatorsareused generally in conjunction with other instrumentation,one must design the system to fit the monochromator. PTl's Monochromators offer a choice four different optical configurations. Two small 45°mirrorsenable the input and output to be in line. By removing one or both of these mirrors, right angle and parallel configuration of input and output beams can be easily achieved. No longer is it necessary to adapt systems to monochromators. You can now design a system to use the optimum light path and geometry defined by your needs. Full range of Accessories PTI providesafull rangeof accessoriesforour instruments, in keeping with our philisophy of high quality at a reasonable cost. These accessories include light sources, power supplies, stepping motor drives, sample chambers, PMT tubes housings. Free Trial Although our instrument is the newest on the market today, it has undergone thousands of hours of field tests involving hundreds of units. Costs prohibit us from demonstrating this instrument to you. However, we feel that you should not be asked to buy a unit on faith or from specification sheets. Therefore, if you place an orderwith us, fora14 day period after you have received the instrument, should you be unhappy with its performance for any reason, return it and we will refund your money.

Photon Technology International Inc. us.:

Europe: Gartnerweg 49 D-2082 Tornesch, West Germany Princeton, N.J. 08542 USA Telephone: (04122) 5 10 61 Telephone: (609) 921-0705 Telex: 238818 RCA Telex: 218500 CIRCLE 166 ON READER SERVICE CARD P. 0.Box AA

1480A

ANALYTICAL CHEMISTRY, VOL. 56, NO. 13, NOVEMBER 1984

agram of a microcomputer-controlled thermal analysis system. The microcomputer controller is in communication with the heater control circuitry to control the voltage ramp to the furnace and is in direct communication with the thermal analyzer to continuously monitor the sample temperature. The emf-vs.-temperature table is contained in read-only memory (ROM) in the microcomputer controller for the particular thermocouple being used with the analyzer. The emf signal is converted to temperature and displayed on the microcomputer controller screen. More than one set of firmware components may be included in the controller so that more than one type of analyzer (DSC, TMA, etc.) can be used with the same controller. The analytical temperature program is developed and entered into random access memory (RAM)though a keyboard. Error messages, actual temperature, etc., are communicated to the user via the screen display. Figure 5 shows a multistep TG analysis with purge gas switching that can be performed automatically with a microcomputer controller. In this rapid proximate analysis procedure, a bituminous coal specimen is rapidly heated to 110.OCunder flowing nitrogen and held for 5 min until all of the moisture is lost. This isothermal step is followed by dynamic heating of the sample a t 100 "C/min to 950 "C; the temperature is then held isothermally until all of the volatile matter has been liberated. At this point, the purge gas is automatically switched from nitrogen to oxygen through the side arm of a two-arm furnace tube through the use of a gas-switching accessory controlled by the microcomputer. As a result, the final weight loss step assigns the percentage of fixed carbon for the coal specimen, and the remaining weight in the pan is read directly as the ash value. The system automatically cools down on completion of the heating program, and the purge gas is automatically changed back to dynamic nitrogen. Analyses such as this required total operator attention prior to the advent of microcomputer controllers. For details of the rapid proximate analysis of coals, see Reference 10. Other examples of TG multistep analyses include the compositional analysis of rubber and elastomer formulations, lube oils, and polymers filled with carbon black. The appropriate analytical heat-hold programs may be developed for these analyses and stored in the memory of the controller along with the valve times for purge gas switching. The box on p. 1482 A lists a number of applications that can be performed by thermogravimetry. For more de-

X-RAY MICROANALYSIS SYSTEM

fa\-

THE X-RAY ANALYSER YOU‘VE ALWAYS PROMISED YOURSELF

‘m’-] ,“.I

132

Obi

W E A x i

itriiLt L A R D

LINK SYSTEMS

LINK SYSTEMS LIMITED, HALIFAX ROAD, HIGH WYCOMBE. BUCKS HP12 3SE. ENGLAND. TEL: 0494 442255 TELEX; 837542 LINK HWG

LINK SYSTEMS(FRANCE), LE MAZIERE, RUE DES MAZIERES, 91033 EVRY CEDEX TEL (6)078-10-20. TELEX: 691 884 F

LINK ANALYTICAL PO BOX 50810, 3290 WEST BAYSHORE ROAD, P A L 0 ALTO, CALIFORNIA 94303, USA. TEL: (415) 856 2726 TELEX: 334420 MCI PLA

Some applications of thermogravlmetry Environmental stability Thermal stability in inert atmospheres Oxidative stability Compositional analysis Rubber and elastomer formulations Proximate analysis of coals and coal products Lubricants and lube oils Polyethylenelcarbon black Moisture determinations Residual solvent determinations Ash and inert filler determinations Characterization and analysis of clays and minerals, oil shales,tar sands, etc.

tails, consult recent literature reviews in this JOURNAL (I1), selected reference hooks (12,13), and instrument manufacturers' literature.

Microcomputer data acquisition Until recently, potentiometric recorders were used with thermal analysis equipment. These varied in type from time base strip chart recorders to dual-pen XYlYp recorders. Due to the ease and speed of microcomputer data systems, the potentiometric recorder is rapidly disappearing from the thermal analysis laboratory. Commercially available instruments containing either built-in or stand-alone microcomputer data acquisition, reduction, and storage became available as early as 1979. There are presently at least six different stand-alone computers and one intelligent recorder that are marketed with, or for use with, thermal analysis equipment. A wide variety of software packages are available with these data systems, easing tasks that were once time-consuming. Among these, differences in both computer hardware and software capabilities exist. These include the presence or absence of a CRT screen, whether or not the data system is operator programmable, use of floppydisk and/or hard-disk storage, and whether or not the system is capable of multi-instrument operation. Most of these differences depend on the size of the central processing unit (CPU) and operating system used. The major advantage of a CRT screen is the ahility to optimize the data and perform data analysis on the screen prior to obtaining a hard copy of the results. Generally spaakiig, in commercial systems, the raw data are smoothed by boxcar averaging. This means that 1482A

Microdistillation of petroleum and syncrude fractions Sublimation Plasticizer content of polymers Sorption studies Catalytic activity Decomposition kinetics Flame retardancy of polymer additives Vacuum TO studies Process simulation Loss on ignition of industrial raw materials Establishment of stoichiometriesof dehydration Characterization of cements

each data point saved in the data set is a result of the averaging of many raw data points. One of the greatest advantages that accrues from the use of these systems is that the operator will seldom, if ever, have to repeat a thermal analysis experiment because of improper choice of ordinate scale sensitivity. In all of the commercial software packages, the operator may rescale the ordinate axis (as well as the temperature axis) after the data for the entire analysis have been received by the CPU. In some commercial systems, the ordinate axis is automatically rescaled upon receipt of the last data point for

the run. To estimate the time saving, one can calculate that a TG characterization run conducted at 10 ' C h i n from ambient to 1025 "C consumes 100 min, and the rerun of such an experiment would require a minimum of 2 h if the cool-down time and sample reloading time are included. Another important feature of these data stations is storage of the original data in a file for recall at a later date. The concept of multitasking It has recently become possible to operate and take data from more than one thermal analyzer both simultaneously and independently. Figure 6 is a block diagram of a system that is capable of multi-instrument operation. This capability depends not only on the size of the CPU but also on the operating system used. As can be seen in the diagram, the CPU undergoes two-way communication with two separate intelligent controllers via RS-232C communication interfaces and data buses, each of which controls a thermal analyzer. These intelligent instrument controllers should not be mistaken for simple communication interfaces; they are microcomputer-based systems themselves. The thermal analyzer units, in this particular case, may he two TG units, two DSCs, or one of each. This system uses the concept of information windows on the video display unit (VDU) to communicate the progress of an analysis being performed by the second analyzer while the other analysis is being monitored. Such a totally computerized system has essentially eliminated control knobs from ther-

Figure 6. A computerized thermal analysis system with multi-instrument operating capability

ANALYTICAL CHEMISTRY, VOL. 56, NO. 13, NOVEMBER 1984

XtraChrom. our new chromatography sohvare for the Series 4400X data systems has so many new features, you need to see it to believe it. Enhancements like our graphics routines to expand and rescale chromatograms instantaneously by cursor position on the CRT. No one else has it. Graphicsaided method and sequence setups to let you mew the run and optimize your parameters. N o o n e else has that either. Plus real-time data acquisition for on. the-fly integration and expanded autocalibrations and sequence programs to make it totally automatic . . .and easy.

With the powerful HP Series 200 desktop computer, you get a

smaller, high quality inkjet printer.

e

The new Series 44OOX supports over 10 different mass-storage devices, from dual-microfloppy drives to 55- megabyte hard disk systems. You can even connect X/Y plotters for color data presentations. No other system is packed with this potential. We're out to make a believer of you. So please call or write 10061 Bubb Road, Cupertino, California, 95014 (408) 725-1107. East Coast: 205 Robin Road, Paramus, New Jersey, 07652 (201) 967-8980 CIRCLE 144 ON READER SERVICE CARD

AllALYTlCAL

Our soft ----?doe~somuchmore, !r1 imystemcomesc'

r

-

TG

urnace area and COllectOr

Flgure 7. Coupling of TG systems for EGA or ECO

mal analyzers. All control parameters, temperature calibration, etc., are entered from the keyboard.

How to increase precision in HPLC sample injection. Rheodyne's Tech Note 5 shows how you can maximize HPLC precision by using skillful sample injection techniques. The information is drawn from hundreds of closely controlled experiments using different types of sample injectors and different loading methods. The 6-page summary of results is written to help both the beginner and the experienced chromatographer. The experiments yielded some surprising answers to hghlv wactiial questions. One question: How much sample must you waste to fill a sample loop 'tompletely"? (It's more than you are likely to think.) Another question: When you are reading a syringe to judge the amount of sample being loaded, how much sample can you inject without impairing accuracy? (It's less than you may think.) Get the answers to these and other salient questions now.

EGD and EGA The analysis of the dynamic purge gas exit stream of DTA, DSC,and TG analyzers has become increasingly popular in recent years. The concept of evolved gas detection (EGD) differs from evolved gas analysis (EGA) in that in the former the absolute identity of the component or components in the exit gas stream need not be determined. EGD and EGA are used for the simulation of many industrial proceeaes in which the materials are heated, for establishing decompwition mechanisms, and for determiningthe stoichiometric relationships for such decompositions. Figure 7 shows a simplified block diagram for performing either EGD or EGA by coupling an appropriate anal p r to the thermogravimetric system. As in any of the hyphenated analytical techniques, the coupling interface often requires careful attention and consideration in design and implementation. Among the most popular and powerful combinationsare T G

mass spectrometry (TG-MS)and TGtandem mass spectrometry (TGMS-MS). In these combined systems, the evolved gas from the TG furnace may be sampled with a capillary or a jet separator. In most cases, a heated line is needed in the coupling to avoid condensation of volatile matter. This need is minimized by keeping the distance between the TG exit system and the analyzer small. T w o recent reviews (14,15) contain details of the TG-MS technique. The schematic given in Figure 8 showsthe interfacing of a TG furnace tube to a commercial triple quadrupole mae~spectrometerthat utilizes an atmospheric pressure chemical ionization (APCI) ion source. Thii T G MS-MS interface has been used (16) with both internal microfurnace TG systems 88 well aa those employing extemal furnaces.The ion source uses air nn the chemical ionization reagent. Molecular species from the TG effluent enter the ion source from the hented transfer line and converted to their quasi-molecular or molecular ions via well-characterized APCI reactions. The ions are then focused into the high-vacuum analyzer portion for

Send for Tech Note #L Contact Rheodyne, Inc.,

PO.Box 996, Cotati, California 94928, USA. Phone f707) . . 664-9050.

s2 i RHEODYNG THE LC CONNECTION COMPANY

Flgure 8. A TG-MS-MS interface(from Reference 16, with permission

CIRCLE 176 ON READER SERVICE CARD ~

1484A

ANALYTICAL CHEMISTRY, VOL. 56. NO, 13. NOVEMBER 1984

-1

_.youcan interface me uu-I witnpopmr E ' s . .. In industrial applicabons, you get fast, fully automated answers without compromising allows unattended analydata quahty The DU-7 sis of over 100 samples with automatic printout of data - including results for quantitative analysis and wavelength scans. Analysis capabilities include linear and non" linear standard curves with statistical analysis of standard curves and storage of calibra~~

Multicomponent analysis is available for -. mixtures of up to five components And you can calibrate with standards that are themselves mixtures. __ Start plotting to acquire a DU-7today by contacting your Beckman representative or Beckman Instruments, Inc.,PO. Box C-19600. Irvine, CA 92713.

6ECKMAN CIRCLE 23 ON READER SERVICE CARD

-

MS or MS-MS analysis. One of the best examples of EGD is the use of a flame ionization detector (FID) in the TG analysis and characterization of source rock materials. For example, the Green River oil shales of the western US.contain a sizable carbonate component. When performing TG pyrolysis studies on these shales one may follow the pyrolysis, particularly at temperatures of 500 ‘C or greater, by connecting the TG dynamic exit stream to an FID. In this m e , the volatile organic pyrolysis producta may be monitored vs. either sample temperature or time, and the carbon dioxide due to decompcaition of the inorganic carbonate component w i l l be ignored by the FID. The concept of EGD and EGA is one of the most active areas in thermal analysis. There are some who feel that when studying new materialsTG and DTG thermal curves represent incomplete data without the inclueion of EGA.Quantitation of sample components by conventional TG may be made only if a stoichiometric relationship exists between the gaseous decomposition producta and the original material. Obviously, quantitation is much simpler when the component is volatilized without decomposition, such as in the assignment of moisture content or residual solvent.

J. F.; Gill, P. S..Eds.Plenum Press: New York, N.Y., 1984, pp. 105-11.

References Nernst, W.;Rieaenfeld, E.H. Ber. Dtsch. Chem. Ges. 1903,36,2086. (2) Wendlandt, W.In “A History of Analytid Chemistry”; Laitinen, H. A.; Ewing, G. W.,Eds.; American C h e m i d Soeiet Washington, D C ,1977. p 48 (3) Macf);enzie,R.C. J. Th;rmLA&.’ (1)

.

1975,B. 197.

(4) MacKenzie, R C. Thermorhim. Acta 1979,28,1-6.

( 5 ) Lombardi, G. “For Better Thermal

Analysis,” 2nd ed.;International Confed-

eration for Thermal Analysis: Rome, 1980.

(6) Honda, K.Sci. Rept. Tohoku Uniu. 1916,4,97.

(7) Norem, S. D.; O’Neill, M.J.; Gray,

A. P. T h e m h i m . Acta 1970, I , 29. (8)McChie, A. R AM^. Chem. 1983.55,

..

Charles M.Earnest is senior researcher for the Thermal and Elemental Analysis Department at the Perkin-Elmer Corporation,Norwalk, Conn. Prior to joininz Perkin-Elmer in 1978, he u&assocTate professor of chemistry at Northeast Louisiana Uniuersity,Monroe, Lo., and abo served on the faculties of the Uniuersity of Rttsburgh ot Johnstown ond Stillman College. H e receiued both his BS in chemistry and his PhD in analytical chemist0 from the Uniuersity of Alobama. Earnest’s research effortshave been in energq and environmental analyses, moterials characterization, and in uarious aspects of chromatographic science.

I

STEP UP COST~kmcIENCY

1MCI Has Just the Moisture

Measurement System You Need. L.

Performance takes a big step forward in our new microproc888or-mtroiCA-05 Moisture MeterNA-05 Water Vaporizer s y s t e m with built-in printer. Moisture measurement is fully automatic. High coulometric Karl Fischer titratim accuracy and sensitivity combine with wide modelcalculation functions for amazing application versatility. Advanced features include an airtight titratlon cell, 16digit display, automatic calendar and optional electronic baimcslcomputer Interfacing. Write for complete details today.

A

! I T

MlTSUBlSHl CHEMICAL INDUSTRIES LIMITED InalmmenIs W., Mlt.ublsh1 Bldp, 5.2 U.ruwudl Z C h m . Chlyoda-ku, T O W tW, Japan 4eLddms.s: MB

Telex 5248

CIRCLE 140 ON I -

1406A

n

_ _ I SERVICE CA

ANALYTICAL CHEMISTRY. VOL 56. NO 13. NOVEMBER 1984