Potentiometric measurements in air-segmented streams without

Biochemistry, University of Toronto, 100 College Street, Toronto, Ontario, M5G 1L5 Canada. INTRODUCTION. The concept of air-segmented continuous-flow ...
0 downloads 0 Views 560KB Size
Anal. Chem. 1989, 61, 504-508

SO4

TECHNICAL NOTES Potentiometric Measurements in Air-Segmented Streams without Debubbling Theodore K. Christopoulos Laboratory of Analytical Chemistry, University of Athens, 104 Solonos Street, Athens 106 80, Greece Eleftherios P. Diamandis* CyberFluor Inc., 179 J o h n Street, Suite 400, Toronto, Ontario, M5T 1 x 4 Canada, and Department of Clinical Biochemistry, University of Toronto, 100 College Street, Toronto, Ontario, M5G 1L5 Canada INTRODUCTION The concept of air-segmented continuous-flow analysis (CFA) was introduced by Skeggs in 1957 (I) and has been used extensively since then (2,3). Although newer approaches like flow injection analysis are emerging (4), CFA is still used, especially in selected areas of industrial and pharmaceutical applications such as quality control. Two approaches have been generally used to solve the problems related to the presence of bubbles in the measuring stream: (a) aspirate the bubbles with a device called a “debubbler” used in the Technicon Autoanalyzer system or recently published in a new configuration (5) or (b) use “bubble gating” techniques to electronically remove signal noise during the passage of the bubbles through the detector (6, 7). Ion-selective electrodes (ISE) are now widely used as detectors in continuous-flow automated systems (8). Technicon succeeded in eliminating the noise generated by air bubbles in their sequential multiple analyzer with computer (SMAC) system, employing ISE for Na+ and K+, by electronic means (9). In most other applications, best results are obtained when the bubbles are removed before reaching the potentiometric detector. In this report, we propose a simple and inexpensive way of carrying out continuous-flow potentiometric measurements in the presence of air bubbles. This is accomplished by measuring the emf of the electrochemical cell simultaneously and continuously a t two points of the flowing stream. The performance of the proposed bubble-through electrode unit was evaluated for iodide (solid-state) and propantheline (PVC-matrix) sensitive membranes. The observed potential oscillations in the presence of bubbles were 10.2 mV in all experiments conducted. EXPERIMENTAL SECTION Reagents. All chemicals used were of analytical-reagent grade. Standard solutions were prepared with deionized, distilled water. Propantheline bromide was obtained from Sigma Chemical Co. (St. Louis, MO), poly(viny1 chloride) (PVC) of very high relative molecular mass from Aldrich Chemical Co. (Milwaukee,WI), and o-nitrophenyl octyl ether (0-NPOE) and sodium tetraphenylborate from Fluka (Buchs, Switzerland). Apparatus. A schematic diagram of the automated analysis flow system used is shown in Figure 1. A Gilson Minipuls variable-speed peristaltic pump facilitated continuous pumping of solutions through the electrode assembly. All flow rates refer to liquid flow only (air flow is not included). The sampler was an A40 Sampler I1 form Hook and Tucker (Surrey, U.K.). The

* To whom correspondence

Inc.

should be addressed at CyberFluor 0003-2700/89/036 1-0504$01.50/0

extemal reference saturated calomel electrode was obtained from Radiometer (Copenhagen, Denmark). All potentiometric measurements were carried out at ambient temperature with a Coming Model 12 Research pH/mV meter and recorded on a Radiometer strip chart recorder (REC 61) through a high-sensitivity unit (REA 112) interface. Construction of the Salt Bridge Electrolyte-Sample Solution Contact Zone. A cylindrical piece of porous ceramic, 6 mm long and 1mm in diameter, was constructed from a standard porous ceramic plate as described previously (IO). This was used in all flow-through cells described below to facilitate contact between the measuring solution and the salt bridge of the extemal reference electrode. Before use, the ceramic plug was placed for 5 min in the appropriate salt-bridge solution (see below) to wet, and any excess was then removed with absorbent paper. Construction of Bubble-Through Flow Cell for SolidState Membrane, ISE. The cell confiiation is shown in Figure 2a. An AgI-Ag,S coprecipitate pellet-sensing membrane was employed, prepared as described elsewhere (1I). After construction, the pellet was ground with emery paper to obtain a semicircular shape (Figure 2a). The solid membrane and the ceramic plug were then embedded in the epoxy resin (Epofix resin; M. Struers Scientific Instruments, Copenhagen, Denmark) by using a plastic mold (dimensions 4.5 X 4 X 1.5 cm) as reported earlier (10). Holes of appropriate diameter were then made on the upper surface of the resulting orthogonal cast block as shown in Figure 2a, with slow-speed d r i i , using regular drill bits. Two flow-through channels, 1.5 mm in diameter and at a 5 m m distaxice between each other, were also drilled vertically to the solid membrane and the ceramic plug, so that each channel contacted both of these components, leaving the ion-sensitive membrane opposite the ceramic junction. Finally, a pump tube (2 cm long and 1mm in internal diameter) was used to join the two channels together by means of two stainless steel nipples. Cell Organization. The internal reference solution (lo-* M AgN03)and the salt bridge (100 g L-’ NaN03) were added in the appropriate holes of the flow-through unit. A silver wire and a saturated calomel electrode were used as internal and external reference electrodes, respectively. The electrochemical cell was mounted in the continuous-flow system by using two stainless steel nipples. Preparation of a PVC Membrane Sensitive to Propantheline. The construction and characteristics of a propantheline liquid-membrane ISE and its application to pharmaceutical analysis were described previously ( I 2 , I 3 ) . The ion exchanger was the tetraphenylborate salt of propantheline dissolved in o-nitrotoluene. The liquid ion exchanger for the PVC membrane used in this work was prepared in the same manner except that o-nitrotoluene was replaced by o-NPOE. The liquid ion exchanger was then entrapped in PVC by the procedure described elsewhere (14). Construction of Bubble-Through Flow Cell for PVCMatrix Membrane ISE. The configuration of the constructed cell is shown in Figure 2b. The unit consists of two orthogonal Plexiglas pieces, A and B, with dimensions of 3 X 1.7 X 1.4 cm. The pieces were assembled as follows: 0 1989 American Chemical Society

ANALYTICAL CHEMISTRY, VOL. 61, NO. 5, MARCH 1, 1989

1 -lH-1

Electrometer

505

1 0 1

Wash whitr-whitr

Recorder

PUMP Figure 1. Schematlc diagram of the continuous-flow system without debubbler: ISE,Ion-selecthre electrode; RE, reference electrode; HO, “cactus”; ISA, Ionic strength adjustlng solution. The color code of the pump tubes is that given by Technicon.

m.

1

3

B

-6

C Y

F a

b

Figure 2. (a) Bubblathrough electrode unit for solid-state membrane ISE: 1, external reference electrode; 2, internal reference electrode; 3, salt bridge; 4, internal reference solution; 5, electrode body; 6, piastic loop connecting the Inlet and outlet channels; C, ceramic plug; M, electrode membrane; PI, P, points of sample contact with the membrane. (b) Bubblathrough electrode unit for PVCmatrix membrane ISE: A,B, orthogonal pieces of Plexiglas; 1-6, C, M, PI,P, as in (a); holes hl-5 are described in detail in the text.

Piece A. The sample stream segmented by air bubbles enters the cell from hole hl (inlet), leaves piece A from hole hz, travels over the loop 6, and then enters again in piece A (but in opposite direction) from hole h3 Finally, it leaves the cell from hole h4 (outlet). All holes (hl-h4) have a diameter of 1.5 mm. Holes hl and h2 coincide at point PI and holes h3 and h4 at point Pzin a distance of 5 mm from P1. In the center of the lower surface of piece A another hole is drilled (h,) 12 mm in diameter and 9 mm deep. Two holes are further drilled from the bottom of hole h5,with a 1.2-mm bit and vertically to the upper surface, in such a way that they encounter holes hz and h4 at points approximately 0.5 mm from points P1 and P2,respectively. In each narrow hole a ceramicjunction plug is inserted so that the one end is positioned approximately in the middle of hole h2 (or h,) and the other lies in h,. Each junction is secured in place with Araldite epoxy adhesive, which is added from points P1or P2and from the bottom of hole &. The adhesive fills the space between the plugs and the Plexiglas and is left for

2 h to harden. The excess of the adhesive is removed by using a 1.5” bit in holes h2and and a 2-mm bit in b.It is essential that the two ends of each ceramic plug are free from an excess of adhesive; otherwise electrolytic contact may not be established. A 4-cm-long glass tube with approximate internal and external diameters of 5 and 7.5 mm, respectively, is inserted in h5and stuck with Araldite adhesive as shown in Figure 2b. This glass tube is connected with a wider “reservoir” tube via PVC tubing (approximately 10 cm long). Piece B. In the center of the upper surface of this piece, hole 4 is drilled, 7 mm in diameter and 8 mm deep. The hole is further drilled with a 2-mm-diameter bit until it reaches the lower surface of the piece. The two pieces can be stacked together by means of two screws that are positioned in holes at the two corners of the pieces. Cell Organization. A 10 X 5 mm piece of PVC membrane is cut and placed on the upper surface of Plexiglas piece A so as to cover the small holes that are created at points P1and P2by

506

ANALYTICAL CHEMISTRY, VOL. 61, NO. 5, MARCH 1, 1989

Flgure 3. Typical recordings of continuous-fiow measurements of dlscrete samples with the iodide bubble-through electrode at flow rates of (A) 1.74, (B) 3.64, and (C) and (D) 6.70 mL min-'. The final concentration of the standards and the sampling rates are shown. Sample to wash ratio was 1:l.

the coincidence of hole hl with h2 and of h3 with hl. Piece B is then placed over piece A and the screws are fastened. Hole hs, the PVC tube, and the reservoir tube are filled with 3 M KCl solution (salt bridge). The level of fluid in the reservoir tube must be at least 5 cm higher than the level of the ceramic junction so that a positive hydrostatic pressure difference exists. Hole 4 of piece B is filled with internal reference solution M propantheline bromide-O.1 M NaCl solution, saturated with AgC1). An Ag-AgC1 wire is dipped in the internal reference solution as shown in Figure 2b. A standard calomel electrode dipped in the salt bridge serves as an external reference electrode.

RESULTS AND DISCUSSION A typical measurement cycle, in the bubble-through units described, consists of four different events: (a) A bubble may be present a t point P1 (Figure 2), thus opening the circuit at this point. However, electrode potential noise is not produced because there is liquid filing loop 6 and electrical contact through Pz, which also contains measuring solution. (b) The bubble moves entirely in loop 6. In this case, both points PI and Pz are free from any bubble effects. The electrode potential is measured through electrical contact a t both points. (c) The bubble moves to point P2 This situation is identical with (a), but electrical contact is facilitated through point PI. (d) The bubble leaves point P2 going to waste. A new bubble has not yet reached PI so that electrical contact is facilitated a t both points without any bubble effects. These events are repeated in an identical fashion throughout the measurements. In this system, noisefree measurements are successful if the length of liquid between two successive bubbles is greater than the length of one bubble, so that a t least one of the points PI, Pz is not in contact with a bubble, throughout. The PIPz distance can be easily adjusted by changing the length of the plastic tube in loop 6 (Figure 2). The proposed bubble-through electrodes for iodide anion and propantheline cation exhibited characteristics similar to those of the corresponding conventional dip-type electrodes. Measurements were carried out with the continuous-flow system shown in Figure 1 without the use of the sampler. Solutions of the measured ions were aspirated continuously into the system via the sampling tube, from each stock solution bottle. The bottles were interchanged manually every few minutes. With this method, the steady-state response of the

Table I. Response Times of the Proposed Bubble-Through Electrodes at Various Flow Rates and Sampling Rates

flow rate, mL min-'

sampling rate, h-'

1.74 3.64 6.70 6.70

60 120 240 360

percentage of steady state propantheline iodide I" IIb In IIb 98 98 95

99 99 96

100

100 99 97

100 99 99

The concentration changes were from 3 X to M (propantheline) and from 10" to 3 X M (iodide). *The concentration changes for both electrodes were from 3 X to M. electrodes is recorded. Na2S04solutions of 0.050 and 0.025 M were used as wash solutions for the iodide and propantheline ISE, respectively. Calibration graphs based on these recordings gave slopes of 59.5 mV (linear range: 1.5 X lo4 to lo-' M; r = 0.9999) and 57.1 mV (linear range: 2.5 X lo* to 8 X M, r = 0.9995) for the iodide and propantheline bubble-through electrodes, respectively. The proposed eledrodes were also tested in continuous-flow measurements on discrete samples where each sample was separated from the next one by wash solution. These measurements were performed a t three flow rates, viz., 1.74,3.64, and 6.70 mL min-', and a t sampling rates of 60,120, and 240 (and 360 for the iodide electrode) determinations per hour, respectively. The wash solutions used contained either constant dilute (Figures 3 and 4) or concentrated (Figure 5) background levels of the measured ions.. With this method, the response of the electrodes from dilute to concentrated solutions and vice versa can be evaluated under continuousflow conditions. The sample to wash ratio was 1:l or 21 when dilute and concentrated background wash solutions were used, respectively. With increasing flow rate, the response time of the electrodes decreases, as has been shown previously (10, 15). With the iodide solid-state ISE as sensor, sampling rates of up to 360 per hour have been obtained with flow rates as high as 9.2 mL mi& with the use of a debubbler (16). Data obtained here with the proposed electrode units for response time studies are given in Table I. The response in all instances is expressed as the percentage of the peak height observed with respect to the peak height of the steady-state

ANALYTICAL CHEMISTRY, VOL. 61, NO. 5, MARCH 1, 1989

t

3 . io-’

aolh

507

120/h

n

Shad .t.ti

Smin

lmin

E

I

c -

scan Figure 4. Typical recordings of continuous-flow measurements of discrete samples with the propantheline bubble-through electrode. Flow rates are as in Figure 3. The final concentrations of the standards and the sampling rates are shown. Sample to wash ratio was 1:l.

60/h

,

240/h

Steady state

-, c 2 min

Smin

c--.--2 .

.-__A

I

----

1 min

-----7r

*

____j

scan

Flgure 5. Typical recordings of continuous-flow measurements wffh the iodide bubble-through ISE using concentrated wash solutions. Flow rates are as in Figure 3. Sample to wash ratio was 2:l.

response. The precision of the measurements can be appreciated by examining the peaks in Figures 3-5. The average coefficient of variation in all instances was better than 2%. In conclusion, we have described new flow-through units for solid-state and PVC-matrix membrane ISE’s. Instead of the segmenting bubbles being debubbled before entering the electrode, they are passed directly through it along with the liquid, without any additional disturbance of the signal measured. The units described are complete electrochemical cells. There are numerou advantages in performing noise-free potentiometric measurements in air-segmented streams

without debubbling: First, all the liquid sample goes through the detector. This is desirable, especially in the case of electrodes with large response times. thus, increasing the sample volume that passes through the electrode may result in improvements in sensitivity. Secondly, the wash of the electrode membrane is improved. Furthermore, the additional sample contamination occurring after removing the bubbles in the CFA systems that make use of a debubbler is avoided. Finally, another advantage is the ability to perform multipoint analysis by using several bubble-through detectors in series, separated by incubation coils.

Anal. Chem. 1989, 6 1 , 508-512

508

LITERATURE CITED (1) Skews, L. T., Jr. Am. J. Clin. Pathol. 1957, 2 8 , 311-322. (2) Snyder, L. R. Anal. Chim. Acta 1980, 174, 3-18. . Anal. Chem. 1987, 17(3), (3) Horvai, G. Pungor, E. CRC C f ~ Rev. 231-264. (4) Ruzicka, J.; Hansen, E. H. Flow Injection Analysis; Wiley: New York, 1981. (5) . . Martin. G. B.: Cho. K. H.: Meverhoff. M. E. Anal. Chem. 1984, 56, 26 12-26 13. (6) Neeley, W. E.; Wardlaw, S. C.; Yates, T.; Holiingsworth. W. G.; Swinnen, M. E. T. Clin. Chem.1978, 22, 227-231. (7) Patton, C. I.; Rabb, M.; Crouch, S. R. Anal. Chem. 1982, 5 4 , 1113-1118. . . .- . . .. . (8) Hadjiioannou, T. P.; Papastathopouios, D. S.; Diamandis, E. P. I n Computerlzation and Automation In Heelthy Facilities; Rubin, M., Ed.; CRC Press: Boca Raton, FL, 1984; pp 169-196.

(9) Rao. K. J. M.; Pelavin. M. H.; Morgenstern, S. Adv. Autom. Anal. Technicon. Int. Cong. 1972-3; 7 , 33-36.

(IO) Christopouios, T. K.; Dlamandis, E. P. Am/yst 1987, 112, 1293-1298. (11) Papastathopouios, D. S.: Karayannis, M. I . J. Chem. Educ. 1980, 57, 904-906. ... .- .

(12) Mitsana-Papazoglou, A.; Christopoulos, T. K.; Diamandis, E. P.; Hadjiioannou. T. P. Analvst 1985. 110. 1091-1094. (13) Mitsana-Papazogbu, A.; Christopoulos, T. K.; Diamandis, E. P.; Koupparis, M. A. J. Pharm. Sci. 1987, 78, 724-730. (14) Graggs, A.; Moody, G. J.; Thomas, J. D. R. J. Chem. Educ. 1974, 51, 541-544. (15) Lindner, E.; Toth, K.; Pungor, E. Anal. Chem. 1976, 48, 1071-88. (16) Alexander. P. W : Seegopaul, P. Anal. Chem. 1980, 52, 2403-2406.

RECEIVED for review July 18, 1988. Accepted November 21, 1988.

Polishable and Robust Modified Graphite Epoxy Electrodes Joseph Wang,* Teresa Golden, Kurian Varughese, and Ibrahim El-Rayes Department of Chemistry, New Mexico State University, Las Cruces, New Mexico 88003 The field of chemically modified electrodes (CMEs) has been the focus of a high level of activity since its inception about 15 years ago ( I , 2). Substantial enhancements of the sensitivity, selectivity, or stability of electroanalytical measurements have been achieved by using electrodes modified with an appropriate preconcentrating agent (3-7), by incorporating redox mediators that accelerate electrochemical reaction rates (8-IO),or by coating the surface with a permselective film (II,I2). Despite the enormous potential of CMEs for chemical analysis, such devices are rarely used for routine quantitative measurements and are not available from commercial sources. This is due to the mode of preparing CMEs, which commonly involves an attachment of the modifier to the bare electrode surface by means of covalent bonding, chemisorption, or polymer coatings. While these approaches yield the desired effects, they suffer from practical difficulties that limit the ease with which the resulting electrodes can be routinely prepared, used, and renewed for subsequent reuse. There is no fundamental reason, however, to expect lesser practicality from CMEs, compared to that of unmodified electrodes. Shaw and Creasy (13-15) described recently a new class of CMEs prepared by copolymerization of styrene and divinylbenzene, in the presence of the modifier and carbon black. Such a scheme results in robust and polishable (i.e., reusable) bulk modified electrodes, as desired for many practical applications. While the concept of bulk modified electrodes is very attractive, the modified carbon composite electrodes of Shaw and Creasy suffer from experimental difficulties (e.g., complexity, length, success rate) associated with their fabrication method. Alternative (and more general) routes for preparing bulk modified electrodes are thus highly desirable. This note describes the fabrication and performance of graphite epoxy modified electrodes, which exhibit desired characteristics for "real-life" applications. The epoxy-bonded graphite is a commercial two-component, graphite-fiied, epoxy resin bonded adhesive filler (Grade RX, Dylon, Cleveland, OH). Unmodified epoxy-bonded graphite electrodes possess desirable electrochemical, mechanical, and chemical propedies and have been used for in vivo voltammetry and microanalysis (16,17). Modification of graphite epoxy is easily accomplished by mixing the desired quantity of the modifier with the two individual (graphite resin and accelerator) components. While both graphite epoxy and carbon composite bulk modification schemes offer similar analytical attributes with respect to

renewability, mechanical stability, and versatility, the graphite epoxy fabrication method is faster, is simpler, and offers a high success rate. EXPERIMENTAL SECTION Apparatus and Reagents. Voltammograms were obtained with either a Bioanalytical Systems Model l00A electrochemical analyzer, EG&G PAR Model 264A voltammetric analyzer, or a Sargent-Welch Model 4001 polarograph. A Bioanalytical Systems Model VC-2 electrochemical cell was used for voltammetric measurements. Two additional 10-mL cells were employed for the preconcentration/medium-exchange/voltammetric procedure for copper (5). The flow injection system, with a thin-layer amperometric detector (Model TL-4, Bioanalytical Systems) and a Rheodyne Model 7010 injection valve (20-pL sample loop), was described previously (12). A Ag/AgCl reference electrode (Model RE-1, Bioanalytical Systems) was used in all experiments. The auxiliary electrode was a platinum wire or stainless steel tube (batch and flow experiments, respectively). All solutions were prepared with double-distilled water. Cobalt phthalocyanine, oxalic acid (Kodak), poly(4-vinylpyridine) (Aldrich),hydrazine sulfate, potassium ferrocyanide, copper nitrate (Baker), penicillamine, and L-cysteine (Sigma) were used without further purification. Electrode Preparation. The electrode material was a commercial epoxy-bonded graphite (Grade RX, Dylon, Cleveland, OH). The unmodified electrode was prepared by thoroughly mixing equal amounts of the epoxy-bonded graphite resin and the accelerator for 5 min. A portion of the electrode material was packed into the end of a 4-mm-i.d. glass tube. Electrical contact was established via a copper wire. The electrode was then cured at 55 "C for 2 h and then at 100 "C for an additional 2 h. The surface was polished with a silicon carbide (No. 320) paper and then with 0.05-pm alumina slurry on a felt pad. Prior to use, the electrode was sonicated for 5 min in double-distilled water. For the cobalt phthalocyanine (CoPC) modified electrode, CoPC was added to the 1:l resin/accelerator mixture (5% (w/w)); a thorough mixing proceeded for 5 min. The cation exchanger modified electrode was prepared by adding the desired amount (usually 11%(w/w)) of the swelled resin (Dowex CGC241 sulfonated polystyrene) to the 1:l graphite/accelerator mixture and mixing for 20 min. Both modified electrodes were then cured and poiished as described above for the unmodified electrode. RESULTS AND DISCUSSION Preconcentrating Modified Graphite Epoxy Electrodes. Voltammetry following nonelectrolytic accumulation represents one of the most useful analytical avenues of CMEs (18). Shaw and Creasy (14) described recently the incorpo-

0003-2700/89/0361-0508$01.50/0 0 1989 American Chemical Society