Trace Metals Debate - C&EN Global Enterprise (ACS Publications)

Aug 19, 2013 - Trace Metals Debate. ANN M. THAYER. C&EN Houston. Chem. Eng. News , 2013, 91 (33), pp 10–13. DOI: 10.1021/cen-09133-cover...
0 downloads 0 Views 203KB Size
R EADIN G SCIEN T IF IC

COVER STORY

READY, TEST, GO

A technician runs mass spectrometry analysis in Reading Scientific’s metal-testing lab.

TRACE METALS DEBATE Contentious new guidelines on PHARMACEUTICAL IMPURITIES will force the drug industry to change testing strategies ANN M. THAYER, C&EN HOUSTON

IN 1905, the U.S. Pharmacopeial Con-

vention, the nonprofit standards-setting organization known as USP, introduced a method to check for heavy metals in U.S. pharmaceuticals. More than 100 years later, drug manufacturers still use the same chemistry to determine whether the level of 10 metal impurities is acceptable in their products. Exposure to metals such as arsenic and mercury can have negative health effects; the substances can also weaken drug stability. But rather than targeting individual metals for their toxicity and setting limits accordingly, the old USP test, described in chapter 231 of the group’s standards, measures all of the metals that can react with a sulfide ion to form a visible precipitate. The aggregate threshold dictates the acceptable limit for all the metals combined.

Regulators and drug manufacturers agree on the need to modernize elemental impurity analysis. However, they disagree on the substance of a new limit-based standard and the timing of its implementation. Tensions have risen in the past three years as deadlines have come and gone. Coming to an agreement on the standard is important because almost every drug sold in the U.S. will have to comply. After nearly two decades of discussions, postings, appeals, and revisions, two USP chapters replacing 231 became official in form on Feb. 1. Chapter 232 sets limits on 15 discrete elemental impurities in drug substances, excipients, and products. Chapter 233 describes procedures for analyticalinstrument-based measuring of impurity levels. USP had intended to implement the new chapters on May 1, 2014, at which time CEN.ACS.ORG

10

AUGUST 1 9, 2013

products must comply, but rising dissension this spring—largely around the timeline for compliance—caused the organization to postpone the move. The deferral gives time to rehash the content and timing for implementation of the chapters, but it doesn’t alter the sense that the changes are inevitable. Despite the uncertainties and unanswered questions, drugs manufacturers don’t want to risk being unprepared. Many companies are putting the needed analytical systems in place. “I haven’t spoken to anyone who thinks this won’t happen, so they show no inclination of slowing down,” says Chuck Schneider, spectroscopy product planning leader at analytical instrument maker PerkinElmer. “But they are all approaching it differently.” Instrumentation suppliers predict that

drug firms initially will play it safe and test for all listed contaminants and later formulate risk-based compliance plans around actual suspected impurities. “They have to be on the side of caution, because they can’t afford to fail a Food & Drug Administration audit,” says Adrian Holley, director for elemental analysis at Thermo Fisher Scientific. Although USP sets standards, FDA is the agency that enforces compliance.

is that everything passes when you never know what actually is present,” Zaidi says. Although USP once considered improving the old method, it concluded that better and more modern approaches to test for elemental impurities are available, Zaidi says. Chapter 233 describes using inductively coupled plasma optical emission spectroscopy (ICP–OES) and ICP-mass spectrometry (ICP-MS), as well as criteria for alternative analytical methods. There’s no reason to think the drug industry has an impurity problem, observers note. FDA testing in 2012 did not identify “unsafe” products, according to Rx-360, a pharmaceutical supply-chain monitoring group. And FDA itself has reported that preliminary studies from 2007 revealed “nothing alarming.”

many haven’t deployed these techniques in manufacturing and quality control. They must also adjust to using recommended sample preps on a range of dosage forms, or matrices, and may have to hire and train technical staff. The process from ordering an instrument through delivery, set up, method development, and validation can take months, Schneider cautions. Depending on the number of products to be analyzed, companies may need to HEIGHTENING CONCERN is that chapters install multiple systems at multiple sites. 232 and 233 would apply to more than 1,200 “Large companies are talking about 40 to monographs, explains Kahkashan Zaidi, 50 installations globally,” Thermo Fisher’s principal scientific liaison at USP. A monoHolley says. “China is already gearing up, graph describes a finished drug or drug inand in India it’s predicted that there will be gredient in terms of its identity, packaging, 50 to 60 ICP-MS and more than 100 ICPlabeling, and the testing and acceptance OES systems.” An ICP-MS system costs criteria required to ensure its strength, about $150,000, and one for ICP-OES runs quality, and purity. about $80,000 to $100,000, ac“We took extra measures to cording to instrument makers. DAILY DOSE develop these new standards Instrumentation providers Toxicologically proposed limits on because of the high impact,” have to be prepared as well. Beelemental impurities differ among standards groups Zaidi says. “USP has been very cause available technology is up flexible and has been listening to the task, no instrument modiORAL PERMITTED to the industry and giving all the fication was required, Schneider a b c d DAILY EXPOSURE (µg/DAY) USP ICH EMA opportunities that are possible says, “but we had to get method Arsenic (inorganic) 1.5 15 na and within our reach to get feeddevelopment, operating proceLead 5 5 na Mercury (inorganic) 15 40 na back and take action wherever dures, standards, consumables, Cadmium 25 5 na we can.” informatics, and validation serPalladium 100 100 100 Nonetheless, industry bevices in place.” Platinum 100 1,000 100 lieves USP hasn’t adequately Schneider has seen the need e Iridium 100 1,000 100 considered the ramifications of for testing on this scale only a Osmium 100 1,000 100e its plans. A two-year-old drug few times, such as when conRuthenium 100 1,000 100e industry group, the Coalition for tamination crises drove the toy Rhodium 100 1,000 100e Molybdenum 100 180 250 Rational Implementation of the and food industries to rapidly Vanadium 100 120 250 USP Elemental Impurities Reramp up. “The good news for the Nickel 500 600 250 quirements, warned in a late-2012 pharma industry is that they are Copper 1,000 1,300 2,500 position paper that “currently apnot responding to a scare,” he Chromium nc 11,000 250 proved, and safe, drug products says. “The industry can be proa Based on a 50-kg (110 lb) person. b As of Feb. 1, 2013. c As of July 26, 2013. List also may be pushed out of compliance active and have good testing in includes antimony, barium, cobalt, gold, lithium, selenium, silver, thallium, and tin. or forced to reformulate, and that place in a timely manner.” d EMA 2008 guideline covers metal residues from catalysts and reagents and also includes iron, manganese, and zinc. e Total limit for subclass of iridium, ruthenium, drug shortages may result.” Timing was an issue the indusrhodium, and osmium. USP = U.S. Pharmacopeial Convention. ICH = International More rigorous analytical tests try coalition raised in its requests Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use. EMA = European Medicines Agency. na = not could potentially reveal that to postpone the May 2014 date. applicable in this guideline. nc = not a safety concern. SOURCES: USP, ICH, EMA drugs can’t meet the new stan“Drug and component manudards. There’s no way to know facturers do not have sufficient on the basis of the old test, Zaidi says. SemiBut good expectations aren’t enough, time to complete method development and quantitative at best, the 231 method is not and drug manufacturers will need anavalidation for the many diverse matrices afspecific for individual metals and indicates lytical systems to provide the proof. With fected, nor for implementation and clinical the overall presence of mixed-metal impuriinstallation under way, many large pharma testing of any consequent reformulations,” ties by comparison to a reference standard. companies will be ready when the time the group wrote in a March letter to USP. The old method’s inability to idencomes, instrument suppliers say. For othIt is a “misconception that several years tify which metals might exceed the limit ers, including many excipient manufacturhave been allowed for this process,” the isn’t its only flaw. The aggressive sample ers, “this will be a burden, and there will coalition argued. Instead, because the USP preparation sometimes required can alter be a steep learning curve,” PerkinElmer’s standards didn’t gel until February 2013, the or destroy what’s present, so the method Schneider says. “Hundreds of small comcoalition calculated that only 15 months redoesn’t actually recover the full amounts of panies have been relying on USP 231.” mained to complete work by May 2014. the different metals in a sample. Relying on Even though big drug firms may be familIt’s unclear how much more time the defalse or misleading results, “the conclusion iar with using ICP-OES or ICP-MS in R&D, ferral will provide to companies that need CEN.ACS.ORG

11

AUGUST 1 9, 2013

COVER STORY

it. For a company to be ready by May 2014, it would have to make a decision to move ahead by this November, Thermo Fisher’s Holley points out. “Not everyone has got it in their heads that they need time to set things up.” ALTHOUGH SOME FIRMS are ahead of

the game, others don’t want to invest until they have to, Holley adds. An alternative is to turn to an outside provider. Contract testing labs have been buying a lot of the necessary equipment to offer services for elemental impurities, suppliers say. In June, Reading Scientific Services Ltd. opened a metals lab in England to prepare for USP standards as well as European ones coming into effect in September. Through a partnership with Agilent Technologies, it has ICP-MS and ICP-OES systems, as well as microwave plasma-atomic emission spectroscopy and atomic absorption spectrometry (AAS) systems. After considering suspected impurities, and types and number of samples, Reading Scientific can advise customers on the choice of method, says Alan Cross, a metals specialist at the lab. Working with a contract lab removes the burden of training staff, qualifying an instrument, and maintaining it, he adds. And because sample prep is critical and time-consuming, a lab that does the work routinely “gets a good idea how to treat

different matrices and get around any problems,” Cross says. “Instrument manufacturers are trying to make things more intuitive, and the software is more powerful, but for us it becomes second nature knowing what works or can go wrong and how to fix things.” The best available methods can recover and identify all the target metals within a couple of minutes in a single run. Despite a high price tag, most customers are opting for ICP-MS systems, instrumentation managers say. The technique can easily achieve the required detection limits across the widest range of concentrations and elements. It also requires much smaller samples than the 2 g called for by the 231 test. “If you are a pharmaceutical company testing an active ingredient, you don’t want to waste your sample, and you want to make sure that you can achieve the lowest detection limits,” says Amir Liba, an applications chemist at Agilent. Another advantage of ICP-MS is the ability to identify different elemental species. Using toxicological assessments, USP has set daily exposure and concentration limits that depend on whether a drug is administered orally, injected, or inhaled. Of the 15 elements on its list, the most toxic four—arsenic, cadmium, lead, and mercury—are environmental contaminants and have the lowest limits. “It is only mandatory to test for the big four,” Liba says.

BRIEF HISTORY

Plans To Change Testing Procedures For Heavy Metals Are A Decade In The Making 2004 U.S. Pharmacopeial Convention (USP) committee begins to look at updating metals testing including the possibility of using advanced sample preparation and analysis. 2005 USP identifies inductively coupled plasma-optical emission spectroscopy (ICPOES) and ICP-mass spectrometry as analysis methods. 2010 After considering

revisions of the existing USP chapter 231 chemistry test and new testing methods, USP drafts new limits and testing procedures. May 2011 Final drafts of new USP chapter 232 on elemental impurity limits and 233 on test procedures are published. They are set to become official on Dec. 1, 2012. November 2012 Industry groups petition USP to

alter plans; USP postpones Dec. 1, 2012, release date. February 2013 USP declares chapters 232 and 233 official and plans to put them into effect by May 1, 2014. May 2013 Industry response leads USP to consider deferring May 1, 2014, implementation. June 2013 USP forms an advisory group to consider the implementation plan.

CEN.ACS.ORG

12

AUGUST 1 9, 2013

The remaining ones might arise from catalysts, reagents, machinery, or equipment. “You test for the others if you think there is possible contamination,” Liba explains, “but the final product should be below the limit on all 15 analytes, regardless of whether you test or not.” USP will accept results from any type of instrument as long as the user validates it for analyzing the elemental impurities. Like ICP-MS, ICP-OES is a rapid multielement technique, generally sensitive enough for almost all the listed elements. Other techniques—such as graphite furnace AAS, which costs one-third as much as an ICP-MS system to buy and operate— work well but may analyze fewer elements and only one or a few at a time. Sample preparation, such as extensive dilution, can increase the need for high sensitivity, Liba says. Although simple dilution or dissolution will work for some materials, others will require digestion. Because drug products are complex mixtures of ingredients, fillers, and coatings, suppliers expect that closed-vessel microwave digestion, which can cost up to $50,000 for a full system, is the best way to ensure complete sample breakdown. Drug manufacturers must comply with the standards, but it is up to them to decide how. “Regulators have left a great degree of freedom in making it risk-based for manufacturers and not prescribed,” says Daniel Kutscher, an ICP-MS application specialist at Thermo Fisher. “But they’ll have to justify why they do their testing their way.” Instrumentation company managers anticipate that, for peace of mind, most drug manufacturers will keep a tight grip on the situation by testing products themselves. “I expect there will be overtesting at first, whenever they start up a new batch,” Holley says. Eventually, manufacturers will determine the frequency of testing required after understanding the risk that specific impurities will appear. Moreover, drug manufacturers are likely to ask excipient, raw material, and active ingredient suppliers to provide data on metals. However, not much is known about metal impurities in excipients, and those sourced from minerals and plants have suppliers the most concerned. It’s not known whether suppliers will be able to generate the data for pharma customers, says the International Pharmaceutical Excipients Council, a member of the coalition. But to facilitate information exchange, the group has created a standard-

ized request letter and form templates. “A manufacturer has to understand its product very, very well before determining what they need to test for, because there is no need to test for everything that is in chapter 232,” USP’s Zaidi says, provided the firm has the knowledge base to support a decision. “We advocate a risk-based strategy in which you evaluate your product, your supplier, your processes, and determine what you need to test and how often you need to test.” IN LIGHT OF ALL THE WORK required to comply with the new standards, the drug industry wants to have to do it only once. Right now, USP standards are not aligned with those under development since 2009 by the International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use, a group known as ICH that aims to align international regulatory requirements. Harmonization between ICH’s Q3D guideline and USP’s chapter 232 will “avoid significant problems throughout the industry and redundant non-value-added work,” the industry coalition wrote in its March letter. ICH completed step 2b of its four-step process to create harmonized guidelines in June, and USP is reviewing the latest ICH version, which was published on Aug. 5. USP revised chapter 232 after previous interim ICH steps, Zaidi points out. ICH is scheduled to reach its final step in June 2014, after which regional implementation would occur in Europe, Japan, and the U.S. In 2008, the European Medicines Agency (EMA) published its own standards that go into effect next month. The industry coalition says it sees “no scientific or safety-based justification” for USP to unilaterally implement the new chapters while ICH is still developing its guidelines. “Because the time is insufficient for properly preparing for compliance, a more rational approach would be to wait for ICH Q3D to be finalized so that 232 can be appropriately harmonized with this international standard,” the coalition says. USP says new chapters are needed now because its become understood that currently used methods cannot differentiate between safe and unsafe products. “We have every intention to go along with the ICH whenever possible,” Zaidi says. “But one has to keep in mind that step two is still not the final document from ICH. One has to wait and see where it finally falls and then make the decision how much the alignment can happen.”

ICH guidelines address only the limits on elemental impurities, and not testing methods. USP, EMA, and ICH have different elements on their lists. Even where the lists do overlap, they have set widely different permitted daily exposure limits for different dosage forms and drug materials. USP says the deferral announced in May will allow it to work closely with ICH

to align activities. In early June, USP created a nine-member advisory group that includes people from FDA and from organizations within the industry coalition. USP says the group is expected to work quickly to provide advice on how to “advance a timely implementation” of chapters 232 and 233 while addressing special impacts on manufacturers. ◾

LUMOS FT-IR microscopy made easy

 Stand-alone

FT-IR microscope with full automation

 Simple

easy-to-use interface

 Motorized

ATR crystal

 Automated

measurements in transmission, reflection and ATR

 Comprehensive

validation with 21 CFR part 11

Bruker‘s new LUMOS is a fully automated FT-IR microscope with an integrated FT-IR spectrometer. LUMOS is designed to combine highest quality in visible inspection and highest performance in infrared spectral analysis. With motorization and networking of all moveable components, the LUMOS provides an ease of use never before seen in FT-IR microscopy. Contact us for more details: www.bruker.com/lumos s www.lumos-ir.com

Innovation with Integrity

CEN.ACS.ORG

13

AUGUST 1 9, 2013

F T-IR